K Number
K100903
Device Name
ANTI-HAV
Date Cleared
2010-10-05

(187 days)

Product Code
Regulation Number
866.3310
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Roche Elecsys Anti-HAV immunoassay is used for the in vitro qualitative detection of total antibodies (IgM and IgG) to hepatitis A virus in human serum and plasma (K2-EDTA). The assay is intended for use as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection. Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with hepatitis A virus in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis A infection, or used as an aid to determine the presence of antibody response to HAV in vaccine recipients. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassav analyzers. Elecsys PreciControl Anti-HAV is used for the quality control of the Elecsys Anti-HAV immunoassay on the Elecsys and cobas e immunoassay analyzers.
Device Description
The Elecsys anti-HAV test is a qualitative assay based on electrochemiluminescence immunoassay "ECLIA" technology. The Elecsys anti-HAV test utilizes a competitive immunoassay format in which sample anti-HAV antibody competes with biotinylated and ruthenvlated anti-HAV monoclonal antibodies for a limited amount of cell culture-derived HAV antigen. The sample antibody and the HAV antigen react in the first incubation. The biotinylated antibodies and ruthenium complex® -labeled antibodies specific for HAV antigen are added in the second incubation together with streptavidin-coated magnetic microparticles. The unbound HAV antigen reacts with the modified antibodies and the resulting immune complexes are bound to the solid phase through a biotinstreptavidin interaction. If all HAV antigens are complexed by sample anti-HAV antibody during the first incubation, no modified/labeled immune complexes are formed and captured during the second incubation. Following the second incubation, the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by elution with ProCell. Application of a voltage to the electrode induces chemiluminescent emission from the captured immune complexes which is measured by a photomultiplier. The level of signal detected by the system decreases as the concentration of the anti-HAV antibody target present in a patient sample increases. Results are determined via a calibration curve which is generated by 2-point calibration on the instrument and a master curve provided via the reagent barcode. The calibration process converts the output so that low levels of sample anti-HAV antibodies are expressed by low output and high levels of antibody are expressed by high output. These outputs are finally interpreted on a qualitative basis around the established cut-off output.
More Information

P780012/S009

K2-EDTA

No
The device description details a standard immunoassay based on electrochemiluminescence (ECLIA) technology. There is no mention of AI, ML, or any related computational techniques for data analysis, interpretation, or decision-making. The results are determined via a calibration curve and interpreted qualitatively around a cut-off, which is a standard immunoassay process.

No

Explanation: This device is an in vitro diagnostic (IVD) immunoassay designed for the qualitative detection of antibodies to hepatitis A virus, used as an aid in diagnosis. It does not provide any form of therapy or treatment.

Yes

The "Intended Use / Indications for Use" section explicitly states that the assay "is intended for use as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection." This clearly indicates its role in diagnosis.

No

The device description clearly outlines a physical immunoassay kit and its use on specific hardware analyzers (Elecsys and cobas e). It involves chemical reactions, magnetic microparticles, and detection of chemiluminescent emission, all of which are hardware-dependent processes.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The "Intended Use / Indications for Use" section explicitly states that the assay is used for the "in vitro qualitative detection of total antibodies (IgM and IgG) to hepatitis A virus in human serum and plasma". The term "in vitro" is a key indicator of an IVD, meaning it's used outside of a living organism.
  • Sample Type: The assay analyzes "human serum and plasma", which are biological specimens commonly used in IVD testing.
  • Purpose: The assay is intended "as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection". This clearly describes a diagnostic purpose performed in a laboratory setting.
  • Technology: The device utilizes an "electrochemiluminescence immunoassay 'ECLIA'", which is a common technology used in IVD assays to detect specific substances in biological samples.
  • Quality Control: The mention of "Elecsys PreciControl Anti-HAV is used for the quality control of the Elecsys Anti-HAV immunoassay" further supports its use as a diagnostic test requiring quality assurance.

All of these points align with the definition and characteristics of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Roche Elecsys Anti-HAV immunoassay is used for the in vitro qualitative detection of total antibodies (IgM and IgG) to hepatitis A virus in human serum and plasma (K2-EDTA). The assay is intended for use as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection.

Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with hepatitis A virus in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis A infection, or used as an aid to determine the presence of antibody response to HAV in vaccine recipients.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassav analyzers.

Elecsys PreciControl Anti-HAV is used for the quality control of the Elecsys Anti-HAV immunoassay on the Elecsys and cobas e immunoassay analyzers.

Product codes (comma separated list FDA assigned to the subject device)

LOL, JJX

Device Description

The Elecsys anti-HAV test is a qualitative assay based on electrochemiluminescence immunoassay "ECLIA" technology. The Elecsys anti-HAV test utilizes a competitive immunoassay format in which sample anti-HAV antibody competes with biotinylated and ruthenvlated anti-HAV monoclonal antibodies for a limited amount of cell culture-derived HAV antigen. The sample antibody and the HAV antigen react in the first incubation. The biotinylated antibodies and ruthenium complex® -labeled antibodies specific for HAV antigen are added in the second incubation together with streptavidin-coated magnetic microparticles. The unbound HAV antigen reacts with the modified antibodies and the resulting immune complexes are bound to the solid phase through a biotinstreptavidin interaction. If all HAV antigens are complexed by sample anti-HAV antibody during the first incubation, no modified/labeled immune complexes are formed and captured during the second incubation.

Following the second incubation, the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by elution with ProCell. Application of a voltage to the electrode induces chemiluminescent emission from the captured immune complexes which is measured by a photomultiplier. The level of signal detected by the system decreases as the concentration of the anti-HAV antibody target present in a patient sample increases.

Results are determined via a calibration curve which is generated by 2-point calibration on the instrument and a master curve provided via the reagent barcode. The calibration process converts the output so that low levels of sample anti-HAV antibodies are expressed by low output and high levels of antibody are expressed by high output. These outputs are finally interpreted on a qualitative basis around the established cut-off output.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Precision/Reproducibility:
Precision and Reproducibility of the Elecsys Anti-HAV assay was evaluated on Elecsys 2010 analyzer and MODULAR ANALYTICS E170 analyzer using Elecsys reagents, three human sera and controls, according to the CLSI EP15-A2/EP5-A2.
For precision study, three human sera (high negative, low positive and moderately positive) and controls (PreciControl Anti-HAV 1 and PreciControl Anti-HAV 2) were tested in replicates of 2 in 2 runs per day for 20 days with one reagent lot spanning at least two cycles of calibration using one Elecsys 2010 analyzer and one Modular Analytics E170 analyzer.
For reproducibility study, three human sera (high negative, low positive and moderately positive) and controls (PreciControl Anti-HAV 1 and PreciControl Anti-HAV 2) were tested in replicates of 3 in 2 runs per day for 5 days with three Elecsys 2010 analyzers and three Modular Analytics E170 analyzers.

