Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K973743
    Date Cleared
    1997-10-31

    (30 days)

    Product Code
    Regulation Number
    862.1260
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ACCESS ESTRADIOL REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Estradiol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of estradiol levels in human serum, using the Access Immunoassay System.

    Device Description

    The Access® Estradiol reagents and the Access® Immunoassay Analyzer comprise the Access® Immunoassay System for the quantitative determination of Estradiol levels in human serum.

    AI/ML Overview

    This document describes the validation of the ACCESS® Estradiol assay, an immunoassay for determining estradiol levels in human serum.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device are implied by the "Summary of Studies" section, where specific performance metrics are reported. There are no explicit "acceptance criteria" stated as thresholds; rather, the study presents the observed performance and concludes that the device is substantially equivalent to a predicate device.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Precision (Within Run)CV within acceptable clinical ranges for estradiol measurementRanges from 19.7% CV (at 39 pg/ml) to 3.1% CV (at 2167 pg/ml)
    Precision (Total)CV within acceptable clinical ranges for estradiol measurementRanges from 20.0% CV (at 39 pg/ml) to 5.0% CV (at 2918 pg/ml)
    Accuracy (Spiking Recovery)Recovery between 80% and 120%Ranging from 80% to 115%
    Accuracy (Dilution Recovery)Mean recovery close to 100%Mean recoveries of 108% to 112%. Average dilution recovery at 1:2 was 106%. For samples >500 pg/ml, average % recovery was 107%.
    Correlation with Predicate DeviceStrong positive correlation (e.g., r > 0.95)r = 0.98
    Analytical Sensitivity (Lowest Detectable Level)Detectable level at or below clinically relevant thresholds20 pg/ml
    Analytical Specificity (Cross-Reactivity with Estrone Sulfate)Negligible or no detectable cross-reactivityNot detectable when spiked at 3600 pg/ml
    Analytical Specificity (Cross-Reactivity with Estriol Sulfate)Low percentage apparent cross-reactivityApparent cross-reactivity of 0.004% when spiked at 10,000,000 pg/ml

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Precision studies: The number of samples for precision studies is not explicitly stated, but typically involves multiple replicates of several control levels or patient samples. The data provenance is not specified (e.g., country of origin, retrospective/prospective).
      • Accuracy (Spiking recovery): Ten serum samples were used. Data provenance not specified.
      • Accuracy (Dilution recovery): 2 patient samples and 2 patient pools were used, along with 10 patient samples with estradiol >500 pg/ml. Data provenance not specified.
      • Correlation: 103 samples were run. Data provenance not specified.
      • Analytical Sensitivity: Implies samples were used to determine the lowest detectable level. Number of samples not specified.
      • Analytical Specificity: S0 calibrator was spiked with estriol sulfate and estrone sulfate. Number of samples/spiked conditions not explicitly detailed beyond the two substances.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This is an in vitro diagnostic (IVD) device (immunoassay). The "accuracy" and "sensitivity" of such devices are typically established against reference methods or through spiked samples with known concentrations, rather than through expert consensus on images or clinical conditions. Therefore, the concept of "experts establishing ground truth for the test set" in the context of interpretation of results (like in image analysis) does not directly apply here. The "ground truth" for quantitative measurements like estradiol is the actual concentration determined by highly accurate methods (e.g., mass spectrometry, or certified reference materials) or by the known concentration of spiked analytes. The document does not specify details about these reference methods or expert panel for ground truth.

    3. Adjudication method for the test set:
      Not applicable (N/A) for this type of quantitative immunoassay study. Adjudication methods like 2+1 or 3+1 are typically used for qualitative or interpretive assessments, such as in image reading where multiple experts might disagree on findings. Here, results are quantitative values compared to known or reference values.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable (N/A). This is an immunoassay, not an AI-assisted diagnostic tool involving human readers interpreting cases.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, this is implicitly a standalone study. The performance characteristics (precision, accuracy, sensitivity, specificity, correlation) are reported for the ACCESS® Estradiol assay system itself (reagents + analyzer) as a device, independent of human interpretation or intervention in the measurement process beyond operating the instrument and following procedures.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      The ground truth used here consists of:

      • Known concentrations: For spiking recovery studies (estradiol spiked into serum samples) and analytical specificity studies (estrone sulfate, estriol sulfate spiked into calibrator).
      • Reference measurements: The Abbott IMx® Estradiol Kit served as a comparator for the correlation study, implying it was considered a widely accepted and validated method for measuring estradiol. For precision and dilution recovery, the ground truth is the inherent concentration of estradiol in the samples, measured multiple times or diluted to known proportions.
    7. The sample size for the training set:
      Not applicable (N/A) as this is a traditional immunoassay validation report, not one describing a machine learning or AI-driven device that requires a distinct "training set." The development of the assay itself involves optimization and characterization over many samples, but these are not formally called "training sets" in the context of this type of document.

    8. How the ground truth for the training set was established:
      Not applicable (N/A). See point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970126
    Date Cleared
    1997-02-25

    (42 days)

    Product Code
    Regulation Number
    862.1260
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    ACCESS ESTRADIOL REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER (33550,33540,33545)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACCESS® Estradiol assay is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of Estradiol levels in human serum using the ACCESS® Immunoassay System.

    Device Description

    The ACCESS® Estradiol reagents and the ACCESS® Immunoassay Analyzer comprise the ACCESS® Immunoassay System for the quantitative determination of Estradiol levels in human serum.

