Search Results
Found 12 results
510(k) Data Aggregation
(88 days)
Wuxi Jiajian Medical Instrument Co., Ltd.
Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Needle Stimulator is an electro-acupuncture device for acupuncture therapy, powered by 6 pieces of 1.5V batteries or AC 100-240V. It is composed of a console and 6 channels of electrode cables with alligator type connectors. The console has the operating controls including function knobs or buttons. Needle Stimulator does not equip with acupuncture needles. The practitioners should select legally marketed needles.
The provided text is an FDA 510(k) summary for a medical device called "Needle Stimulator." It details the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.
However, the provided text does not contain any information about a study that proves the device meets specific performance acceptance criteria for an AI/ML-based device. The information is primarily focused on electrical safety, essential performance, and electromagnetic compatibility tests as per general medical device standards, and not on the performance metrics of an AI system detecting or classifying something.
The recurring phrases like "Similar Note 1," "Similar Note 2," etc., refer to explanations for minor differences between the subject device and the predicate device in terms of hardware specifications and compliance with voluntary standards. They are not related to an AI/ML performance study.
Specifically, the section "6. Clinical Test" states, "Clinical data were not included in this submission." This strongly indicates that no clinical study was performed for this 510(k) submission, and therefore, no AI/ML performance data against ground truth or human readers would be available within this document.
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for an AI/ML-based device based on the information provided. The document describes a standard medical device (an electro-acupuncture stimulator), not an AI/ML product requiring the detailed performance study you've outlined.
If this was a misunderstanding and you expected a generic answer for an AI/ML device based on typical FDA requirements, please clarify. But based solely on the provided text, the requested information is absent.
Ask a specific question about this device
(315 days)
Wuxi Jiajian Medical Instrument Co., Ltd
Transcutaneous Electrical Nerve Stimulator (Model:TENS WMPS2-1) can be used for the symptomatic relief of chronic intractable pain, post traumatic pain adjunctive treatment, and post-surgical pain adjunctive treatment.
Transcutaneous Electrical Nerve Stimulator sends gentle electrical current to underlying nerves and muscle groups via electrodes applied onto the skin to relieve pain. The device has 16 programs (12 standard programs and 4 editable programs). It is a lithium battery-powered device comprising the electronic stimulatory module, the accessories of lead wires, the electrodes and the adapter. Two outlet sockets are used to connect skin electrodes by lead wires. The accessories of electrodes are 510(k) cleared devices (K192568). Size 50*50mm.
Here's a breakdown of the acceptance criteria and the study information based on the provided document:
1. Table of Acceptance Criteria and the Reported Device Performance:
The document describes the acceptance criteria in terms of compliance with various international standards for medical electrical equipment and a comparison to a predicate device. The reported performance is implicitly stated by asserting compliance with these standards and the substantial equivalence to the predicate.
Criteria Category | Acceptance Criteria (from referenced standards or predicate comparison) | Reported Device Performance (as stated in the document) |
---|---|---|
General Safety & Essential Performance | Compliance with Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (ANSI AAMI ES60601-1: 2005/(R) 2012 And A1: 2012, C1: 2009/(R) 2012 And A2: 2010/(R) 2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1: 2005, MOD)) | Bench tests demonstrated the device met all design specifications. Specifically, it passed ANSI AAMI ES60601-1. The differences in power source, output intensity control, average DC current at no load, number of output modes, channel output design and isolation methods, automatic no trip, user override control, net charge, maximum phase charge, maximum average current, maximum current density, and maximum average power density were evaluated and found to not raise any new safety or effectiveness issues, demonstrating compliance with the standard's safety limits. |
Nerve & Muscle Stimulator Specifics | Compliance with Medical Electrical Equipment - Part 2-10: Particular Requirements for the Basic Safety and Essential Performance Of Nerve and Muscle Stimulators (IEC 60601-2-10 Edition 2.1 2016-04) | The device passed IEC 60601-2-10 tests. All programs passed these tests. The differences in power source, output intensity control, average DC current at no load, number of output modes, channel output design and isolation methods, automatic no trip, user override control, net charge, maximum phase charge, maximum average current, maximum current density, and maximum average power density were evaluated and found to not raise any new safety or effectiveness issues, demonstrating compliance. |
Electromagnetic Compatibility (EMC) | Compliance with Medical electrical equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests (IEC 60601-1-2 Edition 4.0 2014-02) | The device passed IEC 60601-1-2. |
Home Healthcare Environment | Compliance with Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety and Essential Performance -- Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Equipment and Medical Electrical Systems Used in Home Healthcare Environment (IEC 60601-1-11 Edition 2.0 2015-01) | The device passed IEC 60601-1-11. |
Power Source(s) | Complies with IEC 60601-1 requirements. | DC3.7 V Li-battery powered. Considered "Similar" to predicate but complies with IEC 60601-1. Differences will not raise safety or effectiveness issues. |
Output Intensity Control | Complies with IEC 60601-1 requirements. | Current control. Considered "Similar" to predicate but complies with IEC 60601-1. Differences will not raise safety or effectiveness issues. |
Average DC current at no load | Similar to predicate device and passes AAMI / ANSI ES60601-1 and IEC 60601-2-10 tests. |
Ask a specific question about this device
(124 days)
Wuxi Jiajian Medical Instrument Co., Ltd
The Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, the TENS WMPS6-1 is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment in the management of post-surgical and post traumatic pain.
Transcutaneous Electrical Nerve Stimulator (model: TENS WMPS6-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. The device has 18 programs (13 standard programs and 5 editable programs), powered by 6 pieces of 1.5V batteries or AC 100-240V, comprising electronic stimulatory module and accessories of lead wires, electrodes. Six outlet sockets are used to connect skin electrodes by lead wires. The accessories of electrodes is 510(k) cleared device (K192568), Size: 50*50mm.
