Search Filters

Search Results

Found 26 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY Industry (Manjung) Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Nittile powder free exam glove is a disposable device, worn on the hand or finger to prevent contamination between examiner and patient. This glove has also been tested for use with Chemotherapy Drugs.

    Device Description

    Non-sterile, Powder Free, ambidextrous, textured, beaded cuff, single use, Doffy® Nitrile Examination Glove, Tested for use with Chemotherapy Drugs

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device: "Non-sterile, Powder Free, ambidextrous, textured, beaded cuff, single use, Doffy® Nitrile Examination Glove, Tested for use with Chemotherapy Drugs".

    The main acceptance criteria for this device, beyond its primary function as an examination glove, is its resistance to permeation by various chemotherapy drugs. This is evaluated by the "Minimum Breakthrough Detection Time (minutes)" for each tested drug.

    Here's the breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Chemotherapy Drug and Concentration)Reported Device Performance (Min Breakthrough Detection Time - minutes)
    Carboplatin, 10 mg/ml>240
    Carmustine, 3.3 mg/ml15.0
    Cisplatin (BCNU), 1.0 mg/ml>240
    Cyclophosphamide (Cytoxan), 20.0 mg/ml>240
    Dacarbazine (DTIC), 10.0 mg/ml>240
    Doxorubicin Hydrochloride, 2.0 m/ml>240
    Etoposide (Toposar), 20.0 mg/ml>240
    Fluorouracil, 5.0 mg/ml>240
    Ifosfamide, 50.0 mg/ml>240
    Methotrexate, 25 mg/ml>240
    MitomycinC, 0.5mg/ml>240
    Mitoxantrone, 2 mg/ml>240
    Paclitaxel (Taxol), 6.0 mg/ml>240
    ThioTEPA, 10.0 mg/ml30.1
    Vincristine Sulfate, 1.0 mg/ml>240

    Note: The document itself specifies a warning based on these results: "WARNING: NOT TO BE USED WITH CARMUSTINE OR THIOTEPA" due to their extremely low permeation times (15.0 minutes and 30.1 minutes respectively). While the provided table states the reported performance, the implied acceptance criteria for safe use would be a breakthrough time considered acceptable for handling (likely much higher than 15 or 30 minutes, commonly 240 minutes per ASTM D6978). The device meets the criteria by reporting its performance, allowing users to make informed decisions and the FDA to determine substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size for the test set (number of gloves tested or repetitions per drug). It also does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. We only know the applicant is YTY Industry (Manjung) Sdn Bhd from Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a physical product (nitrile examination glove) and not an AI or diagnostic tool requiring expert interpretation of images or data. Therefore, the concept of "experts establishing ground truth" in the typical sense for test sets of AI models is not applicable here. The ground truth (breakthrough time) is established through standardized laboratory testing protocols (likely ASTM D6978 for chemotherapy drug permeation), not human expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for a physical product performance test. This concept is relevant for studies involving human interpretation or subjective assessment where consensus among multiple reviewers is needed to establish ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical product (glove), not an AI or assistive diagnostic technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical product (glove), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the chemotherapy drug permeation is established through laboratory testing using standardized methods, likely following a standard like ASTM D6978 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs". This involves precise measurement of the time it takes for a detectable amount of a chemotherapy drug to pass through the glove material.

    8. The sample size for the training set

    Not applicable. This device is a physical product. The concept of a "training set" is relevant for machine learning models, not for traditional manufacturing and testing of physical goods.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of this device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY Industry (Manjung) SDN BHD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

    Aloe Vera: Tested for use with Carboplatin, Carmustine, Cisplatin (BCNU), Cyclophosphamide (Cytoxan), Dacarbazine (DTIC), Doxorubicin Hydrochloride, Etoposide (Toposar), Fluorouracil, Ifosfamide, Methotrexate, Mitomycin C, Mitoxantrone, Paclitaxel (Taxol), ThioTEPA, and Vincristine Sulfate. WARNING: NOT TO BE USED WITH CARMUSTINE OR THIOTEPA.

    Non-Sterile, Powder Free Nitrile Examination Gloves: Tested for use with Carboplatin, Carmustine, Cisplatin (BCNU), Cyclophosphamide (Cytoxan), Dacarbazine (DTIC), Doxorubicin Hydrochloride, Etoposide (Toposar), Fluorouracil, Ifosfamide, Methotrexate, Mitomycin C, Mitoxantrone, Paclitaxel (Taxol), ThioTEPA, Vincristine Sulfate, and Fentanyl Citrate. WARNING: NOT TO BE USED WITH CARMUSTINE OR THIOTEPA.

    Device Description

    Non-sterile, Powder Free Nitrile Examination Aloe Vera Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs and Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs and Fentanyl Citrate.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for nitrile examination gloves and does not contain information about the acceptance criteria or a study that typically involves AI/ML device performance. Instead, it details the indications for use and the gloves' resistance to permeation by chemotherapy drugs and fentanyl citrate, in accordance with ASTM D6978.

