Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K231161
    Date Cleared
    2023-07-21

    (88 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Saline Flush Syringes are intended to be used only for the flushing of indwelling vascular access devices.

    Device Description

    0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Saline Flush Syringe is a polypropylene syringe intended to flush indwelling vascular access devices. It is a sterile, single use syringe prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The device is packaged in a film-on-film blister pack and sterilized by gamma irradiation. External sterility of the device enables it to be used in the sterile field. The subject device is available only in 10mL syringe configuration.

    AI/ML Overview

    This document pertains to the 510(k) premarket notification for the "0.9% Sodium Chloride Injection, USP BD PosiFlush™ SF Saline Flush Syringe," a medical device, not an AI/ML-driven device or an imaging system. Therefore, the questions regarding AI/ML performance metrics, expert consensus for ground truth, MRMC studies, or training/test set details are not applicable to this submission.

    This submission focuses on demonstrating substantial equivalence to a predicate device (K153481) through bench performance testing, packaging integrity, and biocompatibility tests. The "acceptance criteria" here refer to the specifications for these non-clinical tests, rather than diagnostic performance metrics of an AI model.

    Here's the information extracted that aligns with the request, re-interpreting "acceptance criteria" and "study" in the context of this device's submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestPurposeAcceptance CriteriaResult
    Performance/Design Verification Tests
    Tip Cap Removal TorqueMeasure the torque required to remove the tip cap from the syringeTip Cap can be twisted off as per BD validated forcePass
    RefluxMeasure the potential blood reflux into the catheter from PosiFlush™ SF syringeReflux greater than BD validated internal valuePass
    Tip Cap LeakageEvaluate the resistance to leakage between the barrel Luer and the tip cap for pre-filled syringesNo evidence of Tip Cap leakagePass
    Leakage past StopperEvaluate the resistance to leakage between the barrel and stopper ribsNo leakage of solution past the stopper ribsPass
    Breakloose forceMeasure the initial maximum force required to move the plunger rod/stopper in the syringe barrelForce to move plunger rod/stopper is less than the BD validated forcePass
    Breakout ForceMeasure the maximum force required to move the plunger rod/stopper in the syringe barrel a short time after initially moving the plunger rod/stopperForce to move plunger rod/stopper is less than the BD validated forcePass
    Sustaining ForceMeasure the average force required to move the plunger rod/stopper in the syringe barrelForce to move plunger rod/stopper is less than the BD validated forcePass
    Retaining Ring ForceMeasure the force necessary to remove the plunger rod/stopper assembly from the syringe barrel after assemblyForce to move plunger rod/stopper is greater than the BD validated forcePass
    Expelled VolumeMeasure the volume of saline solution expelled from the syringeUSP43-NF38 <697>Container Content for InjectionsPass
    NaCl AssayMeasure the NaCl concentration saline solution in syringe samples0.9% NaCl Injection USP MonographPass
    pHMeasure the pH of saline solution in syringe samples0.9% NaCl Injection USP Monograph and USP43-NF38 <791> pHPass
    IronMeasure the amount of iron in saline solution0.9% NaCl Injection USP Monograph and USP43-NF38 <241> IronPass
    Heavy MetalsMeasure the heavy metals such as Cadmium, Arsenic, Cobalt, Vanadium, Copper, Lead, Nickel, Lithium, Antimony and Mercury in saline solutionUSP43-NF38 <232> Elemental Impurities - LimitsPass
    NaCl solution weight lossMeasure the weight loss of saline solution in pre-filled saline syringesUSP43-NF38 <671> Containers Performance TestingPass
    UV AnalysisMeasure the UV absorbance between 220-360 nm for pre-filled saline syringes< 0.4 AUPass
    Bacterial EndotoxinDetermine the amount of endotoxin in saline pre-filled syringes0.9% NaCl Injection USP Monograph and USP43-NF38 <85> Bacterial Endotoxins TestPass
    SterilityTo verify 10-6 SAL in the fluid path.Sterile; No growthPass
    Surface SterilityTo verify 10-6 SAL on all surface of syringe and inside surface of primary packagingSterile; No growthPass
    Particulate MatterMeasure the number of particulates in saline pre-filled syringesUSP43-NF38 <788> Particular Matter in InjectionsPass
    Appearance and Solution Clarity/Barrel TransparencyExamine the appearance, solution clarity and barrel transparency in pre-filled saline syringesSolution and components are clearPass
    Primary Package Integrity TestsEvaluates the sterile barrier system of the pre-filled syringe throughout shelf lifeNo leaks when tested by Vacuum Bubble Leak (ASTM F2096-11)Pass
    Non-porousPass
    Seal strength and width value as per BD validated force and width measurements respectivelyPass
    No defects in packaging seals when visually inspectedPass
    Clean peel of blister pack with no foreign matterPass
    Biocompatibility Tests
    CytotoxicityISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityGrade ≤ 2Pass
    SensitizationISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin SensitizationNon-SensitizerPass
    IrritationISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for IrritationFinal Test Sample Score ≤ 1Pass
    Acute Systemic ToxicityISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicityNo significantly greater biological reaction than the controlPass
    Material Mediated PyrogenicityISO 10993-11:2017 Biological evaluation of medical devices — Part 11: Tests for systemic toxicity / USP43-NF38 <151> Pyrogen Test (USP Rabbit Test)No temperature rise ≥ 0.5° CPass
    HemocompatibilityISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood / ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials≤ 5% hemolysisPass
    GenotoxicityISO 10993-3: 2014 Biological evaluation of medical devices – Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicityNon-mutagenicPass
    LAL EndotoxinUSP 43-NF38 <161> Medical Devices – Bacterial Endotoxin / USP43-NF38 <85> Bacterial Endotoxins Test / ANSI AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testingBelow the Endotoxin Limit 0.5 EU/devicePass
    Extractable and Leachable AnalysisISO 10993-18: 2020 Biological evaluation of medical devices - Part 18: Chemical characterization of materials / ISO 10993-17: 2002 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substancesUnder the condition of the chemical characterization testing, there were no extractables and leachables identified that indicated significant risk concerns.Note 1