Linearity/Assay Reportable Range:
The detectable range is from 6 IU/L to 60 IU/L.

Analytical Specificity:
Cross-reactivity: 177 samples from 15 potentially cross-reactive sub-groups were used for cross-reactivity study on the Elecsys 2010. 174 Samples were found to be nonreactive (negative) in both the Elecsys anti-HAV assay and the predicate AxSym HAVAB 2.0 assay; 3 samples were found to be discordant between the Elecsys anti-HAV assay and the AxSym HAVAB 2.0 assay.
Interference: The impact of endogenous interfering substances on the Elecsys Anti-HAV assay was determined testing native human serum pools on Elecsys® 2010 Immunoassay Analyzer. The assay is unaffected by icterus (bilirubin

§ 866.3310 Hepatitis A virus (HAV) serological assays.

(a)
Identification. HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.(b)
Classification. Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.

0

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

  • A. 510(k) Number: K100903
  • B. Purpose for Submission: Clearance of New Device
  • C. Measurand: Total Antibody to Hepatitis A Virus in serum and heparinized plasma
  • D. Type of Test: Electrochemiluminescence immunoassay (ECLIA)
  • E. Applicant: Roche Diagnostics
  • F. Proprietary and Established Names: Elecsys® Anti-HAV Assay Elecsys® PreciControl anti-HAV Assay

G. Regulatory Information:

| 1. Regulation section: | 21 CFR §866.3310, Hepatitis A virus Serological Assays
21 CFR §862.1660, Quality Control Material |
|------------------------|-------------------------------------------------------------------------------------------------------------|
| 2. Classification: | Class II |
| 3. Product code: | LOL (Hepatitis A Test – antibody and IgM Antibody)
JJX (Quality control material, assayed and unassayed) |
| 4. Panel: | Microbiology (83) |

H. Intended Use:

    1. Intended use(s):
      The Roche Elecsys Anti-HAV immunoassay is used for the in vitro qualitative detection of total antibodies (IgM and IgG) to hepatitis A virus in human serum and plasma (K2-EDTA). The assay is intended for use as an aid in the laboratory diagnosis of past or acute/recent hepatitis A infection.

Assay results, in conjunction with other laboratory results and clinical information, may be used to provide presumptive evidence of infection with hepatitis A virus in persons with signs or symptoms of hepatitis and in persons at risk for hepatitis A infection, or used as an aid to determine the presence of antibody response to HAV in vaccine recipients.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassav analyzers.

Elecsys PreciControl Anti-HAV is used for the quality control of the Elecsys Anti-HAV immunoassay on the Elecsys and cobas e immunoassay analyzers.

1

This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The users are responsible for establishing their own assay performance characteristics in these populations.

Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician.

    1. Indication(s) for use:
      Same as Intended Use
    1. Special conditions for use statement(s):
      For prescription use only
    1. Special instrument requirements:
      Elecsys 2010 and cobas e 411 analyzers; MODULAR ANALYTICS E170 and cobas e 601 analyzers

I. Device Description:

The Elecsys anti-HAV test is a qualitative assay based on electrochemiluminescence immunoassay "ECLIA" technology. The Elecsys anti-HAV test utilizes a competitive immunoassay format in which sample anti-HAV antibody competes with biotinylated and ruthenvlated anti-HAV monoclonal antibodies for a limited amount of cell culture-derived HAV antigen. The sample antibody and the HAV antigen react in the first incubation. The biotinylated antibodies and ruthenium complex® -labeled antibodies specific for HAV antigen are added in the second incubation together with streptavidin-coated magnetic microparticles. The unbound HAV antigen reacts with the modified antibodies and the resulting immune complexes are bound to the solid phase through a biotinstreptavidin interaction. If all HAV antigens are complexed by sample anti-HAV antibody during the first incubation, no modified/labeled immune complexes are formed and captured during the second incubation.

Following the second incubation, the reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are removed by elution with ProCell. Application of a voltage to the electrode induces chemiluminescent emission from the captured immune complexes which is measured by a photomultiplier. The level of signal detected by the system decreases as the concentration of the anti-HAV antibody target present in a patient sample increases.

2

Results are determined via a calibration curve which is generated by 2-point calibration on the instrument and a master curve provided via the reagent barcode. The calibration process converts the output so that low levels of sample anti-HAV antibodies are expressed by low output and high levels of antibody are expressed by high output. These outputs are finally interpreted on a qualitative basis around the established cut-off output.

