Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    Wuxi Hisky Medical Technologies Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Confocal Microprobe Imaging System can enter the human body cavity or surgical channel through an endoscope, allowing confocal laser imaging of the microstructure of tissues, including but not limited to the identification of cells, vessels and their organization or architecture.

    Device Description

    The working principle of Confocal Microprobe Imaging System is based on probe-based confocal laser endomicroscopy technology (pCLE). The system combines confocal technology and fiber beam imaging technology. The fiber Optic Microprobe can enter the human cavity through the endoscopic working channel and contact the tissue cells through the object lens at the front end of the Fiber Optic Microprobe. The imaging principle of the device is as follows:

    The laser scanning beam emitted by the laser in the Laser Scanning System forms a light source through the grating pinhole and is transmitted to the focal plane of the fluorescent labeled tissue cells through the Fiber Optic Microprobe. The fluorescent substance in the measured tissue cell emits fluorescence under the excitation of the laser. The fluorescence signal is collected by Fiber Optic Microprobe front end object lens and transmitted through the fiber beam microprobe to the detecting hole and then is transmitted to the photomultiplier tube (PMT) of the Photoelectric detector and then to the host for signal analysis and processing. Finally, the image is formed on the computer monitoring screen after software processing.

    Light emitted at the top and bottom of the focal plane of the tested tissue produces a large diameter spot (much larger than that of the detecting hole) at the detecting hole, thus only a very small part of the light can be received by the detector through the detecting hole. Moreover, the larger the distance from the focal plane of the object lens, the larger the diffuse spot produced by the non-targeted tissue in the detecting hole and the lesser energy passes through the detecting hole (from 10% to 1%, slowly close to 0%), thus the weaker the unwanted signal is generated on the detector, and smaller the impact is caused by non-targeted tissue. Because confocal microscopy only images the focal plane of the target tissue, it effectively avoids the interference of diffracting light and scattered light, so that it has a higher resolution than ordinary microscopy and has been widely used in biology.

    AI/ML Overview

    It appears that the provided FDA 510(k) Clearance Letter does not contain detailed information about a clinical study involving human readers or a specific "acceptance criteria" table with reported performance metrics for an AI component.

    The document discusses the Confocal Microprobe Imaging System, which is a hardware device for imaging tissues. While it mentions "software" and "cybersecurity," these sections focus on general software validation and cybersecurity considerations, not the performance evaluation of an AI algorithm designed to interpret or analyze the images beyond the device's basic function.

    The "Performance Testing" section states that "Performance Verification Test has been conducted in accordance with the internal performance requirements stated in the Performance Validation Scheme (HRD0003932 & HRD0004124)" and lists technical performance requirements like "Field of view, Horizontal resolution, Depth of Observation, Frame rate." These relate to the imaging system's hardware performance, not an AI's diagnostic accuracy.

    Therefore,Based on the provided FDA 510(k) clearance letter, I cannot fulfill your request for detailed information regarding acceptance criteria for an AI component and the study that proves the device meets those criteria. The letter primarily addresses the clearance of a Confocal Microprobe Imaging System (hardware), focusing on its substantial equivalence to predicate devices based on technological characteristics, biocompatibility, reprocessing, electrical safety, and general software/cybersecurity validation.

    There is no mention of an AI-specific component, its performance criteria, or any clinical studies (e.g., MRMC studies) pertaining to AI algorithm performance within this document. The "Performance Testing" section refers to the optical and functional performance of the imaging system itself, not the diagnostic performance of an AI that might interpret the images generated.

    If such an AI component exists, its performance evaluation would typically be described in a separate section with specific metrics like sensitivity, specificity, or AUC, and details about the study design (test set, ground truth, expert adjudication, etc.). This information is absent in the provided text.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Wuxi Hisky Medical Technologies Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shear Wave Quantificational Ultrasound Diagnostic System, Transient Elastography based device, is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as Ultrasound Attenuation Parameter (UAP) in internal structures of the body.
    Shear Wave Quantificational Ultrasound Diagnostic System is indicated as a non-invasive aid for the clinical management, diagnosis, and monitoring of adult and pediatric patients with confirmed or suspected liver disease, as part of an overall assessment of the liver. Results in the pediatric population should be interpreted while considering the clinical condition and the overall patient profile.
    Shear Wave Quantificational Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal, including locating of the liver.
    The system must be operated by qualified and appropriately trained healthcare professionals in a professional healthcare facility environment.