The provided text is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) device. It details the device's technical specifications and a comparison to a predicate device to establish substantial equivalence for regulatory clearance. This document does not describe an AI-powered medical device or an associated study for complex acceptance criteria like those involving AI performance metrics (e.g., sensitivity, specificity, MRMC studies, ground truth establishment by experts).
Therefore, I cannot extract the requested information regarding AI device acceptance criteria and study details from the provided text. The device is a physical medical device (TENS) and the "study" referred to in the document is a series of non-clinical bench tests to compare it to a predicate device, and to verify safety and performance standards.
Here's a breakdown of why I cannot fulfill your request based on the provided text:
- No AI Component: The device is a Transcutaneous Electrical Nerve Stimulator, a physical product that applies electrical current. There is no mention of any AI or machine learning component.
- No Diagnostic/Interpretive Task: The device is for pain relief, not for image analysis, diagnosis, risk prediction, or any other task that would typically involve acceptance criteria like sensitivity, specificity, or the need for ground truth established by experts.
- No Human-in-the-Loop Study: Since there's no AI component, there's no MRMC (Multi-Reader Multi-Case) study described for human readers improving with AI assistance.
- No Standalone Algorithm Performance: There is no algorithm to assess in a standalone capacity.
- No Training Set/Test Set (in the AI context): The concepts of training and test sets as they apply to machine learning models are not relevant to the clearance of this physical device. The "test set" here refers to the device itself undergoing bench tests.
- No Ground Truth Establishment by Experts: Ground truth for AI models is often established by expert consensus (e.g., radiologists reviewing images). For this TENS device, "ground truth" would refer to its physical and electrical properties meeting specified standards, validated through engineering tests, not human interpretation.
Summary of what the document DOES describe in terms of performance and comparison:
The document describes acceptance criteria in the context of substantial equivalence to an existing predicate device and compliance with voluntary standards for electrical medical equipment.
-
Acceptance Criteria (Implicit via Comparison to Predicate and Standards Compliance): The acceptance criteria are primarily demonstrated by showing that the new device's specifications (e.g., power source, output channels, waveform, voltage, current, safety features) are either identical, similar with acceptable differences (justified by safety and effectiveness analyses), or meet the requirements of relevant electrical safety and performance standards (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10). The reported device performance is presented as compliance with these standards and the specific electrical output values listed in the comparison table.
-
Study That "Proves" Compliance: The "study" here is a series of non-clinical bench tests.
- Sample Size: This is not a study with a "sample size" in the statistical sense of patient data. It refers to testing the physical device itself.
- Data Provenance: Not applicable as it's not a data-driven AI study. The tests are bench tests performed on the device.
- Experts and Ground Truth: Not applicable in the way you've defined them for an AI study. Electrical engineers and regulatory compliance experts would be involved in setting and verifying the test standards.
- Adjudication Method: Not applicable.
- MRMC Comparative Effectiveness Study: No.
- Standalone Performance: The "standalone" performance here refers to the device successfully passing all required electrical and safety tests independently (i.e., not reliant on human interpretation for its function).
- Type of Ground Truth: The "ground truth" is compliance with established engineering and medical device safety standards for TENS devices (e.g., maximum current density limits, electrical isolation, waveform characteristics).
- Training Set/Ground Truth for Training: Not applicable.
If you have a document describing an AI medical device, please provide that, and I will be able to answer your questions with the level of detail you are seeking.
Ask a specific question about this device
(333 days)
Wuxi Jiajian Medical Instrument Co., Ltd
Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Needle Stimulator is an electro-acupuncture device for acupuncture therapy powered by 6 pieces of 1.5V batteries or AC 100-240V. It is composed of a console and 6 channels of electrode cables with alligator type connectors. The console has the operating controls including function knobs or buttons. Needle Stimulator does not equip with acupuncture needles. The practitioners should select legally marketed needles.
Based on the provided text, the device in question is a "Needle Stimulator," an electro-acupuncture stimulator. The document is a 510(k) Summary, which aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device. This type of submission relies on comparisons to demonstrate safety and effectiveness rather than conducting extensive clinical studies typically associated with novel devices.
Therefore, the acceptance criteria and the "study" proving the device meets them are primarily based on equivalence to a predicate device through engineering and performance testing, rather than a clinical efficacy study with human subjects.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the parameters of the predicate device (K130768) and compliance with relevant safety standards. The reported device performance is presented as a comparison to these predicate specifications.
