Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K220202
    Date Cleared
    2022-03-31

    (66 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uterine ElevatOR PRO with OccludOR Balloon™ is indicated for manipulation of the uterus, and injection of fluids or gases during laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectomy, laparoscopic tubal occlusion, and diagnostic laparoscopy. The Uterine ElevatOR PRO with OccludOR Balloon™ maintains pneumoperitoneum during laparoscopic procedures by sealing the vagina once colpotomy is performed.

    Device Description

    The Uterine ElevatOR PRO with OccludOR Balloon™ is a sterile, single-use medical device used to manipulate the uterus and cervix in surgical and diagnostic procedures. The device includes an insulated and anatomically designed stainless steel shaft with an intrauterine balloon and reference graduations (centimetres) at the proximal [to patient] end and an external handle at the distal end. The intrauterine balloon is inflated with air using a standard syringe (not included). The device has a vaginal cup that is pushed along the shaft to support a green cervical cup. The cervical cup provides a guide for colpotomy. There are four sizes of cervical cups available, 32mm, 37mm, and 40mm, which correspond to the different model numbers. A locking assembly secures the position of the vaginal cup and therefore the cervical cup. Sites for suturing are located on the cervical cup. An Occluder Balloon further reduces leakage of pneumoperitoneum during laparoscopic surgery. A port at the distal end of handle provides a channel for injection of fluids or gases.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called the "Uterine ElevatOR PRO with OccludOR Balloon™", not an AI/ML device. Therefore, the request for acceptance criteria and study details related to AI/ML device performance (such as sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types) is not applicable.

    The document describes performance data for a physical medical device, which includes:

    1. A table of acceptance criteria and the reported device performance:

    The document broadly states that "In all testing, the subject device met the pre-specified acceptance criteria." However, it does not provide a specific table detailing each acceptance criterion and the corresponding reported device performance. It lists the types of tests conducted:

    Test CategorySpecific Tests PerformedReported Performance
    Biocompatibility- Cytotoxicity per ISO 10993-5:2009- Irritation per ISO 10993-10:2010- Sensitization per ISO 10993-5:2010- Acute Systemic toxicity per ISO 10993-11:2017- Pyrogenicity per ISO 10993-11 and USP <151>"The subject device is considered externally communicating device & surface contact device that may come in contact with tissue and breached or compromised surface respectively for a duration of less than 24 hours." "met the pre-specified acceptance criteria."
    Performance & Mechanical- Intrauterine balloon integrity & concentric inspection- Intrauterine balloon inflation stability test for a defined time period.- Intrauterine balloon burst test- Intrauterine balloon fatigue test- Intrauterine balloon adhesion strength test- OccludOR balloon outer diameter inspection & leakage free test- OccludOR balloon deflation time test- OccludOR balloon capacity test- reliability/safety- OccludOR balloon fatigue durability test- OccludOR balloon burst test- Vaginal cup air leakage test- Cervical cup temperature test- Air inflation valve test- Dye-injection port compliance & leakage test- Thumb screw strength test- Locking force test of vaginal cup- Ease of movement of vaginal cup along the shaft- Cervical cup detachment force test- Cervical cup stopper adhesion strength test- Device tensile strength- Handle joint strength- Tip deflection- Ultrasonic welding strength"In all testing, the subject device met the pre-specified acceptance criteria."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify sample sizes for any of the tests, nor does it provide information on data provenance (country of origin, retrospective/prospective). This type of detail is typically found in full study reports, not a 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable, as this is a physical medical device and not an AI/ML device relying on "ground truth" derived from expert consensus. The "truth" for this device's performance is determined by meeting physical and biological specifications through various standard engineering and biocompatibility tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable for a physical medical device as described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is a physical medical device and not an AI/ML device involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable, as this is a physical medical device without an algorithm component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable for a physical medical device. The "ground truth" here is adherence to engineering specifications and biocompatibility standards.

    8. The sample size for the training set:

    Not applicable, as this is a physical medical device and not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set or AI/ML model for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212505
    Device Name
    DUMI ManipulatOR
    Date Cleared
    2022-01-10

    (154 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The use of the DUMI ManipulatOR is indicated in diagnostic laparoscopy, mini laparotomy, fertility exams, and salpingoplastic procedures where manipulation of the uterus is required.

    Device Description

    The DUMI ManipulatOR is a sterile, single use, disposable medical device designed for the intra-operative manipulation of both the anteverted and retroverted uterus as well as facilitating simple and effective dye studies. The device includes an anatomically designed balloon at the distal end that is injected with air by syringe through the luer connector of inflation valve. The three-way hub can accommodate fluid and can be used for dye studies.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines various tests conducted to demonstrate the safety and effectiveness of the DUMI ManipulatOR, primarily by establishing substantial equivalence to a predicate device. The acceptance criteria are generally "successfully met predetermined acceptance criteria" or demonstrating performance "at least the same as the predicate device." Specific numerical criteria are not always explicitly stated in this summary, but the types of tests and their successful outcome are.