a) Tris(2,2'-bipyridyl)ruthenium(II)-complex (Ru(bpy)) 3 2+

J. Substantial Equivalence Information:

    1. Predicate device name(s): Abbott AxSym HAVAB® 2.0
  • Predicate Numbers (s): 2.
    1. Comparison with predicate:
Similarities
Anti-HAV Immunoassay Comparison
FeatureElecsys Anti-HAV AssayPredicate Device
Abbott Axsym HAVAB 2.0
Assay (P780012/S009)
Intended UseThe Roche Elecsys Anti-HAV
immunoassay is used for the in vitro
qualitative detection of total antibodies
(IgM and IgG) to hepatitis A virus in
human serum and plasma (K2-EDTA).
The assay is intended for use as an aid in
the laboratory diagnosis of past or
acute/recent hepatitis A infection.
Assay results, in conjunction with other
laboratory results and clinical information,
may be used to provide presumptive
evidence of infection with hepatitis A
virus in persons with signs or symptoms
of hepatitis and in persons at risk for
hepatitis A infection, or used as an aid to
determine the presence of antibody
response to HAV in vaccine recipients.
The electrochemiluminescence
immunoassay “ECLIA” is intended for
use on Elecsys and cobas e immunoassay
analyzers.Immunoassay for the qualitative
detection of total antibody to
hepatitis A virus (anti-HAV) in
human serum or plasma
(potassium EDTA, sodium
heparin, sodium citrate, or
lithium heparin). A test for anti-
HAV is indicated as an aid in the
laboratory diagnosis of previous
or ongoing hepatitis A viral
infection or in the identification
of HAV-susceptible individuals
for vaccination.
Assay results, in conjunction
with other laboratory results and
clinical information, may be
used to provide presumptive
evidence of infection with
hepatitis A virus in persons with
signs or symptoms of hepatitis
and in persons at risk for
hepatitis A infection.
Indications for
UseSameSame
Sample TypeHuman serum and plasmaSame
Anti-HAV Immunoassay Comparison
FeaturesElecsys Anti-HAV AssayPredicate Device
Abbott Axsym HAVAB 2.0 Assay
(P780012/S009)
Detection
ProtocolElectrochemiluminescence
immunoassay (ECLIA)Microparticle Enzyme Immunoassay
(MEIA)
Traceability/
StandardizationStandardized against the "Second
International Standard for Anti-
Hepatitis A, Immunoglobulin,
Human, NIBSC code: 97/646" of
the NIBSC (National Institute for
Biological Standards and Control).Not Given
Interpretation of
Results≥ 22.0 IU/L Reactive
18.0 ≤ IU/L3.000 Not applicable
Calibration
IntervalOnce per reagent lot and
• After 1 month (28 days) when
using the same reagent lot
• After 7 days (when using the
same reagent kit on the analyzer)
• As required: e.g., quality control
findings outside the specified
limitsA minimum of two replicates of the
AxSYM HAFVAB 2.0 Index Calibrator
must be tested. A single sample of both
the Negative and Positive Controls must
be tested as a means of evaluating the
assay calibration. Once the calibration is
accepted and stored, all subsequent
samples may be tested without further
calibration unless one or more of the
following occur:
• A reagent pack with a new lot number
is used
• Either of the AxSYM HAVAB 2.0
Control values is out of its specified
range
• The MEIA Optics Verification Update
has been performed
ControlsElecsys PreciControl Anti-HAVAbbott AxSYM HAVAB 2.0 Controls

3

Differences

K. Standard/Guidance Document referenced (if applicable):

CLSI EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods"

4

CLSI EP17-A, "Protocols for Determination of Limits of Detection" Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guidan ceDocuments/ucm071055.pdf

L. Test Principle:

The Elecsys Anti-HAV immunoassay utilizes a competitive immunoassay format employing cell culture derived HAV antigen and a monoclonal antibody (modified by biotinylation or ruthenylation) in competition to the polyclonal sample antibody. Capture of formed immune complexes from the reaction mixture is based on biotin in the immune complex binding to streptavidin-coated magnetic microparticles which are collected on a measuring cell electrode. Signal generation is triggered by the application of a voltage to the electrode (electrochemiluminescence technology). The level of signal count detected by the system decreases as the concentration of the anti-HAV total antibody target present in a patient sample increases. Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent barcode

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Precision and Reproducibility of the Elecsys Anti-HAV assay was evaluated on Elecsys 2010 analyzer and MODULAR ANALYTICS E170 analyzer using Elecsys reagents, three human sera and controls, according to the CLSI EP15-A2/EP5-A2.

For precision study, three human sera (high negative, low positive and moderately positive) and controls (PreciControl Anti-HAV 1 and PreciControl Anti-HAV 2) were tested in replicates of 2 in 2 runs per day for 20 days with one reagent lot spanning at least two cycles of calibration using one Elecsys 2010 analyzer and one Modular Analytics E170 analyzer. Results are presented below.

| Sample | Mean
(IU/L) | Repeatability | | Intermediate Precision
(within-laboratory
precision) | | n |
|---------------|----------------|---------------|---------------------|------------------------------------------------------------|---------------------|----|
| | | SD
(IU/L) | CV (%)
(UCL 95%) | SD
(IU/L) | CV (%)
(UCL 95%) | |
| PC Anti-HAV 1 | 12.3 | 0.221 | 1.8 (2.2) | 0.378 | 3.1 (3.7) | 80 |
| PC Anti-HAV 2 | 29.2 | 0.947 | 3.2 (4.0) | 1.325 | 4.5 (5.3) | 80 |
| HSP 1 | 17.6 | 0.534 | 3.0 (3.7) | 0.586 | 3.3 (3.8) | 80 |
| HSP 2 | 21.9 | 0.538 | 2.5 (3.0) | 0.667 | 3.0 (3.6) | 80 |
| HSP3 | 49.3 | 1.655 | 3.4 (4.1) | 2.261 | 4.6 (5.4) | 80 |

  • Precision on Elecsys® 2010 ●

5

| Sample | Mean
(IU/L) | Repeatability | | Intermediate Precision
(within-laboratory
precision) | | |
|---------------|----------------|---------------|---------------------|------------------------------------------------------------|---------------------|----|
| | | SD
(IU/L) | CV (%)
(UCL 95%) | SD
(IU/L) | CV (%)
(UCL 95%) | n |
| PC Anti-HAV 1 | 12.54 | 0.188 | 1.5 (1.8) | 0.688 | 5.5 (7.1) | 80 |
| PC Anti-HAV 2 | 32.12 | 0.724 | 2.3 (2.8) | 1.849 | 5.8 (7.4) | 80 |
| HSP 1 | 18.2 | 0.459 | 2.5 (3.1) | 0.816 | 4.5 (5.6) | 80 |
| HSP 2 | 23 | 0.254 | 1.1 (1.4) | 1.016 | 4.4 (5.9) | 80 |
| HSP3 | 55.37 | 1.088 | 2.0 (2.4) | 2.374 | 4.3 (5.3) | 80 |

Precision on MODULAR® ANALYTICS E170 .

