Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K243158
    Manufacturer
    Date Cleared
    2025-01-23

    (115 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TeraRecon Aorta.CT is intended to provide an automatic 3D segmentation and label anatomical landmarks of the Aorta. The results of TeraRecon Aorta.CT are intended to be used in conjunction with other patient information by trained professionals who are responsible for making any patient management decision per the standard of care. TeraRecon Aorta.CT is a software as a medical device (SaMD) deployed as a containerized application. The device inputs are CT Angiography with contrast DICOM images. The device outputs are DICOM result files which may be viewed utilizing DICOM-compliant systems. The device does not alter the original input data and does not provide a diagnosis.

    TeraRecon Aorta.CT is indicated to generate results from aortic CT Angiography scans taken of adult patients except patients with pre-existing aortic device, bicuspid aortic valve anomaly, aortic dissection, aortic rupture, and abdominal metallic devices. The device is not specific to any gender, ethnic group, or clinical condition.

    Device Description

    The TeraRecon Aorta.CT algorithm is an image processing software device that can be deployed as a containerized application (e.g., Docker container) that runs on off-the-shelf hardware or on a cloud platform.

    The device provides an automatic 3D segmentation of the aorta and landmarks of important aortic anatomy. When TeraRecon Aorta.CT results are used in external viewer devices such as TeraRecon's Intuition or Eureka Clinical Al medical devices, all the standard features offered by the external viewer are employed.

    The TeraRecon Aorta.CT algorithm is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by individuals that have been trained in the software's function, capabilities, and limitations.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the TeraRecon Aorta.CT (1.1.0) device:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    Lumen SegmentationMean DICE score >= 80%Mean DICE score: 88% (Passed)
    Aorta SegmentationMean DICE score >= 80%Mean DICE score: 90% (Passed)
    Landmarking (Overall)Each of the 22 landmarks independently pass class-specific criteria in 80% of cases. Lower bound of the 95% exact binomial confidence interval >= 70%.All landmarks passed the acceptance criteria, all 95% confidence intervals were at least 70%. (Passed)
    Landmarking (Specific Criteria):
    Common Left/Right Iliac Arteries, Left/Right Femoral Arteries (4 landmarks)Correct identification of the vessel in accordance with ground truth.Not explicitly stated with individual percentages, but included in the overall "all landmarks passed" statement.
    Remaining 17 Landmarks (except aortic bifurcation)Euclidean distance between ground truth annotation and medical device output locations within 5mm.Not explicitly stated with individual percentages, but included in the overall "all landmarks passed" statement.
    Aortic Bifurcation (1 landmark)Euclidean distance between ground truth annotation and medical device output locations within 2cm.Not explicitly stated with individual percentages, but included in the overall "all landmarks passed" statement.

    2. Sample Size Used for the Test Set and Data Provenance

    • Initial Sample Size for Test Set: 170 CTA scans for segmentation and landmarking.
    • Adjusted Sample Size for Landmarking: 170 initial studies + 29 supplemental studies = 199 studies (to achieve a target of 70 annotatable landmarks per target).
    • Data Provenance: Retrospective cohort study. At least 50% of the ground truth data is from US patients across 3 geographical regions in the United States. The validation data was enriched with data from patients with clinical diagnosis of aortic dilation/aneurysm and/or aortic valve disease. The final manufacturer distribution of scanner types was 77 Siemens, 33 GE, 35 Philips, and 25 Canon.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: Not explicitly stated as a count, but referred to as "annotators" for landmarking and a "US board certified radiologist" for checking collected datasets.
    • Qualifications of Experts: The individual who checked collected datasets was a "US board certified radiologist, who is currently practicing in the United States and reads similar scans." The qualifications of the "annotators" for landmarking are not specified beyond their task.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It implies that ground truth was established by experts (radiologist and annotators) and then compared to the device output. There is no mention of a process for resolving discrepancies among multiple experts or between expert and device output outside of direct comparison.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study that involves human readers with and without AI assistance was not explicitly described or presented in the provided text. The study focuses on evaluating the standalone performance of the AI device against ground truth.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone study was performed. The described study evaluates the Aorta.CT device's performance (segmentation DICE scores and landmarking accuracy) directly against expert-established ground truth. There is no mention of human-in-the-loop performance evaluation in this specific study.

    7. The Type of Ground Truth Used

    • Segmentation (Lumen and Aorta Wall): Expert annotations as described by "Comparison of Aorta lumen segmentation results from the medical device to aorta lumen segmentation from ground truth" and "This aorta wall to wall segmentation includes aorta lumen + wall for the comparison."
    • Landmarking: Expert annotations as described by "We will also examine the subject device landmarking locations compared to each ground truth annotation."

    8. The Sample Size for the Training Set

    The sample size for the training set is not provided in the given text. The document focuses on the validation study and its test set.

    9. How the Ground Truth for the Training Set was Established

    The method for establishing ground truth for the training set is not provided in the given text. The document only describes how ground truth was established for the retrospective cohort test set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241312
    Manufacturer
    Date Cleared
    2024-11-05

    (180 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon,Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This medical device is intended to segment Cardiac chambers as anatomical structures on contrast or non-contrast MR scans of adult patients that undergo Cardiac MR procedures. The algorithm is not specific to any gender or ethnic group or clinical conditions.

    Device Description

    The TeraRecon Cardiac.Chambers.MR algorithm is comprised of two components:

    1. The TeraRecon Cardiac.Chambers.MR for Cine-ax
    2. The TeraRecon Cardiac.Chambers.MR De-ax

    1: The TeraRecon Cardiac.Chambers.MR for Cine-ax
    The TeraReconCardiac.Chambers.MR algorithm for Cine-ax is an image processing software device that can be deployed as a containerized application (e.g.,Docker container).

    The TeraRecon Cardiac.Chambers.MR algorithm for Cine-ax automatically detects and identifies the heart location and derives left ventricular (LV) and right ventricular (RV) myocardium segmentation on DICOM-compliant cardiovascular MR images of different cardiac imaging sequences.

    The TeraRecon Cardiac.Chambers.MR for Cine-ax algorithm performs a segmentation (or tracing) around the epicardial border as well as the endocardial border wall. For the RV the algorithm segments the endocardial border wall.

    2: The TeraRecon Cardiac.Chambers.MR for De-ax
    The TeraReconCardiac.Chambers.MR algorithm for De-ax is an image processing software device that can be deployed as a containerized application (e.g.,Docker container).

    The TeraRecon Cardiac.Chambers.MR algorithm for De-ax automatically detects and identifies the heart location and derives left ventricular (LV) myocardium segmentation on DICOM-compliant cardiovascular MR images of different cardiac imaging sequences.

    The TeraRecon Cardiac.Chambers.MR for De-ax algorithm performs a segmentation (or tracing) around the epicardial border as well as the endocardial border wall.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the reported performance relative to a threshold (DICE score). While specific thresholds aren't explicitly stated as "acceptance criteria," the text confirms the device passed these criteria. We infer the thresholds from the reported mean DICE scores that met the acceptance criteria.

