Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K253226
    Device Name
    UV5000W Handle
    Manufacturer
    Date Cleared
    2025-10-23

    (24 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UV5000W Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics.

    Device Description

    The UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V connect to the UV5000W Handle to become an endoscopic system that is intended to allow visualization of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics. This 510(k) discloses the modifications made to the UV5000 Handle (K250797) with the capability of transmitting the images captured wirelessly to an external monitor.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250797
    Device Name
    UV5000 Handle
    Manufacturer
    Date Cleared
    2025-09-11

    (181 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UV5000 Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics.

    Device Description

    The Uro-G Cystoscope System is comprised of a cannula and handle that is intended to allow visualization of the urethra, bladder, cervical canal, and uterine cavity for the purpose of performing diagnostics. This 510(k) discloses the modifications made to the handle of the Uro-G HD Cystoscope System (K232837), UV5000 Handle. The Uro-G HD Flexible Cystoscope System is a handheld, battery-operated portable cystoscope consisting of a disposable steerable endoscopic cannula with a 2.2 mm channel for fluid infusion and instrument advancement, and a reusable handle with a video monitor.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243196
    Manufacturer
    Date Cleared
    2024-12-13

    (73 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-G HD Flexible Cystoscope is intended for endoscopic diagnosis and infusion of irrigating fluids within the bladder and urethra.

    Device Description

    This Special 510(k) discloses the modifications to the steps to carry out surface disinfection of the model 4500 handle that was reviewed and cleared in K232837. The Uro-G HD Flexible Cystoscope is a handheld, battery- operated portable cystoscope consisting of a disposable steerable endoscopic cannula with a 2.2 mm channel for fluid infusion and instrument advancement, and a reusable handle with a video monitor. When fully assembled, the Uro-G HD Flexible Cystoscope- cannula and handle-has an overall length of 630 mm (approximately 25 inches). The disposable cannula's working length is 380.2 mm (approx. 15 inches) and a total length of 406 mm, from tip to connector. The cannula's maximum width is 5.6 mm and it has a connector for attaching and detaching to the handle.

    AI/ML Overview

    This document describes a Special 510(k) submission for the Uro-G HD Flexible Cystoscope, focusing on a minor modification to its surface disinfection process. There is NO mention of an AI/ML device or algorithm. Therefore, I cannot provide information on acceptance criteria and a study proving the device meets them in the context of AI/ML.

    The provided text only discusses the equivalence of the modified device to its predicate, based on unchanged intended use, indications for use, and technological characteristics, except for the disinfection process.

    Here's what I can extract regarding the device performance and the "study" mentioned, which is not an AI/ML study but rather microbiology testing for disinfection:

    1. A table of acceptance criteria and the reported device performance:

    Since the document focuses on a change in the disinfection process and asserts that other performance characteristics remain unchanged from the predicate, and does not provide specific acceptance criteria values for the disinfection, I cannot create a table with specific numbers. Instead, I can describe the nature of the change and the demonstrated outcome.

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Surface Disinfection• Efficacy: Achieve sufficient log reductions of tested pathogens. • Compatibility: Allow the use of Super Sani-Cloth or CaviWipe towelettes. • Process Efficiency: Potentially reduce contact and air-dry times while maintaining disinfection efficacy.Efficacy: "Testing found that 3 minutes contact with the disinfecting agent on the wipes was sufficient for log reductions of ≥ 3 and ≥ 6 of tested pathogens." • Compatibility: The Super Sani-Cloth is now an approved option in addition to CaviWipe. • Process Efficiency: Contact time reduced from 6 minutes to 3 minutes, and air dry time reduced from 20 minutes to 5 minutes. The document implies these reduced times are now sufficient for effective disinfection.
    Other Performance(Implicitly aligned with Predicate Device's performance, as no changes were made) • Tip angulation • Tip angulation radius • Lever force for tip angulation • Rotation torque • Rotation angles • Distention flow rate • Cannula leaking • Tip temperature • Field of view • Direction of view • Image quality • Electrical Safety (IEC 60601-2-18 compliant) • EMC Safety (IEC 60601-1-2 and IEC 60601-2-18 compliant)"Performance tests reviewed in K232837 were not repeated for this submission since the minor change in the surface disinfection process of the handle does not affect device performance." The device "continues to meet the compliance requirements" for electrical safety and EMC.

