Search Results
Found 156 results
510(k) Data Aggregation
(68 days)
Irvine Biomedical, a St. Jude Medical Company
The ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.
The ViewFlex™ Xtra ICE Catheter ("ViewFlex™ Catheter") is a temporary intracardiac ultrasound catheter intended for use in patients to accurately visualize cardiac structures, blood flow and other devices within the heart when connected to a compatible intracardiac ultrasound console via the compatible ViewFlex™ Catheter Interface Module. Examples of the types of other devices that can be visualized include, and are not limited to, intracardiac catheters, septal occluders, delivery wires, delivery sheaths, sizing balloons and transseptal needles. The use of these images is limited to visualization with no direct or indirect diagnostic use. The ViewFlex™ Catheter has a useable length of 90 cm, with a 9 French (F) shaft with an ultrasound transducer. A 10F introducer is recommended for use with this catheter for insertion into the femoral or jugular veins. The catheter tip has four-directional deflection allowing for Left-Right and Posterior-Anterior deflection, with an angle of at least 120 degrees in each direction. The ViewFlex™ Catheter is compatible with the ViewMate™ Z, ViewMate™, ViewMate™ Multi, and Philips CX50 ultrasound consoles.
Here's a breakdown of the acceptance criteria and the study information for the ViewFlex™ Xtra ICE Catheter, based on the provided document:
This document is a 510(k) clearance letter and a summary of the special 510(k) submission for the ViewFlex™ Xtra ICE Catheter's compatibility with the ViewMate™ Multi Ultrasound System. It describes the testing performed to ensure the updated device (the ViewFlex™ Xtra ICE Catheter when used with the new ViewMate™ Multi Ultrasound System) does not raise new questions of safety or effectiveness compared to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that "Design verification activities were performed with their respective acceptance criteria to ensure that the use of the ViewFlex™ Xtra ICE Catheter in conjunction with the ViewMate™ Multi Ultrasound System does not affect the safety or effectiveness of the device. All testing performed met the established performance specifications."
While specific, granular acceptance criteria values are not explicitly listed in this summary, the general acceptance criteria are implicitly that the device, when used with the new console, performs safely and effectively within established performance specifications, and meets relevant industry standards.
Acceptance Criteria (General) | Reported Device Performance |
---|---|
Ensure safety and effectiveness when used with ViewMate™ Multi Ultrasound System | "All testing performed met the established performance specifications." |
Compliance with BS EN 60601-2-37 (basic safety & essential performance of ultrasonic medical diagnostic and monitoring equipment) | No adverse findings reported; compliance affirmed implicitly by FDA clearance, and statement "All testing performed met the established performance specifications." |
Compliance with BS EN 60601-1 (general requirements for basic safety and essential performance of medical electrical equipment) | No adverse findings reported; compliance affirmed implicitly by FDA clearance, and statement "All testing performed met the established performance specifications." |
Compliance with BS EN 60601-1-2 (electromagnetic compatibility) | No adverse findings reported; compliance affirmed implicitly by FDA clearance, and statement "All testing performed met the established performance specifications." |
No new or increased hazards identified as a result of modifications | "There were no new or modified hazards identified as a result of the proposed modifications. Abbott has reviewed the applicable risk management documentation... and determined that the changes... do not increase existing risks or create new risks to the user or patient, and that the residual risk remains acceptable following verification and validation activities." |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify the sample size for the test set in terms of number of patients, cases, or images. The testing described is primarily non-clinical design verification and validation testing rather than clinical performance testing with patient data.
The data provenance is for non-clinical testing (bench testing, performance evaluations) related to device compatibility and adherence to standards. It does not involve human patient data from a specific country, nor is it described as retrospective or prospective clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not applicable to this submission. The ground truth for this type of special 510(k) submission for device compatibility is established through objective engineering and performance testing, comparing the device's output and safety metrics against predefined specifications and industry standards, rather than expert interpretation of clinical data to establish a "ground truth" diagnosis.
4. Adjudication Method for the Test Set:
This information is not applicable. Since the testing focuses on technical performance and compliance with standards, there is no need for expert adjudication in the traditional sense of resolving discrepancies in diagnostic interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:
No, an MRMC comparative effectiveness study was not done. This type of study is typically performed to evaluate the diagnostic performance of an AI-powered device or a new imaging modality and compare it to existing methods or human readers. The ViewFlex™ Xtra ICE Catheter is a diagnostic intravascular catheter, and this particular submission focuses on its compatibility with a new ultrasound console, not on comparing its diagnostic performance or an AI component to human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Not applicable. The ViewFlex™ Xtra ICE Catheter is a medical device (a catheter with an ultrasound transducer) that requires a human operator and an ultrasound console to function. It is not an AI algorithm or a standalone diagnostic tool in that sense. The device's performance is inherently tied to its use by a human and with supporting equipment.
7. The Type of Ground Truth Used:
The ground truth used for this type of non-clinical testing is based on established engineering specifications, industry standards, and validated test methods. For example, measuring a deflection angle would use a calibrated measurement system (engineering ground truth), or testing electrical safety complies with specific limits defined in the referenced standards (standard-based ground truth).
8. The Sample Size for the Training Set:
Not applicable. This submission is for a medical device (a catheter) and its compatibility with an ultrasound system. It does not involve machine learning or AI algorithms that would require a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for an AI algorithm in this submission.