    Test CategorySpecific TestAcceptance Criteria (as stated or implied)Reported Device Performance
    General ComplianceCompliance with ISO 13485:2016, 21 CFR 820Requirements for regulatory purposes metSuccessfully met
    ISO 14971:2019 (Risk Management)Application of risk management to medical devicesSuccessfully met
    SterilizationEO Sterilization ValidationRequirements for development, validation, and routine control of sterilization processSuccessfully met
    EO ResiduesAcceptable residue levelsSuccessfully met
    Bacteriostasis and FungistatisNo unacceptable growthSuccessfully met
    BioburdenAcceptable bioburden levelsSuccessfully met
    Sterility Test (ISO 11737-2)Demonstrated sterilitySuccessfully met
    Shelf LifePackaging integrity post 5 year accelerated agingMaintain integritySuccessfully met
    Performance post 5 year accelerated agingMaintain intended performanceSuccessfully met
    PackagingSeal Strength (ASTM F88)Acceptable seal strengthSuccessfully met
    Dye Penetration (ASTM F1929)No leaksSuccessfully met
    Gross Leaks (ASTM F2096)No gross leaksSuccessfully met
    Packaging for terminally sterilized medical devices (ISO 11607-1, ISO 11607-2)Requirements for materials, sterile barrier systems, packaging systems, formation, sealing, assemblySuccessfully met
    TransportationHandling, Loose Load Vibration, Low Pressure (High Altitude) Hazard, Vehicle VibrationWithstand transportation stresses without damage or compromise to deviceSuccessfully met
    BiocompatibilityCytotoxicity (ISO 10993-5)Non-cytotoxicFound to be non-cytotoxic
    Sensitization (ISO 10993-10)Non-sensitizingFound to be non-sensitizing
    Acute Systemic Toxicity (ISO 10993-11)Non-toxicFound to be non-toxic
    Irritation (ISO 10993-10)Non-irritatingFound to be non-irritating
    General Biocompatibility (ISO 10993-1)Evaluation and testing within a risk management processSuccessfully met
    Performance (Device Specific)Intrauterine balloon capacity and diameter testPerformance at least the same as predicate (10cc volume for balloon)Performed at least the same as predicate (same volume)
    Intrauterine balloon fatigue test (repeat inflation)Performance at least the same as predicatePerformed at least the same as predicate
    Intrauterine balloon burst testPerformance at least the same as predicatePerformed at least the same as predicate
    Air leakage test for prolonged timePerformance at least the same as predicatePerformed at least the same as predicate
    Spring load locking mechanism force testPerformance at least the same as predicatePerformed at least the same as predicate
    Component joint strength testPerformance at least the same as predicatePerformed at least the same as predicate
    Tip deflection testPerformance at least the same as predicatePerformed at least the same as predicate
    Balloon adhesion strength testPerformance at least the same as predicatePerformed at least the same as predicate
    Luer connector leakage test (ISO 80369-7)Performance at least the same as predicate; compliant with standard connectorsPerformed at least the same as predicate; compliant with standard
    REACH Regulation (EC) No 1907/2006- Non-DEHP (EN 14372:2004)Non-DEHPSuccessfully met

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state specific sample sizes for each individual test. However, it references several standards that typically dictate sample size requirements for testing and quality control:

    • ISO 2859-1:1999: "Sampling procedures for inspection by attribute - Part 1: Sampling schemes indexed by acceptance quality limit (AQL) for lot-by-lot inspection." This standard would have guided the sample sizes for attribute inspection tests.
    • ASTM F3172-15: "Design verification device size and sample selection for Endovascular devices." While the DUMI ManipulatOR is not an endovascular device, this standard indicates that sample size selection for design verification was a considered aspect of the testing strategy.

    Given the nature of the device (uterine manipulator) and the tests performed (physical, mechanical, biocompatibility, sterilization, packaging), the data would be prospective as the tests were conducted specifically for this submission.

    The provenance of the data is not explicitly stated per test but is implied to be from testing labs associated with The O.R. Company Pty Ltd, an Australian company with offices in Australia, North America, and Hong Kong. The testing was conducted to meet international and US FDA standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The tests described are primarily objective, quantitative engineering, microbiological, and chemical tests. Clinical expert "ground truth" (e.g., radiologists, pathologists) is typically relevant for diagnostic devices or AI algorithms that interpret medical data. For a manual surgical instrument, the "ground truth" for performance is established through defined acceptance criteria in engineering standards, material science, and regulatory guidelines.

    4. Adjudication Method for the Test Set:

    This information is not applicable/provided. Adjudication methods (like 2+1, 3+1) are usually relevant in clinical trials or studies where human interpretation or consensus is required to establish a gold standard for disease presence or other clinical outcomes. The tests performed for the DUMI ManipulatOR are laboratory-based and objective, with outcomes determined by measurement against predetermined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted and is not applicable to this device. The DUMI ManipulatOR is a manual surgical instrument, not an AI-powered diagnostic or assistive tool for image interpretation. Therefore, there is no concept of "human readers" or "AI assistance" in its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm-only performance assessment was not done. The DUMI ManipulatOR is a physical medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is based on:

    • Engineering and Material Standards: Defined specifications and performance limits derived from international and national standards (e.g., ISO, ASTM).
    • Predicate Device Equivalence: The performance of the predicate device (Panpac Uterine Manipulator Injector, Model UMI 4.5) serves as a benchmark for comparison in several performance tests. The new device must perform "at least the same as" the predicate.
    • Regulatory Requirements: Compliance with quality management systems (ISO 13485, 21 CFR 820), risk management (ISO 14971), and biocompatibility guidelines (ISO 10993 series).

    8. The Sample Size for the Training Set:

    This information is not applicable. The DUMI ManipulatOR is not an AI/machine learning device, so there is no concept of a "training set" for an algorithm. The device itself is manufactured, and its performance is verified through testing.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1