K Number
K173595
Date Cleared
2018-06-22

(213 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended to measure the speed of 50Hz shear wave with 2.5MHz ultrasound wave in the liver. The shear wave speed may be used as an aid to clinical management of patients with liver disease.

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal.

Device Description

The Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. The system is equipped with two probes, a fibrosis scanning probe used in elastography mode and an imaging probe used in imaging mode. The fibrosis scanning probe is used for elasticity measurement while the imaging probe is a convex probe used for ultrasound imaging.

Under elastography mode, the system uses transient elastography to measure shear wave speed non-invasively and estimate tissue stiffness in liver. A mechanical vibrator produces low-amplitude shear waves at 50 Hz that travel through the skin and intercostal space into the liver. The propagation speed of the shear wave is measured using ultrasound at 2.5 MHz.

Under imaging mode, the system acquires and displays ultrasound images in B, B/B, B/C (CFM), B/C/D (CPWD) modes. The system uses convex array probe with a frequency range of 2.1MHz to 5 MHz on abdomen for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. The ultrasonic imaging also helps to find a proper location for the transient elastography examination.

AI/ML Overview

The provided document, K173595, focuses on demonstrating substantial equivalence to predicate devices rather than proving device performance against specific clinical acceptance criteria through a dedicated study. Therefore, most of the requested information about acceptance criteria and a study proving their fulfillment is not present in this 510(k) summary.

The document explicitly states: "No clinical study is included in this submission." This means there is no clinical trial data to evaluate device performance against clinical acceptance criteria.

However, based on the information available, here's a breakdown of what can be derived:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of acceptance criteria for clinical performance that the device is tested against. Instead, it compares the technical characteristics and performance specifications of the proposed device ("FT9000 Shear Wave Quantificational Ultrasound Diagnostic System") to two predicate devices: FibroScan® (K123806) and Mindray DC-6 (K072164).

The only "performance" discussed in relation to a quantitative metric is "Stiffness" when comparing FT9000 to FibroScan®. This comparison is likely based on internal testing or technical specifications, not a clinical trial.

Feature/MetricAcceptance Criteria (Not Explicitly Stated as AC)Reported Device Performance (FT9000)Predicate Device 1 (FibroScan® - K123806) Performance
Intended UseImplicit: Match predicate's intended use for shear wave measurement as an aid to clinical management of liver disease and general purpose ultrasound imaging.Measures 50Hz shear wave speed with 2.5MHz ultrasound in the liver as an aid to clinical management of liver disease. General purpose pulse echo ultrasound imaging and Doppler flow analysis for abdominal applications.Provides 50Hz shear wave speed measurements through internal structures of the body; indicated for noninvasive measurement of shear wave speed at 50 Hz in the liver as an aid to clinical management of liver disease.
Operating PrincipleImplicit: Transient elastography.Transient elastography; uses mechanical vibrator (50 Hz shear waves) and ultrasound (2.5 MHz) to measure shear wave speed.Transient elastography, Vibration control elastography imaging.
Probe Frequency (for elasticity)Implicit: Similar to predicate.2.5MHz2.5 MHz (XL+ probe) / 3.5 MHz (M+ probe)
Stiffness BiasExpected to be within acceptable range relative to predicate.(-16.7%) – (8.6%)(-13.9%) – (1.3%)
Stiffness PrecisionExpected to be within acceptable range relative to predicate.(0.9%) – (2.0%)(0%) – (3.1%)
Imaging ModesImplicit: General purpose ultrasound imaging capabilities for abdominal applications, similar to predicate 2 (Mindray DC-6) for general imaging.B, B/B, B/C (CFM), B/C/D (CPWD) modes. Uses convex array probe with 2.1-5 MHz for abdominal imaging.Predicate 2 (Mindray DC-6) has B-Mode, M-Mode, Pulsed (PW) Doppler mode, Color Doppler mode, Continuous Wave (CW) Doppler mode, Amplitude Doppler Mode, 3D Imaging or Harmonic Imaging, 4D Imaging. FT9000 has fewer modes/features than DC-6 (e.g., no CW Doppler, Amplitude Doppler, 3D/4D).
Compliance to StandardsIEC 60601-1, IEC 60601-1-2, NEMA UD 2-2004 (R2009), ISO 10993-5, ISO 10993-10Complied with all listed standards.Complied with all listed standards (or equivalent at the time of predicate clearance).