Ask a specific question about this device
(29 days)
Abbott (St. Jude Medical)
The Jot Dx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
The Jot Dx™ ICM DM4500 System consists of the Jot Dx™ ICM Model DM4500 Implantable Device, Implant Tools (Incision tool Model DM3520 and Insertion tool Model DM3510), Confirm Rx ICM Accessory Kit DM3500A, Magnet, Clinician Programmer (Merlin PCS Programmer Model 3650), myMerlin™ Mobile Application (Model APP1000 (Android) and APP1001 (iOS)), and Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator). The ICM is a minimally invasive, implantable diagnostic monitoring device with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability.
This document is a 510(k) premarket notification for the Abbott Jot Dx™ Insertable Cardiac Monitor (ICM) DM4500. The core of this submission is to demonstrate substantial equivalence to an existing predicate device, the Abbott Confirm Rx™ ICM DM3500 (K202888).
Therefore, the study supporting acceptance criteria isn't a traditional clinical study with human readers and ground truth established by experts. Instead, the "study" is a non-clinical test summary demonstrating that the new device is identical to the predicate device in all relevant aspects affecting safety and effectiveness.
Here's a breakdown based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Since the submission claims identity to a predicate, the acceptance criteria are effectively that the Jot Dx™ ICM DM4500 must perform identically to the Confirm Rx™ ICM DM3500.
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Functional Identity: The Jot Dx™ ICM DM4500 must possess the same intended use, indications for use, operating rules, device/tool functionality, software, packaging materials and process, shelf life, and device longevity as the predicate Confirm Rx™ ICM DM3500. | "The candidate Jot Dx™ ICM DM4500 is identical to the predicate Confirm Rx™ ICM DM3500 (K202888), aside from name, model number, and corresponding laser marking." |
"The fundamental technological characteristics of the candidate Jot Dx™ ICM DM4500 are not changing in comparison to the predicate device, the current Confirm Rx™ ICM DM3500 (510(k) K202888). The candidate Jot Dx™ ICM DM4500 system has the same:
- Intended Use and Indications for Use
- Operating rules
- Device/tool functionality
- Packaging materials and process
- Shelf life
- Device Longevity
- Software"
"The candidate Jot Dx™ ICM DM4500 is identical in design and function and has the same indications and intended use as the predicate Confirm Rx™ ICM DM3500 (K202888)." |
| Safety and Effectiveness: The minor differences (name, model number, laser marking) do not raise new issues of safety and effectiveness. | "The minor differences in name, model number, and laser marking do not raise new issues of safety and effectiveness."
"Completion of all verification and validation activities demonstrated that the candidate Jot Dx™ ICM DM4500 meets its predetermined design and performance specifications and that the product is substantially equivalent to the predicate device (Model DM3500, K202888)."
"The results of the testing show that the candidate Jot Dx ICM DM4500 system performs as intended and is safe for its intended use."
"The candidate Jot Dx™ ICM system is substantially equivalent in terms of safety and technological characteristics to the identified predicate device (Confirm RxTM ICM System; K202888). Product verification and validation testing demonstrate that the candidate Jot Dx ICM system is as safe and as effective and performs as well as the predicate system (K202888)." |
| Risk Assessment: Acceptable risk profile. | "The risk analysis method used to assess the impact of the addition of the candidate Jot Dx™ ICM DM4500 documents the investigation of hazards and mitigation of associated risks and reports the result of the investigation. The risk analysis method used to assess the impact of the modifications was a Failure Mode and Effects Analysis (FMEA/FMECA). It was determined that the overall risk is acceptable." |
2. Sample Size for the Test Set and Data Provenance:
- Test Set Sample Size: Not applicable in a traditional sense. The "test set" here refers to the engineering verification and validation testing performed on the single design of the Jot Dx™ ICM DM4500 to confirm its identity and performance against the predicate. This isn't a statistical sample of independent clinical cases.
- Data Provenance: Not applicable for a traditional clinical study. The "data" comes from internal engineering verification and validation activities. It's retrospective in the sense that they are comparing it to an already cleared predicate device's specifications and performance. No country of origin is specified for these internal engineering tests, but the company is based in the USA.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: Not applicable. Ground truth, in the context of substantial equivalence for an identical device, is established by its adherence to the design and performance specifications of the predicate device, not through expert human interpretation of clinical data on the new device.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not applicable. There is no ambiguous clinical data requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This submission does not involve an AI component or a comparative effectiveness study with human readers. The device is an insertable cardiac monitor that automatically detects events.
6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:
- Stand-alone Performance: While the device automatically detects cardiac arrhythmias (an "algorithm" in a broad sense), the submission doesn't present a standalone performance study with metrics like sensitivity and specificity against a clinical ground truth. Instead, it relies on the established performance of the identical predicate device. The document explicitly states the "Automated triggering of EGM storage when tachycardia, bradycardia, or pauses are detected" and "Automated triggering of EGM storage when atrial fibrillation (AF) is detected." The performance of these detection algorithms is considered equivalent to the predicate.
7. The Type of Ground Truth Used:
- Type of Ground Truth: The "ground truth" for this submission is the established design, performance specifications, and safety/effectiveness profile of the legally marketed predicate device (Confirm Rx™ ICM DM3500, K202888). The new device is asserted to be identical in these respects, making the predicate the reference "truth."
8. The Sample Size for the Training Set:
- Training Set Sample Size: Not applicable. There is no machine learning or AI algorithm being trained for this device submission as it is presented as a re-branding/model number change of an existing technology.
9. How the Ground Truth for the Training Set was Established:
- Ground Truth for Training Set: Not applicable, as there is no training set mentioned in this 510(k) application.
