Search Results
Found 9 results
510(k) Data Aggregation
(64 days)
RENAL SOLUTIONS, INC.
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician.
The 2008 Sorbent System consists of two distinct components:
- the 2008 Machine, and .
- . the SORB™ Module.
The SORB Module is a sorbent dialysate regenerative system that attaches to the 2008 Machine and replaces the machine's existing single-pass dialysate delivery system.
During treatment, used dialysate is chemically reprocessed into fresh, new dialysate and sent back to the dialyzer instead of being sent down a drain. By recirculating and regenerating the dialysate, the 2008 Sorbent System uses less than 15 liters of potable tap water per treatment compared to a standard sinqle-pass system that uses at least 120 liters of purified water during a standard 4-hour dialysis treatment.
The provided text describes a Special 510(k) submission for an updated version of a hemodialysis system, the 2008 Sorbent System. This type of submission focuses on demonstrating substantial equivalence to a predicate device (an already legally marketed device) through verification testing for modifications that do not significantly alter the device's fundamental technology, indications for use, or safety and effectiveness.
Here's an analysis based on the provided text, addressing your questions:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state quantitative "acceptance criteria" in the traditional sense of a performance study with specific outcome measures and thresholds. Instead, it describes a "verification testing summary" aimed at demonstrating that the modified device meets its design requirements and is substantially equivalent to the predicate device.
The "performance" is implicitly defined by the successful completion of these verification activities, indicating that the modified device functions as intended and safely.
Acceptance Criteria (Implicit) | Reported Device Performance (as demonstrated by verification testing) |
---|---|
2008 System Software/Hardware Functionality | Success: Meets performance specifications (Functional SW Verification, Disinfection, Infusate Delivery, Modes of Operation, Alarms testing). |
HISORB+ Cartridge Performance & Equivalence | Success: Dialysate chemistries are comparable, performance (modified cartridge) is acceptable, and passes shipping test. |
Substantial Equivalence to Predicate Device (K093362) | Success: Design verification testing demonstrates that the device meets design requirements and supports substantial equivalence. (Implies all functional aspects are comparable). |
General Safety and Effectiveness (no new questions raised) | Success: Performance and technological characteristics are equivalent, raising no new safety or effectiveness questions. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a numerical sample size for "test sets" in the context of a clinical study. The testing described is non-clinical verification testing (e.g., software testing, hardware testing, cartridge performance), not a study involving human patients. Therefore, the "sample size" would refer to the number of test runs, prototypes, or cartridges used to perform the verification, which is not detailed.
- Data Provenance: The data is likely internal to the manufacturer, Renal Solutions, Inc., and represents retrospective (or in-house) testing focusing on engineering and performance verification of the device itself, rather than prospective clinical data from patients. There is no information regarding the country of origin of data, as it's not a clinical data set.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this submission. The "ground truth" for non-clinical verification testing is typically established by engineering specifications, regulatory standards, and the performance characteristics of the predicate device. The experts involved would be engineers, scientists, and quality assurance personnel within Renal Solutions, Inc., responsible for design, testing, and regulatory affairs. Their specific numbers and qualifications are not detailed here, as this is not a study requiring expert clinical review for ground truth establishment.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies where multiple human readers assess cases and discrepancies need to be resolved. This submission describes non-clinical verification testing of a device's functionality and performance against engineering requirements and established safety standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of an MRMC comparative effectiveness study, human readers, or AI assistance. This device is a hemodialysis system, not an imaging or diagnostic AI-powered device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a hemodialysis machine with a sorbent system. It is a physical medical device for treatment, not an algorithm or AI system for standalone performance evaluation. Its operation still requires a healthcare practitioner ("human-in-the-loop") as stated in its Indications for Use.
7. The Type of Ground Truth Used
The "ground truth" for this submission is based on:
- Engineering Specifications and Design Requirements: The device was tested to ensure it met its own defined performance parameters.
- Predicate Device Performance: The primary ground truth for demonstrating "substantial equivalence" is the established safety and effectiveness of the legally marketed predicate device (2008 Sorbent System, K093362). The modified device's performance was compared to, and intended to be equivalent to, that of the predicate.
