510(k) Data Aggregation
(252 days)
The Non-Sterile, Powder-Free, Latex Examination Gloves, Green and with Protein Labeling Claim (50 micrograms or less total water extractable protein per gram of glove), is a disposable device intended for Medical Purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
Non-Sterile, Powder-Free, Latex Examination Gloves, Green and with Protein Labeling Claim
The provided text describes the acceptance criteria and performance data for "Non-Sterile, Powder-Free, Latex Examination Gloves, Green and with Protein Labeling Claim."
Here's the information parsed according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
Criteria | Acceptance Criteria | Reported Device Performance |
---|---|---|
Material | Latex | Latex |
Cuff | Beaded | Beaded |
Powder Residue | Maximum 2mg/glove | In compliance with ASTM D3578-10 |
Quality Assurance | ISO 2859-1:1999, ISO 9001:2000, ISO 13485 | Manufactured under ISO9001:2000 and ISO 13485 (implicitly meets ISO 2859-1:1999 as per ASTM) |
Dimensions | Inspection Level S-2, AQL 4.0 | (Implicitly met as per ASTM D3578-10 compliance) |
Overall Length | 240 mm minimum | (Implicitly met as per ASTM D3578-10 compliance) |
Width (medium glove) | 95 mm minimum | (Implicitly met as per ASTM D3578-10 compliance) |
Thickness | 0.08mm minimum | (Implicitly met as per ASTM D3578-10 compliance) |
Physical Properties | Inspection Level S-2, AQL 4.0 | (Implicitly met as per ASTM D3578-10 compliance) |
Before Aging-Tensile Strength | 18.0 Mpa minimum | (Implicitly met as per ASTM D3578-10 compliance) |
Before Aging-Ultimate Elongation | 650% minimum | (Implicitly met as per ASTM D3578-10 compliance) |
After Aging-Tensile Strength | 14.0 Mpa minimum | (Implicitly met as per ASTM D3578-10 compliance) |
After Aging-Ultimate Elongation | 500% minimum | (Implicitly met as per ASTM D3578-10 compliance) |
Water Tight Test 1000ml | Inspection Level G-1, AQL 1.5 | Meets the FDA 1000 ml water test both before and after aging |
Visual Major Defects | Inspection Level G-1, AQL 1.5 | (Implicitly met as per ASTM D3578-10 compliance) |
Visual Minor Defects | Inspection Level G-1, AQL 2.5 | (Implicitly met as per ASTM D3578-10 compliance) |
Protein Content |
Ask a specific question about this device
(252 days)
Non-Sterile, Powder-Free, Nitrile Examination Gloves Blue is a disposable device intended for Medical Purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
Non-Sterile, Powder-Free, Nitrile Examination Gloves Blue. A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.
Summary of Technological Characteristics:
Cuff: Beaded
Material: Nitrile
Powder Residue: Maximum 2mg/glove
Quality Assurance: In compliance with ASTM D6319-10, ISO 2859-1, manufactured under ISO9001:2008 and ISO 13485:2003
Physical Properties:
Overall Length: 240 mm minimum
Width: 95 mm minimum (for medium glove)
Thickness: .05 mm minimum
BEFORE AGING: Tensile Strength: 14.0 Mpa minimum, Ultimate Elongation: 500% minimum
AFTER AGING: Tensile Strength: 14.0 Mpa minimum, Ultimate Elongation: 400% minimum
Special Properties: None
Packaging: 150 pcs per dispenser box, 10 boxes per case, 1,500 gloves per case
Sizes: XS — XL
This document describes the acceptance criteria and performance of "Non-Sterile, Powder-Free, Nitrile Examination Gloves Blue" (K112012), not an AI device. Therefore, many of the requested sections related to AI device studies (e.g., sample size for test set, ground truth experts, MRMC studies, standalone performance, training set details) are not applicable.