    Note 1: Result is a qualitative statement, not a numerical pass/fail against a specific threshold listed in the table.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify numerical sample sizes for each of the performance, packaging, or biocompatibility tests. It only states that "appropriate performance tests and biocompatibility tests were performed" and the gamma sterilized test samples were evaluated. The data provenance (country of origin, retrospective/prospective) is also not specified, as these are bench and lab-based tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. This submission is for a physical medical device, not an AI/ML imaging device. Ground truth, in this context, refers to the established standards, monograph requirements (e.g., USP, ISO), and validated internal specifications to which the device performance is measured against. Experts here would be the laboratory personnel and engineers performing the tests and
    interpreting results against these defined criteria, but their specific number or detailed qualifications are not provided in the summary.

    4. Adjudication Method for the Test Set:

    Not applicable. Since this is not an AI/ML diagnostic device, there is no "adjudication" of expert opinions on image interpretation. The tests performed are laboratory or bench-based, with results measured against predetermined quantitative or qualitative acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI/ML device, and no MRMC study was performed. The device's safety and effectiveness are demonstrated through non-clinical performance, packaging, and biocompatibility testing.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical product (saline syringe), not an algorithm or software. No standalone algorithm performance was assessed.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's evaluation is primarily based on:

    • Established Standards: e.g., ISO 10993 series for biocompatibility, ASTM F2096-11 for vacuum bubble leak, ANSI AAMI ST72:2019 for Bacterial Endotoxins.
    • Pharmacopoeia Monographs: e.g., USP43-NF38 for NaCl Injection, pH, Iron, Heavy Metals, Particulate Matter, Bacterial Endotoxin.
    • Validated Internal Specifications: "BD validated force" or "BD validated internal value" for mechanical/physical properties like tip cap removal torque, reflux, breakout force, sustaining force, retaining ring force, and seal strength.
    • Qualitative Observations: e.g., "No evidence of Tip Cap leakage," "No leakage of solution past the stopper ribs," "Solution and components are clear," "No defects in packaging seals when visually inspected."

    Essentially, the ground truth is defined by recognized industry standards, pharmacopoeial requirements, and the manufacturer's own validated specifications for product performance and safety.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device; there is no "training set."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. As there is no training set for an AI/ML model, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161552
    Date Cleared
    2017-02-02

    (241 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe is intended to be used only for the flushing of indwelling vascular access devices. Catalog Number 306547 10 mL BD PosiFlush™ SP Syringes are generally compatible for use with syringe pumps.

    Device Description

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP (also referred to as BD PosiFlush™ SP Syringe) is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The BD PosiFlush™ SP Syringe is provided with a sterile fluid path, which is sterilized via moist heat.

    The change described in this submission is the qualification of a supplier specific formulation for stopper material for 0.9 % Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes.

    AI/ML Overview

    This document describes a Special 510(k) submission for a medical device called "0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringe" (K161552). The submission aims to demonstrate substantial equivalence to a previously cleared predicate device (K141311) due to a change in the supplier specific formulation for the stopper material.

