Search Filters

Search Results

Found 15 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    Wear Safe (Malaysia) Sdn. Bhd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Powder Free Nitrile Surgical Glove, White Colored, Tested for Use with Chemotherapy Drugs, with Non-Pyrogenic Labeling Claim is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. These gloyes were tested for use with cherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Sterile Powder Free Nitrile Surgical Glove, White Colored, Tested for Use with Chemotherapy Drugs, with Non-Pyrogenic Labeling Claim is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    AI/ML Overview

    The document describes the acceptance criteria and performance of a "Sterile Powder Free Nitrile Surgical Glove, White Colored, Tested for Use with Chemotherapy Drugs, with Non-Pyrogenic Labeling Claim." This is a medical device, not an AI/ML device, and thus the acceptance criteria and study data provided are for a physical product, not an algorithm's performance.

    Therefore, many of the requested categories (such as sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set size, and how ground truth for training set was established) are not applicable to this type of device and are not provided in the document.

    1. A table of acceptance criteria and the reported device performance:

    Chemotherapy Drug and ConcentrationAcceptance Criterion: Minimum Breakthrough Detection Time (minutes)Reported Device Performance: Minimum Breakthrough Detection Time (minutes)
    Carmustine (BCNU) (3.3 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.78.4
    Cisplatin (1.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Cyclophosphamide (Cytoxan) (20.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Cytarabine(100 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Dacarbazine (DTIC) (10.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Doxorubicin Hydrochloride (2.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Etoposide (20.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Fluorouracil (50.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Ifosfamide (50.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Methotrexate (25.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Mitomycin C (0.5 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Mitoxantrone (2.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Paclitaxel (Taxol) (6.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240
    Thiotepa (10.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.118.8
    Vincristine Sulfate (1.0 mg/ml)Not explicitly stated as an "acceptance criterion" in this document, but implies a desired performance level based on ASTM D6978-05.>240

    Study Proving Device Meets Acceptance Criteria:

    The study that proves the device meets the performance criteria is described as follows:

    • Methodology: "These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." This standard practice outlines the procedure for testing the resistance of medical gloves to permeation by chemotherapy drugs.

    Details specific to AI/ML device studies (as requested in the prompt but not applicable here):

    1. Sample size used for the test set and the data provenance: Not applicable. For this physical glove, the "test set" would be the number of gloves tested and the specific conditions per the ASTM standard. This information is not provided beyond the reference to the standard.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical glove's chemical resistance is established by laboratory measurements according to a standard.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the performance of the chemotherapy drug permeation is established through laboratory testing following the specific protocols of ASTM D6978-05 (Reapproved 2013). This involves precise measurement of breakthrough detection time.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123469
    Date Cleared
    2013-02-19

    (98 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN BHD - FACILITY II

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored), Non-Sterile meets all of the requirements of ASTM D 6319-10. The gloves are ambidextrous single-use disposable devices that come in five sizes (XS, S, M, L, XL) in blue or white color.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the device, based on the provided text:

    Acceptance Criteria and Device Performance

    CharacteristicAcceptance Criteria (Standards)Reported Device Performance
    DimensionsASTM D 6319-10Meets
    Physical PropertiesASTM D 6319-10Meets
    Freedom from PinholesASTM D 5151-06, ASTM D 6319-10Meets
    Powder Free ResidueASTM D 6124-06, ASTM D 6319-10Meets
    Biocompatibility:
    - Dermal SensitizationISO 10993-10:2010Not a contact skin sensitizer
    - Primary Skin Irritation16 CFR Part 1500 (Primary Skin Irritation Test)Not a primary skin irritant

    Study Information

    The provided document describes a 510(k) Premarket Notification for a medical device (Powder Free Nitrile Patient Examination Glove). For this type of submission, the primary goal is to demonstrate substantial equivalence to a predicate device, rather than conducting extensive clinical studies to prove effectiveness from scratch. Therefore, the "study" mainly refers to the non-clinical performance data demonstrating compliance with established standards.

