Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K042238
    Date Cleared
    2004-09-16

    (29 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80LZA, powder free that meets all the requirements of ASTM Standard D6319-00a53 and FDA Water Leak Test.

    AI/ML Overview

    This document describes the acceptance criteria and the study conducted for "Cashmere Polymer Coated Powder Free Nitrile Examination Gloves".

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (ASTM D6319-00aE3 / FDA)Reported Device Performance (POLYMER COATED POWDER FREE NITRILE EXAM GLOVES)
    Watertight (1000 ml)AQL=2.5% Pass G I (ASTM D6319-00aE3)AQL=2.5% Pass GI
    Length (mm)
    Min Length (all sizes)Min 230240 mm minimum for all sizes
    Palm Width (mm)
    Size XS- (no specific ASTM value)E3)
    Biocompatibility Test- (not specified in ASTM D6319-00aE3)Pass (Primary Dermal Irritation Test, Guinea Pig Sensitization (Buehler) test)

    Note: The asterisk () indicates average results from Attachment C, which is not provided in the given document.*

    Study Details:

    This document is a 510(k) summary for medical gloves, which are considered Class I devices. The studies performed are primarily bench tests and material characterization, aligning with the regulatory requirements for such devices. The concept of "AI" or "algorithms" does not apply to this type of device.

    2. Sample size used for the test set and data provenance:

    • The document implies that testing was conducted on samples of the manufactured gloves to ensure they meet the specified standards. However, the specific sample sizes for each test (e.g., number of gloves for watertight test, length, width, thickness, tensile strength, elongation, powder content, and biocompatibility) are not explicitly stated in the provided text.
    • The data provenance is from Seal Polymer Industries Berhad, Malaysia, presumably from their internal testing facilities. The tests are described as performance data of gloves, which indicates a prospective testing approach on manufactured products.

    3. Number of experts used to establish the ground truth for the test set and their qualifications:

    • This is not applicable as the device is a physical medical glove. The "ground truth" is established by adherence to recognized industrial standards (ASTM D6319-00aE3) and FDA requirements, rather than expert interpretation of complex data like medical images. The tests involve objective measurements and established protocols.

    4. Adjudication method for the test set:

    • This is not applicable for this type of device. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements in expert interpretations, typically in diagnostic or imaging studies. For physical product testing, results are determined by direct measurement against quantitative standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. This is a physical medical device (examination gloves), not an AI/software device involving human readers or interpretation of cases.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This is a physical medical device; there is no algorithm involved.

    7. The type of ground truth used:

    • The "ground truth" for the device's performance is based on established industrial standards (ASTM D6319-00aE3) and specific FDA requirements (e.g., watertight test, powder residual content, pinhole requirements, biocompatibility standards). These standards define the acceptable performance characteristics for patient examination gloves.

    8. The sample size for the training set:

    • This is not applicable. This is a physical medical device; there is no "training set" in the context of machine learning or AI. The manufacturing process of the gloves would involve quality control and process validation, but not a "training set" in the computational sense.

    9. How the ground truth for the training set was established:

    • This is not applicable as there is no training set for this physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042101
    Date Cleared
    2004-08-27

    (23 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80LZA, powder free that meets all the requirements of ASTM Standard D6319-00a53 and FDA requirements.

    AI/ML Overview

    This document describes the performance data for KOH2101, Powder Free Nitrile Examination Gloves, manufactured by SEAL Polymer Industries Berhad. The device is a Class 1 Nitrile Patient Examination Glove intended for medical purposes to prevent contamination between patients and examiners. The performance data is based on ASTM D6319-00a53 and FDA 1000 ml watertight test.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TESTASTM D6319-00aE3 Acceptance Criteria (AQL=2.5%)KOH2101 Powder Free Nitrile Exam Gloves Reported Performance (AQL=2.5%)
    1. Watertight (1000 ml)G IPass G I
    2. Length (mm)
    XSMin 230240 mm minimum for all sizes
    SMin 230
    MMin 230
    LMin 230
    XLMin 230
    3. Palm width (mm)
    XS-
    Ask a Question

    Ask a specific question about this device

    K Number
    K032077
    Date Cleared
    2003-08-19

    (43 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80LZA, powder free that meets all the requirements of ASTM Standard D6319-00a53 and FDA Water Leak Test.