    Therefore, I cannot extract the information required to answer your questions as they pertain to AI/ML device performance studies and acceptance criteria. The document focuses on regulatory clearance for medical gloves based on chemical resistance testing, not on an AI/ML algorithm's performance.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY Industry (Manjung) Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:

    Blue
    Tested Chemotherapy Drug And ConcentrationAverage Breakthrough Detection Time (minutes)
    Carboplatin, 10 mg/ml> 240 minutes
    Carmustine, 3.3 mg/ml11.3 minutes
    Cisplatin (BCNU), 1.0 mg/ml> 240 minutes
    Cyclophosphamide (Cytoxan), 20.0 mg/ml> 240 minutes
    Dacarbazine (DTIC), 10.0 mg/ml> 240 minutes
    Doxorubicin Hydrochloride, 2.0 mg/ml> 240 minutes
    Etoposide (Toposar), 20.0 mg/ml> 240 minutes
    Fluorouracil, 50.0 mg/ml> 240 minutes
    Ifosfamide, 50.0 mg/ml> 240 minutes
    Methotrexate, 25 mg/ml> 240 minutes
    Mitomycin C, 0.5 mg/ml> 240 minutes
    Mitoxantrone, 2 mg/ml> 240 minutes
    Paclitaxel (Taxol), 6.0 mg/ml> 240 minutes
    ThioTEPA, 10.0 mg/ml7.4 minutes
    Vincristine Sulfate, 1.0 mg/ml> 240 minutes

    The following chemotherapy drugs and concentration have extremely low permeation time. Carmustine (3.3 mg/ml): 11.3 minutes ThioTEPA (10.0 mg/ml): 7.4 minutes

    WARNING: NOT TO BE USED WITH CARMUSTINE OR THIOTEPA

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs and Fentanyl Citrate:

    Blue
    Tested Chemotherapy Drug And ConcentrationAverage Breakthrough Detection Time (minutes)
    Carboplatin, 10 mg/ml> 240 minutes
    Carmustine, 3.3 mg/ml30.3 minutes
    Cisplatin (BCNU), 1.0 mg/ml> 240 minutes
    Cyclophosphamide (Cytoxan), 20.0 mg/ml> 240 minutes
    Dacarbazine (DTIC), 10.0 mg/ml> 240 minutes
    Doxorubicin Hydrochloride, 2.0 mg/ml> 240 minutes
    Etoposide (Toposar), 20.0 mg/ml> 240 minutes
    Fluorouracil, 50.0 mg/ml> 240 minutes
    Ifosfamide, 50.0 mg/ml> 240 minutes
    Methotrexate, 25 mg/ml> 240 minutes
    Mitomycin C, 0.5 mg/ml> 240 minutes
    Mitoxantrone, 2 mg/ml> 240 minutes
    Paclitaxel (Taxol), 6.0 mg/ml> 240 minutes
    ThioTEPA, 10.0 mg/ml91.0 minutes
    Vincristine Sulfate, 1.0 mg/ml> 240 minutes
    Busulfan, 6 mg/ml> 240 minutes
    Cytarabine, 100 mg/ml> 240 minutes
    Daunorubicin, 5 mg/ml> 240 minutes
    Docetaxel, 10 mg/ml> 240 minutes
    Epirubicin, 2 mg/ml> 240 minutes
    Gemcitabine, 38 mg/ml> 240 minutes
    Irinotecan Hydrochloride, 20 mg/ml> 240 minutes
    Mechlorethamine HCL, 1.0 mg/ml> 240 minutes
    Melphalan, 5 mg/ml> 240 minutes
    Trisenox, 0.1 mg/ml> 240 minutes
    Fentanyl Citrate, 100mcg/2ml> 240 minutes

    The following chemotherapy drugs and concentration have extremely low permeation time. Carmustine (3.3 mg/ml): 30.3 minutes ThioTEPA (10.0 mg/ml): 91.0 minutes

    CAUTION: CARMUSTINE (3.3 MG/ML) AND THIOTEPA (10 MG/ML) HAVE MUCH LOWER PERMEATION TIMES COMPARE TO OTHER CHEMOTHERAPY DRUGS

    Device Description

    Non-Sterile, Powder Free Nitrile Examination Gloves, Low Dermatitis Potential, And Tested for use with Chemotherapy Drugs Blue; Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs and Fentanyl Citrate - Blue

    AI/ML Overview

    This document describes two types of non-sterile, powder-free nitrile examination gloves:

    1. Non-Sterile, Powder Free Nitrile Examination Gloves, Low Dermatitis Potential, And Tested for use with Chemotherapy Drugs Blue
    2. Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs and Fentanyl Citrate - Blue

    The acceptance criteria for both devices are primarily based on their resistance to permeation by chemotherapy drugs, as tested according to ASTM D6978 Standard Practice. For the second device, resistance to Fentanyl Citrate is also included.

    The study proving the device meets the acceptance criteria is detailed in the "Indications for Use" sections for each glove type.


