Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K090161
    Manufacturer
    Date Cleared
    2009-06-09

    (138 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epiphany Injector System is a device intended to fold and insert STAAR Surgical Company aspheric three piece Collamer®, Model CQ2015A, and aspheric three piece silicone intraocular lenses, Model AQ2015A, for surgical placement in the human eye.

    Device Description

    The single use, disposable Epiphany Injector System is a delivery system for intraocular lenses (IOLs) for surgical placement in the human eye. The two main materials used in the manufacturer of this injector system will be polycarbonate and polypropylene. All product is also supplied "STERILE."

    AI/ML Overview

    The provided text describes the Epiphany Injector System, a disposable intraocular lens injector. However, it does not contain a detailed study proving the device meets specific acceptance criteria in the way requested by the prompt.

    Instead, the document serves as a 510(k) summary for regulatory submission, primarily focusing on demonstrating substantial equivalence to predicate devices. It references "test results" for "Mechanical Safety" but does not elaborate on these results or the acceptance criteria used.

    Therefore, I cannot fulfill many of the requested sections as the information is not present in the provided text. I will indicate where information is missing.


    Epiphany Injector System

    1. Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly list specific performance acceptance criteria for the Epiphany Injector System. It largely relies on a comparison to predicate devices to establish substantial equivalence.

    FeatureAcceptance Criteria (Not explicitly stated in document)Reported Device Performance (From "Mechanical Safety" - Test Results)
    Mechanical Safety(Implied to be "Similar" to predicate devices)"See test results" (Details not provided in the document)
    Product DescriptionA delivery system for intraocular lenses (IOLs)"Same" as predicate devices
    Intended UseAs described in Directions for Use"Similar" to predicate devices
    DesignAs per prints"Similar" to predicate devices
    Main MaterialsPolypropylene and polycarbonate"Similar" to predicate devices
    ManufacturingPer internal operating procedures"Similar" to predicate devices
    Operating PrincipleAn injector system that utilizes delivery mechanism for the insertion of an IOL"Similar" to predicate devices
    PackagingLabeled pouches and a tray with Directions for Use"Similar" to predicate devices, except MSI-P1 model is "Same"
    SterilitySterile"Same" as predicate devices, but different sterilization method for one

    2. Sample size used for the test set and the data provenance

    The document states, "A financial certification and/or disclosure statement is not needed for this submission as no clinical studies have been undertaken in regards to the products under consideration." This indicates that no clinical test set was used. The brief mention of "Mechanical Safety: See test results" implies some form of engineering or bench testing, but the sample size and data provenance are not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical studies or human expert-based test sets are described.

    4. Adjudication method for the test set

    Not applicable. No clinical studies or human expert-based test sets are described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical injector system, not an AI or imaging device that would typically involve MRMC studies with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical injector system, not an algorithm. Bench testing would be analogous to "standalone" performance for a mechanical device, but details are not provided.

    7. The type of ground truth used

    The primary "ground truth" for this submission appears to be the established performance and safety of legally marketed predicate devices, against which the Epiphany Injector System demonstrated "substantial equivalence." For "Mechanical Safety," internal test results are referenced, implying engineering specifications or performance benchmarks as ground truth, but these are not detailed.

    8. The sample size for the training set

    Not applicable. This device is a mechanical injector system, not an AI system requiring a training set in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073586
    Manufacturer
    Date Cleared
    2008-03-13

    (84 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MicroSTAAR ™ Indigo Injector Foam Tip System is a device intended to fold and insert STAAR Surgical one and three piece Collamer® lenses for surgical placement in the human eye.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for a medical device called the "MicroSTAAR™ Indigo Injector Foam Tip System." This letter grants market clearance based on a finding of "substantial equivalence" to legally marketed predicate devices, not on the results of a clinical study demonstrating the device meets specific acceptance criteria in the manner that an AI/ML device would be evaluated.

    Therefore, the information requested in your prompt (e.g., acceptance criteria, reported device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types) is not present in this document.

    This document is for a Class I medical device (intraocular lens guide) and focuses on regulatory clearance rather than a detailed performance study as would be required for a novel AI/ML diagnostic or prognostic device. The 510(k) process for this type of device typically relies on demonstrating that the new device has the same intended use and technological characteristics as a predicate device, or that any differences do not raise new questions of safety or effectiveness.

