(68 days)
A powder free nitrile examination glove is a disposable device made of synthetic material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants. In addition, this product is tested for use with chemotherapy drugs.
Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non Sterile (Chemotherapy Drug Protection Labeling Claim) will meet all the current specification for ASTM D6319.
Here's a breakdown of the acceptance criteria and study information for the "Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim)" based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance:
| Characteristic | Acceptance Criteria / Standard | Device Performance |
|---|---|---|
| Dimensions | ASTM D6319 - 00a(2005)e1 | Meets |
| Physical Properties | ASTM D 412 - 06ae1 | Meets |
| Freedom from pin-holes | ASTM D 5151 - 06 | Meets |
| Powder Free Residue | ASTM D 6124 - 06 | Meets |
| Biocompatibility | Dermal Sensitization (as per ASTM F720-81) | Not a contact skin sensitizer |
| Primary Skin Irritation Test (as per 16CFR Part1500) | Not a primary skin irritant | |
| Chemotherapy Drug Permeation (White Gloves) | Carmustine (BCNU) (3.3mg/ml) breakthrough time | 71.3 minutes |
| Cisplatin (1.0 mg/ml) breakthrough time | >240 minutes | |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) breakthrough time | >240 minutes | |
| Dacarbazine (DTIC) (10.0 mg/ml) breakthrough time | >240 minutes | |
| Doxorubicin Hydrochloride (2.0 mg/ml) breakthrough time | >240 minutes | |
| 5-Fluorouracil (50.0 mg/ml) breakthrough time | >240 minutes | |
| Etoposide (Toposar) (20.0 mg/ml) breakthrough time | >240 minutes | |
| Paclitaxel (Taxol) (6.0 mg/ml) breakthrough time | >240 minutes | |
| ThioTepa (10.0 mg/ml) breakthrough time | 72.36 minutes | |
| Chemotherapy Drug Permeation (Blue Gloves) | Carmustine (BCNU) (3.3mg/ml) breakthrough time | 60.7 minutes |
| Cisplatin (1.0 mg/ml) breakthrough time | >240 minutes | |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) breakthrough time | >240 minutes | |
| Dacarbazine (DTIC) (10.0 mg/ml) breakthrough time | >240 minutes | |
| Doxorubicin Hydrochloride (2.0 mg/ml) breakthrough time | >240 minutes | |
| 5-Fluorouracil (50.0 mg/ml) breakthrough time | >240 minutes | |
| Etoposide (Toposar) (20.0 mg/ml) breakthrough time | >240 minutes | |
| Paclitaxel (Taxol) (6.0 mg/ml) breakthrough time | >240 minutes | |
| ThioTepa (10.0 mg/ml) breakthrough time | 69.0 minutes |
(Note: The document states "Meets" for general characteristics, implying the device performance met the specified ASTM standards. For chemotherapy drug permeation, specific breakthrough times are provided as observed performance.)
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample size used for each specific test (e.g., number of gloves tested for pin-holes, or number of samples for each chemotherapy drug).
The data provenance is from Malaysia, as the submitting company, Wear Safe (Malaysia) Sdn. Bhd., is based in Malaysia. The studies appear to be prospective performance tests conducted on the device to demonstrate compliance with standards and for chemotherapy drug permeation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. This device is a physical product (examination gloves), and its performance is assessed against established physical and chemical standards (ASTM, 16CFR). The "ground truth" is defined by the technical specifications and test methodologies outlined in these standards, not by expert consensus in a clinical diagnostic sense. Therefore, no experts in the context of clinical interpretation or diagnosis were used to establish ground truth.
4. Adjudication Method for the Test Set:
Not applicable. As described above, the evaluation is based on objective measurements against established standards, not on subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No. An MRMC comparative effectiveness study is designed for evaluating diagnostic tools where human readers interpret medical images or data, often with and without AI assistance. This is not relevant for the performance evaluation of examination gloves.
