Search Results
Found 31 results
510(k) Data Aggregation
(363 days)
SGMP CO., LTD.
The Non-Sterile, Powder-Free Yellow Latex Examination Glove With Tutti Flavour and With Protein Labeling Claim of
Non-Sterile, Powder-Free Yellow Latex Examination Gloves With Tutti Frutti Flavour and with Protein Labeling Claim of
I am sorry, but based on the provided text, there is no information about acceptance criteria or a study that proves a device meets those criteria. The document is an FDA 510(k) clearance letter for "Non-Sterile, Powder-Free Yellow Latex Examination Gloves with Tutti Frutti Flavour and with Protein Labeling Claim of
Ask a specific question about this device
(142 days)
SGMP CO., LTD.
A disposable medical glove to be worn on the hand of the healthcare and similar personnel to prevent contamination between healthcare personnel and patient.
Non-sterile , Powder-free Blue Nitrile Examination Gloves , Tested for use with Chemotherapy Drugs.
The provided documents describe the performance of "Non-sterile, Powder-free Blue Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs" (K101822) against various acceptance criteria relevant to medical examination gloves and specifically for their resistance to chemotherapy drugs.
Here’s a breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the device are largely based on established ASTM (American Society for Testing and Materials) standards and a comparison to a legally marketed predicate device (K082957). The key performance indicator for chemotherapy drug resistance is the "Breakthrough Detection Time" as per ASTM D6978-05.
Acceptance Criteria and Reported Device Performance for K101822
Characteristic and Parameter | Acceptance Criteria | Reported Device Performance | Meets Criteria? |
---|---|---|---|
General Glove Characteristics | |||
Device Class | Class I (as per predicate) | Class I | Yes |
Product Code | LZA (as per predicate) | LZA | Yes |
Glove Color | Blue (as per predicate) | Blue | Yes |
Dimensions | Meets ASTM D6319-00a-05 | Meets ASTM D6319-00a-05 | Yes |
Physical Properties | Meets ASTM D6319-00a-05 | Meets ASTM D6319-00a-05 | Yes |
Freedom From Pinholes | Meets ASTM D6319-00a-05 (and/or ISO 2859-1, ASTM D5151-06) | Meets ASTM D6319-00a-05 | Yes |
Powder-free | Meets ASTM D6124-06 | Meets ASTM D6124-06 | Yes |
Biocompatibility | |||
Primary Skin Irritation | Passes Primary Skin Irritation in Rabbits (as per predicate) | Passes Primary Skin Irritation in Rabbits | Yes |
Sensitization | Passes Guinea Pig Maximization (similar to predicate's Guinea Pig Sensitization) | Passes Guinea Pig Maximization | Yes |
Chemotherapy Drugs Permeation (ASTM D6978-05) | |||
Cisplatin | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Cyclophosphamide (Cytoxan) | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Doxorubicin Hydrochloride | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Etoposide (Toposar) | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Flurouracil | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Paclitaxel (Taxol) | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Vincristine Sulfate | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Dacarbazine (DTIC) | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Methotrexate | > 240 minutes (as per predicate) | > 240 minutes | Yes |
Carmustine (BCNU) | Not specified as a pass/fail threshold, but noted as having extremely low permeation time in the predicate | 0.49 minutes | No (Note: Not approved for use with this drug) |
Thiotepa | Not specified as a pass/fail threshold, but noted as having extremely low permeation time in the predicate | 2.61 minutes | No (Note: Not approved for use with this drug) |
Labeling | Meets FDA requirement | Meets FDA requirement | Yes |
Summary of Study Proving Acceptance:
The device K101822 was shown to meet its acceptance criteria through various tests described in the "Substantial Equivalence Discussion" (Document 2, page 2 of 2) and the "CONCLUSION" (Document 4, page 1 of 1). The primary study for chemotherapy drug resistance was conducted according to ASTM D6978-05. For other characteristics like dimensions, physical properties, freedom from pinholes, and powder-free status, the device was tested against and found to meet ASTM D6319-00a-05, ISO 2859-1 (for water leak test), ASTM D5151-06 (for detection of holes), and ASTM D6124-06 respectively. Biocompatibility tests included Primary Skin Irritation in Rabbits and Guinea Pig Maximization tests.
