Search Results
Found 3 results
510(k) Data Aggregation
(116 days)
The Solana Surgical LLC, CrossCheck Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet, wrist, and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.
The Solana Surgical, LLC CrossCHECK Plating System consists of a variety of bone plates and screws to be used in fixation fractures, osteotomies, and fusions in the extremities of the ley and arm including the hand and foot. The various plate designs use either 3mm or 3.5mm screws. Screws include both locking and non-locking designs. Lag screws and cross-hole plates are included for compression. Plate designs include MTP, Lapidus, and 2-7 Hole Utility plates. All plates and screws are manufactured from ASTM F136 Ti-6A1-4V ELI alloy. The design of the Solana Surgical implant is similar to the predicate devices. No new materials or processes are used in the development of this implant.
The provided text describes Solana Surgical, LLC's CrossCheck Plating System, a device for stabilizing and fixing bone fractures, osteotomies, and fusions in the extremities. The 510(k) summary indicates that the device's substantial equivalence determination is based on a comparison to predicate devices, focusing on materials, design, and intended use.
Here's an analysis of the acceptance criteria and the study based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Criterion | Acceptance Metric | Reported Device Performance | Study that Proves Device Meets Acceptance Criteria |
|---|---|---|---|
| Material Properties | The plates and screws must be manufactured from ASTM F136 Ti-6A1-4V ELI alloy. | "All plates and screws are manufactured from ASTM F136 Ti-6A1-4V ELI alloy." | Device description in section 6 confirms the use of specified material. No specific testing report details are provided in this summary, but the statement confirms adherence to the material specification. |
| Mechanical Performance of Plates | Plates must meet the "Standard Specification and Test Method for Metallic Bone Plates" as per ASTM F382. | "The subject device met acceptance criteria..." when evaluated per ASTM F382. | Summary of Device Testing in section 1 states that the plates were evaluated per ASTM F382, and "The subject device met acceptance criteria and is substantially equivalent to the predicates." This implies the mechanical properties (e.g., bending strength, stiffness) were comparable to, or better than, the predicate devices and met the standard. |
| Mechanical Performance of Screws | Screws must meet the "Standard Specification and Test Methods for Metallic Medical Bone Screws" as per ASTM F543-07. | "The subject device met acceptance criteria..." when evaluated per ASTM F543-07. | Summary of Device Testing in section 1 states that the screws were evaluated per ASTM F543-07, and "The subject device met acceptance criteria and is substantially equivalent to the predicates." This implies the mechanical properties (e.g., torque strength, pull-out strength, bending properties) were comparable to, or better than, the predicate devices and met the standard. |
| Substantial Equivalence to Predicate Devices | The device should demonstrate comparable technological characteristics, principles of operation, mechanisms of action, size, materials (no new materials introduced), and intended use as the predicate devices. | "The design of the Solana Surgical implant is similar to the predicate devices. No new materials or processes are used in the development of this implant.""The technological characteristics, principles of operation and mechanisms of action are the same as the predicates.""The subject device met acceptance criteria and is substantially equivalent to the predicates." | Comparison to Predicate Device in section 1, along with the device description and summary of device testing, collectively state that the device is similar in design, materials, principles of operation, and met safety and performance criteria through testing against ASTM standards, using examples from predicate devices for comparison. |
2. Sample Size Used for the Test Set and Data Provenance
The provided 510(k) summary does not specify the sample size used for the mechanical testing (devices tested) or the provenance of any data (country of origin, retrospective/prospective). It only states "Examples from the predicate device systems were used for comparison." This suggests a comparative physical testing approach rather than a data-driven clinical study with a "test set" of patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The study described is primarily a mechanical device performance test against established ASTM standards and a comparison to predicate devices, not a study requiring expert consensus or clinical ground truth establishment in the traditional sense.
4. Adjudication Method for the Test Set
This information is not applicable and therefore not provided. The study described is a laboratory-based mechanical performance test, not a clinical study involving human assessment or expert adjudication of outcomes.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described. The 510(k) summary focuses on the mechanical and material equivalence of the device to existing predicate devices through laboratory testing. It does not involve human readers or AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone (algorithm only) performance study was not done. This device is a bone plate and screw system, a physical medical device, not a software algorithm or an AI-based system.