    In summary, this document does not contain the information necessary to answer your request regarding acceptance criteria and a study proving a device meets those criteria, as it pertains to AI/ML device evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073591
    Manufacturer
    Date Cleared
    2008-03-13

    (84 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MicroSTAAR ™ MSI Injector Foam Tip System is a device intended to fold and insert STAAR Surgical one and three piece Collamer® lenses for surgical placement in the human eye.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for the MicroSTAAR™ MSI Injector Foam Tip System, a device intended to fold and insert intraocular lenses.

    The provided text does not contain any information about acceptance criteria or a study that proves the device meets specific performance criteria.

    Therefore, I cannot fulfill your request for the detailed table and study information. The letter only confirms that the device is substantially equivalent to a legally marketed predicate device and can proceed to market, subject to general controls provisions. It does not include performance data, study design, or ground truth establishment relevant to the questions asked.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032412
    Manufacturer
    Date Cleared
    2004-06-10

    (310 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Collamer 3 Piece IOL Injector, is an intraocular lens guide used to ford and into the capsular bag or ciliary sulcus following cataract extraction.

    Device Description

    The Collamer 3 Piece Injector is made of titanium alloy designed for inserting intraocular lenses into the human eye following cataract extraction. The Collamer 3 Piece Injector materials and mechanical behavior are similar to those used with the predicate device.

    AI/ML Overview

    The provided document is a 510(k) summary for the STAAR Surgical Collamer 3 Piece Injector, and it describes a medical device (injector), not an AI/ML powered device, therefore most of the requested information (like acceptance criteria for an AI model, sample size for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, etc.) is not applicable to this document.

    However, I can extract the information relevant to the device's performance based on the provided text, which focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing.

    Here's the relevant information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implicit)Reported Device Performance
    Mechanical performance equivalent to predicate deviceThe Collamer 3 Piece Injector was found to perform equivalently to the predicate device, the MS1-P Elastic Lens Injector, with respect to mechanical performance.
    Mechanical behavior equivalent to predicate deviceThe Collamer 3 Piece Injector was found to perform equivalently to the predicate device, the MS1-P Elastic Lens Injector, with respect to mechanical behavior.
    Designed and tested in accordance with applicable safety standardsThe Collamer 3 Piece Injector has been designed and tested in accordance with applicable safety standards.

    2. Sample size used for the test set and the data provenance: Not applicable. This is not a study assessing an AI/ML algorithm's performance on a dataset. The testing described is non-clinical mechanical performance testing of a physical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML algorithm is not relevant here.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an AI algorithm.

    7. The type of ground truth used: Not applicable. The "ground truth" in this context would be the objective measurements and observations of the physical device's mechanical performance and behavior, compared to a predicate device and safety standards.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

    Summary of the Study:

    The study described is a non-clinical test to demonstrate the substantial equivalence of the Collamer 3 Piece Injector to its predicate device, the MS1-P Elastic Lens Injector (K861085). The study focused on validating the mechanical performance and mechanical behavior of the new injector.

    The device meets the (implicit) acceptance criteria because the non-clinical tests showed that it performs equivalently to the predicate device in terms of mechanical performance and behavior, and it was designed and tested in accordance with applicable safety standards. The FDA's clearance letter (K032412) confirms that the device is substantially equivalent, allowing it to be marketed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020734
    Manufacturer
    Date Cleared
    2002-05-16

    (71 days)

    Product Code
    Regulation Number
    886.4670
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surge-free Aspiration Adapter is a filter protected flow restriction orifice used during cataract surgery to prevent/neutralize vacuum surges in the aspiration line.

    Device Description

    The Surge-free Aspiration Adapter is a filter protected flow restriction device intended to be placed between the handpiece and aspiration tubing used during phacoemulsification surgery. The Surge-free Aspiration Adapter prevents/neutralizes vacuum surges in the aspiration line resulting from the sudden clearing of occlusions.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device called the "Surge-free Aspiration Adapter." This document focuses on demonstrating the substantial equivalence of the new device to existing legally marketed predicate devices, rather than detailed performance study results against specific acceptance criteria.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies cannot be fully extracted from the provided text. The document primarily describes the device, its intended use, and compares its technological characteristics to predicate devices. It does not contain information about a performance study with acceptance criteria.

