510(k) Data Aggregation

    K Number
    K253399
    Date Cleared
    2025-10-28

    (28 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Visions® PV.014P RX Digital IVUS Catheter:

    The Visions® PV.014P RX Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in the cerebral vessels.

    The Visions® PV.014P RX Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

    Visions® PV .018 Digital IVUS Catheter:

    The Visions® PV .018 Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels.

    The Visions® PV .018 Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

    Device Description

    The Visions® PV .014P RX Digital IVUS Catheter and the Visions® PV .018 PV .018 Digital IVUS Catheter are being bundled into one submission per FDA Guidance for Industry: Bundling Multiple Devices or Multiple Indications in a Single Submission, June 2007 (FDA-2003-D-0376), as the two catheters share the same indications for use, FDA classification product codes, and technological characteristics, and part of the same product line of catheters.

    Visions® PV .014P RX Digital IVUS Catheter

    The Visions® PV .014P RX Digital IVUS Catheter incorporates a cylindrical ultrasound transducer array located near the distal tip of the catheter. The array radiates acoustic energy into the surrounding tissue and detects the subsequent ultrasonic echoes. The information from the echoes is used to generate real-time images of the coronary or peripheral vessels.

    The Visions® PV .014P RX Digital IVUS Catheter utilizes an internal lumen that allows the catheters to track over a 0.018" (0.46 mm) guide wire. The guide wire exits from the guide wire lumen approximately 31 cm proximal to the catheter tip. The PV .018 catheters are introduced either percutaneously or via surgical cut down into the vascular system.

    The Visions® PV .014P RX Digital IVUS Catheter may only be used with Volcano imaging systems, such as the Volcano s5™, Volcano s5i™, Volcano CORE Mobile, and Volcano CORE imaging systems. The catheter will not operate if connected to any other imaging system.

    Visions® PV .018 Digital IVUS Catheter

    The Visions® PV .018 Digital IVUS Catheter incorporates a cylindrical ultrasound transducer array located near the distal tip of the catheter. The array radiates acoustic energy into the surrounding tissue and detects the subsequent ultrasonic echoes. The information from the echoes is used to generate real-time images of the coronary or peripheral vessels.

    The Visions® PV .018 Digital IVUS Catheter utilizes an internal lumen that allows the catheters to track over a 0.018" (0.46 mm) guide wire. The guide wire exits from the guide wire lumen approximately 31 cm proximal to the catheter tip. The PV .018 catheters are introduced either percutaneously or via surgical cut down into the vascular system.

    The Visions® PV .018 Digital IVUS Catheter may only be used with Volcano imaging systems, such as the Volcano s5™, Volcano s5i™, Volcano CORE Mobile, and Volcano CORE imaging systems. The catheter will not operate if connected to any other imaging system.