    I will focus on the provided information to answer your request, specifically looking for acceptance criteria and the study proving it.

    Here's an analysis based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "Design Verification testing was also conducted to ensure the device met the predetermined acceptance criteria for the following tests." It lists the tests but does not provide a specific table enumerating the acceptance criteria values themselves or the quantitative performance measurements obtained for each test. Instead, it states that the device "met predetermined acceptance criteria per BD's internal specification" and "have been verified to meet the established performance criteria above."

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityAll tests (Cytotoxicity, Hemolysis, Acute systemic toxicity, Sub-Chronic Toxicity, Intracutaneous reactivity, Occular irritation, Partial thromboplastin Time, Sensitization, Pyrogenicity, Genotoxicity, Chemical extractable analysis) met the requirements of the FDA Guidance Use of International Standard ISO 10993-1.
    Design VerificationMet predetermined acceptance criteria per BD's internal specification for: - Container Closure Integrity - Break Out/Sustaining force - Stopper Separation - Leakage - Pump Force - Ship Test - Dead Space - Syringe Induced Reflux

    2. Sample Sizes Used for the Test Set and Data Provenance:

    The document does not specify the sample sizes used for any of the tests (biocompatibility or design verification). It also does not mention the country of origin of the data or whether the data was retrospective or prospective. The studies were conducted by Becton, Dickinson and Company.

    3. Number of Experts and Qualifications for Ground Truth:

    This section is not applicable as the clearance is for a pre-filled syringe (a medical device, not an AI or diagnostic tool that requires human expert interpretation for ground truth). The "ground truth" for the performance tests would be based on established engineering and biological standards and verified by internal testing protocols.

    4. Adjudication Method:

    This is not applicable for the same reasons as point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This is not applicable as the device is a pre-filled syringe, not an AI-assisted diagnostic tool.

    6. Standalone (Algorithm Only) Performance:

    This is not applicable as the device is a pre-filled syringe and does not involve an algorithm.

    7. Type of Ground Truth Used:

    For the performance tests, the "ground truth" would be established by:

    • Established industry standards and regulations (e.g., ISO 10993 for biocompatibility).
    • BD's internal specifications and engineering requirements for device functionality and integrity (e.g., for leakage, break force, dead space).

    8. Sample Size for the Training Set:

    This is not applicable as this is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    This is not applicable for the same reasons as point 8.

    Study Proving Device Meets Acceptance Criteria:

    The study that proves the device meets the acceptance criteria is detailed across two main categories:

    1. Biocompatibility Testing:

      • Methodology: Evaluated in accordance with FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," June 16, 2016.
      • Tests Performed: Cytotoxicity (per ISO10993-5), Hemolysis (per ISO10993-4), Acute systemic toxicity (per ISO10993-11), Sub-Chronic Toxicity (Per ISO10993-11), Intracutaneous reactivity (per ISO10993-10), Ocular irritation (per ISO 10993-10), Partial thromboplastin Time (ISO 10993-4), Sensitization (per ISO10993-10), Pyrogenicity (per ISO 10993-11), Genotoxicity (per ISO 10993-3), and Chemical extractable analysis (per ISO 10993-18).
      • Conclusion: The tests confirmed that the device met the requirements outlined in the standards, indicating it is biocompatible.
    2. Design Verification Testing:

      • Methodology: Conducted to ensure the device met "predetermined acceptance criteria per BD's internal specification." The specific methodologies for each test (e.g., how "Container Closure Integrity" was tested) are not detailed but would be standard industry practices for syringe manufacturing.
      • Tests Performed: Container Closure Integrity, Break Out/Sustaining force, Stopper Separation, Leakage, Pump Force, Ship Test, Dead Space, Syringe Induced Reflux.
      • Conclusion: The testing demonstrated that the subject device "met predetermined acceptance criteria."

    Overall Conclusion from the Document:

    The submission concludes that the BD PosiFlush™ SP Syringes "have been verified to meet the established performance criteria above." The results of this design verification and biocompatibility testing "demonstrate that the modifications made to the subject device do not affect the safety and efficacy of the device and demonstrated the same intended use, fundamental scientific technology and operating principles" as the predicate device, thereby proving substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153481
    Date Cleared
    2016-04-04

    (123 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringes are intended to be used only for the flushing of indwelling vascular access devices.

