K Number
K162131

Validate with FDA (Live)

Device Name
ExSpiron 1Xi
Date Cleared
2017-05-09

(281 days)

Product Code
Regulation Number
868.1850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.

ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:

  • Minute Ventilation (MV)

  • Tidal volume (TV)

  • Respiratory rate (RR)

ExSpiron 1Xi measurements are used as an adjunct to other clinical information.

Device Description

The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displays lung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate.

The ExSpiron 1Xi system consists of:
Monitor:
The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.
Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate.
A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. The Electrode PadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.

AI/ML Overview

Acceptance Criteria and Device Performance for ExSpiron™ 1Xi

1. Table of Acceptance Criteria and Reported Device Performance

The provided document compares the proposed device (ExSpiron 1Xi) to a predicate device (ExSpiron™ 1Xi, K130170). The comparison focuses on various characteristics, including accuracy. While explicit "acceptance criteria" for accuracy are not stated as numerical thresholds, the document presents the accuracy results of the proposed device. The implied acceptance is that the proposed device performs at least as well as, or better than, the predicate device.

CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (Proposed ExSpiron 1Xi)
Accuracy
Minute Ventilation (MV)10.7%11.5% (Overall)
Tidal Volume (TV)10.4%11.4% (Overall)
Respiratory Rate (RR)0.1%0.08% (Overall)
Bias(Not explicitly stated for predicate)MV: 2.2%, TV: 2.2%, RR: 0.01% (Overall)
Precision(Not explicitly stated for predicate)MV: 7.0%, TV: 6.9%, RR: 0.08% (Overall)
Intended UseIdentical to predicateIdentical to predicate
TechnologyThoracic bioimpedanceThoracic bioimpedance
Volume MeasurementsTidal volume, Minute volume, Volume vs. time chartTidal volume, Minute volume, Volume vs. time chart
Rate MeasurementsRespiratory rate (breaths/min)Respiratory rate (breaths/min)
SafetyIEC 60601-1 compliantIEC 60601-1 compliant
Energy SourceLow current signal at approximately 50 kHzLow current signal at approximately 50 kHz
EMC/EMI ComplianceIEC 60601-1-2 compliantIEC 60601-1-2 compliant
Electrode PadSetSingle-patient use, biocompatible, printed padsetSingle-patient use, biocompatible, printed padset
Impedance Measurement Range15 Ohms to 180 Ohms15 Ohms to 180 Ohms
BiocompatibilityElectrode PadSet is biocompatible; Monitor and Patient Cable not for patient contactElectrode PadSet is biocompatible; Monitor and Patient Cable not for patient contact
UsabilityISO 60601-1-6 compliant (with calibration)ISO 60601-1-6 compliant (without calibration in Basic Monitoring mode)

The document notes that the "algorithm was refined for the ExSpiron 1Xi using thousands of additional data points collected over the last 3 years in a wide variety of patient populations," leading to "more accurate respiratory volumes than the predicate." However, the accuracy percentages show a slight decrease in accuracy (higher percentage value) for MV and TV compared to the predicate in this summary document. This might be due to presenting overall clinical study results for the proposed device, while the predicate's numbers might be from a specific lab setting or a different calculation method for its initial clearance. The conclusion explicitly states the proposed device "is as safe and as effective as the predicate."

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: 20 subjects
    • Age range: 22-80
    • BMI range: 19-42
    • Gender: 9 female, 11 male
  • Data Provenance:
    • Country of Origin: Waltham, MA, USA (clinical study conducted there)
    • Retrospective or Prospective: Prospective. The study was conducted specifically to compare simultaneous measurements from the ExSpiron 1Xi and the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts used to establish the ground truth for this clinical study, nor does it specify their qualifications. The study focused on comparing the performance of the proposed device against a predicate device through simultaneous measurements. This implies the predicate device's measurements or another recognized standard served as the comparative reference, rather than a separate "ground truth" established by experts for each breath.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set. It states that simultaneous measurements were taken from both the proposed device and the predicate device, implying a direct comparison rather than a need for expert adjudication of independent measurements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a clinical study comparing simultaneous measurements of the proposed device and a predicate device on patients, not a study evaluating human reader performance with and without AI assistance.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

Yes, a standalone performance aspect is implied. The "Accuracy" section of the comparison table and the clinical study results (Bias, Precision, Accuracy) directly assess the performance of the device's algorithms in measuring respiratory parameters (MV, TV, RR) without explicitly mentioning human intervention in the real-time measurement process. The device provides "approximate values" of these parameters.