    Regarding the study that proves the device meets the acceptance criteria:

    The study referenced is a microbiology testing study conducted to evaluate the effectiveness of the revised surface disinfection process for the reusable handle of the Uro-G HD Flexible Cystoscope.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size: Not specified in the provided text. The document refers to "testing" that "found" specific results for log reductions.
    • Data provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This is not applicable to a microbiology disinfection study. The "ground truth" would be established by the standardized protocols and results of the microbiological tests themselves, rather than expert consensus on interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable to a microbiology disinfection study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, and no MRMC study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the disinfection efficacy, the "ground truth" was established through microbiology testing results demonstrating "log reductions of ≥ 3 and ≥ 6 of tested pathogens." This is a quantitative measure of antimicrobial effectiveness.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device; there is no training set mentioned.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device.

    In summary, the provided document describes a regulatory submission for a medical device (a cystoscope) focusing on a change in its disinfection instructions. The "study" mentioned is a microbiology test to validate the effectiveness of the new disinfection protocol. There is no information related to AI/ML or an algorithm in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232837
    Manufacturer
    Date Cleared
    2024-06-07

    (267 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-G HD Flexible Cystoscope is intended for endoscopic diagnosis and infusion of irrigating fluids within the bladder and urethra.

    Generally recognized indications for cystoscopy include:
    • Symptomatic voiding dysfunction
    • Hematuria
    • Bladder tumor surveillance
    • Recurrent lower urinary tract infections
    • Pelvic pain syndromes

    Device Description

    The Uro-G HD Flexible Cystoscope System is a handheld, battery-operated portable cystoscope consisting of a disposable steerable endoscopic cannula with a 2.2 mm channel for fluid infusion and instrument advancement, and a reusable handle with a video monitor. The fully assembled Uro-G HD Cystoscope System—cannula and handle—has an overall length of 630 mm (approximately 25 inches). The disposable cannula's working length is 380.2 mm (approx. 15 inches) and a total length of 406 mm, from tip to connector. The cannula's maximum width is 5.6 mm and it has a connector for attaching and detaching to the handle.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Uro-G HD Flexible Cystoscope, seeking to prove substantial equivalence to a predicate device (Uro-G Flexible Cystoscope, K202921). As such, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to the demonstration that the modified device (Uro-G HD) is as safe and effective as the predicate device, not necessarily a performance study against a specific clinical endpoint. The primary method for proving this is through comparative testing and analysis of technological characteristics.

    Here's an analysis of the provided text in the context of acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative targets in a table format within the provided text. Instead, the document uses a comparative approach against a predicate device. The "reported device performance" is essentially the Uro-G HD's characteristics, and "meeting acceptance criteria" means demonstrating that these characteristics are equivalent to or improve upon the predicate without introducing new risks.

    Here’s a table summarizing the key characteristics and the comparison to the predicate device, highlighting what could be implicitly considered an "acceptance criterion" (i.e., equivalence or improvement):

