(191 days)
The Solana Surgical, LLC FuseForce Implant System is intended to provide fixation for fractures, fusions or osteotomies of the bones of the hand and foot such as: LisFranc arthrodesis, Akin Osteotomy, Scarf and Chevron osteotomies
The Solana Surgical FuseForce Implant is a one-piece device made of Nickel Titanium Alloy intended to be implanted in the bones of the hand or foot. The implant is available in a range of sizes (12) ranging from 8 mm X 8 mm to 25 mm X 22 mm including both straight top and step top configurations. The design of the Solana Surgical implant is similar to the predicate device. No new materials or processes are used in the production of this implant. The device is provided in a sterile package which includes associated single use (disposable) instruments.
Acceptance Criteria and Study Details for Solana Surgical FuseForce Implant System
This document outlines the acceptance criteria for the Solana Surgical FuseForce Implant System and details the studies performed to demonstrate its compliance.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for the FuseForce Implant System were based on its comparison to a legally marketed predicate device, the Memometal Memory Staples (K070031), and the Memometal EasyClip stable. The device was required to perform at least equivalently to the predicate in mechanical performance and corrosion resistance.
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Mechanical Performance | Exceed or be equivalent to the predicate device (Memometal Memory Staples) in: | Exceeded or was equivalent to the predicate device in all evaluated cases. |
| 1. Pullout fixation strength (ASTM F564) | ||
| 2. Static bending (ASTM F564) | ||
| 3. Fatigue bending (ASTM F564) | ||
| Corrosion Resistance | Demonstrate a breakdown potential (Eb) higher than that of the predicate device (Memometal EasyClip). | Demonstrated a breakdown potential (Eb) higher than that of the predicate. |
2. Sample Sizes and Data Provenance
- Test Set Sample Size: For both mechanical and corrosion testing, six subject devices and six predicate devices were evaluated for each test.
- Mechanical Testing: 8mm x 8mm nitinol staples (worst-case for pullout, static, and fatigue bending).
- Corrosion Testing: 25mm x 22mm nitinol staples (largest surface area, worst-case for corrosion susceptibility).
- Data Provenance: The data was generated through bench testing performed by Solana Surgical, LLC. The specific country of origin for the testing itself is not explicitly stated, but the company is based in Memphis, TN, USA. The data is prospective as it involves direct testing of the manufactured devices.
3. Number, Qualifications, and Adjudication of Experts for Ground Truth
- This device is a mechanical implant, and the ground truth for its performance is established through objective, standardized mechanical and material testing protocols (ASTM F564 and ASTM F2129).
- Therefore, the concept of "experts establishing ground truth" in the cognitive or interpretive sense, as seen in imaging studies, is not applicable here. The "ground truth" is defined by the objective measurements obtained from the testing.
- No human experts were used to establish ground truth in the context of interpretation or consensus for these purely mechanical and material property tests.
4. Adjudication Method
- Not applicable as the ground truth is established through objective, quantitative measurements from mechanical and material testing, not through expert interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not performed. This type of study is typically relevant for diagnostic imaging devices where human interpretation plays a role and the AI's impact on reader performance is being evaluated. The FuseForce Implant System is a mechanical implant, not a diagnostic AI.
6. Standalone (Algorithm Only) Performance Study
- Not applicable. The FuseForce Implant System is a physical medical device, specifically a bone fixation staple. It does not employ an AI algorithm, and therefore, a "standalone algorithm performance" study is irrelevant.
