K Number
K233058

Validate with FDA (Live)

Date Cleared
2023-12-22

(88 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VivaChek™ Link Plus Blood Glucose Monitoring System is comprised of the VivaChek™ Link Plus Blood Glucose Meter and the VivaChek™ Ino Blood Glucose Test Strips.

VivaChek™ Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Device Description

VivaChek Link Plus Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

VivaChek Link Plus Blood Glucose Monitoring System contains 4G module, the device complies with US federal guidelines, FCC Part 15 Subpart B, FCC Part 2, FCC Part 24 Subpart E, FCC Part 27 Subpart C, and FCC 47 CFR§ 2.1093 based on the test reports.

AI/ML Overview

The provided text describes the VivaChek™ Link Plus Blood Glucose Monitoring System, an in-vitro diagnostic device. Here's an analysis of the acceptance criteria and study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of quantitative acceptance criteria and corresponding reported device performance metrics in the format requested. Instead, it states that various laboratory studies were performed, and for these studies, "the test results indicated that the acceptance criteria were met."

Therefore, based on the provided text, a generic representation of this would be:

Acceptance Criteria CategorySpecific Acceptance Criteria (Not explicitly detailed in text)Reported Device Performance
Accuracy (User Evaluation)[e.g., ISO 15197:2013 standards for accuracy]"able to obtain blood glucose readings"
Precision[e.g., CV% within specified limits]"acceptance criteria were met"
Linearity[e.g., R-squared > 0.99]"acceptance criteria were met"
Interference[e.g., no significant bias from common interferents]"acceptance criteria were met"
Usability/Ease of Operation[e.g., users can successfully perform tests, satisfaction]"satisfied with the ease of operation... and the overall performance"
Stability[e.g., device functions effectively over time]"acceptance criteria were met"
Environmental Conditions[e.g., functional across specified temperature/humidity]"acceptance criteria were met"
Safety (Electrical, EMC, etc.)[e.g., complies with relevant standards]"acceptance criteria were met"
Cybersecurity[e.g., controls in place, management plan]"FCC, cybersecurity control DFMEA, Cybersecurity Management Plan and Web App software validation were conducted."

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document mentions "clinical study (user evaluation) was conducted with intended users." However, it does not specify the number of users or samples included in this evaluation.
  • Data Provenance: The document does not explicitly state the country of origin for the data. It also directly states that the user evaluation was conducted as a "clinical study," implying it was a prospective study, though not explicitly labelled as such. The other laboratory studies listed are implicitly prospective tests performed by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. For a blood glucose monitoring system, the "ground truth" for glucose levels would typically be established using a laboratory reference method (e.g., a YSI analyzer), not by human experts interpreting results. The user evaluation focuses on the usability and ability of lay persons to obtain readings, not on expert interpretation of the device's output.

4. Adjudication method for the test set

This information is not applicable/provided in the context of this device. Adjudication methods (like 2+1, 3+1) are typically used in studies where multiple human readers interpret images or complex data to establish a consensus ground truth, often for AI-assisted diagnostic tools. For a blood glucose meter, the reference method (e.g., YSI) provides the objective ground truth, and device readings are compared directly to this.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for imaging or diagnostic AI where human readers interact with AI assistance. The VivaChek™ Link Plus Blood Glucose Monitoring System is a standalone measurement device for self-testing; it does not involve human readers interpreting AI-generated insights.
  • Effect size of improvement with AI vs without AI assistance is not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone performance evaluation was done. The entire suite of laboratory studies (e.g., precision, linearity, interference, stability, environmental, safety) represents the standalone performance evaluation of the device system (meter and test strips) without human interpretation in the loop. The "user evaluation" section primarily assesses usability rather than the intrinsic accuracy of the algorithm/device itself. The device is designed for "self-testing outside the body (in vitro diagnostic use)," operating without professional human-in-the-loop interpretation of its results for diagnosis.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the specific reference method used for establishing ground truth for glucose levels. However, in the context of blood glucose monitoring systems, the ground truth is typically established using a laboratory reference method, such as a YSI glucose analyzer, which provides highly accurate and precise glucose concentration measurements. This is an objective chemical measurement, not based on expert consensus, pathology, or outcomes data.

8. The sample size for the training set

This information is not provided and is generally not applicable in this context. Blood glucose monitoring systems are typically designed and calibrated based on biochemical principles and rigorous testing, rather than "training" a machine learning algorithm with a dataset in the way an AI diagnostic tool would be trained. The "training set" concept is usually associated with AI/ML development.

