Search Results
Found 6 results
510(k) Data Aggregation
(88 days)
0.9% Sodium Chloride Injection, USP BD PosiFlush SF Saline Flush Syringe
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices.
0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Saline Flush Syringe is a polypropylene syringe intended to flush indwelling vascular access devices. It is a sterile, single use syringe prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The device is packaged in a film-on-film blister pack and sterilized by gamma irradiation. External sterility of the device enables it to be used in the sterile field. The subject device is available only in 10mL syringe configuration.
This document pertains to the 510(k) premarket notification for the "0.9% Sodium Chloride Injection, USP BD PosiFlush™ SF Saline Flush Syringe," a medical device, not an AI/ML-driven device or an imaging system. Therefore, the questions regarding AI/ML performance metrics, expert consensus for ground truth, MRMC studies, or training/test set details are not applicable to this submission.
This submission focuses on demonstrating substantial equivalence to a predicate device (K153481) through bench performance testing, packaging integrity, and biocompatibility tests. The "acceptance criteria" here refer to the specifications for these non-clinical tests, rather than diagnostic performance metrics of an AI model.
Here's the information extracted that aligns with the request, re-interpreting "acceptance criteria" and "study" in the context of this device's submission:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Purpose | Acceptance Criteria | Result |
---|---|---|---|
Performance/Design Verification Tests | |||
Tip Cap Removal Torque | Measure the torque required to remove the tip cap from the syringe | Tip Cap can be twisted off as per BD validated force | Pass |
Reflux | Measure the potential blood reflux into the catheter from PosiFlush™ SF syringe | Reflux greater than BD validated internal value | Pass |
Tip Cap Leakage | Evaluate the resistance to leakage between the barrel Luer and the tip cap for pre-filled syringes | No evidence of Tip Cap leakage | Pass |
Leakage past Stopper | Evaluate the resistance to leakage between the barrel and stopper ribs | No leakage of solution past the stopper ribs | Pass |
Breakloose force | Measure the initial maximum force required to move the plunger rod/stopper in the syringe barrel | Force to move plunger rod/stopper is less than the BD validated force | Pass |
Breakout Force | Measure the maximum force required to move the plunger rod/stopper in the syringe barrel a short time after initially moving the plunger rod/stopper | Force to move plunger rod/stopper is less than the BD validated force | Pass |
Sustaining Force | Measure the average force required to move the plunger rod/stopper in the syringe barrel | Force to move plunger rod/stopper is less than the BD validated force | Pass |
Retaining Ring Force | Measure the force necessary to remove the plunger rod/stopper assembly from the syringe barrel after assembly | Force to move plunger rod/stopper is greater than the BD validated force | Pass |
Expelled Volume | Measure the volume of saline solution expelled from the syringe | USP43-NF38 Container Content for Injections | Pass |
NaCl Assay | Measure the NaCl concentration saline solution in syringe samples | 0.9% NaCl Injection USP Monograph | Pass |
pH | Measure the pH of saline solution in syringe samples | 0.9% NaCl Injection USP Monograph and USP43-NF38 pH | Pass |
Iron | Measure the amount of iron in saline solution | 0.9% NaCl Injection USP Monograph and USP43-NF38 Iron | Pass |
Heavy Metals | Measure the heavy metals such as Cadmium, Arsenic, Cobalt, Vanadium, Copper, Lead, Nickel, Lithium, Antimony and Mercury in saline solution | USP43-NF38 Elemental Impurities - Limits | Pass |
NaCl solution weight loss | Measure the weight loss of saline solution in pre-filled saline syringes | USP43-NF38 Containers Performance Testing | Pass |
UV Analysis | Measure the UV absorbance between 220-360 nm for pre-filled saline syringes | Bacterial Endotoxins Test | Pass |
Sterility | To verify 10-6 SAL in the fluid path. | Sterile; No growth | Pass |
Surface Sterility | To verify 10-6 SAL on all surface of syringe and inside surface of primary packaging | Sterile; No growth | Pass |
Particulate Matter | Measure the number of particulates in saline pre-filled syringes | USP43-NF38 Particular Matter in Injections | Pass |
Appearance and Solution Clarity/Barrel Transparency | Examine the appearance, solution clarity and barrel transparency in pre-filled saline syringes | Solution and components are clear | Pass |
Primary Package Integrity Tests | Evaluates the sterile barrier system of the pre-filled syringe throughout shelf life | No leaks when tested by Vacuum Bubble Leak (ASTM F2096-11) | Pass |
Non-porous | Pass | ||
