(28 days)
The ACUSON P500 ultrasound system is intended for diagnostic ultrasound imaging and fluid flow analysis for the following applications: Fetal, Abdominal, Small Parts, OB/GYN, Pelvic, Cardiovascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The ACUSON P500 is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Tissue Doppler Image, Power (Amplitude) Doppler Mode and a combination of modes on a Flat Panel Display.
This document is a 510(k) Premarket Notification from Siemens Medical Solutions USA, Inc. for their ACUSON P500 Ultrasound System, submitted to the FDA on January 12, 2015. The purpose of the submission is to demonstrate substantial equivalence to legally marketed predicate devices, the ACUSON X300 ultrasound system (K121699) and the ACUSON S2000 ultrasound system (K140959).
Based on the provided document, here's an analysis of the acceptance criteria and study information:
1. A table of acceptance criteria and the reported device performance:
The document does not explicitly present a table of "acceptance criteria" with quantitative performance metrics for the ACUSON P500 ultrasound system itself. Instead, it demonstrates substantial equivalence to predicate devices based on:
- Identical Indications for Use: The ACUSON P500 shares the same intended uses as the predicate devices across various applications (Fetal, Abdominal, Small Organ, Cardiac, Transrectal, Transvaginal, Peripheral vessel, Musculo-skeletal (conventional), Musculo-skeletal (superficial), Emergency Medicine).
- Similar Technological Characteristics: The document provides a detailed comparison table (Page 5) showing that the ACUSON P500 largely shares:
- Supported Center Frequencies (2.0 MHz to 11.0 MHz across the board for all three devices).
- Imaging Modes (B, M, PWD, SCW, Color Doppler, Power Doppler, Combined modes, THI, M-THI are all present in the P500 and the predicates).
- Many Features (e.g., Dual-Beam Processing, Quad-Beam Processing, Clip Capture, DIMAQ, Multiple Frequency Imaging, Digital Architecture, DICOM, Dynamic TCE, Wireless functionality).
- Compliance with Safety Standards: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety, conforming to applicable medical device safety standards like UL 60601-1, IEC 60601-2-37, CSA C22.2 No. 601-1, AIUM/NEMA UD-3, AIUM/NEMA UD-2, 93/42/EEC Medical Devices Directive, EN/IEC 60601 series, EN/IEC 62304, EN/IEC 62366, and ISO 10993-1.
Reported Device Performance (by inference of substantial equivalence):
Based on the assertion of substantial equivalence, the implied "performance" of the ACUSON P500 is considered to be equivalent to that of the predicate devices. The document does not provide specific numerical performance metrics (e.g., sensitivity, specificity, resolution) for the ACUSON P500 or its predicates. It focuses on functional equivalence.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
The document explicitly states under "F. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence": "Since the P500 uses the same technology and principles as existing devices, clinical data is not required."
Therefore, no clinical test set was required or used for this 510(k) submission. The entire evaluation for substantial equivalence was based on non-clinical tests (engineering, safety, and performance characteristic comparisons to predicates).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
Not applicable. As no clinical data/test set was required, no experts were needed to establish ground truth for this submission.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. No clinical test set was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The document is for an ultrasound system, not an AI-assisted diagnostic device. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a diagnostic ultrasound system, not a standalone AI algorithm. Its performance is inherent to its hardware and software for image acquisition and display, which are deemed substantially equivalent to existing devices.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
Not applicable. As no clinical data/test set was required, no ground truth was established for this submission. The "ground truth" for the FDA's substantial equivalence determination here relies on the established safety and effectiveness of the predicate devices and the P500's compliance with recognized safety standards through non-clinical testing.
8. The sample size for the training set:
Not applicable. The document describes a medical device seeking 510(k) clearance based on substantial equivalence to existing predicate devices, not an AI/machine learning device that would typically involve a training set.
