Search Results
Found 4 results
510(k) Data Aggregation
(105 days)
Disposable Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. In addition, this surgical gown meets the requirements of AAMI Level 4 barrier protection for a surgical gown per ANSU AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. It is single use, disposable medical devices, provided sterile.
The subject devices are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate materials.
In addition, this surgical gown meets the requirements of AAMI Level 4 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.
The proposed device is single use, disposable medical devices and provided sterile.
The main materials of device are SMS nonwoven +TPU membrane, PP and Polyester fiber, and the main manufacturing technique are Ultrasonic welding and Sewing.
The provided text is a 510(k) Summary for a medical device (Disposable Surgical Gown) and does not describe an AI/ML-driven device or study. Therefore, most of the requested information regarding acceptance criteria, training/test sets, ground truth establishment, expert adjudication, or MRMC studies for an AI/ML device is not applicable.
Instead, the document details the device's physical properties, material composition, and performance against established industry standards for surgical gowns. The "study" mentioned is a series of non-clinical tests to verify the device meets design specifications and relevant standards.
Here's a breakdown of the relevant information from the document, tailored to the non-AI/ML context:
Device Description: Disposable Surgical Gown
Purpose of the "Study" (Non-Clinical Testing): To verify that the proposed device (Disposable Surgical Gown) meets all design specifications and complies with relevant standards for surgical gowns, demonstrating its safety and effectiveness compared to a legally marketed predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
The document provides a table of non-clinical tests conducted, their purpose, acceptance criteria, and results.
Test Method | Purpose | Acceptance Criteria | Results |
---|---|---|---|
AATCC 127 | Water resistance Hydrostatic Pressure | >50 cm | Passed |
AATCC 42 | Water Resistance impact penetration Nonwoven and plastic Barriers | ≤1.0 g | Passed |
ASTM D5034-09 | Tensile Strength | MD Mean≥30 lbs; CD Mean≥25 lbs | Passed |
ASTM D5587-19 | Tearing Strength | MD Mean≥9 lbs; CD Mean≥18 lbs | Passed |
ASTM F1868-17 | Evaporative Resistance of fabrics | Mean Evaporative Resistance (Ref)≥0.06(kPa·m2/W). | Passed |
16 CFR 1610 | Flammability testing | Meets requirements of Flame Resistant CPSC 1610 Class 1 | Meets requirements |
ASTM D1683-17 | Seam Strength | >50N | Passed |
ASTM F1671 | Resistance to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage | No virus through, “None Seen” | Passed |
ISO 9073-10:2003 | Lint and Other particles generation in the dry state | Log 10 | Bacterial Endotoxin Test |
Ask a specific question about this device
(51 days)
Sterilization Wraps is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:
• Pre-vacuum steam at 270°F/132°C for 4 minutes
Types of medical devices to be sterilized in the pre-vacuum cycle:
General purpose reusable metal and nonmetal devices(No lumen) including devices with stainless steel diffusionrestricted spaces such as forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.
The size of the Devices to be sterilization Wraps pack and form a closed space. Validated for dry time is 30 minutes for Sterilization Wraps.
Color of wrap: Blue Size of wrap: 48 in x 48 in
The maximum validated weight of load for Sterilization Wraps is 12kg.
The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until used.
Sterilization Wraps is a 63gsm, latex-free, 3-layer (SMS) non-woven sterile wrap, manufactured with spun-bonded / meltblown polypropylenem. Sterilization Wraps provides a strong barrier which protects against cuts, tears with particularly heavy orthopedic sets. Sterilization Wraps is designed to be implemented as an outer sterilization wrap which can be used in combination with Clinipak choice. Sterilization Wraps is made of polypropylene and blue pigment by non-woven process.
The provided document describes the acceptance criteria and the results of the study for the Sterilization Wraps device (K220133). This is a non-clinical study focused on the physical and biological characteristics of the wrap, rather than a clinical study involving human or AI performance.