    Device Description

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush" SF Syringe is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringe is sterilized via moist heat and provided externally sterile for use on a sterile field.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringe:

    This document is a 510(k) summary for a medical device that claims substantial equivalence to a predicate device, rather than a study proving the device meets novel acceptance criteria through a comparison against a primary ground truth. It focuses on demonstrating that the subject device (with new packaging) performs identically to the predicate device, which has already been deemed safe and effective.

    Therefore, many of the typical questions asked for an AI/diagnostic study (like sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) are not applicable to this type of submission. This is a submission for a modification to an existing device (packaging change), where the primary goal is to show that the modified device maintains the performance characteristics of the original device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the performance of the predicate device (K121050). The reported device performance for the subject device is consistently "Identical to predicate" or "Equivalent to predicate".

    Performance CharacteristicAcceptance Criteria (Predicate Device K121050)Reported Device Performance (Subject Device)
    Functional Testing
    Container Closure IntegrityNo dye in solution; No leakage in luer well or tip threads; No leakage past stopper ribs; No dye between stopper ribsIdentical to predicate
    Break Loose ForceNo leakage beyond stopper ribsIdentical to predicate
    Break Out ForceAcceptable forces required to remove plunger rodIdentical to predicate
    Sustaining ForceConsistent plunger rod forcesIdentical to predicate
    Pump Force (10ml/hr)20NIdentical to predicate
    Pump Force (1ml/hr)13NIdentical to predicate
    Pump Force (0.1ml/hr)9NIdentical to predicate
    Dead Space / Expelled VolumeEquivalent to labeled volume on barrelIdentical to predicate
    Syringe Induced Reflux0 average reflux when connected to a 4 Fr catheterIdentical to predicate
    Package IntegrityPer ISO 11607Equivalent to predicate
    Package StabilityPer ISO 11607Equivalent to predicate
    Sodium Chloride Injection, USP Testing
    Bacterial EndotoxinPer USP <85>Identical to predicate
    Particulate MatterPer USP <788>Identical to predicate
    Assay of NaClPer USP <11>Identical to predicate
    Heavy MetalsPer USP <231>Identical to predicate
    IronPer USP <241>Identical to predicate
    UV/visPer USP <851>Identical to predicate
    pHPer USP <791>Identical to predicate
    Biocompatibility Testing
    CytotoxicityPer ISO10993-5:1999, Non-ToxicIdentical to predicate
    HemolysisPer ISO10993-4:2002/A:2006, Non-ToxicIdentical to predicate
    Acute Systemic ToxicityPer ISO10993-11:2006, Non-ToxicIdentical to predicate
    Intracutaneous ReactivityPer ISO10993-10:2002/A1:2006, Non-IrritantIdentical to predicate
    SensitizationPer ISO10993-10:2002/A1:2006, Non-SensitizerIdentical to predicate
    Bacterial MutagenicityPer ISO10993-3, Non-MutagenicIdentical to predicate
    In Vitro Mouse LymphomaPer ISO10993-3, Non-MutagenicIdentical to predicate
    Mouse Embryo AssayPer ISO10993-3, Non-MutagenicIdentical to predicate
    Occular IrritationPer ISO10993-10:2002/A1:2006, Non-IrritantIdentical to predicate
    Rabbit PyrogenPer ISO10993-11:2006, Non-PyrogenicIdentical to predicate
    Subchronic Intracutaneous ToxicityPer ISO10993-11:2006, Non-ToxicIdentical to predicate
    Chemical Extractables AnalysisLC/DAD/MS & GC/MS, No significant extractablesIdentical to predicate

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document states that "Performance Testing conducted to support substantial equivalence." It details the types of tests performed (Functional, Chemical, Biocompatibility) but does not specify the sample sizes for these tests or the data provenance. These tests would typically be laboratory-based rather than involving patient data.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is not a study requiring expert interpretation of results for ground truth establishment. The performance criteria are objective measurements as per recognized standards (e.g., USP, ISO).


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. The tests performed are objective, quantitative or qualitative measurements against established specifications/standards, not subjective assessments requiring adjudication.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a pre-filled saline syringe, not an AI-powered diagnostic or assistive tool.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This device is a manual medical device, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" or acceptance criteria for this medical device are based on:

    • Established standards and regulations: e.g., USP (United States Pharmacopeia) for chemical properties and bacterial endotoxin, ISO (International Organization for Standardization) for biocompatibility and package integrity.
    • Performance of the predicate device: The subject device is designed to perform "Identical to predicate" or "Equivalent to predicate" in all tested characteristics. The predicate device's performance already demonstrates its safety and effectiveness.

    8. The sample size for the training set

    This is not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set.