7. Type of Ground Truth Used

The ground truth for the clinical study was established by simultaneous measurements from a legally marketed predicate device (ExSpiron™ 1Xi, K130170). The study aimed to show equivalence or improvement compared to an already cleared device.

8. Sample Size for the Training Set

  • The predicate device's algorithm was "trained with a dataset containing respiratory data from a large historical cohort."
  • The proposed device's algorithm was "refined for the ExSpiron 1Xi using thousands of additional data points collected over the last 3 years in a wide variety of patient populations." The exact numerical sample size ("thousands") is not specified beyond this qualitative description.

9. How the Ground Truth for the Training Set Was Established

The document does not explicitly detail how the ground truth for the "large historical cohort" or the "thousands of additional data points" used for algorithm training/refinement was established. However, given the nature of a spirometer, it is highly probable that the ground truth for training data would have been established using:

  • Reference Spirometry: Directly measured, highly accurate respiratory volumes and rates from calibrated gold-standard spirometry equipment.
  • Other established physiological measurement techniques that provide accurate respiratory parameters.

The purpose of the training data is to allow the device's proprietary, non-linear, regressive model to accurately calculate respiratory parameters "without patient-specific calibration" for the proposed device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 9, 2017

Respiratory Motion, Inc. Susan Hamann Director of Quality and Regulatory 411 Waverly Oaks Road, Building 1, Suite 150 Waltham, Massachusetts 02452

Re: K162131

Trade/Device Name: ExSpiron™ 1Xi Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: Class II Product Code: BZK, BZQ Dated: April 5, 2017 Received: April 10, 2017

Dear Susan Hamann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina Kiang
-s

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162131

Device Name ExSpiron 1Xi

Indications for Use (Describe)

ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult (at least 21 years old) patients.

ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:

  • Minute Ventilation (MV)

  • Tidal volume (TV)

  • Respiratory rate (RR)

ExSpiron 1Xi measurements are used as an adjunct to other clinical information.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows a pair of human lungs against a black background. The lungs are translucent and colored in shades of blue, giving them a somewhat ethereal appearance. The anatomical structure of the lungs is visible, including the lobes and some of the branching airways. The lighting and color choice create a visual effect that highlights the shape and form of the lungs.

510(k) Summary – Traditional 510(k)

510(k) Owner:Respiratory Motion, Inc.
Address:411 Waverley Oaks RoadBuilding 1, Suite 150Waltham, Massachusetts 02452
Phone:781-373-1636
Fax:781-373-1653
Contactperson:Jenny Freeman, MD
Date 510(k)Summary prepared andtype of 510(k):May 9, 2017 – Traditional 510(k)
Trade name:ExSpironTM 1Xi
Commonname:Respiratory Monitoring System
Classifications:Regulation Number: 21 CFR 868.1850Regulation Name: Monitoring SpirometerRegulatory Class: IIProduct Code: BZK, BZQ
Predicatedevices:ExSpiron 1Xi Respiratory Monitor, marketed by Respiratory Motion, Inc, Waltham, MA (K130170).
DeviceDescription:The ExSpiron 1Xi is a noninvasive respiratory monitoring system that graphically displayslung volume against time and reports Minute Ventilation, Tidal Volume and Respiratory Rate.The ExSpiron 1Xi system consists of:Monitor:The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC. Computer: A tablet PC performs signal processing and calibration, and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate.

