Search Results
Found 3 results
510(k) Data Aggregation
(29 days)
21HQ513D, 32HL512D, 31HN713D, 32HQ713D
21HQ513D, 32HL512D: This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
31HN713D, 32HO713D: This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.
The provided document describes the acceptance criteria and the results of the study for the medical monitors 21HQ513D, 32HL512D, 31HN713D, and 32HQ713D.
Here's the requested information:
1. A table of acceptance criteria and the reported device performance
The document refers to the "performance items suggested in the FDA guidance 'Display Devices for Diagnostic Radiology'" as the acceptance criteria. The performance for each measurement is uniformly reported as "Pass" for the tested items.
Measurements | Acceptance Criteria (Implied by FDA Guidance) | Reported Device Performance |
---|---|---|
a. Spatial resolution | Meet FDA guidance standards | Pass |
b. Pixel defects | Meet FDA guidance standards | Pass |
c. Artifacts | Meet FDA guidance standards | Pass |
d. Temporal response | Meet FDA guidance standards | Pass |
e. Luminance | Meet FDA guidance standards | Pass |
f. Conformance to a grayscale-to-luminance function | Meet FDA guidance standards | Pass |
g. Luminance at 30° and 45° in diagonal, horizontal, and vertical directions at center and four corners | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
h. Luminance uniformity or Mura test | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
i. Stability of luminance and chromaticity response with temperature and time of operation (on-time) | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
j. Spatial noise | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
k. Reflection coefficient | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
l. Veiling glare or small-spot contrast | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
m. Color tracking | Meet FDA guidance standards | Pass |
n. Gray tracking | Meet FDA guidance standards | Pass (for 31HN713D, 32HQ713D) / N/A (for 21HQ513D, 32HL512D) |
Note: The "N/A" for certain measurements for models 21HQ513D and 32HL512D might indicate that these tests were not applicable or not performed for these specific models, possibly due to their differing indications for use (not intended for mammography, unlike 31HN713D and 32HQ713D). The document does not explicitly state the acceptance numerical values for each criterion but implies compliance with the FDA guidance "Display Devices for Diagnostic Radiology".
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in terms of patient data or images. The study described is a non-clinical bench test on the display devices themselves. The sample size for the test set would be the number of devices tested, which is implied to be one of each model (21HQ513D, 32HL512D, 31HN713D, and 32HQ713D). The data provenance is not applicable in the typical sense of clinical data, as it's a technical performance test of hardware.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable. The study is a non-clinical bench test of display performance against predefined technical standards and FDA guidance, not a study involving expert interpretation of medical images.
4. Adjudication method for the test set
This information is not applicable. There was no expert adjudication process as it was a technical performance test.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. The device is a medical monitor, not an AI-powered diagnostic tool. The document explicitly states: "No clinical studies were considered necessary and performed."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is not applicable as the device is a medical monitor, not an algorithm or AI system. The study focused on the technical performance of the monitors.
7. The type of ground truth used
For the bench test, the ground truth was based on the performance items suggested in the FDA guidance "Display Devices for Diagnostic Radiology." This refers to quantifiable technical specifications and standards for medical image displays.
8. The sample size for the training set
This is not applicable. The device is a medical monitor, not a machine learning algorithm that requires a training set. The software components underwent validation according to IEC 62304.
9. How the ground truth for the training set was established
This is not applicable, as there was no training set for an AI algorithm. The validation of the software was done according to IEC 62304, which involves verifying the software against its design specifications and requirements.
Ask a specific question about this device
(30 days)
32HL512D, 31HN713D, 32HQ713D
32HL512D: This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
31HN713D, 32HQ713D: This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.
The provided text describes a medical monitor device (32HL512D, 31HN713D, 32HQ713D) and its associated software, "LG Calibration Studio Medical," which is used for calibration. The acceptance criteria and the study performed relate to the performance of the medical monitor and the calibration software, not an AI algorithm for image analysis. Therefore, there is no information about AI-specific aspects such as training sets, ground truth establishment for AI, or comparative effectiveness studies with human readers assisted by AI.
Here's the information derived from the text regarding the device and its software:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the medical monitors are derived from the FDA guidance "Display Devices for Diagnostic Radiology." The reported performance for all three models (32HL512D, 31HN713D, 32HQ713D) is that they "Pass" all tested measurements.
Measurements | Acceptance Criteria | Reported Device Performance (32HL512D, 31HN713D, 32HQ713D) |
---|---|---|
a. Conformance to a grayscale-to-luminance function | Conformance to a grayscale-to-luminance function (implied by FDA guidance requirements) | Pass |
b. Luminance uniformity or Mura test | Luminance uniformity or Mura test (implied by FDA guidance requirements) | Pass |
c. Stability of luminance and chromaticity response with temperature and time of operation (on-time) | Stability of luminance and chromaticity response with temperature and time of operation (on-time) (implied by FDA guidance requirements) | Pass |
d. Spatial noise | Spatial noise (implied by FDA guidance requirements) | Pass |
e. Veiling glare or small-spot contrast | Veiling glare or small-spot contrast (implied by FDA guidance requirements) | Pass |
Note: The document states, "All display characteristics of the 32HL512D, 31HN713D and 32HQ713D have met the predefined criteria. Therefore, the performance of 32HL512D, 31HN713D and 32HQ713D were verified through the performance test." The FDA guidance "Display Devices for Diagnostic Radiology" serves as the basis for these predefined criteria.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated as the testing is on the physical device's characteristics rather than a dataset of medical images. The tests were performed on one or more units of each model (32HL512D, 31HN713D, 32HQ713D).
