Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K233789
    Manufacturer
    Date Cleared
    2024-04-10

    (134 days)

    Product Code
    Regulation Number
    878.5070
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Surnic Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surnic Surclear Smoke Plume Evacuation System with accessories is intended to remove smoke created in surgical procedures at the surgical site.

    Device Description

    The Surclear 8010 have been designed with a high suction, high flow rate vacuum motor. The ultra-quiet motor is used to draw the surgical smoke from the surgical site through the vacuum tubing and into the Surclear 8010 filter 8QF1 where the surgical smoke is processed by a series of filters. A single disposable filter is used to simplify the installation and removal during filter changes. The filter is completely enclosed to protect the healthcare personnel from potential contamination during filter changes. One 8QF1 filter contains four different stages within to capture the smoke plume.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of non-clinical tests for the 8Q10 Surclear Smoke Plume Evacuation System. It does not contain information about a study proving the device meets the acceptance criteria in the context of an AI/human-in-the-loop performance evaluation, which is typically found in submissions for AI-powered devices.

    The information provided pertains to the safety and functional performance of a medical device (a smoke plume evacuation system), not an AI/software as a medical device (SaMD). Therefore, many of the requested criteria (like sample size for test set, data provenance, number of experts, MRMC studies, etc.) are not applicable or not provided in this document.

    However, I can extract and present the information that is available:


    Acceptance Criteria and Reported Device Performance for the 8Q10 Surclear Smoke Plume Evacuation System

    This document describes the non-clinical testing conducted to demonstrate the safety and functionality of the 8Q10 Surclear Smoke Plume Evacuation System.

    1. Table of Acceptance Criteria and Reported Device Performance

    Title of TestPurpose of TestAcceptance CriteriaResults
    Electrical Safety TestingEvaluate Electrical SafetyFulfil the requirement of IEC 60601-1:2012+2020 as applicablePassed
    Electromagnetic CompatibilityEvaluate Electromagnetic CompatibilityFulfil the requirements of IEC 60601-1-2:2014+2020 as applicablePassed
    Software ValidationEvaluate Device SoftwareAll test cases shall pass or deviations explained as to why it is acceptablePassed
    Flow Rate TestingEvaluate flow rate against design requirementThe flow rates at each flow mode are within the defined tolerances.Passed
    Plume Evacuation EvaluateEvaluate System for evacuation of plume generated by medical devicesFulfil the requirements of ISO 16571:2014 as applicablePassed

    Further Technological Characteristics and Comparison to Predicate Device:

    CharacteristicSubject Device (K233789)Predicate Device (K131402)Discussion
    Indications for UseRemove smoke created in surgical procedures at the surgical site.Remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures.Similar
    Static Motor Suction25.69 kPa25.69 kPaSame
    Flow rates (Max)25 cfm (707 lpm) @7/8(22mm) tubing30 cfm (839 lpm) @7/8(22mm) tubingSimilar, (cfm) 25vs30 is due to different manufacturers' understanding of the maximum flow limit required in actual applications.
    Suction control level10 Level10 LevelThere are 10 control levels of suction rate from 10% to 100%.
    Suction blockage warningYesYesAn occlusion warning safety feature alerts the operator when an occlusion has occurred.
    Foot controlYesYesThe operator can foot-control the suction of the device to turn on or off.
    Filter life indicationYesYesCounter indicates filter status and lifetime.
    Filter performance99.99% efficiency at 0.1-to-0.2-micron particle size with a filter life of up to 40 hours.99.99% efficiency at 0.1-to-0.2-micron particle size with a filter life of up to 35 hours.Same filter efficiency, larger filters can increase application time from 35 to 40 hrs.
    Power Supply100-120VAC, 50/60Hz, 9A max; 220-240VAC, 50/60Hz, 4.5A max100/240VAC, 50/60Hz; 220/240 VACSame

    Information Not Found in the Provided Text:

    The following specific requested information is not available in the given document, as it pertains to AI/SaMD performance studies, which is not the subject of this 510(k) submission:

    1. Sample size used for the test set and the data provenance: Not applicable/not provided for a hardware device's functional testing. The tests are against specific engineering and safety standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a smoke evacuator's performance is based on physical measurements and compliance with engineering standards.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used: For this device, ground truth is based on established international (IEC, ISO) and industry standards for electrical safety, electromagnetic compatibility, software validation, flow rates, and plume evacuation.
    7. The sample size for the training set: Not applicable (not an AI/ML device).
    8. How the ground truth for the training set was established: Not applicable (not an AI/ML device).
    Ask a Question

    Ask a specific question about this device

    K Number
    K203522
    Manufacturer
    Date Cleared
    2021-02-22

    (83 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Surnic Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surns Electrosurgical Argon Beam Generator and Accessories are indicated for monopolar or bipolar surgery to achieve argon beam and cut or coagulation for the tissue. It is intended to be used with monopolar handpiece and dispersive electrode or bipolar handpiece and footswitches.