    MetricAcceptance Criteria (Inferred from passing results)Reported Device Performance (Mean DICE Score)95% Confidence Interval
    Cine-ax Algorithm
    LV Myocardium DICE score≥ 0.810.82(0.81, 0.83)
    LV Chambers DICE score≥ 0.900.90(0.89, 0.91)
    RV Chamber DICE score≥ 0.820.84(0.82, 0.85)
    De-ax Algorithm
    LV Myocardium DICE score≥ 0.750.79(0.75, 0.83)
    LV Chambers DICE score≥ 0.840.88(0.84, 0.92)

    Notes on Acceptance Criteria:

    • The document states: "The results of the Cine-ax algorithm showed the LV Myocardium mean DICE scores for were within the acceptance criteria at 0.82 (0.81,0.83)... The results indicated the LV Chambers mean DICE Scores were within the acceptance criteria and were at or greater than 0.90 (0.89, 0.91)... and the mean DICE Score for RV Chamber was 0.84 (0.82, 0.85) and thus within the 95% confidence interval and passing the DICE limit score." This implies the lower bound of the 95% CI or a value just below the reported mean was the threshold. For simplicity and clarity, I've listed the lower bound of the reported 95% CI where available as the inferred minimum acceptance criterion, or the stated passing value.
    • For the De-ax algorithm, it states: "The results of the De-ax algorithm showed the LV myocardium mean DICE scores were within the acceptance criteria at 0.79 (0.75, 0.83), and the LV Chamber mean DICE scores were within the acceptance criteria at 0.88 (0.84, 0.92)." This implies the lower bound of the 95% CI was the threshold.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): 100 adult patients (50 for the Cine-ax algorithm, 50 for the De-ax algorithm).
    • Data Provenance: Retrospective cohort study.
      • 82% of studies came from the United States.
      • 18% of studies came from Europe.
      • Data was collected from different sites than the training data.
      • Included data from three MR equipment manufacturers: GE, Philips, or Siemens (covering 74% of US scanners).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: At least one (referred to as "a board certified radiologist").
    • Qualifications: A board-certified radiologist practicing in the United States with experience in reviewing cMR (cardiac MR) studies.

    4. Adjudication Method for the Test Set

    • The document states: "All collected datasets were reviewed by a board certified radiologist practicing in the United States with experience in reviewing cMR studies. The radiologists evaluated whether each cMR scan met the inclusion/exclusion criteria, and if the study did not meet the inclusion exclusion criteria then the study was replaced."
    • This suggests a single-reader read for inclusion/exclusion criteria, followed by implicit establishment of ground truth by this single expert (or group of experts if "radiologists" implied more than one but written singular elsewhere). An explicit adjudication method (e.g., 2+1, 3+1) for the actual segmentation ground truth is not detailed, implying either a single expert's consensus or a pre-established "ground truth" that the single radiologist reviewed for study suitability. The sentence "The next phase of the study was to collect and compare the TeraRecon Cardiac.Chambers.MR cine-ax algorithm output to the created ground truth as described below" further suggests either a pre-existing ground truth or one established by the single reviewing radiologist, but not a multi-reader adjudication process for the segmentation itself.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No, the provided text does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The study focused on the standalone performance of the algorithm against established ground truth.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone study was performed. The study compared the "TeraRecon Cardiac.Chambers.MR [algorithm] output to the created ground truth." This indicates an evaluation of the algorithm's performance without direct human interaction or assistance during the segmentation process.

    7. The Type of Ground Truth Used

    • The ground truth was established by expert review ("All collected datasets were reviewed by a board certified radiologist practicing in the United States with experience in reviewing cMR studies"). The ground truth itself consisted of segmented cardiac chambers (LV Myocardium, LV Chamber, RV Chamber, LV only myocardium and chamber for de-ax) against which the algorithm's DICE scores were calculated. While "expert consensus" is often multi-reader, the text describes a single expert reviewing the dataset for suitability and implying their role in the "created ground truth." Pathology or outcomes data were not mentioned as ground truth sources.

    8. The Sample Size for the Training Set

    • The sample size for the training set is not specified in the provided text. The document only states: "It is ensured that patient data received for training of the algorithm is from different sites than the validation data utilized for this study."

    9. How the Ground Truth for the Training Set was Established

    • The method for establishing ground truth for the training set is not specified in the provided text. It only mentions that the training data came from different sites than the validation data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220349
    Device Name
    TeraRecon Neuro
    Manufacturer
    Date Cleared
    2022-08-12

    (186 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TeraRecon Neuro Algorithm is an algorithm for use by trained professionals, including but not limited to physicians, surgeons and medical clinicians.

    The TeraRecon Neuro Algorithm is a standalone image processing software device that can be deployed as a Microsoft Windows executable on off-the-shelf hardware or as a containerized application (e.g., a Docker container) that runs on off-the-shelf hardware or on a cloud platform. Data and images are acquired via DICOM compliant imaging devices. DICOM results may be exported, combined with, or utilized by other DICOM-compliant systems and results.

    The TeraRecon Neuro Algorithm provides analysis capabilities for functional, dynamic, and derived imaging datasets acquired with CT or MRI. It can be used for the analysis of dynamic brain perfusion image data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to brain tissue perfusion, vascular assessment, tissue blood volume, and other parametric maps with or without the ventricles included in the calculation. The algorithm also include volume reformat in various orientation, rotational MIP 3D batch while removing the skull. This "tumble view" allows qualitative review of vascular structure in direct correlation to the perfusion maps for comprehensive review.

    The results of the TeraRecon Neuro Algorithm can be delivered to the end-user through image viewers such as TeraRecon's Aquarius Intuition system, TeraRecon's Eureka AI Results Explorer, TeraRecon's Eureka Clinical AI Platform, or other image viewing systems like PACS that can support DICOM results generated by the TeraRecon Neuro Algorithm.

    The TeraRecon Neuro Algorithm results are designed for use by trained healthcare professionals and are intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions.

    Device Description

    The TeraRecon Neuro algorithm version 2.0.0 is a modification of the predicate device Neuro.AI Algorithm (K200750), which was a modification of the predicate device, Intuition-TDA, TVA, Parametric Mapping (which was cleared under K131447). The predicate device Intuition -TDA, TVA, Parametric Mapping is an optional module/workflow for the Intuition system (K121916). The TeraRecon Neuro algorithm is an image processing software device that can be deployed as a Microsoft Windows executable on off-the-shelf hardware or as a containerized application (e.g., Docker container) that runs on off-the-shelf hardware or on a cloud platform. The device has limited network connectivity or external medical support.