Seal strength and width value as per BD validated force and width measurements respectively | Pass | ||
No defects in packaging seals when visually inspected | Pass | ||
Clean peel of blister pack with no foreign matter | Pass | ||
Biocompatibility Tests | |||
Cytotoxicity | ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Grade ≤ 2 | Pass |
Sensitization | ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization | Non-Sensitizer | Pass |
Irritation | ISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation | Final Test Sample Score ≤ 1 | Pass |
Acute Systemic Toxicity | ISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity | No significantly greater biological reaction than the control | Pass |
Material Mediated Pyrogenicity | ISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity / USP43-NF38 Pyrogen Test (USP Rabbit Test) | No temperature rise ≥ 0.5° C | Pass |
Hemocompatibility | ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood / ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials | ≤ 5% hemolysis | Pass |
Genotoxicity | ISO 10993-3: 2014 Biological evaluation of medical devices – Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity | Non-mutagenic | Pass |
LAL Endotoxin | USP 43-NF38 Medical Devices – Bacterial Endotoxin / USP43-NF38 Bacterial Endotoxins Test / ANSI AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing | Below the Endotoxin Limit 0.5 EU/device | Pass |
Extractable and Leachable Analysis | ISO 10993-18: 2020 Biological evaluation of medical devices - Part 18: Chemical characterization of materials / ISO 10993-17: 2002 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances | Under the condition of the chemical characterization testing, there were no extractables and leachables identified that indicated significant risk concerns. | Note 1 |
Note 1: Result is a qualitative statement, not a numerical pass/fail against a specific threshold listed in the table.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify numerical sample sizes for each of the performance, packaging, or biocompatibility tests. It only states that "appropriate performance tests and biocompatibility tests were performed" and the gamma sterilized test samples were evaluated. The data provenance (country of origin, retrospective/prospective) is also not specified, as these are bench and lab-based tests, not clinical studies involving patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This submission is for a physical medical device, not an AI/ML imaging device. Ground truth, in this context, refers to the established standards, monograph requirements (e.g., USP, ISO), and validated internal specifications to which the device performance is measured against. Experts here would be the laboratory personnel and engineers performing the tests and
interpreting results against these defined criteria, but their specific number or detailed qualifications are not provided in the summary.
4. Adjudication Method for the Test Set:
Not applicable. Since this is not an AI/ML diagnostic device, there is no "adjudication" of expert opinions on image interpretation. The tests performed are laboratory or bench-based, with results measured against predetermined quantitative or qualitative acceptance criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI/ML device, and no MRMC study was performed. The device's safety and effectiveness are demonstrated through non-clinical performance, packaging, and biocompatibility testing.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a physical product (saline syringe), not an algorithm or software. No standalone algorithm performance was assessed.
7. The Type of Ground Truth Used:
The "ground truth" for this device's evaluation is primarily based on:
- Established Standards: e.g., ISO 10993 series for biocompatibility, ASTM F2096-11 for vacuum bubble leak, ANSI AAMI ST72:2019 for Bacterial Endotoxins.
- Pharmacopoeia Monographs: e.g., USP43-NF38 for NaCl Injection, pH, Iron, Heavy Metals, Particulate Matter, Bacterial Endotoxin.
- Validated Internal Specifications: "BD validated force" or "BD validated internal value" for mechanical/physical properties like tip cap removal torque, reflux, breakout force, sustaining force, retaining ring force, and seal strength.
- Qualitative Observations: e.g., "No evidence of Tip Cap leakage," "No leakage of solution past the stopper ribs," "Solution and components are clear," "No defects in packaging seals when visually inspected."
Essentially, the ground truth is defined by recognized industry standards, pharmacopoeial requirements, and the manufacturer's own validated specifications for product performance and safety.
8. The Sample Size for the Training Set:
Not applicable. This is not an AI/ML device; there is no "training set."
9. How the Ground Truth for the Training Set was Established:
Not applicable. As there is no training set for an AI/ML model, this question does not apply.