Here's a breakdown of the requested information based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by meeting the requirements of recognized industry standards. The study demonstrates that the proposed device complies with these standards.
Study/Test | Standard(s) Used | Description/Criteria (Acceptance Criteria) | Reported Device Performance (Results) |
---|---|---|---|
Sterilization Validation - Steam PREVACUUM | ANSI AAMI ISO 17665-1:2006/(R)2013 | A method of steam sterilization was validated to a sterility assurance level (SAL) of 10^-6 | Pass |
Sterilant Penetration - Steam PREVACUUM | ANSI AAMI ISO 17665-1:2006/(R)2013 | The testing details the methods used in determining the internal temperature profile for wrapped sterilization packs when processed in a steam sterilization pre-vacuum cycle at 132°C (270°F) for four (4) minutes exposure. | Testing has demonstrated adequate sterilant penetration |
Validation - Dry Time | ANSI AAMI ISO 17665-1:2006/(R)2013 | Determining the proper drying time required | Test samples meet or exceed the minimum criteria for dry time. |
90 Day Real Time Maintenance of Sterility Validation -Steam PREVACUUM | AATCC 127-18, ISO 11607:2019, ASTM D3776/D3776M, ASTM D3786/D3786M-18 | The study details the methods used in verifying the test samples can maintain the integrity of its contents for an extended period of time following exposure to a steam sterilization process. | No growth |
Package Integrity Test - Steam PREVACUUM | ASTM D5587-15, ASTM D5034-2009, ASTM D737-18, ASTM F2101-14, AATCC 127-18, ISO 11607:2019, ASTM D5587-15, ASTM D5034-2009 | The testing details the methods of the test sample in maintaining package integrity. The purpose of the physical properties testing was to demonstrate passing results for the physical properties (contain Hydrostatic pressure, Weight, Bursting Strength, Tear Resistance, Tensile Strength ) for the wrap. | The subject wrap, were found to be effective barriers when processed in a Steam Pre-Vacuum cycle. The physical properties testing met the acceptance criteria and demonstrated passing results. |
Bacterial Filtration Efficiency (BFE) of Non-Woven Sterilization Wrap When Processed In a Steam Sterilization Cycle | ISO 11607:2019, ASTM F2101-14 | BFE testing is a type of test used to determine the efficiency of filter materials to provide protection against microbial organisms. | Pass |
Dimension testing | ISO 11607:2019 | 48 in X 48 in (This is the specified dimension, implying it's the acceptance criterion) | Pass |
Weight testing | ASTM D3776/D3776M | According to the standard test, the weight should meet the requirements | Pass |
Air Permeability Test | ASTM D737 | Air permeability should meet the requirements of ASTM D737. | Pass |
Shelf Life Testing | AATCC 127-18, ISO 11607:2019, ASTM D3786/D3786M-18, ASTM D5587-15, ASTM D5034-2009, ASTM D737-18, ASTM F2101-14 | Whole package integrity test of real time shelf life samples (The acceptance criterion is implied to be capable of maintaining sterility and package integrity) | Sterilization Wraps was capable of maintaining sterility and package sterilization wrapper) following an approximate 18 month period of real time shelf life prior to being sterilized in the steam sterilization cycles. |
Cytotoxicity testing | ISO 10993-5: 2009 | No cytotoxic potential | Pass |
Irritation testing | ISO 10993-10: 2010 | No irritation on the skin | Pass |
Sensitization testing | ISO 10993-10: 2010 | No skin sensitization | Pass |
Additional Information
2. Sample size used for the test set and the data provenance:
- The document does not explicitly state numerical sample sizes for each test. Instead, it refers to "test samples" or "subject wrap."