    TeraRecon Neuro allows motion correction and processes, calculates and outputs brain perfusion analysis results for functional, dynamic, and derived imaging datasets acquired with CT or MRI. TeraRecon Neuro results are used for the analysis of dynamic brain perfusion image data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to brain tissue perfusion, vascular assessment and tissue blood volume.

    Outputs include parametric map of measurements including time to peak (TTP), take off time (TOT), recirculation time (RT), mean transit time (MTT), blood volume (BV/CBV), blood flow (BF/CBF), time to maximum (Tmax) and penumbra/umbra maps that are derived from combinations of measurement parameters, such as mismatch maps and hypoperfusion maps with volumes and ratios, as well as 2D and 3D visualization of brain tissues and brain blood vessels (Note: Tmax, mismatch and hypoperfusion maps are only available for images of CT modality).

    When TeraRecon Neuro results are used in external viewer devices such as TeraRecon's Intuition or Eureka medical devices, all the standard features offered by Intuition or Eureka are employed such as image manipulation tools like drawing the region of interest, manual or automatic segmentation of structures, tools that support creation of a report, transmitting and storing this report in digital form, and tracking historical information about the studies analyzed by the software.

    The TeraRecon Neuro algorithm outputs can be used by physicians to aid in the diagnosis and for clinical decision support including treatment planning and post treatment evaluation. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by individuals that have been trained in the software's function, capabilities and limitations. The device is intended to provide supporting analytical tools to a physician, to speed decision-making and to improve communication, but the physician's judgment is paramount, and it is normal practice for physicians to validate theories and treatment decisions multiple ways before proceeding with a risky course of patient management.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Software Acceptance CriteriaAll pre-defined acceptance criteria for the Neuro.AI Algorithm were met, and all software test cases passed during software development and testing in accordance with IEC 62304:2006/AI:2015.
    Qualitative Clinical User EvaluationThe generated maps of TeraRecon Neuro were confirmed through qualitative assessment to be at least 85% substantially equivalent or better than the predicate and reference devices.
    Quantitative Tmax Measurement AccuracySubject device limit of agreement for both absolute error and absolute percent error (of Tmax measurements compared to ground truth, defined as the average Tmax of two reference devices) was less than or equal to the limit of agreement of each predicate device compared to the ground truth.
    Safety and EffectivenessThe TeraRecon Neuro device meets its qualified requirements, performs as intended, and is as safe and effective as the predicate device. No new or different questions of safety or efficacy have been raised. All risks were analyzed, and there are no new risks or modified risks that could result in significant harm which are not effectively mitigated in the predicate device. The device is determined to be Substantially Equivalent to the predicate device in terms of safety, efficacy, and performance.

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the test set used in the qualitative clinical user evaluation or the quantitative Tmax measurement accuracy study. It refers to "comparison maps generated by the subject device, the predicate device and two additional reference devices." Without specific numbers, it's impossible to determine the precise size of the test set cases.

    Regarding data provenance, the document does not provide details on the country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: One expert was used.
    • Qualifications: Dr. Robert Falk, MD. No additional details about his specific experience or sub-specialty (e.g., radiologist with X years of experience) are provided in the text.

    4. Adjudication Method for the Test Set

    The adjudication method used for the clinical user evaluation was not explicitly specified as 2+1, 3+1, or any other formal method. The study involved a single evaluator (Dr. Robert Falk, MD) who was "asked to confirm through qualitative assessment." This suggests a single-expert review, rather than a multi-expert adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Improvement with AI vs. Without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described. The evaluation involved a single expert providing a qualitative assessment. The study was focused on demonstrating substantial equivalence to predicate and reference devices, not on measuring the improvement of human readers with AI assistance. Therefore, there is no reported effect size of how much human readers improve with AI vs. without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was conducted for the quantitative Tmax measurement. The acceptance criteria for Tmax accuracy were based on comparing the subject device's measurements directly against the ground truth (average of reference devices) in ROIs, without explicit human intervention in the measurement process for the test cases. While the "ground truth" itself is derived from other devices (which are used by humans), the comparison of the algorithm's output to this ground truth represents a standalone assessment of the algorithm's quantitative accuracy.

    7. The Type of Ground Truth Used

    • Qualitative Clinical User Evaluation: The ground truth for this evaluation appears to be the performance of the predicate and reference devices, as the subject device's maps were compared to these for substantial equivalence. It's a comparative assessment rather than an absolute ground truth (e.g., pathology).
    • Quantitative Tmax Measurement Accuracy: The ground truth for Tmax measurements was defined as the average Tmax measurement of the two reference devices (GE Medical Systems FastStroke CT Perfusion 4D (K193289) and ISchemaView RAPID (K182130)) for a given ROI.

    8. The Sample Size for the Training Set

    The document does not provide any information regarding the sample size used for the training set for the TeraRecon Neuro algorithm.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information on how the ground truth for the training set was established. Training set details are not discussed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200750
    Manufacturer
    Date Cleared
    2020-11-06

    (228 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neuro.AI Algorithm is an algorithm for use by trained professionals, including but not limited to physicians, surgeons and medical clinicians.

    The Neuro.Al Algorithm is a standalone image processing software device that can be deployed as a Microsoft Windows® executable on off-the-shelf hardware or as a containerized application (e.g., a Docker container) that runs on off-the-shelf hardware or on a cloud platform. Data and images are acquired via DICOM compliant imaging devices. DICOM results may be exported, combined with or utilized by other DICOM-compliant systems and results.

    The Neuro.AI algorithm provides analysis capabilities for static, functional, dynamic and derived imaging datasets acquired with CT or MRI. It can be used for the analysis of dynamic brain image data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to brain tissue perfusion, vasular assessment and tissue blood volume and other parametric maps with or without the ventricles included in the calculation. The algorithm also includes volume reformat in various orientions, rotational MIP 3D batch while removing the skull. This "tumble view" allows qualitative review of vascular structure in direct correlation to the perfusion maps for comprehensive review.

    The results of the Neuro.AI Algorithm can be delivered to the end-user through image viewers such as TeraRecon's Aquarius iNtuition system, TeraRecon's Northstar AI Results Explorer, or other image viewing systems like PACS that can support DICOM results generated by Neuro.AI.

    The Neuro.AI Algorithm results are designed for use by trained healthcare professionals and are intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions.

    Device Description

    The Neuro.Al Algorithm is a modification of the predicate device, iNtuition-TDA, TVA, Parametric Mapping which was cleared under K131447. The predicate device is an optional module/workflow for the iNtuition system (K121916). The Neuro.Al Algorithm is a standalone image processing software device that can be deployed as a Microsoft® Windows executable on off-the-shelf hardware or as a containerized application (e.g., Docker container) that runs on off-the-shelf hardware or on a cloud platform. The device has limited network connectivity or external medical support.

