Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143534
    Date Cleared
    2015-08-06

    (237 days)

    Product Code
    Regulation Number
    866.6010
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ROCHE PROFESSIONAL DIAGNOSTICS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Elecsys CA 125 II is an immunoassay for the in vitro quantitative determination of OC 125 reactive determinants in human serum, Li heparin, K2-EDTA, as well as Li-heparin plasma tubes containing separating gel on the cobas e 411 analyzer.

    These determinants are associated with a high molecular weight glycoprotein in serum and plasma of women with primary epithelial invasive ovarian cancer (excluding those with cancer of low malignant potential).

    This immunoassay is indicated for use as an aid in the detection of recurrent ovarian carcinoma. This immunoassay is further indicated for use in monitoring patients for disease progress or response to therapy.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzers.

    For use in the verification of the callbration established by the Elecsys CA 125 II reagent on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    The CA 125 II assay employs a sandwich test principle using biotinylated monoclonal CA 125-specific antibody and a monoclonal CA 125-specific antibody labeled with a ruthenium complex to form a sandwich complex. The use of streptavidin-coated microparticles serves as the solid phase for the electrochemiluminescence detection.

    Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve (5-point-calibration) provided with the reagent bar code.

    The CA 125 II application is identical to the predicate assay (K972162). This submission is being done to modernize the labeling by adding the LoB, LoD and LoQ data and to change the sample:reagent ratio from 40:60μL to 20:70μL. Additionally, based on internal stability data the calibration frequency has been extended from 4 to 8 weeks.

    The Elecsys CA 125 II CalCheck is a lyophilized product consisting of equine serum in level 1 and human serum matrix for levels 2 and 3. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Elecsys CA 125 II Assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comparison table (Table 1) between the predicate device (Elecsys CA 125 II, K972162) and the candidate device (Elecsys CA 125 II Assay), highlighting both similarities and differences, including labeled performance characteristics. Since explicit "acceptance criteria" are not given in a numerical form that can be directly compared to "reported performance" for each item, I will present the key performance characteristics detailed for the candidate device as its reported performance, implicitly indicating what was demonstrated to FDA for substantial equivalence.

    Acceptance Criteria (Implied / Predicate)Reported Device Performance (Candidate Device)
    General Assay Features
    Intended Use/Indications for UseLargely similar indications, but explicitly mentions K2-EDTA and K3-EDTA, as well as Li-heparin plasma tubes with separating gel. Specific to cobas e 411 analyzer.
    Assay ProtocolSame (sandwich test principle)
    Detection ProtocolSame (Electrochemiluminescent Assay)
    ApplicationsSame (18 minute application)
    Instrument Platformcobas e 411 analyzer (Predicate: Elecsys 2010, cobas e 411, MODULAR Analytics E170, cobas e 601 and cobas e 602 immunoassay analyzers)
    Sample: Reagent Ratio20:70 μL (Predicate: 40:60 μL)
    Sample TypeHuman serum and Li-heparin, K2-EDTA and K3-EDTA, as well as Li-heparin plasma tubes containing separating gel (Predicate: Broader, including Na-NH4+-heparin, K2-EDTA, K3-EDTA, sodium citrate plasma).
    ReagentsSame
    CalibratorElecsys CA 125 II CalSet II (K140112) (Predicate: Elecsys CA 125 II CalSet (K003969))
    Calibration IntervalAfter 8 weeks when using the same reagent lot (Predicate: After 1 month (28 days)). Other conditions are similar.
    ControlsSame (Elecsys PreciControl Tumor Marker)
    Traceability/StandardizationSame (standardized against Enzymun-Test CA 125 II, which was standardized against CA 125 II RIA from Fujirebio Diagnostics).
    Reagent Stability (on analyzers)6 weeks (Predicate: 4 weeks)
    Labeled Performance Characteristics
    Measuring Range2 (LoQ) - 3000 U/mL (Predicate: 0.6 (LDL) - 5000 U/mL)
    Precision (cobas e411 analyzers)Intra-Assay/Within-run (Repeatability):
    3.1% CV @ 14.70 U/mL
    3.0% CV @ 3.07 U/mL
    2.6% CV @ 2399 U/mL
    1.9% CV @ 34.95 U/mL
    0.9% CV @ 120.7 U/mL
    1.1% CV @ 329.6 U/mL
    Total (Intermediate):
    4.1% CV @ 14.70 U/mL
    4.2% CV @ 3.07 U/mL
    3.4% CV @ 2399 U/mL
    3.0% CV @ 34.95 U/mL
    1.3% CV @ 120.7U/mL
    1.3% CV @ 329.6U/mL
    LoB0.6 U/mL (Predicate: Not Reported)
    LoD1.2 U/mL (Predicate: Not Reported)
    LoQ2 U/mL (Predicate: Not Reported)
    Lower Detection LimitN/A (Functional Sensitivity 0.6 U/mL for Predicate, but now explicit LoD/LoQ are reported for candidate).
    Performance Characteristics
    Hook EffectNo high-dose hook effect at CA 125 concentrations up to 50,000 U/mL (Predicate: up to 20,000 U/mL).
    Limitations (Interferences)Unaffected by: Hemolysis
    Ask a Question