    CharacteristicPredicate Device (Uro-G, K202921) PerformanceSubject Device (Uro-G HD) PerformanceAcceptance Criteria (Implicit)Met?
    Cannula
    Working length380.2 mm380.2 mmNot Changed (Equivalent)Yes
    Total length (tip to connector)406 mm406 mmNot Changed (Equivalent)Yes
    Number of ports22Not Changed (Equivalent)Yes
    Working channel minimum width2.2 mm2.2 mmNot Changed (Equivalent)Yes
    Maximum insertion width (OD)5.5 mm5.6 mmEquivalent (Increased by 0.1mm, deemed not to introduce new risk, as human urethral diameter is larger)Yes
    Tip OD4 mm4 mmNot Changed (Equivalent)Yes
    Camera
    CMOS module pixels0.3 million pixels0.7 million pixelsEquivalent (Higher pixels for HD imaging, does not introduce new risk)Yes
    Focal length5 mm to 50 mm5 mm to 50 mmNot Changed (Equivalent)Yes
    Field of view (in air)120 ° ± 5 °100 ° ± 5 °Equivalent (Smaller FOV but sufficient for HD imaging in confined spaces, does not introduce new risk)Yes
    Direction of view from center axis0 ° (Forward Viewing)0 ° (Forward Viewing)Not Changed (Equivalent)Yes
    Image resolution3 lp/mm on 1551 USAF test chart in air3 lp/mm on 1551 USAF test chart in airNot Changed (Equivalent)Yes
    Image distortion< 15% at x and y direction< 15% at x and y directionNot Changed (Equivalent)Yes
    Sensor on CMOS camera640 x 480 pixels1280 x 720 pixelsEquivalent (Higher resolution for HD, does not introduce new risk or increase established risk)Yes
    Tip Angulation
    RangeUp 210 ° ± 5 °, Down 130 ° ± 5 °Up 210 ° ± 5 °, Down 130 ° ± 5 °Not Changed (Equivalent)Yes
    RadiusUp < 15 mm, Down < 20 mmUp < 15 mm, Down < 20 mmNot Changed (Equivalent)Yes
    Lever force for angulation< 2.2 lbs< 2.2 lbsNot Changed (Equivalent)Yes
    Rotation torque< 3.5 inch-lb< 3.5 inch-lbNot Changed (Equivalent)Yes
    Rotation angle< 90 ° (clockwise and counter-clockwise)< 90 ° (clockwise and counter-clockwise)Not Changed (Equivalent)Yes
    Fluid & Light
    Fluid lumen cross-section4.84 mm²4.84 mm²Not Changed (Equivalent)Yes
    Distension flow through channel> 120 mL/min> 120 mL/minNot Changed (Equivalent)Yes
    Light sourceLED, 520 lumens at 25 mm focal lengthLED, 520 lumens at 25 mm focal lengthNot Changed (Equivalent)Yes
    Tip temperature when LEDs are on< 41° C< 41° CNot Changed (Equivalent)Yes
    Power rating of source90 mW90 mWNot Changed (Equivalent)Yes
    Sterilization & Biocompatibility
    Provided SterileYesYesNot Changed (Equivalent)Yes
    Sterilization methodEthylene oxideEthylene oxideNot Changed (Equivalent)Yes
    Sterility Assurance Level10^-610^-6Not Changed (Equivalent)Yes
    Biocompatibility of MaterialsMeets ISO 10993 requirementsMeets ISO 10993 requirementsNot Changed (Equivalent)Yes
    Single UseYesYesNot Changed (Equivalent)Yes
    Handle
    Device Identity35004500N/A (New model but functional equivalence demonstrated)Yes
    Length135 mm176 mmEquivalent (Increased length does not affect performance)Yes
    Diameter37 mm36 mmEquivalent (1 mm reduction does not impact risk or performance)Yes
    Weight220 g264 gEquivalent (Increased weight does not introduce new risk or affect usability)Yes
    Video Monitor
    Screen size3.5 inches diagonal4.3 inches diagonalEquivalent (Larger size makes viewing easier, no new risk)Yes
    Input formatRGBRGBNot Changed (Equivalent)Yes
    Effective pixels640 x 4801280 x 720Equivalent (Higher resolution for HD, no new risk)Yes
    LCD resolution480 x 320800 x 480Equivalent (Higher resolution for HD, no new risk, sharper image)Yes
    LCD display resolution0.15 million pixels0.4 million pixelsEquivalent (Higher resolution for HD, no new risk, sharper image)Yes
    Power
    Power supply18650 Lithium Ion rechargeable18650 Lithium Ion rechargeableNot Changed (Equivalent)Yes
    Voltage of fully charged battery3.7 V3.7 VNot Changed (Equivalent)Yes
    Operating time after fully charged2 hours2 hoursNot Changed (Equivalent)Yes
    Battery capacity3100 mA3100 mANot Changed (Equivalent)Yes
    Operational Characteristics
    Electrical SafetyIEC 60601-2-18 compliantIEC 60601-2-18 compliantNot Changed (Equivalent)Yes
    EMCIEC 60601-1-2 compliantIEC 60601-1-2 compliantNot Changed (Equivalent)Yes
    Adjust brightnessBy depressing button on handleBy depressing button on handleNot Changed (Equivalent)Yes
    Image/video captureBy depressing camera button on handleBy depressing camera button on handleNot Changed (Equivalent)Yes
    Cleaning/disinfectionHandle cleaned/disinfected per IFUHandle cleaned/disinfected per IFUEquivalent (Modified instruction for air dry time, but efficacy remains)Yes
    DisposalCannula sterile, single use, disposed after procedureCannula sterile, single use, disposed after procedureNot Changed (Equivalent)Yes
    Frequency of useHandle reusable, Cannula single useHandle reusable, Cannula single useNot Changed (Equivalent)Yes
    Duration of use< 24 hours< 24 hoursNot Changed (Equivalent)Yes
    Site of UseHospitals and physician officesHospitals and physician officesNot Changed (Equivalent)Yes
    Software
    SW version2.04.1N/A (New version, but updates do not introduce new risks to operation)Yes
    Level of ConcernModerateModerateNot Changed (Equivalent)Yes
    General FunctionsInitialize/setup video, LCD, UI, recordingInitialize/setup video, LCD, UI, recording, storage, playbackNot Changed (Equivalent)Yes
    General Parameters of ActionTransfer of sensor data to LCD displayTransfer of sensor data to LCD displayNot Changed (Equivalent)Yes
    Block DiagramMain Board performs imaging functionalityMain Board performs imaging functionalityEquivalent (Same except for specific sensor model, which handles HD output)Yes
    Specific Parameters of Actionv2.0 updatesv4.1 updates (including v2.0)Equivalent (Updates do not introduce a new risk; e.g., improved logo, zoom, file saving, etc.)Yes

