Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K171550
    Date Cleared
    2018-07-13

    (413 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sanrea Healthcare Products Pvt Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Latex Surgical Gloves, Powder free, is a disposable device made of natural rubber; intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The subject device is a surgical glove made of natural rubber latex compound. It is sterile, Powder free. The device is anatomic and is hand specific. The device meets ASTM D3577 – 09 : Standard specification for Rubber Surgical Gloves. The device is for over-the counter single use.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Sterile Latex Surgical Gloves, Powderfree. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness or performance for an AI/ML-based device.

    Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device (e.g., sample size for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) is not applicable to this document.

    The document primarily addresses the following for medical devices like surgical gloves:

    • Acceptance Criteria & Device Performance: This section is presented as a comparison table against the predicate device and relevant ASTM/ISO standards.
    • Non-Clinical Testing: The study proves the device meets acceptance criteria through various standardized tests.

    Here's the relevant information extracted from the document concerning the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicsTest StandardAcceptance Criteria (from Standard)Reported Device Performance (Sterile Latex Surgical Gloves, Powderfree)Result
    Freedom from HolesASTM D3577-09 / ASTM D5151-06 (Re-approved 2011) / FDA 1000ml water leak testSampling plan/Inspection level/AQL: ISO 2859-1 / G1/AQL 2.5 (for FDA 1000 ml water leak test)PASS (for FDA 1000ml water leak test)PASS
    Dimensions: LengthASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Overall length min 265 mm> 265 mmPASS
    Dimensions: WidthASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Varies from 76 mm for 6.0 size to 114 mm for 9.0 size (76±6 mm to 114±6 mm for sizes 6.0 to 9.0)76±6 mm to 114±6 mm (sizes 6.0 to 9.0)PASS
    Dimensions: ThicknessASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Minimum 0.10mm (Cuff, palm & finger)> 0.10 mm (Cuff, palm& finger)PASS
    Physical Properties: Before agingASTM D3577-09 and ASTM D412-06Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Tensile strength 24 Mpa min; Ultimate elongation 750% minTensile strength : > 24 Mpa ; Ultimate Elongation : >750%PASS
    Physical Properties: After Accelerated agingASTM D3577-09 and ASTM D573-04Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Tensile strength 18 Mpa min; Ultimate elongation 560 % minTensile strength : > 18 Mpa ; Ultimate Elongation : > 560%PASS
    Powder-free residueASTM D3577-09 and ASTM D6124-06Sampling plan: N=5 ; Less than 2 mg per gloveLess than 2 mg per glovePASS
    Biocompatibility: Primary skin irritationISO 10993-10Under the conditions of the study the device is not an irritantUnder the conditions of the study the device is not an irritantPASS
    Biocompatibility: Skin SensitizationISO 10993-10Under the conditions of the study the device is not a sensitizerUnder the conditions of the study the device is not a sensitizerPASS
    SterilityISO-11737-2SterileSterilePASS
    EO ResidueISO 10993-07:2008Maximum limit of EO Residue : 0.0025mg/deviceMeetsMeets
    ECH ResidueISO 10993-07:2008Maximum limit of ECH Residue : 0.0025mg/deviceMeetsMeets
    Expiration dating/Shelf lifeASTM D7160-05Not explicitly stated in criteria table, but mentioned as "Three years"Three yearsMeets
    SterilizationISO 11135-2014(E)Requirements for the development, validation and routine control of a sterilization process for medical devicesMeetsMeets