    AI/ML Overview

    The provided document is a 510(k) summary for the ACCESS® Estradiol assay, which is an in-vitro diagnostic device (IVDD). The questions appear to be tailored for AI/Machine Learning (ML) medical devices, which typically require different types of studies and acceptance criteria than IVDDs.

    Given the nature of the device (an immunoassay for quantifying Estradiol levels) and the document (510(k) submission from 1997), many of the requested fields are not applicable or the information is not provided in the same format as for AI/ML device studies.

    However, I will extract and describe the available information as best as possible, indicating when a question is not applicable to this type of device or when the information is not present.


    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVDD product like the ACCESS® Estradiol assay, "acceptance criteria" are typically defined by performance specifications and comparisons to a predicate device, rather than explicit thresholds of diagnostic accuracy (e.g., AUC, sensitivity, specificity) against a ground truth in a clinical setting. The "performance" refers to the analytical performance of the assay itself.

    Acceptance Criteria (Implied from Study Goals)Reported Device Performance (ACCESS® Estradiol assay)
    Precision: Demonstrate acceptable reproducibility of results.Within-run CV: 11.6% (at 117 pg/ml) to 2.7% (at 3041 pg/ml)
    Total Imprecision CV: 21.1% (at 72 pg/ml) to 5.2% (at 1456 pg/ml)
    Accuracy (Spiking Recovery): Recover spiked analyte within an acceptable range.Recovery: 92% to 106% (mean recovery of 105% ± 7% SD for dilution recovery)
    Accuracy (Dilution Recovery): Recover diluted analyte within an acceptable range.Mean recovery: 105% with a standard deviation of 7%
    Correlation with Predicate Device: Strong linear correlation with the legally marketed predicate device.Correlation (r): 0.958
    Linear Regression: y = -39 + 0.983x (compared to Abbott IMx® Estradiol Kit)
    Analytical Sensitivity (LOD): Lowest detectable level distinguishable from zero.Lowest detectable level: 23 pg/ml (with 95% confidence)
    Analytical Specificity: Minimal cross-reactivity with structurally similar compounds.Estrone sulfate: Not detectable at 3600 pg/ml
    Estriol sulfate: Apparent cross-reactivity of 0.04% at 10,000,000 pg/ml

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Correlation Study: 214 samples were used for comparison between the ACCESS® Estradiol assay and the Abbott IMx® Estradiol Kit.
      • Spiking Recovery: 10 serum samples.
      • Dilution Recovery: 18 patient samples.
      • Analytical Sensitivity: Implied use of Calibrator S0 (zero calibrator) multiple times to determine the limit of detection, but a specific number of samples/replicates is not given.
      • Analytical Specificity: Not specified, but involved spiking into S0 calibrator.
      • Precision Studies: Number of samples not explicitly stated, but typically involves multiple replicates of control materials or pooled patient samples run over several days/runs.
    • Data Provenance: The document does not explicitly state the country of origin for the samples. Given the applicant's address in Chaska, MN, USA, it is highly probable the samples were from the USA.
    • Retrospective or Prospective: Unspecified, but for analytical performance studies of IVDDs, these are typically laboratory-based studies using samples collected specifically for method validation or archived samples, rather than prospective clinical trials in the sense of AI/ML devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable in the context of diagnostic accuracy for an AI/ML device. For an IVDD, the "ground truth" for analytical studies is typically established by:
      • The known concentration of an analyte in a spiked sample.
      • The calculated expected concentration in a diluted sample.
      • The results from a well-established, legally marketed predicate device (as in the correlation study).
      • Known pure reference materials for specificity.
      • There are no "experts" in the sense of human readers/interpreters establishing a ground truth for imaging or clinical diagnosis here.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are common in AI/ML studies where multiple human readers interpret data to reach a consensus ground truth. For an IVDD's analytical performance studies, this concept does not apply.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study is not relevant for an immunoassay kit. This device directly measures a biomarker concentration and does not involve human readers interpreting complex data that could be augmented by AI.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in spirit. The performance described (precision, accuracy, correlation, sensitivity, specificity) is the standalone performance of the assay system (reagents + analyzer). There is no "human-in-the-loop" component in the direct measurement process of this IVDD; the analyzer outputs a quantitative result. The human then interprets that numerical result, which is distinct from human readers interpreting images or clinical data where AI might assist.

    7. The Type of Ground Truth Used

    • Reference Method Comparisons / Known Concentrations:
      • Correlation Study: The results from the Abbott IMx® Estradiol Kit, a legally marketed predicate device, served as the comparative reference (analogous to a ground truth for comparison).
      • Spiking Recovery: Known amounts of estradiol spiked into serum.
      • Dilution Recovery: Expected concentrations based on dilution of patient samples.
      • Analytical Sensitivity/Specificity: Defined concentrations of the analyte or interferent.

    8. The Sample Size for the Training Set

    • Not Applicable / Not Provided for this type of device. Immunoassays are not "trained" like AI/ML algorithms. The assay's parameters (e.g., antibody concentrations, incubation times, calibrator values) are developed and optimized during product development. The concept of a "training set" for an AI algorithm does not apply here. Calibrators are used to establish a standard curve for each run, which is different from a "training set" for an AI model.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As explained in point 8, there is no "training set" in the AI/ML sense. For the calibrators included with the assay, their "ground truth" (i.e., their assigned estradiol concentration) would have been established through a rigorous value assignment process, often involving reference methods or gravimetric preparation of highly pure standards. This is part of the manufacturing process for IVDDs.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1