- Regulatory Standards: Implicitly, the verification testing would also adhere to relevant industry standards and regulatory guidance for hemodialysis devices.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical hemodialysis system, not a machine learning model that requires a "training set."
9. How the Ground Truth for the Training Set was Established
This information is not applicable for the same reason as above (not a machine learning model).
Ask a specific question about this device
(88 days)
RENAL SOLUTIONS INC.
The Venofer Pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer® to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.
The Venofer Pump is an optional module for use on Fresenius 2008 Series Hemodialysis Machines and is designed to administer Venofer® during dialysis treatments and consists of a control panel, vial holder, fluid detector, and a peristaltic pump. The module is a self contained microprocessor controlled device that receives its power from the Fresenius 2008 Hemodialysis machine.
The provided text describes the Venofer Pump Version 1.3, an optional accessory for hemodialysis machines. This submission is a "Special 510(k)" for modifications to support the use of an additional blood tubing set. The document focuses on demonstrating substantial equivalence to a previously cleared device (K103564) and confirming that the modifications do not raise new questions of safety or effectiveness.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Verification Tests Conducted | Reported Device Performance/Outcome |
---|---|---|
Venofer Pump Performance | Functional testing to demonstrate the device performs as designed and expected. | The performance (verification) test results support the performance characteristics of the Venofer Pump. It was confirmed that the device does not perform in an unexpected and/or unsafe manner (system-level hazard analysis). Specific details of the performance metrics are not explicitly provided beyond this general statement. |
Venofer Pump Safety | Electromagnetic compatibility (EMC) testing (IEC 60601-1-2: 2007) | Product safety testing demonstrates that the device performs per the FDA Consensus Standards. This specifically includes EMC testing. |
Biocompatibility testing (AAMI / ANSI / ISO 10993-1:2009) | Biocompatibility testing was performed on all new materials that are patient-fluid contacting. The verification test results support the safety characteristics of the Venofer Pump. (No specific numerical results or pass/fail criteria are provided, only that testing was performed and supported safety). | |
Venofer Pump Software | Software architecture SDD (Software Design Description) Unit test System and software requirements Traceability | Software testing demonstrates the device software meets the design input requirements. The testing includes required documentation as described in the guidance titled "Guidance for the Content of Premarket Submissions for Software." (No specific metrics like bug rates, code coverage, or pass/fail percentages are provided, only a general statement of meeting design input requirements and review of documentation). The phrase "Software release history" is also listed as documentation reviewed, suggesting a process was followed. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the test set for any of the verification activities (functional, safety, or software).
The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. The studies described are non-clinical (verification) testing, not clinical trials on human subjects.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. Given that the studies are non-clinical verification tests, the "ground truth" would likely be established by engineering specifications, regulatory standards, or consensus standards, rather than expert clinical judgment in the traditional sense for medical imaging or diagnostic devices.
4. Adjudication Method for the Test Set
This information is not applicable and therefore not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human readers/interpreters, not for non-clinical device verification testing.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No MRMC comparative effectiveness study was done. This device is a pump for administering medication, not an AI-powered diagnostic or assistive device for human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
No standalone algorithm performance study was done in the context of AI. The "standalone" performance described would be the functional performance of the pump itself, which was tested during "Venofer Pump Performance Testing." The device is a physical pump, not a diagnostic algorithm.
7. The Type of Ground Truth Used
For the non-clinical verification testing described:
- Performance Testing: The "ground truth" would be the device's design specifications and expected functional behavior.
- Safety Testing: The "ground truth" would be established by FDA Consensus Standards (e.g., IEC 60601-1-2 for EMC, AAMI / ANSI / ISO 10993-1 for biocompatibility).
- Software Testing: The "ground truth" would be the software design input requirements and functional specifications.
8. The Sample Size for the Training Set
This information is not applicable and therefore not provided. The Venofer Pump is a physical medical device, not an AI/machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable and therefore not provided. As mentioned above, the device is not an AI/machine learning algorithm.
Ask a specific question about this device
(66 days)
RENAL SOLUTIONS, INC.
The intended use of the Venofer Pump 1.1 is identical to the unmodified device.
The Venofer® Pump Is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.
Fresenius 2008T is indicated for acute and chronic dialysis therapy.
The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.