    1. Table of Acceptance Criteria and Reported Device Performance

    Device 1: Non-Sterile, Powder Free Nitrile Examination Gloves, Low Dermatitis Potential, And Tested for use with Chemotherapy Drugs Blue

    Tested Chemotherapy Drug And ConcentrationAcceptance Criterion (Minimum Breakthrough Detection Time)Reported Device Performance (Average Breakthrough Detection Time)
    Carboplatin, 10 mg/mlNot explicitly stated (implied for medical use performance)> 240 minutes
    Carmustine, 3.3 mg/mlNot explicitly stated11.3 minutes
    Cisplatin (BCNU), 1.0 mg/mlNot explicitly stated> 240 minutes
    Cyclophosphamide (Cytoxan), 20.0 mg/mlNot explicitly stated> 240 minutes
    Dacarbazine (DTIC), 10.0 mg/mlNot explicitly stated> 240 minutes
    Doxorubicin Hydrochloride, 2.0 mg/mlNot explicitly stated> 240 minutes
    Etoposide (Toposar), 20.0 mg/mlNot explicitly stated> 240 minutes
    Fluorouracil, 50.0 mg/mlNot explicitly stated> 240 minutes
    Ifosfamide, 50.0 mg/mlNot explicitly stated> 240 minutes
    Methotrexate, 25 mg/mlNot explicitly stated> 240 minutes
    Mitomycin C, 0.5 mg/mlNot explicitly stated> 240 minutes
    Mitoxantrone, 2 mg/mlNot explicitly stated> 240 minutes
    Paclitaxel (Taxol), 6.0 mg/mlNot explicitly stated> 240 minutes
    ThioTEPA, 10.0 mg/mlNot explicitly stated7.4 minutes
    Vincristine Sulfate, 1.0 mg/mlNot explicitly stated> 240 minutes

    Note: The document explicitly warns: "WARNING: NOT TO BE USED WITH CARMUSTINE OR THIOTEPA" due to their extremely low permeation times (11.3 minutes and 7.4 minutes, respectively). This implicitly sets an acceptance criterion that usage with these drugs is not recommended, rather than a performance threshold for them. For other drugs, a permeation time of "> 240 minutes" appears to be the target performance for safe use.


    Device 2: Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs and Fentanyl Citrate - Blue

    Tested Chemotherapy Drug And ConcentrationAcceptance Criterion (Minimum Breakthrough Detection Time)Reported Device Performance (Average Breakthrough Detection Time)
    Carboplatin, 10 mg/mlNot explicitly stated (implied for medical use performance)> 240 minutes
    Carmustine, 3.3 mg/mlNot explicitly stated30.3 minutes
    Cisplatin (BCNU), 1.0 mg/mlNot explicitly stated> 240 minutes
    Cyclophosphamide (Cytoxan), 20.0 mg/mlNot explicitly stated> 240 minutes
    Dacarbazine (DTIC), 10.0 mg/mlNot explicitly stated> 240 minutes
    Doxorubicin Hydrochloride, 2.0 mg/mlNot explicitly stated> 240 minutes
    Etoposide (Toposar), 20.0 mg/mlNot explicitly stated> 240 minutes
    Fluorouracil, 50.0 mg/mlNot explicitly stated> 240 minutes
    Ifosfamide, 50.0 mg/mlNot explicitly stated> 240 minutes
    Methotrexate, 25 mg/mlNot explicitly stated> 240 minutes
    Mitomycin C, 0.5 mg/mlNot explicitly stated> 240 minutes
    Mitoxantrone, 2 mg/mlNot explicitly stated> 240 minutes
    Paclitaxel (Taxol), 6.0 mg/mlNot explicitly stated> 240 minutes
    ThioTEPA, 10.0 mg/mlNot explicitly stated91.0 minutes
    Vincristine Sulfate, 1.0 mg/mlNot explicitly stated> 240 minutes
    Busulfan, 6 mg/mlNot explicitly stated> 240 minutes
    Cytarabine, 100 mg/mlNot explicitly stated> 240 minutes
    Daunorubicin, 5 mg/mlNot explicitly stated> 240 minutes
    Docetaxel, 10 mg/mlNot explicitly stated> 240 minutes
    Epirubicin, 2 mg/mlNot explicitly stated> 240 minutes
    Gemcitabine, 38 mg/mlNot explicitly stated> 240 minutes
    Irinotecan Hydrochloride, 20 mg/mlNot explicitly stated> 240 minutes
    Mechlorethamine HCL, 1.0 mg/mlNot explicitly stated> 240 minutes
    Melphalan, 5 mg/mlNot explicitly stated> 240 minutes
    Trisenox, 0.1 mg/mlNot explicitly stated> 240 minutes
    Fentanyl Citrate, 100mcg/2mlNot explicitly stated> 240 minutes

    Note: The document states: "CAUTION: CARMUSTINE (3.3 MG/ML) AND THIOTEPA (10 MG/ML) HAVE MUCH LOWER PERMEATION TIMES COMPARE TO OTHER CHEMOTHERAPY DRUGS". Similar to Device 1, this indicates that while tested, their permeation times (30.3 minutes and 91.0 minutes, respectively) are lower than the target for other drugs, suggesting a cautionary use or implicit lower acceptance for these specific substances.