    AI/ML Overview

    The device being discussed is "Cashmere Non-Sterile, Powder Free Nitrile Examination Gloves (Violet)". The study conducted to prove the device meets acceptance criteria is based on the performance data of the gloves according to ASTM D6319-00a3 standards and the FDA 1000 ml watertight test, as well as biocompatibility tests.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Test/CharacteristicAcceptance Criteria (ASTM D6319-00a3)Reported Device Performance (Cashmere Powder Free Nitrile Exam Gloves)
    Watertight (1000 ml)G I, AQL=2.5%Pass G I, AQL=2.5%
    Length (mm) - All SizesMin 230240 mm minimum for all sizes
    Palm width (mm) - Size XS-110
    Thickness (mm) - FingerMin 0.080.08 minimum
    Thickness (mm) - PalmMin 0.080.08 minimum
    Before Aging: Tensile Strength (Mpa)Min 14.023.9*
    Before Aging: Ultimate Elongation (%)Min 500760*
    After Aging: Tensile Strength (Mpa)Min 14.024.3*
    After Aging: Ultimate Elongation (%)Min 400640*
    Powder Content- (FDA Requirement)Below 2mg / glove
    BiocompatibilityPass (implied by FDA review)Pass Primary Dermal Irritation Test, Pass Guinea Pig Sensitization (Buehler) test
    • Indicates average results from Attachment B2, which is not provided in the document.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each test in the provided performance data. However, for the "Watertight (1000 ml)" test, an AQL (Acceptable Quality Level) of 2.5% is mentioned, which implies a sampling plan was used according to ASTM D6319-00a3.

    The data provenance is from Malaysia, where Seal Polymer Industries SDN. BHD. is located. It is
    prospective testing conducted by the manufacturer to demonstrate compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is a medical glove, and the "ground truth" is established by adherence to physical and chemical standards and biocompatibility tests, not subjective expert evaluation of medical images or conditions. The standards themselves (ASTM D6319-00a3 and FDA requirements) define the objective criteria.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The tests are objective measurements based on established standards, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-enabled diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI-enabled diagnostic device. The performance data presented refers to the physical and chemical properties of the glove itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on established industry standards and regulatory requirements:

    • ASTM D6319-00a3 Standard: Defines the physical properties and performance requirements for nitrile examination gloves.
    • FDA 1000 ml watertight test: A specific regulatory requirement for barrier integrity.
    • FDA minimum powder residual content: A specific regulatory limit for powder on gloves.
    • Biocompatibility test results: Objective laboratory tests (Primary Dermal Irritation Test and Guinea Pig Sensitization (Buehler) test) to assess the material's interaction with biological systems.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set. The manufacturing process for gloves is subject to quality control and batch testing, rather than an AI training paradigm.

    9. How the ground truth for the training set was established

    Not applicable, as no training set is relevant for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022792
    Date Cleared
    2002-09-26

    (35 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    Class 1 Latex Patient Examination Glove 80 LYY, powder free that meets all the requirements of ASTM Standard D3578-01 and FDA Water Leak Test.

    AI/ML Overview

    The provided document describes the performance data for "Cashmere Powder Free Latex Examination Gloves" based on ASTM D3578-01 and FDA 1000 ml watertight test.

    Here's an analysis of the acceptance criteria and study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (ASTM D3578-01)Reported Device Performance (CASHMERE POWDER FREE LATEX EXAM GLOVES)
    Watertight (1000 ml)AQL=4.0%Pass GI AQL=4.0%
    Length (mm)Min 230 (for all sizes S, M, L)240 mm minimum for all sizes
    Palm width (mm) - Size S80 +/- 1082 - 88
    Palm width (mm) - Size M95 +/- 1092 - 98
    Palm width (mm) - Size L111 +/- 10102 - 108
    Thickness (mm) (Single Layer) - FingerMin 0.080.10 minimum
    Thickness (mm) (Single Layer) - PalmMin 0.080.10 minimum
    Physical Properties - Before Aging
    Tensile Strength (Mpa)Min 1418.7*
    Ultimate Elongation (%)Min 650956*
    Physical Properties - After Aging
    Tensile Strength (Mpa)Min 1418.3*
    Ultimate Elongation (%)Min 500938*
    Powder Content(Implied FDA Requirement: Below 2 mg/glove)Below 2 mg / glove
    Protein Content(Implied FDA Requirement: Below 50 ug/gram)Below 50 microgram / gram

    * The document states these are "The average result taken from Attachment B2," which is not included in the provided text.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for each test. However, it mentions "AQL=4.0%" for the watertight test, which refers to Acceptable Quality Level, a sampling standard. The data provenance is Malaysia and the study appears to be a prospective testing of manufactured gloves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a physical device performance test based on established ASTM standards, not a diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication method for the test set

    Not applicable. The tests are objective measurements against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone device in the sense that its performance is measured directly against physical and chemical standards, without human interaction with an algorithm.

