Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221411
    Date Cleared
    2023-06-23

    (403 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Terumo Medical Products (Hangzhou) Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surflo Winged Infusion Set is indicated for venipuncture and intravenous administration of fluids for up to 24 hours using compatible/appropriate disposable infusion devices for medical purposes. It may be used for any patient population with consideration given to patient size.

    Device Description

    The Surflo Winged Infusion Set consists of a protector, needle, wing, tubing, connector, and connector cap. The connector is 6% female Luer. It is primarily made from medical grade PVC (the whole device uses TOTM as plasticizer instead of DEHP) and medical grade stainless steel. The Surflo Winged Infusion Set is operated manually. The device is intended to be used by trained healthcare professionals in a healthcare facility. The Surflo Winged Infusion Set is available in different gauge sizes and tubing lengths. The Surflo Winged Infusion Set is a disposable device intended for single use only. The device is packaged and sterilized by ethylene oxide gas.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Surflo Winged Infusion Set, seeking substantial equivalence to a legally marketed predicate device. As such, it describes performance testing conducted to demonstrate product safety and effectiveness, rather than a study proving a device meets acceptance criteria for an AI/ML product. Therefore, several points in your requested output, such as those related to AI/ML specific criteria (e.g., sample size for training set, number of experts for ground truth establishment, MRMC study, standalone performance) are not applicable to this type of medical device submission.

    However, I can extract and present the relevant information regarding acceptance criteria and performance testing for this traditional medical device:

    Acceptance Criteria and Device Performance for Surflo Winged Infusion Set

    The Surflo Winged Infusion Set underwent non-clinical performance testing to ensure it met predetermined specifications, conformed to applicable ISO and non-ISO standards, and demonstrated substantial equivalence to its predicate device.

    1. Table of Acceptance Criteria (Standard Compliance) and Reported Device Performance

    For tests conducted against specific standards, the document implies that compliance with the standard is the acceptance criterion. For additional performance tests, the acceptance criterion was "Meets acceptance criteria," established based on the predicate device's performance.

    TestStandard / Acceptance CriteriaReported Device Performance
    Flow rateGB 18671-2009 Intravenous needles for single useComplies with standard
    Leak testJIS T 3222:2011 Sterile winged intravenous devicesComplies with standard
    Wing to tubing connection strength (static pull force)ISO 8536-4:2019 Infusion equipment for medical useComplies with standard
    Tubing to connector connection strength (static pull force)ISO 8536-4:2019 Infusion equipment for medical useComplies with standard
    Needle to wing connection strength (static pull force)ISO 8536-4:2019 Infusion equipment for medical useComplies with standard
    Needle curvatureJIS T 3222:2011 Sterile winged intravenous devicesComplies with standard
    Wing to tubing connection strength (dynamic pull force)Meets acceptance criteria (based on predicate)Meets acceptance criteria
    Tubing to connector connection strength (dynamic pull force)Meets acceptance criteria (based on predicate)Meets acceptance criteria
    Needle to wing connection strength (dynamic pull force)Meets acceptance criteria (based on predicate)Meets acceptance criteria
    Wing-needle protector fitMeets acceptance criteria (based on predicate)Meets acceptance criteria
    Needle penetration resistanceMeets acceptance criteria (based on predicate)Meets acceptance criteria
    Blockage testMeets acceptance criteria (based on predicate)Meets acceptance criteria
    Priming volumeMeets acceptance criteria (based on predicate)Meets acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each performance test. It only mentions that "Performance testing was conducted to ensure the Surflo Winged Infusion Set met the predetermined specifications." The data provenance is implied to be from the manufacturer's internal testing facilities in Hangzhou, China, as it is manufactured by "Terumo Medical Products (Hangzhou) Co., Ltd." The testing is described as non-clinical, implying a prospective testing approach conducted specifically for this 510(k) submission.

    3. Number of Experts Used and Qualifications

    Not applicable. This is a non-clinical performance test for a traditional medical device, not a study requiring expert readers for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. This is a non-clinical performance test for a traditional medical device, not a study requiring adjudication of expert readings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a non-clinical performance test for a traditional medical device, not a study evaluating human reader performance with or without AI assistance.