7. The Type of Ground Truth Used
The "ground truth" for this device's evaluation is primarily based on established engineering standards and material specifications, specifically:
- ASTM F382: Standard Specification and Test Method for Metallic Bone Plates.
- ASTM F543-07: Standard Specification and Test Methods for Metallic Medical Bone Screws.
- Material Specification: ASTM F136 Ti-6A1-4V ELI alloy for the manufacturing material.
The performance of predicate devices also serves as a benchmark for comparison during the testing. This is a technical, rather than a clinical or pathological, ground truth.
8. The Sample Size for the Training Set
This information is not applicable and therefore not provided. As this is a physical medical device and not an AI/software system, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable and therefore not provided. There is no training set for this type of device submission.
Ask a specific question about this device
(179 days)
The Solana Surgical LLC, Lesser Metatarsal Head Implant System is intended for use as a hemiarthroplasty implant for the metatarsophalangeal joint, for the treatment of degenerative and post-traumatic arthritis, hallux limitus, hallux valgus, hallux rigidus, and an unstable or painful metatarsophalangeal (MTP) joint. The device is intended for single use to be used with bone cement or press fit without bone cement.
The Solana Surgical Implant is a one-piece device made of Cobalt Chromium (with titanium plasma spray coated stem) intended to replace the articulating surface of the metatarsal bone at the lesser metatarsal phalangeal (MTP) joint. The implant is available in a range of sizes (3) to match the geometry of the lesser metatarsal phalangeal joint. Design features include an articulating surface and a stem which extends proximally in the intramedullary canal of the distal lesser metatarsal. The design of the Solana Surgical implant is similar to the predicate devices. No new materials or processes are used in the development of this implant.
This looks like a 510(k) premarket notification for a medical device called the "Lesser Metatarsal Head Implant System." This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study data with acceptance criteria for device performance as would be expected for a novel device or a PMA.
Therefore, the information requested in your prompt related to acceptance criteria, specific performance metrics, sample sizes for test/training sets, expert adjudication, MRMC studies, or standalone performance studies is not typically found in a 510(k) summary like this one.
Here's a breakdown of why and what information is available:
Response to your Request:
1. A table of acceptance criteria and the reported device performance
- Not Applicable. This 510(k) summary does not include specific acceptance criteria for a device performance study in the way a clinical trial or performance study for an AI/diagnostic device would. The "performance" being evaluated is primarily the mechanical strength and material properties in comparison to predicate devices, not clinical efficacy or diagnostic accuracy.
- The summary states: "An engineering analysis was performed to evaluate the mechanical strength of subject and predicate device implant stems." and "The titanium plasma spray coating process as applied to the Solana Surgical device has been thoroughly evaluated for morphological and strength characteristics." These are the "reported device performance" aspects, but without specific quantitative acceptance criteria disclosed in this summary.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable / Not Provided. No clinical test set data from actual patients is described or used in this 510(k) summary to prove performance against specific clinical acceptance criteria. The evaluation is focused on engineering analysis and comparison to predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. No ground truth establishment by experts is described, as there are no clinical test sets or diagnostic performance claims requiring expert interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No test set requiring expert adjudication is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a mechanical implant, not an AI software or diagnostic tool. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a mechanical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. No clinical ground truth is established or used for performance evaluation in this 510(k) summary. The "ground truth" for the engineering evaluation would be established engineering principles, material science standards, and comparison with the known properties of the predicate devices.
8. The sample size for the training set
- Not Applicable. This is a mechanical implant; there is no "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
- Not Applicable. No training set exists for this type of device submission.
Summary of Available Information from the Provided Text:
- Device Name: Lesser Metatarsal Head Implant System
- Purpose: Hemiarthroplasty implant for the metatarsophalangeal joint.
- Materials: Cobalt Chromium (with titanium plasma spray coated stem).
- Design: One-piece device with an articulating surface and a stem for intramedullary canal insertion.
- Comparison Basis: Substantially equivalent to predicate devices (OsteoMed: K073065, Vilex: K102401) based on:
- Similar materials
- Similar indications for use
- Comparable sizing
- Similar intended use (single use, surgical implantation longer than 30 days)
- Engineering analysis of mechanical strength (comparison of subject and predicate device implant stems).
