510(k) Data Aggregation

    K Number
    K192595
    Device Name
    ExSpiron 2Xi
    Manufacturer
    Respiratory Motion
    Date Cleared
    2019-12-17

    (88 days)

    Product Code
    BZK, BZO
    Regulation Number
    868.1850
    Why did this record match?
    Applicant Name (Manufacturer) :

    Respiratory Motion

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ExSpiron 2Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in patients at least 1 - year of age. ExSpiron 2Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of: - Minute Ventilation (MV) - Tidal volume (TV) - Respiratory Rate (RR) ExSpiron 2Xi measurements are used as an adjunct to other clinical information.
    Device Description
    The ExSpiron 2Xi Respiratory Monitor System consists of: Monitor: The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure. Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate. Patient Cables and Electrode Padsets that are also included in the system but were cleared in previous 510(k)s: K130170, K162131, K173181 ExSpiron Patient Cable: A reusable cable that connects the ExSpiron 2Xi Monitor to the Electrode PadSet. Single patient use ExSpiron Electrode PadSet: A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin.
    Ask a Question
    K Number
    K173181
    Device Name
    ExSpiron 1Xi
    Manufacturer
    Respiratory Motion Inc.
    Date Cleared
    2018-03-16

    (168 days)

    Product Code
    BZK, BZQ
    Regulation Number
    868.1850
    Why did this record match?
    Applicant Name (Manufacturer) :

    Respiratory Motion Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in patients at least one year of age. ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of: * Minute Ventilation (MV) * Tidal volume (TV) * Respiratory rate (RR) ExSpiron 1Xi measurements are used as an adjunct to other clinical information.
    Device Description
    The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate. The ExSpiron 1Xi system consists of: • Monitor: The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure. o Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. o Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate. There are no hardware changes included in this submission. Single-Patient Use ExSpiron Electrode PadSet: • A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. The Electrode PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.
    Ask a Question
    K Number
    K162131
    Device Name
    ExSpiron 1Xi
    Manufacturer
    Respiratory Motion Inc.
    Date Cleared
    2017-05-09

    (281 days)

    Product Code
    BZK, BZQ
    Regulation Number
    868.1850
    Why did this record match?
    Applicant Name (Manufacturer) :

    Respiratory Motion Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients. ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of: * Minute Ventilation (MV) * Tidal volume (TV) * Respiratory rate (RR) ExSpiron 1Xi measurements are used as an adjunct to other clinical information.
    Device Description
    The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate. The ExSpiron 1Xi system consists of: Monitor: The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure. Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate. A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. The Electrode PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.
    Ask a Question
    K Number
    K130170
    Device Name
    EXSPIRON 1XI
    Manufacturer
    RESPIRATORY MOTION, INC.
    Date Cleared
    2013-05-29

    (125 days)

    Product Code
    BZK, BZQ
    Regulation Number
    868.1850
    Why did this record match?
    Applicant Name (Manufacturer) :

    RESPIRATORY MOTION, INC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ExSpiron 1xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients. ExSpiron 1xi is a non-invasive system that graphically displays lung volume against time and reports an approximate value of: Tidal volume, Respiratory rate, and Minute ventilation ExSpiron 1xi measurements are used as an adjunct to other clinical information sources.
    Device Description
    The ExSpiron 1xi system consists of: Bioimpedance measurement system: A stabilized high frequency current generator is connected to two outer electrodes. The inner four electrodes are connected to an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware performs signal acquisition and relays data to the panel PC. Computer: A Windows 7 PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute volume, tidal volume, and respiratory rate. Single Patient Use ExSpiron™ 1xi Electrode PadSet: An electrode assembly containing six electrodes to be placed on the torso. It delivers current and records impedance measurements. The electrode PadSet is also used to perform subsystem checks prior to patient measurements.
    Ask a Question
    K Number
    K120087
    Device Name
    EXSPIRON
    Manufacturer
    RESPIRATORY MOTION, INC.
    Date Cleared
    2012-09-26

    (259 days)

    Product Code
    BZK, BZQ
    Regulation Number
    868.1850
    Why did this record match?
    Applicant Name (Manufacturer) :

    RESPIRATORY MOTION, INC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    ExSpiron is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years of age) patients. ExSpiron is a non-invasive system that graphically displays lung volume against time and reports an approximate value of: - Tidal volume, . - . Respiratory rate, and - Minute ventilation. ● ExSpiron measurements are used as an adjunct to other clinical information sources.
    Device Description
    The ExSpiron™ consists of: Bioimpedance measurement system: A stabilized high frequency current generator is connected to two outer electrodes. The inner four electrodes are connected to an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware performs signal acquisition and relays data to the panel PC. Panel PC: A Windows 7 PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute volume, tidal volume, and respiratory rate. Single Patient Use ExSpiron™ Electrode Lead Set: An electrode lead set containing six electrodes to be placed on the torso. It delivers current and records impedance measurements. The electrode lead set is also used to perform subsystem checks prior to patient measurements.