Ask a specific question about this device
(64 days)
Abbott (St. Jude Medical)
The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as the following: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
The Confirm Rx™ ICM System consists of the following key features and components: Confirm Rx™ ICM Model DM3500 Implantable Device, Implant Tools (Incision tool model DM3520 and Insertion tool model DM3510), Confirm Rx ICM Accessory Kit DM3500A, Magnet, Clinician Programmer (Merlin PCS Programmer Model 3650), myMerlin™ Mobile Application (Model APP1000 (Android) and APP1001 (iOS)), and Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator). The ICM is a minimally invasive, implantable diagnostic monitoring device with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability.
The provided text does not contain information about acceptance criteria for device performance or a study proving that the device meets specific performance criteria. Instead, it details a 510(k) premarket notification for an Abbott (St. Jude Medical) Confirm Rx™ Insertable Cardiac Monitor system (K202876).
This submission is for an Accessory Kit DM3500A and an updated incision tool DM3520 blade design for an already cleared device (predicate K193310). The document asserts substantial equivalence to the predicate device, meaning it argues that the new components do not raise new questions of safety or effectiveness and perform as well as the predicate.
The document states:
- "Completion of all verification and validation activities demonstrated that the Confirm Rx ICM accessory kit meets its predetermined design and performance specifications and that the product is substantially equivalent to the current device (Model DM3500, K193310)." (Page 9)
- "The results of the verification and validation tests and the risk analysis have demonstrated the candidate Confirm Rx ICM system with Accessory Kit DM3500A and updated incision tool DM3520 blade design functions in accordance with product specifications." (Page 9)
- "Product verification and validation testing demonstrate that the candidate Confirm Rx ICM system is as safe and as effective and performs as well as the predicate system (K193310)." (Page 9)
However, it does not provide a table of acceptance criteria and reported device performance for the new components or the system as a whole. It also does not describe a specific study with a test set, training set, expert ground truth, or adjudication methods in the context of device performance metrics (e.g., sensitivity, specificity for arrhythmia detection). The non-clinical test summary mentions a "risk analysis method... a Failure Mode and Effects Analysis (FMEA/FMECA)," but this is for assessing hazards and mitigating risks, not for evaluating performance against clinical acceptance criteria.
Therefore, I cannot fulfill your request for the detailed information as it is not present in the provided FDA document.
Ask a specific question about this device
(56 days)
Abbott (St. Jude Medical)
The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
The Confirm RxTM ICM System consists of the following key features and components:
- Confirm RxTM ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias.
- Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool are used to implant the device subcutaneously.
- Magnet: The magnet facilitates faster startup of Bluetooth connection and provides user authentication (required for programmer sessions).
- Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information.
- myMerlin™ Mobile Application (Model APP1000 (Android) and APP1001 (iOS)): The mobile application provides the means for the patient to activate EGM recording in the Confirm Rx™ device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network.
- Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net system allows physicians to remotely monitor and diagnose patients' cardiac events.
This document is a 510(k) summary for the Abbott Confirm Rx Insertable Cardiac Monitor (ICM) with updated MR Conditional labeling. It focuses on demonstrating substantial equivalence to a previously cleared version of the device, specifically for MR Conditional labeling for 3T MRI environments. The acceptance criteria and study described primarily relate to this MR Conditional aspect, not the core cardiac monitoring functionality.
Here's the information extracted from the provided text, addressing your specific points:
1. Table of Acceptance Criteria and Reported Device Performance
The document states that the testing performed supports additional 3T MR Conditional labeling. It explicitly mentions that the "3T testing meets the same clinical acceptance criteria that was previously evaluated for predicate 1.5T MR Conditional labeling demonstration of safety."
While the specific clinical acceptance criteria for MR Conditional labeling (e.g., maximum temperature rise, induced voltage, artifact size, device malfunction rates) are not detailed within this summary, the outcome is reported as:
Acceptance Criteria Category | Reported Device Performance |
---|---|
MR Conditional for 3T MRI environments | The device demonstrates safety for MRI scans in a 3T scanning environment. |
Functional Integrity Post-MRI | The device functions in accordance with product specifications after 3T MRI exposure. |
Safety for Intended Use Post-MRI | The device is safe for its intended use after 3T MRI exposure. |
The document concludes that: "The results of the MRI verification testing have demonstrated the candidate Confirm Rx™ ICM DM3500 with updated MR Conditional labeling functions in accordance with product specifications." and "The results of the testing show that the candidate Confirm Rx™ ICM performs as intended and is safe for its intended use."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size for the test set used for the MR Conditional testing. It generically refers to "all necessary device and system verification testing."
The provenance of the data (e.g., country of origin, retrospective/prospective) is not mentioned. Given the nature of MR Conditional testing, it would typically be conducted in a controlled lab environment.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the document. For MR Conditional testing, ground truth typically relates to engineering measurements against pre-defined safety limits (e.g., temperature probes, E-field/H-field sensors, functional checks), rather than expert medical interpretation of diagnostic outputs.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method. For MR Conditional testing, results are typically objective measurements against engineering standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC study was not done as this submission focuses on updated MR Conditional labeling for a device, not the interpretation of its diagnostic output by human readers. Therefore, there is no effect size of human readers improving with AI vs. without AI assistance.
6. Standalone Performance Study
The document implies a standalone (algorithm only) performance study related to MR Conditional labeling, as it focuses on the device itself and its interaction with MRI fields. The "results of the MRI verification testing" demonstrate the device's functional integrity and safety. This testing is conducted on the device without human interpretation of its outputs during the test itself, only checking if the device continues to function correctly after the MRI exposure.