Ask a specific question about this device
(211 days)
0.9% Sodium Chloride Injection, USP, BD PosiFlush SafeScrub
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SafeScrub prefilled flush syringe with an integrated disinfection unit is intended to be used as a disinfection cleaner for needleless access devices attached to indwelling vascular access devices (VADs) and flushing of these VADs.
BD PosiFlush™ SafeScrub is a sterile, single use pre-filled saline syringe with integrated Disinfection Unit (DU). The polypropylene syringe contains 0.9% sodium chloride (USP) solution with a tip cap that is modified at the distal end to accommodate DU. The DU has high density polyethylene housing with 70% Isopropyl Alcohol (IPA) solution in low density polyethylene foam. The pre-filled syringe with modified syringe tip cap is sterilized by moist heat and the DU is sterilized by gamma irradiation. The subject device is available only in 10mL syringe configuration.
The provided text describes the acceptance criteria and the results of various tests conducted for the BD PosiFlush™ SafeScrub device (K213955). This device is a pre-filled saline syringe with an integrated disinfection unit.
Here's the breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Performance/Design Verification Tests | ||
Container Closure Integrity | No dye within the syringe | Pass |
Leakage Test | No leakage from the syringe | Pass |
Torque Removal Test | Tip Cap can be twisted off as per BD validated force | Pass |
Sterile Fluid Path | SAL: 10-6 | Pass |
Axial Pull Force | DU cannot be pulled off per BD validated force | Pass |
Torque | DU cannot be twisted off as per BD validated force | Pass |
Particulate Ingress | USP | Pass |
Antimicrobial Efficacy | ≥ 4-log reduction | Pass |
70% IPA Concentration | 70±7% | Pass |
Foam Rotation | Foam should not rotate >90 degrees within the DU housing during use | Pass |
Foam Retention (before and during use) | Foam must be retained within the DU | Pass |
Foam Retention (after use) | Foam must be retained within DU after scrubbing | Pass |
Foam Durability | No ripped or ragged material and debris or particulate | Pass |
Foam Compressibility and Wetness | Foam must be wet and compressible | Pass |
70% IPA Ingress | Maximum dose of 2 mg IPA/kg body mass/day per US EPA | Pass |
Package integrity of DU | ||
Bubble Leak | as per ASTM F2096 | Pass |
Seal Width | ≥ 0.58 mm | Pass |
Delamination | No delamination | Pass |
Peel Force | USL: ≤ 12.9 N, LSL: ≥ 3.69 N | Pass |
Microbial properties | as per ISO 11607-1:2019 | Pass |
Biocompatibility | ||
Cytotoxicity | Grade ≤ 2 (ISO 10993-5:2009) | Pass |
Sensitization | Non-Sensitizer (ISO 10993-10:2010) | Pass |
Irritation or Intracutaneous Activity | Final Test Sample Score ≤ 1 (ISO 10993-10:2010) | Pass |
Acute Systemic Toxicity | No significantly greater biological reaction than the control (ISO 10993-11:2017) | Pass |
Material Mediated Pyrogenicity | No temperature rise ≥ 0.5° C (ISO 10993-11:2017) | Pass |
Hemocompatibility | ≤ 5% hemolysis (ISO 10993-4:2017, ASTM F756-17) | Pass |
LAL Endotoxin | Below the Endotoxin Limit 20 EU/device (USP 43-NF38 ) | Pass |
Extractable and Leachable Analysis | N/A (Result is a Toxicological Risk Assessment) | Toxicological Risk Assessment |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes for each test in the provided table. It mentions "BD validated force" for several tests, implying internal validation studies by the manufacturer. The document details that "The subject device is also evaluated throughout its shelf life by bench performance testing to ensure that the device meets the predetermined acceptance criteria."
The data provenance is not specified regarding country of origin or whether it's retrospective or prospective, but it implies a prospective testing approach as part of the device's development and regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable. The tests performed are primarily bench performance, engineering, chemical, and biological evaluations, not requiring human expert interpretation in the way, for example, a medical imaging AI product would. The "ground truth" for these tests is based on established scientific and regulatory standards (e.g., ISO, ASTM, USP guidelines) rather than expert consensus on diagnostic interpretations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As described in point 3, these are objective physical and chemical tests, not requiring human interpretation or adjudication panels. The "Pass" results are based on meeting predetermined quantitative or qualitative criteria from recognized standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical product (a pre-filled syringe with a disinfection unit), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and the concept of "human readers improve with AI" are not relevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. As mentioned, this is not an AI algorithm. The performance of the device itself (standalone) is what was tested through the various bench performance tests, antimicrobial efficacy, and biocompatibility studies.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the various tests relies on:
- Established scientific and engineering principles: e.g., for container closure integrity, leakage, torque, pull force.