PC = PreControl

UCL = Upper Confidence Limit

For reproducibility study, three human sera (high negative, low positive and moderately positive) and controls (PreciControl Anti-HAV 1 and PreciControl Anti-HAV 2) were tested in replicates of 3 in 2 runs per day for 5 days with three Elecsys 2010 analyzers and three Modular Analytics E170 analyzers. Results are presented below.

| Clinical | Sample | N | Mean
IU/L | Repeatability | | Between Run | | Between Day | | Within Site
Reproducibility | |
|----------|---------------|----|--------------|---------------|------|-------------|------|-------------|------|--------------------------------|------|
| Site | | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Site 1 | HSP 01 | 30 | 22.629 | 0.523 | 2.3 | 0.275 | 1.2 | 0.176 | 0.8 | 0.616 | 2.7 |
| | HSP 02 | 30 | 49.903 | 2.469 | 4.9 | 0.000 | 0.0 | 0.123 | 0.2 | 2.472 | 5.0 |
| | HSP 03 | 30 | 18.287 | 0.527 | 2.9 | 0.052 | 0.3 | 0.264 | 1.4 | 0.592 | 3.2 |
| | PC A-
HAV1 | 30 | 12.678 | 0.321 | 2.5 | 0.000 | 0.0 | 0.201 | 1.6 | 0.379 | 3.0 |
| | PC A-
HAV2 | 30 | 29.198 | 0.707 | 2.4 | 0.516 | 1.8 | 0.505 | 1.7 | 1.011 | 3.5 |
| Site 2 | HSP 01 | 30 | 23.196 | 0.302 | 1.3 | 0.452 | 2.0 | 0.000 | 0.0 | 0.544 | 2.3 |
| | HSP 02 | 30 | 50.965 | 0.688 | 1.4 | 0.55 | 1.1 | 0.000 | 0.0 | 0.881 | 1.7 |
| | HSP 03 | 30 | 18.499 | 0.417 | 2.3 | 0.336 | 1.8 | 0.000 | 0.0 | 0.535 | 2.9 |
| | PC A-
HAV1 | 30 | 12.942 | 0.278 | 2.2 | 0.309 | 2.4 | 0.000 | 0.0 | 0.416 | 3.2 |
| | PC A-
HAV2 | 30 | 30.814 | 0.376 | 1.2 | 0.441 | 1.4 | 0.000 | 0.0 | 0.58 | 1.9 |
| Site 3 | HSP 01 | 30 | 22.634 | 0.371 | 1.6 | 0.000 | 0.0 | 0.266 | 1.2 | 0.456 | 2.0 |
| | HSP 02 | 30 | 48.03 | 1.356 | 2.8 | 1.451 | 3 | 0.000 | 0.0 | 1.985 | 4.1 |
| | HSP 03 | 30 | 17.892 | 0.398 | 2.2 | 0.268 | 1.5 | 0.292 | 1.6 | 0.561 | 3.1 |
| | PC A-
HAV1 | 30 | 12.515 | 0.248 | 2.0 | 0.106 | 0.9 | 0.372 | 3.0 | 0.459 | 3.7 |
| | PC A-
HAV2 | 30 | 29.46 | 1.144 | 3.9 | 0.962 | 3.3 | 0.503 | 1.7 | 1.578 | 5.4 |

Elecsys anti-HAV: Within Site Reproducibility: Elecsys® 2010 Analyzer

6

| Clinical
Site | Sample | N | Mean
IU/L | Repeatability | | Between Run | | Between Day | | Within Site
Reproducibility | |
|------------------|---------------|----|--------------|---------------|------|-------------|------|-------------|------|--------------------------------|------|
| | | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| Site 1 | HSP 01 | 30 | 23.433 | 0.237 | 1.0 | 0.419 | 1.8 | 0.000 | 0.0 | 0.481 | 2.1 |
| | HSP 02 | 30 | 51.725 | 0.454 | 0.9 | 0.707 | 1.4 | 0.000 | 0.0 | 0.84 | 1.6 |
| | HSP 03 | 30 | 18.918 | 0.158 | 0.8 | 0.167 | 0.9 | 0.135 | 0.7 | 0.266 | 1.4 |
| | PC A-
HAV1 | 30 | 13.205 | 0.197 | 1.5 | 0.211 | 1.6 | 0.132 | 1.0 | 0.317 | 2.4 |
| | PC A-
HAV2 | 30 | 31.477 | 0.339 | 1.1 | 0.787 | 2.5 | 0.000 | 0.0 | 0.857 | 2.7 |
| Site 2 | HSP 01 | 30 | 22.541 | 0.705 | 3.1 | 0.090 | 0.4 | 0.231 | 1.0 | 0.747 | 3.3 |
| | HSP 02 | 30 | 51.871 | 1.344 | 2.6 | 0.000 | 0.0 | 0.302 | 0.6 | 1.378 | 2.7 |
| | HSP 03 | 30 | 18.455 | 0.642 | 3.5 | 0.000 | 0.0 | 0.355 | 1.9 | 0.733 | 4 |
| | PC A-
HAV1 | 30 | 12.676 | 0.695 | 5.5 | 0.000 | 0.0 | 0.627 | 4.9 | 0.936 | 7.4 |
| | PC A-
HAV2 | 30 | 31.438 | 0.733 | 2.3 | 0.231 | 0.7 | 0.000 | 0.0 | 0.769 | 2.4 |
| Site 3 | HSP 01 | 30 | 23.946 | 0.288 | 1.2 | 0.22 | 0.9 | 0.000 | 0.0 | 0.363 | 1.5 |
| | HSP 02 | 30 | 52.166 | 0.299 | 0.6 | 0.829 | 1.6 | 0.000 | 0.0 | 0.881 | 1.7 |
| | HSP 03 | 30 | 19.426 | 0.203 | 1.0 | 0.336 | 1.7 | 0.000 | 0.0 | 0.392 | 2.0 |
| | PC A-
HAV1 | 30 | 12.974 | 0.205 | 1.6 | 0.225 | 1.7 | 0.000 | 0.0 | 0.304 | 2.3 |
| | PC A-
HAV2 | 30 | 31.236 | 0.241 | 0.8 | 1.038 | 3.3 | 0.000 | 0.0 | 1.066 | 3.4 |