The device's performance was directly compared to a legally marketed predicate device, Siam Sempermed Corp. Ltd.'s Non-sterile, Powder-free Nitrile Examination Glove (K082957), showing substantial equivalence in all tested parameters, except for highlighting the short permeation times for Carmustine (BCNU) and Thiotepa, for which the gloves are explicitly not approved.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: The specific sample sizes for each test (e.g., number of gloves tested for chemotherapy permeation, number of rabbits/guinea pigs for biocompatibility, or specific counts for physical properties) are not explicitly stated in the provided documents. The documents only state that the device "Meets ASTM D6319-00a-05," "Meets ASTM D6124-06", etc., which implies that the testing was performed according to the methodologies and sample size requirements specified in those respective ASTM/ISO standards.
- Data Provenance: The data is presented as part of a 510(k) Premarket Notification submission to the FDA by SGMP Company Limited, located in Thailand (Document 0, page 1 of 2). This indicates the testing was conducted by or on behalf of the manufacturer, and generally, such data would be generated in a controlled laboratory environment. The documents do not specify the country of origin of the specific lab that conducted these tests, but the manufacturer is based in Thailand. The studies were retrospective in the sense that they were conducted to characterize the finished medical device for regulatory submission, rather than as part of a clinical trial.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to this type of device submission. The "ground truth" for the performance of these examination gloves is established through standardized laboratory tests (e.g., ASTM standards) measuring physical and chemical properties, not through expert human interpretation or clinical ground truth. Therefore, there are no "experts" in the sense of medical professionals establishing ground truth for diagnostic accuracy, nor are their qualifications relevant here.
4. Adjudication method for the test set
This question is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human judgment (e.g., for image interpretation) to resolve disagreements among experts. The performance of these gloves is determined by objective, quantitative laboratory measurements according to published standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. An MRMC study is relevant for diagnostic devices (e.g., AI algorithms for image interpretation) where human readers evaluate cases with and without AI assistance. This submission describes an examination glove, which is a physical barrier device, and does not involve human readers or AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not applicable. This describes a physical medical device (examination gloves), not an algorithm or AI system.
7. The type of ground truth used
The ground truth used for performance validation is based on objective, quantitative measurements obtained through standardized laboratory testing protocols, primarily from ASTM standards (D6319-00a-05, D6124-06, D6978-05) and ISO 2859-1. For biocompatibility, the ground truth is established by the observed biological responses in animal models (rabbits and guinea pigs) according to standardized tests for skin irritation and sensitization. The device's performance is then compared against the requirements specified in these standards and against the performance of a legally marketed predicate device.
8. The sample size for the training set
This question is not applicable. This is a physical medical device, not an AI model or algorithm that requires a "training set" of data.
9. How the ground truth for the training set was established
This question is not applicable as there is no training set for this type of device.
Ask a specific question about this device
(89 days)
SGMP CO., LTD.
The Non-sterile Powder Free Black Nitrile Examination Gloves with Cherry Flavor, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Doves I Nitrile examination gloves 80LZA, powder free and meeting all the requirements of ASTM D6319-00a (2005) Standard Specification for Nitrile Examination Gloves for Medical Application.
This document describes the acceptance criteria and the study proving the device meets those criteria for Non-Sterile, Powder Free Black Nitrile Examination Gloves with Cherry Flavor.
It's important to note that this is a premarket notification (510(k)) for a Class I medical device (patient examination glove). The "study" here refers to demonstrating compliance with established performance standards and biocompatibility testing, rather than a clinical trial with human subjects in the way one might think of for more complex medical devices or pharmaceuticals.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Standard/Test | Reported Device Performance (as implied by "meets" or "equivalent") |
---|---|
ASTM D6319-00a(2005) | Met: Standard Specification for Nitrile Gloves. |
ISO 2859-1 (Water Leak Test) | Met: Standard for Water Leak Test. |
ASTM D5151-06 (Hole Detection) | Met: Test Method for Detection of Holes in Medical Gloves (alternative to ISO 2859-1). |
ASTM D6124-06 | Met: Standard Test Method for Residual Powder on Medical Gloves. |
Biocompatibility Testing | Met: Testing performed on White Rabbits and Guinea Pigs. |
Labeling | Met: FDA Specifications. |
Equivalence to K072400 | Met: Except for the cherry flavor, this glove is equivalent to the predicate device K072400 (POWDER FREE BLACK NITRILE EXAM GLOVES). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test (e.g., number of gloves tested for water leak, number of residual powder tests). However, for standards like ISO 2859-1, specific sampling plans (sample sizes and acceptable quality limits - AQLs) are defined within the standard itself for batch testing.