The Venofer Pump Module fits into the module compartment of Fresenius 2008T Machines and consists of a control panel, vial holder, fluld detector, and a peristaltic pump.
This document is a 510(k) Summary for the Venofer Pump Version 1.1, a device intended to administer Venofer (iron sucrose) during hemodialysis. This filing is a "Special 510(k)" for a modification to an existing device (Venofer Pump K093964).
Crucially, this document focuses on demonstrating substantial equivalence through non-clinical verification testing (performance, safety, and software) and does NOT describe a study involving human patients or a process for establishing ground truth from expert consensus for diagnostic accuracy.
Therefore, most of the requested information regarding clinical studies, sample sizes, expert involvement, and ground truth establishment will not be applicable or available in this specific document.
Here's an analysis based on the provided text:
1. Table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Reported Device Performance (Summary from this document) |
---|---|
Venofer Pump Performance Testing | Functional testing demonstrated that the device performs as designed and expected. This included specific verification tests with defined objectives, test articles, test methods and procedures, and pre-defined acceptance or pass/fail criteria. A system-level hazard analysis confirmed the device does not perform in an unexpected and/or unsafe manner. (Specific quantitative performance metrics are not detailed in this summary.) |
Venofer Pump Safety Testing | Product safety testing demonstrated that the device performs per FDA Consensus Standards. This included: |
- Electromagnetic Compatibility (EMC): Tested to IEC 60601-1-2: 2007.
- Biocompatibility: Tested to AAMI / ANSI / ISO 10993-1:2009 for all new patient-fluid contacting materials. (Specific quantitative results or pass/fail criteria for these tests are not detailed in this summary.) |
| Venofer Pump Software Testing | Software testing demonstrated the device software meets the design input requirements. This included: - Structure chart/flow chart of software architecture.
- Summary of software development procedures, including changes.
- Software requirements specification with traceability to hazard analysis.
- Verification and validation test plans, including pass/fail criteria and traceability to requirements.
- System-level test results.
- Current software version number and date of latest revision. (Specific quantitative test results or detailed pass/fail criteria are not detailed in this summary.) |
Study Proving Device Meets Acceptance Criteria:
The study referenced is the "510(k) Verification Testing Non-Clinical Summary" which includes "Performance, Safety, and Software Testing." The document emphasizes that "The conclusions drawn from this testing demonstrates that the Venofer Pump is as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices to which it was compared."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: This refers to the number of devices or components tested, not patient data. The document does not specify the exact number of units or test iterations for the performance, safety, or software verification tests. It generally refers to "the device" implying tests were conducted on one or more representative units.
- Data Provenance: Not applicable. This is not a study involving patient data. The testing was non-clinical (laboratory/engineering) and would have been conducted by the manufacturer, Renal Solutions, Inc. (USA).
- Retrospective or Prospective: Not applicable. This applies to clinical data collection, not non-clinical device verification.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This device is a pump for administering medication, not an AI/diagnostic device that requires expert-established ground truth from medical imaging or patient data. The "ground truth" for this device's performance is its adherence to engineering specifications, safety standards, and software requirements, which are verified through the non-clinical testing described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. Adjudication methods like 2+1 or 3+1 refer to how discrepancies in expert opinions are resolved to establish a definitive ground truth in diagnostic studies. This is a non-clinical device verification, so such adjudication is not performed. Acceptance was based on pre-defined pass/fail criteria for engineering and safety tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study is relevant for diagnostic devices that assist human readers (e.g., radiologists interpreting images). This device is a pump, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Applicable in the traditional sense of an AI algorithm. This is a hardware device with embedded software. The "standalone" performance was effectively what the document refers to as "Venofer Pump Performance Testing" and "Venofer Pump Software Testing" – verifying the device and its embedded software performed according to specifications and safety requirements without human intervention during the test cycles, but it's not an AI algorithm performing a diagnostic task.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable/Defined Differently. For this type of device, "ground truth" is defined by:
- Engineering Specifications: The pre-defined functional requirements and performance parameters of the pump (e.g., flow rate accuracy, pressure limits).
- Regulatory Standards: Compliance with recognized FDA Consensus Standards (e.g., IEC 60601-1-2 for EMC, ISO 10993-1 for Biocompatibility).