7. Type of Ground Truth Used
The ground truth used for the MR Conditional testing is based on engineering standards and design specifications for safe performance in an MRI environment. This includes, but is not limited to, parameters like temperature rise, induced voltage, and device functionality post-exposure, which are objectively measurable and compared against established safety thresholds.
8. Sample Size for the Training Set
The document describes pre-market notification for an updated label for an existing device. It does not mention any "training set" in the context of an AI/ML algorithm. This device is an implantable cardiac monitor, and the current submission is for an updated MR Conditional label, not for a new or updated AI algorithm.
9. How the Ground Truth for the Training Set Was Established
As there is no mention of a training set related to an AI/ML algorithm in this document, this information is not applicable.
Ask a specific question about this device
(76 days)
Abbott (St. Jude Medical)
The Agilis HisPro Steerable Catheter With Electrodes is indicated to provide a pathway for delivery and support of transvenous devices within the chambers and vasculature of the heart and can be used for electrogram recording and stimulation.
The Agilis HisPro Steerable Catheter With Electrodes is a deflectable, slittable catheter with two distal tip electrodes. It serves as a delivery conduit for devices such as cardiac leads. The distal portion can be formed into a "U" shape, when fully deflected, to facilitate positioning of the tip at a desired location in the heart. The 10.5 Fr catheter has a braided shaft with a multi-durometer outer Pebax jacket. The catheter has a 7 Fr diameter lumen that allows for delivery of a 6Fr cardiovascular transvenous device, including a 6Fr pacing lead. The Agilis HisPro catheter's proximal handle consists of a rotating actuator with passive locking mechanism for deflecting the distal portion, a hemostasis valve for leak-proof delivery, and an integrated cable incorporating a 4-pin electrical connection for connecting to external electrical systems via a commercially available electrophysiology cable. The tip electrodes provide the ability to sense intracardiac electrogram (EGM) and pace when connected to the Merlin™ Pacing System Analyzer (PSA) or the WorkMate Claris™ Recording System. The sensing and pacing feature of the Agilis HisPro catheter provides the ability to identify the desired location in the heart. Accessories packaged for use with the Agilis HisPro catheter include dilator, guidewire for vascular access, cable adaptor pins for connection to the Merlin PSA via the electrophysiology cable, valve bypass tools for lead insertion, three-way stopcock and slitter. The catheter and the packaged accessories are provided as a sterile single-use device.
The provided FDA 510(k) summary for the Agilis HisPro Steerable Catheter With Electrodes does not contain information about a study proving the device meets specific acceptance criteria in the context of AI/machine learning or diagnostic performance metrics.
The document describes the device's technical specifications, indications for use, comparison to predicate devices, and a list of performance bench testing, packaging and shelf-life testing, biocompatibility assessment, and some animal testing. However, it does not mention:
- Acceptance criteria directly related to diagnostic performance (e.g., sensitivity, specificity, accuracy).
- Any study with a "test set" and "training set" of data, ground truth establishment, or expert adjudication that would be typical for validating an AI/ML medical device.
- Multi-reader multi-case (MRMC) comparative effectiveness studies.
- Standalone algorithm performance.
The listed tests are primarily focused on the physical, electrical, functional, and biological safety and performance of a medical catheter, not on a diagnostic or prognostic algorithm's performance.
Therefore, I cannot populate the requested table or answer the specific questions related to AI/ML device validation based on the provided text.
The information that is available relates to the overall performance testing for regulatory clearance of a physical medical device. Below is a summary of the performance testing performed, but it doesn't align with the requested format for AI/ML device performance.
Summary of Performance Testing (as described in the document):
The device underwent various performance tests to demonstrate it meets design specifications and is as safe and effective as predicate devices.
1. A table of acceptance criteria and the reported device performance:
- The document lists categories of tests but does not provide specific numerical acceptance criteria (e.g., "deflection force must be X N ± Y N") or quantified "reported device performance" values for each criterion. It only states that testing "demonstrated that the subject device is as safe and effective as the predicate devices."
Summary of Performance Tests Performed:
Test Category | Description / Outcome |
---|---|
Design Verification | |
Physical and Dimensional Characteristics | Performed. Outcome: Demonstrated device meets design specification. |
Electrical Characteristics | Performed. Outcome: Demonstrated device meets design specification. |
Functional Characteristics | Performed. Outcome: Demonstrated device meets design specification. |
Connector/Lead/Slitter Compatibility | Performed. Outcome: Demonstrated device meets design specification. |
Design Validation | |
Simulated-use Bench Testing | Performed. Outcome: Demonstrated device meets design specification (implies successful performance in simulated use). |
Packaging and Shelf Life | |
Nominal T=0 | Performed. |
Accelerated Aging 12-month | Performed. Outcome: Assessed stability and integrity over simulated shelf life. |
Biocompatibility | |
Hemocompatibility | Performed. |
Cytotoxicity | Performed. |
Systemic Toxicity | Performed. |
Sensitization/Irritation | Performed. |
Particulate Matter | Performed. |
Electromagnetic Compatibility and Electrical Safety | Performed. |
Animal Testing | |
Safety evaluation in comparison to Medtronic SelectSite C304 | Performed. Outcome: Presumed comparable safety profile to predicate. |
Sensing/Pacing performance via compatibility with PSA and WorkMate Claris | Performed. Outcome: Demonstrated compatibility and functionality for electrogram recording and pacing. |
Since the document does not relate to an AI/ML device, the following points cannot be addressed:
- 2. Sample sized used for the test set and the data provenance
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- 4. Adjudication method for the test set
- 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- 7. The type of ground truth used
- 8. The sample size for the training set
- 9. How the ground truth for the training set was established
Ask a specific question about this device
(109 days)
Abbott (St. Jude Medical)
The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
The Abbott Confirm RxTM (ICM) System consists of the following key features and components:
- Confirm Rx™ ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias.
- Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool to implant the device subcutaneously.
- Model 3111 Magnet: SJM donut magnet, facilitates faster startup of Bluetooth connection and provides user authentication (required for programmer sessions).
- Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information.
- myMerlin™ Patient App (Model APP1000 (Android) and APP1001 (iOS)): The mobile application provides the means for the patient to activate EGM recording in the Confirm RxTM device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network.
- Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net MN5000 system allows physicians to remotely monitor and diagnose patients' cardiac events.
The provided text describes a 510(k) premarket notification for the "myMerlin Mobile Application (Android)" and "myMerlin Mobile Application (iOS)" by Abbott (St. Jude Medical). This application is part of the Confirm Rx™ Insertable Cardiac Monitor System. The submission claims substantial equivalence to a previously cleared predicate device (K192593, Confirm Rx™ Insertable Cardiac Monitor System).
The submission focuses on updates to the myMerlin™ mobile application, primarily related to scheduled background workflows and telemetry connectivity improvements, as well as the ability to transfer a patient's remote monitoring support profile on Merlin.net.
The document states that the fundamental technological characteristics, intended use, indications for use, operating rules, cybersecurity, and usability testing remain the same as the predicate device. Therefore, the testing performed was to demonstrate that the updated mobile application maintains its safety and effectiveness for its intended use, rather than proving performance against new acceptance criteria for a novel functionality.
Given this context, the acceptance criteria and study details are presented as follows:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Source of Criteria / Performance Standard | Reported Device Performance |
---|---|---|
Software Functionality & Performance | Predetermined design and performance specifications (consistent with predicate) | "Completion of all verification and validation activities demonstrated that the updated mobile application meets its predetermined design and performance specifications." |
Telemetry Connectivity | Predetermined design and performance specifications for device communication | Update to improve myMerlin™ mobile application scheduled background workflows and update of the telemetry connectivity of the Confirm Rx™ ICM System, including the myMerlin™ mobile application and Confirm Rx™ ICM device. |
Data Transfer (Remote Monitoring Profile) | Predetermined design and performance specifications | "The candidate myMerlin™ mobile application… is being updated to allow the patient's remote monitoring support profile on Merlin.net to be transferred from the implant clinic to the remote monitoring clinic." |
Safety and Effectiveness | Equivalent to predicate device (K192593) | "The results of the testing show that the candidate Confirm Rx™ ICM System with updated myMerlin™ mobile application performs as intended and is safe for its intended use." |
Risk Mitigation | Risk analysis | "The results of the verification and validation tests and the risk analysis have demonstrated the myMerlin™ mobile application within the Confirm Rx™ ICM System functions in accordance with the product specifications." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size of a "test set" in terms of patient data or scenarios used. The testing described is primarily focused on software verification, system verification, design validation, and Bluetooth Low Energy proxy software verification. This implies testing involved the software application itself and its interaction with the ICM device, rather than a clinical study with a patient cohort.
- Data Provenance: Not applicable in the context of a clinical test set. The provenance of the software updates is from the manufacturer, Abbott. The nature of the changes (software updates) suggests internal testing and technical verification.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This information is not provided. Given the nature of the submission (updates to a mobile application interacting with an already cleared device), the "ground truth" would likely be established by engineering and software performance standards, rather than expert clinical consensus on patient data.
4. Adjudication Method for the Test Set
This information is not provided. As noted in point 3, the testing appears to be technical in nature rather than clinical, so a traditional clinical adjudication method would not be expected.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was mentioned. The myMerlin™ mobile application is primarily a patient interface for activating EGM recording and data pass-through. It is not an AI-assisted diagnostic tool for "human readers," so this type of study would not be relevant.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
The submission describes the myMerlin™ mobile application (APP1000/APP1001) as a component of the Confirm Rx™ ICM System. The application enables patients to activate EGM recording and facilitates data pass-through to the Merlin.net Patient Care Network for physician review. The focus is on the functional performance of the application and its connectivity, not on a standalone algorithm for diagnosis without human input. Therefore, a standalone algorithm performance study as typically understood for diagnostic AI was not done or described. The "standalone" performance here refers to the application's ability to fulfill its intended technical functions.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the specific updates to the myMerlin™ mobile application, the "ground truth" for verification and validation would be defined by predetermined design and performance specifications or product specifications. This means the software was tested against its expected behavior and technical requirements, rather than against clinical outcomes, pathology, or expert consensus on a clinical dataset. The objective was to confirm that the updated software performed as intended and did not negatively impact the already cleared system's functionality.
8. The Sample Size for the Training Set
This information is not applicable. The myMerlin™ mobile application is not described as involving a machine learning or AI algorithm that would require a "training set" in the conventional sense (i.e., for learning to identify patterns in data). The changes are described as software updates to workflows and connectivity.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable, as no training set for a machine learning algorithm is mentioned or implied.
Ask a specific question about this device
(58 days)
Abbott (St. Jude Medical)
The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
The Confirm RxTM ICM has not been specifically tested for pediatric use.
The Confirm Rx™ Insertable Cardiac Monitor (ICM) Model DM3500 is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias. It has a two-year longevity, MR conditional labeling, sensing and detection algorithms, and Bluetooth communication. Specific features include:
- Patient-initiated triggering of EGM storage using the myMerlin™™ Patient App for mobile devices. This includes capability for the patient to identify symptoms, which are stored with the EGM for physician review.