    7. The type of ground truth used

    The ground truth is established by the specified ASTM D3578-01 Standard and FDA requirements for watertightness, powder content, and protein content. These are objective, quantitative standards.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device, so there is no training set in that context. The "training" for the device would be the manufacturing process itself, and the goal is to consistently produce gloves that meet the specifications.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set mentioned in the context of an AI/ML algorithm. For the manufacturing process, the "ground truth" for the processes (making the gloves) is the ability to consistently achieve the stated performance criteria for the final product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021388
    Date Cleared
    2002-06-10

    (39 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    Class 1 Polymer Coated Patient Examination Glove 80 LYY, powder free that meets all the requirements of ASTM Standard D3578-00a and FDA Water Leak Test.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for Cashmere Powder Free Latex Examination Gloves.

    1. Acceptance Criteria and Reported Device Performance

    TESTAcceptance Criteria (ASTM D3578-00a)Reported Device Performance (Cashmere Powder Free Latex Exam Gloves)
    Watertight (1000 ml)GI AQL=4.0%Pass GI AQL=4.0%
    Length (mm)
    - Size XSMin 230240 mm minimum for all sizes
    - Size SMin 230
    - Size MMin 230
    - Size LMin 230
    - Size XLMin 230
    Palm width (mm)
    - Size XS-75 - 78
    - Size S80 +/- 1082 - 88
    - Size M95 +/- 1092 - 98
    - Size L111 +/- 10102 - 108
    - Size XL-111 - 115
    Thickness (mm)
    - Finger (Single Layer)Min 0.080.10 minimum
    - Palm (Single Layer)Min 0.080.10 minimum
    Physical Properties
    - Before Aging
    - Tensile Strength (Mpa)Min 1425.1
    - Ultimate Elongation (%)Min 650940
    - After Aging
    - Tensile Strength (Mpa)Min 1418
    - Ultimate Elongation (%)Min 500950
    Powder Content-Below 2 mg / glove
    Protein Content-Below 50 microgram / gram
    Biocompatibility- (implicitly, pass tests)Pass Primary Dermal Irritation Test and Guinea Pig Sensitization (Buehler) test

    2. Sample size used for the test set and the data provenance

    The document specifies "GI AQL=4.0%" for the watertight test, which refers to General Inspection Level 1 with an Acceptance Quality Limit of 4.0% according to ASTM D3578-00a. This standard typically involves sampling plans where the sample size is determined by the lot size, but the exact number of units tested is not explicitly stated in the provided text.

    The data provenance is from Malaysia, specifically from SEAL POLYMER INDUSTRIES SDN. BHD. The study appears to be prospective as it involves the testing of a manufactured product against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the document describes performance testing of a physical product (gloves) against established engineering and material standards (ASTM D3578-00a and FDA requirements), not the establishment of ground truth by human experts for a diagnostic algorithm.

    4. Adjudication method for the test set

    This information is not applicable as the document describes performance testing of a physical product against established objective standards, not a process requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the document describes the performance of a physical medical device (examination gloves) and does not involve AI, human readers, or case studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the document pertains to the performance of a physical medical device, not an algorithm.

    7. The type of ground truth used

    The ground truth used consists of established industry standards and regulatory requirements:

    • ASTM D3578-00a Standard for Latex Examination Gloves
    • FDA Water Leak Test requirements (1000 ml watertight test)
    • FDA minimum powder residual content requirements
    • Biocompatibility test standards (Primary Dermal Irritation Test and Guinea Pig Sensitization (Buehler) test)

    8. The sample size for the training set

    This information is not applicable. The document describes the performance of a physical product, not a statistical model or algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set mentioned or implied for this physical product performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020868
    Date Cleared
    2002-04-22

    (35 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patients' body, fluids, -.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80LZA, powder free that meets all the requirements of ASTM Standard D6319-00 and FDA Water Leak Test.

    AI/ML Overview

    The provided document describes the acceptance criteria and study for "Cashmere Non-Sterile, Powder Free Nitrile Examination Gloves (White)".

    Description of the Device and Intended Use:
    The device is a Class 1 Nitrile Patient Examination Glove, powder-free, intended to be worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste, or environment.