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

There is no clinical test set or data described in this submission, as it explicitly states "No clinical study is included in this submission." The reported "performance" metrics for stiffness bias and precision likely come from internal lab testing on phantoms or simulated data, rather than human subjects. Therefore, provenance information is not applicable for a clinical test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical study or test set with expert-established ground truth is presented.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical study or test set is presented.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a diagnostic ultrasound system for shear wave quantification and general imaging. It is an aid to clinical management, but there is no mention of "AI assistance" or a MRMC study involving human readers with and without such assistance. The submission focuses on device functionality and safety, not on improving human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The device itself, in its shear wave measurement mode, provides a quantitative output (shear wave speed) without human interpretation in the measurement step itself. The "stiffness bias" and "precision" figures likely represent standalone performance against a known standard (e.g., a tissue phantom). However, the overall interpretation of that shear wave speed for clinical management is still human-dependent. There is no specific study described to quantify this standalone performance in a clinical context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the "Stiffness" metrics (bias and precision), the ground truth would likely be a known, quantifiable stiffness value of a phantom material used in laboratory testing. For the general ultrasound imaging functionality, compliance with NEMA UD 2-2004 (R2009) implies performance against physical phantom targets and established acoustic output standards.

8. The sample size for the training set

Not applicable. There is no mention of a machine learning component or a "training set" in the context of the device's development as described in this 510(k) summary. The device relies on established ultrasound physical principles and signal processing, not a trained AI model.

9. How the ground truth for the training set was established

Not applicable. As there's no mention of a training set, the establishment of its ground truth is irrelevant here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Wuxi Hisky Medical Technologies Co., Ltd. % Mr. Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. 5-402, Building #27, No. 56, LiangXiang East Rd., FangShan District Beijing, 102401 CHINA

Re: K173595

Trade/Device Name: Shear Wave Quantificational Ultrasound Diagnostic System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX, IYN Dated: May 23, 2018 Received: May 29, 2018

Dear Mr. Wang:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

June 22nd, 2018

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Bargas

for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K173595

Device Name

Shear Wave Quantificational Ultrasound Diagnostic System

Indications for Use (Describe)

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended to measure the speed of 50Hz shear wave with 2.5MHz ultrasound wave in the liver. The shear wave speed may be used as an aid to clinical management of patients with liver disease.

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal.

Details please refer to the attached pages.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Diagnostic Ultrasound Indications For Use

System: Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalNNNNNNote 1, 2
Intra-operative(Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Urology)
Other (OB/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralPeripheral vessel
VesselOther (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: A mode

Note 2: Vibration Controlled Transient Elastography at 50Hz

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription Use: ✓OROver-the Counter Use: __
(Per 21 CFR 801.109)(Per 21 CFR 807)

{4}------------------------------------------------

Diagnostic Ultrasound Indications For Use

System: Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000

Transducer: FT-2.5D9

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerPowerDoppler(Amplitude)Combined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalNNote 1, 2
Intra-operative(Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Specify)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: A mode

Note 2: Vibration Controlled Transient Elastography at 50Hz

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription Use: √OROver-the Counter Use: __
(Per 21 CFR 801.109)(Per 21 CFR 807)

{5}------------------------------------------------

Diagnostic Ultrasound Indications For Use

System: Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000 Transducer: FT-3.5R65, Convex

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Vibration Controlled Transient Elastography at 50Hz

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerPowerDoppler(Amplitude)Combined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (OB/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use: OR Over-the Counter Use: (Per 21 CFR 801.109) (Per 21 CFR 807)

{6}------------------------------------------------

Tab #3 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K173595

    1. Date of Preparation: 5/23/2018
    1. Sponsor Identification

Wuxi Hisky Medical Technologies Co., Ltd. Room B401, 530 Plaza, University Science Park, Taihu International Science & Technology Park, 214135 Wuxi, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA Establishment Registration Number: not registered

Contact Person: Jinhua Shao Position: General Manager Tel: 86-10-82151572 Fax: 86-10-82151571 Email: shaojh@fibrotouch.com

    1. Designated Submission Correspondent
      Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd. 5-402, Building #27, No. 56, LiangXiang East Rd.,

FangShan District, BeiJing, China 102401

Tel: +86-18910677558 Fax: +86-10-52214696 Email: ray.wang@believe-med.com

{7}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Shear Wave Quantificational Ultrasound Diagnostic System Common Name: Diagnostic Ultrasound System with Accessories Model(s): FT9000

Regulatory Information

Classification Name: 1) Ultrasonic Pulsed Echo Imaging System; 2) Diagnostic Ultrasound Transducer; 3) Ultrasonic Pulsed Doppler Imaging System; Classification:II Product Code:IYO & ITX&IYN Regulation Number:21 CFR 892.1550 & 21 CFR 892.1560&21 CFR 892.1570 Review Panel:Radiology:

Intended Use Statement:

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended to measure the speed of 50Hz shear wave with 2.5MHz ultrasound wave in the liver. The shear wave speed may be used as an aid to clinical management of patients with liver disease.

Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal.

