Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202932
    Date Cleared
    2021-05-28

    (241 days)

    Product Code
    Regulation Number
    886.5928
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ABT12 multi-purpose solution

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bausch + Lomb ABT12 multi-purpose solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.

    Device Description

    Bausch + Lomb ABT12 multi-purpose solution is a sterile isotonic aqueous solution [containing polyaminopropyl biguanide (0.0005%), polyquaternium (0.00015%), and alexidine (0.00025%)] for disinfecting, cleaning, conditioning, rinsing, protein removal, and storing soft (hydrophilic) and silicone hydrogel contact lenses. The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.

    AI/ML Overview

    Based on the provided text, the device in question is a multi-purpose contact lens solution (ABT12 multi-purpose solution), not an AI-powered medical device for image analysis. Therefore, many of the typical acceptance criteria and study details related to AI performance (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of product and approval process.

    The document describes the regulatory approval of a contact lens solution (Bausch + Lomb ABT12 multi-purpose solution) for which the primary claims are related to its safety and effectiveness in cleaning, disinfecting, and storing contact lenses. The "acceptance criteria" here are based on demonstrating substantial equivalence to a predicate device (COMPLETE multi-purpose solution Easy Rub Formula) through a combination of preclinical (biocompatibility, microbiology, lens compatibility, cleaning efficacy) and clinical studies.

    Here's a breakdown of the requested information, adapted to the context of this non-AI medical device:


    Acceptance Criteria and Study for Bausch + Lomb ABT12 Multi-Purpose Solution

    As the device is a contact lens solution, not an AI-powered diagnostic tool, the typical "acceptance criteria" table for AI performance metrics (e.g., sensitivity, specificity, AUC) is not applicable. Instead, the acceptance criteria are based on demonstrating substantial equivalence to a predicate device in terms of safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for a contact lens solution primarily revolve around meeting established safety and effectiveness standards, often demonstrated through non-inferiority to a legally marketed predicate device.

    Criteria CategorySpecific Acceptance Criteria (Demonstrated via Non-Inferiority to Predicate)Reported Device Performance (ABT12 Multi-Purpose Solution)
    Clinical SafetyNon-inferiority in ophthalmological findings (e.g., slit lamp findings) compared to predicate. Absence of serious adverse events (SAEs) or device-related adverse events (AEs). No significant changes to corneal health.The ABT12 group demonstrated non-inferiority to the COMPLETE group Over All Follow-up Visits for slit lamp findings greater than Grade 2. No serious adverse events, adverse device effects, or significant non-serious adverse events were reported in the ABT12 multi-purpose solution group. No eyes were discontinued due to an AE. None of the AEs were considered related to the study solution. No corneal infiltrates were reported, and no changes to pre-existing corneal scars. Visual acuities were similar between groups. Conclusion: Safe for use.
    Clinical EffectivenessNon-inferiority in user-reported comfort and dryness. Non-inferiority in lens deposit prevention/removal.For the three primary effectiveness endpoints (overall comfort averaged over all follow-up visits; dryness averaged over all follow-up visits; and optimal [{none or light} lens deposits at all follow-up visits), the treatment differences indicated that the ABT12 solution is non-inferior to the COMPLETE solution. No notable differences were found regarding symptoms/complaints, worn lens characteristics, and dispensed lens characteristics. Conclusion: Effective for use.
    BiocompatibilityNo significant cytotoxicity, ocular irritation, oral toxicity, or sensitization. Compatibility with various lens materials.Studies demonstrated the biocompatibility of ABT12 multi-purpose solution, including tests on the solution itself, with maximum preservative concentrations, and with extracts from different lens materials. Leveraged existing clearance for packaging components. Predicate K014202 was used for side-by-side comparison in relevant studies.
    Microbiological EfficacyMeets or exceeds criteria for disinfection and preservative efficacy against relevant microorganisms (e.g., as per EN ISO 14729:2001 and EN ISO 14730:2014).The solution exceeds the criteria for disinfection and preservative efficacy.
    Lens CompatibilityCompatible with soft contact lenses, including silicone hydrogel contact lenses.Studies demonstrated compatibility with soft contact lenses, including silicone hydrogel contact lenses.
    Cleaning EfficacyDemonstrates effective cleaning properties (e.g., protein removal).In-vitro laboratory studies demonstrated cleaning properties with artificially protein-deposited contact lenses. Surfactant concentrations are well above the Critical Micelle Concentration (CMC).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study - Test Set): A total of 252 subjects were enrolled. 240 subjects completed the study. These subjects were randomized into two groups:
      • ABT12 multi-purpose solution (Test): 127 subjects
      • COMPLETE multi-purpose solution (Control/Predicate): 125 subjects
    • Data Provenance: The clinical study was a multicenter study conducted in the USA.
    • Retrospective or Prospective: This was a prospective, randomized, masked, parallel, bilateral clinical study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable in the context of this device. The "ground truth" for a contact lens solution's performance is established through direct clinical observations (e.g., slit lamp findings by ophthalmologists/optometrists) and patient-reported outcomes, as well as laboratory testing results, rather than expert consensus on image interpretation. The study was masked, implying that the clinicians assessing outcomes were blinded to the treatment assignment.

    4. Adjudication Method for the Test Set

    This concept (adjudication of interpretations) is not applicable to this type of study. Clinical outcomes were measured directly by investigators in a masked fashion and patient-reported outcomes were collected.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was not done. MRMC studies are typically used to evaluate the impact of a diagnostic tool (like AI) on physician performance in interpreting medical images. This study focused on the safety and effectiveness of a physical product (contact lens solution) in a clinical trial setting.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. There is no algorithm for this device. The "standalone performance" of the contact lens solution is demonstrated through the preclinical studies (biocompatibility, microbiology, cleaning efficacy, lens compatibility) which evaluate the product's intrinsic properties irrespective of human handling errors (though proper usage instructions are provided).

    7. The Type of Ground Truth Used

    The ground truth was established through a combination of:

    • Clinical Measurements: Objective clinical assessments by masked investigators (e.g., slit lamp findings, visual acuities).
    • Patient-Reported Outcomes: Subjective data gathered directly from subjects regarding comfort, dryness, and other symptoms.
    • Laboratory Testing: In-vitro and ex-vivo studies to assess intrinsic properties like microbiological efficacy, cleaning efficacy, and biocompatibility.
    • Comparison to Predicate: The "truth" of effectiveness and safety was established by demonstrating non-inferiority to a known, legally marketed predicate device.

    8. The Sample Size for the Training Set

    This question is not applicable as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as this is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1