    2. Sample size used for the test set and the data provenance

    The document specifies that the gloves were "tested for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

    • Sample Size for the Test Set: The document does not explicitly state the sample size (e.g., number of gloves or tests performed per drug) used for these tests. The ASTM D6978 standard would likely specify the required number of replicates.
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. The testing was conducted by YTY Industry (Manjung) Sdn Bhd, located in Malaysia, suggesting the tests were performed in Malaysia. The nature of laboratory testing like this is inherently prospective for the specific batch or design of gloves being evaluated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable in the context of this document. The "ground truth" here is the physical measurement of chemical permeation, determined by laboratory testing methods described in ASTM D6978, rather than expert interpretation of medical images or diagnostic data. There is no indication of experts being used to establish a ground truth for the test results.


    4. Adjudication method for the test set

    This question is not applicable. The test results are objective chemical permeation measurements (breakthrough detection time), not subjective interpretations that require adjudication by experts.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device is a medical glove, not an AI-powered diagnostic tool requiring human interpretation or MRMC studies.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. The device is a medical glove, not an algorithm. The performance described is that of the physical product itself.


    7. The type of ground truth used

    The ground truth used is laboratory measurement of chemical permeation according to the ASTM D6978 standard. This standard describes methods to evaluate the resistance of medical gloves to permeation by chemotherapy drugs, typically by detecting the breakthrough of a specific chemical through the glove material over time.


    8. The sample size for the training set

    This question is not applicable. The device consists of physical gloves, not a machine learning model, and therefore does not have a "training set" in the computational sense. The manufacturing process of gloves might involve quality control and material testing, but this is distinct from training a dataset for an algorithm.


    9. How the ground truth for the training set was established

    This question is not applicable, as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182383
    Date Cleared
    2019-03-19

    (200 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY Industry (Manjung) Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Non-Sterile, PF Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs - Blue

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for "Non-Sterile, PF Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs - Blue." This document does not describe an AI/ML powered device, but rather a standard medical device.

    Therefore, the requested information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, multi-reader multi-case studies, standalone performance, and ground truth for an AI device is not applicable to this document.

    The document does include a table indicating the performance of the gloves with various chemotherapy drugs, specifically the "Average Breakthrough Detection Time." This table could be interpreted as describing device performance against acceptance criteria for chemotherapy drug permeation, but it is not related to AI/ML performance.

    Acceptance Criteria and Reported Device Performance (as related to chemotherapy drug permeation for the gloves):

    Tested Chemotherapy Drug And ConcentrationAcceptance Criteria (Implicit: > 240 minutes where stated otherwise)Reported Device Performance (Average Breakthrough Detection Time)
    Carboplatin, 10 mg/ml(Not explicitly stated, but common industry standard for high resistance)>240 minutes
    Carmustine 3.3 mg/ml(Not explicitly stated)11.3 minutes
    Cisplatin (BCNU), 1.0 mg/ml(Not explicitly stated)> 240 minutes
    Cyclophosphamide (Cytoxan), 20.0 mg/ml(Not explicitly stated)> 240 minutes
    Dacarbazine (DTIC), 10.0 mg/ml(Not explicitly stated)> 240 minutes
    Doxorubicin Hydrochloride, 2.0 mg/ml(Not explicitly stated)> 240 minutes
    Etoposide (Toposar), 20.0 mg/ml(Not explicitly stated)> 240 minutes
    Fluorouracil, 50.0 mg/ml(Not explicitly stated)> 240 minutes
    Ifosfamide , 50.0 mg/ml(Not explicitly stated)> 240 minutes
    Methotrexate, 25 mg/ml(Not explicitly stated)> 240 minutes
    Mitomycin C, 0.5 mg/ml(Not explicitly stated)> 240 minutes
    Mitoxantrone, 2 mg/ml(Not explicitly stated)> 240 minutes
    Paclitaxel (Taxol), 6.0 mg/ml(Not explicitly stated)> 240 minutes
    ThioTEPA (10.0 mg/ml)(Not explicitly stated)7.4 minutes
    Vincristine Sulfate, 1.0 mg/ml(Not explicitly stated)> 240 minutes

    The document states that "Carmustine (3.3 mg/ml): 11.3 minutes" and "ThioTEPA (10.0 mg/ml): 7.4 minutes" have "extremely low permeation time," which implies that for these drugs, the gloves do not offer extended protection. This information would be crucial for users to understand the limitations of the device.

    All other requested information about an AI/ML study is not present in this document as it is not a submission for an AI/ML medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171104
    Date Cleared
    2017-08-25

    (134 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN BHD

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. This glove is also tested for use against Chemotherapy Drugs. The Chemotherapy Drugs and its permeation time is listed as below.

    Device Description

    Non-Sterile, Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs - Blue and Non-Sterile, Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs - Cobalt Blue meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application and D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs and FDA 21 CFR 880.6250. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for two types of nitrile examination gloves: "Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs (Blue)" and "Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs (Cobalt Blue)". The information is presented as a 510(k) summary, comparing the subject devices to a predicate device (K161215).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The acceptance criteria for the gloves are primarily based on ASTM standards (ASTM D6319-10 for physical and dimensional properties, ASTM D6124-06 for powder-free claims, and ASTM D6978-05 for chemotherapy drug permeation), as well as ISO 10993-10 for biocompatibility. The reported device performance is compared against these standards and the predicate device.