- Evaluation of the titanium plasma spray coating process for morphological and strength characteristics.
This document serves to demonstrate that the new device is as safe and effective as a legally marketed predicate device, primarily through engineering comparisons and material analysis, rather than extensive clinical efficacy studies with predefined acceptance criteria.
Ask a specific question about this device
(191 days)
The Solana Surgical, LLC FuseForce Implant System is intended to provide fixation for fractures, fusions or osteotomies of the bones of the hand and foot such as: LisFranc arthrodesis, Akin Osteotomy, Scarf and Chevron osteotomies
The Solana Surgical FuseForce Implant is a one-piece device made of Nickel Titanium Alloy intended to be implanted in the bones of the hand or foot. The implant is available in a range of sizes (12) ranging from 8 mm X 8 mm to 25 mm X 22 mm including both straight top and step top configurations. The design of the Solana Surgical implant is similar to the predicate device. No new materials or processes are used in the production of this implant. The device is provided in a sterile package which includes associated single use (disposable) instruments.
Acceptance Criteria and Study Details for Solana Surgical FuseForce Implant System
This document outlines the acceptance criteria for the Solana Surgical FuseForce Implant System and details the studies performed to demonstrate its compliance.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for the FuseForce Implant System were based on its comparison to a legally marketed predicate device, the Memometal Memory Staples (K070031), and the Memometal EasyClip stable. The device was required to perform at least equivalently to the predicate in mechanical performance and corrosion resistance.
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Mechanical Performance | Exceed or be equivalent to the predicate device (Memometal Memory Staples) in: | Exceeded or was equivalent to the predicate device in all evaluated cases. |
| 1. Pullout fixation strength (ASTM F564) | ||
| 2. Static bending (ASTM F564) | ||
| 3. Fatigue bending (ASTM F564) | ||
| Corrosion Resistance | Demonstrate a breakdown potential (Eb) higher than that of the predicate device (Memometal EasyClip). | Demonstrated a breakdown potential (Eb) higher than that of the predicate. |
2. Sample Sizes and Data Provenance
- Test Set Sample Size: For both mechanical and corrosion testing, six subject devices and six predicate devices were evaluated for each test.
- Mechanical Testing: 8mm x 8mm nitinol staples (worst-case for pullout, static, and fatigue bending).
- Corrosion Testing: 25mm x 22mm nitinol staples (largest surface area, worst-case for corrosion susceptibility).
- Data Provenance: The data was generated through bench testing performed by Solana Surgical, LLC. The specific country of origin for the testing itself is not explicitly stated, but the company is based in Memphis, TN, USA. The data is prospective as it involves direct testing of the manufactured devices.
3. Number, Qualifications, and Adjudication of Experts for Ground Truth
- This device is a mechanical implant, and the ground truth for its performance is established through objective, standardized mechanical and material testing protocols (ASTM F564 and ASTM F2129).
- Therefore, the concept of "experts establishing ground truth" in the cognitive or interpretive sense, as seen in imaging studies, is not applicable here. The "ground truth" is defined by the objective measurements obtained from the testing.
- No human experts were used to establish ground truth in the context of interpretation or consensus for these purely mechanical and material property tests.
4. Adjudication Method
- Not applicable as the ground truth is established through objective, quantitative measurements from mechanical and material testing, not through expert interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not performed. This type of study is typically relevant for diagnostic imaging devices where human interpretation plays a role and the AI's impact on reader performance is being evaluated. The FuseForce Implant System is a mechanical implant, not a diagnostic AI.
6. Standalone (Algorithm Only) Performance Study
- Not applicable. The FuseForce Implant System is a physical medical device, specifically a bone fixation staple. It does not employ an AI algorithm, and therefore, a "standalone algorithm performance" study is irrelevant.
7. Type of Ground Truth Used
- The ground truth used was objective performance data derived from standardized mechanical and material property testing against established ASTM (American Society for Testing and Materials) standards:
- ASTM F564 (for mechanical properties: pullout, static bending, fatigue bending)
- ASTM F2129 (for corrosion resistance)
8. Sample Size for the Training Set
- Not applicable. This device is a physical implant, not an AI algorithm that requires a training set. The performance is based on the material properties and design of the manufactured product.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1