Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K983512
    Manufacturer
    Date Cleared
    1999-04-28

    (203 days)

    Product Code
    Regulation Number
    876.5665
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WaterVault Cube Water Treatment System for HemoDialysis is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate.

    Device Description

    Organic, inorganic and microbial contaminants are removed from water supplied to hemodialysis clinics for use in the dilution and preparation of dialysate solutions used in hemodialysis machines. The water treatment system includes a means of tempering the water and pretreatment, followed by reverse osmosis, ultrafiltration and UV sterilization. Pretreatment includes water softening to remove hardness and Granular Activated Carbon treatment to remove Chloramines in addition to filament cartridge filters to control particulate solids. Reverse Osmosis utilizes a pressure gradient to reverse the process water diffusion across a semipermeable membrane, concentrating the contaminants which are subsequently rejected as concentrate. Water forced through the membrane known as "Permeate" is then passed through a hollow fiber ultra filter to remove microorganisms, pyrogens and endotoxins. A UV light source that provides a minimum of 30,000 microwatts/cm2 sec limits further bacterial growth in the treated water.

    AI/ML Overview

    The Water Vault Cube, Water Treatment System for Hemodialysis, underwent nonclinical testing to demonstrate its performance against AAMI (Association for the Advancement of Medical Instrumentation) standards for water used in the preparation of dialysate.

    1. Table of Acceptance Criteria and Reported Device Performance

    ComponentAcceptance Criteria (AAMI Standard mg/l)Reported Device Performance (Permeate mg/l)Meets or Exceeds Acceptance Criteria
    Calcium21Yes
    Magnesium41Yes
    Sodium701Yes
    Potassium81Yes
    Fluoride0.20.1Yes
    Chlorine0.50.02Yes
    Chloramines0.10.02Yes
    Nitrate20.1Yes
    Sulfate1005Yes
    Copper0.10.01Yes
    Barium0.10.01Yes
    Zinc0.10.02Yes
    Aluminum0.010.01Yes
    Arsenic0.0050.005Yes
    Lead0.0050.003Yes
    Silver0.0050.005Yes
    Cadmium0.0010.001Yes
    Chromium0.0140.005Yes
    Selenium0.050.005Yes
    Mercury0.00020.0002Yes

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The test results are based on a single sample test of a "demonstration unit."
    • Data Provenance: The data is retrospective, derived from tests performed on a specific demonstration unit (Osmonics Solorss reverse osmosis unit, Serial Number 96-61650-11). The country of origin is not explicitly stated but implied to be the United States, given the FDA filing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this device is based on quantifiable chemical analysis against established industry standards (AAMI Standards), not expert human assessment.

    4. Adjudication Method for the Test Set

    Not applicable. The device performance is assessed through direct chemical analysis (quantitative measurements), not through human adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a water treatment system, not an AI-powered diagnostic device, so an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone test was performed. The device's performance was evaluated by analyzing the "Permeate" water it produced, independently of any human intervention or interpretation, against the AAMI standards.

    7. The Type of Ground Truth Used

    The ground truth used is AAMI Standards for water quality in hemodialysis. This is an objective, established standard for chemical contaminant levels.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical water treatment system, not a machine learning algorithm that requires a training set. The "design" of the system is based on established water purification principles.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI/ML device, there is no training set or ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1