- Software Requirements: The predefined design input requirements for the device's embedded software.
- Hazard Analysis: Confirmation that the device does not perform in an unexpected and/or unsafe manner.
8. The sample size for the training set
- Not Applicable. This device does not use machine learning or AI that requires a "training set" of data. Its functionality is based on deterministic programming and electro-mechanical design, not learned patterns from data.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set for an AI algorithm, there is no ground truth to establish for it.
Ask a specific question about this device
(289 days)
RENAL SOLUTIONS INC.
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician.
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician. The 2008 Sorbent System operates in a similar manner to standard hemodialysis systems currently approved for use; it is substantially equivalent to the Fresenius 2008T hemodialysis machine (K080964) and the Renal Solutions Allient® Sorbent Hemodialysis System (K070739). The significant difference between the 2008 Sorbent System and standard single-pass hemodialysis machines is that the 2008 Sorbent System utilizes a sorbent cartridge to purify dialysate made from potable tap water and then regenerates fresh dialysate from spent dialysate (like the Allient Sorbent Hemodialysis System). The 2008 Sorbent System consists of two distinct components: the 2008 Machine, and the SORB™ Module. The SORB Module is a sorbent dialysate regenerative system that attaches to the 2008 Machine and replaces the machine's existing single-pass dialysate delivery system. During treatment, used dialysate is chemically reprocessed into fresh, new dialysate and sent back to the dialyzer instead of being sent down a drain. By recirculating and regenerating the dialysate, the 2008 Sorbent System uses less than 15 liters of potable tap water per treatment compared to a standard single-pass system that uses at least 120 liters of purified water during a standard 4-hour dialysis treatment.
The provided text describes the 2008 Sorbent Hemodialysis System, focusing on its substantial equivalence to predicate devices and the verification testing performed. However, it does not include the specific details required to fully answer your request regarding acceptance criteria and a study proving the device meets those criteria, particularly for an AI-powered medical device.
The document is a 510(k) summary for a Sorbent Hemodialysis System, which is a medical device for dialysate delivery, not an AI or software-intensive diagnostic device in the sense your questions imply (e.g., test sets, ground truth, expert consensus, MRMC studies). The "software testing" mentioned is to ensure the device's control software meets its design input requirements, not for evaluating an AI algorithm's diagnostic or predictive performance.
Therefore, many of your requested items cannot be extracted from this document as they are not relevant to the type of device being described.
However, I can extract what is available:
1. A table of acceptance criteria and the reported device performance:
The document outlines categories of testing and general objectives, implying acceptance criteria were defined for each. However, it does not explicitly state the quantitative acceptance criteria or the specific numerical performance results. Instead, it refers to "pre-defined acceptance or pass/fail criteria" and states that testing "demonstrates the device meets its performance specifications."
Acceptance Criteria Category | Reported Device Performance (as inferred) |
---|---|
2008 Sorbent System Performance Testing | Functional testing conducted to demonstrate the device performs as designed and expected, meeting its performance specifications. (Specific verification tests, protocols, objectives, test articles, methods, procedures, and pre-defined acceptance criteria were documented in the 510(k) submission, but not detailed here). System-level hazard analysis confirmed the device does not perform in an unexpected and/or unsafe manner. |
2008 Sorbent System Safety Testing | Product safety testing conducted demonstrating compliance with FDA Consensus Standards, including: Electrical safety (IEC 60601-1), Electromagnetic compatibility (IEC 60601-2-16), Biocompatibility (AAMI / ANSI / ISO 10993-1) for new patient-fluid contacting materials, and Risk Analysis (ISO 14971). |
2008 Sorbent System Software Testing | Software testing conducted demonstrating the device software meets the design input requirements. This included structural charts, development procedures, requirements specifications with traceability to hazard analysis, verification and validation test plans (with pass/fail criteria and traceability), system-level test results, and current software version documentation. |
Overall Substantial Equivalence to Predicate Devices | Demonstrated the device is as safe, as effective, and performs at least as safely and effectively as the Allient Sorbent Hemodialysis System and Fresenius 2008T Hemodialysis System. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the given text. The testing described is non-clinical verification testing of a physical medical device, not a study involving patient data in the context of AI performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable/provided. The ground truth for this type of device (a hemodialysis system) is established through engineering and performance specifications, adherence to standards, and risk analysis, not expert consensus on diagnostic interpretations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/provided. Adjudication methods like 2+1 are typically used in clinical studies involving multiple readers or interpretations, which is not what is described here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not done, as this device is a hemodialysis system and not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable/provided. The "software testing" refers to the control software of the device, not a standalone diagnostic algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device is based on engineering specifications, regulatory standards (e.g., IEC, ISO), and performance requirements related to the physical and functional operation of the hemodialysis system. The document lists "pre-defined acceptance or pass/fail criteria" as part of the test protocols.