    1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify exact sample sizes for each test conducted (e.g., how many gloves were tested for pinholes, dimensions, etc.). It generally states that the device "Meets" the respective ASTM and ISO standards, implying that the number of samples tested was sufficient to establish conformity according to those standards.
    • Data Provenance: The testing was performed by the manufacturer, WSM WEAR SAFE (MALAYSIA) SDN BHD, located in Malaysia. The data is retrospective, as it refers to performance tests conducted on the device to demonstrate compliance with standards for the 510(k) submission.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This type of information is not applicable to this submission. The "ground truth" for glove performance is established by widely accepted international and national standards (ASTM, ISO, CFR). The experts involved would be the engineers and quality assurance personnel at the manufacturing facility who conducted the tests and verified compliance with these objective standards. Their qualifications are inherent in their ability to perform and interpret these standardized tests.

    3. Adjudication method for the test set:

    • Not applicable in the context of expert adjudication for clinical outcomes. The "adjudication" here is the comparison of the device's measured performance against the quantitative and qualitative requirements outlined in the specified standards. Compliance is determined by whether the test results fall within the acceptable ranges or pass/fail criteria of those standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, this was not done. The device in question is a patient examination glove, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to human reader improvement with or without AI assistance is irrelevant and not performed.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this was not done. Again, this is a physical medical device (glove), not a software algorithm. The concept of "standalone" algorithm performance does not apply.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for this device's performance is based on established industry standards and regulatory requirements. For example:
      • Dimensions, Physical Properties, Freedom from Pinholes, Powder Free Residue: The ground truth is defined by the specifications and test methods outlined in ASTM D 6319-10, ASTM D 5151-06, and ASTM D 6124-06.
      • Biocompatibility: The ground truth for biocompatibility is defined by the test methods and acceptance criteria specified in ISO 10993-10:2010 (for dermal sensitization) and 16 CFR Part 1500 (for primary skin irritation).

    7. The sample size for the training set:

    • Not applicable. This device is a physical medical product, not a machine learning model, so there is no "training set." The manufacturing process would have internal quality control and verification procedures, but these are distinct from AI model training.

    8. How the ground truth for the training set was established:

    • Not applicable. As there is no training set, the question of how its ground truth was established is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K101799
    Date Cleared
    2010-09-03

    (67 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free latex examination glove is a disposable device made of natural rubber material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants.

    Device Description

    Powder Free Natural Rubber Latex Examination Gloves, Blue Color, Non-Sterile will meet all the current specification for ASTM D3578.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Powder Free Natural Rubber Latex Examination Gloves:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicStandards / Acceptance CriteriaDevice Performance
    DimensionsASTM D3578 - 05Meets
    Physical PropertiesASTM D 412 - 06ae1Meets
    Freedom from pin-holesASTM D 5151 - 06Meets
    Powder Free ResidueASTM D 6124 - 06Meets
    Soluble Protein LevelASTM D 5712 - 05 (also protein labeling claim of 50 micrograms per dm² of glove or less)Meets (with protein labeling claim of 50 micrograms per dm² of glove or less)
    Biocompatibility:
    - Dermal SensitizationDermal Sensitization (as per ASTM F720-81)Not a contact skin sensitizer
    - Skin IrritationPrimary Skin Irritation Test (as per 16CFR Part1500)Not a primary skin irritant

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each of the tests (Dimensions, Physical Properties, Freedom from pin-holes, Powder Free Residue, Soluble Protein Level, Biocompatibility).

    The data provenance is not specified beyond indicating these are "performance test data." It is highly likely these were conducted as prospective tests by the manufacturer, specifically for this submission. The country of origin of the data would be Malaysia, where WEAR SAFE (MALAYSIA) SDN. BHD. is located and conducted the testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The tests performed are objective, laboratory-based physical and chemical assays, and biocompatibility tests. They rely on standardized methods rather than expert interpretation of data like in an imaging study. Therefore, the concept of "experts establishing ground truth" in the way it might apply to medical imaging or diagnostics is not directly relevant here. The ground truth for these tests is defined by the objective measurement criteria of the specified ASTM and CFR standards.