Parameter | Acceptance Criteria (Predicate Device K130768) | Reported Device Performance (Model CMNS6-1) | Reported Device Performance (Model CMNS6-2) | Remarks (from text) |
---|---|---|---|---|
Intended Use | Electro-acupuncture stimulator for qualified practitioners | Same | Same | Same |
Type of Use | Prescription use | Same | Same | Same |
Power Source(s) | DC 1.5Vx6 Type R14 | DC 1.5Vx6 Type R14 or AC 100-240V | DC 1.5Vx6 Type R14 or AC 100-240V | Note 1: AC power passed IEC 60601-1 and EMC tests |
Method of Line Current Isolation | Type BF | Type BF | Type BF | Same |
Patient Leakage Current (Normal) | 2µA | 2µA | 2µA | Same |
Patient Leakage Current (Single Fault) | ≤50µA | ≤50µA | ≤50µA | Same |
Avg DC current through electrodes (no pulses) | N/A | N/A | N/A | Same |
Number of Output Modes | 3 (continuous/interrupted/dense-disperse wave) | Same | Same | Same |
Number of Output channels | 6 (3 channels at most work together on single patient) | Same | Same | Same |
Synchronous or Alternating | Synchronous | Synchronous | Synchronous | Same |
Method of Channel Isolation | Transformer | Transformer | Transformer | Same |
Regulated Current or Voltage? | Voltage Control | Voltage Control | Voltage Control | Same |
Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Same |
Automatic Overload Trip? | No | No | No | Same |
Automatic No-Load Trip? | No | No | No | Same |
Automatic Shut Off? | Yes | Yes | Yes | Same |
User Override Control? | Yes | Yes | Yes | Same |
On/Off Status Indicator? | Yes | Yes | Yes | Same |
Low Battery Indicator? | Yes | Yes | Yes | Same |
Voltage/Current Level Indicator? | Yes | Yes | Yes | Same |
Timer Range (minutes) | 0-60min | 0-60min | 1-99min | Note 2 (for CMNS6-2): Difference insignificant |
Compliance with Voluntary Standards? | IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2 | Same | Same | Same |
Compliance with 21 CFR 898? | Yes | Yes | Yes | Same |
Weight (grams) | 740g | 740g | approx. 657g | Note 2 (for CMNS6-2): Difference insignificant |
Dimensions [W x H x D] | 23015555mm | 23015555mm | 23818475mm | Note 2 (for CMNS6-2): Difference insignificant |
Housing Materials & Construction | ABS; Injection molded | ABS; Injection molded | ABS; Injection molded | Same |
Waveform | Biphasic | Biphasic | Biphasic | Same |
Shape | Asymmetric biphasic square wave | Asymmetric biphasic square wave | Asymmetric biphasic square wave | Same |
Max Output Voltage (@500Ω) | 27V±10% | 27V±10% | 27V±10% | Same |
Max Output Voltage (@2kΩ) | 60.4V±10% | 60.4V±10% | 60.4V±10% | Same |
Max Output Voltage (@10kΩ) | 75V±10% | 75V±10% | 75V±10% | Same |
Max Output Current (@500Ω) | 54mA±10% | 54mA±10% | 54mA±10% | Same |
Max Output Current (@2kΩ) | 30.2mA±10% | 30.2mA±10% | 30.2mA±10% | Same |
Max Output Current (@10kΩ) | 7.5mA±10% | 7.5mA±10% | 7.5mA±10% | Same |
Pulse width (Positive) | 200µs±10% | 200µs±10% | 175µs±10% | Note 3 (for CMNS6-2): Passed IEC 60601-2-10 tests |
Pulse width (Negative) | 1030µs (5.15 x (+Phase)) | 1030µs (5.15 x (+Phase)) | 1051µs (6 x (+Phase)) | Note 3 (for CMNS6-2): Passed IEC 60601-2-10 tests |
Pulse Period (msec) | 10~1000ms | 10~1000ms | 10~1000ms | Same |
Max. pulse frequency (Rate) | 1~100Hz±10% | 1~100Hz±10% | 1~100Hz±10% | Same |
Net Charge (µC per pulse) | 0µC@500Ω, + and – pulses cancel | 0µC@500Ω, + and – pulses cancel | 0µC@500Ω, + and - pulses cancel | Same |
Maximum Phase Charge (µC) | 9.4µC @500Ω | 9.4µC @500Ω | 8.225µC @500Ω | Note 4 (for CMNS6-2): Complies with standards |
Maximum Average Current (mA) | 1.08mA @500Ω | 1.08mA @500Ω | 0.945mA @500Ω | Note 4 (for CMNS6-2): Better safety |
Maximum Current Density (mA/cm², r.m.s.) | 13.3mA/cm²@500Ω | 9.4mA/cm²@500Ω | 8.225mA/cm² @500Ω | Note 2 (for K130768 vs CMNS6-1), Note 4 (for K130768 vs CMNS6-2). Calculated by different needle surface areas, within safety limit. |
Maximum Average Power Density (W/cm²) | 0.2W/cm²@500Ω | 0.141W/cm²@500Ω | 0.1234W/cm²@500Ω | Note 2 (for K130768 vs CMNS6-1), Note 4 (for K130768 vs CMNS6-2). Calculated by different needle surface areas, within safety limit. |
Biocompatibility | ISO10993-5, ISO 10993-10 | Same | Same | Same |
Accessories | Lead wires, Alligator type connectors | Same | Same | Same |
2. Sample size used for the test set and the data provenance
The document does not describe a "test set" in the context of clinical data for performance evaluation. Instead, the "testing" refers to engineering and compliance testing of the device prototypes themselves. Therefore, the "sample size" is the manufactured device unit or units tested for compliance.
- Sample Size: Not applicable in the context of a clinical test set. The testing was performed on the device models themselves (CMNS6-1 and CMNS6-2).
- Data Provenance: Not specified as clinical data. The tests were likely conducted by the manufacturer in China, given their location. The studies are retrospective in the sense that they are proving equivalency to a device already on the market, but the testing on the new device would have been done prospectively for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the ground truth for this type of submission is established through:
- Engineering specifications and design requirements.
- Compliance with recognized national and international safety standards (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10).
- Demonstration of substantial equivalence to a legally marketed predicate device.
No human expert interpretation of clinical data (like radiology reads) is involved in establishing "ground truth" for this device's safety and performance claim.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable as there is no clinical test set requiring human adjudication. The "adjudication" is done through engineering measurements and conformity assessments to established standards and predicate specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is an electro-acupuncture stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or required for this type of device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical instrument, not a software algorithm. Its performance is inherent in its electrical and mechanical properties, not an algorithm's output.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this submission is:
- Engineering specifications and performance characteristics of the predicate device (K130768).
- International and national safety standards (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10), which define acceptable limits for various electrical and physical parameters, ensuring patient and user safety.