    The Neuro.Al Algorithm allows motion correction and processes, calculates and outputs brain perfusion analysis results for static, functional, dynamic and derived imaging datasets acquired with CT or MRI. Neuro.Al results are used for visualization and analysis of dynamic brain perfusion image data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to brain tissue perfusion, vascular assessment displayed in rotational Maximum Intensity Projection (MIP) called the tumble view, and tissue blood volume and other parametric maps with or without brain ventricles included in the calculation.

    Outputs include text and parametric map displays of measurements including time to peak (TTP), take off time (TOT), recirculation time (RT), mean transit time (MTT), blood volume (BV/CBV), blood flow (BF/CBF), classification maps, reformatted images and rotational MIPs for 2D and 3D visualization of brain tissues and blood vessels, and for correlation to the perfusion maps.

    The results of the Neuro.Al Algorithm can be delivered to the end-user through image viewers such as TeraRecon's iNtuition system, TeraRecon's Northstar Al Results Explorer ("Northstar"), or other third-party image viewing systems like PACS that can display the DICOM results generated by Neuro.Al output does not depend on the viewing system's capabilities as the results are self-contained and the only interface is through DICOM.

    When the Neuro.Al Algorithm results are used on iNtuition, all the standard features offered by iNtuition are employed such as image manipulation tools like drawing the region of interest, manual or automatic segmentation of structures, tools that support creation of a report, transmitting and storing this report in digital form, and tracking historical information about the studies analyzed by the software.

    The Neuro.Al algorithm can be used by physicians to aid in the diagnosis. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by individuals that have been trained in the software's function, capabilities and limitations. The device is intended to provide supporting analytical tools to a physician, to speed decision-making and to improve communication, but the physician's judgment is paramount, and it is normal practice for physicians to validate theories and treatment decisions multiple ways before proceeding with a risky course of patient management.

    AI/ML Overview

    The provided document describes the Neuro.AI Algorithm and its substantial equivalence to a predicate device, iNtuition-TDA, TVA, Parametric Mapping (K131447). However, it does not contain a detailed performance study with specific acceptance criteria and reported device performance in the format requested. The document focuses on regulatory compliance, outlining the device's indications for use, technological characteristics, and a general statement about software verification and validation.

    Therefore, many of the requested items cannot be extracted directly from this document.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Not provided in the document. The text states: "During software testing, all predefined acceptance criteria for the Neuro.Al Algorithm were met and all software test cases passed." However, it does not specify what those acceptance criteria were or provide a table of performance metrics.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not provided in the document. The document mentions "software testing and performance evaluation" but does not detail the test set's sample size or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not provided in the document. The document describes the device's intended use by "trained professionals, including but not limited to physicians, surgeons and medical clinicians" but doesn't specify how ground truth was established for testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not provided in the document. The document does not describe a comparative effectiveness study involving human readers with and without AI assistance. The focus is on the device's substantial equivalence to a predicate device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, implicitly. The document describes the "Neuro.AI Algorithm as a standalone image processing software device." The testing mentioned ("software testing and performance evaluation") would inherently be evaluating the algorithm's standalone performance against its predefined acceptance criteria, even if those criteria aren't explicitly detailed. The statement "The Neuro.AI Algorithm is as safe and effective as the predicate device" implies standalone testing for functional equivalence.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not explicitly provided in the document. While the device assists in diagnosis, the method for establishing ground truth for testing is not described.

    8. The sample size for the training set

    Not provided in the document. The document describes a "510(k) summary," which focuses on demonstrating substantial equivalence to a predicate device rather than detailing AI model development specifics like training set size.

    9. How the ground truth for the training set was established

    Not provided in the document. Similar to the training set size, the method for establishing ground truth for training is not included in this regulatory summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191585
    Manufacturer
    Date Cleared
    2019-07-12

    (28 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iNtuition-Structural Heart Module is a software solution that is intended to assist Cardiologists, and Clinical Specialists with the visualization and measurements of the heart and vessels.

    iNtuition-Structural Heart Module enables the user to:

    • · Visualize and measure (diameters, lengths, angles, areas and volumes) structures of the heart and vessels for pre-operative planning and sizing for cardiovascular interventions and for post-operative evaluation.
    • · Quantify calcium (volume, density)

    iNtuition-Structural Heart Module has the following tools and features that facilitate:

    • · Automatic and manual centerline detection.
    • · Segmentation of cardiovascular structures.
    • · Measurement tools (diameters, lengths, areas, volumes, angles) for the dimensions of vessels and structures.
    • · Calcium quantification and scoring.
    • · Various visualization techniques: 2D/3D/4D visualization, MPR, Stretched MPR, MIP, MinlP, Raysum and MAR.
    • · Capture and Report.
    Device Description

    iNtuition-Structural Heart Module is an optional image post-processing software module using iNtuition (K121916) standard features and part of its optional features. It is a software device generally used with the off-the-shelf hardware, offered in various configurations, with the simplest configuration being a stand-alone workstation capable of image review, communications, archiving, database maintenance, remote review, reporting and basic 3D capabilities. It can also be configured as a server with some, or with all, or none of its optional features.

    Whether provided as a workstation or a server, the iNtuition-Structural Heart Module software is designed to provide access by a local user physically sitting at the computer hosting the iNtuition server software, and/or by one or more remote users who concurrently connect to the server using a freely-downloadable thin client application or through a zero-footprint web viewer (with conference capabilities) over local network or internet.

    iNtuition-Structural Heart Module is iNtuition (K121916) based optional feature and employs all standard features offered by iNtuition such as convenient image manipulation tools like drawing of region of interests, manual and automatic segmentation of structures, image assessment and measurement tools - linear, diameter, angle, area and volume and tools that support the creation of reports, transmitting and storing this report in digital form and tracking historical information about the studies analyzed by the software. iNtuition Vessel analysis and calcium scoring features are utilized to support automatic and manual centerline extraction and analysis and calcium quantification.

    iNtuition-Structural Heart Module:

    • . Supports the visualization and quantification of coronary vessels and cardiac structures for anatomic and pre- or post-operative evaluations through guided clinical workflows.
    • . Enables the assessment and measurement of different structures of the heart, e.g. aorta, aortic valves, mitral valve, pulmonic valve, atria and atrial appendages, and ventricles.
    • . Provides analysis of the feasibility of a transapical, transfemoral or subclavian approach to structures for replacement or repair procedures via 3D measurements.
    • Uses the same iNtuition (K121916) Vessel Analysis and Calcium modules.Enables .
      assessment and measurement of vessels and can help identify calcifications, aneurysms and other anomalies to quickly and reliably prepare for various types of vascular procedures.Supports the creation, transmission and storage of a report in digital form. It can also track historical information about the studies analyzed by the software.Displays results analysis, that can be printed as hardcopy or saved in a variety of formats to a hard disk, network, PACS system or CD/DVD/USB.
    AI/ML Overview

    The provided text is a 510(k) Premarket Notification submission for the TeraRecon iNtuition-Structural Heart Module. It is a regulatory document declaring substantial equivalence to predicate devices, rather than a detailed study report proving the device meets specific acceptance criteria based on performance metrics.

