K Number
K132384

Validate with FDA (Live)

Date Cleared
2014-07-02

(336 days)

Product Code
Regulation Number
884.1690
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The U-Scope 8000 HSC+EMB Cannula is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures and to obtain an endometrial tissue sample (biopsy) in an outpatient or office setting. The sample is used for cytologic and histologic diagnosis. Generally recognized indications for diagnostic hysteroscopy include: abnormal bleeding, infertility and pregnancy wastage, evaluation of abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, and pelvic pain.

Device Description

The U-Scope 8000 HSC+EMB Cannula is a handheld, battery-operated portable hysteroscope. It includes a sterile single use disposable endometrial biopsy cannula (the U-Scope 8000 HSC+EMB cannula) and a reusable HandTower (U-Scope 8000). The disposable U-Scope 8000 HSC+EMB cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at its tip and two channels for infusion of irrigating fluid and aspiration of tissue. Adjacent to the tip of the cannula is a curette that facilitates collection of tissue during the biopsy procedure. The disposable cannula is sterilized and packaged in a sealed pouch. The HandTower is lightweight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the U-Scope 8000 HSC+EMB cannula. The HandTower contains the remaining electronics, including a power on/off button, a button to adjust the brightness of the LED, a button to allow capture of single images or start/stop a video of the procedure, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers, and firmware.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the U-Scope 8000 HSC+EMB Cannula, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test CategoryAcceptance Criteria (Specified Standard)Reported Device Performance
Mechanical TestingN/A (implied by "meet design" and comparison to cleared devices for stiffness/force-to-tip deflection as a surrogate for biopsy sampling quality)Found to meet design after testing for bending, pulling, torque, and presence of leaks. Also tested for stiffness/force-to-tip deflection in comparison to other cleared endometrial sampling devices (met this, serving as a surrogate for biopsy sampling quality assessment).
Temperature TestingIEC 60601-2-18Found to meet requirements specified in IEC 60601-2-18 (surface temperatures at various points over time).
Field of View TestingISO 8600-3Found to meet requirements specified in ISO 8600-3 (accuracy of field of view).
Direction of Viewing TestingISO 8600-3Found to meet requirements specified in ISO 8600-3 (accuracy of directions of view).
Image Quality TestingISO 8600-5Found to meet requirements of ISO 8600-5.
Electrical SafetyIEC 60601-1, IEC 60601-1-2:2007 (EMC), IEC 61000-4-2:2008 (EMC), IEC 60601-2-18:2009 (thermal safety)Met compliance requirements for electrical safety as specified in IEC 60601-1, including provisions for EMC safety in IEC 60601-1-2:2007 and IEC 61000-4-2:2008, as well as testing to IEC 60601-2-18:2009. Thermal safety included in IEC 60601-2-18.
BiocompatibilityISO 10993 (implied, for tissue contact materials)Demonstrates biocompatibility through cytotoxicity, sensitization, irritation, and acute systemic toxicity testing for device components in contact with the patient.
Sterilization ValidationISO 11135-1:2007Found in compliance with ISO 11135-1:2007 (requirements for development, validation and routine control of a sterilization process for medical devices using Ethylene oxide).
Packaging (Sterile Devices)ISO 11607-1:2006Found in compliance with ISO 11607-1:2006 (packaging for terminally sterilized medical devices).
Reprocessing (Reusable)AAMI TIR12:2010, AAMI TIR 30:2011Found in compliance with AAMI TIR12:2010 (designing, testing, and labeling reusable medical devices for reprocessing) and AAMI TIR 30:2011 (compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices), likely referring to the HandTower which is reusable.
Shelf LifeN/A (validation for claimed shelf life)Sterilization validation and shelf life testing confirmed the label shelf life.

2. Sample Size Used for the Test Set and Data Provenance

The provided summary does not specify sample sizes for the test set for any of the performance tests (mechanical, temperature, field of view, image quality, electrical safety, biocompatibility, sterilization, or shelf life).

The data provenance is also not explicitly stated (e.g., country of origin, retrospective/prospective). These appear to be laboratory and bench testing rather than clinical study data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided as the document describes technical performance testing rather than studies involving human interpretation or clinical outcomes that would require ground truth established by experts.

4. Adjudication Method

This information is not applicable as the tests described are technical performance tests that do not involve expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. A MRMC comparative effectiveness study was not conducted or reported in this summary. The summary focuses on the technical performance and safety of the device itself, not its impact on human reader performance.

6. Standalone (Algorithm Only) Performance Study

No. This device is a hysteroscope with an endometrial biopsy cannula, a physical medical device. It is not an algorithm or AI-driven system, so a standalone algorithm performance study is not applicable.