- Data Provenance: Not specified, but generally, these types of non-clinical tests are performed in a laboratory setting. There is no information regarding country of origin for the test data, nor if it's retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This is a non-clinical study involving material and biological performance testing (e.g., sterilization effectiveness, physical properties, biocompatibility). It does not involve "experts" establishing a ground truth in the way a clinical diagnostic study would (e.g., radiologists interpreting images). The "ground truth" is established by the technical specifications and performance requirements outlined in the referenced international standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. This is a non-clinical performance study against predetermined technical standards, not a diagnostic accuracy study requiring adjudication of interpretations.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not a study involving human readers, AI assistance, or diagnostic effectiveness.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical sterilization wrap, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth is based on the requirements and methodologies defined by internationally recognized standards (e.g., ISO, ASTM, AAMI, AATCC, USP) for sterilization wraps and medical device packaging. Compliance with these standards serves as the "ground truth" for acceptable performance.
8. The sample size for the training set:
- Not applicable. This is a physical product, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(84 days)
The Nitrile Examination Gloves (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device, Nitrile Examination Gloves (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.
The proposed device is provided non-sterile. The proposed device is made of Nitrile.
The provided text describes the acceptance criteria and the study results for "Nitrile Examination Gloves (Powder free, Blue)". This is a physical medical device, not a software-driven AI device. Therefore, the specific questions regarding AI-related performance metrics (like sample size for test set/training set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance) are not applicable to this document.
However, I can extract the acceptance criteria and reported device performance from the provided tables.
Acceptance Criteria and Reported Device Performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance | Comments |
---|---|---|---|---|
Physical Performance (ASTM D6319, D412, D573) | ||||
Freedom from Holes (ASTM D5151) | Testing for Freedom from holes | AQL 2.5 | No water leakage is inspected from 200 samples | Meets acceptance criteria. |
Residual Powder (ASTM D6124) | Determine the powder residue for powder free gloves |
Ask a specific question about this device
(91 days)
The Surgical Masks is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Surgical Masks is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided non-sterile.
The Model YH-EZ-004(White Color, Blue Color) is Level 2 barrier as ASTM F2100 requirements.
The proposed device(s) includes 1 model, which is YH-EZ-004. This model has two colors, Blue color and White color. This model is Flat Pleated type mask, utilizing Ear Loops' way for wearing, and it has Nose Piece design for fitting the facemask around the nose.
All two colors of the model YH-EZ-004 of proposed device(s) share same materials and structure, they all are manufactured with three layers, the inner and outer layers are made of spunbonded non-woven fabric, and the middle layer is made of melt-blown non-woven fabric.
The nose piece contained in the proposed device(s) is in the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polyethylene.
The proposed device(s) is held in place over the user's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are made with terylene and spandex.
The proposed device(s) are sold non-sterile and are intended to be single-use, disposable devices.
Based on the provided text, the device in question is a Surgical Mask, and the submission is a 510(k) premarket notification. This type of submission typically relies on demonstrating substantial equivalence to a legally marketed predicate device rather than extensive clinical studies for novel AI/ML-driven diagnostics.
Therefore, the requested information regarding AI/ML-specific acceptance criteria, test sets, expert ground truth, MRMC studies, and standalone algorithm performance is not applicable to this document. This document describes the testing for physical and biological properties of the surgical masks to show they perform as well as a predicate device.
Here's the information that is applicable based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the surgical masks are based on ASTM F2100 Level 2 requirements and other relevant standards.
ITEM | Acceptance Criteria | Results (Blue Color) | Results (White Color) |
---|---|---|---|
Performance Characteristics (ASTM F2100 Level 2 Requirement) | |||
Fluid Resistance (ASTM F1862) | 120 mmHg | 120 mmHg | 120 mmHg |
Particulate Filtration Efficiency (ASTM F2299) at 0.1 micron | ≥ 98% | ≥ 99.30% | ≥ 99.77% |
Bacterial Filtration Efficiency (ASTM F2101) | ≥ 98% | ≥ 98.54% | ≥ 99.04% |
Differential Pressure (Delta P) (EN 14683 Annex C) |
Ask a specific question about this device
Page 1 of 1