    The document explicitly states: "iNtuition-Structural Heart Module did not require clinical studies to demonstrate its safety and effectiveness." This means that the information you've requested regarding performance data, sample sizes, ground truth establishment, expert adjudication, and MRMC studies for demonstrating the device meets acceptance criteria via a pre-defined study is not present in this regulatory filing.

    Therefore, I cannot extract structured information about specific acceptance criteria and a study proving the device meets them from this document. The FDA clearance is based on a claim of substantial equivalence to existing predicate devices, meaning it has similar indications for use and technological characteristics, and therefore does not raise new questions of safety or effectiveness.

    The "Performance Data" section (Page 10) only mentions:

    • "The verification and Validation tests have been performed to address the indication for use, the technological characteristics claims, requirement specifications and risk management results."
    • "Software testing and validation were done according to written test protocols established before testing was conducted."
    • "Test results were reviewed by designated technical professionals before being formalized and after ensuring that the software fully satisfies all expected and previously defined system requirements and features. Test results support the conclusion that iNtuition- Structural Heart Module performance satisfies the design intent and is equivalent to its predicate devices."

    This describes internal software verification and validation activities, which are typically for functional correctness and adherence to design specifications, not a comparative clinical performance study against specific, quantified acceptance criteria for diagnostic accuracy or clinical outcomes that would usually involve human readers or external ground truth.

    In summary, the document states that a clinical performance study was not required. Thus, I cannot provide the details you requested in the structured format, as the information is not available in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180916
    Manufacturer
    Date Cleared
    2018-09-24

    (168 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TeraRecon Inc.,

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iNtuition-T1 Mapping and T2/T2* Mapping are software modules that support the derivation and quantification of T1, T2 and T2* values from MR DICOM image pixel intensities and header information. The quantification of these parameters can be used to characterize tissues. Support is provided to overlay the T1, T2, and T2* values using colormaps on related MR images.

    Support is provided for using different colormaps to overlay different ranges of T1, T2 or T2* values and restrict the overlay to region of interest on the images can be of simple planar scan like a single slice or volumetric or 4D scans of a body part. iNtuition-T1 Mapping and T2/T2* Mapping are iNtuition software features that can be used in multiple workflows or be used as basic tools for cardiac functionality, the overlaid images can be captured and forwarded to other systems using standards such as DICOM or http protocol. Quantitative analysis is derived and available as text and graphical display.

    iNtuition- T1 Mapping and T2/T2* Mapping qualitation can be used in a clinical setting on MR images of an individual patient and can be used to support the clinical decision making for the patient. iNtuition- T1 Mapping and T2/T2* Mapping are designed for use by healthcare professionals and are intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions.

    Device Description

    iNtuition-T1 Mapping and T2/T2* Mapping is an optional image post-processing module, part of iNtuition (K121916), which is software only device generally used with the off-the-shelf hardware, offered in various configuration, with the simplest configuration being a stand-alone workstation capable of image review, communications, archiving, database maintenance, remote review, reporting and basic 3D capabilities. It can also be configured as a server with some, all, or none of its optional features disabled. Whether provided as a workstation or a server, the iNtuition software is designed to provide access by a local user physically sitting at the computer hosting the iNtuition server software, and/or by one or more remote users who concurrently connect to the server using a freely-downloadable thin client application or through a zero-footprint web viewer (with conference capabilities) over local network or internet.

    iNtuition-T1 Mapping and T2/T2* Mapping feature can derive quantitative values from intensities and header information of specific MR scan sequences that are specifically coded to enable such derivation (such as Look-Locker and MOLLI for T1.) The quantification of these parameters can be used to derive clinical value such as T2*.. Post-processing such as computation of statistics like volume, area, min/max or various combinations of the derived values, over regions of interests or overlay the derived values using a colormap on related images or a region of the images. The region of interests can be defined by the user through manual, semi-automatic or automatic segmentation techniques provided by iNtuition. The derivation and post-processing can be used with planar, volumetric or 4D scan sequences for cardiac functionality.

    iNtuition-T1 Mapping and T2/T2* Mapping is an iNtuition based optional features, and employ all standard features offered by iNtuition such as convenient image manipulation tools like drawing region of interest, manual or automatic segmentation of structures and tools that support creation of a report, transmitting and storing this report in digital form, and tracking historical information about the studies analyzed by the software.

    This device is intended only to assists the operator in making decisions. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people that have been trained in the software's function (iNtuition), capabilities and limitations. The device is intended to provide supporting analytical tools to a physician, to speed decision-making and to improve communication, but the physician's judgment is paramount and it is normal practice for physicians to validate theories and treatment decisions multiple ways before proceeding with a risky course of patient management.

    iNtuition-T1 Mapping and T2/T2* Mapping modules may be sold separately or as an extension of iNtuition.

    AI/ML Overview

    This submission for K180916 does not provide specific acceptance criteria or a study demonstrating that the device meets those criteria, as it is a Traditional 510(k) stating the product is substantially equivalent to a predicate device and did not require clinical studies.

    Therefore, many of the requested details cannot be extracted from the provided text.

    However, based on the principle of substantial equivalence, the "acceptance criteria" are generally that the device performs as well as the predicate device across its intended use and technological characteristics.

    Here's a breakdown of what can be inferred or stated from the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since no specific performance metrics or acceptance criteria are listed, this table cannot be populated directly. The document repeatedly states that the device is "substantially equivalent" to its predicate and "performs as well as the predicate device" in terms of its intended use and technological characteristics.

    Acceptance CriteriaReported Device Performance
    Not specified in the documentNot specified in the document

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The document mentions "Software verification and validation was completed in accordance with internal processes" and "Performance testing was carried out according to internal company procedures." This implies internal testing rather than a large-scale external test set with specific patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. Software testing was reviewed by "designated technical professionals."

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not specified.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • MRMC Study: No. The document explicitly states: "The subject of this traditional 510k notification, iNtuition-T1 Mapping and T2/T2* Mapping, did not require clinical studies to show safety and effectiveness of the software." Therefore, no MRMC study was conducted.
    • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: Not applicable, as no MRMC study was performed.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Standalone Study: Not explicitly detailed as a separate "study" with performance metrics. The document describes the software's functionality in deriving and quantifying T1, T2, and T2* values, and mentions "Software verification and validation" and "Performance testing" were conducted internally to ensure it met design intent and was equivalent to the predicate. This would constitute standalone testing of the algorithm's output against expected results, but the specifics are not provided.

    7. Type of Ground Truth Used:

    • Type of Ground Truth: Not explicitly stated. For internal performance testing of quantitative measurements (like T1/T2/T2* values), ground truth would likely be established through:
      • Reference standards/phantoms: Using known values.
      • Comparison to predicate device's output: Ensuring the new device's output matches that of the already cleared predicate.
      • Manual calculations/expert evaluation: For regions of interest.