Elecsys anti-HAV: Within Site Reproducibility: MODULAR® ANALYTICS E170

Elecsys anti-HAV: Between Site Reproducibility: Elecsys® 2010 Analyzer

| | N | Mean
IU/L | Repeatability | | Between Run | | Between Day | | Between Site | | Reproducibility | |
|---------------|----|--------------|---------------|------|-------------|------|-------------|------|--------------|------|-----------------|------|
| Sample | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| HSP 01 | 90 | 22.819 | 0.409 | 1.8 | 0.3 | 1.3 | 0.000 | 0.0 | 0.304 | 1.3 | 0.591 | 2.6 |
| HSP 02 | 90 | 49.633 | 1.674 | 3.4 | 0.697 | 1.4 | 0.000 | 0.0 | 1.462 | 2.9 | 2.33 | 4.7 |
| HSP 03 | 90 | 18.226 | 0.451 | 2.5 | 0.25 | 1.4 | 0.166 | 0.9 | 0.276 | 1.5 | 0.608 | 3.3 |
| PC A-
HAV1 | 90 | 12.712 | 0.284 | 2.2 | 0.179 | 1.4 | 0.22 | 1.7 | 0.176 | 1.4 | 0.438 | 3.4 |
| PC A-
HAV2 | 90 | 29.824 | 0.807 | 2.7 | 0.68 | 2.3 | 0.394 | 1.3 | 0.808 | 2.7 | 1.386 | 4.6 |

7

| Sample | N | Mean
IU/L | Repeatability | | Between Run | | Between Day | | Between Site | | Reproducibility | |
|---------------|----|--------------|---------------|------|-------------|------|-------------|------|--------------|------|-----------------|------|
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| HSP 01 | 90 | 23.307 | 0.46 | 2 | 0.278 | 1.2 | 0.000c | 0.0 | 0.7 | 3 | 0.883 | 3.8 |
| HSP 02 | 90 | 51.921 | 0.837 | 1.6 | 0.542 | 1 | 0.000c | 0.0 | 0.168 | 0.3 | 1.011 | 1.9 |
| HSP 03 | 90 | 18.933 | 0.399 | 2.1 | 0.148 | 0.8 | 0.17 | 0.9 | 0.472 | 2.5 | 0.658 | 3.5 |
| PC A-
HAV1 | 90 | 12.952 | 0.433 | 3.3 | 0.000c | 0 | 0.368 | 2.8 | 0.193 | 1.5 | 0.6 | 4.6 |
| PC A-
HAV2 | 90 | 31.384 | 0.487 | 1.6 | 0.764 | 2.4 | 0.000c | 0 | 0.000c | 0 | 0.906 | 2.9 |

Elecsys anti-HAV: Between Site Reproducibility: MODULAR® ANALYTICS E170

b. Linearity/assay reportable range:

The qualitative assay has a quantitative measurement. The detectable range is from 6 IU/L to 60 IU/L.

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.
    Traceability: This method has been standardized against the "Second International Standard for Anti-Hepatitis A, Immunoglobulin, Human, NIBSC code: 97/646" of the NIBSC (National Institute for Biological Standards and Control).

Controls:

PC A-HAV 1: anti-HAV antibodies (human), with approximately 12 IU/L

PC A-HAV 2: anti-HAV antibodies (human), with approximately 30 IU/L

Calibrators:

Calibration of the Elecsys anti-HAV is performed by using the 2 calibrators provided with the reagent kit. Calibrator 1 is an anti-HAV negative serum; Calibrator 2 is an anti-HAV positive serum with approx. 46 IU/L.

A reagent lot specific master standard curve is provided on the barcode of each reagent kit. The calibrator master standard curve is established by diluting an anti-HAV positive human serum with an analyte free human serum matrix. The concentration value was assigned to it by reading the concentration from the reference calibrator curve using four E2010 and four E170 instruments with 2 runs and 2 replicates per run and per instrument. The master curve is stored in the barcode and is automatically read in by the instrument when the barcode is scanned for the new reagent.

The calibration must be performed once per reagent lot. Acceptance criteria for the assigned values are 40 IU/L for Cal 2. Recalibration is recommended if the same reagent lot was used more than 1 month (28 days), or if the same reagent kit has been left on the analyzer for 7 days.

8

Stability:

Reagents:
-----------
unopened at 2-8 °Cup to the stated expiration date
M, R1, R2 after opening at 2-8 °C8 weeks
on Elecsys® 2010 and cobas e® 411
at
20-25 °C8 weeks
on MODULAR® ANALYTICS E170 and cobas e® 601 at 20-25 °C
Reconstituted calibrators at -20 °C12 weeks
Reconstituted calibrators at 2-8 °C2 weeks
Reconstituted calibrators on
Elecsys® 2010 and cobas e® 411 at
20-25 °Cup to 6 hours
Reconstituted calibrators on
MODULAR® ANALYTICS E170
and cobas e® 601 at 32°C

PreciControls:

unopened at 2-8 °Cup to the stated expiration date
after opening at 2-8 °C8 weeks
After opening at -20 °C3 months (freeze only once)
on Elecsys® 2010 and cobas e® 411
at 20-25 °Cup to 7 hours
on MODULAR® ANALYTICS
E170 and cobas e® 601 at 20-25 °Cup to 7 hours

Samples:

Sample stability was determined using different sample materials e.g. Serum, Serum with separating gel, K2-EDTA Plasma. Four negative samples, one sample around cutoff, and three low positive samples were tested and compared with fresh reference materials.

Samples stored at -20°Cup to 3 months
Samples stored at 2-8°CUp to 7 days
Freeze and thaw cyclesUp to 5 cycles
on Elecsys® 2010 and cobas e® 411
at 20-25 °Cup to 8 hours

9

d. Detection limits:

Limit of Blank (LoB) of the Elecsys® Anti-HAV assay was determined according to CLSI EP17-A. Limit of Blank determines the highest observed measurement values for samples free of analyte. The LoB was determined as the 95 percentile of measurements of blank samples. The distribution of values for five zero-level human serum has been determined on two Elecsys® 2010 / cobas e 411 Analyzers and on two MODULAR® ANALYTICS E170 / cobas e® 601 Immunoassay Analyzers over 3 days, 2 runs per day. The LoB for Elecsys® 2010 / cobas e 411 Analyzer is 3.03 IU/L and 2.099 IU/L for MODULAR® ANALYTICS E170 / cobas e® 601 Immunoassay Analyzers.

Limit of detection (LoD) of the Elecsys® Anti-HAV assay was determined according to CLSI EP17-A. The LoD was determined as the lowest amount of analyte in a sample that can be detected with 95% probability. The distribution of values for five low-level human serum samples has been determined on two Elecsys® 2010 / cobas e® 411 Analyzers and on two MODULAR® ANALYTICS E170 / cobas e 601 Immunoassay Analyzers over 3 days, 2 runs per day. Samples were measured in one-fold determination in each run. 30 measuring points were collected per instrument.