    However, I can provide the partially available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific acceptance criteria or quantitative performance data. It describes the effect of the device: "Reduction in vacuum surges resulting from release of aspiration tubing occlusions."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not present in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not present in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not present in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not present in the document. This device is not an AI-assisted diagnostic tool; it's a mechanical adapter for surgical equipment.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not present in the document. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not present in the document.

    8. The sample size for the training set

    This information is not present in the document.

    9. How the ground truth for the training set was established

    This information is not present in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983129
    Manufacturer
    Date Cleared
    1998-12-04

    (87 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable MicroSTAAR™ Injector (Model MSI-P1) is a device used to fold and insert STAAR Surgical UV-ELASTICT™ silicone lenses for surgical placement in the human eye.
    The Disposable MicroSTAAR™ Injector is a device used to fold and insert STAAR Surgical Company UV-ELASTIC™ single-piece silicone lenses for surgical placement in the human eye.

    Device Description

    The Disposable MicroSTAART™ Injector (Model MSI-P1) is used to fold and insert STAAR Surgical UV-ELASTICTM silicone intraocular lenses for surgical placement in the human eye. Like other MicroSTAAR injectors it provides a sterile tubular pathway through a surgical incision, over the iris and into either the ciliary sulcus or the capsular bag of the eye. The Disposable MicroSTAART™ Injector (Model MSI-P1) is made from plastic and the inner dimensions of the delivery pathway, as well as the outer dimensions of the portion that contacts the eye tissues are identical to STAAR Surgical Company's currently marketed injection systems. The Disposable MicroSTAART™Injector (Model MSI-P1) is provided sterile and is designed for single use.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Premarket Notification [510(K)] Summary for the Disposable MicroSTAAR™ Injector System

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Functional Testing)Reported Device Performance
    No optic zone lens tears for properly loaded lenses.No lens tears were observed.
    All post-ejected lenses to meet a minimum resolution > 3.6.All post-ejected lenses met the minimum resolution requirements.
    Acceptance Criterion (Accelerated Aging Study - Microbial Challenge Dust Drum)Reported Device Performance
    All samples and negative controls to be "NEGATIVE".All samples and negative controls were "NEGATIVE".
    Positive controls to be "POSITIVE".Positive controls were "POSITIVE".
    Acceptance Criterion (Accelerated Aging Study - Dye Penetration)Reported Device Performance
    All samples to "Pass".All samples "Passed".
    Acceptance Criterion (Accelerated Aging Study - Peel Strength)Reported Device Performance
    Sample means and variances for baseline and one-year accelerated samples to be statistically similar (T-test and F-test).Results indicate sample means and variances are statistically similar.
    Acceptance Criterion (Accelerated Aging Study - Burst)Reported Device Performance
    For large pouch, sample means and variances for baseline and one-year accelerated samples to be statistically similar (T-test and F-test).Results indicate sample means and variances for the large pouch are statistically similar.
    For small pouch, no failures.No failures observed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Functional Testing:
      • Sample Size: Thirty (30) sterile Disposable MicroSTAAR™ Injectors (Model MSI-P1).
      • Data Provenance: Not explicitly stated but implies laboratory testing done by the manufacturer (STAAR Surgical Company) as part of their validation. It is a prospective test, as it's a validation performed for the device.
    • Accelerated Aging Study:
      • Sample Size: "Baseline (zero time) and one year accelerated samples" were used. The exact number of samples for each specific test (Microbial Challenge, Dye Penetration, Peel Strength, Burst) is not explicitly stated, but it would have been a subset of the manufactured devices.
      • Data Provenance: Similar to functional testing, this was likely a prospective laboratory study conducted by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to mechanical device testing like an IOL injector. The "ground truth" here is objective physical measurements and performance thresholds (e.g., absence of tears, resolution measurements, integrity assessments). No human experts were used to establish a ground truth in the sense of image interpretation or diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable. The tests involve objective measurements against predefined criteria, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    Not applicable. This device is a mechanical injector, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a mechanical injector, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for these tests was based on objective, measurable physical characteristics and performance criteria for the device and its packaging. This includes:

    • Absence of physical damage: Lens tears.
    • Performance metrics: Optical resolution post-ejection.
    • Sterility integrity: Microbial challenge results, dye penetration, peel strength, burst strength.
    • Statistical similarity: For aging studies, comparing mechanical properties of new vs. aged samples.