    Ask a specific question about this device

    K Number
    K100903
    Device Name
    ANTI-HAV
    Date Cleared
    2010-10-05

    (187 days)

    Product Code
    Regulation Number
    866.3310
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ROCHE PROFESSIONAL DIAGNOSTICS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Roche Elecsys Anti-HAV immunoassay is used for the in vitro qualitative detection of total antibodies (IgM and IgG) to hepatitis A virus in human serum and plasma (K2-EDTA). The assay is intended for use as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection.

    Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with hepatitis A virus in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis A infection, or used as an aid to determine the presence of antibody response to HAV in vaccine recipients.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassav analyzers.

    Elecsys PreciControl Anti-HAV is used for the quality control of the Elecsys Anti-HAV immunoassay on the Elecsys and cobas e immunoassay analyzers.

    Device Description

    The Elecsys anti-HAV test is a qualitative assay based on electrochemiluminescence immunoassay "ECLIA" technology. The Elecsys anti-HAV test utilizes a competitive immunoassay format in which sample anti-HAV antibody competes with biotinylated and ruthenvlated anti-HAV monoclonal antibodies for a limited amount of cell culture-derived HAV antigen. The sample antibody and the HAV antigen react in the first incubation. The biotinylated antibodies and ruthenium complex® -labeled antibodies specific for HAV antigen are added in the second incubation together with streptavidin-coated magnetic microparticles. The unbound HAV antigen reacts with the modified antibodies and the resulting immune complexes are bound to the solid phase through a biotinstreptavidin interaction. If all HAV antigens are complexed by sample anti-HAV antibody during the first incubation, no modified/labeled immune complexes are formed and captured during the second incubation.

    Following the second incubation, the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by elution with ProCell. Application of a voltage to the electrode induces chemiluminescent emission from the captured immune complexes which is measured by a photomultiplier. The level of signal detected by the system decreases as the concentration of the anti-HAV antibody target present in a patient sample increases.

    Results are determined via a calibration curve which is generated by 2-point calibration on the instrument and a master curve provided via the reagent barcode. The calibration process converts the output so that low levels of sample anti-HAV antibodies are expressed by low output and high levels of antibody are expressed by high output. These outputs are finally interpreted on a qualitative basis around the established cut-off output.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Elecsys® Anti-HAV Assay, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the performance metrics reported, as the submission concludes that the information "supports a substantial equivalence decision." This means the reported performance met the FDA's expectations for equivalence to the predicate device.