6. If a Standalone (algorithm only without human-in-the-loop performance) was done:
Yes, in essence. The performance tests conducted for the gloves are "standalone" in the sense that they assess the intrinsic physical, chemical, and barrier properties of the glove material itself, without human intervention in the interpretation of results beyond standard laboratory procedures. There is no "algorithm" involved, but the device's performance is measured directly.
7. The Type of Ground Truth Used:
The ground truth used is based on established industry standards and regulatory specifications. Specifically:
- ASTM (American Society for Testing and Materials) Standards: For dimensions, physical properties, freedom from pin-holes, and powder-free residue.
- 16CFR Part 1500: For primary skin irritation.
- ASTM F720-81: For dermal sensitization.
- Chemotherapy Drug Permeation Testing: The "ground truth" for the chemotherapy drug claim is the measured breakthrough time of each specified drug through the glove material, which is an objective measurement based on a defined test methodology (though the specific standard for this permeation testing is not explicitly cited in the provided text, it is common practice to follow recognized standards like ASTM D6978).
8. The Sample Size for the Training Set:
Not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as no training set was used.
{0}------------------------------------------------
V S Wear Safe (Malaysia) Sdn.Bhd. (Company No.204396-X)
FDA 510(k), Premarket Notification : 510(k) Summary of Safety and Effectiveness Information
1.0 Submitter:
WEAR SAFE (MALAYSIA) SDN. BHD. Lot PT 13726, Jalan Haji Salleh, Off Jalan Meru, 41050 Klang, Selangor Darul Eshan, Malaysia.
+603 3392 3088 Telephone No .: +603 3392 2118 Fax No .:
2.0 Contact Person:
| Contact: | Mr. SH TAN |
|---|---|
| E-mail: | shtan@kossan.com.my |
| Telephone No.: | +603 3291 2657 |
| Fax No.: | +603 3291 0584 |
3.0 Name of Device:
.
.
.
Trade Name:
Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored). Non-Sterile (Chemotherapy Drug Protection Labeling Claim) Patient Examination Glove Common Name: Classification Name: Patient Examination Glove.
Identification of The Legally Marketed Device: 4:0
The Powder Free Nitrile Examination Gloves, White (non colored) or Blue (colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim), Class I Patient Examination gloves, 80LZC. meets all of the requirements of ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical Application.
5.0 Description of Device:
Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non Sterile (Chemotherapy Drug Protection Labeling Claim) will meet all the current specification for ASTM D6319.
Lot 13726, Jalan Haji Salleh, Batu 5% Off Jalan Meru, 41050 Klang, Selangor Darul Ehsan Malaysia. Tel: 603-33923088, Fax: 603-33922599
Page 1 of 3
{1}------------------------------------------------
Wear Safe (Malaysia) Sdn.Bhd. (Company No.204396-X)
6.0 Intended Use of the Device:
Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non Sterile (Chemotherapy Drug Protection Labeling Claim) is a disposable device intended for medical purposes that is intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants. In addition, this product is tested for use with chemotherapy drugs.
7.0 Summary of The Technological Characteristics of The Device:
Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non Sterile (Chemotherapy Drug Protection Labeling Claim) possesses the following technological characteristic (as compared to ASTM or equivalent standards):
| Characteristic | Standards | Device Performance |
|---|---|---|
| Dimensions | ASTM D6319 -00a(2005)e1 | Meets |
| Physical Properties | ASTM D 412 - 06ae1 | Meets |
| Freedom from pin-holes | ASTM D 5151 - 06 | Meets |
| Powder Free Residue | ASTM D 6124 - 06 | Meets |
| Biocompatibility | Dermal Sensitization(as per ASTM F720-81)Primary Skin Irritation Test(as per 16CFR Part1500) | Not a contact skin sensitizerNot a primary skin irritant |
8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
The performance test data that support a determination of substantial equivalence are described above.
9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data are not needed for market cleared examination gloves.