    CharacteristicAcceptance Criteria (Standard)Blue Glove Performance (Reported)Cobalt Blue Glove Performance (Reported)
    Chemotherapy Permeation (ASTM D6978-05)(Comparison to predicate and general clinical expectation of protection)See detailed table belowSee detailed table below
    Carboplatin, 10mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Carmustine, 3.3mg/mlNo specific numerical acceptance listed15.0 minutes2.4 minutes
    Cisplatin, 1.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Cyclophosphamide, 20.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Dacarbazine, 10.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Doxorubicin HCl, 2.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Etoposide, 20.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Fluorouracil, 50.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Ifosfamide, 50.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Methotrexate, 25mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Mitomycin C, 0.5mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Mitoxantrone, 2mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Paclitaxel, 6.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    ThioTEPA, 10.0mg/mlNo specific numerical acceptance listed30.1 minutes1.9 minutes
    Vincristine Sulfate, 1.0mg/mlNo specific numerical acceptance listed, but >240 min is ideal> 240 minutes> 240 minutes
    Physical (ASTM D6319-10)Meets ASTM D6319-10 specificationsMeetsMeets
    Dimension (Length)Min 240mm (ASTM D6319-10)Min 240mmMin 240mm
    Thickness (Finger)No specific numerical acceptance for subject devices listed, predicate 0.03-0.05mm0.03mm min0.03mm min
    Thickness (Palm)No specific numerical acceptance for subject devices listed, predicate 1.5-0.08mm0.05mm min0.05mm min
    Thickness (Cuff)No specific numerical acceptance for subject devices listed, predicate 1.6-0.09mm0.05mm min0.05mm min
    Powder-free
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN. BHD.,

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. This glove is also tested for use against Chemotherapy Drugs. The Chemotherapy Drugs and its permeation time is listed as below.

    Device Description

    Non-Sterile, Powder Free Blue Nitrile Patient Examination Gloves, Tested for use with Chemotherapy Drugs (Original Blue, Cobalt Blue) meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application and D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs and FDA 21 CFR 880.6250. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for medical gloves, focusing on their resistance to chemotherapy drugs. It details the performance of "Non-Sterile, Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs (Original Blue, Cobalt Blue)" against a range of chemotherapy agents.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for chemotherapy drug permeation are defined by the "Average Breakthrough Detection Time" for each drug. While the specific numerical acceptance criteria (e.g., minimum time) are not explicitly stated as distinct "criteria" separate from the results, the implication is that the longer the breakthrough time, the better the performance. For many drugs, a breakthrough time of "> 240 minutes" is presented, suggesting this is a desirable outcome or a de facto acceptance benchmark for long-term protection. The device performance is the directly measured average breakthrough detection time for each drug.

    Below are the tables summarizing the performance for both "Original Blue" and "Cobalt Blue" glove variations, as tested against various chemotherapy drugs.

    Device Name: Non-Sterile, Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs - Original Blue

    Tested Chemotherapy Drug And ConcentrationAverage Breakthrough Detection Time (minutes)
    Carboplatin, 10 mg/ml> 240
    Carmustine (3.3 mg/ml)1.84
    Cisplatin (BCNU), 1.0 mg/ml> 240
    Cyclophosphamide (Cytoxan), 20.0 mg/ml> 240
    Dacarbazine (DTIC), 10.0 mg/ml> 240
    Doxorubicin Hydrochloride, 2.0 mg/ml> 240
    Etoposide (Toposar), 20.0 mg/ml> 240
    Fluorouracil, 50.0 mg/ml> 240
    Ifosfamide, 50.0 mg/ml> 240
    Methotrexate, 25 mg/ml> 240
    Mitomycin C, 0.5 mg/ml> 240
    Mitoxantrone, 2 mg/ml> 240
    Paclitaxel (Taxol), 6.0 mg/ml> 240
    ThioTEPA (10.0 mg/ml)0.76
    Vincristine Sulfate, 1.0 mg/ml> 240
    Warning for drugs with extremely low permeation time:Carmustine (3.3 mg/ml): 1.84 minutes
    ThioTEPA (10.0 mg/ml): 0.76 minutes

    Device Name: Non-Sterile, Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs - Cobalt Blue

    Tested Chemotherapy Drug And ConcentrationAverage Breakthrough Detection Time (minutes)
    Carboplatin, 10 mg/ml> 240
    Carmustine (3.3 mg/ml)1.82
    Cisplatin (BCNU), 1.0 mg/ml> 240
    Cyclophosphamide (Cytoxan), 20.0 mg/ml> 240
    Dacarbazine (DTIC), 10.0 mg/ml> 240
    Doxorubicin Hydrochloride, 2.0 mg/ml> 240
    Etoposide (Toposar), 20.0 mg/ml> 240
    Fluorouracil, 50.0 mg/ml> 240
    Ifosfamide, 50.0 mg/ml> 240
    Methotrexate, 25 mg/ml> 240
    Mitomycin C, 0.5 mg/ml> 240
    Mitoxantrone, 2 mg/ml> 240
    Paclitaxel (Taxol), 6.0 mg/ml> 240
    ThioTEPA (10.0 mg/ml)0.93
    Vincristine Sulfate, 1.0 mg/ml> 240
    Warning for drugs with extremely low permeation time:Carmustine (3.3 mg/ml): 1.82 minutes
    ThioTEPA (10.0 mg/ml): 0.93 minutes