- Recognized industry standards and guidelines: ISO 10993 series for biocompatibility, ASTM F2096 for package integrity, USP for particulate ingress, USP 43-NF38 for LAL Endotoxin.
- Predetermined chemical concentrations: 70±7% for IPA concentration.
- Microbiological reduction targets: ≥ 4-log reduction for antimicrobial efficacy, based on established efficacy standards for disinfectants.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
(241 days)
0.9% Sodium Chloride Injection, USP BD PosiFlush SP Syringe
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices. Catalog Number 306547 10 mL BD PosiFlush™ SP Syringes are generally compatible for use with syringe pumps.
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP (also referred to as BD PosiFlush™ SP Syringe) is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The BD PosiFlush™ SP Syringe is provided with a sterile fluid path, which is sterilized via moist heat.
The change described in this submission is the qualification of a supplier specific formulation for stopper material for 0.9 % Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes.
This document describes a Special 510(k) submission for a medical device called "0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe" (K161552). The submission aims to demonstrate substantial equivalence to a previously cleared predicate device (K141311) due to a change in the supplier specific formulation for the stopper material.
I will focus on the provided information to answer your request, specifically looking for acceptance criteria and the study proving it.
Here's an analysis based on the document:
1. Table of Acceptance Criteria and Reported Device Performance:
The document broadly states that "Design Verification testing was also conducted to ensure the device met the predetermined acceptance criteria for the following tests." It lists the tests but does not provide a specific table enumerating the acceptance criteria values themselves or the quantitative performance measurements obtained for each test. Instead, it states that the device "met predetermined acceptance criteria per BD's internal specification" and "have been verified to meet the established performance criteria above."
Acceptance Criteria Category | Reported Device Performance |
---|---|
Biocompatibility | All tests (Cytotoxicity, Hemolysis, Acute systemic toxicity, Sub-Chronic Toxicity, Intracutaneous reactivity, Occular irritation, Partial thromboplastin Time, Sensitization, Pyrogenicity, Genotoxicity, Chemical extractable analysis) met the requirements of the FDA Guidance Use of International Standard ISO 10993-1. |
Design Verification | Met predetermined acceptance criteria per BD's internal specification for: |
- Container Closure Integrity
- Break Out/Sustaining force
- Stopper Separation
- Leakage
- Pump Force
- Ship Test
- Dead Space
- Syringe Induced Reflux |
2. Sample Sizes Used for the Test Set and Data Provenance:
The document does not specify the sample sizes used for any of the tests (biocompatibility or design verification). It also does not mention the country of origin of the data or whether the data was retrospective or prospective. The studies were conducted by Becton, Dickinson and Company.
3. Number of Experts and Qualifications for Ground Truth:
This section is not applicable as the clearance is for a pre-filled syringe (a medical device, not an AI or diagnostic tool that requires human expert interpretation for ground truth). The "ground truth" for the performance tests would be based on established engineering and biological standards and verified by internal testing protocols.
4. Adjudication Method:
This is not applicable for the same reasons as point 3.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
This is not applicable as the device is a pre-filled syringe, not an AI-assisted diagnostic tool.
6. Standalone (Algorithm Only) Performance:
This is not applicable as the device is a pre-filled syringe and does not involve an algorithm.
7. Type of Ground Truth Used:
For the performance tests, the "ground truth" would be established by:
- Established industry standards and regulations (e.g., ISO 10993 for biocompatibility).
- BD's internal specifications and engineering requirements for device functionality and integrity (e.g., for leakage, break force, dead space).
8. Sample Size for the Training Set:
This is not applicable as this is not an AI/machine learning device that requires a training set.
9. How the Ground Truth for the Training Set was Established:
This is not applicable for the same reasons as point 8.