Device Description

The Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. The system is equipped with two probes, a fibrosis scanning probe used in elastography mode and an imaging probe used in imaging mode. The fibrosis scanning probe is used for elasticity measurement while the imaging probe is a convex probe used for ultrasound imaging.

Under elastography mode, the system uses transient elastography to measure shear wave speed non-invasively and estimate tissue stiffness in liver. A mechanical vibrator produces low-amplitude shear waves at 50 Hz that travel through the skin and intercostal space into the liver. The propagation speed of the shear wave is measured using ultrasound at 2.5 MHz.

Under imaging mode, the system acquires and displays ultrasound images in B, B/B, B/C (CFM), B/C/D (CPWD) modes. The system uses convex array probe with a frequency range of 2.1MHz to 5 MHz on abdomen for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. The ultrasonic imaging also helps to find a proper location for the transient elastography examination.

{8}------------------------------------------------

Transducer ModelTypeFrequencyApplication
FT-2.5D9Round probe2.5MHz(deviation≤±15%)Liver, elasticity measurement
FT-3.5R65convex probe3.5MHz(nominal),2.1 , 2.5, 3.1, 5.0 (broadbandfrequency)Abdomen, general purpose pulseecho ultrasound imaging andDoppler flow analysis

Table 7-1 Transducer List

Fig 7-1 Working Frame of FT9000

Image /page/8/Figure/4 description: The image shows a block diagram of an ultrasound and elastography system. The system consists of two main parts: the elastography system and the ultrasound imaging system. Both systems are connected to a system control CPU, which also connects to a keyboard, monitor, and power supply. The elastography system includes modules for transmitting/receiving, signal excitation, pre-amplification, signal processing, and numerical calculation, while the ultrasound imaging system includes modules for real-time control, signal and image processing, digital beam synthesis, high-speed A/D conversion, and pre-amplification.

  • Identification of Predicate Device(s) ર.
    Predicate Device 1 : 510(k) Number: K123806 Product Name: FibroScan® Manufacturer: Echosens

Predicate Device 2 :

510(k) Number: K072164 Product Name: Diagnostic Ultrasound System Model Name: DC-6 Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

{9}------------------------------------------------

6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic a. safety and essential performance
  • IEC 60601-1-2:2007, Medical Electrical Equipment-Part 1-2: General Requirements For Basic b. Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests
  • NEMA UD 2-2004 (R2009), Acoustic Output Measurement Standard For Diagnostic Ultrasound c. Equipment Revision 3. (Radiology)
  • d. ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity.
  • ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation e. and delay-type hypersensitivity
    1. Clinical Test Conclusion

No clinical study is included in this submission.

{10}------------------------------------------------

8. Substantially Equivalent (SE) Comparison

Table 7-1 Comparison of Technology Characteristics to K123806

ItemProposed Device(s)Predicate Device 1(K123806)
Device nameFT9000 Shear Wave Quantificational Ultrasound Diagnostic SystemFibroScan®
Classification Name1) Ultrasonic Pulsed Echo Imaging System;2) Diagnostic Ultrasound Transducer;3) Ultrasonic Pulsed Doppler Imaging System;1) Ultrasonic pulsed echo imaging system2) Diagnostic Ultrasonic Transducer
Product CodeIYOITXIYNIYOITX
Regulation Number892.1550;892.1560;892.1570;892.1560;892.1570;
Intended UseShear Wave Quantificational Ultrasound Diagnostic System, ModelFT9000, is intended to measure the speed of 50Hz shear wave with2.5MHz ultrasound wave in the liver. The shear wave speed may beused as an aid to clinical management of patients with liver disease.Shear Wave Quantificational Ultrasound Diagnostic System, ModelFT9000, is intended for general purpose pulse echo ultrasoundimaging and Doppler flow analysis of the human body. It can beused in the following applications: Abdominal.The FibroScan® system is intended to provide 50Hz shear wave speedmeasurements through internal structures of the body.FibroScan® is indicated for noninvasive measurement of shear wavespeed at 50 Hz in the liver. The shear wave speed may be used as anaid to clinical management of patients with liver disease.