    The document indicates the software supports "derivation and quantification of T1, T2 and T2* values from MR DICOM image pixel intensities and header information," suggesting the ground truth for these values would be based on the principles of MRI physics and potentially established clinical methods for calculating these parameters.

    8. Sample Size for the Training Set:

    • Sample Size for Training Set: Not specified. The document describes the software as an "optional image post-processing module" that derives quantitative values from specific MR scan sequences. This doesn't inherently suggest a machine learning model that requires a "training set" in the common sense (i.e., for supervised learning). It's more about algorithmic derivation and quantification. If any machine learning components were involved, the training set details are not disclosed.

    9. How the Ground Truth for the Training Set was Established:

    • How Ground Truth for Training Set was Established: Not applicable/not specified, as training set details are not provided and the primary function described is algorithmic derivation rather than machine learning inference.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131447
    Manufacturer
    Date Cleared
    2013-12-24

    (218 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TERARECON, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iNtuition-TDA, TVA, and Parametric Mapping are software modules which supports assessment of time-dependent behavior of image intensity, dimensions or volume of regions of interest over time, for volumetric or planar dynamic image types such as CT or MR. Parametric mapping tools encode in color various parameters derived from the temporal or spatial characteristics of the planar or volumetric data.

    Support is provided for digital image processing to derive metadata or new images from input image sets, for internal use or for forwarding to other devices using the DICOM protocol. Image processing tools are provided to extract metadata to derive parametric images from combinations of multiple input images.

    iNtuition-TDA, TVA and Parametric Mapping are iNtution based software features with dedicated workflows and basic tools and thus support post-processing, displaying and manipulation of reports and medical images from acquisition devices and visualization in 2D, 3D and 4D for single or multiple datasets, or combinations thereof.

    iNtuition-TDA, TVA, Parametric Mapping are designed for use by healthcare professionals and are intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions.

    Device Description

    iNtuition - TDA. IVA. Parametric Mapping are post-processing modules, part of iNtuition, which is a software device generally used with off-the-shelf hardware. offered in various configurations, with the simplest configuration being a stand-alone workstation capable of image review, communications, archiving, database maintenance, remote review, reporting and basic 3D capabilities. It can also be configured as a server with some, all, or none of its optional features disabled. A fully-configured iNtuition system is capable of various image processing and visualization functions to support the physician in medical image reviewing.

    iNtuition - TDA, TVA, Parametric Mapping intended used is to provide solutions to various medical image analysis and viewing problems, which come about as modalities generate more and more images. They also support image distribution over networks, and are DICOM compliant.

    iNtuition Time-Dependent Analysis (TDA) and Time-Volume Analysis (TVA) features can obtain quantitative information relating to the evolution of the intensity, density or dimensions of certain regions of CT. MR or other images over time. Statistical analysis such as a histogram representation of the image density values in an image is supported, and analysis of changes in volume over time from multi-phase volumetric images; for example, eiection fraction and stroke volume measurement calculation can be performed using the Time-Volume Analysis tools.

    iNtuition Parametric Mapping tools encode in color various parameters derived from the temporal or spatial characteristics of the planar or volumetric data.

    iNtuition - TDA, TVA and Parametric Mapping are iNtuition-based optional features, and employ all standard features offered by iNtuition, such as convenient tools to support creation of a report, transmitting and storing this report in digital form, and tracking historical information about the studies analyzed with the software.

    These three modules can be sold separately or as a part of the bigger iNtuition package.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria for quantitative device performance or a study explicitly proving the device meets such criteria. Instead, it focuses on demonstrating substantial equivalence to predicate devices and adherence to internal company procedures and voluntary industry standards.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    No specific, measurable acceptance criteria with corresponding performance metrics are reported in the document. The general acceptance is that the device "fully satisfies all expected and previously defined system requirements and features" and is "substantially equivalent to and perform as well as the predicate devices."

    Acceptance Criteria (Not Explicitly Stated as Measurable Metrics)Reported Device Performance
    Satisfies all expected and previously defined system requirements and features"Test results support the conclusion that actual device performance satisfies the design intent and is equivalent to its predicate devices."
    Substantially equivalent to predicate devices for intended use and technological characteristics"In all material aspects, iNtuition-TDA, TVA, Parametric Mapping is substantially equivalent to the predicate devices."
    No new significant safety and effectiveness concerns"The introduction of iNtuition-TDA, TVA, Parametric Mapping has no significant concerns of safety and efficacy."
    Adheres to internal company procedures for software testing and validation"Performance testing was carried out according to internal company procedures. Software testing and validation were done according to written test protocols established before testing was conducted."
    Adheres to voluntary standards (e.g., DICOM)"voluntary standards such as DICOM, various in-house standard operating procedures are in place and utilized in the production of the software."

    2. Sample size used for the test set and the data provenance:

    • Sample size for the test set: Not specified. The document states "Software testing and validation were done according to written test protocols." It doesn't mention a specific test set size (e.g., number of cases or images).
    • Data provenance: Not specified. Since clinical studies were not required, it's unlikely that the "test set" involved patient data in a formal clinical trial sense. It likely refers to internal software testing using simulated or previously acquired anonymized data that were part of the predicate device's validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The document states that clinical studies were not required. The "ground truth" for the software testing would have been based on the expected outputs as defined by the design requirements, rather than expert interpretation of medical images for diagnostic accuracy. "Test results were reviewed by designated technical professionals." Their qualifications are not specified beyond "technical professionals."

    4. Adjudication method for the test set:

    Not applicable. No mention of an adjudication method, as it wasn't a study involving human interpretation of medical images with a diagnostic endpoint.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "The subject of this traditional 510k notification, iNtuition-TDA. TVA, Parametric Mapping, did not require clinical studies to show safety and effectiveness of the software." Therefore, there is no information on the effect size of human reader improvement with or without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The performance testing described is likely a standalone assessment of the software's functionality and accuracy against its design specifications. The document states, "Test results support the conclusion that actual device performance satisfies the design intent and is equivalent to its predicate devices." However, it doesn't quantify this performance in medical terms (e.g., sensitivity, specificity for a particular pathology), but rather in terms of meeting functional requirements. The device is intended to "assist the physician in diagnosis, who is responsible for making all final patient management decisions," implying it is not a standalone diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Given the lack of clinical studies, the "ground truth" for the software's internal testing would have been defined by the expected computational results based on the software's design and algorithms. For example, if the software calculates ejection fraction, the ground truth would be the mathematically correct ejection fraction from a given input, based on a reference method or calculation. It would not be based on expert medical consensus, pathology, or outcomes data in a clinical validation context.