    Device Description

    The proposed device is a general purpose, mobile, software-controlled, diagnostic ultrasound system. It consists of a fibrosis scanning probe, an imaging probe, a main unit, a probe holder, an AC/DC adapter and a foot switch. The fibrosis scanning probe is a single element probe used for elasticity measurement, while the imaging probe is a convex probe used for ultrasound imaging. The system uses transient elastography to measure shear wave speed non-invasively and estimates of tissue stiffness as well as Ultrasound Attenuation Parameter (UAP) in internal structures of the body. A mechanical vibrator in the fibrosis scanning probe produces low-frequency shear waves at 50 Hz that travel through the skin and intercostal space into the liver. The propagation speed of the shear wave is captured using ultrasound at 2.5 MHz. Under imaging mode, the system acquires and displays ultrasound images in B,M,CFM,PWD modes. The system uses imaging probe with nominal Frequency of 2.5-4.5 MHZ on abdomen for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. The ultrasonic imaging also helps to find a proper location for the transient elastography examination.
    There are a total of 9 models in this product, all of which share the same intended use, physical design and principle of technology. The only differences among these models are software functions configuration and availability of imaging probe.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided FDA 510(k) Clearance Letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device seem to be implicitly established by demonstrating comparable performance to the predicate device, K200136. The provided document focuses on reporting the device's performance metrics and comparing them to those of the predicate.

    Metric (Acceptance Criteria implied by predicate performance)Reported Device Performance (Subject Device)Predicate Device Performance (K200136)
    Liver Stiffness
    TE Bias Liver Stiffness(-2.1%) – (3.5%)FT9000: (-4.7%) - (2.4%)
    Mini800: (-5.3%) – (1.2%)
    FT100: (-2.1%) – (3.5%)
    TE Precision Liver Stiffness(0.0%) – (3.8%)FT9000: (0.0%) – (1.6%)
    Mini800: (0.0%) – (1.8%)
    FT100: (0.0%) – (3.8%)
    UAP (Ultrasound Attenuation Parameter)
    UAP Bias(-1.6%) – (6.5%)FT9000: (-3.3%) – (2.0%)
    Mini800: (-6.9%) - (4.8%)
    FT100: (-1.6%) – (6.5%)
    UAP Precision(0.3%) – (2.0%)FT9000: (0.2%) – (1.5%)
    Mini800: (0.2%) – (2.0%)
    FT100: (0.3%) – (2.0%)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the sample size used for the test set or the data provenance (e.g., country of origin of the data, retrospective or prospective). It simply states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information regarding the number of experts used to establish ground truth or their qualifications. The non-clinical testing appears to be based on technical specifications and comparisons rather than expert-derived ground truth for patient data.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method for a test set. This type of information is typically relevant for studies involving human interpretation or subjective assessment, which is not clearly described as part of these non-clinical tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or, at least, is not reported in this document. The submission explicitly states "Clinical investigation: Not applicable," which further confirms the absence of such a study.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    The device, a "Shear Wave Quantificational Ultrasound Diagnostic System," is an imaging system that provides measurements (shear wave speed, tissue stiffness, UAP) and imaging capabilities. Its performance metrics (Bias and Precision for liver stiffness and UAP) are reported as technical measurements, suggesting a standalone assessment of the algorithm/device's measurement capabilities. The context of the "non-clinical tests" reinforces this. However, it's important to note that while the measurements are standalone, a human operator is still required to acquire the data with the ultrasound probe.

    7. Type of Ground Truth Used

    Based on the nature of the reported performance metrics (Bias and Precision for physical measurements like liver stiffness and UAP) and the statement "Non-clinical tests were conducted to verify that the proposed device met all design specifications," the ground truth was likely established using phantom studies or other controlled technical measurements with known stiffness and attenuation properties, rather than expert consensus, pathology, or outcomes data from human subjects.