- The premise that if the new device operates within the same safety and performance parameters as a previously cleared predicate and complies with relevant standards, it is "safe and effective."
8. The sample size for the training set
This is not applicable. This is not an AI/ML device that requires a "training set."
9. How the ground truth for the training set was established
This is not applicable. As it is not an AI/ML device, there is no training set or associated ground truth establishment process.
Ask a specific question about this device
(263 days)
Wuxi Jiajian Medical Instrument Co., Ltd
The Transcutaneous Electrical Nerve Stimulator is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment of post surgical and post traumatic pain.
Transcutaneous Electrical Nerve Stimulator is Transcutaneous Electrical Nerve Stimulator for pain relief. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The device has standard programs and edit programs. It is a battery-powered portable device, comprising electronic stimulatory module and accessories of lead wires, electrodes and battery. Two outlet sockets are used to connect skin electrodes by lead wires. The accessories of electrodes is 510(k) cleared device (K192568), Size: 50*50mm.
This document is a 510(k) Premarket Notification for a Transcutaneous Electrical Nerve Stimulator (TENS) device. The purpose of this submission is to demonstrate that the new device, "TENStem eco basic," is substantially equivalent to a legally marketed predicate device (K112288), also a TENS device manufactured by the same company, Wuxi Jiajian Medical Instrument Co., Ltd.
Here's an analysis of the provided information regarding acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the parameters of the predicate device (K112288) and the relevant IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10). The reported device performance is presented as a comparison between the proposed device ("Modified Device") and the predicate device in the "Substantially Equivalent Comparison Conclusion" table.
Parameter | Acceptance Criteria (Predicate Device K112288) | Reported Device Performance (TENStem eco basic) | Remark (Meeting Criteria) |
---|---|---|---|
Intended Use | Same | Same | Met |
Type of Use | Prescription use | Prescription use | Met |
Power Source(s) | 9V Battery type 6F22 | 1.5Vx4 AAA alkaline battery | Similar (Note 1) |
Method of Line Current Isolation | NA | N/A | Met |
Patient Leakage Current - Normal Condition (μΑ) | 2μΑ | 2μΑ | Met |
Patient Leakage Current - Single Fault Condition (μΑ) | NA | NA | Met |
Average DC current through electrodes when device is on but no pulses are being applied (μΑ) |
Ask a specific question about this device
(504 days)
Wuxi Jiajian Medical Instrument Co., Ltd
The Self-adhesive electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.
The Self-adhesive electrode, lead wire type and snap button type, are non-sterile flexible structures, composed of materials commonly used in this application: First layer: Non-woven fabric tape Second layer: Polyurethane (PU) electrically conductive carbon cloth (Hebei Kangshengda Electronic Technology Co., Ltd) Third layer: Biocompatible conductive hydrogel coupling media (ValueTrode Carbon, K970426), which has passed the required skin sensitivity testing criteria as specified in ISO 10993-10 and cytotoxicity testing criteria as specified in ISO ISO 10993-5. The electrodes are designed for single patient / multiple application use. It can be used for low-frequency or medium-frequency nerve or muscle stimulators, as the conduction film adhered to body skin. There are nine shapes of round, rectangle, elliptical, calabash, butterfly, saddle shape, rhombus, meniscus shape and palm shape. For the electrical connection, Jiajian provides lead wire type and snap button type: Lead wire assembly - at least 40mm long wire with 2mm/2.5mm diameter female socket. Snap button assembly - with 2.5~5mm diameter male socket. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR part 898 by testing under ANSI/AAMI ES60601-1, subclause 8.5.2.3.
The provided document is a 510(k) summary for the Jiajian Self-Adhesive Electrode, which is a medical device. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving a device meets specific performance criteria through clinical studies in the same way a PMA (Premarket Approval) would require.
Therefore, many of the requested points, particularly those related to clinical studies, MRMC studies, training set details, or complex ground truth establishment, are not applicable to this type of regulatory submission (a 510(k)). The manufacturer's conclusion explicitly states: "Clinical data was not including in this submission."
Instead, the submission relies on bench testing and a comparison to an existing predicate device to demonstrate substantial equivalence, focusing on technical specifications, materials, and non-clinical performance.
Here's an analysis based on the available information:
1. Acceptance Criteria and Reported Device Performance (as demonstrated by testing and comparison):
The document doesn't present a formal table of "acceptance criteria" paired with "reported performance" in the way one might expect from a clinical study. Instead, it outlines the tests performed to show the device's adherence to relevant standards and its substantial equivalence to a predicate device. The implicit acceptance criterion is that the device performs as well as or similarly to the predicate device across key parameters and meets relevant safety standards.