8. The sample size for the training set
This information is not applicable/provided. This device is a physical medical device, not an AI model trained on a dataset.
9. How the ground truth for the training set was established
This information is not applicable/provided. As above, there is no AI model or training set in the context of this device.
In summary, the provided document describes a 510(k) submission for a Sorbent Hemodialysis System, focusing on its functional, safety, and software verification to demonstrate substantial equivalence to predicate devices. It does not contain the detailed information typically associated with the evaluation of AI-powered diagnostic or predictive medical devices.
Ask a specific question about this device
(197 days)
RENAL SOLUTIONS INC.
Fresenius Model 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.
The Venofer® Pump Module is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.
The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.
The Venofer Pump Module is intended to provide ease and consistent delivery of Venofer, an iron sucrose supplement, indicated for the treatment of iron deficiency anemia for Hemodialysis dependent-chronic kidney disease patients. Anemia commonly occurs in patients with chronic kidney disease undergoing Hemodialysis and effective anemia management is recognized as an important factor in improving the outcomes of these patients. Iron deficiency is a frequent contributing factor which complicates the treatment of anemia in chronic kidney disease patients. Iron sucrose is commonly delivered to improve iron status in patients who have chronic kidney disease. Venofer is an injectable iron preparation drug approved for treatment of iron deficiency and is the most commonly used drug to treat iron deficiency anemia in dialysis patients. Renal Solutions has developed the Venofer Pump Module that provides the means to deliver Venofer when undergoing Hemodialysis on a Fresenius 2008T Dialysis System.
The Venofer Pump Module fits into the module compartment of existing Fresenius 2008T Machines and consists of a control panel, vial holder, fluid detector, and a peristaltic pump.
The provided text describes a 510(k) premarket notification for the "Fresenius Model 2008T Hemodialysis Machine with Venofer Pump Module." This document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with acceptance criteria and performance data for a novel device. Therefore, much of the requested information regarding acceptance criteria, sample sizes, expert ground truth, adjudication methods, and comparative effectiveness studies is not available in the provided text.
Based on the information given, here's what can be extracted:
1. A table of acceptance criteria and the reported device performance:
This information is not provided in the document. The submission is for a medical device accessory and states that non-clinical testing was performed, but it doesn't detail specific acceptance criteria or quantitative performance results in a table format. It generally claims the device is "as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices."
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the document. The text mentions "non-clinical testing" but does not specify sample sizes or data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the document. The nature of the device (a pump for administering medication during hemodialysis) suggests that ground truth would likely be established through engineering and medical performance standards, rather than expert-derived ground truth in the way a diagnostic imaging AI might.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable and not provided in the document. The Venofer Pump Module is an accessory for a hemodialysis machine, administering medication. It is not an AI-assisted diagnostic tool that would typically involve human "readers" or an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This information is not applicable in the context of an AI algorithm. The device is a physical pump module. The document states, "Like the Heparin Pump, the Venofer Pump Module is not a stand alone device. The Venofer Pump Module cannot operate unless it is connected to the 2008T Hemodialysis Machine." This refers to its operational dependency on the larger hemodialysis machine, not a standalone AI performance evaluation.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
This information is not explicitly stated in the document. Given the device's function, ground truth for its performance would likely be established through engineering specifications, flow rate accuracy, absence of adverse events, and efficacy of drug delivery as measured by standard medical parameters (e.g., successful iron administration without pump-related complications), rather than categorical labels like pathology or expert consensus on image interpretation. The submission relies on "verification, safety, performance and software testing."
8. The sample size for the training set:
This information is not provided in the document. The device is a physical pump, not an AI algorithm that undergoes a "training set" in the machine learning sense.