    6. Standalone Performance (Algorithm Only)

    Not applicable. This is a traditional medical device, not an AI/ML algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance tests were adherence to established industry standards (ISO, JIS, GB) and internal company specifications derived from the predicate device's performance.

    8. Sample Size for the Training Set

    Not applicable. This is a traditional medical device, not an AI/ML product requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is a traditional medical device, not an AI/ML product requiring a training set.

    Additional Test Information:

    • Biocompatibility: Tests were conducted on non-aged and accelerated-aged devices according to ISO 10993-1, with results demonstrating biocompatibility throughout the product's shelf life.
    • Sterilization: The device is sterilized to a Sterility Assurance Level (SAL) of 10^-6, validated in accordance with ISO 11135:2014.
    • Clinical Tests: The 510(k) submission states that it "does not include data from clinical tests."
    Ask a Question

    Ask a specific question about this device

    K Number
    K220934
    Date Cleared
    2022-06-29

    (90 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Terumo Medical Products (Hangzhou) Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RADIFOCUS Torque Device is intended to facilitate guidewire manipulation during interventional procedures.

    Device Description

    RADIFOCUS Torque Device is a supportive device intended to facilitate guide wire manipulation during interventional procedures. This product consists of a cap that screws onto the main body of the device (handle) to apply a clamping force on the guide wire. The cap is made of polypropylene and the handle is made of polyoxymethylene. The torque device can be used for the guide wires with diameters of 0.010" to 0.038". RADIFOCUS Torque Device is sterilized with ethylene oxide.

    AI/ML Overview

    The provided text describes a 510(k) submission for the RADIFOCUS Torque Device, a medical device intended to facilitate guidewire manipulation. It specifically states that no clinical tests were performed or deemed necessary for this submission. The evaluation relies on non-clinical performance testing and comparison to a predicate device.

    Therefore, the requested information regarding acceptance criteria and a study proving device performance cannot be fully provided in the context of a clinical study with human subjects, given that no such study was conducted or required.

    However, based on the provided text, I can infer the "acceptance criteria" and "study" refers to the non-clinical performance testing conducted.

    Here's the breakdown of what can be extracted from the provided text regarding the non-clinical performance testing:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "The RADIFOCUS Torque Device tested met the predetermined acceptance criteria." While the specific numerical or qualitative acceptance criteria for each test are not detailed, the results are uniformly positive.

    Performance TestAcceptance Criteria (Implied)Reported Device Performance
    AppearanceMeets predetermined criteriaMeets acceptance criteria
    Fixed strength of guide wireMeets predetermined criteriaMeets acceptance criteria
    Fixed strength of capMeets predetermined criteriaMeets acceptance criteria
    Fixable dimensionMeets predetermined criteriaMeets acceptance criteria
    Guide wire damageMeets predetermined criteriaMeets acceptance criteria
    Collet releaseMeets predetermined criteriaMeets acceptance criteria
    Torque slip forceMeets predetermined criteriaMeets acceptance criteria

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified. The document only mentions "Performance testing was conducted."
    • Data provenance: Not specified. It's implied to be internal testing by the manufacturer (Terumo Medical Products (Hangzhou) Co., Ltd.) given the context of a 510(k) submission and the lack of external study details. The testing is retrospective in the sense that it was conducted before the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as there was no clinical study involving human subjects or expert assessment of clinical outcomes. The "ground truth" for the non-clinical performance tests would be the established engineering specifications and measurement standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable for non-clinical performance testing of a medical device. Adjudication methods are typically used in clinical studies for interpretation of imaging or clinical events.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device is a physical torque device, not an AI-assisted diagnostic tool. No MRMC study was mentioned or implied.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the device is not an algorithm or AI. The performance testing was for the physical device.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" was established by predetermined specifications and engineering standards for mechanical performance (e.g., strength, dimensions, torque).

    8. The sample size for the training set:

    • This is not applicable as there was no AI/algorithm that required a training set. This is a physical medical device.

    9. How the ground truth for the training set was established:

    • This is not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K031279
    Date Cleared
    2003-05-30

    (38 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TERUMO MEDICAL PRODUCTS HANGZHOU CO., LTD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is a winged blood collection needle intended for venipuncture to collect blood specimens from patients.

    The TERUMO® SURSHIELD™ SAFETY WINGED BLOOD COLLECTION SET is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device. This device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy.