    Performance MetricAcceptance Criteria (Implied/General)Reported Device Performance
    Analytical Performance
    PrecisionAcceptable %CV for repeatability and intermediate precision, typically within established guidelines (e.g., CLSI EP15-A2/EP5-A2).Elecsys 2010:
    • PC Anti-HAV 1: Repeatability CV 1.8%, Int. Precision CV 3.1%
    • PC Anti-HAV 2: Repeatability CV 3.2%, Int. Precision CV 4.5%
    • HSP 1, 2, 3: Repeatability CVs 2.5-3.4%, Int. Precision CVs 3.0-4.6%
      MODULAR ANALYTICS E170:
    • PC Anti-HAV 1: Repeatability CV 1.5%, Int. Precision CV 5.5%
    • PC Anti-HAV 2: Repeatability CV 2.3%, Int. Precision CV 5.8%
    • HSP 1, 2, 3: Repeatability CVs 1.1-3.5%, Int. Precision CVs 4.3-5.6%
      Reproducibility (Overall within-site and between-site CVs also provided, e.g., Overall PPA/NPA > 90%) |
      | Detection Limits (LoD) | Lowest amount of analyte detectable with 95% probability, set at 6.00 IU/L. | Elecsys 2010 / cobas e 411: LoD = 5.155 IU/L
      MODULAR ANALYTICS E170 / cobas e 601: LoD = 2.994 IU/L
      Reported in labeling: 6.00 IU/L for both types. |
      | Analytical Specificity (Cross-reactivity) | Minimal false positives with common interfering conditions/diseases. | 177 samples from 15 subgroups tested: 174 non-reactive, 3 discordant with predicate. No HAMA effect found. |
      | Analytical Specificity (Interference) | Recovery of positive samples within ± 20 % of initial value. | Unaffected by icterus (90-95%) with a legally marketed predicate device. | Overall PPA: 98.12% (96.46% to 99.14% CI) |
      | Overall Negative Percent Agreement (NPA) with Predicate | High agreement (e.g., >90-95%) with a legally marketed predicate device. | Overall NPA: 97.37% (95.70% to 98.52% CI) |
      | Seroconversion Sensitivity | Agreement with predicate and vendor data for seroconversion panels. | Elecsys Anti-HAV assay showed similar performance to comparator (Abbott AxSym HAV AB-2.0) based on post bleed day of earliest reactive/last positive result for 3 panels. |
      | Method Comparison (between platforms) | High PPA and NPA (e.g., >95%) between the Elecsys 2010 and MODULAR ANALYTICS E170. | PPA: 96.8% (91.0% to 99.3% CI)
      NPA: 98.0% (93.0% to 99.8% CI)
      Pearson's regression correlation: 0.9951 (slope 1.024, intercept -0.555). |

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Clinical Performance Study (Comparison with Predicate):
      • Total Samples: 1050 samples (after excluding one concordantly equivocal sample).
      • Cohort Breakdown:
        • Subjects for whom routine hepatitis A testing had been ordered.
        • Hospitalized patients.
        • Subjects at increased risk for hepatitis.
        • Subjects with signs and symptoms of hepatitis.
        • Subjects characterized with acute hepatitis A.
        • Subjects below the age of 21 years (pediatric/adolescents).
        • Specific numbers per cohort are detailed in the tables (e.g., 197 for Routine HAV Testing, 219 for Hospitalized, etc., summing to the overall total).
      • Data Provenance: Multi-center study conducted in the U.S. (across 3 sites: BW, WU, JH). Samples were a mix of prospective and retrospective collections. All samples stored frozen before shipment.
    • Analytical Specificity (Cross-reactivity): 177 samples from 15 potentially cross-reactive subgroups.
    • Analytical Specificity (Interference): Native human serum pools, and 18 commonly used pharmaceuticals and folic acid.
    • Prevalence Study (Expected Values): 602 patients (300 from high prevalence region, Western U.S. (New Mexico); 302 from low prevalence region, Eastern U.S. (Indiana)). This was a prospective study.
    • Seroconversion Sensitivity: 3 seroconversion panels.
    • Method Comparison (between platforms): 202 samples (100 non-reactive from prevalence cohort, 53 reactive from clinical cohorts, 49 reactive from Roche R&D archive).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not explicitly provided in the summary. The "ground truth" for clinical performance is established by comparing the Elecsys Anti-HAV assay results against an "FDA-cleared reference method" (the predicate device, Abbott AxSym HAVAB® 2.0). However, the summary does not specify if or how many human experts were involved in interpreting results from either the new device or the predicate to establish a "ground truth" in cases of discordance, or to confirm clinical status. The study relies on the predicate device's established performance and classification of samples.

    4. Adjudication Method for the Test Set

    The summary does not explicitly describe an adjudication method involving human experts for the test set. The clinical performance section states that the assay's performance was determined by agreement with the predicate device, which indicates a direct comparison rather than a human adjudication process to "resolve" discordant results. For "borderline" results from the Elecsys Anti-HAV assay (18.0

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1