9. How the ground truth for the training set was established:
This information is not provided in the document, as it's not an AI device with a training set.
Ask a specific question about this device
(70 days)
RENAL SOLUTIONS INC.
The Allient Sorbent Hemodialysis System, including the SORB™ series and HISORB™ series of cartridges is to be used for the treatment of acute and chronic uremic patients where hemodialysis is prescribed by the physician.
Not Found
This document is a 510(k) clearance letter for the Allient Sorbent Hemodialysis System. It does not contain information on the acceptance criteria or a study that proves the device meets specific acceptance criteria in the format requested. The letter confirms that the device is substantially equivalent to legally marketed predicate devices, but it does not detail the technical performance data that led to this determination.
Therefore, I cannot provide the requested table or information based on the content of this document.
Ask a specific question about this device
(99 days)
RENAL SOLUTIONS, INC.
The Allient Sorbent Hemodialysis System, including the SORB series and HISORB series of cartridges is to be used for the treatment of acute and chronic uremic patients where hemodialysis is prescribed by the Physician.
The Allient Sorbent Hemodialysis System is intended for the treatment of THE Allent Oorbont Homoutients where hemodialysis is prescribed by the ohysician. The Allient System functions as a traditional re-circulating sorbent hemodialysis system. Either single-needle or dual-needle access to the nemodialysis bystem . The system consists of the Allient hemodialysis machine, SORB™ series of sorbent cartridges, a single-use, sterile, disposable blood tubing set, a single-use disposable dialysate set, and various dialysate and infusate chemicals. The patient's blood is pumped from the access through a dialyzer and is returned to the patient.
This is a 510(k) summary for a medical device (Allient® Sorbent Hemodialysis System) and not a study describing acceptance criteria and device performance. Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and comparative effectiveness studies is not available in the provided text.
The document primarily focuses on establishing substantial equivalence to a previously cleared predicate device, rather than presenting a performance study with detailed acceptance criteria and results.
Here's what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance
- Not Available. The document does not describe specific acceptance criteria (e.g., performance metrics, thresholds) for the device. Instead, it asserts "substantial equivalence" to a predicate device.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Available. There is no mention of a "test set" or any specific clinical data used to evaluate the device's performance against predefined criteria. The assessment is based on comparison to a predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable / Not Available. Since no "test set" or performance study is detailed, there's no mention of experts establishing ground truth for such a study.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable / Not Available. No test set or adjudication method is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a hemodialysis system, not an AI-assisted diagnostic tool involving human readers. Therefore, an MRMC study is not relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a hardware medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable / Not Available. As no detailed performance study with a test set is described, the concept of "ground truth" in this context is not relevant. The substantial equivalence is based on similar design, operating principles, and intended use as the predicate device.
8. The sample size for the training set
- Not Applicable / Not Available. This device is not an AI/ML algorithm that would typically involve a "training set."
9. How the ground truth for the training set was established
- Not Applicable / Not Available. As above, this concept doesn't apply to the reported device.
Summary of available information related to the device:
- Device Name: Allient® Sorbent Hemodialysis System
- Intended Use: Treatment of acute and chronic uremic patients where hemodialysis is prescribed by the Physician.
- Predicate Device: Allient Sorbent Hemodialysis System K043572 (or K043574 as cited later, which appears to be the same predicate).
- Basis for Equivalence:
- Same indicated use.
- Same operating principles.
- Same basic system design.
- Incorporates the same accessories and generic materials, manufactured, packaged, and sterilized using the same processes.
- Regulatory Classification: Class II, Product Code FKT, Regulation Number 21 CFR §876.5600 (Sorbent regenerated dialysate delivery system for hemodialysis).
In conclusion, the provided text is a regulatory submission for a 510(k) clearance, which aims to demonstrate substantial equivalence to an already legally marketed device rather than providing a detailed performance study with acceptance criteria and results as typically seen for novel diagnostic or AI devices.
Ask a specific question about this device
(158 days)
RENAL SOLUTIONS, INC.
The Allient® Sorbent Hemodialysis System, including the SORB™ series and HISORB™ series of cartridges, is to be used for the treatment of acute and chronic uremic patients where hemodialysis is prescribed by the physician.