    2. Sample Size Used for the Test Set and Data Provenance

    The document primarily describes bench testing and comparative analysis of the device's technological characteristics. It does not refer to a "test set" in the sense of a dataset of clinical cases for an AI algorithm. The performance data listed (e.g., measurement of outer diameter, image quality, image distortion) are conducted on the device hardware itself.

    • Sample Size for Test Set: Not applicable in the context of an AI/algorithm-based "test set." The testing involved physical devices and their components. The number of physical units tested is not specified but is typically determined by manufacturing quality control and regulatory submission requirements for non-AI devices.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not applicable for the type of testing described (bench testing of a medical device). This is a hardware device submission, not an AI/software submission that relies on clinical image data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: Not applicable. Ground truth, in the context of diagnostic performance typically associated with AI/algorithm submissions, is not directly established through expert consensus on clinical cases for this device. The "ground truth" for this device's performance data relates to engineering specifications and physical measurements (e.g., true dimensions, known test chart patterns).
    • Qualifications of Experts: Not applicable beyond the implicit expertise of engineers and technicians performing the stated bench tests, and the FDA's regulatory reviewers.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There is no clinical "test set" requiring adjudication by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its Effect Size.

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a hardware cystoscope, not an AI-assisted diagnostic tool for which such studies are typically conducted to evaluate reader performance with and without AI assistance.
    • Effect Size of Human Readers Improvement: Not applicable, as no MRMC study was conducted.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

    • Standalone Performance: Not applicable. This is a hardware device. While it contains software, the "software verification testing" mentioned is to ensure the software properly controls the device's functions (e.g., displaying HD images, recording). There is no standalone algorithm whose diagnostic performance is being evaluated independent of human use.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Engineering Specifications: The design parameters and expected performance of the device (e.g., specified focal length, field of view, angulation range, lumen cross-section, electrical safety standards).
    • Physical Measurements: Direct measurements of the device's components against design specifications (e.g., outer diameter, image quality against a known test chart, image distortion against a known pattern).
    • Biocompatibility Standards: Compliance with ISO 10993 requirements, verified through specific biocompatibility tests (Cytotoxicity, Irritation, Sensitization, Systemic Toxicity).
    • Predicate Device Performance: The established performance and safety characteristics of the legally marketed predicate device (Uro-G Flexible Cystoscope, K202921) serve as the primary reference for demonstrating equivalence.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This submission is for a hardware medical device with embedded software controlling its functions, not an AI/machine learning algorithm trained on a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    • Ground Truth for Training Set: Not applicable, as there is no training set for an AI/ML model for this hardware device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202921
    Device Name
    Uro-G Cystoscope
    Manufacturer
    Date Cleared
    2021-05-04

    (217 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-G flexible cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra.

    Device Description

    The Uro-G cystoscope is a handheld, battery-operated portable cystoscope consisting of a sterile, disposable steerable endoscopic cannula and a reusable handle with a video monitor. The disposable cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at its tip and one channel for infusion of irrigating fluid. The handle is lightweight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the disposable cannula. The handle contains the remaining electronics, including a power on/off button, a button to adjust the brightness of the LED, a button to allow capture of single images or to start/stop video of the procedure, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers, and firmware.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Uro-G Cystoscope. This document primarily focuses on establishing substantial equivalence to a predicate device (Uro-V Cystoscope) rather than presenting a performance study with acceptance criteria in the typical sense of a clinical trial for an AI/ML device.