Study Proving Device Meets Acceptance Criteria:
The study that proves the device meets the acceptance criteria is detailed across two main categories:
-
Biocompatibility Testing:
- Methodology: Evaluated in accordance with FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," June 16, 2016.
- Tests Performed: Cytotoxicity (per ISO10993-5), Hemolysis (per ISO10993-4), Acute systemic toxicity (per ISO10993-11), Sub-Chronic Toxicity (Per ISO10993-11), Intracutaneous reactivity (per ISO10993-10), Ocular irritation (per ISO 10993-10), Partial thromboplastin Time (ISO 10993-4), Sensitization (per ISO10993-10), Pyrogenicity (per ISO 10993-11), Genotoxicity (per ISO 10993-3), and Chemical extractable analysis (per ISO 10993-18).
- Conclusion: The tests confirmed that the device met the requirements outlined in the standards, indicating it is biocompatible.
-
Design Verification Testing:
- Methodology: Conducted to ensure the device met "predetermined acceptance criteria per BD's internal specification." The specific methodologies for each test (e.g., how "Container Closure Integrity" was tested) are not detailed but would be standard industry practices for syringe manufacturing.
- Tests Performed: Container Closure Integrity, Break Out/Sustaining force, Stopper Separation, Leakage, Pump Force, Ship Test, Dead Space, Syringe Induced Reflux.
- Conclusion: The testing demonstrated that the subject device "met predetermined acceptance criteria."
Overall Conclusion from the Document:
The submission concludes that the BD PosiFlush™ SP Syringes "have been verified to meet the established performance criteria above." The results of this design verification and biocompatibility testing "demonstrate that the modifications made to the subject device do not affect the safety and efficacy of the device and demonstrated the same intended use, fundamental scientific technology and operating principles" as the predicate device, thereby proving substantial equivalence.
Ask a specific question about this device
(123 days)
0.9% Sodium Chloride Injection, USP BD PosiFlush SF Syringe
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringes are intended to be used only for the flushing of indwelling vascular access devices.
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush" SF Syringe is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringe is sterilized via moist heat and provided externally sterile for use on a sterile field.
Here's an analysis of the provided text regarding the acceptance criteria and study for the 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringe:
This document is a 510(k) summary for a medical device that claims substantial equivalence to a predicate device, rather than a study proving the device meets novel acceptance criteria through a comparison against a primary ground truth. It focuses on demonstrating that the subject device (with new packaging) performs identically to the predicate device, which has already been deemed safe and effective.
Therefore, many of the typical questions asked for an AI/diagnostic study (like sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) are not applicable to this type of submission. This is a submission for a modification to an existing device (packaging change), where the primary goal is to show that the modified device maintains the performance characteristics of the original device.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the performance of the predicate device (K121050). The reported device performance for the subject device is consistently "Identical to predicate" or "Equivalent to predicate".
Performance Characteristic | Acceptance Criteria (Predicate Device K121050) | Reported Device Performance (Subject Device) |
---|---|---|
Functional Testing | ||
Container Closure Integrity | No dye in solution; No leakage in luer well or tip threads; No leakage past stopper ribs; No dye between stopper ribs | Identical to predicate |
Break Loose Force | No leakage beyond stopper ribs | Identical to predicate |
Break Out Force | Acceptable forces required to remove plunger rod | Identical to predicate |
Sustaining Force | Consistent plunger rod forces | Identical to predicate |
Pump Force (10ml/hr) | 20N | Identical to predicate |
Pump Force (1ml/hr) | 13N | Identical to predicate |
Pump Force (0.1ml/hr) | 9N | Identical to predicate |
Dead Space / Expelled Volume | Equivalent to labeled volume on barrel | Identical to predicate |
Syringe Induced Reflux | 0 average reflux when connected to a 4 Fr catheter | Identical to predicate |
Package Integrity | Per ISO 11607 | Equivalent to predicate |
Package Stability | Per ISO 11607 | Equivalent to predicate |
Sodium Chloride Injection, USP Testing | ||
Bacterial Endotoxin | Per USP | Identical to predicate |
Particulate Matter | Per USP | Identical to predicate |
Assay of NaCl | Per USP | Identical to predicate |