    8. Sample Size for the Training Set

    The document does not provide information about the sample size for a training set. Given that "Clinical investigation: Not applicable" is stated, and the focus is on non-clinical testing and comparison to a predicate, it's possible that the device's algorithms were developed and validated internally using proprietary data or phantom studies, but these details are not disclosed in the clearance letter.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for a training set was established. As mentioned in point 8, the "Not applicable" for clinical investigation suggests that the development and validation of the device's algorithms might have relied on non-clinical data, potentially using phantoms with known properties.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200136
    Date Cleared
    2020-03-17

    (56 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Wuxi Hisky Medical Technologies Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shear Wave Quantificational Ultrasound Diagnostic System (Models: FT9000, FT100 and Mini800), Transient Elastography based device, is intended to provide 50Hz shear wave speed measurements and estimates of tissue stiffness as well as Ultrasound Attenuation Parameter (UAP ) in internal structures of the body.

    Shear Wave Quantificational Ultrasound Diagnostic System (Models: FT9000, FT100 and Mini800), is indicated for noninvasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as Ultrasound Attenuation Parameter ( UAP ).

    The shear wave speed and stiffness, and UAP may be used as an aid to diagnosis and monitoring of patients with liver disease, as part of an overall assessment of the liver.

    Shear Wave Quantificational Ultrasound Diagnostic System (Models: FT9000), is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal, including location of the liver.

    Device Description

    The Shear Wave Quantificational Ultrasound Diagnostic System, Models: FT9000, FT100 and Mini800, iLivTouch brand, is a general purpose, mobile, software-controlled, diagnostic ultrasound system. FT9000 is equipped with two probes, a fibrosis scanning probe used in elastography mode and an imaging probe used in imaging mode. The fibrosis scanning probe is used for elasticity measurement while the imaging probe is a convex probe used for ultrasound imaging. But the model FT100 and Mini800 is equipped with only one fibrosis scanning probe.FT100 and Mini800 have the same appearance structure and hardware, but the software is different.

    Under elastography mode, the system uses transient elastography to measure shear wave speed non-invasively and estimates of tissue stiffness as well as Ultrasound Attenuation Parameter (UAP) in internal structures of the body. A mechanical vibrator produces low-amplitude shear waves at 50 Hz that travel through the skin and intercostal space into the liver. The propagation speed of the shear wave is measured using ultrasound at 2.5 MHz.

    Under imaging mode, the system acquires and displays ultrasound images in B, B/B, B/C (CFM), B/C/D (CPWD) modes. The system uses convex array probe with a frequency range of 2.1MHz to 5 MHz on abdomen for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. The ultrasonic imaging also helps to find a proper location for the transient elastography examination.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study information for the Shear Wave Quantificational Ultrasound Diagnostic System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary directly compares the proposed device to predicate devices. The "acceptance criteria" are implied by the similarities to the predicate devices and the performance characteristics stated. The reported device performance is given in terms of bias and precision for shear wave speed, stiffness, and UAP measurements.

    CategoryAcceptance Criteria (Implied by Predicate)Reported Device Performance (Proposed Device)
    Shear Wave Speed / Stiffness Measurement
    Display RangeShear wave speed (0.8-5.0 m/s), Stiffness (2.0-75 kPa), Interquartile range (IQR) and IQR/median ratio (Predicate: FibroScan® 502 Touch)FT9000: Shear wave speed (0.8-5.2 m/s), Stiffness (2.0-80 kPa), Interquartile range (IQR) and IQR/median ratio
    FT100 & Mini800: Similar to FT9000, but exact ranges not explicitly separated for TE display.
    Bias(-14.3%) - (3.6%) (Predicate: FibroScan® 502 Touch)FT9000: (-4.7%) - (2.4%)
    FT100: (-2.1%) - (3.5%)
    Mini800: (-5.3%) - (1.2%)
    Precision(0.2%) - (1.9%) (Predicate: FibroScan® 502 Touch)FT9000: (0.0%) - (1.6%)
    FT100: (0.0%) – (3.8%)
    Mini800: (0.0%) - (1.8%)
    UAP Display RangeCAP (100-400 dB/m), Interquartile range (IQR) and IQR/median ratio (Predicate: FibroScan® 502 Touch)UAP (90-450 dB/m), Interquartile range (IQR) and IQR/median ratio
    UAP Bias(0.0%) - (10.0%) (Predicate: FibroScan® 502 Touch)FT9000: (-3.3%) - (2.0%)
    FT100: (-1.6%) – (6.5%)
    Mini800: (-6.9%) - (4.8%)
    UAP Precision(0.0%) - (1.0%) (Predicate: FibroScan® 502 Touch)FT9000: (0.2%) - (1.5%)
    FT100: (0.3%) - (2.0%)
    Mini800: (0.2%) – (2.0%)