- Automated triggering of EGM storage when tachycardia, bradycardia, or pauses are detected; with physician-programmable values for pause duration, bradycardia rate, tachycardia rate, and number of tachycardia intervals.
- Automated triggering of EGM storage when atrial fibrillation (AF) is detected, with physician programmable values for AF duration.
- The ability to identify EGM anomalies as a consequence of noise or vigorous activity and inhibit EGM storage as applicable.
- Remote care monitoring.
The provided document is a 510(k) premarket notification for the Abbott Confirm Rx™ Insertable Cardiac Monitor (ICM) System, Model DM3500, with updated firmware. The purpose of this submission is to demonstrate substantial equivalence to a predicate device (K182981).
The document details the device description, technological characteristics, and the testing conducted to support the substantial equivalence claim. However, it does NOT contain a detailed study proving the device meets specific acceptance criteria for AI/algorithm performance as would be typically found for new AI-driven diagnostic devices. Instead, the focus is on demonstrating that the updated firmware, particularly its detection algorithms, functions as intended without raising new issues of safety and effectiveness compared to the predicate device.
Specifically, the acceptance criteria and performance data are described in terms of verification and validation activities showing the device meets its "predetermined design and performance specifications" and "functions in accordance with product specifications." The changes in the firmware relate to "detection algorithms for the diagnosis of bradycardia, asystole (pause), and atrial fibrillation (AF)," including "second pass undersensing discriminators" and a "P-wave detection discriminator."
Given the information provided, it's not possible to present a table of acceptance criteria and reported device performance in the context of an AI study as you described.
However, I can extract the relevant information regarding the firmware update validation and
address the closest approximations to your questions for this type of medical device submission.
Analysis of the Provided Document Regarding Acceptance Criteria and Study:
The document describes an update to the firmware of an existing, already cleared device. The "study" here is the validation of the firmware update to ensure it doesn't negatively impact safety or effectiveness and that the new algorithms perform as intended.
1. A table of acceptance criteria and the reported device performance
The document states that the "Completion of all verification and validation activities demonstrated that the device with updated firmware meets its predetermined design and performance specifications and that the product is substantially equivalent to the predicate Confirm Rx™ ICM device (K182981)." It also states, "The results of the testing show that the candidate Confirm RxTM ICM performs as intended and is safe for its intended use."
The updated firmware includes changes to detection algorithms for bradycardia, asystole (pause), and atrial fibrillation (AF). The acceptance criteria would inherently be related to the accuracy and reliability of these detections compared to established benchmarks or the predicate device, although specific numerical performance metrics (e.g., sensitivity, specificity, accuracy percentages) are not detailed in this summary.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Device meets predetermined design and performance specifications for detection algorithms (bradycardia, asystole, AF) including new discriminators. | Device with updated firmware meets its predetermined design and performance specifications. |
Device performs as intended for its specified use (monitoring and diagnostic evaluation of arrhythmias). | Device performs as intended and is safe for its intended use. |
Updated firmware does not raise new issues of safety and effectiveness compared to the predicate device. | The minor differences (firmware update) do not raise new issues of safety and effectiveness. |
Maintenance of device longevity with updated firmware. | Accounted for the current drain and time used to execute the undersensing and P-wave discriminators within the battery longevity calculation. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the sample size (number of patients, number of EGM recordings) used for the verification and validation (V&V) testing. It also does not specify the data provenance (country of origin, retrospective/prospective). This type of detail is typically found in the full testing report, not usually summarized in this section of a 510(k) summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
The document does not specify the number or qualifications of experts used to establish ground truth. For embedded device algorithm testing, ground truth might often be established through simulated signals, pre-recorded clinical data with confirmed diagnoses, or expert review of EGMs, but this detail is not provided.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
The document does not specify any adjudication method for ground truth establishment.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not conducted or described. This is a device firmware update, not a new AI-assisted diagnostic tool where human reader performance would be compared. The focus is on the device's internal algorithm performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the testing described appears to be a standalone validation of the algorithm's performance within the device, given the context of "Software/Firmware Verification and System Verification" and "Design Validation." The purpose is to confirm the algorithm's internal logic and detection capabilities. The document states:
- "Second pass undersensing discriminators added to the Asystole and Bradycardia detection algorithms to reject false Pause and Bradycardia detections."
- "Second pass P-wave detection discriminator to reject false AF detections."
This implies that the algorithm itself was tested for its ability to correctly identify and reject false detections, which is a standalone performance metric.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The type of ground truth is not explicitly stated. For cardiac rhythm detection algorithms, ground truth typically involves a highly accurate reference ECG/EGM interpretation, which could be established via:
- Expert Consensus: Review by multiple cardiologists or electrophysiologists.
- Manual Annotation: Beat-by-beat or episode-by-episode annotation of EGM data by trained personnel, often adhering to specific criteria.
- Validation against high-fidelity recordings: Comparing device output to a gold-standard recording from a different, validated system.
Given the focus on "reject[ing] false detections," it implies that there was a reference against which the algorithm's detection (true positive/false positive) was measured.
8. The sample size for the training set
The document does not provide any information about a training set size. This indicates that this submission is about validating a firmware update for an already cleared device, not seeking de novo clearance for a new machine learning algorithm that typically requires a distinct training phase. While the algorithms were likely "trained" or designed using data at some point in their development, that information is not part of this 510(k) summary for a firmware update.
9. How the ground truth for the training set was established
Since there is no mention of a training set in this document, there is no information on how its ground truth was established.