9. How the ground truth for the training set was established:
Not applicable. No training set was used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a design that resembles an abstract bird or wing shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2015
Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K150050
Trade/Device Name: Acuson P500 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 5, 2015 Received: January 12, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150050
Device Name Acuson P500 Ultrasound System
Indications for Use (Describe)
The ACUSON P500 ultrasound system is intended for diagnostic ultrasound imaging and fluid flow analysis for the following applications: Fetal, Abdominal, Small Parts, OB/GYN, Pelvic, Cardiovascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary Prepared December 15, 2014
| Sponsor: | Siemens Medical Solutions, Inc.,Ultrasound Division685 East Middlefield RoadMountain View, California 94043 | ||
|---|---|---|---|
| Manufacturing Facility | SIEMENS LTD SEOUL2nd -30 floor, 143, Sunhwan-ro,Jungwon-gu, Seongnam-si, Gyeonggi-do,Republic of Korea | ||
| Contact Person: | Shelly PearceTelephone:(650) 279-0134(650) 694-5580Fax: | ||
| Device Name: | ACUSON P500 Ultrasound System | ||
| Common Name: | Diagnostic Ultrasound System with Accessories | ||
| Classification:a market of the comments of the comments of |
Regulatory Class: = Review Category: Tier II Classification Panel: Radiology
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|---|---|---|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
A. Legally Marketed Predicate Devices
The ACUSON P500 Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to our current product, the ACUSON X300 ultrasound system (K121699) and ACUSON S2000 ultrasound system (K140959).
B. Device Description:
The ACUSON P500 is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Tissue Doppler Image, Power (Amplitude) Doppler Mode and a combination of modes on a Flat Panel Display.
{4}------------------------------------------------
C. Intended Use
The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal. Small Parts. OB/GYN. Pelvic. Cardiovascular. Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small orqan, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
D. Substantial Equivalence
The ACUSON P500 is substantially equivalent to the ACUSON X300, cleared via K121699, and ACUSON S2000. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.
| Feature / Characteristic | Predicate DeviceACUSON X300(K121699) | Predicate DeviceACUSON S2000(K140959) | Submission DeviceACUSON P500 | |
|---|---|---|---|---|
| Indications for Use: | ||||
| ■ | Fetal | √ | √ | √ |
| ■ | Abdominal | √ | √ | √ |
| ■ | Small Organ | √ | √ | √ |
| ■ | Cardiac | √ | √ | √ |
| ■ | Transrectal | √ | √ | √ |
| ■ | Transvaginal | √ | √ | √ |
| ■ | Peripheral vessel | √ | √ | √ |
| ■ | Musculo-skeletal (conventional) | √ | √ | √ |
| ■ | Musculo-skeletal (superficial) | √ | √ | √ |
| ■ | Emergency Medicine | √ | √ | √ |
| Center Frequencies Supported: | ||||
| ■ | 2.0 MHz | √ | √ | √ |
| ■ | 2.5 MHz | √ | √ | √ |
| ■ | 3.0 MHz | √ | √ | √ |
| ■ | 3.5 MHz | √ | √ | √ |
| ■ | 4.0 MHz | √ | √ | √ |
| ■ | 5.0 MHz | √ | √ | √ |
| ■ | 5.5 MHz | √ | √ | √ |
| ■ | 6.0 MHz | √ | √ | √ |
| ■ | 6.5 MHz | √ | √ | √ |
| ■ | 7.5 MHz | √ | √ | √ |
| ■ | 8.0 MHz | √ | √ | √ |
| ■ | 9.0 MHz | √ | √ | √ |
| ■ | 10.0 MHz | √ | √ | √ |