Heavy Metals | Per USP | Identical to predicate |
Iron | Per USP | Identical to predicate |
UV/vis | Per USP | Identical to predicate |
pH | Per USP | Identical to predicate |
Biocompatibility Testing | ||
Cytotoxicity | Per ISO10993-5:1999, Non-Toxic | Identical to predicate |
Hemolysis | Per ISO10993-4:2002/A:2006, Non-Toxic | Identical to predicate |
Acute Systemic Toxicity | Per ISO10993-11:2006, Non-Toxic | Identical to predicate |
Intracutaneous Reactivity | Per ISO10993-10:2002/A1:2006, Non-Irritant | Identical to predicate |
Sensitization | Per ISO10993-10:2002/A1:2006, Non-Sensitizer | Identical to predicate |
Bacterial Mutagenicity | Per ISO10993-3, Non-Mutagenic | Identical to predicate |
In Vitro Mouse Lymphoma | Per ISO10993-3, Non-Mutagenic | Identical to predicate |
Mouse Embryo Assay | Per ISO10993-3, Non-Mutagenic | Identical to predicate |
Occular Irritation | Per ISO10993-10:2002/A1:2006, Non-Irritant | Identical to predicate |
Rabbit Pyrogen | Per ISO10993-11:2006, Non-Pyrogenic | Identical to predicate |
Subchronic Intracutaneous Toxicity | Per ISO10993-11:2006, Non-Toxic | Identical to predicate |
Chemical Extractables Analysis | LC/DAD/MS & GC/MS, No significant extractables | Identical to predicate |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document states that "Performance Testing conducted to support substantial equivalence." It details the types of tests performed (Functional, Chemical, Biocompatibility) but does not specify the sample sizes for these tests or the data provenance. These tests would typically be laboratory-based rather than involving patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as this is not a study requiring expert interpretation of results for ground truth establishment. The performance criteria are objective measurements as per recognized standards (e.g., USP, ISO).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. The tests performed are objective, quantitative or qualitative measurements against established specifications/standards, not subjective assessments requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This device is a pre-filled saline syringe, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. This device is a manual medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" or acceptance criteria for this medical device are based on:
- Established standards and regulations: e.g., USP (United States Pharmacopeia) for chemical properties and bacterial endotoxin, ISO (International Organization for Standardization) for biocompatibility and package integrity.
- Performance of the predicate device: The subject device is designed to perform "Identical to predicate" or "Equivalent to predicate" in all tested characteristics. The predicate device's performance already demonstrates its safety and effectiveness.
8. The sample size for the training set
This is not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
This is not applicable. As above, no training set is relevant for this type of device.
Ask a specific question about this device
(66 days)
0.9% SODIUM CHLORIDE INJECTION, USP BD POSIFLUSH; SP SYRINGE, SF SYRINGE
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes and 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringes are intended to be used only for the flushing of indwelling vascular access devices. Catalog Number 306547 10 mL 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes are generally compatible for use with syringe pumps.
The 0.9% Sodium Chloride Injection, USP, BD PosiFlush" SP and SF Syringe (also referred to as BD PosiFlush " SP and SF Syringe) is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The BD PosiFlush™ SP Syringe is provided with a sterile fluid path, and the BD PosiFlush™ SF Syringe is provided externally sterile for use on a sterile field. Both configurations are sterilized via moist heat. The modified BD PosiFlush™ SP Syringe and BD PosiFlush™ SF Syringe, the subject of this 510(k), includes a new resin material in the barrel of the prefilled syringe.
The provided text describes a 510(k) summary for the BD PosiFlush™ SP and SF Syringes, which are prefilled syringes containing 0.9% sodium chloride for flushing vascular access devices. The submission focuses on a change in the resin material for the syringe barrel and asserts substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text, while noting the absence of information typically found in AI/algorithm-focused submissions:
1. Table of Acceptance Criteria and Reported Device Performance
This product is a medical device (prefilled syringe), not an AI/algorithm. Therefore, the "device performance" here refers to the physical and chemical properties of the syringe and its contents, not a diagnostic accuracy or algorithmic output. The performance is assessed relative to a predicate device or established standards.
