Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212924
    Date Cleared
    2021-12-13

    (90 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The nitrile examination gloves are intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

    AI/ML Overview

    The provided text describes the regulatory filing for Nitrile Examination Gloves and includes non-clinical testing results. It does not involve a medical device with an AI algorithm, a diagnostic purpose, or human-in-the-loop performance. Therefore, many of the requested fields are not applicable.

    Here's the information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical DimensionsLength: S: ≥220mm; M/L/XL: ≥230mmLength: >230mm / Pass
    Width: S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mmWidth: S: 83-85mm / Pass; M: 94-96mm / Pass; L: 110-112mm / Pass; XL: 113-117mm / Pass
    Thickness: Finger: ≥0.05mm; Palm: ≥0.05mmThickness: Finger: 0.12-0.15mm / Pass; Palm: 0.08-0.10mm / Pass
    ASTM D5151Watertightness (Holes)Meet the requirements of ASTM D5151 AQL 2.50/200 / Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124