    8. The sample size for the training set:

    Not applicable. This is a post-processing software module, not a machine learning or AI algorithm that typically requires a large training set for model development. The document does not mention any machine learning components, and thus, no training set or its size is provided. The comparison is based on "similar technological characteristics" to predicate devices.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no mention of a training set for machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121916
    Device Name
    INTUITION
    Manufacturer
    Date Cleared
    2013-04-02

    (274 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TERARECON, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To receive, store, transmit, post-process, display and allow manipulation of reports and medical images from acquisition devices, including optical or other non-DICOM format images. DICOM images with modality type XA, US, CR, DR, SPECT, NM and MG, and images from volumetric medical scanning devices such as EBT, CT, PET or MRI. To provide access to images derived data and derived images via client-server software, web browser and mobile technology.

    Visualization in 2D, 3D and 4D are supported for single or multiple datasets, or combinations thereof. Tools are provided to define and edit paths through structures such as centerlines, which may be used to analyze cross-sections of structures, or to provide flythrough visualizations rendered along such a centerline. Segmentation of regions of interest and quantitative analysis tools are provided, for images of vasculature, pathology and morphology, including distance, angle, volume, histogram, ratios thereof, and tracking of quantities over time. A database is provided to track and compare results using published comparison techniques such as RECIST and WHO. Calcium scoring for quantification of atherosclerotic plaque is supported.

    Support is provided for digital image processing to derive metadata or new images from input image sets, for internal use or for forwarding to other devices using the DICOM protocol. Image processing tools are provided to extract metadata to derive parametric images from combinations of multiple input images, such as temporal phases, or images co-located in space but acquired with different imaging parameters, such as different MR pulse sequences, or different CT image parameters (e.g. dual energy).

    iNtuition is designed for use by healthcare professionals and is intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions.

    Interpretation of mammographic images or digitized film screen images is supported only when the software is used without compression and with an FDA-Approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.

    iNtuitionMOBILE provides wireless and portable access to medical images. This device is not intended to replace full workstations and should be used only when there is no access to a workstation. Not intended for diagnostic use when used via a web browser or mobile device.

    Device Description

    iNtuition is a software device generally used with off-the-shelf hardware, offered in various configurations, with the simplest configuration being a stand-alone workstation capable of image review, communications, archiving, database maintenance, remote review, reporting and basic 3D capabilities described elsewhere in this document. The system can also be configured as a server with some, all, or none of its optional features disabled. Whether provided as a workstation or a server, the iNtuition software is designed to provide access by a local user physically sitting at the computer hosting the iNtuition server software, and/or by one or more remote users who concurrently connect to the server using a freely-downloadable thin client application (with conference capabilities). iNtuition supports the physician in medical image viewing.

    A fully-configured iNtuition system is capable of various image processing and visualization functions, including full-color Volume Rendering, Calcium Scoring, Segmentation Analysis and Tracking (SAT), Vessel Analysis, Flythrough, Multi-phase review, CT/ CTA Subtraction, Lobular Decomposition (LD), iGENTLE, Maxillo-Facial, Volumetric Histogram, Findings Workflow, Fusion CT/ MR/ PET/ SPECT, MultiKV etc. Each of these features may be offered as an independent upgrade option to the basic configuration.

    The intended use of the device is to provide solutions to various medical image analysis and viewing problems, which come about as modalities generate more and more images. It also supports image distribution over networks, and is DICOM compliant.

    AI/ML Overview

    The provided 510(k) summary for the iNtuition device (K121916) explicitly states that no clinical studies were required or performed to prove the safety and effectiveness of the software. This is a critical piece of information. The assessment relies on non-clinical performance tests and a comparison to predicate devices to establish substantial equivalence.

    Therefore, many of the requested categories related to clinical studies and ground truth establishment will be "Not Applicable" or "Not Reported" based on the provided document.

    Here's the breakdown based on the given text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with internal company proceduresPerformance testing carried out according to internal company procedures.
    Compliance with voluntary standards (e.g., DICOM)Voluntary standards such as DICOM are in place and utilized in the production of the software.
    Software testing and validation according to written test protocolsSoftware testing and validation were done according to written test protocols established before testing was conducted.
    Software fully satisfies all expected and previously defined system requirements and featuresTest results were reviewed by designated technical professionals, ensuring the software fully satisfies all expected and previously defined system requirements and features.
    Actual device performance satisfies design intentTest results support the conclusion that actual device performance satisfies the design intent.
    Substantial equivalence to predicate devicesDevice is substantially equivalent to predicate devices (Aquarius Workstation (K011142), AquariusNET Server (K012086), AquariusAPS Server (K061214), VitreaView (K122136), IQQA-Liver Software (K061696)).
    No significant concerns of safety and efficacy"The introduction of iNtuition has no significant concerns of safety and efficacy."
    Performs as well as predicate devices"iNtuition... performs as well as the predicate devices."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not reported. The document states "no clinical studies were required to show safety and effectiveness of the software." Performance testing was non-clinical.
    • Data Provenance: Not reported, as no clinical data was used for direct safety and effectiveness demonstrations. Non-clinical performance tests would likely use synthetic or internal test data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable/Not reported. Ground truth in a clinical sense was not established for non-clinical performance tests. "Designated technical professionals" reviewed test results for software validation, but their qualifications are not specified beyond that title.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/Not reported. This relates to clinical studies for establishing ground truth, which were not performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. The summary explicitly states: "The subject of this traditional 510k notification, iNtuition, did not require clinical studies to show safety and effectiveness of the software." Therefore, no MRMC study comparing human readers with or without AI assistance was performed.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not explicitly detailed as a separate study. The device is a "software device" and "offers convenient tools to support creation of a report," but its performance metrics are established through non-clinical software validation and comparison to predicate devices, not through a standalone performance study with specific metrics like sensitivity/specificity against a gold standard. The device is "intended to assist the physician in diagnosis, who is responsible for making all final patient management decisions," implying a human-in-the-loop context for diagnostic use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: Not applicable/Not reported for demonstrating safety and effectiveness. For non-clinical software performance tests, the "ground truth" would be established by the expected output based on the defined system requirements and internal test protocols.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/Not reported. The device is a general medical imaging system, not an AI/ML device in the modern sense that requires a specific training set to learn from data for a particular diagnostic task. Its functionality is based on established algorithms and image processing techniques.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable/Not reported, as there is no mention of a training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K061214
    Manufacturer
    Date Cleared
    2006-05-15

    (13 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TERARECON, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AquariusAPS server receives medical images from medical imaging acquisition devices adhering to the DICOM protocol for image transfer such as EBT, CT, MRI, and other volumetric or planar medical imaging modalities, and performs digital image processing to derive certain information or new images from these image sets. The information or new images thus derived is transmitted using the DICOM protocol to other devices supporting this standard protocol.

    Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

    Device Description

    The AquariusAPS server receives medical images from medical imaging acquisition devices adhering to the DICOM protocol for image transfer such as EBT, CT, MRI, and other volumetric or planar medical imaging modalities, and performs digital image processing to derive certain information or new images from these image sets. The information or new images thus derived is transmitted using the DICOM protocol to other devices supporting this standard protocol. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA."

    The intended use of the device is to provide time-saving pre-processing of images to remove the need for an image review system to perform these activities while a user is waiting for processing to complete, to optimize the use of the user's time.

    The AquariusAPS Server utilizes standard "off the shelf" personal computer systems as its hardware platform. The software requires the use of the Windows 2000 operating system, and a Pentium III - class processor or equivalent.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the TeraRecon AquariusAPS Server. It is a regulatory document and does not contain information about acceptance criteria or a specific study designed to prove the device meets those criteria.

    The document primarily focuses on:

    • Device identification: Trade name, common name, classification, establishment name, and contact information.
    • Substantial Equivalence: Listing equivalent devices (predicates) and claiming equivalence based on basic design, features, and intended use.
    • Device Description: What the device does (receives and processes medical images, transmits derived information).
    • Intended Use Statement: To provide time-saving pre-processing of images.
    • Hardware & Software Information: Operating system, processor requirements, and compliance with FDA guidance for software.
    • Feature Comparison Table: A table comparing the AquariusAPS Server's features to those of the predicate devices. This table highlights what features the AquariusAPS Server possesses, but it doesn't provide performance metrics or acceptance criteria for those features.
    • FDA Approval Letter: Officially confirming the substantial equivalence determination.

    Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving device performance because that data is not present in the provided text. The document is primarily a statement of equivalence for regulatory purposes, not a performance study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K012086
    Manufacturer
    Date Cleared
    2001-09-13

    (72 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TERARECON, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AquariusNET Server acquires, stores, transmits, and enables compatible computers on a network to display medical images from medical scanning devices and patient reports of various types. Teleradiology, such as MRI, CT or NM and archiving, image manipulation, 3D and 4D visualization are supported. Calcium scoring from whole body computed tomography derived measurements, for non-invasive detection and quantification of atherosclerotic plaque. Tools for histogram analysis of the density distribution of certain regions of interest are provided. A database management and report generation tool is included.

    Device Description

    AquariusNET is a device consisting of a DICOM server that receives and stores images from a PACS or other image giving modalities. It archives images in a scalable storage medium and delivers them in response to DICOM Query/Retrieve requests from other DICOM devices on the network (not part of AquariusNET). It also serves image requests to its remote "thin clients", which act as the graphical user interface to the AquariusNET server. The server can host multiple concurrent sessions from remote "thin clients". AquariusNET features an integrated 2D/3D streaming engine which allows regular PCs or notebooks to control the server, and to review 2D images and 3D reconstructions interactively over a network. AquariusNET is capable of image review, communications, archiving, database maintenance, reporting and basic 3D capabilities described elsewhere in this document. It is also capable of full-color Volume Rendering and Calcium Scoring.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Summary for the AquariusNET Server. It describes the device, its features, and its intended use, focusing on its substantial equivalence to predicate devices. However, the document does not contain the following information typically found in a study proving a device meets acceptance criteria:

    1. A table of acceptance criteria and the reported device performance: The document includes a "Feature Comparison Table" that lists features present in the AquariusNET Server and its predicate devices. This table serves to demonstrate functional equivalence, not performance against specific, quantitative acceptance criteria.
    2. Sample size used for the test set and the data provenance: No information about a test set, its size, or data origin is provided.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No mention of experts or ground truth establishment for a test set.
    4. Adjudication method for the test set: Not applicable as no test set evaluation is described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No MRMC study is mentioned.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The device is an image communication and storage system with advanced visualization; its performance is described in terms of features rather than an algorithmic standalone performance study.
    7. The type of ground truth used: Not applicable as no ground truth is mentioned for performance evaluation.
    8. The sample size for the training set: No training set is mentioned as this is not a machine learning algorithm claim in the modern sense.
    9. How the ground truth for the training set was established: Not applicable.

    Summary of the document's content relevant to acceptance criteria and study:

    The document demonstrates substantial equivalence to predicate devices (Imatron Ultra Access Workstation with Cardiac Software Extensions (K972903) and TeraRecon IiVS™ Integrated Image Viewing Station (K994329)) by comparing features.

    Acceptance Criteria (Implied by Substantial Equivalence):

    The implied acceptance criteria are that the AquariusNET Server possesses equivalent features and performance to the predicate devices for its intended use. The "Feature Comparison Table" acts as the primary evidence for meeting these implied criteria.

    FeatureAquariusNET Server Performance (Reported as Present)Predicate Device 1 (Imatron K972903)Predicate Device 2 (IiVS K994339)
    2D Image ReviewYesYesYes
    Multiplanar reformattingYesYesYes
    3D Volume RenderingYesYesYes
    Maximum Intensity ProjectionYesYesYes
    Image ArchivingYesYesYes
    Image FilmingYesYesYes
    Image Transfer or Network ConnectivityYesYesYes
    Examination of 2D image data from a calcium scanYesYesYes
    Examination of calcium scan as a 3D volumeYesYesYes
    Semi-automated identification of regions considered calciumYesYesNo
    User override of automatically identified regionsYesYesNo
    Automatic calculation of calcium scoreYesYesNo
    Ability to measure CT numbers on a 2D imageYesYesYes
    Saving of calcium data with patient exam dataYesYesNo
    Creation of a paper calcium reportYesYesNo
    Comparison of multiple scansYesYesYes
    Indications for use - general medical imaging workstationsYesYesYes

    Study Proving Acceptance Criteria (as presented in the 510(k) Summary):

    The "study" is a feature comparison study against two predicate devices. The document implies that by demonstrating the presence of these features, the device is "substantially equivalent" to legally marketed devices, thereby meeting the regulatory requirements for market clearance.

    • Sample Size for Test Set & Data Provenance: Not applicable. The submission relies on a feature-by-feature comparison rather than performance testing on a specific dataset.
    • Number of Experts & Qualifications for Ground Truth: Not applicable. No ground truth for a test set is established.
    • Adjudication Method: Not applicable.
    • MRMC Comparative Effectiveness Study: Not performed or reported.
    • Standalone Performance Study: The document focuses on the functional capabilities of the system rather than an isolated algorithmic performance evaluation.
    • Type of Ground Truth: Not applicable.
    • Sample Size for Training Set: Not applicable, as this is not an AI/ML model in the modern sense.
    • Ground Truth for Training Set: Not applicable.

    In essence, the "study" demonstrating acceptance is the comparison table itself, showing that the AquariusNET Server either matches or exceeds the features of its predicate devices, especially regarding calcium scoring functionalities compared to the IiVS, and matches the Imatron Ultra Access Workstation. The FDA's issuance of the 510(k) clearance acts as the formal acceptance that the device is substantially equivalent based on this comparison.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2