7. Type of Ground Truth Used
- The ground truth used was objective performance data derived from standardized mechanical and material property testing against established ASTM (American Society for Testing and Materials) standards:
- ASTM F564 (for mechanical properties: pullout, static bending, fatigue bending)
- ASTM F2129 (for corrosion resistance)
8. Sample Size for the Training Set
- Not applicable. This device is a physical implant, not an AI algorithm that requires a training set. The performance is based on the material properties and design of the manufactured product.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
SIO(K) Summary
5. K124045 Summary (Updated July 9, 2013)
Summary of 510 (k) safety and effectiveness information upon which the substantial equivalence determination is based:
| Prepared: | December 28, 2012 | Revised July 9, 2013 |
|---|---|---|
| Applicant: | Solana Surgical, LLC | Contact: Joe Clift |
| 6363 Poplar Ave, Suite 312 | (901) 818-1860 | |
| Memphis, TN 38119 | ||
| Common Name: | Staple, Fixation, Bone | |
| Device Trade Name: | FuseForce Implant System | |
| Device Classification Name: | Single/Multiple component metallic bone | |
| fixation appliances and accessories | ||
| Device Classification: | Class II | |
| Reviewing Panel: | Orthopedic | |
| Regulation Number: | 21 CFR 888.3030 | |
| Product Code: | JDR | |
| Predicate Device: | K070031 Memometal Memory Staples |
Device Description:
The Solana Surgical FuseForce Implant is a one-piece device made of Nickel Titanium Alloy intended to be implanted in the bones of the hand or foot. The implant is available in a range of sizes (12) ranging from 8 mm X 8 mm to 25 mm X 22 mm including both straight top and step top configurations. The design of the Solana Surgical implant is similar to the predicate device. No new materials or processes are used in the production of this implant. The device is provided in a sterile package which includes associated single use (disposable) instruments.
Indications for Use:
The Solana Surgical, LLC FuseForce Implant System is intended to provide fixation for fractures, fusions or osteotomies of the bones of the hand and foot such as: LisFranc arthrodesis, Akin Osteotomy, Scarf and Chevron osteotomies
Summary Comparison to Predicate Devices:
Similarities of the Solana Surgical device to its predicates include these devices being intended for single use only and for surgical implantation longer than 30 days. The Solana Surgical device and the predicate device have similar design characteristics and intended use. The stated indications for use add specificity to that of the predicate by listing examples of procedures within the broader context of the predicate's statement. The indications for use of the subject device are, therefore, not new but merely better defined. Comparative bench testing per ASTM F564 indicates that the subject staple performance exceeds that of the predicate device in all cases evaluated. Original mechanical testing was performed at room temperature. Additional mechanical testing at body temperature was performed to characterize the performance of the subject device and a legally marketed predicate, the Memometal EasyClip stable.
JUL 1 0 2013
{1}------------------------------------------------
The devices tested were 8mm x 8mm nitinol staples which represent the worstcase for 1.) pullout fixation strength, 2.) static bending and 3.) fatigue bending. For both devices, production components were used. Six subject devices and six predicate devices were evaluated in each test. See Attachment K "VR-029 - "Staple Mechanical Testing Report -- FuseForce Implant System".
The Solana subject device exceeded or was equivalent to the predicate in mechanical performance and therefore demonstrates acceptable mechanical characteristics.
Comparative testing characterized the corrosion resistance of the subject device and a legally marketed predicate, the Memometal EasyClip staple. The testing was performed per ASTM F 2129. The devices tested were 25mm x 22mm nitinol staples which represent the largest surface area and, therefore, worst case for corrosion susceptibility. For both devices, production components were used and a simulated implantation in bone substitute was performed. Six subject devices and six predicate devices were tested. See Attachment J "VR-026 - Staple Implant Corrosion Validation".
The subject device demonstrates a breakdown potential (Eb) that is higher than that of the predicate suggesting that the subject device exhibits an acceptable level of corrosion resistance.
The subject device is substantially equivalent to the predicates devices and should not introduce new concerns in terms of safety or effectiveness,
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID 20993-0002
July 10, 2013
Solana Surgical, LLC % Mr. Joc Clift Senior Vice President Technical Operations and Compliance 6363 Poplar Avenue, Suite 312 Memphis, Tennessee 38119
Re: K124045
Trade/Device Name: FuseForce Implant System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: May 29, 2013 Reccived: June 03, 2013
Dear Mr. Clift:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Joe Clift
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin Do ith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K124045 Device Name: FuseForce Implant System Indications for Use:
The Solana Surgical, LLC FuseForce Implant System is intended to provide fixation for fractures, fusions or osteotomies of the bones of the hand and foot such as: LisFranc arthrodesis, Akin Osteotomy, Scarf and Chevron osteotomies
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Page 7 of 32
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.