7. Type of Ground Truth Used

The "ground truth" for the performance tests described can be considered the established requirements and specifications of the referenced standards (ISO, IEC, AAMI). For example, for temperature testing, the ground truth is the allowable temperature range in IEC 60601-2-18. For image quality, it's the criteria specified in ISO 8600-5. For biocompatibility, it's the pass/fail criteria for the specific tests based on ISO 10993.

8. Sample Size for the Training Set

This information is not applicable. The U-Scope 8000 HSC+EMB Cannula is a physical device, not an AI or machine learning model that requires a training set.

9. How Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

1111 0 2 2014

K132384 Pg. 1 of 5

510(k) Summary

Sponsor:EndoSee Corporation
Contact Person:Xiaolong (Bruce) Ou Yang
Telephone:(650) 397-5174
Fax:(888) 388-1420
510(k) Summary Date:July 23, 2013
Device Name:U-Scope 8000 HSC+EMB Cannula
Common Name:Hysteroscope
Classification:
Regulatory Class:II
Review Category:21 CFR 884.1690
Classification Panel:Obstetrics and Gynecology
Product Code:HIH (hysteroscope and accessories)

Legally Marketed Predicate Devices

The U-Scope Model 8000, EndoSee Corporation (K123151)

Endosampler, MedGyn Products, Inc. (K840383)

Device Description

The U-Scope 8000 HSC+EMB Cannula is a handheld, battery-operated portable hysteroscope. It includes a sterile single use disposable endometrial biopsy cannula (the U-Scope 8000 HSC+EMB cannula) and a reusable HandTower (U-Scope 8000). The disposable U-Scope 8000 HSC+EMB cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at its tip and two channels for infusion of irrigating fluid and aspiration of tissue. Adjacent to the tip of the cannula is a curette that facilitates collection of tissue during the biopsy procedure. The disposable cannula is sterilized and packaged in a sealed pouch. The HandTower is lightweight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the U-Scope 8000 HSC+EMB cannula. The HandTower contains the remaining electronics, including a power on/off button, a button to adjust the brightness of the LED, a button to allow capture of single images or start/stop a video of the procedure, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers, and firmware.

{1}------------------------------------------------

Intended Use

The U-Scope 8000 HSC+EMB Cannula is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures and to obtain an endometrial tissue sample (biopsy) in an outpatient or office setting. The sample is used for cytologic and histologic diagnosis. Generally recognized indications for diagnostic hysteroscopy include: abnormal bleeding, infertility and pregnancy wastage, evaluation of abnormal hysterosalpingogram, intrauterine foreign body, amenorrhea, and pelvic pain.

Substantial Equivalence

The submission device is substantially equivalent to the U-Scope Model 8000 manufactured by EndoSee Corporation (K123151) in terms of technological characteristics of the hand tower and to the Endosampler manufactured by MedGyn Products, Inc. (K840383) in terms of indication of use and technological characteristics of a biopsy cannula with curette.

Subject devicePredicate devicePredicate device
U-Scope 8000 HSC+EMBCannula(this submission)EndosamplerMedGyn Products Inc.(K840383)U-Scope Model 8000EndoSee Corporation(K123151)
Indication for useTo permit viewing of thecervical canal and uterinecavity for the purpose ofperforming diagnosticprocedures and to obtainan endometrial tissuesample (biopsy) in anoutpatient or an officesetting. The sample isused for cytologic andhistologic diagnosis.Generally recognizedindications for diagnostichysteroscopy include:abnormal bleeding,infertility and pregnancywastage, evaluation ofabnormalhysterosalpingogram,intrauterine foreign body,To obtain clearlydifferentiatedendometrial tissuesample withoutanesthesia in anoutpatient setting or inan office setting. Thesample is used forcystologic andhistologic diagnosisTo permit viewing ofthe cervical canal anduterine cavity for thepurpose of performingdiagnostic procedures.Generally recognizedindications fordiagnostichysteroscopy include:abnormal bleeding,infertility andpregnancy wastage,evaluation of abnormalhysterosalpingogram,intrauterine foreignbody, amenorrhea, andpelvic pain.
amenorrhea, and pelvic pain.
ProceduresUterine diagnostic and endometrial biopsyUterine diagnostic usesDiagnostic hysteroscopy
Site of useHospitals and physician officesHospitals and physician officesHospitals and physician offices
Device Features
Components(1) image-capturing hand tower; (2) attachable cannula with outflow and inflow ports in which suction can be created and with a curette near its tipCannula with outflow port in which suction can be created and with a curette opening near its tip(1) image-capturing hand tower; (2) attachable cannula with inflow port
Cannula Outer Diameter4.3 mm3 mm3.8 x 4.6 mm (oval)
Cannula length27 cm21 cm26 cm
Illumination light sourceLEDsN/ALEDs
Image transmissionImage transmitted from a video camera to a video monitor on the handleNo imageImage transmitted from a video camera to a video monitor on the handle
LCD display size3.5 inches (diagonal)N/A3 inches (diagonal)
Battery charge indicationCharge indication as an icon on the LCD monitorN/ANo charge indication on the LCD monitor
Battery power3.7 VN/A3.3 V
Adjust brightness of LEDsAdjust by depressing a button on the hand towerN/ABrightness cannot be adjusted
Capture still or video images during procedureCapture still or video during procedure by depressing a camera button on the hand towerN/ACannot capture still or video during procedure
Enter patient ID information prior to procedureUser interface on monitor allows physician to add patient informationN/ACannot input patient information via the interface with the monitor
Duration of use≤ 24 hours≤ 24 hours≤ 24 hours
SterilizationHandTower is not sterile. The U-Scope 8000 HSC+EMB cannula is sterile following exposure to ethylene oxide (EO)Exposure to ethylene oxide (EO)Hand tower is not sterile. Diagnostic cannula sterile following exposure to ethylene oxide (EO)
Frequency of useHandTower is reusable. The U-Scope 8000 HSC+EMB cannula is single patient useSingle patient useHand tower is reusable Diagnostic cannula is single patient use
Tissue contact materialsCompliant with ISO 10993Compliant with ISO 10993Compliant with ISO 10993