9. How the ground truth for the training set was established

As noted above, a "training set" in the AI/ML sense is likely not applicable here. Device calibration and verification are based on established analytical chemistry methods where the "ground truth" for glucose concentrations in control solutions or reference samples would be determined using highly accurate laboratory reference instruments.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 22, 2023

VivaChek Biotech (Hangzhou) Co., Ltd. Mark Qian Quality Director Level 2, Block 2, 146 East Chaofeng Rd. Yuhang Economy Development Zone Hangzhou, Zhejiang 311100 China

Re: K233058

Trade/Device Name: VivaChek™ Link Plus Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: September 20, 2023 Received: September 25, 2023

Dear Mark Oian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yiduo Wu -S

for Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233058

Device Name

VivaChek™ Link Plus Blood Glucose Monitoring System

Indications for Use (Describe)

VivaChek™ Link Plus Blood Glucose Monitoring System is comprised of the VivaChek™ Link Plus Blood Glucose Meter and the VivaChek™ Ino Blood Glucose Test Strips.

VivaChek™ Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section F

Section F: 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The Assigned 510(k) number is K233058.

Submitter's Identification:

VivaChek Biotech (Hangzhou) Co., Ltd Level 2, Block 2, 146 East Chaofeng Rd., Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China.

Date Updated: Dec 19, 2023

Contact Person:

Name: Mark Qian Position: Quality Director Email: mark.qian@vivachekbio.com

Proprietary Name of the Device:

VivaChek Link Plus Blood Glucose Monitoring System

Common Name: Glucose Test System

Classification Name:

Class II §862.1345 Glucose Test System Product Code: NBW

Predicate Device:

VivaChek Ino Smart Blood Glucose Monitoring System VivaChek Laboratories, Inc. 510(k) Number: K173140

{4}------------------------------------------------

Applicant: VivaChek Biotech (Hangzhou) Co., Ltd Address: Level 2, Block 2, 146 East Chaofeng Rd., Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China Contact: Mark Qian Email: mark.qian@vivachekbio.com

Section F

Device Name:
Proprietary NameMeterModelTest StripsModelClassificationProductCodeDescriptionCommonName
VivaChek Link PlusBlood GlucoseMonitoring SystemVGM90VGS01862.1345Class IINBWSystem, Test,Blood Glucose,Over The CounterGlucoseTest System

Intended Use:

VivaChek Link Plus Blood Glucose Monitoring System is comprised of the VivaChek Link Plus Blood Glucose Meter and the VivaChek Ino Blood Glucose Test Strips. VivaChek Link Plus Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Description:

VivaChek Link Plus Blood Glucose Monitoring System is designed to quantitatively measure the glucose concentration in fresh capillary whole blood. The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose oxidase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

VivaChek Link Plus Blood Glucose Monitoring System contains 4G module, the device complies with US federal guidelines, FCC Part 15 Subpart B, FCC Part 2, FCC Part 24 Subpart E, FCC Part 27 Subpart C, and FCC 47 CFR§ 2.1093 based on the test reports.

{5}------------------------------------------------

Section F

Comparison to Predicate Device:

The VivaChek Link Plus Blood Glucose Monitoring System is substantially equivalent to VivaChek Ino Smart Blood Glucose Monitoring System (K173140).

FeaturesPredicate: VivaChek Ino Smart BloodGlucose Monitoring System (K173140)Candidate: VivaChek LinkPlus Blood GlucoseMonitoring System
Similarities
Intended UseIt is intended to quantitatively measurethe glucose concentration in freshcapillary whole blood samples drawnfrom the fingertips. It is intended for useby persons with diabetes at home as anaid to monitor the effectiveness ofdiabetes control. It is not intended forneonatal use or for the diagnosis of orscreening for diabetes. It is intended forself-testing outside the body (in vitrodiagnostic use), and should only beused by a single person and should notbe shared.Same
Operation PrincipleElectrochemical biosensorSame
Detection MethodAmperometricSame
Strip ChemicalCompositionGlucose oxidaseSame
Measurement ResultPlasma equivalentSame
SampleFresh capillary whole bloodSame
Memory500 recordsSame
Unit of Measuremg/dLSame
Measurement Range20-600 mg/dLSame
Sample Volume0.8µLSame
Test Time5 secondsSame
Operating RelativeHumidity10-90% (non-condensing)Same
Operating Temperature41-113°FSame
Hematocrit Range20-70%Same
Automatic Shutoff2 minutes after last actionSame
Power SourceRechargeable 3.7 Volt Lithium IonbatterySame
Differences
Battery TypeRechargeable, non-serviceable,250mAh, 3.7 Volt DC nominal, lithiumRechargeable, 800 mAh, 3.7Volt DC nominal, lithiumpolymer battery (5V input)
polymer battery (5V input chargevoltage)charge voltage)
Data TransmissionBluetooth4G
Dimensions83 mm x 52 mm x 18.7 mm100.9mm x 61.6mm x 23.7mm
Display Size32mm x 32 mm42 mm x 41 mm
WeightApproximately 53gApproximately 85g