{11}------------------------------------------------

510(k) Summary

Probe TypesFibrosis scanning probe FT-2.5D9M+ probeXL+ probe
Probe frequencyFibrosis scanning probe (FT-2.5D9):2.5MHzM+ probe: 3.5 MHzXL+ probe: 2.5 MHz
Modes of OperationM-modeA-modeM-modeA-mode
Applied Standards:
BiocompatibilityISO10993-5&ISO10993-10ISO10993-5&ISO10993-10
Electrical SafetyIEC60601-1IEC60601-1
EMCIEC60601-1-2IEC60601-1-2
PerformanceUD2UD2

The subject device is similar to the predicate devices:

  • Has the same intended use and indications for use >
  • Utilizes the same operating principle A
  • Incorporates the same basic design A
  • Incorporates the same technological characteristics A
  • Tested to the same electrical and electromagnetic safety standards for medical electrical equipment >
  • A Manufactured under a quality system

{12}------------------------------------------------

ItemProposed Device(s)Predicate Device 2 (K072164)
Device nameFT9000 Shear Wave Quantificational Ultrasound Diagnostic SystemDC-6 Diagnostic Ultrasound System
Classification NameUltrasonic Pulsed Echo Imaging System;Diagnostic Ultrasound Transducer;Ultrasonic Pulsed Doppler Imaging System;Ultrasonic pulsed echo imaging system
Product CodeIYOITXIYNIYOITXIYN
Regulation Number892.1550;892.1560;892.1570;892.1550;892.1560;892.1570;
Intended UseShear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended to measure the speed of 50Hz shear wave with 2.5MHz ultrasound wave in the liver. The shear wave speed may be used as an aid to clinical management of patients with liver disease.Shear Wave Quantificational Ultrasound Diagnostic System, Model FT9000, is intended for general purpose pulse echo ultrasound imaging and Doppler flow analysis of the human body. It can be used in the following applications: Abdominal.The device is intended for use by a qualified physician for ultrasound evaluation of abdominal, cardiac, small parts (breast, testes, thyroid, etc.), peripheral vascular, fetal, transrectal, transvaginal, pediatric,neonatal cephalic, musculoskeletal (general and superficial), Urology/Prostate and intraoperative (liver, gallbladder, pancreas)
Probe TypesConvex probeConvex probeLinear probeMicro-Convex Probe
Modes of OperationB Mode,B-Mode
N/AM-Mode
Pulsed (PW) Doppler ModePulsed (PW) Doppler mode
N/AContinuous wave (CW) Doppler mode
Color Doppler ModeColor Doppler Mode
N/AAmplitude Doppler Mode
N/A3D Imaging or Harmonic Imaging
N/A4D Imaging
Applied Standards:
BiocompatibilityISO10993-5&ISO10993-10ISO10993-5&ISO10993-10
Electrical SafetyIEC60601-1IEC60601-1
EMCIEC60601-1-2IEC60601-1-2
PerformanceUD2IEC60601-2-37
UD2
Acoustic OutputTrack 1 & Track 3Track 3

Table 7-2 Comparison of Technology Characteristics to K072164

{13}------------------------------------------------

510(k) Summary

{14}------------------------------------------------

  1. Substantially Equivalent (SE) Conclusion

SE Analysis :

The subject device has same classification information, same intended use, same indication for use, similar product design, similar specification, same safety elements, similar applied Standards as predicate device.

The differences are included as followings:

  1. Difference between Fibroscan® (K123806) and FT9000

FT9000 has one additional imaging probe, FT-3.5R65, which is a convex Ultrasonic Imaging Probe for image guiding.

Instead of FibroScan's two elasticity probes(M*, XL*), FT9000 come with one fibrosis scanning probe (FT-2.5D9).

  1. Difference between Mindray DC-6 (K072164) and FT9000

No Continuous Wave (CW) Doppler Mode, Amplitude Doppler Mode, 3D/4D Imaging or Harmonic Imaging Mode for FT9000, due to the less variety of ultrasound probes for FT9000, one convex probe, other than Mindray's DC-6 multiple ultrasound probes.

FT9000 follows Track 1 and Track 3 of FDA ultrasound submission, and Mindray DC-6 follows Track 3, but this difference does not raise issues in safety since FT9000 follows UD2 acoustic output standard and FDA guidance: Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (Issued on 9/9/2008).

FibroScan®K123806FT9000Conclusion / Discussion
Transient elastographyTransient elastographySame
Vibration control elastography imagingVibration control elastography imagingSame
Frequency range(2.5MHz,3.5MHz)Frequency: 2.5MHzSimilar
M-mode, A-modeM-mode, A-modeSame
N/AUltrasound-B imagingAdditional 2D imaging.
StiffnessBias: (-13.9%) – (1.3%)Precision: (0%) – (3.1%)StiffnessBias: (-16.7%) – (8.6%)Precision: (0.9%) – (2.0%)Similar

{15}------------------------------------------------

510(k) Summary

Analyse 1: The device and predicate device have difference in performance specification, But the propose device have tested for measurement accuracy by accuracy testing and software validation, so these difference can prove the effectiveness of propose device.

Analyse 2: FT9000 follows Track 1 of FDA ultrasound submission, and Mindray DC-6 follows Track 3, but this difference does not raise issues in safety since FT9000 follows UD2 acoustic output standard.

Conclusion: The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.