Lot 13726, Jalan Haji Salleh, Batu 5% Off Jalan Meru, 41050 Klang, Selangor Darul Ehsan Malaysia. Tel: 603-33923088, Fax: 603-33922599
{2}------------------------------------------------
10.0 Conclusion
It can be concluded that the Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored) or, Non Sterile (Chemotherapy Drug Protection Labeling Claim) is safe and effective for use and will perform according to the glove performance standards referenced in Section 7 above, thereby meeting ASTM standards, FDA requirements, and the labeling claims for the product.
Consequently, this device is substantially equivalent to current marketed devices. This summary will include any other information reasonably deemed necessary by the FDA.
Lot 13726, Jalan Haji Salleh, Batu 5¼ Off Jalan Meru, 41050 Klang, Selangor Darul Ehsan Malaysia. Tel: 603-33923088, Fax: 603-33922599
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure with outstretched arms, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
JAN 1 9 2010
Mr. S.H. Tan Regulatory Affairs Department Manager Wear Safe MALAYSIA Sdn. Bhd. Lot PT 13726, Jalan Haji Salleh off Jalan Meru Klang Selangor Darul Ehsan MALAYSIA 41050
Re: K093500
Trade/Device Name: Powder Free Nitrile Examination Gloves, White (non-colored) or Bule (colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim) Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZC Dated: December 23, 2009 Received: December 29, 2009
Dear Mr. Tan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device. Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties: We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Susan Gunner
Anthony D. Watson, BS, MS, MBA Director
Division of Anesthesiology, General Hospital, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Wear Safe (Malaysia) Sdn.Blid. (Company No.204396-X)
revised copy
INDICATIONS FOR USE
Applicant: WEAR SAFE (MALAYSIA) SDN. BHD.
510(k) Number (if known): K093500
Device Name: Powder Free Nitrile Examination Gloves, White (non-colored) or Blue (colored), Non-Sterile (Chemotherapy Drug Protection Labeling Claim)
Indication For Use: A powder free nitrile examination glove is a disposable device made of synthetic material intended to be worn on the hand for medical purposes to provide barrier against potentially infectious materials and other contaminants. In addition, this product is tested for use with chemotherapy drugs.
| Powder Free Nitrile Patient Examination Gloves, White (Non-Colored) Non Sterile (Chemotherapy Drugs | |
|---|---|
| Protection Claim): | |
| Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes | |
| The following chemotherapy drugs have been tested: | |
| Carmustine (BCNU) (3.3mg/ml) | 71.3 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240 |
| Dacarbazine (DTIC) (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| 5-Fluorouracil (50.0 mg/ml) | >240 |
| Etoposide (Toposar) (20.0 mg/ml) | >240 |
| Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| ThioTepa (10.0 mg/ml) | 72.36 |
Powder Free Nitrile Patient Examination Gloves, Blue (Colored) Non sterile (Chemotherapy Drugs Protection Claim):
Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes The following chemotherapy drugs have been tested:
| The following chemotherapy drugs have been tested. | |
|---|---|
| Carmustine (BCNU) (3.3mg/ml) | 60.7 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240 |
| Dacarbazine (DTIC) (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| 5-Fluorouracil (50.0 mg/ml) | >240 |
| Etoposide (Toposar) (20.0 mg/ml) | >240 |
| Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| ThioTepa (10.0 mg/ml) | 69.0 |
. Lot 13726, Jalan Haji Salleh, Batu 5% Off Jalan Meru, 41050 Klang, Selangor Darul Ehsan Malaysia. Tel: 603-33923088, Fax: 603-33922599
{6}------------------------------------------------
KD93500
WSM Wear Safe (Malaysia) Sdn.Bhd. (Company No.204396-X)
revised copy
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth P. Clavier Will
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number. K093500
Page 2 of 2
Lot 13726, Jalan Haji Salleh, Batu 5% Off Jalan Meru, 41050 Klang, Selangor Darul Ebsan Malaysia. Tel: 603-33923088, Fax: 603-33922599
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.