Here's a table summarizing the relevant data points and how they compare to the predicate, as presented in the "Substantially Equivalent Comparison Conclusion" table:
Parameter | Acceptance Criteria (Implied / Predicate Value) | Reported New Device Performance (Wuxi Jiajian Self-Adhesive Electrode) | Remark |
---|---|---|---|
Indications for Use | Intended to be used to apply electrical stimulation current to the patient's skin (TENS and EMS). | Intended to be used to transmit electrical stimulation current to the patient's skin (TENS and EMS). | Same |
Target population | Single patient use and multiple application | Single patient use and multiple application | Same |
Type of use | OTC and Prescription use | OTC and Prescription use | Same |
Design (shapes) | Round, Rectangle, Oval, Gourd, Butterfly, Saddle (according to customized specification). | Round, Rectangle, Elliptical, Calabash, Butterfly, Saddle shape, Rhombus, Meniscus shape, Palm shape (according to customized specification). | Similar (More shapes) |
Design (connection) | Lead wire with female socket. | Lead wire with female socket, or snap button with male snap connector. | Similar (Additional) |
Materials | Non-woven fabric tape; Electrically conductive carbon cloth (ValueTrode Carbon, K970426); Biocompatible conductive hydrogel coupling media (Value Trode Carbon. K970426). | Non-woven fabric tape; Polyurethane (PU) electrically conductive carbon cloth (Hebei Kangshengda Electronic Technology Co., Ltd); Biocompatible conductive hydrogel coupling media (ValueTrode Carbon, K970426). | Similar |
Electrode Pad Size | Predicate range specified per shape (e.g., Round: Min.Ø20mm; Max.Ø80mm). | New Device range specified per shape (e.g., Round: Min.Ø20mm; Max.Ø85mm). | Similar (Slightly larger range for some shapes) |
Electrode Impedance | Predicate Round: 290~1000Ω | New Device Round: 302 | Similar (Within predicate range) |
Patient contact area | Predicate range specified per shape (e.g., Round: Min.3.14cm²; Max.50.24cm²). | New Device range specified per shape (e.g., Round: Min.3.14cm²; Max.56.716cm²). | Similar (Slightly larger range for some shapes) |
Max. Current Density | Predicate Round: 3.18mA/cm² (Ø20mm: 3.14cm²) | New Device Round: 3.18mA/cm² (Ø20mm: 3.14cm²). Other shapes also calculated, e.g., 0.24mA/cm² for 90x47mm). Max current density of all electrodes except round shapes does not exceed 2mA/cm² (IEC 60601-2-2). | Similar (Round matches, others are below the general limit) |
Hydrogel Thickness | 35mils ± 5mils (0.89mm ± 0.13mm) | 35mils ± 5mils (0.89mm ± 0.13mm) | Same |
Hydrogel pH | 4.2 ± 1.0 | 4.2 ± 1.0 | Same |
Hydrogel Volume Resistivity | 1500 ohm-cm max | 1500 ohm-cm max | Same |
Standards Met | ISO14971; ISO 13485; EN 980; ANSI/AAMI EC12; IEC 60601-1. | Lead wires test per 8.5.2.3 of AAMI/ANSI ES 60601-1; Impedance test, Conformability test and Fluid tolerance test per 201.15.101.6 and 201.15.101.7 of ANSI AAMI IEC 60601-2-2:2009; Impedance Test (Dispersion Test) per FDA's requirement; Peel strength test according to manufacturer's requirement; Shelf life test per ASTM F1980:2016; Reuse test per FDA's requirement. | Similar (New device tested to more specific relevant standards) |
Biocompatibility | ISO10993-5; ISO10993-10 | ISO10993-5; ISO10993-10 | Same |
Sterility Status | Non-sterile | Non-sterile | Same |
Electrical Safety | Lead wire meets IEC 60601-1 | Lead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1 | Same |
Other Performance | Good electrical conductivity, good adhesive property | Good electrical conductivity, good adhesive property | Same |
2. Sample size used for the test set and data provenance:
- Test Set Sample Size: For the bench tests (e.g., impedance, conformability, fluid tolerance, peel strength, shelf life, reuse), the document states: "All test samples were drawn from batches, with 3 samples for each specification."
- Data Provenance: The tests were conducted internally by Wuxi Jiajian Medical Instrument Co., Ltd. in China, as indicated by the submitter/manufacturer address. The data is retrospective, as it's part of a premarket submission.
3. Number of experts used to establish the ground truth for the test set and qualifications:
- This is Not Applicable (N/A). For a 510(k) submission based on bench testing and substantial equivalence, "ground truth" established by human experts (like radiologists for image analysis) is not relevant. Performance is measured against engineering specifications and industry standards.
4. Adjudication method for the test set:
- N/A. As there are no human experts establishing a "ground truth" for a medical imaging/AI dataset, no adjudication method is performed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- N/A. The document states "Clinical data was not including in this submission." MRMC studies apply to performance of diagnostic devices often involving human interpretation of results, which is not the case for this external electrode.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- N/A. This device is a passive electrode, not an algorithm or AI. Standalone performance refers to the accuracy of an AI algorithm.
7. The type of ground truth used:
- For the non-clinical tests, the "ground truth" or reference was based on engineering specifications, established industry standards (e.g., AAMI/ANSI ES 60601-1, IEC 60601-2-2), and the performance characteristics of the legally marketed predicate device (K090198). Essentially, the device's performance was compared against these defined benchmarks.
8. The sample size for the training set:
- N/A. This is not an AI/ML device requiring a training set.
9. How the ground truth for the training set was established:
- N/A. This is not an AI/ML device requiring a training set or its associated ground truth establishment.
Ask a specific question about this device
(311 days)
WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD.
The Jiajian® Pointoselect Digital is intended for use in the symptomatic relief of chronic intractable pain, postoperative pain, and acute pain.
Jiajian® Pointoselect Digital is a newly designed and easy to operate hand held stimulator. It is 9V battery powered device, with an LCD in the console. which can display the selected operation mode, intensity, frequency, pulse width, and battery level of the device, and so on. The device composes of a console, a hand held probe, a hand grip electrode, and lead wires. The hand probe can be used for treatment at individual sites. When using the hand probe, the patient must hold the hand grip electrode in order to electrically ground the device.
The provided text does not describe a study that proves the device meets specific acceptance criteria in the context of clinical performance or diagnostic accuracy. Instead, it details a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) device, the Jiajian® Pointoselect Digital, focusing on its substantial equivalence to a predicate device (Pointer Excel) based on technical characteristics and compliance with safety and biocompatibility standards.
Therefore, many of the requested categories (sample size, data provenance, ground truth, experts, adjudication, MRMC studies, standalone performance, training set) are not applicable or cannot be extracted from this document, as they relate to clinical studies or performance evaluations that are not present.