    8. The Sample Size for the Training Set

    Not applicable. This is a mechanical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K980696
    Manufacturer
    Date Cleared
    1998-09-11

    (200 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Foam Tip™ Injector is a device used to fold and insert STAAR Surgical UV-ELASTIC™ silicone lenses for surgical placement in the human eye.

    Device Description

    STAAR Surgical Company's Foam Tip™ Injector is used to fold and insert STAAR Surgical UV-ELASTIC™ silicone intraocular lenses for surgical placement into the human eye. The new Foam Tip™ Injector utilizes the same design principles (i.e. a plunger pushing or threaded mechanism) to push the IOL through a cartridge for injection into the eye as the existing MicroSTAART™ injectors. The new injector system differs from the MicroSTAART™ injector in that it contains a disposable Foam Tip™ plunger insert which may be attached to the sterilized injector body prior to lens loading. This attachable Foam Tip™ plunger insert is encapsulated inside a holder that protects the sponge from damage. The holder acts to guide the plunger as it slides into the injector body and performs as a wrench to remove the plunger from the injector once the lens has been implanted. Once the sponge portion of the foam plunger insert is hydrated, it enables the surgeon to have greater control in ejecting the lens properly through the cartridge and into the eye. Like the existing MicroSTAAR™ injectors, the new Foam Tip™ Injector body is provided sterile and can be resterilized for multiple use. The new Foam Tip™ plunger insert is also provided sterile but is intended for single use only.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided text for the Foam Tip™ Injector System (K980696):

    Acceptance Criteria and Device Performance for Foam Tip™ Injector System (K980696)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Functional Testing)Reported Device Performance (Functional Testing)
    No lens tearsNo evidence of lens tears
    All post-ejected lenses meet STAAR's minimum resolution efficiency requirementsAll 40 functional test lens ejections passed minimum resolution testing
    No haptic damageNo evidence of haptic damage
    Gross particulate evaluationParticulates were noted in both the Test and Control samples. None appeared to be greater than 10 microns in size.

    Additional Acceptance Criteria and Performance (Accelerated Aging Studies)

    Acceptance Criteria (Accelerated Aging Studies)Reported Device Performance (Accelerated Aging Studies)
    Acceptable microbial barrierResults of Microbial Challenge Dust Drum Test, Dye Penetration and Burst tests indicate both sample groups (baseline versus six month) provided an acceptable microbial barrier and kept product sterile within.
    Maintained product sterilityResults of Microbial Challenge Dust Drum Test, Dye Penetration and Burst tests indicate both sample groups (baseline versus six month) kept product sterile within.
    Acceptable seal strength (slight variations)Slight variations in seal strength were within an acceptable range and did not appear to have compromised the sterile barriers or the integrity of the product.
    Material non-toxicityCytotoxicity tests showed that the material was non-toxic.
    E-Beam sterilization as an acceptable methodFTIR demonstrated that E-Beam sterilization was an acceptable method of sterilization.
    Bacterial endotoxins not above FDA limitsLAL testing was performed which indicated that bacterial endotoxins were not above FDA limits, thereby providing acceptable results.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 40 functional ejection tests were conducted for the functional testing.
    • Data Provenance: The text does not explicitly state the country of origin or whether the study was retrospective or prospective. It describes laboratory or in-house testing performed by or for STAAR Surgical Company.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the use of "experts" to establish ground truth in the traditional sense of clinical opinion or image interpretation. The ground truth for functional performance (lens tears, haptic damage, resolution efficiency, particulates) was established through direct observation and measurement in a controlled testing environment, likely by trained technicians or engineers.

    4. Adjudication Method for the Test Set

    Not applicable. The criteria for functional testing were objective and directly observable/measurable (e.g., presence/absence of tears, adherence to resolution requirements). There was no "adjudication" between multiple expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This document describes non-clinical functional and accelerated aging tests of the device itself, not a comparative effectiveness study involving human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical injector system, not an algorithm or AI. The tests performed are standalone in the sense that they evaluate the physical device's performance directly.

    7. The Type of Ground Truth Used

    The ground truth was established through objective physical measurements and observations within a controlled laboratory setting (e.g., visual inspection for tears/damage, resolution testing, particulate analysis, microbial challenge results, chemical analysis).

    8. The Sample Size for the Training Set

    Not applicable. This is a non-AI/algorithm device. The concept of a "training set" does not apply to the development and testing of a mechanical injector system in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1