    Acceptance Criteria and Reported Device Performance:

    TESTASTM D6319-00 / FDA RequirementCASHMERE POWDER FREE NITRILE EXAM GLOVES (Reported Performance)Meets Criteria?
    1. Watertight (1000 ml)GI AQL=2.5%Pass GI AQL=2.5%Yes
    2. Length (mm) - All sizesMin 230240 mm minimum for all sizesYes
    3. Palm width (mm)
    XS-110Not directly comparable, but implied acceptable.
    4. Thickness (mm) (Single Layer)
    FingerMin 0.080.10 minimumYes
    PalmMin 0.080.10 minimumYes
    5. Physical Properties - Before Aging
    Tensile Strength (Mpa)Min 14.015.2Yes
    Ultimate Elongation (%)Min 500670Yes
    6. Physical Properties - After Aging
    Tensile Strength (Mpa)Min 14.019.7Yes
    Ultimate Elongation (%)Min 400760Yes
    7. Powder ContentBelow 2mg / glove (FDA requirement)Below 2mg / gloveYes
    8. Biocompatibility TestPass (implied requirement for medical devices)Pass (Primary Dermal Irritation, Guinea Pig Sensitization)Yes

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact sample sizes for each test mentioned in the table. It refers to "GI AQL=2.5%" for the watertight test, which indicates a sampling plan based on Acceptance Quality Limit, but not the specific number of gloves tested.
      • Data Provenance: The studies were conducted by SEAL POLYMER INDUSTRIES SDN. BHD. in Malaysia. The nature of the tests (physical property measurements, watertight tests, biocompatibility) implies prospective testing on manufactured gloves.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is not applicable as the studies are technical performance tests against established standards (ASTM D6319-00, FDA requirements) and not diagnostic or interpretive tasks requiring expert ground truth establishment.
    3. Adjudication method for the test set:

      • Not applicable for these types of performance tests. The results are objective measurements against defined minimums or ranges.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case comparative effectiveness study was not done. This type of study is not relevant for the performance evaluation of examination gloves, which are physical medical devices.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device is a physical glove, not an algorithm.
    6. The type of ground truth used:

      • The "ground truth" for the performance tests consists of established industry standards and regulatory requirements:
        • ASTM D6319-00 Standard for Nitrile Patient Examination Gloves.
        • FDA 1000 ml watertight test requirements.
        • FDA minimum powder residual content requirements.
      • For the biocompatibility tests, the ground truth is "Pass" based on established biological safety evaluation methods (Primary Dermal Irritation Test and Guinea Pig Sensitization test).
    7. The sample size for the training set:

      • Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K014055
    Date Cleared
    2002-01-28

    (49 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

    Device Description

    Class 1 Powdered Latex Patient Examination Glove 80 LYY, powdered with absorbable dusting powder, which meets all the requirements of ASTM Standard D3578-99 and FDA Water Leak Test.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for "Cashmere Non-Sterile, Powdered Latex Examination Gloves". Here's the information requested, extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TESTAcceptance Criteria (based on ASTM D3578-99 and FDA 1000 ml watertight test)Reported Device Performance (CASHMERE LIGHTLY POWDERED LATEX EXAM GLOVES)
    1. Watertight (1000 ml)GI AQL=2.5%Pass GI AQL=2.5%
    2. Length (mm) - All SizesMin 230240 mm minimum for all sizes
    3. Palm width (mm) - Size XS-110
    4. Thickness (mm) (Single Layer) - FingerMin 0.080.10 minimum
    4. Thickness (mm) (Single Layer) - PalmMin 0.080.10 minimum
    5. Physical Properties (Before Aging) - Tensile Strength (Mpa)Min 1427.1*
    5. Physical Properties (Before Aging) - Ultimate Elongation (%)Min 700917.5*
    5. Physical Properties (After Aging) - Tensile Strength (Mpa)Min 1424.9*
    5. Physical Properties (After Aging) - Ultimate Elongation (%)Min 500880*
    6. Powder ContentBelow 200 mg / glove (FDA requirement)Below 200 mg / glove
    7. Protein ContentBelow 200 microgram / gram (As per the device name's claim)Below 200 microgram / gram
    8. Biocompatibility TestPass Primary Dermal Irritation Test and Guinea Pig Sensitization (Buchler) testPass Biocompatibility Tests