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission." This means there is no test set of human data for the proposed device analyzed in this submission. The performance metrics (bias and precision) likely refer to phantom studies or bench testing rather than clinical data. The document does not provide details on the sample size for these non-clinical tests or their provenance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since no clinical study was included, there were no experts used to establish ground truth for a human test set. For the phantom/bench testing, the ground truth would typically be established by the known physical properties of the phantom materials and calibrated measurement equipment. No specific details about such experts or their qualifications are provided for non-clinical testing.

    4. Adjudication Method for the Test Set

    As there was no human test set, no adjudication method was applicable or described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done as "no clinical study is included in this submission." Therefore, no effect size of human readers improving with or without AI assistance is provided.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance metrics (bias and precision) presented in the tables are indicative of standalone performance (algorithm only) as applied to non-clinical tests (e.g., phantoms). The document states the proposed devices passed "related performance test (eg UD2 test)" and "measurement accuracy by accuracy testing and software validation." These imply standalone testing, but no specific study details are provided beyond these results.

    7. The Type of Ground Truth Used

    For the performance metrics listed, the ground truth was based on known physical properties of phantoms or calibrated measurement systems used in non-clinical performance and safety testing (e.g., UD2 test).

    8. The Sample Size for the Training Set

    The document does not provide any information regarding a training set sample size. This suggests that the device's algorithms were either established using pre-existing knowledge (e.g., physics-based models for shear wave propagation) or that any internal development/training data was not part of this 510(k) submission, which focused on demonstrating substantial equivalence through technical characteristics and non-clinical performance to existing predicate devices.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned, no information on how its ground truth was established is provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173595
    Date Cleared
    2018-06-22

    (213 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Wuxi Hisky Medical Technologies Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended to measure the speed of 50Hz shear wave with 2.5MHz ultrasound wave in the liver. The shear wave speed may be used as an aid to clinical management of patients with liver disease.

    Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal.

    Device Description

    The Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. The system is equipped with two probes, a fibrosis scanning probe used in elastography mode and an imaging probe used in imaging mode. The fibrosis scanning probe is used for elasticity measurement while the imaging probe is a convex probe used for ultrasound imaging.

    Under elastography mode, the system uses transient elastography to measure shear wave speed non-invasively and estimate tissue stiffness in liver. A mechanical vibrator produces low-amplitude shear waves at 50 Hz that travel through the skin and intercostal space into the liver. The propagation speed of the shear wave is measured using ultrasound at 2.5 MHz.

    Under imaging mode, the system acquires and displays ultrasound images in B, B/B, B/C (CFM), B/C/D (CPWD) modes. The system uses convex array probe with a frequency range of 2.1MHz to 5 MHz on abdomen for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. The ultrasonic imaging also helps to find a proper location for the transient elastography examination.

    AI/ML Overview

    The provided document, K173595, focuses on demonstrating substantial equivalence to predicate devices rather than proving device performance against specific clinical acceptance criteria through a dedicated study. Therefore, most of the requested information about acceptance criteria and a study proving their fulfillment is not present in this 510(k) summary.

    The document explicitly states: "No clinical study is included in this submission." This means there is no clinical trial data to evaluate device performance against clinical acceptance criteria.

    However, based on the information available, here's a breakdown of what can be derived:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a formal table of acceptance criteria for clinical performance that the device is tested against. Instead, it compares the technical characteristics and performance specifications of the proposed device ("FT9000 Shear Wave Quantificational Ultrasound Diagnostic System") to two predicate devices: FibroScan® (K123806) and Mindray DC-6 (K072164).