However, I can extract the acceptance criteria related to technical specifications and safety standards compliance and the reported performance based on the substantial equivalence discussion and safety and effectiveness section.
Here's the information that can be extracted based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The "acceptance criteria" here are defined by the comparison to the predicate device and compliance with established safety/biocompatibility standards.
Feature | Acceptance Criteria (Predicate Device K060517 Performance) | Reported Device Performance (Jiajian® Pointoselect Digital) | Notes |
---|---|---|---|
Intended Use | Symptomatic relief of chronic intractable pain, postoperative pain, and acute pain. | Symptomatic relief of chronic intractable pain, postoperative pain, and acute pain. | Met: Identical intended use. |
Power Source | DC 9V battery, Type 6F22 | 9 Volt battery type 6F22 (Carbon-zinc), or 9 Volt battery type 6LR61 (Alkaline) | Met: Similar DC battery power source. |
Waveform | Biphasic | Biphasic | Met: Identical waveform. |
Shape | Asymmetric biphasic square wave | Asymmetric biphasic square wave | Met: Identical wave shape. |
Net Charge | 0μC@500Ω | 0μC @500Ω, + and - pulses cancel | Met: Identical net charge. |
Frequency Range | 1-16Hz | 2~18Hz | Met: Similar frequency range, subject to "differences would not raise new safety concerns." |
Maximum Output Voltage | 11V ±15% @500Ω | 6.6V ±15% @500Ω | Different: Lower for new device. Document asserts "differences would not raise new safety concerns." |
Maximum Output Current | 22mA ±15% @500Ω | 13.2mA ±15% @500Ω | Different: Lower for new device. Document asserts "differences would not raise new safety concerns." |
Pulse Duration | 220μS | 60-120μS | Different: Shorter for new device. Document asserts "differences would not raise new safety concerns." |
Maximum Phase Charge | 4.8μC@500Ω | 2.4μC@500Ω | Different: Lower for new device. Document asserts "differences would not raise new safety concerns." |
Maximum Current Density (r.m.s.) | 10.35mA/cm²@500Ω | 12.08mA/cm²@500Ω | Similar: Slightly higher for new device. Document states "similar maximum output current density." |
Safety Standard Compliance | IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2 (for predicate) | IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2 (for new device) | Met: New device was tested and found to meet these safety standards. |
Biocompatibility Standards | ISO 10993-5, ISO 10993-10 (for predicate) | ISO 10993-5, ISO 10993-10 (for new device, specifically pointer probe) | Met: New device's pointer probe was tested and found to comply with these biocompatibility standards. Lead wires also tested to IEC 60601-1. |
Battery Life Impact | Not explicitly stated | Stimulus parameters not significantly affected (less than ±10%) with reduced battery level. | Met: Specific testing for performance under reduced battery conditions. |
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable. The document describes a comparison of technical specifications and compliance with standards, not a clinical trial or performance study involving patient data.
- Data Provenance: Not applicable, as no human subject data (retrospective or prospective) is discussed. The device is manufactured in Wuxi, Jiangsu, China.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. There is no mention of "ground truth" established by experts in the context of clinical performance or diagnostic accuracy. The assessment is based on engineering specifications and adherence to standards.
4. Adjudication method for the test set
- Not applicable. No adjudication method is described as there is no clinical test set requiring it.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical medical device (TENS) for pain relief, not an algorithm or AI system.
7. The type of ground truth used
- Not applicable. The document assesses the device against technical specifications and safety/biocompatibility standards, not against a "ground truth" derived from clinical outcomes or expert consensus.
8. The sample size for the training set
- Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
- Not applicable. No training set is involved.
Ask a specific question about this device
(149 days)
WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD.
Jiajian® CMN Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Jiajian® CMN Stimulator is an electro-acupuncture device for acupuncture therapy, powered by 6 pieces of 1.5V batteries. It is composed of a console and 6 channels of electrode cables with alligator type connectors. Only 3 channels at most could work together on single patient. The console has the operating elements of Wave choosing knob, Frequency adjust knob, Intensity adjust knobs, and Timer. Jiajian® CMN Stimulator does not equip with acupuncture needles. The practitioners should select 510(k) cleared needles (with minimum diameter of 0.30mm and insertion depth of 15mm) for use.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Jiajian® CMN Stimulator:
Overview:
The document provided is a 510(k) Summary for the Jiajian® CMN Stimulator, an electro-acupuncture device. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (ES-130, K081943) by comparing technical characteristics and compliance with recognized safety standards. The document does not describe a clinical study for performance or effectiveness in the way an AI/ML device submission would. Instead, the "study" demonstrating it meets acceptance criteria refers to technical testing against safety standards and a comparison of its electrical output parameters to those of the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) for an electro-acupuncture stimulator, the "acceptance criteria" are primarily established by compliance with recognized electrical safety and EMC standards, and by demonstrating that its operational parameters are either within established safe limits or are substantially equivalent to a legally marketed predicate device without raising new safety or effectiveness concerns.