- Data Provenance: Not applicable in the context of device performance testing. The "data" here refers to measurements taken directly from the physical monitors during bench testing. The tests adhere to international standards (e.g., IEC standards) and FDA guidance for medical display devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. The ground truth for the performance of the medical monitors is established by objective physical measurements against established technical standards and FDA guidance, not by expert interpretation of images.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for expert consensus in image interpretation. This study involves objective technical measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. There is no mention of an MRMC study. The devices are medical monitors for displaying images, not AI algorithms.
6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done
A standalone performance test was done for the device, focusing on its technical display characteristics. The "LG Calibration Studio Medical" software (a moderate level of concern software) was also verified and validated, but this is a calibration tool, not an image analysis algorithm. Therefore, "standalone (i.e. algorithm only without human-in-the-loop performance)" is applicable to the technical performance of the monitor as a device, and to the verification and validation of the calibration software, but not in the context of AI for medical image interpretation.
7. The Type of Ground Truth Used
The ground truth used for the device's performance testing is based on objective physical measurements and technical specifications set forth by international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 62304) and the FDA guidance "Display Devices for Diagnostic Radiology."
8. The Sample Size for the Training Set
Not applicable, as this is not an AI algorithm for image analysis. The "LG Calibration Studio Medical" is software, validated according to IEC 62304, but the text does not describe a "training set" in the machine learning sense for this calibration software.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for an AI algorithm mentioned in relation to the devices or their calibration software.
Ask a specific question about this device
(149 days)
32HL512D
This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images
Here's a breakdown of the acceptance criteria and study information based on the provided document, addressing each of your points.
It's important to note that this document is a 510(k) summary for a medical monitor (32HL512D) intended for displaying radiological images. It is not an AI/algorithm-based diagnostic device. Therefore, many of your questions related to AI performance, ground truth, expert consensus, MRMC studies, and training sets will not be applicable to this submission. The "device" in this context is a monitor, not a diagnostic algorithm.
Acceptance Criteria and Device Performance for LG 32HL512D Medical Monitor
The acceptance criteria for this device (a medical monitor) are primarily related to its technical specifications for displaying medical images and compliance with electrical safety and EMC standards. The "performance" is demonstrated by meeting these specifications and showing compliance.
1. Table of Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria (from predicate or general medical display requirements) | Reported Device Performance (LG 32HL512D) |
---|---|---|
Indications for Use | Display radiological images for review, analysis, and diagnosis by trained medical practitioners. Not intended for mammography. | This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. (Matches exactly) |
Electrical Safety | Compliance with IEC 60601-1:2005/AMD1:2012 | Complies with IEC 60601-1:2005/AMD1:2012 |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 Edition 4.0:2014 | Complies with IEC 60601-1-2 Edition 4.0:2014 |
Usability | Compliance with IEC 60601-1-6:2010/A1:2013 | Complies with IEC 60601-1-6:2010/A1:2013 |
Software Validation (if applicable) | Designed and developed according to a software development process; verified and validated (for MODERATE level of concern software) | Software was designed and developed according to a software development process and was verified and validated. (MODERATE level of concern) |
Resolution | (Implied comparison to predicate/PACS needs) | 3,840 x 2,160 pixels |
Screen Size | (Implied comparison to predicate/PACS needs) | 718.2 x 414.3 mm (or 31.5" diagonal through module description) |
Display Technology | TFT LCD with specific display capabilities for medical images. | TFT Liquid Crystal Display module with LED Backlight unit, 30-pin eDP port, 4 lane eDP interface, supports 3840 x 2160 UHD mode, 10-bit drivers. |
Pixel Pitch | (Implied comparison to predicate/PACS needs) | 0.18159 x 0.18159 mm |
Input Video Signals | Standard video inputs for medical imaging systems. | DisplayPort x 2, HDMI x 1 |
Note: The document focuses on demonstrating substantial equivalence to a predicate device rather than setting explicit quantifiable acceptance thresholds for all performance parameters in the context of a new, unique claim. Its performance is accepted if it is shown to be as safe and effective as the predicate, despite technological differences. The "reported device performance" are its specifications and compliance with standards.
Regarding the Study Proving Device Meets Acceptance Criteria:
The document states: "No clinical studies were considered necessary and performed." This is crucial. This is a submission for a medical display monitor, not a diagnostic algorithm or device that makes a clinical determination. Therefore, the "study" proving it meets acceptance criteria is primarily non-clinical testing and a comparison to a predicate device.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not applicable in the context of patient data or algorithm performance. The "test set" for a monitor would be the monitor itself undergoing various electrical, EMC, and display characteristic tests.
- Data Provenance: Not applicable. The testing is on the physical monitor, not a dataset.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
- Not applicable. There is no "ground truth" to establish in the sense of medical diagnoses for this device. The device itself is the display. The "experts" involved would be engineers and metrologists conducting the technical measurements and ensuring compliance with display standards and safety regulations.
4. Adjudication Method for the Test Set:
- Not applicable. This concept pertains to resolving discrepancies in expert interpretations of medical data, which is not relevant for a medical monitor.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, not performed. This type of study is relevant for evaluating the impact of AI algorithms or new diagnostic methods on human reader performance. As this is a medical monitor, an MRMC study is not part of its substantial equivalence demonstration.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, not applicable. This device is a monitor, not a standalone diagnostic algorithm.
7. The Type of Ground Truth Used:
- Not applicable. There is no 'ground truth' in the clinical diagnostic sense for a medical display monitor. The "ground truth" for its performance is its measured technical specifications and compliance with relevant industry and safety standards (e.g., DICOM Part 14 for display characteristics, IEC standards for safety).
8. The Sample Size for the Training Set:
- Not applicable. This device does not use a training set as it is not an AI/machine learning algorithm.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. No training set exists for this device.
Ask a specific question about this device
Page 1 of 1