    2. Sample size used for the test set and the data provenance

    • Sample Sizes: Sample sizes are indicated within the "Sampling plan / Inspection level / AQL" columns in the performance data table (e.g., ISO 2859-1 / G1 / AQL 2.5, ISO 2859-1 / S2 / AQL 4.0, N=5 for powder-free residue). For the specific Freedom from Pinholes test, it notes: "Lot Size : 35001 – 150000 pair; Sample size : 200 pair".
    • Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective" in the context of clinical studies, as this is laboratory/manufacturing performance data. The manufacturer is SANREA HEALTHCARE PRODUCTS PVT LTD, located in Palakkad, Kerala, India.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: This type of testing involves laboratory and physical property measurements against established international standards (ASTM, ISO, FDA). "Ground truth" in this context is defined by the technical specifications and methodologies of these standards, not expert adjudication of medical images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: As above, this pertains to physical and chemical testing against defined standard metrics.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a premarket notification for a Class I medical device (surgical gloves), not an AI/ML-based diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standardized Test Methods and Specifications: The "ground truth" for the performance of these gloves is defined by the requirements and test methodologies outlined in the referenced ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards (e.g., ASTM D3577-09 for rubber surgical gloves, ASTM D5151-06 for freedom from holes, ISO 10993 for biocompatibility, ISO 11135 for sterilization). The FDA 1000ml water leak test is also mentioned.

    8. The sample size for the training set

    • Not Applicable: There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not Applicable: There is no "training set" as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171367
    Date Cleared
    2018-01-26

    (262 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sanrea Healthcare Products Pvt Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Latex Patient Examination gloves, Powder Free, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device is a patient examination glove made of latex compound. It is sterile, Powder Free. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM D3578-05: Standard specification for Rubber Examination Gloves. The subject device is substantially equivalent to legally marketed Latex Examination Powder Free gloves identified as Product code LYY. The device is for over-the counter single use.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: SANCARE STERILE LATEX EXAMINATION GLOVES.

    It describes the device's characteristics and compares its performance against a legally marketed predicate device (K052728, Powder free Latex Sterile Patient Examination Gloves from M/S. KANAM LATEX INDUSTRIES PVT. LTD.) to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the SANCARE STERILE LATEX EXAMINATION GLOVES are primarily based on established ASTM standards and FDA water leak tests, as compared to the performance of the predicate device. The document explicitly states that the device "Meets" all specified standards.

    CharacteristicTest / StandardAcceptance Criteria (from Standard)Reported Device Performance (SANCARE STERILE LATEX EXAM GLOVES)Result
    Freedom from PinholesFDA 1000 ml water leak test, ASTM D5151-06 (Re-approved 2011)ISO 2859-1 / G1 / AQL 2.5 (PASS/FAIL)PASSPASS
    Dimensions
    LengthASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (min 230 mm)> 230 mmPASS
    WidthASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (70±10 mm to 120±10 mm (sizes XS to XL))70±10 mm to 120±10 mm (sizes XS to XL)PASS
    ThicknessASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (min 0.08 mm (Cuff, palm & finger))> 0.08 mm (Cuff, palm& finger)PASS
    Physical Properties
    Before aging (Tensile Strength)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (> 18 Mpa)> 18 MpaPASS
    Before aging (Ultimate Elongation)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (> 650%)>650%PASS
    Before aging (Stress at 500% elongation)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (5.5 Mpa max)5.5 Mpa maxPASS
    After Accelerated aging (Tensile Strength)ASTM D3578-05 / ASTM D573-04ISO 2859-1 / S2 / AQL 4.0 (> 14 Mpa)> 14 MpaPASS
    After Accelerated aging (Ultimate Elongation)ASTM D3578-05 / ASTM D573-04ISO 2859-1 / S2 / AQL 4.0 (> 500%)> 500%PASS
    Powder-free residueASTM D3578-05 / ASTM D6124-06N=5 (Less than 2 mg per glove)Less than 2 mg per glovePASS
    Residual ProteinASTM D3578-05 / ASTM D5712-15N=3 (Less than 50µg/dm²)Less than 50µg/dm²PASS
    Biocompatibility
    Primary skin irritationISO 10993-10Under the conditions of the study, the device is not an irritantNon-irritantPASS
    Skin SensitizationISO 10993-10Under the conditions of the study, the device is not a sensitizerNon-sensitizerPASS
    SterilityISO 11737-2SterileSterilePASS
    Expiration dating/Shelf lifeASTM D7160-05(Not explicitly stated as Pass/Fail but confirmed)Three yearsMeets
    SterilizationISO 11135-2014(E)(Not explicitly stated as Pass/Fail but confirmed)Meets (Ethylene Oxide)Meets

