Search Filters

Search Results

Found 14 results

510(k) Data Aggregation

    K Number
    K102096
    Date Cleared
    2010-10-04

    (69 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Chemotherapy DrugAverage BDT
    Fluorouracil>240 min.
    Etoposide (Toposar)>240 min.
    Cyclophosphamide (Cytoxan)>240 min.
    Carmustine2.03 min.
    Thiotepa25.23 min.
    Paclitaxel (Taxol)>240 min.
    Doxorubicin Hydrochloride>240 min.
    Dacarbazine (DTIC)>240 min.
    Cisplatin>240 min.
    Please note that Carmustine and Thiotepa have extremely low permeation times of less than 30 minutes.
    Device Description

    Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Polymer Patient Examination Gloves, 80 LZA, and meets all requirements of ASTM standard D 6319-00a (2005)e1.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for "Syntex Healthcare Products Co., Ltd Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs (Blue)". This is a medical device approval application, not a study of an AI/ML device. Therefore, the specific questions related to AI/ML device performance, such as sample sizes for test sets, expert ground truth, MRMC studies, and standalone algorithm performance, are not applicable.

    However, I can extract the acceptance criteria and performance data for this device as it relates to the non-clinical tests conducted.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally established by recognized standards like ASTM D 6319-00a (2005)e1, ASTM D6124-06, ISO 10993 Part 10, and 16CFR 1500.41/1500.3. The "Performance Results" column indicates if the device meets these standards.

    CharacteristicsApplicable Recognized StandardsReported Device Performance
    DimensionsASTM D 6319-00a (2005)e1Meets
    Physical PropertiesASTM D 6319-00a (2005)e1Meets
    Freedom from holes(FDA Pinhole requirements)Meets
    Residual Powder TestASTM D 6319-00a (2005)e1, ASTM D6124-06Meets
    Primary Skin Irritation and Skin SensitizationISO 10993 Part 10, 16CFR 1500.41, 16CFR 1500.3Meets
    Resistance to Permeation (Chemotherapy Drugs)ASTM D6978-05See specific breakthrough detection times below

    Breakthrough Detection Times (BDT) for Chemotherapy Drugs (Resistance to Permeation):

    Chemotherapy DrugAverage BDT
    Fluorouracil>240 min.
    Etoposide (Toposar)>240 min.
    Cyclophosphamide (Cytoxan)>240 min.
    Carmustine2.03 min.
    Thiotepa25.23 min.
    Paclitaxel (Taxol)>240 min.
    Doxorubicin Hydrochloride>240 min.
    Dacarbazine (DTIC)>240 min.
    Cisplatin>240 min.

    Important Note: The document explicitly states: "Please note that Carmustine and Thiotepa have extremely low permeation times of less than 30 minutes." This implies that while the device was tested against these drugs, their low BDTs are a critical factor for users to consider, and the labeling states "Do not use with Carmustine and Thiotepa."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each non-clinical test (e.g., how many gloves were tested for dimensions, physical properties, or permeation). It mentions the tests were performed by Syntex Healthcare Products Co., Ltd in China (the submitter's location), so the data likely originates from China. The tests are non-clinical and would be considered prospective for the purpose of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    These questions are designed for AI/ML device evaluations. The presented document describes a traditional medical device (examination gloves) and its non-clinical testing. Therefore, these questions are not applicable to this submission. The "ground truth" for this device is determined by physical and chemical testing against established standards (e.g., ASTM standards for glove properties and permeation resistance). No AI/ML components are mentioned or evaluated in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052877
    Date Cleared
    2005-11-08

    (27 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
    A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power-Free Nitrile Patient Examination Glove, 80 LZA, and meets all requirements of ASTM standard D-6319-00€3

    AI/ML Overview

    This document describes the validation of a medical device, specifically a "Powder Free Nitrile Blue Textured Examination Glove." However, it focuses on the physical and biological characteristics of the glove and its substantial equivalence to a predicate device, rather than the performance of an AI or software-based medical device.

    Therefore, many of the requested categories (e.g., sample size for test set, ground truth experts, MRMC studies, standalone algorithm performance, training set details) are not applicable to this type of device submission.