    The only "performance" discussed in relation to a quantitative metric is "Stiffness" when comparing FT9000 to FibroScan®. This comparison is likely based on internal testing or technical specifications, not a clinical trial.

    Feature/MetricAcceptance Criteria (Not Explicitly Stated as AC)Reported Device Performance (FT9000)Predicate Device 1 (FibroScan® - K123806) Performance
    Intended UseImplicit: Match predicate's intended use for shear wave measurement as an aid to clinical management of liver disease and general purpose ultrasound imaging.Measures 50Hz shear wave speed with 2.5MHz ultrasound in the liver as an aid to clinical management of liver disease. General purpose pulse echo ultrasound imaging and Doppler flow analysis for abdominal applications.Provides 50Hz shear wave speed measurements through internal structures of the body; indicated for noninvasive measurement of shear wave speed at 50 Hz in the liver as an aid to clinical management of liver disease.
    Operating PrincipleImplicit: Transient elastography.Transient elastography; uses mechanical vibrator (50 Hz shear waves) and ultrasound (2.5 MHz) to measure shear wave speed.Transient elastography, Vibration control elastography imaging.
    Probe Frequency (for elasticity)Implicit: Similar to predicate.2.5MHz2.5 MHz (XL+ probe) / 3.5 MHz (M+ probe)
    Stiffness BiasExpected to be within acceptable range relative to predicate.(-16.7%) – (8.6%)(-13.9%) – (1.3%)
    Stiffness PrecisionExpected to be within acceptable range relative to predicate.(0.9%) – (2.0%)(0%) – (3.1%)
    Imaging ModesImplicit: General purpose ultrasound imaging capabilities for abdominal applications, similar to predicate 2 (Mindray DC-6) for general imaging.B, B/B, B/C (CFM), B/C/D (CPWD) modes. Uses convex array probe with 2.1-5 MHz for abdominal imaging.Predicate 2 (Mindray DC-6) has B-Mode, M-Mode, Pulsed (PW) Doppler mode, Color Doppler mode, Continuous Wave (CW) Doppler mode, Amplitude Doppler Mode, 3D Imaging or Harmonic Imaging, 4D Imaging. FT9000 has fewer modes/features than DC-6 (e.g., no CW Doppler, Amplitude Doppler, 3D/4D).
    Compliance to StandardsIEC 60601-1, IEC 60601-1-2, NEMA UD 2-2004 (R2009), ISO 10993-5, ISO 10993-10Complied with all listed standards.Complied with all listed standards (or equivalent at the time of predicate clearance).

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    There is no clinical test set or data described in this submission, as it explicitly states "No clinical study is included in this submission." The reported "performance" metrics for stiffness bias and precision likely come from internal lab testing on phantoms or simulated data, rather than human subjects. Therefore, provenance information is not applicable for a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical study or test set with expert-established ground truth is presented.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study or test set is presented.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a diagnostic ultrasound system for shear wave quantification and general imaging. It is an aid to clinical management, but there is no mention of "AI assistance" or a MRMC study involving human readers with and without such assistance. The submission focuses on device functionality and safety, not on improving human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device itself, in its shear wave measurement mode, provides a quantitative output (shear wave speed) without human interpretation in the measurement step itself. The "stiffness bias" and "precision" figures likely represent standalone performance against a known standard (e.g., a tissue phantom). However, the overall interpretation of that shear wave speed for clinical management is still human-dependent. There is no specific study described to quantify this standalone performance in a clinical context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "Stiffness" metrics (bias and precision), the ground truth would likely be a known, quantifiable stiffness value of a phantom material used in laboratory testing. For the general ultrasound imaging functionality, compliance with NEMA UD 2-2004 (R2009) implies performance against physical phantom targets and established acoustic output standards.

    8. The sample size for the training set

    Not applicable. There is no mention of a machine learning component or a "training set" in the context of the device's development as described in this 510(k) summary. The device relies on established ultrasound physical principles and signal processing, not a trained AI model.

    9. How the ground truth for the training set was established

    Not applicable. As there's no mention of a training set, the establishment of its ground truth is irrelevant here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1