LOD = LOB + 1.6529 x SD total (of low analyte samples) The LoD for Elecsys® 2010 / cobas e® 411 Analyzer is 5.155 IU/L, and 2.994 IU/L for MODULAR® ANALYTICS E170 / cobas e® 601 Immunoassay Analyzers.

The LOD is set at 6.00 IU/L for both types of analyzers and is reported in the labeling.

Calibrators: Range for the electrochemiluminescence signals for the calibrators:

Negative calibrator (Call): 5 mg/day), no sample should be taken until at least 8 hours after the last biotin administration. One aliquot of each serum sample was spiked with the interfering substance; another aliquot was spiked with the same volume of isotonic NaCl solution (dilution pool). The interfering pool was then diluted into the dilution pool in 10% increments. The % recovery was determined by dividing the mean value of the measured concentration by the expected concentration. Criterion: Recovery of positive samples within ± 20 % of initial value.

In vitro tests were performed on 18 commonly used pharmaceuticals (Acetylcysteine, Ampicillin-Na, Ascorbic acid. Ca-Dobesilate, Cyclosporine, Cefoxitin, Heparin, Intralipid, Levodopa, Methyldopa, Metronidazole, Phenylbutazone, Tetracycline, Acetylsalicylic acid, Rifampicin, Acetaminophen, Ibuprofen, and Theophylline) and in addition on folic acid. No interference with the assay was found.

  • f. Assay cut-off:
    The cutoff was set at 20 IU/L according to the WHO recommendation for vaccination. It was established by histogram analysis of the predicate and was confirmed by measuring a total of 736 samples from 2 European centers including hospitalized patients, acute HAV infection patient, blood doors, pregnant women and the other samples that suspected for HAV infection. The 20 IU/L in the Elecsys anti-HAV assay corresponds to 1.0 of S/CO in the predicate.

Non-reactive: 22.0 IU/L

Samples with a concentration 15% |
| K2-EDTA | 24 | 1 | 0 |
| Serum Gel | 23 | 2 | 0 |

| Plasma matrix | Number of borderline specimens showing recovery
to serum within various ranges | | |
|---------------|-----------------------------------------------------------------------------------|--------|-------|
| | 15% |
| K2-EDTA | 3 | 0 | 0 |
| Serum Gel | 3 | 0 | 0 |

| Plasma matrix | Number of negative specimens showing recovery
to serum within various ranges | | |
|---------------|---------------------------------------------------------------------------------|---------------|------------|
| | 2.0 IU/L |
| K2-EDTA | 29 | 1 | 0 |
| Serum Gel | 30 | 0 | 0 |

13

3. Clinical studies:

  • a. Clinical Sensitivity: N/A
  • b. Clinical specificity: N/A
  • c. Other clinical supportive data (when a. and b. are not applicable):

Clinical Performance

Clinical Study Cohorts:

A multi-center study was conducted in the U.S. to characterize the performance of the Elecsys Anti-HAV Immunoassay. All subjects were tested with the Elecsys Anti-HAV assay on the Elecsys 2010 analyzer and with an FDAcleared reference method in strict accordance with the manufacturer's package insert instructions.

A total of 1050 samples were obtained from multiple specimen sources, representing subjects for whom routine hepatitis A testing had been ordered. hospitalized patients, subjects at increased risk for hepatitis, subjects with signs and symptoms of hepatitis, subjects characterized with acute hepatitis A, and subjects below the age of 21 years (pediatric/adolescents). All samples (prospective and retrospective) are stored frozen before shipment to Roche and to the respective sites for testing. Among the 1050 samples, 487 were (46.38%) were from males and 563 (53.62%) were from females. 0.38% were from American Indian/Alaska Native, 2.76% were Asian, 15.90% were African American /Black, 65.14% were Caucasian/White, 14.95% unknown and 0.86% others.

Among all specimens tested, one specimen subjected to routine Hepatitis A testing was concordantly equivocal, and was excluded from the calculations.

The positive and negative percent agreement for prospectively collect samples is 98.13% (95.96% to 99.31%) and 97.37% (95.06% to 98.79%), respectively.

The positive and negative percent agreement for retrospectively collect samples is 98.11% (94.59% to 99.61%) and 97.37% (94.36% to 99.03%), respectively.

The positive percent agreement and the negative percent agreement results for the different clinical population are presented in the following table:

14

| Elecsys

2010 ResultPredicate Result
Site 1 BWSite 2 WUSite 3 JHAll Sites
RXEQNRRXEQNRRXEQNRRXEQNR
Reactive600200016007602
Equivocal102000100202
Negative0097000001700114
Total61010100017017780118
PPA98.36 (60/61)0.00 (0/0)94.12 (16/17)97.44 (76/78)
95% CI91.20 to 99.960.00 to 100.0071.31 to 99.8591.04 to 99.69
NPA96.04 (97/101)0.00 (0/0)100.00 (17/17)96.61 (114/118)
95% CI90.17 to 98.910.00 to 100.0080.49 to 100.0091.55 to 99.07

Percent Agreement for Elecsys and Predicate anti-HAV Assay Results from Symptomatic Individuals

Percent Agreement for Elecsys and Predicate anti-HAV Assay Results from Specimens Subjected to Routine Hepatitis A Testing

Predicate Result
Elecsys
2010 ResultSite 1 BWSite 2 WUSite 3 JHAll Sites
RXEQNRRXEQNRRXEQNRRXEQNR
Reactive9002000360112603
Equivocal001000011012
Negative205400000212075
Total9205700036123128180
PPA97.83 (90/92)0.00 (0/0)100.00 (36/36)98.44 (126/128)
95% CI92.37 to 99.740.00 to 100.0090.26 to 100.0094.47 to 99.81
NPA94.74 (54/57)0.00 (0/0)91.30 (21/23)93.57 (75/80)
Percent Agreement for Elecsys and Predicate anti-HAV Assay Results from
Hospitalized Patients (Retrospective and Prospective)
Predicate Result
Site 1 BWSite 2 WUSite 3 JHAll Sites
Elecsys
2010 ResultRXEQNRRXEQNRRXEQNRRXEQNR
Reactive420000028007000
Equivocal000000100100
Negative0090000105110141