| Feature / Characteristic | Predicate DeviceACUSON X300(K121699) | Predicate DeviceACUSON S2000(K140959) | Submission DeviceACUSON P500 | |
| 11.0 MHz | √ | √ | √ | |
| Modes: | ||||
| B | √ | √ | √ | |
| M | √ | √ | √ | |
| PWD (Pulsed Wave Doppler) | √ | √ | √ | |
| SCW (Steerable CW) | √ | √ | √ | |
| D (Color Doppler) | √ | √ | √ | |
| Power Doppler(Amplitude Doppler) | √ | √ | √ | |
| Combined (BM, BC, BCM, BCD) | √ | √ | √ | |
| THI (Tissue Harmonic Imaging) | √ | √ | √ | |
| M-THI | √ | √ | √ | |
| Features: | ||||
| 3D Imaging (3-Scape) | √ | √ | -- | |
| 3D Measurements | √ | √ | -- | |
| 4D Basic Imaging (fourSight 4D) | √ | √ | -- | |
| Panoramic 2D Imaging (SieScape) | √ | √ | -- | |
| Cardiac Imagingphysiological signal displaymultiplane TEEfourSight TEE Imaging | √ | √ | -- | |
| Dual-Beam Processing | √ | √ | √ | |
| Quad-Beam Processing | √ | √ | √ | |
| SynAps | √ | √ | -- | |
| Clip Capture | √ | √ | √ | |
| Tissue Grayscale Optimization (TGO) | √ | √ | -- | |
| Spectral DTI | √ | √ | -- | |
| Axius OB Automated Calipers | √ | √ | -- | |
| Steerable CW | √ | √ | √ | |
| Stress Echo | √ | √ | -- | |
| 2D(color) DTI | √ | √ | -- | |
| Image Arena reporting S/W | √ | √ | -- | |
| DIMAQ (PIMS Workplace) | √ | √ | √ | |
| Vascular Enhancement (Clarify VE) | √ | √ | -- | |
| eSielmage | -- | √ | √ | |
| Advance SieClear | -- | √ | √ | |
| Multiple Frequency Imaging(MultiHertz) | √ | √ | √ | |
| Digital Architecture | √ | √ | √ | |
| Fully integrated DICOM | √ | √ | √ | |
| Dynamic TCE | √ | √ | √ | |
| AHP | √ | √ | -- | |
| Monitor: 15.4"WXGA (FPD) | √ (17" FPD) | √ (19" FPD) | √ | |
| Wireless | √ | √ | √ | |
| Feature / Characteristic | Predicate DeviceACUSON X300(K121699) | Predicate DeviceACUSON S2000(K140959) | Submission DeviceACUSON P500 | |
| # Channels | 128 | 192 | 64 | |
| Output Display Standard(Track 3) | √ | √ | √ | |
| Patient Contact Materials | Tested to ISO 10993-1 | Tested to ISO 10993-1 | Tested to ISO 10993-1 | |
| UL60601-1 Certified | √ | √ | √ |
The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics.
{5}------------------------------------------------
{6}------------------------------------------------
E. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
- UL 60601-1, Safety Requirements for Medical Equipment 国
- IEC 60601-2-37 Diagnostic Ultrasound Safety Standards 보
- CSA C22.2 No. 601-1. Safety Requirements for Medical Equipment l
- AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical 전 Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 피 AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound
- 93/42/EEC Medical Devices Directive 별
- 트 Safety and EMC Requirements for Medical Equipment
- EN/IEC 60601-1 L
- I EN/IEC 60601-1-1
- 트 EN/IEC 60601-1-2
- 트 EN/IEC 60601-1-6
- . EN/IEC 62304
- EN/IEC 62366 家
- I EN/IEC 60601-2-18
- 트 EN/IEC 60601-2-25
- ISO 10993-1 Biocompatibility 피
Cleared patient contact materials, electrical and mechanical safety are unchanged.
F. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
Since the P500 uses the same technology and principles as existing devices, clinical data is not required.
G. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms with 21 CFR 820 Quality System Requlation and ISO 13485:2003 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Siemens Medical that the P500 is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{7}------------------------------------------------
The ACUSON P500 is verified and validated according to the company's design control process.
N/A