    9. How the ground truth for the training set was established

    This is not applicable. As above, no training set is relevant for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141311
    Date Cleared
    2014-07-25

    (66 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes and 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SF Syringes are intended to be used only for the flushing of indwelling vascular access devices. Catalog Number 306547 10 mL 0.9% Sodium Chloride Injection, USP, BD PosiFlush™ SP Syringes are generally compatible for use with syringe pumps.

    Device Description

    The 0.9% Sodium Chloride Injection, USP, BD PosiFlush" SP and SF Syringe (also referred to as BD PosiFlush " SP and SF Syringe) is a three-piece, sterile, single use syringe with a 6% (luer) connector prefilled with 0.9% sodium chloride injection, USP, and sealed with a tip cap. The BD PosiFlush™ SP Syringe is provided with a sterile fluid path, and the BD PosiFlush™ SF Syringe is provided externally sterile for use on a sterile field. Both configurations are sterilized via moist heat. The modified BD PosiFlush™ SP Syringe and BD PosiFlush™ SF Syringe, the subject of this 510(k), includes a new resin material in the barrel of the prefilled syringe.

    AI/ML Overview

    The provided text describes a 510(k) summary for the BD PosiFlush™ SP and SF Syringes, which are prefilled syringes containing 0.9% sodium chloride for flushing vascular access devices. The submission focuses on a change in the resin material for the syringe barrel and asserts substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text, while noting the absence of information typically found in AI/algorithm-focused submissions:

    1. Table of Acceptance Criteria and Reported Device Performance

    This product is a medical device (prefilled syringe), not an AI/algorithm. Therefore, the "device performance" here refers to the physical and chemical properties of the syringe and its contents, not a diagnostic accuracy or algorithmic output. The performance is assessed relative to a predicate device or established standards.

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Functional Testing
    Container Closure IntegrityNo Dye in SolutionNo Leakage in the luer well or tip threadsNo Leakage Past the Stopper RibsNo Dye Between Stopper RibsPerformed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met. Specific results are not detailed in this summary for each test).
    Break Loose ForceEquivalence to PredicatePerformed in an equivalent manner to the predicate device.
    Break Out ForceEquivalence to PredicatePerformed in an equivalent manner to the predicate device.
    Sustaining ForceEquivalence to PredicatePerformed in an equivalent manner to the predicate device.
    Retaining Ring ForceEquivalence to PredicatePerformed in an equivalent manner to the predicate device.
    Tip Cap Removal ForceEquivalence to PredicatePerformed in an equivalent manner to the predicate device.
    Sodium Chloride Injection, USP, TestingPer USP Monograph, Sodium Chloride InjectionPerformed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met. Specific results are not detailed in this summary for each test).
    Bacterial EndotoxinMet USP Monograph requirements.
    Particulate MatterMet USP Monograph requirements.
    Assay of NaClMet USP Monograph requirements.
    Heavy MetalsMet USP Monograph requirements.
    IronMet USP Monograph requirements.
    pHMet USP Monograph requirements.
    Biocompatibility TestingPer ISO10993 SeriesLC/DAD/MS, GC-MS, ICP; acceptable extractable test.Performed in an equivalent manner to the predicate device and is safe and effective when used as intended. (Implies all criteria were met, specifically acceptable extractable test as specified).

    Note: The reported device performance generally states "performed in an equivalent manner to the predicate device and is safe and effective when used as intended" or implies meeting USP Monograph requirements. Specific quantitative results for each test are not detailed in this summary.

    Absent Information (Typical for AI/Algorithm Studies)

    The following information is not applicable or not provided in this 510(k) summary because the device is a physical medical device (syringe with saline), not an AI/algorithm.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for a physical device. Testing involves batches of syringes, not a "test set" of data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical device testing is based on established engineering and chemical standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For a physical device, ground truth is based on established engineering specifications, chemical analysis standards (e.g., USP Monograph), and biocompatibility standards (e.g., ISO 10993).
    7. The sample size for the training set: Not applicable (no AI/algorithm).
    8. How the ground truth for the training set was established: Not applicable (no AI/algorithm).

    Conclusion of the Study (Non-Clinical Testing)

    The company performed non-clinical testing based on a risk analysis. The results of these tests demonstrated that the modified BD PosiFlush™ SP and SF Syringe performed in an equivalent manner to the predicate device. This equivalence, combined with adherence to established standards (USP Monograph, ISO 10993 series), is deemed sufficient to prove the device's safety and effectiveness.

    Clinical testing was explicitly stated as "not required" for this submission.

    The conclusion is that the device has been validated to meet the established performance criteria based on non-clinical tests, and performs as safely and effectively as the legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1