K Number
K980696
Manufacturer
Date Cleared
1998-09-11

(200 days)

Product Code
Regulation Number
886.4300
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Foam Tip™ Injector is a device used to fold and insert STAAR Surgical UV-ELASTIC™ silicone lenses for surgical placement in the human eye.

Device Description

STAAR Surgical Company's Foam Tip™ Injector is used to fold and insert STAAR Surgical UV-ELASTIC™ silicone intraocular lenses for surgical placement into the human eye. The new Foam Tip™ Injector utilizes the same design principles (i.e. a plunger pushing or threaded mechanism) to push the IOL through a cartridge for injection into the eye as the existing MicroSTAART™ injectors. The new injector system differs from the MicroSTAART™ injector in that it contains a disposable Foam Tip™ plunger insert which may be attached to the sterilized injector body prior to lens loading. This attachable Foam Tip™ plunger insert is encapsulated inside a holder that protects the sponge from damage. The holder acts to guide the plunger as it slides into the injector body and performs as a wrench to remove the plunger from the injector once the lens has been implanted. Once the sponge portion of the foam plunger insert is hydrated, it enables the surgeon to have greater control in ejecting the lens properly through the cartridge and into the eye. Like the existing MicroSTAAR™ injectors, the new Foam Tip™ Injector body is provided sterile and can be resterilized for multiple use. The new Foam Tip™ plunger insert is also provided sterile but is intended for single use only.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided text for the Foam Tip™ Injector System (K980696):

Acceptance Criteria and Device Performance for Foam Tip™ Injector System (K980696)

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Functional Testing)Reported Device Performance (Functional Testing)
No lens tearsNo evidence of lens tears
All post-ejected lenses meet STAAR's minimum resolution efficiency requirementsAll 40 functional test lens ejections passed minimum resolution testing
No haptic damageNo evidence of haptic damage
Gross particulate evaluationParticulates were noted in both the Test and Control samples. None appeared to be greater than 10 microns in size.

Additional Acceptance Criteria and Performance (Accelerated Aging Studies)

Acceptance Criteria (Accelerated Aging Studies)Reported Device Performance (Accelerated Aging Studies)
Acceptable microbial barrierResults of Microbial Challenge Dust Drum Test, Dye Penetration and Burst tests indicate both sample groups (baseline versus six month) provided an acceptable microbial barrier and kept product sterile within.
Maintained product sterilityResults of Microbial Challenge Dust Drum Test, Dye Penetration and Burst tests indicate both sample groups (baseline versus six month) kept product sterile within.
Acceptable seal strength (slight variations)Slight variations in seal strength were within an acceptable range and did not appear to have compromised the sterile barriers or the integrity of the product.
Material non-toxicityCytotoxicity tests showed that the material was non-toxic.
E-Beam sterilization as an acceptable methodFTIR demonstrated that E-Beam sterilization was an acceptable method of sterilization.
Bacterial endotoxins not above FDA limitsLAL testing was performed which indicated that bacterial endotoxins were not above FDA limits, thereby providing acceptable results.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 40 functional ejection tests were conducted for the functional testing.
  • Data Provenance: The text does not explicitly state the country of origin or whether the study was retrospective or prospective. It describes laboratory or in-house testing performed by or for STAAR Surgical Company.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the use of "experts" to establish ground truth in the traditional sense of clinical opinion or image interpretation. The ground truth for functional performance (lens tears, haptic damage, resolution efficiency, particulates) was established through direct observation and measurement in a controlled testing environment, likely by trained technicians or engineers.

4. Adjudication Method for the Test Set

Not applicable. The criteria for functional testing were objective and directly observable/measurable (e.g., presence/absence of tears, adherence to resolution requirements). There was no "adjudication" between multiple expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This document describes non-clinical functional and accelerated aging tests of the device itself, not a comparative effectiveness study involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a mechanical injector system, not an algorithm or AI. The tests performed are standalone in the sense that they evaluate the physical device's performance directly.

7. The Type of Ground Truth Used

The ground truth was established through objective physical measurements and observations within a controlled laboratory setting (e.g., visual inspection for tears/damage, resolution testing, particulate analysis, microbial challenge results, chemical analysis).