    Note: The document states these figures were the average among the eight tested samples.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document explicitly mentions "eight tested samples" for the physical properties (tensile strength and ultimate elongation). For watertightness, it references "GI AQL=2.5%", which implies a sampling plan defined by ASTM D3578-99 and FDA 1000 ml watertight test, but the specific number of gloves tested is not provided in this summary.
    • Data Provenance: The study appears to be retrospective, as it's a summary of performance data collected to demonstrate compliance with existing standards (ASTM D3578-99 and FDA requirements). The manufacturer is Seal Polymer Industries Sdn. Bhd., located in Lahat, Perak, Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a performance claim for a physical device based on standardized tests, not an AI/diagnostic device requiring expert ground truth establishment for medical image interpretation or similar. The "ground truth" is established by the specifications in ASTM D3578-99 and FDA requirements.

    4. Adjudication method for the test set

    Not applicable. As this involves physical testing against established standards, adjudication by experts for discordant results is not mentioned or typically relevant. The results are objective measurements or pass/fail criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a medical glove, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document describes a medical glove, not an algorithm.

    7. The type of ground truth used

    The ground truth for this device's performance is based on established industry standards and regulatory requirements:

    • ASTM D3578-99 Standard
    • FDA 1000 ml watertight test requirements
    • FDA minimum powder residual content requirements
    • Biocompatibility test standards (Primary Dermal Irritation Test and Guinea Pig Sensitization (Buchler) test).

    8. The sample size for the training set

    Not applicable. This is a medical glove, not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a medical glove, not an AI device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993134
    Date Cleared
    1999-11-17

    (58 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patients' body, fluids, waste or environment.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80 LZA, powder free that meets all the requirements of ASTM Standard D3578-95 and FDA Water Leak Test.

    AI/ML Overview

    This document describes the acceptance criteria and performance of the Cashmere Non-Sterile, Powder Free Nitrile Examination Gloves (Blue and Green).

    1. Table of Acceptance Criteria and Reported Device Performance:

    TESTAcceptance Criteria (ASTM D3578-95)Reported Device Performance (Cashmere Powder Free Nitrile Exam Gloves)
    Watertight (1000 ml)GI AQL=4.0%Pass GI AQL=4.0%
    Length (mm)
    Size XSMin 230240 mm minimum for all sizes
    SMin 230
    MMin 230
    LMin 230
    XLMin 230
    Palm width (mm)
    Size XS-75 - 78
    S80 +/- 1082 - 88
    M95 +/- 1092 - 98
    L111 +/- 10102 - 108
    XL-111 - 115
    Thickness (mm) (Single Layer)
    FingerMin 0.080.10 minimum
    PalmMin 0.080.10 minimum
    Physical Properties - Before Aging
    Tensile Strength (Mpa)Min 1416.5
    Ultimate Elongation (%)Min 500550
    Physical Properties - After Aging
    Tensile Strength (Mpa)Min 1423.1
    Ultimate Elongation (%)Min 500520
    Powder Content- (FDA Requirement)Below 2 mg / glove

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each test. However, the performance data is based on the ASTM D3578-95 standard and the FDA 1000 ml watertight test, which would typically specify sampling plans. The data provenance is from SEAL POLYMER INDUSTRIES SDN. BHD. in Lahat, Perak, Malaysia. The study appears to be a retrospective analysis of product performance against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For product performance testing against standard specifications like ASTM D3578-95, the "ground truth" is typically defined by the objective measurement procedures within the standard, not by expert consensus on individual cases.

    4. Adjudication method for the test set:

    Not applicable. The performance data is based on objective measurements and compliance with a standard (ASTM D3578-95) and FDA requirements, not on subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for a medical glove, a physical device, and not an AI-powered diagnostic or decision support tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    The ground truth used for performance evaluation is based on established industry standards and regulatory requirements, specifically:

    • ASTM D3578-95 Standard for Rubber Examination Gloves.
    • FDA 1000 ml watertight test for pinhole requirements.
    • FDA requirements for minimum powder residual content.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993326
    Date Cleared
    1999-11-17

    (44 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patients' body, fluids, waste or environment.

    Device Description

    Class 1 Polymer Coated Patient Examination Glove 80 LYY, powder free that meets all the requirements of ASTM Standard D3578-95 and FDA Water Leak Test.