Performance Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|
Functional Testing | ||
Container Closure Integrity | No Dye in Solution | |
No Leakage in the luer well or tip threads | ||
No Leakage Past the Stopper Ribs | ||
No Dye Between Stopper Ribs | Performed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met. Specific results are not detailed in this summary for each test). | |
Break Loose Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device. |
Break Out Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device. |
Sustaining Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device. |
Retaining Ring Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device. |
Tip Cap Removal Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device. |
Sodium Chloride Injection, USP, Testing | Per USP Monograph, Sodium Chloride Injection | Performed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met. Specific results are not detailed in this summary for each test). |
Bacterial Endotoxin | Met USP Monograph requirements. | |
Particulate Matter | Met USP Monograph requirements. | |
Assay of NaCl | Met USP Monograph requirements. | |
Heavy Metals | Met USP Monograph requirements. | |
Iron | Met USP Monograph requirements. | |
pH | Met USP Monograph requirements. | |
Biocompatibility Testing | Per ISO10993 Series | |
LC/DAD/MS, GC-MS, ICP; acceptable extractable test. | Performed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met, specifically acceptable extractable test as specified). |
Note: The reported device performance generally states "performed in an equivalent manner to the predicate device and is safe and effective when used as intended" or implies meeting USP Monograph requirements. Specific quantitative results for each test are not detailed in this summary.
Absent Information (Typical for AI/Algorithm Studies)
The following information is not applicable or not provided in this 510(k) summary because the device is a physical medical device (syringe with saline), not an AI/algorithm.
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for a physical device. Testing involves batches of syringes, not a "test set" of data.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical device testing is based on established engineering and chemical standards.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): For a physical device, ground truth is based on established engineering specifications, chemical analysis standards (e.g., USP Monograph), and biocompatibility standards (e.g., ISO 10993).
- The sample size for the training set: Not applicable (no AI/algorithm).
- How the ground truth for the training set was established: Not applicable (no AI/algorithm).
Conclusion of the Study (Non-Clinical Testing)
The company performed non-clinical testing based on a risk analysis. The results of these tests demonstrated that the modified BD PosiFlush™ SP and SF Syringe performed in an equivalent manner to the predicate device. This equivalence, combined with adherence to established standards (USP Monograph, ISO 10993 series), is deemed sufficient to prove the device's safety and effectiveness.
Clinical testing was explicitly stated as "not required" for this submission.
The conclusion is that the device has been validated to meet the established performance criteria based on non-clinical tests, and performs as safely and effectively as the legally marketed predicate devices.
Ask a specific question about this device
(307 days)
0.9% SODIUM CHLORIDE INJECTION USP BD POSIFLUSH FLUSH SYRINGE, SF FLUSH SYRINGE
BD PosiFlush™ Normal Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices. 10ml BD PosiFlush™ Normal Saline. Flush Svringes are generally compatible for use with syringe pumps.
The modified BD PosiFlush Flush Syringe is a three-piece, sterile, single use syringe with a 6% (Luer) connector pre-filled with 0.9% Sodium Chloride Injection, USP and sealed with a tip cap. BD has altered the modified device from the predicate by changing the stopper material and adding an extra thread to the plunger rod, threaded stopper insert. All performance characteristics are equivalent to the predicate device. The 0.9% Sodium Chloride Injection, USP BD PosiFlush Flush Syringes are intended for use in maintaining patency of vascular access devices (VAD's). This submission represents two 2 devices: the BD PosiFlush Flush Syringe, which is provided with a sterile fluid path; and the BD PosiFlush Flush SF Syringe, which is provided externally sterile for use on a sterile field. Both configurations are sterilized via moist heat.
This document describes the 510(k) summary for the BD PosiFlush Flush Syringe, which is a modified version of a predicate device. The primary change is the stopper material and an added extra thread to the plunger rod. The submission focuses on demonstrating substantial equivalence to the predicate device through performance testing.
Here's the breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and the Reported Device Performance
The document states that "Design Verification tests were performed based on the risk analysis conducted, and the results of these tests demonstrate that the BD PosiFlush Flush Syringe performed in an equivalent manner to the predicate device and is safe and effective when used as intended."
The table below summarizes the acceptance criteria and indicates that the reported device performance met these criteria, demonstrating equivalence to the predicate device.