The data provenance is from Thailand (SGMP Company Limited is in Thailand). The data is retrospective in the sense that the testing was conducted prior to submission for regulatory clearance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This type of submission for examination gloves does not typically involve "experts" establishing a "ground truth" in the clinical sense (e.g., radiologists interpreting images). Instead, the "ground truth" is adherence to the performance specifications defined by the ASTM and ISO standards themselves. The "experts" involved would be the qualified laboratory personnel performing the standardized tests according to the written protocol of each standard. Their qualifications would be in laboratory practices, quality control, and the specific test methodologies. The number of such personnel is not specified.
4. Adjudication Method
Not applicable for this type of device and testing. Test results against established standards do not typically involve adjudication among multiple human readers. Outcomes are binary: pass/fail the standard's defined limits.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This is not the type of device where MRMC studies are performed. MRMC studies are typically for diagnostic imaging devices where human interpretation is a key component.
6. Standalone Performance Study
Yes, in a sense. The compliance with ASTM and ISO standards demonstrates the standalone performance of the device (the glove itself, without a human "in-the-loop" in the way an AI diagnostic tool would have). The biocompatibility testing also evaluates the standalone biological performance of the glove material.
7. Type of Ground Truth Used
The ground truth used is primarily performance specifications defined by internationally recognized standards (ASTM and ISO), along with biocompatibility testing results.
- ASTM D6319-00a(2005): Defines physical properties (e.g., tensile strength, elongation, force at break) for nitrile examination gloves.
- ISO 2859-1 / ASTM D5151-06: Specifies acceptable levels of pinholes in gloves.
- ASTM D6124-06: Specifies limits for residual powder.
- Biocompatibility Testing: Involves a series of tests (often based on ISO 10993 standards, though not explicitly stated here for gloves) to assess the biological response to the device, conducted on animals (White Rabbits and Guinea Pigs).
8. Sample Size for the Training Set
Not applicable. This device is not an AI/ML algorithm requiring a "training set." The testing performed is to demonstrate compliance of manufactured batches with established performance standards.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(213 days)
SGMP CO., LTD.
The Non-Sterile Powdered Pink Latex Examination Gloves with Cherry Flavour and with Protein Labeling claims (200 Micrograms or less) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-05 Standard Specification for Latex Examination Gloves for Medical Application.
The provided text describes the acceptance criteria and the study conducted for "Non-Sterile, Powdered Pink Latex Examination Gloves with Cherry Flavour and with Protein Labeling Claims (200 Micrograms or Less)."
Here's the breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Metric (ASTM D3578-05) | Reported Device Performance (SGMP) | Pass/Fail |
---|---|---|---|
Dimension | |||
X-Small (Width) | 70 mm +/- 10 mm | 70 - 80 mm | Pass |
Small (Width) | 80 mm +/- 10 mm | 80 - 85 mm | Pass |
Medium (Width) | 95 mm +/- 10 mm | 90 - 97 mm | Pass |
Large (Width) | 111 mm +/- 10 mm | 105 - 111 mm | Pass |
Length (all sizes) | 230 mm minimum | 240 mm | Pass |
Thickness - Finger | 0.08 mm min | 0.13 mm min | Pass |
Thickness - Palm | 0.08 mm min | 0.10 mm min | Pass |
Physical Properties | |||
Tensile Strength (Mpa) | |||
Before Aging | 18.0 Mpa | X-Small: 23.5 Mpa | |
Small: 28.8 Mpa | |||
Medium: 22.7 Mpa | |||
Large: 25.5 Mpa | Pass | ||
After Aging | 14.0 Mpa | X-Small: 21.8 Mpa | |
Small: 25.5 Mpa | |||
Medium: 22.3 Mpa | |||
Large: 23.8 Mpa | Pass | ||
Ultimate Elongation (%) | |||
Before Aging | 650 % | X-Small: 780 % | |
Small: 920 % | |||
Medium: 835 % | |||
Large: 810 % | Pass | ||
After Aging | 500 % | X-Small: 740 % | |
Small: 850 % | |||
Medium: 780 % | |||
Large: 830 % | Pass | ||
Water Tight Test | 2.5% AQL (max. 25 leaky gloves per 1000) | UN-AGED: 2-2-1-2 leaky gloves (total 7/500) | |
AGED: 0-3-2-1 leaky gloves (total 6/500) | Pass | ||
Total Residual Powder Content | 10 mg/dm² max (ASTM D 6124-06) | Range: 5.2 - 8.5 mg/dm² | |
Mean: 7.5 mg/dm² | Pass | ||
Residual Protein Level |
Ask a specific question about this device
(213 days)
SGMP CO., LTD.