    4. Adjudication method for the test set:

    Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments where multiple readers might disagree on findings (e.g., radiologists interpreting images). Since these are objective physical, chemical, and biocompatibility tests, no adjudication method in this sense would have been used. The results are based on direct measurements and observations according to the test protocols.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. This device is a patient examination glove, a physical product, not an AI-powered diagnostic or assistive technology. The concept of "human readers improving with AI assistance" is not applicable to this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    No standalone algorithm performance study was done. This is a physical non-AI device.

    7. The type of ground truth used:

    The ground truth for these tests is based on objective measurement criteria and thresholds defined by established international standards:

    • ASTM D3578-05 for Dimensions
    • ASTM D 412-06ae1 for Physical Properties
    • ASTM D 5151-06 for Freedom from pin-holes
    • ASTM D 6124-06 for Powder Free Residue
    • ASTM D 5712-05 for Soluble Protein Level
    • ASTM F720-81 for Dermal Sensitization
    • 16CFR Part 1500 for Primary Skin Irritation Test

    These standards specify the methodology and acceptable limits for each characteristic.

    8. The sample size for the training set:

    Not applicable. This device is a physical product, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093500
    Date Cleared
    2010-01-19

    (68 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free nitrile examination glove is a disposable device made of synthetic material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants. In addition, this product is tested for use with chemotherapy drugs.

    Device Description

    Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non Sterile (Chemotherapy Drug Protection Labeling Claim) will meet all the current specification for ASTM D6319.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the "Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim)" based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance Criteria / StandardDevice Performance
    DimensionsASTM D6319 - 00a(2005)e1Meets
    Physical PropertiesASTM D 412 - 06ae1Meets
    Freedom from pin-holesASTM D 5151 - 06Meets
    Powder Free ResidueASTM D 6124 - 06Meets
    BiocompatibilityDermal Sensitization (as per ASTM F720-81)Not a contact skin sensitizer
    Primary Skin Irritation Test (as per 16CFR Part1500)Not a primary skin irritant
    Chemotherapy Drug Permeation (White Gloves)Carmustine (BCNU) (3.3mg/ml) breakthrough time71.3 minutes
    Cisplatin (1.0 mg/ml) breakthrough time>240 minutes
    Cyclophosphamide (Cytoxan) (20.0 mg/ml) breakthrough time>240 minutes
    Dacarbazine (DTIC) (10.0 mg/ml) breakthrough time>240 minutes
    Doxorubicin Hydrochloride (2.0 mg/ml) breakthrough time>240 minutes
    5-Fluorouracil (50.0 mg/ml) breakthrough time>240 minutes
    Etoposide (Toposar) (20.0 mg/ml) breakthrough time>240 minutes
    Paclitaxel (Taxol) (6.0 mg/ml) breakthrough time>240 minutes
    ThioTepa (10.0 mg/ml) breakthrough time72.36 minutes
    Chemotherapy Drug Permeation (Blue Gloves)Carmustine (BCNU) (3.3mg/ml) breakthrough time60.7 minutes
    Cisplatin (1.0 mg/ml) breakthrough time>240 minutes
    Cyclophosphamide (Cytoxan) (20.0 mg/ml) breakthrough time>240 minutes
    Dacarbazine (DTIC) (10.0 mg/ml) breakthrough time>240 minutes
    Doxorubicin Hydrochloride (2.0 mg/ml) breakthrough time>240 minutes
    5-Fluorouracil (50.0 mg/ml) breakthrough time>240 minutes
    Etoposide (Toposar) (20.0 mg/ml) breakthrough time>240 minutes
    Paclitaxel (Taxol) (6.0 mg/ml) breakthrough time>240 minutes
    ThioTepa (10.0 mg/ml) breakthrough time69.0 minutes

    (Note: The document states "Meets" for general characteristics, implying the device performance met the specified ASTM standards. For chemotherapy drug permeation, specific breakthrough times are provided as observed performance.)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample size used for each specific test (e.g., number of gloves tested for pin-holes, or number of samples for each chemotherapy drug).