    The "acceptance criteria" for a medical device like a cystoscope typically refer to the performance standards and safety requirements it must meet to be considered safe and effective, often demonstrated through engineering bench testing, biocompatibility, and electrical safety. The "study that proves the device meets the acceptance criteria" in this context refers to these non-clinical tests.

    There is no mention of AI/ML components in the Uro-G Cystoscope from the provided text, nor are there any acceptance criteria, sample sizes, or ground truth establishment relevant to AI/ML or comparative effectiveness studies with human readers. The device is a physical endoscope.

    Therefore, I will interpret "acceptance criteria" and "study" in the context of a medical device submission focused on equivalence for a non-AI physical device.

    Here's the breakdown based on the provided document:


    Acceptance Criteria and Study for Uro-G Cystoscope (Non-AI Device)

    The Uro-G Cystoscope is a physical medical device. The "acceptance criteria" are based on demonstrating substantial equivalence to a legally marketed predicate device (Uro-V Cystoscope) and compliance with relevant safety and performance standards through non-clinical testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criterion (Implicit/Explicit)Reported Device Performance/Comparison
    Predicate EquivalenceIntended Use: "The Uro-G flexible cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra."EXACTLY THE SAME as predicate (Uro-V Cystoscope).
    Indications for Use: Symptomatic voiding dysfunction, Hematuria, Bladder tumor surveillance, Recurrent lower urinary tract infection, Pelvic pain syndromes.EXACTLY THE SAME as predicate.
    Route of Advancement: Advanced to the bladder via the urethra.SAME as predicate.
    Site of Use: Hospitals and physician offices.SAME as predicate.
    Components: Reusable handle with video screen; Attachable cannula with working channel, illumination source, and camera at tip.**SAME** as predicate (handle is identical, cannula has modifications but core components are the same).
    Image Transmission/Display: Image transmitted from video camera at tip to video monitor on handle; 3.5-inch LCD Display.SAME as predicate.
    Field of View: 140 degrees.SAME as predicate.
    Direction of View: 0 degrees (forward viewing).SAME as predicate.
    Operational Characteristics: Adjustable illumination brightness, capture still images/video, cleaning/disinfection protocol, frequency/duration of use.SAME as predicate for these operational aspects.
    Tissue Contact Materials: Compliant with ISO 10993.SAME as predicate, and cannula materials passed all applicable biocompatibility tests (see details below).
    Differences and Rationale: Changes in Outer Diameter (5.5mm vs 4.2mm); Working Length (380mm vs 254mm); Working Channel ID (2.2mm); Flexible Tip Deflection (up to 210° up and 130° down); Focal Length (5-50mm vs 3-50mm). Rationale provided for each difference to demonstrate no new safety/effectiveness issues.Acceptable; Differences are addressed and do not raise new issues of safety or effectiveness. For example, increased OD is still within anatomical limits for patients, increased working length allows better access, improved tip deflection enhances ease of use without compromising safety due to continuous imaging.
    Performance Data (Bench/Non-Clinical)Design Verification & Validation: Ensure device performs as intended and meets specifications.Conducted with successful results.
    Packaging & Shelf-Life: Maintain sterility and integrity over time.Conducted for the sterile cannula, confirming labeled shelf life compliance.
    Biocompatibility: Meet ISO 10993 Part 1 requirements for short-duration (<24 hours) indwelling device concerning Cytotoxicity, Irritation, Sensitization, and Systemic Toxicity (acute).Device passed all tests.
    Sterilization Procedure Validation: Compliance with ISO 11135-1 (Ethylene Oxide), ISO 11607 (Packaging), AAMI TIR12 (Reusable device reprocessing), AAMI TIR30 (Cleaning reusable devices).Conducted for the sterile cannula; compliance confirmed.
    Software Verification & Validation: Ensure firmware and software function correctly and safely.Conducted with successful results.
    Electrical Safety & Electromagnetic Compatibility (EMC): Compliance with ISO 60601-1, ISO 60601-1-2 (EMC), and IEC 60601-2-18 (Endoscopic equipment, including thermal safety).Device found to meet compliance requirements. (Note: The handle component's electrical safety and EMC testing from the U-Scope 8000 submission K132384 was applicable).
    Mechanical Characteristics: Proper functioning of mechanical components (e.g., deflection mechanism).Tested with successful results.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of an AI/ML test set. For physical device testing, sample sizes would refer to the number of units tested for each specific bench test (e.g., number of cannulas tested for biocompatibility, sterility, or mechanical deflection). These specific numbers are not provided in this summary, but the document states "All necessary performance testing was conducted with bench testing."
    • Data Provenance: The data comes from the manufacturer's (UroViu Corporation) internal bench testing and verification/validation activities. The document does not specify the country of origin for the data or whether any retrospective/prospective studies (which are clinical by nature) were performed. Since no clinical studies were deemed necessary, the data is entirely from non-clinical (benchtop) sources.