8. The Sample Size for the Training Set

Not applicable. This is a non-AI/algorithm device. The concept of a "training set" does not apply to the development and testing of a mechanical injector system in this context.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

Kg80696

PREMARKET NOTIFICATION [510(k)] SUMMARY SEP 1 1 1998 FOR FOAM TIP™ INJECTOR SYSTEM K980696 :

Trade name:Foam Tip™ Injector System
Common name:Foldable Intraocular Lens Injector System
Classification name:Intraocular Lens Guide per 21 CFR, Section 886.4300
Ophthalmic sponge per 21 CFR, Section 886.4790
Submitted by:STAAR Surgical Company1911 Walker AvenueMonrovia, CA 91016Phone: (626) 303 - 7902FAX: (626) 630-1423
Contact Person:Steven L. ZiémbaVice-President, Regulatory AffairsPhone: (626) 303 - 7902, ext. 2308
Date Summary Prepared:8/12/98
Legally Marketed PredicateDevice:MicroSTAAR™ Injector System (reusable anddisposable) manufactured by STAAR Surgical Company.
Intended Use of theDevice:The Foam Tip™ Injector is a device used to fold and insertSTAAR Surgical UV-ELASTIC™ silicone lenses for surgicalplacement in the human eye.
Device Description:STAAR Surgical Company's Foam Tip™ Injector is used to foldand insert STAAR Surgical UV-ELASTIC™ silicone intraocularlenses for surgical placement into the human eye. The new FoamTip™ Injector utilizes the same design principles (i.e. a plungerpushing or threaded mechanism) to push the IOL through acartridge for injection into the eye as the existing MicroSTAART™injectors. The new injector system differs from theMicroSTAART™ injector in that it contains a disposable FoamTip™ plunger insert which may be attached to the sterilizedinjector body prior to lens loading. This attachable Foam Tip™plunger insert is encapsulated inside a holder that protects thesponge from damage. The holder acts to guide the plunger as itslides into the injector body and performs as a wrench to removethe plunger from the injector once the lens has been implanted.Once the sponge portion of the foam plunger insert is hydrated, it

10111

.

{1}------------------------------------------------

K980694

enables the surgeon to have greater control in ejecting the lens properly through the cartridge and into the eye. Like the existing MicroSTAAR™ injectors, the new Foam Tip™ Injector body is provided sterile and can be resterilized for multiple use. The new Foam Tip™ plunger insert is also provided sterile but is intended for single use only.

Nonclinical Performance Testing:

Functional Testing

I unctional of 40 functional ejection tests were conducted on STAAR Surgical single-piece UV-ELASTIC™ silicone lenses. A matrix was used to define the injector/lens/cartridge combinations, with a minimum of ten lenses per injector/cartridge combination. Two types of injectors were tested with the Foam Tip™ plunger inserts, the plunger design and the threaded design. (The plunger type injector requires that the user push on the knob cap to move the plunger. The threaded plunger requires that the user turn the threaded cap clockwise or counterclockwise to move the plunger.)

Acceptance Criteria for the Lens Injection Test for Foam Tip™ Injectors included:

  • No lens tears (an indication that the lens was properly loaded)
  • All post-ejected lenses meet STAAR's minimum resolution efficiency requirements
  • No haptic damage
  • Gross particulate evaluation

All 40 functional test lens ejections passed minimum resolution testing. There was no evidence of haptic damage or lens tears. Particulates were noted in both the Test and Control samples. None appeared to be greater than 10 microns in size.

Accelerated Aging Studies

The following tests were performed :

  • Dust Drum Microbial Challenge
  • Dye Penetration tests:
    • Seal Strength -
    • Burst Test -
    • Cytotoxicity per USP 23 <87> -
    • -FTIR
    • Sponge degradation Analysis -
    • Package integrity -

The results of the accelerated aging tests, specifically the Microbial Challenge Dust Drum Test, Dye Penetration and Burst tests indicate that both sample groups (baseline versus six month) provided an acceptable microbial barrier and kept product sterile within. Slight variations in seal strength were within an acceptable range and did not appear to have compromised the sterile barriers or the integrity of the product.

11712

{2}------------------------------------------------

K980696

Cytotoxicity tests showed that the material was non-toxic and along with the FTIR demonstrated that E-Beam sterilization was an acceptable method of sterilization.

LAL testing was also performed which indicated that bacterial endotoxins were not above FDA limits, thereby providing acceptable results.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 1 1998

Mr. Steven Ziémba Vice-President, Regulatory Affairs STAAR Surgical Company 1911 Walker Avenue Monrovia, CA 91016

Re: K980696 Trade Name: Foam Tip™ Injector System Regulatory Class: I Product Code: MSS Dated: August 12, 1998 Received: August 14, 1998

Dear Mr. Ziémba:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Mr. Steven L. Ziémba

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Roerl lorentthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: K980696

Foam Tip™ Injector Device Name:

Indications For Use:

The Foam Tip™ Injector is a device used to fold and insert STAAR Surgical UV-The Fount Tip - Infoscee silicone lenses for surgical placement in the human eye.

There are no marketing claims for this device at this time.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

. .

Prescription Use ﺳ (Per 21 CFR 801.109) OR

Over-The-Counter Use __

(Optional Format 1-2-96)

Claudine d. Krawczyk for DRL
(Division Sign. Off.)

Division of Ophthalmic Devices
510(k) Number K980696

§ 886.4300 Intraocular lens guide.

(a)
Identification. An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.(b)
Classification. Class I (general controls). Except when used as folders or injectors for soft or foldable intraocular lenses, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9.