The Allient Sorbent Hemodialysis System (Allient System) is intended for the treatment of acute or chronic uremic patients where hemodialysis is prescribed by the physician. The system consists of the Allient hemodialysis machine, SORB™ series of Sorbent cartridges, a single-use, sterile, disposable blood tubing set, a singleuse disposable dialysate set, and various dialysate and infusate chemicals. The system is recommended for use with a variety of dialyzers, access needles or access cannulae, which will be supplied by other manufacturers. The Allient System operates in a similar manner to standard hemodialysis systems currently approved for use. The only significant differences between the Allient System and standard hemodialysis machines are as follows: The Allient System utilizes a Sorbent cartridge to purify dialysate made from potable water. After passing through the dialyzer, the dialysate is recirculated to the dialyzer instead of being disposed of. The Allient System utilizes the Pulsar™ air-activated pump in contrast to the roller-activated pump commonly employed in traditional hemodialysis systems. The Pulsar blood pump is a dual-chamber pneumatically operated pump. All other aspects of Renal Solutions' Allient Sorbent Hemodialysis System process are identical to those employed in the standard "single pass" method of dialysis: there are no significant differences in the technology employed in Sorbent dialysis therapy between the Allient System and traditional methods. The Allient System functions as a traditional recirculating Sorbent hemodialysis system, similar to the REDY 2000 System. Either single-needle or dual-needle access to the patient is permitted. The patient's blood is pumped from the access through a dialyzer and is returned to the patient. Heparin is administered via a flow-controlled heparin pump capable of delivering rates of 0 to 3 milliliter per hour (mL/hr) in 0.5 mL/hr increments as well as an initial bolus of 0 to 10 mL. Blood-flow rates are settable in 10 milliliter per minute (mL/min) increments in the range of 200 to 400 mL/min in the dual-needle mode and 200 to 300 mL/min in the singleneedle mode. The dialysate delivery system consists of a container for dialysate. Sorbent regenerative cartridges, chemical infusion system to replace chemicals removed by the Sorbent cartridges, and an ultrafiltrate control system. The dialysate delivery system operates in a batch recirculating mode and continuously recirculates the dialysate through the cartridge. The cartridge removes dialyzed toxins from the dialysate and buffers the sodium and bicarbonate levels in the dialysate to maintain normal levels in the patient. The Acute mode of operation is intended for use in a hospital, acute or chronic-care facility, or dialysis clinic where the patient is continually monitored. The Acute mode of operation allows the clinician or dialysis technician to enter prescription parameters and machine operating parameters for each patient. The Allient System incorporates a Graphical User Interface (GUI) comprised of a touch-screen display, which controls the System software. The GUI is used to select the mode of operation of the Allient System and enter prescription data and machine control parameters during setup. This information is stored for future reference and verification prior to subsequent treatments. Treatment status and progress, in addition to machine status, is continually collected and displayed on the GUI screen to support treatment. The GUI's touch-screen technology also provides ease of use for Home patients. The Allient System incorporates a control system that contains a number of safety features and audible and visible alert and alarm functions to warn the operator or patient of unsafe conditions.
The provided document describes the "Allient® Sorbent Hemodialysis System" and its substantial equivalence to a predicate device, the REDY 2000. It focuses on the device's specifications and compliance with various domestic and international standards, rather than a clinical study with acceptance criteria in the typical sense of accuracy, sensitivity, or specificity for a diagnostic or AI-driven device.
However, I can extract the relevant information from the document to construct an answer regarding acceptance criteria based on performance specifications and the study that "proves" (or verifies, in this context) the device meets them.
Acceptance Criteria and Device Performance based on provided specifications:
1. Table of Acceptance Criteria and Reported Device Performance
For a medical device like a hemodialysis system, "acceptance criteria" are more about meeting technical specifications and safety standards rather than diagnostic performance metrics (like sensitivity/specificity). The document details various operating parameters, flow rates, and safety features as the de facto acceptance criteria. The "reported device performance" is implicitly stated by saying the system "meets its specifications" and various standards.