    AI/ML Overview

    The provided documentation describes the acceptance criteria and performance of the "Cashmere Non-Sterile, Polymer Coated Powder Free Latex Examination Gloves." This is a Class I medical device, and the evaluation is based on established ASTM standards and FDA requirements for such gloves, not an AI/ML device. Therefore, the questions related to AI/ML specific criteria (e.g., test set sample size, expert ground truth, MRMC study, standalone performance, training set) are not applicable.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TESTAcceptance Criteria (ASTM D3578-95 & FDA)Reported Device Performance (Cashmere Polymer Coated Powder Free Latex Exam Gloves)
    1. Watertight (1000 ml)G I, AQL=4.0%Pass GI, AQL=4.0%
    2. Length (mm)Min 230 mm (all sizes)240 mm minimum for all sizes
    - Size XSMin 230240 mm
    - Size SMin 230240 mm
    - Size MMin 230240 mm
    - Size LMin 230240 mm
    - Size XLMin 230240 mm
    3. Palm width (mm)
    - Size XS-75 - 78
    - Size S80 +/- 1082 - 88
    - Size M95 +/- 1092 - 98
    - Size L111 +/- 10102 - 108
    - Size XL-111 - 115
    4. Thickness (mm)
    - Finger (Single Layer)Min 0.080.10 minimum
    - Palm (Single Layer)Min 0.080.10 minimum
    5. Physical Properties
    Before Aging
    - Tensile Strength (Mpa)Min 1416.5
    - Ultimate Elongation (%)Min 700550
    After Aging
    - Tensile Strength (Mpa)Min 1423.1
    - Ultimate Elongation (%)Min 500520
    6. Powder Content- (No specific ASTM value provided)Below 2 mg / glove
    7. Protein Content- (No specific ASTM value provided)Below 70 microgram / gram

    2. Sample size used for the test set and the data provenance

    The document states that performance data is based on "ASTM D3578-95 and FDA 1000 ml watertight test." These standards specify sampling plans (e.g., AQL for watertightness), but the exact sample sizes used for this specific submission are not explicitly provided in the excerpt. The data provenance is for a manufacturing company located in Malaysia (SEAL POLYMER INDUSTRIES SDN. BHD., Lot 72706, Jalan Lahat, Kawasan Perindustrian Bukit Merah, 31500 Lahat, Perak, Malaysia). The nature of these tests (physical and chemical properties) indicates a prospective evaluation of manufactured samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical glove, and its performance is evaluated against objective physical and chemical standards (ASTM D3578-95, FDA watertight test). It does not involve expert interpretation or ground truth establishment in the context of diagnostic AI/ML devices.

    4. Adjudication method for the test set

    Not applicable. Device performance is measured against quantifiable standards for physical and chemical properties.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used

    The "ground truth" for this device's performance is objective measurement against established industry and regulatory standards. For example, for "Watertightness," the ground truth is whether the glove passes the FDA's 1000 ml watertight test at an AQL of 4.0%. For dimensions, it's the measured length and width compared to the specified range. For physical properties, it's the measured tensile strength and elongation compared to minimum requirements.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993133
    Date Cleared
    1999-11-17

    (58 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    SEAL POLYMER INDUSTRIES BHD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patients' body, fluids, waste or environment.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80 LZA, powdered with absorbable dusting powder, that meets all the requirements of ASTM Standard D3578-95 and FDA Water Leak Test.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of "Cashmere Non-Sterile, Powdered Nitrile Examination Gloves (Blue)" and "Cashmere Non-Sterile, Powdered Nitrile Examination Gloves (Green)" as detailed in the K993133 510(k) submission.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (ASTM D3578-95 & FDA)Reported Device Performance (Cashmere Powdered Nitrile Exam Gloves)
    Watertight (1000 ml)GI AQL=4.0%Pass GI AQL=4.0%
    Length (mm) - Min. for all sizesMin 230240 mm minimum for all sizes
    Palm width (mm) - Size XS-75 - 78
    Palm width (mm) - Size S80 +/- 1082 - 88
    Palm width (mm) - Size M95 +/- 1092 - 98
    Palm width (mm) - Size L111 +/- 10102 - 108
    Palm width (mm) - Size XL-111 - 115
    Thickness (mm) - Finger (Single Layer)Min 0.080.10 minimum
    Thickness (mm) - Palm (Single Layer)Min 0.080.10 minimum
    Before Aging:
    Tensile Strength (Mpa) Before AgingMin 1430.4
    Ultimate Elongation (%) Before AgingMin 500640
    After Aging:
    Tensile Strength (Mpa) After AgingMin 1430.9
    Ultimate Elongation (%) After AgingMin 500610
    Powder Content
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2