    3. Number of Experts and Qualifications for Ground Truth Establishment

    • Not Applicable: This pertains to AI/ML model ground truth, which is not relevant for this physical device. The "ground truth" for this device's performance is established by direct physical measurements, chemical tests, electrical tests, and adherence to engineering specifications and international standards (e.g., ISO, IEC).

    4. Adjudication Method for the Test Set

    • Not Applicable: This pertains to expert review and consensus for AI/ML ground truth. For this physical device, "adjudication" is not needed in the same sense; rather, test results are assessed against predefined engineering and regulatory specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No: An MRMC comparative effectiveness study was not performed. This type of study is relevant for AI/ML diagnostic aids to assess human reader improvement, which is not applicable to a physical endoscope.

    6. Standalone Performance (Algorithm Only)

    • Not Applicable: There is no AI/ML algorithm in this device. Its performance is entirely dependent on its physical and electrical characteristics.

    7. Type of Ground Truth Used

    • Engineering Specifications and Standard Compliance: The "ground truth" for the device's performance relies on meeting pre-defined engineering specifications, international safety and performance standards (e.g., ISO 10993 for biocompatibility, ISO 60601-1 for electrical safety, ISO 11135-1 for sterilization), and demonstrating physical functionality (e.g., image quality, fluid infusion, tip deflection).

    8. Sample Size for the Training Set

    • Not Applicable: This device does not use an AI/ML training set. Its design and manufacturing process are based on engineering principles and established medical device production.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As there is no AI/ML training set, this question is not relevant.

    Summary for this specific product (Uro-G Cystoscope):

    The FDA 510(k) clearance process for the Uro-G Cystoscope primarily relied on demonstrating substantial equivalence to a predicate device (Uro-V Cystoscope) and meeting general safety and performance requirements through a series of benchtop engineering tests. Questions related to AI/ML-specific validation (e.g., "AI acceptance criteria," "sample sizes for test/training sets," "expert ground truth," "MRMC studies") are not relevant for this particular device as described in the provided document, as it is a physical, non-AI medical instrument.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182876
    Device Name
    Uro-N Cystoscope
    Manufacturer
    Date Cleared
    2019-07-02

    (263 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-N Cystoscope is intended for direction of therapeutic agents and solutions into target areas of the bladder and the lower urinary tract via a cystoscope.

    Device Description

    The UroViu Uro-N Cystoscope System consists of (1) a hand-held, battery operated portable endoscope contained within a reusable handle and (2) a disposable cannula with an injection needle. The fully assembled Uro-N system has overall length of 522 mm (~21 inches). The disposable cannula working length is 269 mm (~11 inches) and total length is 388 mm (~15 inches). The injection needle is 23 Ga (OD 0.635mm, ID 0.318mm). The combined handle and cannula weigh less than 1 pound. The reusable handle has a connector and locking mechanism for attaching and detaching the cannula. The catheter is sterilized by electron beam and is intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the Uro-N Cystoscope, demonstrating its substantial equivalence to predicate devices rather than proving it meets specific acceptance criteria through a standalone study with performance metrics. Therefore, many of the requested sections regarding acceptance criteria, performance, sample sizes, expert ground truth, and comparative effectiveness against human readers are not directly addressed in this document.

    However, I can extract information related to the device's testing and safety/effectiveness claims based on comparison to its predicates.