    Additionally, after withdraw of the needle from the patient's vein, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

    Device Description

    The Surshield Safety Winged Blood Collection Set manufactured by Terumo Medical Products in Hangzhou, China is a sterile, single-use device consisting of a needle attached to a winged type hub, tubing, female connector with a multi-sample luer adapter attached for blood collection.

    A hinged shield cover is attached to the wing just below the needle-to-wing junction. The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a one- or two- handed technique. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk of needle stick injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the TERUMO® SURSHIELD™ Safety Winged Blood Collection Set, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria for each test. Instead, it states that the performance tests were performed and the conclusion is that the device is "substantially equivalent" to a predicate device, which implies the test results either met the predicate's performance or fell within acceptable manufacturing tolerances. For some tests, the document describes the design intent or functionality, which implicitly forms the basis of acceptance.

    Test PerformedImplied Acceptance Criteria (Based on Substantial Equivalence and Design)Reported Device Performance
    Shield cover/Needle Locking StrengthThe shield cover must effectively and securely lock the needle after activation, preventing accidental exposure.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Break strength of the shield cover jointThe joint must withstand normal handling during activation without breaking, ensuring the safety mechanism remains intact.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Force to lock the needle in the shield cover (Force to activate safety feature)The activation force should be reasonable for user operation while ensuring positive locking.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Puncture Resistance of Shield CoverThe shield cover must adequately resist needle puncture once locked, preventing accidental needlesticks.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Flow RateThe device should allow for efficient blood collection and fluid administration.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Wing to Tubing Connection StrengthConnections must be robust to prevent leaks or detachment during use.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Tubing to Connector Connection StrengthConnections must be robust to prevent leaks or detachment during use.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Needle to Wing Connection StrengthConnections must be robust to prevent leaks or detachment during use.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Needle Penetration ResistanceNot explicitly defined for the safety feature (likely applies to the needle itself, ensuring proper venipuncture efficacy).Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Wing Needle Protector FitThe protector must fit securely on the wing, likely referring to the design and integration of the safety feature.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Leak TestThe device must not leak during normal use for blood collection or fluid administration.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Blockage TestThe device must remain unblocked, allowing for proper fluid flow.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Connection Strength of Winged Infusion Set and Luer Adaptor JointThe connection between the set and luer adaptor must be secure.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Valve Protector FitLikely refers to the fit and function of any valve mechanisms, ensuring proper seal and operation.Not explicitly quantified, but considered "substantially equivalent" to predicate.
    Sterility Assurance Level (SAL)SAL of 10^-6 (implied by EN550 validation to 10^-9 means it meets this common standard).Validated to provide a Sterility Assurance Level (SAL) of 10^-9.
    Ethylene Oxide Residual LevelsEthylene Oxide: 25 ppm; Ethylene Chlorohydrin: 25 ppm.Ethylene Oxide: Does not exceed 25 ppm; Ethylene Chlorohydrin: Does not exceed 25 ppm.
    BiocompatibilityMaterials must be biocompatible according to ISO-10993 Part-1.Blood contacting materials are biocompatible.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the sample size for any of the performance tests.
    • Data Provenance: The tests were performed on the "Surshield Safety Winged Blood Collection Set manufactured by Terumo Medical Products in Hangzhou, China." The data is thus prospective for this specific device model from that manufacturing location.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The study described is primarily a device performance and substantial equivalence assessment, not one relying on expert interpretation of results. Physical and chemical tests were performed.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the nature of the tests (mechanical, chemical, flow rates), an adjudication method as typically used for clinical or imaging data (e.g., 2+1) would not be applicable. The tests likely involved standardized methods with objective pass/fail criteria or measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document is for a physical medical device (blood collection set), not an AI-powered diagnostic or interpretive system. Therefore, the concept of "human readers" or "AI assistance" is not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    No, this is not an algorithm-only or AI-based device. It is a physical medical device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests was based on engineering specifications, industry standards (e.g., EN550, ISO-10993), and the performance characteristics of the predicate device. For example, the ground truth for "Shield cover/Needle Locking Strength" would be defined by the predicate device's measured locking strength or a specific force threshold required by relevant standards for safety features.

    8. The Sample Size for the Training Set

    Not Applicable. This is a physical device, not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As above, no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1