    2. Sample Sizes Used for Test Set and Data Provenance

    Specific sample sizes for each test are provided within the table under "Sampling plan / Inspection level / AQL":

    • Freedom from Pinholes: ISO 2859-1 / G1 / AQL 2.5
    • Dimensions (Length, Width, Thickness): ISO 2859-1 / S2 / AQL 4.0 for each
    • Physical Properties (Before & After aging): ISO 2859-1 / S2 / AQL 4.0 for each
    • Powder-free residue: N=5
    • Residual Protein: N=3

    The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that it's a 510(k) submission for a finished medical device (gloves), the performance data would typically be generated by the manufacturer through in-house testing or contracted labs, following international standards (ASTM, ISO).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable to this type of device and submission. This is a submission for examination gloves, which are evaluated based on physical, chemical, and biological performance characteristics against recognized standards, not clinical diagnostic accuracy requiring expert interpretation of images or patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable. The device's performance is determined by objective laboratory tests against pre-defined thresholds from standards, not through adjudication of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (e.g., AI algorithms for medical imaging) that involve human readers' interpretation of outputs. For examination gloves, the evaluation criteria are entirely different, focusing on physical properties, safety, and sterility.

    6. If a Standalone Study Was Done (Algorithm Only Without Human-in-the-loop Performance)

    No, a standalone "algorithm only" study was not done. This concept is specific to AI or software as a medical device (SaMD). The SANCARE STERILE LATEX EXAMINATION GLOVES are a physical, Class I medical device. The performance data presented (e.g., tensile strength, freedom from holes) are inherent characteristics of the gloves themselves, independent of any algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation is based on established international performance standards (ASTM D3578-05, ASTM D5151-06, ASTM D412-06, ASTM D6124-06, ASTM D5712-15, ISO 10993-10, ASTM D7160-05, ISO 11135-2014(E), ISO 11737-2) and FDA-specific tests (FDA 1000ml water leak test). These standards define acceptable ranges or conditions for physical properties, biocompatibility, and sterility.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" in the context of examination gloves. This term typically refers to data used to train machine learning models, which are not involved in the manufacturing or evaluation of this device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171093
    Date Cleared
    2017-07-19

    (98 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sanrea Healthcare Products Pvt Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Nitrile Patient Examination gloves, Powder free, Blue color, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device is a patient examination glove made of synthetic nitrile latex compound. It is sterile, Powder free and is Blue in colour. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM D6319-10: Standard specification for Nitrile Examination Gloves for Medical Application. The subject device is substantially equivalent to legally marketed Nitrile examination gloves identified as Product code LZA. The device is for over-the counter single use.