    Here's the summary based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not provide a table of quantitative acceptance criteria with corresponding device performance metrics. Instead, it relies on demonstrating substantial equivalence to predicate devices, implying that if it functions similarly and passes specified engineering and biocompatibility tests, it meets the necessary safety and effectiveness profiles.

    The performance evaluation is primarily comparative and based on the established safety and effectiveness of the predicate devices.

    | Aspect | Acceptance Criteria (Implied by Predicate) | Reported Device Performance (Summary) |
    |:--------------------------------|:---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
    | Biocompatibility | Meets ISO 10993 requirements for medical devices with less than 24 hours of patient tissue contact. Materials must be biocompatible (e.g., stainless steel for the needle). | The Uro-N Cystoscope and cannula are made from materials reviewed and cleared in K171500. The only new material, stainless steel for the needle, passed cytotoxicity testing. |
    | Electrical Safety | Compliance with IEC 60601-1 (medical electrical equipment - general requirements for basic safety and essential performance), specifically Class I, Type CF, defibrillation proof. | The device complies with IEC 60601-1, as reviewed and cleared in K171500 (predicate device). |
    | Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 (medical electrical equipment - electromagnetic disturbances - requirements and tests). | The device complies with IEC 60601-1-2, as reviewed and cleared in K171500 (predicate device). |
    | Software Verification and Validation | Software should be verified and validated to ensure proper function and safety. The software level of concern is "Moderate." | No change was made to the software from the predicate device (K171500), which was previously reviewed and cleared. |
    | Mechanical Performance | The device should be mechanically sound, withstand simulated use, and maintain structural integrity (e.g., resistance to bending, pulling, torque, and leaks). It should also maintain acceptable temperature at the tip surface, field of view, direction of view, and image quality comparable to the predicate diagnostic cystoscope. | Mechanical testing included: simulated use testing, mechanical testing (bending, pulling, torque, and presence of leaks), temperature at the surface of the tip testing, field of view testing, direction of view testing, and image quality testing. The results of these tests are not provided in detail, but the conclusion states "The information submitted... confirms that the Uro-N Cystoscope raises no new questions of safety and effectiveness and that it is substantially equivalent to the predicate devices." |
    | Sterility | Disposable components (cannula, needle) must achieve a Sterility Assurance Level (SAL) of 10^-6. | Disposable cannula and needle are sterile following exposure to ethylene oxide (EO) with an SAL of 10^-6. |
    | Cleanability (Reusable Handle) | The reusable handle, not provided sterile, must be cleanable and disinfectable according to company instructions to prevent cross-contamination. | The handle is cleaned and disinfected following company instructions (implies this is met). |
    | Duration of Use | Less than 24 hours for patient contact components. | Less than 24 hours. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of collected patient data or images for performance evaluation. The substantial equivalence argument relies on engineering and bench testing, as well as the prior clearance of the predicate devices.

    • Sample Size: Not applicable as no clinical or image-based test set was used for de novo performance evaluation against acceptance criteria.
    • Data Provenance: Not applicable. The testing was laboratory/bench-based.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. As no clinical test set requiring expert interpretation was used for de novo performance evaluation, no experts were involved in establishing ground truth in this context.

    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was reportedly done. The submission is a 510(k) for substantial equivalence, not a clinical efficacy study involving human readers with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The Uro-N Cystoscope is a medical device, not an AI algorithm. Its performance is evaluated through physical, electrical, and materials testing, and by demonstrating equivalence in function and safety to predicate devices.

    7. Type of Ground Truth Used

    The "ground truth" for the device's acceptable performance is based on demonstrating compliance with regulatory standards (e.g., ISO, IEC), confirmed biocompatibility, and functional equivalence to legally marketed predicate devices through various engineering and bench tests. There is no biological or diagnostic "ground truth" (like pathology or outcomes data) established for a test set in this summary.

    8. Sample Size for the Training Set

    Not applicable. The Uro-N Cystoscope is a physical device, not an AI/machine learning model, so there is no training set in the context of data for algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171500
    Device Name
    Uro-V Cystoscope
    Manufacturer
    Date Cleared
    2018-01-30

    (252 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uro-V cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra.

    Device Description

    The UroViu Uro-V cystoscope is a handheld, battery-operated portable cystoscope. It includes a sterile, single-use disposable cannula and a reusable handle. The disposable cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at its tip and one channel for infusion of irrigating fluid. The handle is lightweight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the disposable cannula. The handle contains the remaining electronics, including a power on/off button, a button to adjust the brightness of the LED, a button to allow capture of single images or to start/stop a video of the procedure, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers, and firmware.