    The data provenance is from Malaysia, as the submitting company, Wear Safe (Malaysia) Sdn. Bhd., is based in Malaysia. The studies appear to be prospective performance tests conducted on the device to demonstrate compliance with standards and for chemotherapy drug permeation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is a physical product (examination gloves), and its performance is assessed against established physical and chemical standards (ASTM, 16CFR). The "ground truth" is defined by the technical specifications and test methodologies outlined in these standards, not by expert consensus in a clinical diagnostic sense. Therefore, no experts in the context of clinical interpretation or diagnosis were used to establish ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. As described above, the evaluation is based on objective measurements against established standards, not on subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No. An MRMC comparative effectiveness study is designed for evaluating diagnostic tools where human readers interpret medical images or data, often with and without AI assistance. This is not relevant for the performance evaluation of examination gloves.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    Yes, in essence. The performance tests conducted for the gloves are "standalone" in the sense that they assess the intrinsic physical, chemical, and barrier properties of the glove material itself, without human intervention in the interpretation of results beyond standard laboratory procedures. There is no "algorithm" involved, but the device's performance is measured directly.

    7. The Type of Ground Truth Used:

    The ground truth used is based on established industry standards and regulatory specifications. Specifically:

    • ASTM (American Society for Testing and Materials) Standards: For dimensions, physical properties, freedom from pin-holes, and powder-free residue.
    • 16CFR Part 1500: For primary skin irritation.
    • ASTM F720-81: For dermal sensitization.
    • Chemotherapy Drug Permeation Testing: The "ground truth" for the chemotherapy drug claim is the measured breakthrough time of each specified drug through the glove material, which is an objective measurement based on a defined test methodology (though the specific standard for this permeation testing is not explicitly cited in the provided text, it is common practice to follow recognized standards like ASTM D6978).

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092492
    Date Cleared
    2009-10-21

    (69 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free polymer coated latex examination glove is a disposable device made of natural rubber material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants.

    Device Description

    The Powder Free Polymer Coated Latex Examination Gloves, Non-Sterile, with Protein Labeling claim of 50 micrograms per dm2 of glove or less of water soluble protein will meet all the current specification for ASTM D3578.

    AI/ML Overview

    This document is a 510(k) summary for a medical device: "Powder Free Polymer Coated Latex Examination Gloves, Non-Sterile, with Protein Labeling claim of 50 micrograms per dm2 of glove or less of water soluble protein."

    Here's an analysis of the provided information, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standards)Device Performance
    DimensionsASTM D3578 - 05Meets
    Physical PropertiesASTM D 412 - 06ae1Meets
    Freedom from pin-holesASTM D 5151 - 06Meets
    Powder Free ResidueASTM D 6124 - 06Meets
    Soluble Protein LevelASTM D 5712 - 05 (claim of
    Ask a Question

    Ask a specific question about this device

    K Number
    K090066
    Date Cleared
    2009-05-15

    (126 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free polymer coated latex surgical glove is a disposable device made of natural rubber material intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The Powder Free Polymer Coated Latex Surgical Gloves, Sterile, with Protein Labeling claim of 50 micrograms per dm2 of glove or less of water soluble protein will meet all the current specification for ASTM D3577.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the "Powder Free Polymer Coated Latex Surgical Gloves" based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance Criteria (Standards)Reported Device Performance
    DimensionsASTM D3577 - 06Meets
    ASTM D 412 - 06aelMeets
    Freedom from pin-holesASTM D 5151 - 06Meets
    Powder Free ResidueASTM D 6124 - 06Meets
    Biocompatibility: Dermal SensitizationDermal Sensitization (as per ISO 10993-10)Not a contact skin sensitizer
    Biocompatibility: Primary Skin IrritationPrimary Skin Irritation Test (as per ISO 10993-10)Not a primary skin irritant
    Protein Labeling Claim50 micrograms per dm² of glove or less of water soluble proteinMeets (implied by overall conclusion of safety and effectiveness)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not provide specific details on the sample size used for the tests to demonstrate compliance with the ASTM and ISO standards. It states that "The performance test data that support a determination of substantial equivalence are described above," referring to the "Device Performance" column in the table.