The Non-sterile Powder Free Black Nitrile Examination Gloves , is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I Nitrile examination gloves 80LZA, powder free and meeting all the requirements of ASTM D6319-00aE3 Standard Specification for Nitrile Examination Gloves for Medical Application.
The provided document describes the safety and effectiveness summary for Non-Sterile, Powder Free Black Nitrile Examination Gloves, which are classified as a Class I medical device. The document focuses on demonstrating substantial equivalence to a predicate device (K000868) by meeting established industry standards for examination gloves.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria / Standard | Required Performance | Reported Device Performance | Study Type/Method |
---|---|---|---|
ASTM D6319-00aE3, Standard Specification For Nitrile Gloves. | Meeting all requirements of ASTM D6319-00aE3 | Meets the requirements of ASTM D6319-00aE3 | Implied through compliance statement. Specific tests include: |
- Tensile Strength
- Elongation
- Dimensions (Length, Palm Width, Finger Thickness) |
| 1SO 2859-1, Standard for Water Leak Test and/or ASTM D5151-06, Standard Test Method for Detection of Holes in Medical Gloves. | Detection of holes in medical gloves | Meets the requirements of ISO 2859-1 and/or ASTM D5151-06 (Water Leak Test) | Water Leak Test (specifics of sample size/method not detailed beyond standard reference) |
| ASTM D6124-06, Standard Test Method for Residual Powder on Medical Gloves. | Measurement of residual powder | Meets the requirements of ASTM D6124-06 (Residual Powder Test) | Residual Powder Test (specifics of sample size/method not detailed beyond standard reference) |
| Biocompatibility Testing | Biocompatibility on White rabbits and Guinea pigs (absence of irritation/sensitization) | Meets Biocompatibility Testing on White rabbits and Guinea Pigs | Biocompatibility testing (specifics of tests/sample size not detailed beyond animal models) |
| Labeling | Meets FDA Specifications | Meets FDA Specifications | Labeling review against FDA guidelines |
| Equivalence to Predicate Device (K000868) | Substantially equivalent to K000868 (except for color) | Substantially equivalent to K000868 (except for color) | Comparative review against predicate device specifications |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each test (e.g., water leak, residual powder, tensile strength). It refers to compliance with the standards, implying that the sample sizes and methods prescribed within those standards were followed.
- Data Provenance: The manufacturing company is based in Thailand (SGMP Company Limited, 181 Moo 6, Tambol Kampacngpotch, Rattaphum, Songkhla 90180, Thailand). It is likely that the testing data was generated in Thailand or by laboratories compliant with international standards, but the exact country of origin for the data is not explicitly stated. The studies are retrospective in the sense that they were conducted on the manufactured product to demonstrate compliance with existing standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable in this context. The "ground truth" for medical gloves is established by meeting predefined physical and chemical performance standards (e.g., ASTM, ISO standards) rather than by expert clinical consensus or interpretation of diagnostic images. The standards themselves define the acceptable parameters for mechanical properties, barrier integrity, and biocompatibility.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication typically refers to resolving discrepancies in expert opinions, which is not relevant for testing physical properties of a device against objective standards. The outcome of the tests (e.g., does it leak or not, what is the tensile strength) is typically a quantifiable measurement or a pass/fail determination based on predefined thresholds in the standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This is not applicable. The device is a physical medical glove, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI assistance metrics are irrelevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable. The device is a physical medical glove, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance of these gloves is defined by:
- Performance Standards: The specifications outlined in ASTM D6319-00aE3, ISO 2859-1, ASTM D5151-06, and ASTM D6124-06. These standards define the acceptable physical properties (e.g., tensile strength, elongation), barrier integrity (water leak test), and chemical characteristics (residual powder).