15

Total4209000030051720141
PPA100.00 (42/42)0.00 (0/0)93.33 (28/30)97.22 (70/72)
95% CI91.59 to 100.000.00 to 100.0077.93 to 99.1890.32 to 99.66
NPA100.00 (90/90)0.00 (0/0)100.00 (51/51)100.00 (141/141)
95% CI95.98 to 100.000.00 to 100.0093.02 to 100.0097.42 to 100.00

Percent Agreement for Elecsys and Predicate anti-HAV Assay Results from Subjects Characterized as Acute Hepatitis A (Retrospective and Prospective)

Predicate Result
Site 1 BWSite 2 WUSite 3 JHAll Sites
Elecsys 2010 ResultRXEQNRRXEQNRRXEQNRRXEQNR
Reactive8100000500013100
Equivocal000000000000
Negative003000000003
Total8103000500013103
PPA100.00 (81/81)0.00 (0/0)100.00 (50/50)100.00 (131/131)
95% CI95.55 to 100.000.00 to 100.0092.89 to 100.0097.22 to 100.00
NPA100.00 (3/3)0.00 (0/0)0.00 (0/0)100.00 (3/3)
95% CI29.24 to 100.000.00 to 100.000.00 to 100.0029.24 to 100.00

Percent Agreement for Elecsys and Predicate anti-HAV Assay Results from Pediatric and Adolescent Subjects

Predicate Result
Site 1 BWSite 2 WUSite 3 JHAll Sites
Elecsys
2010 ResultRXEQNRRXEQNRRXEQNRRXEQNR
Reactive00000019021902
Equivocal000000000000
Negative00000000650065
Total0000001906719067
PPA0.00 (0/0)0.00 (0/0)100.00 (19/19)100.00 (19/19)
95% CI0.00 to 100.000.00 to 100.0082.35 to 100.0082.35 to 100.00
NPA0.00 (0/0)0.00 (0/0)97.01 (65/67)97.01 (65/67)
95% CI0.00 to 100.000.00 to 100.0089.63 to 99.6489.63 to 99.64

16

Predicate Result
Site 1 BWSite 2 WUSite 3 JHAll Sites
Elecsys
2010 ResultRXEQNRRXEQNRRXEQNRRXEQNR
Reactive100100038024803
Equivocal000000001001
Negative1050002015230157
Total1106000400155510161
PPA90.91 (10/11)0.00 (0/0)95.00 (38/40)94.12 (48/51)
95% CI58.72 to 99.770.00 to 100.0083.08 to 99.3983.76 to 98.77
NPA83.33 (5/6)0.00 (0/0)98.06 (152/155)97.52 (157/161)
95% CI35.88 to 99.580.00 to 100.0094.45 to 99.6093.76 to 99.32

Agreement for Elecsys and Predicate anti-HAV Assay Results from Subjects at Increased Risk for Hepatitis

RX - Reactive, positive, EQ - equivocal, borderline, NR - non-reactive, negative, PPA - positive percent agreement, NPA negative percent agreement, 95% CI - 95% Exact confidence interval

CohortPositive Percent AgreementNegative Percent Agreement
PPA (x/n)95 % confidence intervalNPA (x/n)95 % confidence interval
Routine HAV Testing98.44 (126/128)94.47 to 99.8193.75 (75/80)86.01 to 97.94
Hospitalized97.22 (70/72)90.32 to 99.66100.00 (141/141)97.42 to 100.00
Signs and Symptoms97.44 (76/78)91.04 to 99.6996.61 (114/118)91.55 to 99.07
Increased Risk for Hepatitis94.12 (48/51)83.76 to 98.7797.52 (157/161)93.76 to 99.32
Characterized Acute HAV100.00 (131/131)97.22 to 100.00100.00 (3/3)29.24 to 100.00
Pediatric/Adolescent100.00 (19/19)82.35 to 100.097.01 (65/67)89.63 to 99.64
Overall98.12 (470/479)96.46 to 99.1497.37 (555/570)95.70 to 98.52

Summary of the Percent Agreements for the Various Specimen Cohorts

HAV Vaccination:

Fifty-four specimens that were collected both pre- and post-HAV vaccination was evaluated by the Elecsys anti-HAV assay and the predicate. The post-vaccination specimens were obtained at least four weeks but not more than 10 weeks after the completion of the vaccine regimen. Three HAV vaccines which are currently licensed in the US were used in this study: VAQTA (16), HAVRIX (20) and TWINRIX (18). Two vaccination studies are represented: one conducted in the Eastern region of the US and the other in Penzberg Germany. There was no antibody response observed in pre-vaccination specimens with the exception of two

17

specimens which were just above the equivocal/borderline (- 23.23 IU/L) and just reactive (30.34 IU/L), respectively.

anti-HAV Assay Results
Predicate Result
Elecsys
ResultHAVRIXTWINRIXVAQTAAll Vaccines
RXEQNRRXEQNRRXEQNRRXEQNR
Reactive000001000001
Equivocal000000000000
Non-
Reactive0020001700160053
Total0020001800160054
PPA0.00 (0/0)0.00 (0/0)0.00 (0/0)0.00 (0/0)
95% CI*0.00 to 100.000.00 to 100.000.00 to 100.000.00 to 100.00
NPA100.00 (20/20)94.44 (17/18)100.00 (16/16)98.15 (53/54)
95% CI*83.16 to 100.0072.71 to 99.8679.41 to 100.0090.11 to 99.95
Percent Agreement for Elecsys and Predicate anti-HAV Assays Results from
Pre-HAV Vaccination Specimens

Percent Agreement for Elecsys and Predicate anti-HAV Assays Results from Post-HAV Vaccination Specimens

anti-HAV Assay Results
Predicate Result
Elecsys
ResultHAVRIXTWINRIXVAQTAAll Vaccines
RXEQNRRXEQNRRXEQNRRXEQNR
Reactive2000170116005301
Equivocal000000000000
Non-
Reactive000000000000
Total2000170116005301
PPA100.00 (20/20)100.00 (17/17)100.00 (16/16)100.00 (53/53)
95% CI*83.16 to 100.0080.49 to 100.0079.41 to 100.0093.28 to 100.00
NPA0.00 (0/0)0.00 (0/1)0.00 (0/0)0.00 (0/1)
95% CI*0.00 to 100.000.00 to 97.500.00 to 100.000.00 to 97.50