Here's the available information based on the provided text, adapted to the closest relevant categories:
1. Table of Acceptance Criteria and Reported Device Performance
Criteria | Acceptance Criteria | Reported Device Performance |
---|---|---|
Powder Residue | Maximum 2mg/glove | Not explicitly stated, but implies meeting this |
Quality Assurance | In compliance with ASTM D6319-10, ISO 2859-1, ISO9001:2008, ISO 13485:2003 | Stated as being in compliance |
Dimensions (Inspection) | AQL 4.0 (S-2 level) | Not explicitly stated as pass/fail, but implies meeting this |
Physical Properties (Inspection) | AQL 4.0 (S-2 level) | Not explicitly stated as pass/fail, but implies meeting this |
Water Tight Test (1000ml, Inspection) | AQL 1.5 (G-1 level) | Met both before and after aging |
Visual Major Defects (Inspection) | AQL 1.5 (G-1 level) | Not explicitly stated as pass/fail, but implies meeting this |
Visual Minor Defects (Inspection) | AQL 2.5 (G-1 level) | Not explicitly stated as pass/fail, but implies meeting this |
Overall Length | 240 mm minimum | Not explicitly stated as a measured value, but implies meeting this |
Width (medium glove) | 95 mm minimum | Not explicitly stated as a measured value, but implies meeting this |
Thickness | 0.05 mm minimum | Not explicitly stated as a measured value, but implies meeting this |
BEFORE AGING | ||
Tensile Strength | 14.0 Mpa minimum | 14.0 Mpa minimum |
Ultimate Elongation | 500% minimum | 500% minimum |
AFTER AGING | ||
Tensile Strength | 14.0 Mpa minimum | 14.0 Mpa minimum |
Ultimate Elongation | 400% minimum | 400% minimum |
Protein Labeling Claim Level |
Ask a specific question about this device
(83 days)
The Powder-Free Black Examination Glove with Protein Labeling Claim (50 micrograms or less total water extractable protein per gram of glove), is a disposable device intended for Medical Purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
Trade Name: Powder-Free Black Latex Examination Glove With Protein Labeling Claim
Common Name: Examination Gloves
Classification Name: Patient Examination Glove (per 21 CFR 880.6251)
Class 1: Powder-Free Latex examination glove 80LYY that meets all of the requirements of ASTM D3578-05.
Powder Residue: Maximum 2mg/glove
Cuff: Beaded
Material: Latex Material. Latex Quality ASSurancer 11: Com859-1:1999 and manufactured under ISO9001:2000
Overall Length: 240 mm minimum
Width: 95 mm minimum (for medium glove)
Palm Thickness: 0.13 to 0.18mm (at center of palm)
Finger Thickness: 0.15 to 0.20 mm (at 15mm from tip of center finger)
Cuff Thickness: 0.10 to 0.15 mm (at 40mm from the beaded end)
BEFORE AGING Tensile Strength: 21.0 Mpa minimum
BEFORE AGING Ultimate Elongation: 700% minimum
BEFORE AGING Pinhole AQL 1.5 minimum
AFTER AGING 16.0 Mpa minimum
AFTER AGING 500% minimum
AFTER AGING 1.5 minimum
Special Properties: Black Colorant
Packaging: 100 pcs per dispenser box, 10 boxes per case, 1,000 gloves per case
Sizes: XXS — XXL
The provided document describes the acceptance criteria and performance of a Powder-Free Black Latex Examination Glove. It does not describe a study involving an AI-powered device or human readers. Therefore, several requested sections, such as those pertaining to AI model performance, human expert involvement, and training/test set details, are not applicable.
Here's the available information presented in the requested format:
1. A table of acceptance criteria and the reported device performance
Criteria | Acceptance Criteria (Minimum/Maximum) | Reported Device Performance |
---|---|---|
Physical Properties | Meets all requirements of ASTM D3578-05 | Meets all requirements of ASTM D3578-05 |
Powder Residue | Maximum 2 mg/glove | Not explicitly stated, but "Powder-Free" implies meeting this or better. |
Overall Length | 240 mm minimum | 240 mm minimum |
Width (for medium glove) | 95 mm minimum | 95 mm minimum |
Palm Thickness | 0.13 to 0.18 mm (at center of palm) | 0.13 to 0.18 mm (at center of palm) |
Finger Thickness | 0.15 to 0.20 mm (at 15mm from tip of center finger) | 0.15 to 0.20 mm (at 15mm from tip of center finger) |
Cuff Thickness | 0.10 to 0.15 mm (at 40mm from the beaded end) | 0.10 to 0.15 mm (at 40mm from the beaded end) |
Before Aging | ||
Tensile Strength | 21.0 Mpa minimum | 21.0 Mpa minimum |
Ultimate Elongation | 700% minimum | 700% minimum |
Pinhole AQL | 1.5 minimum | 1.5 minimum |
After Aging | ||
Tensile Strength | 16.0 Mpa minimum | 16.0 Mpa minimum |
Ultimate Elongation | 500% minimum | 500% minimum |
Pinhole AQL | 1.5 minimum | 1.5 minimum |
Water Tightness | Meets FDA 1000 ml water test (both before and after aging) | Meets FDA 1000 ml water test (both before and after aging) |
Protein Labeling Claim |
Ask a specific question about this device