{6}------------------------------------------------

Applicant: VivaChek Biotech (Hangzhou) Co., Ltd

Address: Level 2, Block 2, 146 East Chaofeng Rd., Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China Contact: Mark Qian Email: mark.qian@vivachekbio.com

Section F

Laboratory and Clinical Testing:

The performance characteristics of the VivaChek Link Plus Blood Glucose Monitoring System were evaluated by performing the following studies:

No.Test/Validation Item
1Within-Run Precision Evaluation
2Intermediate Precision Evaluation
3Linearity Evaluation Study
4User Evaluation
5User Evaluation - Accuracy at Extreme Glucose Values
6Usability Evaluation
7Interference Agents Study
8Accelerated Closed Vial Stability Study
9Accelerated Open Vial Stability Study
10System Operating Conditions Evaluation
11Altitude Effect Evaluation
12Error Messages Validation
13Short Sample Detection Study
14Sample Perturbation Study
15Intermittent Sampling Study
16Testing with Used Test Strips
17Test Strip Early Removal Validation
18Incorrect Strip Insertion Validation

{7}------------------------------------------------

Applicant: VivaChek Biotech (Hangzhou) Co., Ltd

Address: Level 2, Block 2, 146 East Chaofeng Rd.,

Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China

Contact: Mark Qian Email: mark.qian@vivachekbio.com

Section F

19Shipping and Handling Study
20Oxygen Interference Study
21Hematocrit Effect Study
22Meter Environmental Temperature Test
23Vibration Testing
24Shock Testing
25Low Battery Study
26Recharging Temperature Limit Validation
27Cleaning and Disinfection Study
28Meter Robustness Study
29EMC Testing including the wireless connection testing
30Electrical Safety Testing
31FCC Testing
32Meter Software (Firmware) Validation
33Cybersecurity Control DFMEA
34Cybersecurity Management Plan

To confirm the 4G module has not brought any unexpected functional failure or adverse effect, FCC, cybersecurity control DFMEA, Cybersecurity Management Plan and Web App software validation were conducted.

Discussion of Laboratory Studies:

Above laboratory studies were performed on VivaChek Link Plus Blood Glucose Monitoring System in accordance with the applicable guidance or standards, and the test results indicated that the acceptance criteria were met. Therefore, the performances from these laboratory studies were acceptable.

{8}------------------------------------------------

Section F

Discussion of Clinical Study:

Clinical study (user evaluation) was conducted with intended users using the VivaChek Link Plus Blood Glucose Monitoring System. Study results indicated that non-professional, inexperienced lay persons were able to obtain blood glucose readings when using the VivaChek Link Plus Blood Glucose Monitoring System. In addition, the participated lay persons were questioned and responded as satisfied with the ease of operation by following the User Manual and the overall performance of the VivaChek Link Plus Blood Glucose Monitoring System.

Conclusion:

The laboratory studies and user evaluation study results demonstrate that the VivaChek Link Plus Blood Glucose Monitoring System is safe, effective and easy-to-use. It also demonstrates that the VivaChek Link Plus Blood Glucose Monitoring System meets FDA Guidance SMBG for OTC Use.

Based on the same intended use and work principle, and the same technological characteristics listed in the section Comparison to Predicate Device - Similarities, meanwhile the different technological characteristics listed in the section Comparison to Predicate Device do not raise safety and effectiveness questions according to the completed performance testing and validation reports, therefore the candidate device VivaChek Link Plus Blood Glucose Monitoring System is substantially equivalent to the predicate device VivaChek Ino Smart Blood Glucose Monitoring System (K173140).

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.