RX - Reactive, positive, EQ - equivocal, borderline, NR - non-reactive, negative, PPA - positive percent agreement, NPA negative percent agreement, 95% CI - 95% Exact confidence interval

Prevalence Studies:

18

The Elecsys Anti-HAV assay was used to evaluate the prevalence of HAV total antibodies in an apparently healthy population (normal, healthy individuals without symptoms). The prospective study population for the Elecsys Anti-HAV assay consisted of 602 patients. Of these 602 patients, 300 patients were from the high prevalence region, Western states of the U.S. (New Mexico) and 302 patients were from the low risk region Eastern states of the U.S. (Indiana). The prospective study population was 208 (34.6 %) males and 394 (65.4 %) females (total n = 602) including 493 (81.9 %) Caucasians, 32 (5.3 %) African Americans, 6 (1.0 %) Asians, 69 (11.5 %) American Indians and 2 (0.3 %) Others. The results of prevalence population are summarized according to age groups in decades, gender, geographical area and the number of reactive, non-reactive and equivocal results.

Prevalence rate for reactive anti-HAV assay in specimens collected in a low prevalence region, Eastern states of the US (Indiana), was 35.10 %. Prevalence rate for reactive anti-HAV assay in specimens collected in a high prevalence region, Western states of the US (New Mexico), was 37.33%. The results of prevalence study are summarized according to age groups and gender.

Expected results for the Elecsys Anti-HAV assay in subjects from low prevalence
areas for Hepatitis A

| Age
range

(years)GenderElecsys Anti-HAV result
ReactiveEquivocalNon-reactive
NPercentNPercentNPercentTotal
11 - 20Female00.000.01100.01
11 - 20Male00.000.01100.01
21-30Female114.300.0685.77
21-30Male00.000.06100.06
31-40Female29.514.81885.721
31-40Male00.000.02100.02
41-50Female313.614.61881.822
41-50Male17.700.01292.313
51-60Female716.712.43481.042
51-60Male947.400.01052.619
61-70Female2345.100.02854.951
61-70Male1242.900.01657.128
71-80Female2450.000.02450.048
71-80Male1446.700.01653.330
> 80Female480.000.0120.05
> 80Male6100.000.000.06
All
agesFemale6432.531.513066.0197
All
agesMale4240.000.06360.0105
Total10635.131.019363.9302

19

Age range (years)GenderNPercentNPercentNPercentTotal
11-20Female450.000450.08
11-20Male240.000.0360.05
21-30Female423.500.01376.517
21-30Male218.200.0981.811
31-40Female311.100.02488.927
31-40Male323.100.01076.913
41-50Female1426.900.03873.152
41-50Male633.300.01266.718
51-60Female2138.900.03361.154
51-60Male1145.800.01354.224
61-70Female1872.000.0728.025
61-70Male936.000.01664.025
71-80Female975.000.0325.012
71-80Male571.400.0228.67
> 80Female00.000.01100.01
> 80Male00.000.000.00
UnknownFemale1100.000.000.01
UnknownMale00.000.000.00
All agesFemale7437.600.012362.4197
All agesMale3836.900.06563.1103
Total11237.300.018862.7300

Expected results for the Elecsys Anti-HAV assay in subjects from high prevalence areas for Hepatitis A

Seroconversion Sensitivity:

Seroconversion sensitivity of the Elecsys anti-HAV assay was shown by testing 3 seroconversion panels in comparison to that of the comparator assay (Abbott AxSym HAV AB- 2.0). Seroconversion panel results were also compared with the data provided by the vendor for the Abbott AxSym HAV AB 2.0 assay. The comparator assay and vendor assay are based on the Abbott AxSym HAV AB-2.0. The results are summarized in the following table:

20

Panel IDElecsys 2010 assay1Comparator anti-HAV assay2Comparator anti-HAV assay
Post bleed day of earliest reactive resultPost bleed day of last positive resultNon-ReactiveReactiveNon-ReactiveReactive
HAV-01091091091
3PHT 90216163Not testedNot tested16163
RP0136190Not testedNot tested9190

Elecsys anti-HAV assay Seroconversion Panel Results

The comparator results were shown by Roche using the Abbott AxSym HAV AB-2.0.

َ The comparator results were provided by Vendor i.e. Abbott AxSym HAV AB-2.0

3 The panel was not tested with the reference assay due to the limited sample size tested.

Method Comparison studies on two instruments:

The method comparison for the Elecsys anti-HAV assay between the two platforms E2010 and E170 was demonstrated by testing the prevalence and other clinical cohort specimens on MODULAR Analytics E170 and Elecsys 2010 analyzers. 100 non-reactive specimens were obtained from the Prevalence cohort. 53 reactive specimens were collected in the US from various clinical cohorts and another 49 reactive samples were provided by Roche R & D from sample archive. Among those samples, 96 were low negatives; 6 were high negatives; 12 were low reactive and 88 were high reactive samples. The regression analyses for the comparisons using the prevalence cohort and the combined clinical cohorts were carried out using the least squares and Passing-Bablok regression methods. The Pearson's regression correlation for the 202 data pairs was 0.9951, with slope at 1.024 and intercept at -0.555. Positive percent agreement was 91/94 = 96.8 % (95 % exact confidence limits of 91.0 to 99.3). Negative percent agreement was 98/100 = 98.0 % (95 % exact confidence limits of 93.0 to 99.8). 8 specimens were concordantly equivocal.

4. Clinical cut-off:

Refer to assay cutoff section above for additional details.

5. Expected values/Reference range:

The Elecsys anti-HAV assay was used to evaluate the prevalence of HAV antibodies in an apparently healthy population (normal, healthy Individuals without symptoms). A statistically significant number of subjects were prospectively collected in a "high prevalence" region, the Western states, and a "low prevalence" region, the Eastern states, and were tested using only the Elecsys anti-HAV assay and were not compared with the predicate device.

21

Expected Results from low and high Prevalence based on the test device are presented under the prevalence study section. Prevalence rate for reactive anti-HAV total antibody in specimens collected in a high prevalence region, Western states of the US (New Mexico), was 37.33 %. Prevalence rate for reactive anti-HAV total antibody in specimens collected in a low prevalence region, Eastern states of the US (Indiana), was 35.10 %.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

22