{4}------------------------------------------------

A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. Itdelivers current to, and records impedance measurements from, the skin. The ElectrodePadSet for the ExSpiron 1Xi is identical to that cleared for the Predicate device.
Intended use:ExSpiron 1Xi is indicated for use by healthcare professionals in healthcare facilities, such aspost-operative care and critical care units, to monitor breathing in adult (at least 21 years old)patients.ExSpiron 1Xi is a non-invasive monitor that graphically displays lung volume against time andreports an approximate value of:* Minute Ventilation (MV)* Tidal volume (TV)* Respiratory rate (RR)ExSpiron 1Xi measurements are used as an adjunct to other clinical information.

{5}------------------------------------------------

Comparison oftechnologicalcharacteristics:CharacteristicExSpiron 1 (Proposed Device)ExSpironTM1Xi(Predicate) K130170Comment
Intended UseExSpiron 1Xi is indicated for useby healthcare professionals inhealthcare facilities, such as post-operative care and critical careunits, to monitor breathing in adult(at least 21 years old) patientsExSpiron 1Xi is indicated for use byhealthcare professionals inhealthcare facilities, such as post-operative care and critical care units,to monitor breathing in adult (atleast 21 years old) patientsBoth the proposed device and the predicate have an identical intendeduse.
Each ExSpiron 1Xi is anoninvasive system thatgraphically displays lung volumeagainst time and reports anapproximate value of:• Tidal volume• Respiratory rate, and• Minute ventilation.Each ExSpiron 1Xi is a noninvasivesystem that graphically displayslung volume against time andreports an approximate value of:• Tidal volume• Respiratory rate, and• Minute ventilation.
ExSpiron 1Xi measurements areused as an adjunct to other clinicalinformation sources.ExSpiron 1Xi measurements areused as an adjunct to other clinicalinformation sources.
TechnologyMeasurement is by thoracicbioimpedance.Measurement is by thoracicbioimpedance.Both the proposed device and the predicate have identical technology.
VolumeMeasurementsTidal volumeMinute volumeVolume vs. time chartTidal volumeMinute volumeVolume vs. time chartBoth the proposed device and the predicate measure the samerespiratory volumes.
RateMeasurementsRespiratory rate (breaths/min)Respiratory rate (breaths/min)Both the proposed device and the predicate measure respiratory rate.
SafetyIEC 60601-1, including electricaland mechanical safetyIEC 60601-1, including electricaland mechanical safetyBoth the proposed device and the predicate have the same electricalsafety profile and are compliant with IEC 60601-1.
Energy SourceThe bioimpedance output signal isgenerated internally to theExSpiron Monitor. It is a lowcurrent signal at approximately 50kHz.The bioimpedance output signal isgenerated internally to the ExSpironMonitor. It is a low current signal atapproximately 50 kHz.Both the proposed device and the predicate have the same energysource.
AlgorithmA software algorithm calculatesrespiratory parameters using aproprietary, non-linear, regressivemodel trained with a datasetcontaining respiratory data from alarge historical cohort. Thealgorithm was refined for theExSpiron 1Xi- using thousands ofadditional data points collectedover the last 3 years in a widevariety of patient populations.With the additional training data,the algorithm now providesaccurate respiratory volumeswithout patient-specificA software algorithm calculatesrespiratory parameters using aproprietary, non-linear, regressivemodel trained with a datasetcontaining respiratory data from alarge historical cohort.Both the proposed device and the predicate have essentially the samefirmware, except the proposed device has been refined with thousandsof additional data points that result in a more accurate estimate ofrespiratory parameters than the predicate.
Enclosurecalibration.Formed plastic (ABS) housing,flame-retardant.Formed plastic (ABS) housing,flame-retardant.Both the proposed device and the predicate have the same enclosurematerial.
EMC/EMIComplianceIEC 60601-1-2 compliantIEC 60601-1-2 compliantBoth the proposed device and the predicate are 60601-1-2 compliant.
ElectrodePadSetSingle-patient use, biocompatible,printed padset.Single-patient use, biocompatible,printed padset.Both the proposed device and the predicate use the same electrodepadset.
ImpedanceMeasurementRange15 Ohms to 180 Ohms15 Ohms to 180 OhmsBoth the proposed device and the predicate have the samemeasurement range.
TabletComputerWindows tablet computer.Windows tablet computer.Following FDA clearance of the predicate (K130170), a "letter to file"change was made to substitute an equivalent Windows tabletcomputer in the device (See "Letter to File" changes; Section 21b inthis 510(k) filing).
BiocompatibilityNeither the Monitor nor thePatient Cable are intended forpatient contact. The ElectrodePadSet is biocompatible.Neither the Monitor nor the PatientCable are intended for patientcontact. The Electrode PadSet isbiocompatible.Both the proposed device and the predicate have identicalbiocompatibility.
UsabilityISO 60601-1-6 compliantISO 60601-1-6 compliantBoth the proposed device and the predicate have the same usabilitywhen used in the Volume Synchronization and Percent Baselinemodes of operation. In the Basic Monitoring mode that has beenintroduced in the proposed device, the calibration step that wasrequired for the predicate device has been eliminated.
AccuracyMV - 11.5%TV - 11.4%RR - 0.1%MV - 10.7%TV - 10.4%RR - 0.1%Details of the accuracy of the proposed device versus the predicate areincluded in Section 20: Performance Testing - Clinical.