Feature / Criteria (Derived from Substantial Equivalence and Standards) | Acceptance Criteria (Predicate Device K081943 or Safety Standards) | Reported Device Performance (New Device: Jiajian® CMN Stimulator) | Assessment |
---|---|---|---|
Intended Use | For use in the practice of acupuncture by qualified practitioners. | Same as predicate. | Met |
Power Source | DC 9V battery, Type 6F22 | DC 1.5Vx6, Type R14 (different but both DC) | Met (Equivalent) |
Method of Line Current Isolation | N/A for DC current | N/A for DC current | Met |
Patient Leakage Current (Normal Condition) | Not Stated (Predicate) | 2μΑ | Met (Within implicit safety limits) |
Patient Leakage Current (Single Fault Condition) | Not Stated (Predicate) | ≤50μΑ | Met (Within implicit safety limits) |
Waveform | Biphasic Asymmetric biphasic square wave | Biphasic Asymmetric biphasic square wave | Met |
Maximum Output Voltage @500Ω | 18.0V ±15% | 24.4V ±10% | Differs (Higher), justified by overall charge and safety. |
Maximum Output Current @500Ω | 36.0mA ±15% | 48.8mA ±10% | Differs (Higher), justified by overall charge and safety. |
Pulse Duration | 100μs | 200μs ±10% | Differs (Higher), justified by overall charge and safety. |
Frequency | 1~500Hz | 1~100Hz | Differs (Lower upper limit). |
Net Charge @500Ω | 0µC | 0µC (+ and - pulses cancel) | Met |
Maximum Phase Charge, (µC) | 7.2µC | 17.8µC | Differs (Higher), justified as within safety limit. |
Maximum Current Density (r.m.s.) | 25.5mA/cm2 | 12.6mA/cm2 | Met (Lower, thus safer). |
Maximum Average Power Density | 0.24W/cm2 | 0.18W/cm2 | Met (Lower, thus safer). |
EMC Compliance | IEC 60601-1-2 | IEC 60601-1-2 | Met |
Electrical Safety (General) | IEC 60601-1 | IEC 60601-1 | Met |
Electrical Safety (Nerve/Muscle Stimulators) | IEC 60601-2-10 | IEC 60601-2-10 | Met |
Stimulus Parameters Stability (Reduced Battery) | Not explicitly stated for predicate; implied stability. | Not significantly affected (less than ±10%) | Met (Specific test performed for new device.) |
Note on "Acceptance Criteria" for this device: For this type of electro-acupuncture stimulator pursuing 510(k) clearance, acceptance criteria are generally met by (1) demonstrating compliance with recognized safety standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2) and (2) showing that differences in technical specifications from the predicate device do not raise new questions of safety or effectiveness. There isn't a "performance" criterion in the sense of a clinical outcome effectiveness measure for these devices in a 510(k) context.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable in the context of clinical trial/performance data. The "test set" here refers to the device itself being subjected to electrical and mechanical testing for safety standards compliance. There is no mention of human subject data or a test set of data points derived from patients.
- Data Provenance: The 'data' proving compliance comes from internal testing by the manufacturer (Wuxi Jiajian Medical Instrument Co., Ltd), likely performed in China, against the specified international standards. This is not derived from patient studies, retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable. There is no "ground truth" established by experts in the context of a clinical test set for this 510(k) submission. The evaluation is technical and regulatory, centered on compliance with engineering standards and comparison to a predicate device.
4. Adjudication Method for the Test Set
Not applicable. There is no expert adjudication process described for the technical testing of the device for safety standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. An MRMC comparative effectiveness study was not performed. This type of study is relevant for diagnostic imaging devices or devices where human interpretation/performance is augmented by AI. This device is a therapeutic stimulator, and its clearance relies on technical safety and equivalence to an existing device, not a comparison of human interpretation.
6. Standalone (Algorithm Only) Performance Study
No. This is not an AI/ML device, so a standalone algorithm performance study is not applicable. The "standalone" performance here would refer to the device operating according to its technical specifications.
7. Type of Ground Truth Used
The "ground truth" for this submission are the established international safety standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2) and the technical specifications of the cleared predicate device (ES-130, K081943). The device's performance is compared against these benchmarks to establish substantial equivalence.
8. Sample Size for the Training Set
Not applicable. This device is not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no AI/ML training set.
Ask a specific question about this device
(210 days)
WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD.
Jiajian® Electro-acupuncture Stimulator, Model WQ-10D1 and WQ-6F are electro-acupuncture stimulator devices, which are indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Jiajian® Electro-acupuncture Stimulator, Model WQ-10D1, is a battery powered instrument with 3 channel outputs for acupuncture needle stimulation. The predicate device is ES-130, K081943. The output frequency and intensity for stimulation are adjustable. The stimulation time can be set. The 3 channels can output stimulation simultaneously. The device is powered by DC 9V battery, Type 6F22. When the battery has insufficient voltage (
The provided document describes the Wuxi Jiajian Medical Instrument Co., Ltd.'s Electro-acupuncture Stimulators (Model WQ-10D1 and WQ-6F) and its 510(k) summary for substantial equivalence to a predicate device. This submission focuses on comparing the new device's technical specifications and safety standards to those of the predicate device, not on clinical performance or specific acceptance criteria met through clinical studies with human subjects.
Therefore, many of the requested categories related to clinical study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, training set) are not applicable to this type of regulatory submission, which primarily relies on engineering performance testing and comparison to a legally marketed predicate device.
However, the document does list product specifications and safety standard compliance, which could be considered criteria for acceptance in a technical and regulatory context.
Here's the information that can be extracted from the provided text according to your request categories:
1. A table of acceptance criteria and the reported device performance
Since this is a 510(k) submission for substantial equivalence based on technical characteristics and safety standards, the "acceptance criteria" are generally that the new device's parameters and safety performance are equivalent to or within the acceptable limits established by the predicate device and relevant standards. The "reported device performance" refers to the new device's measured specifications. The tables below synthesize this information for key parameters, highlighting where differences exist and how they are addressed in the substantial equivalence discussion.