- Biocompatibility Standards: Established protocols for testing biological reactions on animal models (White rabbits and Guinea pigs).
8. The Sample Size for the Training Set
This is not applicable. This is a physical medical device, not a machine learning model. There is no "training set." The manufacturing process is subject to quality control, and batches of gloves are sampled for testing against the standards.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as point 8.
Ask a specific question about this device
(207 days)
SGMP CO., LTD.
The Non-Sterile, Powder Free Pink Latex Examination Gloves with Cherry Flavour and with protein labeling claims (50 Micrograms or less) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-05 Standard Specification for Latex Examination Gloves for Medical Application.
This document describes the acceptance criteria and performance data for "Non-Sterile, Powder Free Pink Latex Examination Gloves with Cherry Flavour and with Protein Labeling Claims (50 Micrograms or Less)" submitted by SGMP Company Limited.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Acceptance Criteria (ASTM D3578-05 Standard / FDA Requirement) | Reported Device Performance (SGMP) |
---|---|---|
Dimension (Width) | X-Small: 70 mm +/- 10 mm | 70 – 80 mm |
Small: 80 mm +/- 10 mm | 80 – 85 mm | |
Medium: 95 mm +/- 10 mm | 90 – 97 mm | |
Large: 111 mm +/- 10 mm | 105 - 111 mm | |
Dimension (Length) | 230 mm minimum (for all sizes) | 240 mm |
Thickness - Finger | 0.08 mm min | 0.13 mm min |
Thickness - Palm | 0.08 mm min | 0.10 mm min |
Tensile Strength (Before Aging) | 18.0 Mpa | X-Small: 25.1 Mpa |
Small: 22.8 Mpa | ||
Medium: 28.0 Mpa | ||
Large: 26.3 Mpa | ||
Tensile Strength (After Aging) | 14.0 Mpa | X-Small: 24.5 Mpa |
Small: 23.5 Mpa | ||
Medium: 26.0 Mpa | ||
Large: 25.2 Mpa | ||
Ultimate Elongation (Before Aging) | 650 % | X-Small: 830 % |
Small: 870 % | ||
Medium: 850 % | ||
Large: 855 % | ||
Ultimate Elongation (After Aging) | 500 % | X-Small: 790 % |
Small: 805 % | ||
Medium: 780 % | ||
Large: 770 % | ||
Water Tight Test (AQL) | 2.5% AQL (ASTM D3578-05 requirements) | Un-aged (Batch #7023): |
X-Small: 0 leaked (0%) | ||
Small: 2 leaked (1.6%) | ||
Medium: 1 leaked (0.8%) | ||
Large: 1 leaked (0.8%) | ||
Aged (Batch #7023): | ||
X-Small: 2 leaked (1.6%) | ||
Small: 0 leaked (0%) | ||
Medium: 1 leaked (0.8%) | ||
Large: 0 leaked (0%) | ||
All reported values are ≤ 2.5% | ||
Biocompatibility | Passed tests for examination gloves (as per APPENDIX K) | Passed |
Residual Powder Content | 10 mg/glove max (ASTM D 6124-06) | Ranger: 0.8-1.7 mg/glove |
Mean: 1.4 mg/glove | ||
Presence of Cornstarch | Negative | Negative |
Residual Protein Level |
Ask a specific question about this device
(101 days)
SGMP CO., LTD.
This Non-Sterile, Latex Powder Free Patient Examination Gloves (Multicolored with Xylitol and Calcium Carbonate, with Protein Labeling claims (50 Micrograms or less) is a disposable device intended for medical purposes, that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-01aE2 Standard Specification for Latex Examination Gloves for Medical Application.
Acceptance Criteria and Device Performance for Powder-Free Latex Examination Gloves
This report summarizes the acceptance criteria and device performance for the "POWDER-FREE LATEX EXAMINATION GLOVES (MULTICOLORED WITH XYLITOL AND CALCIUM CARBONATE)" developed by SGMP Company Limited. The information is extracted from the provided 510(k) summary (K07/060).
1. Table of Acceptance Criteria and Reported Device Performance
The device's performance was evaluated against the ASTM D3578-01aE2 and ASTM D3578-00aE2 standards for latex examination gloves, as well as FDA requirements for powder-free gloves and protein labeling.