Table I. Comparison of the U-Scope 8000 HSC+EMB Cannula to the Predicate Devices

{2}------------------------------------------------


.

.

・ ·

{3}------------------------------------------------

Performance Data


:

:

The U-Scope 8000 HSC+EMB Cannula was subjected to mechanical, temperature, field of view and image quality testing, in accordance with the FDA's "Hysteroscopic and Gynecologic Laparoscopes Submission Guidance" (1996).

Mechanical testing: The device was tested for bending, pulling, torque, and presence of leaks and following testing was found to meet design. The device was also tested for stiffness/force-to-tip deflection in comparison to other cleared endometrial sampling devices. This test served as a surrogate for an assessment of biopsy sampling quality.

Temperature testing: the device was measured for surface temperatures at various points over time and found to meet requirements specified in IEC 60601-2-18.

Field of View and Direction of Viewing testing: The device was tested for field of view accuracy and directions of view accuracy and found to meet requirements specified in ISO 8600-3.

Image Quality testing: The device was tested for image quality and found to meet requirements of ISO 8600-5.

The U-Scope 8000 HSC+EMB Cannula was tested for electrical and safety testing by a certified test laboratory and met the compliance requirements for electrical safety as specified in IEC 60601-1, including provisions for EMC safety in IEC 60601-1-2:2007 and IEC 61000-4-2:2008, as well as testing to IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: "Particular requirements for the basic safety and essential performance of endoscopic equipment," including thermal safety. Mechanical characteristics were also tested with successful results.

A' series of biocompatibility testing, including cytotoxicity, sensitization, irritation, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.

{4}------------------------------------------------

Due to its labeling as sterile, the U-Scope 8000 HSC+EMB Cannula underwent sterilization validation and shelf life testing to confirm the label shelf life and was found in' compliance with the following:

  • ISO 11135-1:2007 Sterilization of health care products --. Ethylene oxide -Part 1: Requirements for development, validation and routine control of a sterilization process for medical device;
  • Packaging for Terminally Sterilized Medical . ISO 11607-1:2006 Devices;
  • Designing, testing, and labeling . AAMI TIR12:2010 reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers; and
  • . AAMI TIR 30:2011 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices.

Conclusion

U-Scope 8000 HSC+EMB Cannula is substantially equivalent to its proposed predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 2, 2014

EndoSee Corporation Pin Yu Vice President, Regulatory Affairs 4546 El Camino Road, Suite 215 Los Altos. CA 94022

Re: K132384

Trade/Device Name: U-Scope™ 8000 HSC+EMB Cannula Regulation Number: 21 CFR§ 878.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: Class II Product Code: HIH Dated: June 3, 2014 Received: June 5, 2014

Dear Pin Yu,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it . may be subject to additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be idvised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{6}------------------------------------------------

Page 2 - Pin Yu

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/6/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a bold, sans-serif font. The text is horizontally oriented and appears to be a title or heading. The background is plain and does not distract from the text.

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K132384

Device Name: U-Scope 8000 HSC+EMB Cannula

Indications for Use:

The U-Scope 8000 HSC+EMB Cannula is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic procedures and to obtain an endometrial tissue sample (biopsy) in an outpatient or office setting. The sample is used for cytologic and histologic diagnosis. Generally recognized indications for diagnostic hysteroscopy include: abnormal bleeding, infertility and pregnancy wastage, evaluation of abnormal hysterosalpingogram, intrauterine foreign body, amenorthea, and pelvic pain.

Prescription Use: ____________________________________________________________________________________________________________________________________________________________ X (21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use: · (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Attachment 13

2 ol 2

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.