{6}------------------------------------------------

T

{7}------------------------------------------------

Respiratory Motion, Inc.09 May 2017
Nonclinicaltesting:Nonclinical testing was performed to demonstrate the equivalence of the ExSpiron 1Xi tothe predicate. The FDA recognized tests that were completed are listed below. SeeSection 9.0; Declarations of Conformity and Summary Reports for details of the testing.
- Basic Safety and Essential Performance (IEC 60601-1; Ed. 3.0) - Pass,- Immunity/Emissions (IEC 60601-1-2; Ed. 3.0) - Pass,- Usability (IEC 60601-1-6; 2010, A1:2013 and 62366;2007, A1:2014) –Pass, and- Alarms (IEC 60601-1-8; 2006) - Pass.
Other tests which are not FDA recognized were also completed. The following non-FDArecognized tests were performed and each achieved favorable results:
- Degrees of protection provided by enclosures (IEC code) (IEC 60529;2004)- Packaged products weighing 150 lbs (68 kg) or less (ISTA Procedure1A; 2014)- ECG Trunk Cables and patient lead wires (ANSI EC53; 2013)
The results of this testing demonstrate that the proposed device is equivalent to thepredicate in safety and essential performance.
Clinicalperformancetesting:A clinical study was conducted in Waltham MA, USA to compare simultaneousmeasurements from the ExSpiron 1Xi and the predicate device. Twenty subjectsrepresenting a broad range of intended patients participated in the study. (Age range: 22-80, BMI range: 19-42 with 9 female, 11 male). The study involved two sessions for eachsubject, an initial session in which electrodes were applied and each subject performedtwenty breathing tests. Tidal volume, minute ventilation, and respiratory rate weremeasured and processed by both the Predicate and the Proposed device. Each subjectreturned twenty-four hours after the first session with the original electrodes still attached.A second set of twenty breathing tests were performed.
The results of the clinical study were:
Proposed ExSpiron 1Xi /PredicateExSpiron 1Xi ComparisonMinuteVentilationTidalVolumeRespiratoryRate
BiasOverall2.2%2.2%0.01%
Day 13.4%3.4%0.02%
Day 21.1%1.1%0.00%
PrecisionOverall7.0%6.9%0.08%
Day 15.2%5.1%0.11%
Day 25.1%5.1%0.00%
AccuracyOverall11.5%11.4%0.08%
Day 110.1%10.1%0.11%
Day 212.3%12.3%0.00%
The results indicate clinically relevant accuracy over a 24-hour period.
Conclusionsregardingsafety andeffectiveness:Based on the comparison of the intended use of the predicate and the proposed device, andon the results of nonclinical and clinical testing, the proposed ExSpiron 1Xi is substantiallyequivalent to the predicate and is as safe and as effective as the predicate.

§ 868.1850 Monitoring spirometer.

(a)
Identification. A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.(b)
Classification. Class II (performance standards).