Acceptance Criteria Category | Specific Test / Requirement | Acceptance Criteria (Standard / FDA) | Reported Device Performance (SGMP) |
---|---|---|---|
Dimension | X-Small Width | 70 mm +/- 10 mm | 70 - 80 mm |
Small Width | 80 mm +/- 10 mm | 80 - 85 mm | |
Medium Width | 95 mm +/- 10 mm | 90 - 97 mm | |
Large Width | 111 mm +/- 10 mm | 105 - 111 mm | |
Length (all sizes) | 230 mm minimum | 241 mm | |
Thickness - Finger | 0.08 mm min | 0.13 mm min | |
Thickness - Palm | 0.08 mm min | 0.10 mm min | |
Physical Properties (Before Aging) | Tensile Strength (Mpa) | 18.0 Mpa | X-Small: 26.5, Small: 23.7, Medium: 27.0, Large: 28.5 |
Ultimate Elongation (%) | 650 % | X-Small: 875, Small: 805, Medium: 830, Large: 840 | |
Physical Properties (After Aging) | Tensile Strength (Mpa) | 14.0 Mpa | X-Small: 23.0, Small: 24.5, Medium: 25.8, Large: 25.5 |
Ultimate Elongation (%) | 500 % | X-Small: 750, Small: 800, Medium: 730, Large: 830 | |
Water Tight Test (Pinhole) | AQL (Acceptable Quality Level) | 2.5% AQL (ASTM D3578-00aE2) | Less than 2.5% AQL (as shown by individual leak numbers per batch) |
Biocompatibility | - | Passed tests for examination gloves (as per APPENDIX K) | Passed tests for examination gloves (as per APPENDIX K) |
Residual Powder Content | ASTM D 6124-00 | 2 mg/glove max | Ranger: 0.6 - 1.4 mg/glove, Mean: 1.1 mg/glove |
Presence of Cornstarch | - | Negative | Negative |
Residual Protein Level | ASTM D 5712-99 |
Ask a specific question about this device
(37 days)
SGMP CO., LTD.
The Non-Sterile, Latex, Powder-Free Patient Examination Gloves (Multicolored), with Protein Labeling Claim (50 Micrograms or less) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I latex patient examination gloves 80LYY, powder free and mecting all the requirements of ASTM-D3578-01aE2 Standard Specification for Latex Examination Glowes for Medical Application.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Acceptance Criteria (ASTM D3578-01aE2 or FDA/Internal Requirements) | Reported Device Performance (SGMP / Lot# 4144) |
---|---|---|
Dimension | ||
X-Small Width | 70 mm +/- 10 mm | 70 – 80 mm |
Small Width | 80 mm +/- 10 mm | 80 – 85 mm |
Medium Width | 95 mm +/- 10 mm | 90 – 97 mm |
Large Width | 111 mm +/- 10 mm | 105 - 111 mm |
Length (all sizes) | 230 mm minimum | 242 mm |
Thickness - Finger | 0.08 mm min | 0.13 mm min |
Thickness - Palm | 0.08 mm min | 0.1 mm min |
Physical Properties | ||
Ten. Strength Before Aging | 18.0 Mpa | X-Small: 23.5 Mpa; Small: 26.8 Mpa; Medium: 22.5 Mpa; Large: 29.2 Mpa |
Ult. Elong. Before Aging | 650 % | X-Small: 890 %; Small: 800 %; Medium: 850 %; Large: 870 % |
Ten. Strength After Aging | 14.0 Mpa | X-Small: 22.8 Mpa; Small: 27.5 Mpa; Medium: 28.3 Mpa; Large: 26.8 Mpa |
Ult. Elong. After Aging | 500 % | X-Small: 780 %; Small: 820 %; Medium: 750 %; Large: 820 % |
Water Tight Test | ASTM D3578-00aE2 requirements for latex exam gloves of 2.5% AQL (Acceptance Quality Limit) | Un-aged (X-Small: 0, Small: 1, Medium: 2, Large: 0 leaks); Aged (X-Small: 0, Small: 1, Medium: 1, Large: 2 leaks) - All within 2.5% AQL |
Biocompatibility | Passed tests for examination gloves (as per APPENDIX K) | Passed (results as per APPENDIX K) |
Residual Powder Content | FDA Requirement: 2 mg/glove max (ASTM D 6124-00) | Range: 0.4 - 1.0 mg/glove; Mean: 0.8 mg/glove |
Presence of Cornstarch | Negative | Negative |
Residual Protein Level |
Ask a specific question about this device
(87 days)
SGMP CO., LTD.