    Here's the information that can be extracted and presented, with clarifications for the questions that don't apply:

    Acceptance Criteria and Reported Device Performance

    CriteriaReported Device Performance
    Physical and Dimensions TestingMeets requirements of ASTM standard D-6319-00€3. Inspection level S-2, AQL 4.0.
    Water Fill Test (Pinhole)Meets requirements of the FDA 1000 ml. Water Fill Test with samplings of AQL 2.5, inspection level G-1.
    Primary Skin IrritationResults show no primary skin irritant reactions.
    Skin Sensitization (Allergic Contact Dermatitis)Results show no sensitization reactions.
    Residual PowderMeets "powder-free" claims (contains no more than 2 mg powder per glove) based on ASTM D-6124-01 for Starch at finished inspection.
    Compliance with ASTM StandardConforms fully to ASTM D-6319-00E standard.
    Compliance with 21 CFR ReferencesConforms fully to applicable 21 CFR references (specifically 21 CFR 880.6250, specifying Class I, Power-Free Nitrile Patient Examination Glove, 80 LZA).
    Labeling ClaimsNo special labeling claims; does not claim gloves as hypoallergenic. (This is a statement of compliance with labeling requirements rather than a performance metric, but it confirms adherence to the device description).

    Study Details (as applicable for a physical medical glove)

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a single "test set" for performance evaluation in the way a software algorithm would be. Instead, testing was conducted on samples from manufacturing lots based on AQLs (Acceptable Quality Levels).
        • Physical and Dimensions Testing: Inspection level S-2, AQL 4.0.
        • Water Fill Test: AQL 2.5, inspection level G-1.
      • Data Provenance: Not specified, but generally, such testing would be conducted by the manufacturer (Syntex Healthcare Products Co., Ltd.) in China, where the device is manufactured. The testing is likely retrospective in the sense that it's performed on finished products, but it represents ongoing quality control.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. The "ground truth" for a physical glove is determined by established physical and chemical test methods against defined standards (ASTM, FDA). It does not involve expert consensus in the way an AI diagnostic tool would for clinical images.
    3. Adjudication method for the test set:

      • Not Applicable. Pertains to expert review for ambiguous cases in diagnostic studies, not physical product testing.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This is a physical, disposable medical device, not an AI software.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable. This is a physical, disposable medical device, not an AI algorithm.
    6. The type of ground truth used:

      • Established Standards and Test Methods: The ground truth for this device's performance is based on recognized consensus standards (ASTM D-6319-00€3, ASTM D-6124-01) and regulatory test methods (FDA 1000 ml. Water Fill Test), as well as biocompatibility testing for skin irritation and sensitization.
    7. The sample size for the training set:

      • Not Applicable. This refers to machine learning algorithms, which are not involved here.
    8. How the ground truth for the training set was established:

      • Not Applicable. As above, no training set for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042226
    Date Cleared
    2004-09-30