Jiajian® Electro-acupuncture Stimulator, Model WQ-10D1 vs. Predicate Device ES-130 (K081943)
Parameter | Acceptance Criteria (Predicate Device ES-130) | Reported Device Performance (Jiajian® WQ-10D1) | Substantial Equivalence Discussion / Justification |
---|---|---|---|
Intended Use | Practice of acupuncture by qualified practitioners | Practice of acupuncture by qualified practitioners | Similar. Same intended use. |
Power Source | DC 9V battery, Type 6F22 | DC 9V battery, Type 6F22 | Similar. |
Compliance with Voluntary Standards | IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2 | IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2 | Similar. Both comply with the same safety standards. |
Compliance with 21 CFR 898 | Yes | Yes | Similar. |
Waveform | Biphasic, Asymmetric biphasic square wave | Biphasic, Asymmetric biphasic square wave | Similar. |
Net Charge (per pulse @500Ω) | 0μC | 0μC | Similar. Both devices have zero net charge, ensuring no DC current accumulation. |
Max Output Voltage (Low Intensity @500Ω) | 18V ± 15% | 11V ± 15% | Difference. New device has lower max output voltage. Justified by physiological effectiveness depending primarily on delivered charge, and the new device's maximum phase charge is slightly higher while within safety limits. |
Max Output Current (Low Intensity @500Ω) | 36.0mA ± 15% | 22mA ± 15% | Difference. New device has lower max output current. Justified by physiological effectiveness depending primarily on delivered charge, and the new device's maximum phase charge is slightly higher while within safety limits. |
Pulse Duration (Positive) | 100μS | 480 ± 10% μS | Difference. New device has higher pulse width. Justified by physiological effectiveness depending primarily on delivered charge, and the new device's maximum phase charge is slightly higher while within safety limits. Duty cycle is similar. Does not raise new safety or effectiveness concerns. |
Frequency (Hz) | 1~500Hz | 0~100Hz | Difference. New device has a lower frequency range. Justified by physiological effectiveness depending primarily on delivered charge, and the new device's maximum phase charge is slightly higher while within safety limits. Duty cycle is similar. Does not raise new safety or effectiveness concerns. |
Maximum Phase Charge (μC) | 7.2μC | 17.2μC | Difference. New device has a higher maximum phase charge. Justified that this enables effectiveness and is within safety limits. Addressed as not raising new safety or effectiveness concerns. |
Max Current Density (mA/cm², r.m.s.) | 25.5mA/cm² | 12.2mA/cm² | Difference. New device has lower current density. Justified as being within safety limits per IEC 60601-2-10 and FDA guidance. |
Max Average Power Density (W/cm²) | 0.24W/cm² | 0.09W/cm² | Difference. New device has lower average power density. Justified as being within safety limits per IEC 60601-2-10 and FDA guidance. |
Burst Mode | N/A (continuous train of impulses) | Present (various intermittent, rise-fall, sawtooth waves) | Difference. The new device has burst modes. Justified that the basic characteristics of burst mode are consistent with continuous wave, so this difference does not raise new types of safety or effectiveness questions. |
CA-AM knob | Constant amplitude only | CA-AM knob (Constant Amplitude and Amplitude Modulated wave) | Difference. The new device offers amplitude modulation. Justified that physiological effectiveness can also be achieved through modulated amplitude, not just constant pulse trains, thus not raising new types of safety or effectiveness questions. |
Battery Level Indicator | Check button | Auto shut off |
Ask a specific question about this device
(267 days)
WUXI JIAJIAN MEDICAL INSTRUMENT CO., LTD.
Jiajian® TENS is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment in the management of post surgical and post traumatic pain.
Jiajian® TENS, is Transcutaneous Electrical Nerve Stimulator for pain relief. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. It is a battery-powered portable device, comprising electronic stimulatory module and accessories of lead wires and 9 volt type 6F22 battery. When using this device, the physician should select and use 510(k) cleared electrodes. The area of electrodes must be larger than 8cm2. The electronic stimulatory module has the operating elements of (1) Display screen, 2 Menu keys, 3 Modification keys, 4 On/Off key, 5 Battery compartment and (6) Outlet socket. The display screen can show (a) battery power, (b) selected program, (c) lasting time or left time of a program, (d) current intensity for each channel, (e) program phase and (f) locking state. The menu key "P" is for selecting standard program or user-program, and for locking; the menu key "E" is for editing program when the device is not being locked. The modification key "3A" and "3B" are for intensity level adjustment during stimulation. The outlet socket is used to connect skin electrodes by lead wires. The device has 12 selectable programs, which can be grouped into 4 output modes, i.e. Normal mode (P1, P2, P3, P4, P5, P6), Burst mode (P7, P12), Rate & width modulation (P8), and Intensity modulation (P9, P10, P11).
The provided document is a 510(k) summary for the Jiajian® TENS device and primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study of the device's performance against specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment is not available in the provided text.
However, based on the information provided, here's what can be extracted and what is explicitly missing:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in terms of clinical performance metrics (e.g., pain reduction scores, patient satisfaction). Instead, it focuses on demonstrating technical equivalence and compliance with safety standards as a basis for being "substantially equivalent" to a predicate device.
The table below summarizes the comparison of technical specifications between the new device and the predicate device. While not acceptance criteria in the sense of clinical study endpoints, these indicate the technical specifications that the new device meets and how they compare to the predicate.
Parameter | Acceptance Criteria (from predicate) | Reported Device Performance (Jiajian® TENS) |
---|---|---|
Functional/Safety Characteristics | ||
Intended Use | For symptomatic relief and management of chronic intractable pain, adjunctive treatment in the management of post surgical and post traumatic pain | For symptomatic relief of chronic intractable pain and adjunctive treatment in the management of post surgical and post traumatic pain. |
Power Source | 9V Battery type 6F22 | 9V Battery type 6F22 |
Average DC current through electrodes (no pulses) |
Ask a specific question about this device
Page 1 of 2