This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.
Class I nitrile patient examination gloves 80LZA, powder free and meeting all the requirements of ASTM-D6319-00aE1 Standard Specification for Nitrile Examination Gloves for Medical Application.
This document describes the "NON-STERILE POWDER FREE BLUE NITRILE EXAMINATION GLOVES WITH VANILLA SCENTING," a patient examination glove. The provided text outlines the device's classification, intended use, and its equivalence to a predicate device, but it lacks detailed information regarding specific performance acceptance criteria and study data typical for AI/ML device evaluations.
Therefore, for aspects related to AI/ML device evaluation, the information is not available in the provided text. The following describes the available information focused on the
glove's conformity to standards.
Acceptance Criteria and Device Performance for Powder-Free Blue Nitrile Examination Gloves with Vanilla Scenting
The acceptance criteria for this device are based on its conformity to recognized industry standards for patient examination gloves. The performance of the device is reported as meeting these standards.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Compliance with ASTM-D6319-00aE1 Standard Specification for Nitrile Examination Gloves for Medical Application | Device meets all requirements of ASTM-D6319-00aE1 |
Powder-Free | Confirmed to be powder-free |
Non-Sterile | Confirmed to be non-sterile |
Patient examination glove (per 21 CFR 880.6250) intended for medical applications | Confirmed for intended use |
2. Sample Size and Data Provenance for the Test Set
The provided summary does not detail a specific "test set" in the context of sampling a batch of gloves for evaluation. Instead, the device's performance is stated as meeting the requirements of a recognized standard, ASTM-D6319-00aE1. This standard outlines various tests (e.g., physical properties, barrier integrity) that examination gloves must pass. The glove manufacturer is responsible for ensuring ongoing compliance through their quality system, which would involve sampling and testing according to the standard's guidelines. The document does not specify the country of origin of the data or whether the data is retrospective or prospective, as it implicitly refers to adherence to a standard rather than a specific study.
3. Number of Experts and Qualifications for Ground Truth
This information is not applicable. For examination gloves, "ground truth" is established by adherence to specified physical and chemical properties defined in standards like ASTM-D6319-00aE1, rather than through expert consensus on diagnostic images or clinical outcomes. The experts would be professionals in materials testing, quality control, and regulatory compliance who ensure the manufacturing process and product meet the standard's specifications.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjectively assessed data, such as medical image interpretation, to resolve discrepancies among experts. For glove performance, objective measurements against technical specifications dictate acceptance.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not applicable. An MRMC study is relevant for evaluating the effectiveness of a diagnostic tool, such as how AI assistance improves human interpretation of medical images. This device is a medical glove, not a diagnostic tool requiring interpretation.
6. Standalone (Algorithm Only) Performance Study
This information is not applicable. This device is a physical medical product, not an algorithm or software. Therefore, there is no "standalone" algorithm performance to evaluate.
7. Type of Ground Truth Used
The ground truth used for this device is based on objective measurements against established technical specifications and performance criteria outlined in the ASTM-D6319-00aE1 standard. This includes physical properties (e.g., tensile strength, elongation), dimensions, and barrier integrity (e.g., pinhole detection). It does not involve expert consensus, pathology, or outcomes data in the way these terms are typically used for diagnostic devices.
8. Sample Size for the Training Set
This information is not applicable. The concept of a "training set" is relevant for machine learning algorithms or AI models. This product is a medical glove, manufactured according to established processes and tested against standards. There is no AI model or training involved in its development as described.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reasons mentioned in point 8.
Ask a specific question about this device
(395 days)
SGMP CO., LTD.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-01ar2 Standard Specification for Latex Examination Gloves for Medical' Application.
The document provided is a 510(k) Premarket Notification from the FDA for a medical device: "Non-Sterile Powder Free Green Latex Patient Examination Gloves with Nopal/Aloe Vera/Vitamin E and Donning Aid Coating, and With Protein Labeling Claim (
Ask a specific question about this device
Page 1 of 4