    Regarding data provenance, the document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. The manufacturer is based in Malaysia.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable to this type of device and study. The testing for surgical gloves involves physical and chemical property assessments based on established international standards (ASTM, ISO), not expert interpretation of diagnostic images or patient outcomes.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation, such as in medical imaging, to resolve discrepancies among experts establishing ground truth. The tests for surgical gloves are objective measurements against defined standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, which is not applicable to a physical medical device like surgical gloves.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Yes, a standalone study was done in the sense that the device's performance was evaluated independently against the specified standards without human intervention influencing the device itself. However, this is not an "algorithm-only" study as the device is a physical product, not a software algorithm. The "Device Performance" column reflects the standalone performance of the gloves.

    7. The Type of Ground Truth Used:

    The ground truth used for this device is based on established objective measurements and performance specifications defined by international standards (ASTM and ISO). For example:

    • Dimensions: Measured directly and compared against ASTM D3577.
    • Freedom from pin-holes: Evaluated according to ASTM D5151.
    • Powder Free Residue: Measured according to ASTM D6124.
    • Biocompatibility (Dermal Sensitization, Primary Skin Irritation): Assessed using standardized biological tests as per ISO 10993-10, with defined criteria for "not a contact skin sensitizer" and "not a primary skin irritant."

    8. The Sample Size for the Training Set:

    This information is not applicable. Surgical gloves are physical devices tested against predefined standards, not machine learning models that require training sets.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090327
    Date Cleared
    2009-04-03

    (53 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE (MALAYSIA) SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free polymer coated latex examination glove is a disposable device made of natural rubber material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants.

    Device Description

    The Powder Free Polymer Coated Latex Examination Gloves, Sterile, with Protein Labeling claim of 50 micrograms per dm- of glove or less of water soluble protein will meet all the current specification for ASTM D3578.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:

    Acceptance Criteria and Device Performance for Wear Safe (Malaysia) Sdn.Bhd. Powder Free Polymer Coated Latex Examination Gloves

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standards)Reported Device Performance
    DimensionsASTM D3578 - 05Meets
    Physical PropertiesASTM D 412 - 06ae1Meets
    Freedom from pin-holesASTM D 5151 - 06Meets
    Powder Free ResidueASTM D 6124 - 06Meets
    Biocompatibility:Dermal Sensitization (as per ISO 10993-10)Not a contact skin sensitizer
    Primary Skin Irritation Test (as per ISO 10993-10)Not a primary skin irritant
    Protein Labeling Claim50 micrograms per dm² of glove or less of water soluble proteinClaimed to meet

    Study Details:

    Based on the provided document, the study conducted is a non-clinical performance evaluation against established standards for patient examination gloves.

    1. Sample size used for the test set and the data provenance:
      The document does not specify the exact sample size used for each test. However, it indicates that the tests were performed according to specified ASTM and ISO standards, which inherently include sampling methodologies. The provenance of the data is Malaysia, as the manufacturer is "Wear Safe (Malaysia) Sdn.Bhd." The data is retrospective in the sense that it reflects testing performed on product samples to demonstrate compliance at the time of the 510(k) submission.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is not applicable to this type of device and study. The "ground truth" for the performance characteristics is established by adherence to the highly standardized and objective testing methods outlined in the ASTM and ISO standards. These standards themselves are developed by expert committees, but the immediate study does not involve human expert adjudication of individual test results in the way it would for, say, diagnostic image interpretation.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      None in the context of human expert adjudication. The "adjudication" is met by following the prescribed methodology of the ASTM and ISO standards. For instance, pinhole testing involves visual inspection of water-filled gloves, and if a leak is observed, it's a pinhole. There isn't an "expert consensus" on whether a pinhole exists; it's a direct observation.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted, and this type of study is not relevant for examination gloves. MRMC studies are typically performed for diagnostic devices (e.g., AI for radiology) where human interpretation is a key component. This device is a barrier protection product.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      Not applicable. This device is a physical product (gloves), not an algorithm or AI system. Therefore, standalone algorithm performance is not a relevant concept here.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      The "ground truth" is defined by the objective measurement and observation protocols outlined in the referenced ASTM and ISO standards. For example:

      • Dimensions: Measured against specified ranges in ASTM D3578.
      • Physical Properties: Tensile strength, elongation, etc., measured using ASTM D 412.
      • Freedom from pin-holes: Evaluated according to ASTM D 5151 (e.g., water leak test).
      • Powder Free Residue: Measured using ASTM D 6124.
      • Biocompatibility: Determined through specific in vivo (e.g., animal or human patch tests, though the document doesn't specify if human or animal for this submission) tests following ISO 10993-10 guidelines, where the "ground truth" is the observed biological response (e.g., absence of sensitization or irritation).
    7. The sample size for the training set:
      Not applicable. There is no "training set" as this is a physical product, not a machine learning model.

    8. How the ground truth for the training set was established:
      Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062917
    Date Cleared
    2006-11-02

    (35 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE MALAYSIA SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Sterile and Non-Sterile Powder-Free Latex Patient Examination Glove, with Extractable Protein Content Labeling Claim (50 microgram per Gram of Glove or Less)

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) clearance letter from the FDA for "Sterile and Non-Sterile Powder-Free Latex Patient Examination Glove, with Extractable Protein Content Labeling Claim (50 microgram per Gram of Glove or Less)".

    This document simply states that the device is substantially equivalent to legally marketed predicate devices and outlines general regulatory requirements. It does not contain any information about acceptance criteria for a study, reported device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided input. This type of information would typically be found in a separate performance study report or design validation documentation, not in an FDA clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062778
    Date Cleared
    2006-10-27

    (39 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE MALAYSIA SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Sterile Powder-Free Nitrile Patient Examination Glove, Blue Colored,
    Non-sterile Powder-free Nitrile Patient Examination Glove, Blue Colored,

    AI/ML Overview

    This document is a 510(k) premarket notification letter from the FDA for a medical device: "Sterile Powder-Free Nitrile Patient Examination Glove, Blue" and "Non-Sterile Powder-Free Nitrile Patient Examination Glove, Blue." This type of document does not contain the detailed acceptance criteria or the study that proves the device meets those criteria in the way described in the request.

    A 510(k) submission primarily demonstrates substantial equivalence to a legally marketed predicate device. The letter confirms that the FDA has reviewed the submission and found the device substantially equivalent for its intended use. It does not provide performance study details.

    Therefore, I cannot extract the requested information as it is not present in the provided text. The document confirms market clearance based on substantial equivalence, but it does not detail:

    1. A table of acceptance criteria or reported device performance.
    2. Sample sizes or data provenance for a test set.
    3. Number or qualifications of experts for ground truth.
    4. Adjudication method for a test set.
    5. MRMC comparative effectiveness study or human reader improvement with AI.
    6. Standalone (algorithm only) performance.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How the ground truth for the training set was established.

    This information would typically be found in the actual 510(k) submission, specifically sections related to performance data, or in a separate clinical study report, not in the FDA's clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992675
    Date Cleared
    2000-01-20

    (163 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WEAR SAFE MALAYSIA SDN. BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient Examination Powder glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Not Found

    AI/ML Overview

    This is an FDA 510(k) clearance letter for Patient Exam Nitrile Powdered Gloves. It is not a study of an AI/ML device, and therefore does not contain information about acceptance criteria or a study proving device performance as requested.

    The document is a regulatory approval for a medical glove, confirming its substantial equivalence to previously marketed devices. It does not involve any AI components or performance metrics suitable for the detailed questions about AI device studies.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2