    (44 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Patient Examination Glove, 80LZA, and meets all requirements of ASTM Standard D6319-00aE3.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device: "Syntex Powder-Free Nitrile Examination Glove, Tested for Chemotherapy." Based on the provided document, the device is a Class I patient examination glove. The document describes non-clinical tests to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / StandardReported Device Performance
    Physical and Dimensions Testing:
    ASTM-D-6319-00aE3 (General Glove Standard)"All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 4.0." And "Syntex Healthcare Products Co., Ltd. Syntex Powder-Free Nitrile Examination Glove, Tested for Chemotherapy conform fully to ASTM-D-6319-00a€3 standard..."
    Pinhole / Water Leak Test:
    FDA 1000 ml. Water Fill Test (AQL 2.5, Inspection Level G-1)"The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level G-1, meeting these requirements." And "...meets pinhole FDA requirements."
    Biocompatibility (Skin Irritation/Sensitization):
    Primary Skin Irritation and Skin Sensitization testing"Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions." And "...biocompatibility requirements..."
    Powder Residue (for "powder-free" claim):
    ASTM D6124-01 for Starch (no more than 2 mg powder per glove)"A Residual Powder Test that based on ASTM D6124-01 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free' claims (contain no more than 2 mg powder per glove)."
    Chemotherapy Permeation (Implied based on device name, but not explicitly detailed in the summary)The device is named "Tested for Chemotherapy," implying specific testing for this use. However, the 510(k) summary does not explicitly detail the acceptance criteria or results for chemotherapy permeation testing. It only states that the device "conform[s] fully to ASTM-D-6319-00a€3 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements. biocompatibility requirements and labeling claims as shown by data in Section 7." The lack of specific details for chemotherapy testing in Section 7 (Discussion of Non-Clinical tests) and Section 8 (Discussion of Clinical Tests) is a notable omission in this summary regarding the "Tested for Chemotherapy" aspect.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Physical and Dimensions Testing: Inspection Level S-2, AQL 4.0. (This refers to a sampling plan for quality control rather than a specific sample size for a "test set" in an AI/Software context).
    • FDA 1000 ml. Water Fill Test: AQL 2.5, Inspection Level G-1. (Again, a sampling plan for quality control).
    • Biocompatibility (Skin Irritation/Sensitization), Residual Powder Test: No specific sample sizes are mentioned for these tests, only that testing was "conducted."
    • Data Provenance: Not applicable in the context of this device (a physical glove). The tests are performed on manufactured gloves.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this type of device (a physical examination glove). Ground truth in this context is established by physical measurements and laboratory analyses against defined standards, not by expert interpretation of data.

    4. Adjudication Method for the Test Set

    This information is not applicable to this type of device. Adjudication methods are typically used in studies involving human interpretation or subjective assessment.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable to this type of device. An MRMC study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) use the AI for assistance. This device is a physical glove.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable to this type of device. "Standalone performance" refers to the performance of an algorithm without human intervention, which is not relevant for a physical glove.

    7. The Type of Ground Truth Used

    The ground truth for this device's performance is based on defined physical and chemical standards as outlined in relevant ASTM standards and FDA requirements. This includes:

    • Physical dimensions and tensile strength (ASTM-D-6319-00aE3)
    • Barrier integrity (FDA 1000 ml. Water Fill Test for pinholes)
    • Biocompatibility (Primary Skin irritation and Skin Sensitization testing via laboratory methods)
    • Powder content (ASTM D6124-01)
    • (Implied: Chemotherapy permeation data based on the device name, but methods/results are not detailed in this summary).

    8. The Sample Size for the Training Set

    This information is not applicable to this type of device. Training sets are relevant for machine learning algorithms, not for physical products like gloves.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable to this type of device, as it does not involve a training set for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033777
    Date Cleared
    2004-02-05

    (63 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between health care personnel and the patient.

    Device Description

    Ultrasense AF Blue Power-free Nitrile Examination Gloves

    AI/ML Overview

    This document pertains to the clearance of "Ultrasense AF Blue Power-Free Nitrile Examination Gloves" (K033777). As such, it is a device submission for a physical product (gloves), not a diagnostic AI/ML algorithm or software as a medical device (SaMD).

    Therefore, the requested information regarding acceptance criteria, study design, ground truth, expert involvement, and sample sizes for an AI/ML device is not applicable to this document.

    The document discusses the substantial equivalence determination for examination gloves based on regulatory requirements for a physical medical device, not a software-based diagnostic tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023826
    Date Cleared
    2003-01-24

    (67 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device made of natural rubber latex that may bear a trace amount of glove powder and is intended to be worn on the hand or finger(s) for medical purposes to provide a barrier against potentially infectious materials and other contaminants.

    Device Description

    Syntex Powder-Free Latex Examination Gloves - Blue Color. A powder-free patient examination glove is a disposable device made of natural rubber latex that may bear a trace amount of glove powder.

    AI/ML Overview

    The provided text is a scanned FDA 510(k) clearance letter for "Syntex Powder-Free Latex Examination Gloves". This document pertains to the regulatory clearance of a medical device based on substantial equivalence to a predicate device, rather than a study proving performance against specific acceptance criteria.

    The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976..."