    In addition to chemotherapy drug permeation, the gloves were also tested against other specifications, including:

    CharacteristicStandardAcceptance CriteriaDevice Performance (Original Blue)Device Performance (Cobalt Blue)
    DimensionASTM D6319-10Minimum 240mm lengthMinimum 240mmMinimum 240mm
    Physical PropertiesASTM D6319-10Meets standard requirementsMeetsMeets
    ThicknessASTM D6319-10Finger 0.07-0.10mm (Predicate 1), 0.10min (Predicate 2); Palm 0.07-0.09mm; Cuff 0.06-0.08mm (Predicate 1), Not specified (Cuff, Predicate 2)Finger 0.09-0.14mm; Palm 0.07-0.10mm; Cuff 0.06-0.09mmFinger 0.10min; Palm 0.07-0.09mm; Cuff 0.05-0.08mm
    Powder-freeASTM D6124-06 (
    Ask a Question

    Ask a specific question about this device

    K Number
    K143289
    Date Cleared
    2015-08-18

    (274 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    Non-Sterile, Powder Free Nitrile Examination Gloves – Orange, Green, Blue and Violet Color Color meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for "Non-Sterile, Powder Free Nitrile Examination Gloves - Orange, Green, Blue and Violet Color," which is seeking substantial equivalence to a predicate device (K052502).

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance (Ranges for Orange, Green, Blue, Violet)
    DimensionASTM D6319-10Meets (Specific ranges provided below)
    Length (Size M)Min 230 mm (Predicate: 240-251 mm)Orange: 242-257 mm
    Green: 240-249 mm
    Blue: 240-250 mm
    Violet: 240-244 mm
    Palm Width (Size M)95 ± 10 mm (Predicate: 94-96 mm)Orange: 96-100 mm
    Green: 95-99 mm
    Blue: 95-99 mm
    Violet: 96-98 mm
    Thickness Finger (Single Layer)Min 0.05 mm (Predicate: 0.15-0.19 mm)Orange: 0.11-0.14 mm
    Green: 0.11-0.14 mm
    Blue: 0.09-0.10 mm
    Violet: 0.09-0.11 mm
    Thickness Palm (Single Layer)Min 0.05 mm (Predicate: 0.12-0.16 mm)Orange: 0.09-0.11 mm
    Green: 0.07-0.08 mm
    Blue: 0.05-0.06 mm
    Violet: 0.06-0.07 mm
    Thickness Cuff (Single Layer)Not explicitly stated in tableOrange: 0.07-0.09 mm
    Green: 0.06-0.07 mm
    Blue: 0.04-0.05 mm
    Violet: 0.05-0.06 mm
    Physical PropertiesASTM D6319-10Meets (Specific ranges provided below)
    Before Aging: Tensile Strength (MPa)Min 14 MPa (Predicate: 26.00-30.00 MPa)Orange: 21.21-29.63
    Green: 24.15-29.21
    Blue: 28.46-33.44
    Violet: 23.22-29.11
    Before Aging: Ultimate Elongation (%)Min 500% (Predicate: 750-800%)Orange: 540-600
    Green: 580-620
    Blue: 520-580
    Violet: 520-580
    After Aging: Tensile Strength (MPa)Min 14 MPa (Predicate: 25.00-28.00 MPa)Orange: 29.23-33.56
    Green: 28.72-34.84
    Blue: 29.76-34.18
    Violet: 27.39-30.82
    After Aging: Ultimate Elongation (%)Min 400% (Predicate: 670-730%)Orange: 480-520
    Green: 520-540
    Blue: 440-520
    Violet: 440-500
    Freedom from Holes (Barrier)ASTM D6319-10 (AQL = 2.5, FDA 1000ml Water Leak Test)Meets (Holes found: 0 for all colors; Accept 1, Reject 7 criteria)
    AQLAQL 2.5 (Predicate: AQL 2.5)AQL 2.5 (Result: 0 for all colors)
    Powder-freeASTM D6319-10 (Max 2.0 mg/glove)Meets (Specific values below)
    Residual PowderMax 2.0 mg/glove (Predicate: 0.20 mg/glove)Orange: 0.16 mg/glove
    Green: 0.24 mg/glove
    Blue: 0.20 mg/glove
    Violet: 0.14 mg/glove
    Bio-compatibilityISO 10993-10 (2010), Non-irritant, Non-sensitizerMeets (Under the conditions of the study, the device is non-irritant or non-sensitizer for all colors)

    Study Proving Device Meets Acceptance Criteria:

    The document explicitly states that the device "meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application." The tables provided, particularly on pages 5 and 7, show the performance of the "Non-Sterile, Powder Free Nitrile Examination Gloves - Orange, Green, Blue and Violet Color" against these ASTM D6319-10 standards, and also against the FDA 1000ml Water Leak Test and ISO 10993-10 (2010) for biocompatibility.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: The specific sample sizes for each test (e.g., how many gloves were tested for length, how many for freedom from holes) are not explicitly detailed. For the "Watertight (1000ml)" test, it mentions "Multiple Normal GII AQL = 2.5", and the acceptance criterion is "(Accept 1, Reject 7)", which implies a sampling plan based on AQL principles, but not a fixed number of samples. This is a common statistical sampling approach in quality control of manufactured goods.
    • Data Provenance: The manufacturer is YTY Industry (Manjung) Sdn. Bhd., located in Sitiawan, Perak, Malaysia. Therefore, the data provenance is Malaysia. The study is a prospective test of the manufactured gloves against established standards, not a retrospective analysis.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided as it's not relevant for this type of device (nitrile examination gloves). The "ground truth" for glove performance is established by objective, standardized tests (ASTM, ISO, FDA methods), not by expert interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • None. As mentioned above, the performance of examination gloves is determined by objective physical and chemical testing against established standards, not by human adjudication of "ground truth".