    AI/ML Overview

    The provided text describes the Uro-V Cystoscope and its substantial equivalence to a predicate device, focusing on its design, intended use, and performance data from various bench tests. However, it does not contain information typically found in a clinical study report or a detailed validation study, especially regarding human-in-the-loop performance, ground truth establishment with experts, or comparative effectiveness studies with human readers.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a quantitative table for performance metrics like sensitivity, specificity, or accuracy. It lists various tests conducted to ensure compliance with standards and safety. The "reported device performance" is primarily a statement of compliance with these standards.

    Acceptance Criteria CategorySpecific Tests/Standards MetReported Device Performance
    Design & FunctionalityDesign verification and validation studies"All necessary performance testing was conducted with bench testing and included: Design verification and validation studies." (General statement of completion)
    Packaging & Shelf-LifePackaging and shelf-life studies"Packaging and shelf-life studies" (General statement of completion)
    TransitTransit testing"Transit testing" (General statement of completion)
    BiocompatibilityISO 10993 - Part 1 (for short duration indwelling device)"Testing was conducted on the device to verify that it is compliant with biocompatibility requirements for a short duration indwelling device, as specified in ISO 10993 - Part 1. The device was found to meet the compliance requirements..."
    SterilizationSterilization procedure validation; ISO 11135-1; ISO 11607; AAMI TIR12:2010; AAMI TIR30:2011"Due to the cannula being labeled as sterile, the cannula underwent sterilization validation and shelf life testing to confirm the label shelf life complies with the following standards: ISO 11135-1, ISO 11607, AAMI TIR12:2010, and AAMI TIR30:2011." (Implied compliance as part of validation)
    SoftwareSoftware verification and validation"Software verification and validation" (General statement of completion)
    Physical CharacteristicsSurface and edges evaluation; Cannula dimensions (maximum insertion width, minimum instrument channel width)"Surface and edges evaluation; Cannula dimensions (maximum insertion width, minimum instrument channel width)" (General statement of completion)
    Electrical Safety & EMCISO 60601-1; ISO 60601-1-2; IEC 60601-2-18 ("Medical electrical equipment -- Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment," including thermal safety)"The device was found to meet the compliance requirements for electrical safety as specified in ISO 60601-1 including provisions for EMC safety in ISO 60601-1-2 and IEC 60601-2-18 Medical electrical equipment -- Part 2-18: 'Particular requirements for the basic safety and essential performance of endoscopic equipment,' including thermal safety."
    MechanicalMechanical characteristics testing"Mechanical characteristics were also tested, with successful results."
    OpticalField of view accuracy; Direction of view accuracy; Optical resolution"Optical tests: Field of view accuracy, Direction of view accuracy, Optical resolution." (General statement of completion for these tests)

    2. Sample size used for the test set and the data provenance

    The document specifies "bench testing" and "design verification and validation studies." It does not mention a test set in the context of patient data, clinical images, or AI performance evaluation. Therefore, there is no information on sample size or data provenance (country of origin, retrospective/prospective). The testing focused on device specifications and compliance rather than clinical performance with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a hardware cystoscope, and the testing described is engineering and regulatory compliance testing, not a study involving human experts establishing ground truth for diagnostic accuracy (e.g., in an AI context).

    4. Adjudication method for the test set

    Not applicable for the same reasons as #3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted. This document describes a traditional 510(k) submission for a medical device (cystoscope), which involves demonstrating substantial equivalence primarily through technical specifications, safety, and performance bench testing. It is not an AI-powered device, and therefore, an MRMC study related to AI assistance for human readers is irrelevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm-only performance study was not done. The device itself is a physical endoscope, not an algorithm, and the testing described focuses on its physical and electrical characteristics, sterilization, and biocompatibility.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the various bench tests, the "ground truth" would be established by engineering specifications, validated test methods, and regulatory standards. For example, for biocompatibility, the ground truth is whether the materials meet ISO 10993. For electrical safety, it's whether the device passes the tests outlined in ISO 60601-1. This is not clinical ground truth from patient pathology or expert consensus.

    8. The sample size for the training set

    Not applicable. This document pertains to a physical medical device (cystoscope), not an AI algorithm. Therefore, there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1