Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233058
    Device Name
    VivaChek Link Plus Blood Glucose Monitoring System
    Manufacturer
    VivaChek Biotech (Hangzhou) Co., Ltd.
    Date Cleared
    2023-12-22

    (88 days)

    Product Code
    NBW
    Regulation Number
    862.1345
    Why did this record match?
    Applicant Name (Manufacturer) :

    VivaChek Biotech (Hangzhou) Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    VivaChek™ Link Plus Blood Glucose Monitoring System is comprised of the VivaChek™ Link Plus Blood Glucose Meter and the VivaChek™ Ino Blood Glucose Test Strips. VivaChek™ Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.
    Device Description
    VivaChek Link Plus Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed. VivaChek Link Plus Blood Glucose Monitoring System contains 4G module, the device complies with US federal guidelines, FCC Part 15 Subpart B, FCC Part 2, FCC Part 24 Subpart E, FCC Part 27 Subpart C, and FCC 47 CFR§ 2.1093 based on the test reports.
    Ask a Question