Ask a specific question about this device
(123 days)
St. Jude Medical (now Abbott Medical)
The OPTIS imaging system with a compatible Dragonfly™ imaging catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal procedures. The compatible Dragonfly™ imaging catheters are intended for use in vessels 2.0 to 3.5 mm in diameter. The compatible Dragonfly™ imaging catheters are not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The OPTIS imaging system is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
OPTISTM with AptiVueTM Software (version E.5) perform Optical Coherence Tomography (OCT), Fractional Flow Reserve (FFR), and Resting Full-cycle Ratio (RFR) procedures and provide images of the coronary arteries in patients who are candidates for transluminal interventional procedures. FFR, Pd/Pa at rest, and RFR physiological waveforms measured by the system are used to assess the severity of a coronary lesion by measuring the pressure drop across the lesion (distal vs proximal pressure).
Here's an analysis of the acceptance criteria and study information for the ILUMIEN OPTIS with AptiVue Software version E.5, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance:
The provided document describes the validation of the Resting Full-cycle Ratio (RFR) feature, comparing its diagnostic performance against Fractional Flow Reserve (FFR) and Instantaneous Wave-free Ratio (iFR). The acceptance criteria are implicitly defined by achieving comparable diagnostic accuracy and agreement metrics.
Metric (vs FFR) | Acceptance Criteria (Implicit: Comparable to iFR-FFR) | Reported Device Performance (RFR-FFR) |
---|---|---|
Diagnostic Accuracy | Comparable to 92.2% [89.5%, 94.4%] | 93.6% [91.1%, 95.6%] |
Percent Positive Agreement | Comparable to 88.8% [84.1%, 92.5%] | 91.3% [86.9%, 94.5%] |
Percent Negative Agreement | Comparable to 95.4% [92.1%, 97.6%] | 95.8% [92.6%, 97.9%] |
Positive Predictive Value (PPV) | Comparable to 94.7% [90.9%, 97.2%] | 95.2% [91.6%, 97.6%] |
Negative Predictive Value (NPV) | Comparable to 90.2% [86.1%, 93.5%] | 92.3% [88.4%, 95.1%] |
Diagnostic Accuracy Outside the Grey Zone | Comparable to 86.8% [82.4%, 90.4%] | 88.5% [84.1%, 92.0%] |
Lesions free from Hyperemic Agents | Comparable to 58.9% [54.4%, 63.2%] | 55.5% [51.0%, 59.9%] |
Patients free from Hyperemic Agents | Comparable to 54.3% [49.5%, 59.1%] | 50.8% [46.0%, 55.6%] |
2. Sample Size and Data Provenance:
The document does not explicitly state the exact sample size for the test set. However, it refers to a "prospective study... of RFR for the physiological assessment of coronary artery disease in real-world patients." The confidence intervals provided in the table, such as "91.1%, 95.6%" for diagnostic accuracy, indicate a substantial number of cases. Given the context of a 510(k) summary, specific details on country of origin are typically not provided, but the language suggests a clinical trial setting. The study was prospective.
3. Number of Experts and Qualifications:
The document does not specify the number of experts used to establish the ground truth for the test set, nor their qualifications.
4. Adjudication Method:
The document does not explicitly describe an adjudication method for the test set. It outlines a "RFR-FFR Hybrid Method Result Interpretation" which includes a grey zone where the decision would be "based on FFR." This implies FFR serves as the primary ground truth, and the RFR values are being compared against it.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study is mentioned. The document focuses on the performance of the RFR algorithm itself compared to established physiological indices (FFR and iFR). There is no information provided regarding the effect size of human readers improving with AI assistance vs. without AI assistance.
6. Standalone Performance:
Yes, a standalone performance study was done. The "Summary of RFR Validation Study" directly compares the diagnostic accuracy and agreement metrics of the RFR algorithm (standalone, or hybrid with FFR for the grey zone) against FFR as the gold standard, and against iFR-FFR for equivalence.
7. Type of Ground Truth Used:
The ground truth used is primarily Fractional Flow Reserve (FFR). For the RFR-FFR hybrid method, when RFR falls within a defined grey zone (0.86 ≤ RFR ≤ 0.93), the decision is explicitly stated to be "based on FFR." Therefore, FFR serves as the reference standard for determining ischemia-causing lesions.
8. Sample Size for the Training Set:
The document does not provide information about the sample size for the training set used to develop the AptiVue Software.
9. How the Ground Truth for the Training Set was Established:
The document does not provide information on how the ground truth for the training set was established. It describes the validation study of the RFR feature, not the development process of the software.
Ask a specific question about this device
(113 days)
St. Jude Medical (now Abbott Medical)
The QUANTIEN™ Measurement System is indicated to provide hemodynamic information for use in the diagnosis and treatment of coronary or peripheral artery disease.
The QUANTIEN measurement system is intended for use in the catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices.
The QUANTIENTTM Measurement System is a diagnostic computer designed to record, compute, display and store data from PressureWireTM guidewire and other external transducers. The information is displayed as graphs as well as numerical values on the screen. Relevant display data includes: systolic, diastolic and mean blood pressure, heart rate, Fractional Flow Reserve (FFR), Resting Full-Cycle Ratio (RFR) index, Pd/Pa and data from ECG.
Information on the display screen may also be transferred to the cathlab monitoring system or an offsite video monitor. Recorded procedures can be viewed on a PC, with application specific viewing software installed such as RadiView, for post procedural review and analysis.
Quantien allows for importing of a patient work list from the hospital DICOM system, exporting of recorded measurement data to DICOM or to an external server location or portable (USB) memory device.