    Device Description

    Surn5 electrosurgical argon beam Generator and accessories have Monopolar Pure cut, Monopolar Bland Cut, Argon Beam, Bipolar functions. Surn5 electrosurgical argon beam Generator and accessories consists of HF generator and accessories. The device generates HF power, which is transmitted to the tissue through the output Pencil. Operator selects the output function and power with the front panel. The indicator is used to indicate the setup status and output status of the device. The output is controlled by foot or hand switch.

    AI/ML Overview

    This document K203522 is a 510(k) premarket notification for an electrosurgical argon beam generator. This specific device is regulated under 21 CFR 878.4400, "Electrosurgical Cutting and Coagulation Device and Accessories," which falls under Class II.

    The document does not describe a study involving human subjects, AI/algorithm performance, or the establishment of ground truth by expert consensus for medical imaging. Instead, it discusses the safety and effectiveness testing for an electrosurgical device, primarily focusing on non-clinical performance data to demonstrate substantial equivalence to a predicate device.

    Therefore, many of the requested criteria regarding AI/ML performance, sample sizes for data sets, expert adjudication, MRMC studies, and ground truth establishment are not applicable to the content provided in this 510(k) summary.

    However, I can extract the information that is present regarding the non-clinical acceptance criteria and the tests performed.

    1. Table of acceptance criteria and the reported device performance:

    The document outlines the non-clinical performance tests performed. While it doesn't provide specific quantitative acceptance criteria or detailed numerical results, it lists the standards to which the device was tested and states that the tests were "successfully completed."

    Test CategoryAcceptance Standard (Implicitly met if "successfully completed")Reported Device Performance
    General Safety TestingIEC 60601-1:2005+C1+C2+A1:2012 Medical electrical equipment, Part 1 General requirements for basic safety and essential performance"Medrange completed successfully the following non-clinical performance tests" - implying compliance with the standard for basic safety and essential performance.
    EMC Safety TestingIEC 60601-1-2:2014 (Edition 4.0) Medical electrical equipment, Part 1-2 General requirements for basic safety and essential performance Collateral Standard: Electromagnetic disturbances - Requirements and tests"Medrange completed successfully the following non-clinical performance tests" - implying compliance with the standard for electromagnetic compatibility.
    HF Safety TestingIEC 60601-2-2:2017 Medical electrical equipment, Part 2-2 Particular requirements for the basic safety and essential performance of high frequency - Surgical equipment and high frequency surgical accessories"Medrange completed successfully the following non-clinical performance tests" - implying compliance with the standard for high-frequency surgical equipment safety and performance.
    Performance TestingApplicable parts of FDA Guidance Document "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery”, May 2016"Medrange completed successfully the following non-clinical performance tests" - implying the device met the performance requirements outlined in the FDA guidance for electrosurgical devices.
    Usability TestingIEC 60601-1-6:2010, AMD1:2013 Medical electrical equipment Part 1-6 General requirements for safety - Collateral Standard: Usability"Medrange completed successfully the following non-clinical performance tests" - implying the device met the usability requirements, ensuring that the device can be used safely and effectively as intended. The "software data" mentioned in the statement of substantial equivalence would likely fall under this, or be part of performance testing.

    Summary Judgement on Acceptance: The document states that the tests were "successfully completed" and concludes that the device is "as safe and effective as the predicate device," thus fulfilling "the requirements of a substantially equivalent device."

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not applicable in the context of this 510(k). The tests are typically conducted on a single prototype or representative units of the device in a laboratory setting, not on a "test set" of patient data or images.
    • Data provenance: Not applicable. The tests are in vitro or ex vivo engineering and electrical performance tests, not based on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is an electrosurgical generator, not an AI/ML algorithm requiring expert annotation of medical images for ground truth. The "ground truth" here refers to established engineering standards and device specifications.

    4. Adjudication method for the test set:

    • Not applicable. Testing is against predefined electrical, mechanical, and safety standards, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results, often with and without AI assistance. This document concerns a surgical energy device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML algorithm. The performance of the device itself (e.g., power output, frequency, safety features) is tested in a standalone manner against engineering standards.

    7. The type of ground truth used:

    • The "ground truth" for this device's performance is based on International Electrotechnical Commission (IEC) standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6) and applicable sections of FDA Guidance Documents ("Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery”).
    • Additionally, comparison to a legally marketed predicate device's technological characteristics (e.g., power outputs, waveforms, system voltage) serves as a comparative "ground truth" to establish substantial equivalence.

    8. The sample size for the training set:

    • Not applicable. There is no AI/ML component described, so no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable, for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1