    Therefore, the document does not contain the information requested regarding acceptance criteria, reported device performance, study details (sample size, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance), or information about training sets for an AI/algorithm-based device.

    The questions are designed for a scenario where an AI or algorithm-based device undergoes performance evaluation and validation, which is not the subject of this 510(k) clearance for a physical medical product like examination gloves.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022784
    Date Cleared
    2002-11-13

    (83 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder-Free Nitrile Examination Glove -- Purple Color

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a device called "Syntex Powder-Free Nitrile Examination Glove, Purple Color." This document does not contain information about acceptance criteria, device performance studies, sample sizes, expert ground truth establishment, or any details related to AI or comparative effectiveness studies.

    The letter explicitly states that the device is "substantially equivalent" to legally marketed predicate devices. This means that the FDA has determined the new device is as safe and effective as a device already on the market, based on its indications for use, technological characteristics, and performance data. However, the performance data itself, including specific acceptance criteria and study results, is typically found in the 510(k) submission summary, which is not part of this letter.

    Therefore, I cannot provide the requested information from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022092
    Date Cleared
    2002-09-03

    (68 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Syntex Pre-Powdered Latex Examination Glove

    AI/ML Overview

    The provided text is a FDA 510(k) clearance letter for a medical device: "Syntex Pre-Powdered Latex Examination Glove."

    This type of document does not contain the detailed technical information requested about acceptance criteria and study designs that prove device performance.

    FDA 510(k) clearances are based on demonstrating substantial equivalence to a predicate device, not necessarily on new, extensive clinical studies proving efficacy or specific performance metrics in the way a PMA (Premarket Approval) would.

    Therefore, I cannot provide the requested information from the given input. The letter essentially states that the device is "substantially equivalent" for its indicated use based on the information provided in the 510(k) submission, and it lists various regulatory requirements for the manufacturer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022088
    Date Cleared
    2002-09-03

    (68 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A glove is worn on the hand of healthcare worker and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

    Device Description

    Syntex Powder-Free Latex Examination Glove

    AI/ML Overview

    I am sorry, but based on the provided text, I cannot extract the information required to describe the acceptance criteria and the study proving the device meets them. The document is a 510(k) premarket notification letter from the FDA to Syntex Healthcare Product Company Limited regarding their "Syntex Powder-Free Latex Examination Glove".

    This letter confirms that the device has been found substantially equivalent to a legally marketed predicate device and can be marketed. However, it does not include details about:

    1. Acceptance criteria: There is no table of specific performance metrics or a detailed description of what constitutes "acceptance."
    2. Reported device performance: The letter does not contain any test results, data, or performance statistics for the Syntex glove.
    3. Study details: There is no mention of a specific study, its design, sample sizes, data provenance, expert involvement, adjudication methods, MRMC studies, standalone performance, or how ground truth was established for either training or test sets.

    The letter primarily focuses on the regulatory decision of substantial equivalence and general compliance requirements. To answer your request, I would need a different type of document, such as a summary of safety and effectiveness data (SSD), a clinical study report, or a more detailed 510(k) submission document that typically includes performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022090
    Date Cleared
    2002-09-03

    (68 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A glove is worn on the hand of healthcare worker and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

    Device Description

    Pre-Powdered Nitirile Examination Glove

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about acceptance criteria, device performance, or any studies conducted to prove the device meets acceptance criteria. The text provided is an FDA clearance letter for a medical device (Syntex Pre-Powdered Nitrile Examination Glove), outlining its substantial equivalence to a predicate device and regulatory compliance. It does not contain details about specific performance metrics or studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022146
    Date Cleared
    2002-09-03

    (63 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder-Free Polymer-Coated Latex Examination Glove

    AI/ML Overview

    This document is a 510(k) premarket notification letter for a Syntex Powder-Free Polymer-Coated Latex Examination Glove. It is from the FDA to Syntex Healthcare Products Co Ltd. and states that the device is substantially equivalent to legally marketed predicate devices.

    This document does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria. It is a regulatory approval letter based on substantial equivalence, not a performance study report.

    Therefore, I cannot provide the requested table or answer any of the questions about a performance study, sample sizes, ground truth, or expert qualifications based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2