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • A MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic medical devices where expert human readers interpret images or data, often with or without AI assistance. This device (examination gloves) does not involve human interpretation or AI assistance in its intended use or performance evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • A standalone algorithm performance study was not done. This device is a physical medical product, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for the performance of these gloves is defined by established international and national standards and test methods:
      • ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application)
      • FDA 1000ml Water Leak Test
      • ISO 10993-10 (2010) (Biological evaluation of medical devices - Tests for irritation and skin sensitization)
      • ASTM D6124-10 (Reapproved 2011) (Residual Powder on Medical Gloves)
      • ASTM D5151-06 (Reapproved 2011) (Detection of Holes in Medical Gloves)

    8. The sample size for the training set:

    • This information is not applicable. This is a physical manufactured product, not an AI/ML algorithm that requires a "training set". The manufacturing process is subject to quality control and adherence to standards.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, there is no "training set" for this type of medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143055
    Date Cleared
    2015-02-12

    (112 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves - Grape, Apple, Peppermint, and Vanilla meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the "Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves - Grape, Apple, Peppermint, Vanilla" based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicStandard / Acceptance CriteriaReported Device Performance (Grape, Apple, Peppermint, Vanilla)
    DimensionASTM D6319-10 (Min 230mm length for Size M)Meets (240-251mm length for Size M in all scented variations)
    Physical PropertiesASTM D6319-10 (Before Aging: Min 14 MPa Tensile Strength, Min 500% Elongation; After Aging: Min 14 MPa Tensile Strength, Min 400% Elongation)Meets (Reported ranges consistently exceed minimums for all scented variations)
    Freedom from holesASTM D6319-10 (AQL = 2.5)Meets (0 holes found in samples for all scented variations)
    Powder-freeASTM D6319-10 (Max 2.0 mg/glove residual powder)Meets (0.10 - 0.16 mg/glove residual powder for all scented variations)
    BiocompatibilityISO 10993-10 (Non-irritant, Non-sensitizer)Meets (Device is non-irritant and non-sensitizer under study conditions)
    Watertight (1000ml)ASTM D5151-06 (2011) / ASTM D6319-10 (AQL = 2.5)Meets (0 holes found in samples for all scented variations)
    Length (Size M)Min 230 mm (ASTM D6319-10)240-251 mm (all scented variations)
    Palm Width (Size M)95 +/- 10 mm (ASTM D6319-10)Ranges like "93-96" or "94-96" (all scented variations)
    Thickness (Single Layer)Min 0.05 mm (Finger, Palm) (ASTM D6319-10)Finger: 0.10-0.14 mm; Palm: 0.07-0.09 mm (all scented variations)

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not explicitly state the exact sample size used for each specific test (e.g., number of gloves tested for watertightness, physical properties, etc.). It refers to meeting general ASTM standards which typically define sample sizes for such tests.
    • The data provenance is Malaysia, as the manufacturing company (YTY Industry (Manjung) SDN. BHD.) is based there.
    • The study appears to be prospective as it's a submission for product clearance, implying these tests were conducted on the device being submitted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This is not applicable as the ground truth is established through adherence to standardized international and FDA-recognized test methods and specifications (ASTM D6319-10, ISO 10993-10, ASTM D5151-06, ASTM D6124-10). These standards define objective, measurable criteria, and therefore do not require expert consensus for interpretation in the same way, for example, diagnostic image analysis might.

    4. Adjudication Method for the Test Set:

    • None in the typical sense of expert adjudication. The tests (e.g., water leak, tensile strength, irritation) produce objective, quantitative results that are compared directly against the predefined acceptance criteria from the relevant ASTM and ISO standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic or screening devices where human interpretation is a key component, and the AI's effect on human performance is being evaluated. This submission is for examination gloves, which do not involve human interpretation of AI output.

    6. Standalone (Algorithm Only) Performance Study:

    • Not applicable. This device is a physical product (gloves), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant. The performance evaluation is based on the physical and chemical properties of the gloves themselves.

    7. Type of Ground Truth Used:

    • The ground truth is based on objective, measurable criteria defined by established international and national standards for medical gloves. Specifically:
      • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application.
      • ASTM D5151-06 (2011): Standard Test Method for Detection of Holes in Medical Gloves.
      • ASTM D6124-10 (2011): Standard Test Method for Residual Powder on Medical Gloves.
      • ISO 10993-10: Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization (for biocompatibility).