Here's a summary of the acceptance criteria and study details for the QUANTIEN™ Measurement System with Software Version 1.12.1. as presented in the provided document:
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Acceptance Criteria (Threshold) | Reported Device Performance (RFR-FFR Hybrid Method) |
---|---|---|
Diagnostic Accuracy | Not explicitly stated as a numerical threshold, but implicitly defined by the comparative equivalence to iFR-FFR. | 93.6% [91.1%, 95.6%] |
Percent Positive Agreement | Not explicitly stated as a numerical threshold. | 91.3% [86.9%, 94.5%] |
Percent Negative Agreement | Not explicitly stated as a numerical threshold. | 95.8% [92.6%, 97.9%] |
Positive Predictive Value (PPV) | Not explicitly stated as a numerical threshold. | 95.2% [91.6%, 97.6%] |
Negative Predictive Value (NPV) | Not explicitly stated as a numerical threshold. | 92.3% [88.4%, 95.1%] |
Diagnostic Accuracy Outside the Gray Zone | Not explicitly stated as a numerical threshold. | 88.5% [84.1%, 92.0%] |
Note: The acceptance criteria are implicitly defined by demonstrating equivalence to the iFR-FFR hybrid approach, which itself serves as a benchmark for diagnostic utility in assessing coronary artery disease. The study aims to show that the RFR-FFR hybrid method performs comparably to the established iFR-FFR method.
Study Details
-
Sample Size used for the test set and the data provenance:
- Sample Size: Not explicitly stated as a numerical value for the overall study. However, the results are presented with 95% confidence intervals, which usually implies a sufficient sample size was used for statistical significance.
- Data Provenance: Prospective study. The country of origin of the data is not specified in the provided document.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document does not specify the number of experts or their qualifications used to establish the ground truth.
- Instead, the "ground truth" for the comparison is RFR-FFR and iFR-FFR hybrid methods, where FFR is typically considered a reference standard in fractional flow reserve studies. The document does not describe an independent "expert consensus" or manual ground truth labeling for the images/data used. The comparison is between two algorithmic approaches.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not describe a specific adjudication method like 2+1 or 3+1. The study is a comparison of two physiological assessment methods (RFR-FFR vs. iFR-FFR), not a reader-based adjudication process for image interpretation.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done to evaluate human reader improvement with AI assistance. This study focuses on the comparison of two physiological indices (RFR and iFR) against FFR as the reference, not on human reader performance.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance metrics (Diagnostic Accuracy, Agreement, PPV, NPV) presented for "RFR-FFR" and "iFR-FFR" represent the standalone performance of these hybrid methods. They involve algorithmic computation and interpretation based on defined thresholds (e.g., RFR
Ask a specific question about this device
(56 days)
St Jude Medical
The GuardianTM burr hole cover system is intended for use following cranial surgery as an implantable 14-mm (0.55-in) burr hole cover for the skull. It can also be used to secure a lead with a 1.29-mm (0.051-in) or 1.39-mm (0.055-in) diameter
The GuardianTM burr hole cover system is used to close a cranial burr hole and secure an implanted, compatible lead, when applicable. The burr hole cover system is nonpyrogenic and has three main features: base, clip, and cover. The base is intended for burr holes with a 14-mm (0.55-in) diameter. It contains two grooved slots to hold a lead in place. The clip fits into the base to hold the lead. The locking mechanism temporarily holds a lead in place before the burr hole cover is secured. The cover snaps onto the base, closing the burr hole and locking a lead in place.
The document describes a 510(k) premarket notification for a modification to the Guardian™ Burr Hole Cover System to include an "MR Conditional" statement in the labeling. The acceptance criteria and the study that proves the device meets them are related to this MR Conditional status.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Related to MR Conditional Status) | Reported Device Performance (Guardian™ Burr Hole Cover System) |
---|---|
Device establishes "MR Conditional" status (as per FDA Guidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment") | Demonstrated to be "MR Conditional" |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state a "test set" in the context of medical imaging or diagnostic device performance evaluation. Instead, the testing was for the "MR Conditional" status. This typically involves physical testing of the device itself according to specific MR safety standards. Therefore, the "sample size" would refer to the number of physical devices tested to assess their behavior in an MR environment. The document does not specify the exact number of units tested.
- Data Provenance: The testing was conducted "following the FDA Guidance, 'Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment'." This implies a laboratory or testing facility setting, rather than clinical patient data. The country of origin of the data is not specified, but the submission is to the US FDA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This question is not directly applicable to this type of submission. Establishing "MR Conditional" status for a passive implant primarily involves engineering and physics expertise for conducting MR safety tests and interpreting their results according to established international standards (e.g., ASTM F2052, F2119, F2182). It does not typically involve a panel of medical experts establishing "ground truth" on patient images or outcomes. The "ground truth" in this context is the objective measurement of MR-related effects (e.g., heating, artifact, force).
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
Not applicable. MR safety testing data is typically objective and quantitative, and does not require adjudication by medical experts in the way clinical diagnostic study results might.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No, an MRMC comparative effectiveness study was not done. This submission is for a burr hole cover system (a passive implant), not a diagnostic imaging AI algorithm.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done
Not applicable. This is a medical device, not an algorithm. The "standalone" performance refers to the device's inherent physical properties and behavior in an MR environment.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for determining MR Conditional status is based on objective, quantifiable measurements from MR safety testing protocols defined in recognized standards (e.g., SAR measurements, temperature rise, displacement force, torque, image artifact assessment). This is not derived from expert consensus, pathology, or outcomes data in the traditional sense of a diagnostic study.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not an AI/machine learning device. The testing involved physical devices.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set.
Ask a specific question about this device
Page 1 of 16