Performance Characteristic | Acceptance Criteria | Reported Device Performance (as stated in document) |
---|---|---|
Functional Testing | ||
Container Closure Integrity | No Dye in Solution; No Leakage in the luer well or tip threads; No Leakage Past the Stopper Ribs; No Dye Between Stopper Ribs | Performed in an equivalent manner to the predicate device |
Break Loose Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device |
Break Out Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device |
Sustaining Force | Equivalence to Predicate | Performed in an equivalent manner to the predicate device |
Pump Force | 10ml/hr – 20N; 1ml/hr - 13N; 0.1ml/hr - 9N | Performed in an equivalent manner to the predicate device |
Dead Space / Expelled Volume | Equivalent to labeled volume (3ml, 5ml or 10ml) | Performed in an equivalent manner to the predicate device |
Syringe Induced Reflux | 0 average reflux when connected to a 4 Fr catheter | Performed in an equivalent manner to the predicate device |
Sodium Chloride Injection, USP Testing | ||
Bacterial Endotoxin | Per USP | Performed in an equivalent manner to the predicate device |
Particulate Matter | Per USP | Performed in an equivalent manner to the predicate device |
Assay of NaCl | Per USP | Performed in an equivalent manner to the predicate device |
Heavy Metals | Per USP | Performed in an equivalent manner to the predicate device |
Iron | Per USP | Performed in an equivalent manner to the predicate device |
UV/vis | Per USP | Performed in an equivalent manner to the predicate device |
pH | Per USP | Performed in an equivalent manner to the predicate device |
Biocompatibility Testing | ||
Cytotoxicity | Per ISO10993-5:1999, Non-Toxic | Performed in an equivalent manner to the predicate device |
Hemolysis | Per ISO10993-4:2002/A:2006, Non-Toxic | Performed in an equivalent manner to the predicate device |
Acute Systemic Toxicity | Per ISO10993-11:2006, Non-Toxic | Performed in an equivalent manner to the predicate device |
Intracutaneous Reactivity | Per ISO10993-10:2002/A1:2006, Non-Irritant | Performed in an equivalent manner to the predicate device |
Sensitization | Per ISO10993-10:2002/A1:2006, Non-Sensitizer | Performed in an equivalent manner to the predicate device |
Bacterial Mutagenicity | Per ISO10993-3, Non-Mutagenic | Performed in an equivalent manner to the predicate device |
In Vitro Mouse Lymphoma | Per ISO10993-3, Non-Mutagenic | Performed in an equivalent manner to the predicate device |
Mouse Embryo Assay | Per ISO10993-3, Non-Mutagenic | Performed in an equivalent manner to the predicate device |
Occular Irritation | Per ISO10993-10:2002/A1:2006, Non-Irritant | Performed in an equivalent manner to the predicate device |
Rabbit Pyrogen | Per ISO10993-11:2006, Non-Pyrogenic | Performed in an equivalent manner to the predicate device |
Subchronic Intracutaneous Toxicity | Per ISO10993-11:2006, Non-Toxic | Performed in an equivalent manner to the predicate device |
Chemical Extractables Analysis | LC/DAD/MS & GC/MS. No significant extractables | Performed in an equivalent manner to the predicate device |
2. Sample size used for the test set and the data provenance
The document states that "Design Verification testing included the following" and lists various tests. However, it does not specify the sample sizes used for each of these tests. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective. The testing is described as "Design Verification testing," which implies it was conducted specifically for this submission to prove equivalence.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The tests performed are primarily laboratory or analytical tests against established standards (e.g., USP, ISO 10993) rather than interpretations requiring expert consensus on clinical data.
4. Adjudication method for the test set
This information is not applicable and therefore not provided in the document, as the tests outlined are objective measurements against defined standards, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable to this device. An MRMC study or AI assistance is not mentioned or relevant for a pre-filled saline flush syringe.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable to this device. This is a medical device, not an algorithm or AI system.
7. The type of ground truth used
The ground truth used for the performance evaluation is based on established regulatory and industry standards. This includes:
- USP (United States Pharmacopeia) for chemical and particulate testing (e.g., USP , , , , , , ).
- ISO 10993 (Biological evaluation of medical devices) for biocompatibility testing (e.g., ISO 10993-5, -4, -11, -10, -3).
- Performance specifications of the predicate device for functional characteristics like force measurements and reflux.
8. The sample size for the training set
This is not applicable and therefore not provided in the document. This device does not involve machine learning algorithms that require a training set.
9. How the ground truth for the training set was established
This is not applicable and therefore not provided in the document, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1