    AI/ML Overview

    The provided document describes the Sanrea Healthcare Products Pvt Ltd's Sterile Nitrile Patient Examination Gloves, Powder free, Blue color, and its equivalence to a predicate device. The information details the performance testing and acceptance criteria for this Class I medical device.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsTest Standard / Acceptance CriteriaReported Device Performance
    Freedom from HolesASTM D6319-10 / ASTM D5151-06 (Reapproved 2011) with Sampling plan / Inspection level / AQL: ISO 2859-1 / G1/AQL 2.5; FDA 1000 ml water leak testMeets / PASS
    Dimensions: LengthASTM D6319-10; ISO 2859-1 / S2/AQL 4.0; > 230 mmMeets / PASS (> 230 mm)
    Dimensions: WidthASTM D6319-10; ISO 2859-1 / S2/AQL 4.0; 70±10 mm to 120±10 mm (sizes XS to XL)Meets / PASS (70±10 mm to 120±10 mm - sizes XS to XL)
    Physical Properties: ThicknessASTM D6319-10; ISO 2859-1 / S2/AQL 4.0; > 0.05 mm (palm & finger)Meets / PASS (> 0.05 mm - palm & finger)
    Physical Properties: Tensile strength (Before aging)ASTM D6319-10 and ASTM D412-06; ISO 2859-1 / S2/AQL 4.0; > 14 MpaMeets / PASS (> 14 Mpa)
    Physical Properties: Ultimate Elongation (Before aging)ASTM D6319-10 and ASTM D412-06; ISO 2859-1 / S2/AQL 4.0; > 500%Meets / PASS (> 500%)
    Physical Properties: Tensile strength (After Accelerated Aging)ASTM D6319-10 and ASTM D412-06; ISO 2859-1 / S2/AQL 4.0; > 14 Mpa (Aging done at 70 ±2 deg C for 166±2 hrs or 100±2deg C for 22±0.3 hrs)Meets / PASS (> 14 Mpa)
    Physical Properties: Ultimate Elongation (After Accelerated Aging)ASTM D6319-10 and ASTM D412-06; ISO 2859-1 / S2/AQL 4.0; > 400% (Aging done at 70 ±2 deg C for 166±2 hrs or 100±2deg C for 22±0.3 hrs)Meets / PASS (> 400%)
    Powder-free residueASTM D6319-10 and ASTM D6124-06; N=5; Less than 2 mg per gloveMeets / PASS (Less than 2 mg per glove)
    Biocompatibility: Primary skin irritationISO 10993-10; Under the conditions of the study, the device is not an irritant.Non-irritant / PASS (Under the conditions of the study, the device is not an irritant.)
    Biocompatibility: Skin SensitizationISO 10993-10; Under the conditions of the study, the device is not a sensitizer.Non-sensitizer / PASS (Under the conditions of the study, the device is not a sensitizer.)
    Expiration dating/Shelf lifeASTM D7160-05Three years
    SterilizationISO 11135-2014(E) (Ethylene Oxide)Meets
    SterilityISO 11737-2; Test of sterility performed in the definition, validation and maintenance of sterilization process.Sterile / PASS

    2. Sample size used for the test set and the data provenance

    • Sample Size for Physical Properties and Freedom from Holes: The sample sizes are generally cited within the "Sampling plan / Inspection level / AQL" column. For example, for "Freedom from Holes," it refers to "ISO 2859-1 / G1/AQL 2.5." For "Dimensions" and "Physical Properties," it refers to "ISO 2859-1 / S2/AQL 4.0." For "Powder-free residue," N=5 is used. The specific number of gloves tested within these AQL standards is not explicitly stated as a single number but is dictated by the standard.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. The manufacturer is SANREA HEALTHCARE PRODUCTS PVT LTD, located in Palakkad, Kerala, India. It can be inferred that the testing was conducted either internally by the manufacturer or by a third-party laboratory, likely in India, to demonstrate compliance with the cited ASTM and ISO standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable to this type of device (patient examination gloves). The "ground truth" for these tests is established by objective physical, chemical, and biological measurements performed according to recognized international standards (ASTM, ISO). There is no interpretation by human experts required to establish a diagnostic ground truth, unlike with image-based AI devices.

    4. Adjudication method for the test set

    This is not applicable as the tests are objective measurements against defined physical, chemical, and biological criteria, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This document describes a medical device (examination gloves) that does not incorporate AI or involve human readers for diagnostic interpretation. Therefore, an MRMC study is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an inanimate object (nitrile gloves) and does not involve any algorithms.

    7. The type of ground truth used

    The "ground truth" for the performance claims of these gloves is based on:

    • Standardized Physical Measurements: Length, width, thickness, tensile strength, ultimate elongation performed according to ASTM D6319-10 and ASTM D412-06.
    • Standardized Chemical Measurements: Powder-free residue performed according to ASTM D6124-06.
    • Standardized Biological Testing: Biocompatibility (primary skin irritation and skin sensitization) performed according to ISO 10993-10, and sterility testing according to ISO 11737-2.
    • Performance Tests: Freedom from holes measured by FDA 1000 ml water leak test and ASTM D5151-06.

    8. The sample size for the training set

    This is not applicable. There is no "training set" for this type of medical device as it does not involve machine learning or AI algorithms. The performance is assessed through direct physical, chemical, and biological testing of manufactured samples.

    9. How the ground truth for the training set was established

    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1