    8. Sample Size for the Training Set:

    • Not applicable. This is a physical product, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The manufacturing process of gloves is subject to Quality Management Systems (e.g., 21 CFR Part 820) which involve process control and quality assurance, not machine learning training.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, for the same reasons as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142283
    Date Cleared
    2015-01-16

    (154 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Non Sterile. Powder Free Nitrile Examination Gloves [{8603F (AD) Blue Color, 8611F (M) Blue Color, 8603F (Y4) Blue Color, 8613F Blue Color, 8603F (H) Blue Color, 8612F (C) Blue Color, 8604T Black Color and 8617F White Color] is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Non Sterile, Powder Free Nitrile Examination Gloves [(8603F (AD) Blue Color, 8611F (M)Blue Color, 8603F (Y4) Blue Color, 8613F Blue Color, 8603F (H) Blue Color, 8612F (C) Blue Color, 8604T Black Color and 8617F White Color] meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. This device is for over-the counter single use. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    The provided document describes the performance data for various models of Non-Sterile, Powder Free Nitrile Examination Gloves from YTY Industry (MANJUNG) SDN. BHD. based on ASTM and FDA standards. This is not an AI/ML device, therefore, the information requested in points 2, 3, 4, 5, 6, 7, 8, and 9 are not applicable.

    Here's the summary of the acceptance criteria and reported device performance:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicAcceptance Criteria (Standards)Reported Device Performance (Non-Sterile, Powder Free Nitrile Examination Gloves)
    DimensionASTM D6319-10Meets
    Physical PropertiesASTM D6319-10Meets
    Freedom from holesASTM D6319-10 (GII AQL = 2.5) & FDA 1000ml Water Leak TestMeets (0 holes found in sample size of 80 for all models, with accept criteria of 5 and reject criteria of 6)
    Length (mm) for Size MMin 230240-253 (range across all models)
    Palm Width (mm) for Size M95 ± 1095-99 (range across all models)
    Thickness (mm) Finger (single layer)Min 0.050.08-0.16 (range across all models)
    Thickness (mm) Palm (single layer)Min 0.050.06-0.12 (range across all models)
    Powder-freeASTM D6319-10Meets
    Moisture Content (%)Max 2.0%0.8% - 1.2% (Average across all models)
    Residual Powder (mg/glove)Max 2.0mg/glove (ASTM-D6124-10)0.10mg/l - 0.20mg/glove (range across all models; presented as a percentage in one instance for residual powder and mg/glove for moisture content)
    BiocompatibilityPrimary skin irritation: ISO 10993-10 (Non-Irritant)
    Dermal Sensitization: ISO 10993-10 (Non-sensitizer)Non-Irritant
    Non-sensitizer (for all models)

    Study Proving Device Meets Acceptance Criteria:

    The study summarized in the document is a series of non-clinical tests performed on the Non-Sterile, Powder Free Nitrile Examination Gloves (across various models and colors).

    • Standards Used: The performance of the devices was evaluated against the following standards:

      • ASTM Specification D6319-10, Standard Specification for Nitrile Gloves for Medical Application.
      • FDA 1000ml Water Leak Test (ASTM D5151-06 (2011) for watertightness with AQL GII 2.5).
      • ISO 10993-10 for Primary Skin Irritation and Dermal Sensitization (Biocompatibility).
      • ASTM-D6124-10 (Reapproved 2011) for Residual Powder.
    • Test Results: The table above details the specific performance results for each characteristic, demonstrating that all tested models of the gloves met or exceeded the specified acceptance criteria for dimensions, physical properties (tensile strength, ultimate elongation, before and after aging), freedom from holes, powder-free status (moisture content, residual powder), and biocompatibility.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    The document states for "Freedom from holes" that the sample size for the water leak test was 80 gloves per lot, with a lot size of 16,000 gloves. The data provenance is not explicitly stated in terms of country of origin of the data or whether it was retrospective or prospective. However, the manufacturer is stated as YTY INDUSTRY (MANJUNG) SDN. BHD. from Malaysia. The tests are non-clinical, implying they were likely conducted in a controlled lab environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    This information is not applicable as the device is not an AI/ML device. The ground truth is established by standardized testing protocols (ASTM, ISO, FDA tests), not by expert consensus in the human interpretation sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    This information is not applicable as the device is not an AI/ML device. The "ground truth" is determined by objective measurements and standardized testing procedures.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This information is not applicable as the device is not an AI/ML device and human "readers" or interpretations are not part of its function or evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This information is not applicable as the device is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" for the performance evaluation of these gloves is defined by objective measurements and adherence to international and national standards (ASTM, ISO, FDA specifications) for physical properties, barrier integrity, and biocompatibility. This includes measurements of length, width, thickness, tensile strength, elongation, freedom from holes (water leak test), moisture content, residual powder, and biological response (skin irritation and sensitization).

    8. The sample size for the training set
    This information is not applicable as the device is not an AI/ML device. There is no concept of a "training set" for physical products like examination gloves in this context.

    9. How the ground truth for the training set was established
    This information is not applicable as the device is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133123
    Date Cleared
    2014-05-30

    (242 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    YTY INDUSTRY (MANJUNG) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 3