Feature / Parameter | Acceptance Criteria (Specification) | Reported Device Performance |
---|---|---|
Dialysate Flow Rate | 0 to 400 +/- 10% mL/min (into 25 psi max back pressure from Sorbent cartridge); Measuring Tolerance: +/- 10% of set point; Increment: 10 mL/min (200-400 mL/min range) | "The Allient System has been tested to verify that it meets its specifications as listed above." (Implicitly, the device performs within these specified tolerances.) |
Infusate Flow Rate | 0 to 3 mL/hr against +/- psi back pressure; Measuring Tolerance: +/- 10% of set point | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Heparin Flow Rate | 0 to 3 mL/hr; Measuring Tolerance: +/- 0.1mL/hr (1 mL/hr); Increment: 0.25 mL/hr; Accuracy: 10% of set point | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Blood Flow Rates | Single Needle: 150-200 mL/min; Dual Needle: 150-400 mL/min | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Air Detectors | Detection Range: > 1.0 ml bubble or microbubbles > 1.5 ml | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Dialysate Temperature | Control Range: 34°C to 37°C | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Blood Leak Detector | Leak Detection Range: 0.35 ml/min of 0.25% hematocrit Blood in dialyzer output flow | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Saline Flow Rate | Controlled at 200 mL/min; Measuring Tolerance: +/- 10% of set point | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Net Ultrafiltration | Min. 20g/hour to 2000 g/hour (based on acceptable Dialyzer); Measuring Tolerance: +/- 20 g/hr or 5% (whichever is greater) | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Degassing Flow Rate | 750 mL/min (with SORB cartridge bypassed) | "The Allient System has been tested to verify that it meets its specifications as listed above." |
Compliance with Standards | ANSI/AAMI RD17, ANSI/AAMI RD5, EN 556, IEC 60529, IEC 60601 series, ISO 10993 series, ISO 11135, ISO 8638 | "The Allient System has also been tested and meets the following domestic and international standards." |
Substantial Equivalency | To predicate REDY 2000 system | "All testing performed on the Allient Sorbent Hemodialysis System verifies the substantial equivalency of the Allient System to the predicate REDY 2000 system." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a "sample size" for a test set in the traditional sense of a clinical trial with patient data. The "testing" referred to is engineering verification and validation against technical specifications and recognized standards. This typically involves testing multiple units of the device and its components in a controlled laboratory environment.
- Test Set Sample Size: Not specified as a number of patients or cases. It would refer to the number of devices or components tested.
- Data Provenance: The testing is internal (by Renal Solutions Inc.) for verification and validation against engineering specifications and international standards (e.g., IEC, ISO, AAMI). This is a technical validation, not clinical data provenance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This concept is not applicable to the type of device and study described. For a hemodialysis machine, the "ground truth" is defined by the physical laws of operation and the engineering specifications that ensure safety and effectiveness. Experts (e.g., engineers, quality control personnel) would be involved in designing, measuring, and verifying these specifications against established benchmarks and regulatory requirements. There is no "ground truth" established by clinical experts reading or interpreting output in the way one would for an AI-driven diagnostic device.
4. Adjudication Method for the Test Set
This concept is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical trials or studies where there is subjective interpretation (e.g., image reading) and a need to resolve disagreements among multiple experts to establish a definitive ground truth. For the Allient System, performance is measured against objective, quantifiable engineering specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done, and the document does not mention any clinical trials comparing human performance with and without AI assistance. This type of study is typically for evaluating diagnostic accuracy or decision support systems, which the Allient System is not.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The Allient System is a physical medical device, not an algorithm or AI system. Its performance is inherent in its electromechanical and chemical operations. Therefore, the concept of "standalone" performance for an algorithm is not applicable here. Its operation is standalone in the sense that it performs its function (dialysis) according to its design.
7. The Type of Ground Truth Used
The "ground truth" for the Allient System is based on:
- Engineering Specifications: Quantifiable parameters and tolerances for flow rates, temperatures, leak detection, etc.
- Compliance with Recognized Standards: Adherence to established domestic and international standards for medical devices (e.g., ANSI/AAMI, IEC, ISO) which define safety, electrical, biological, and sterilization requirements.
- Predicate Device Performance: The predicate device, REDY 2000, serves as a benchmark for "substantial equivalency," implying its established safety and effectiveness forms part of the "ground truth" or acceptable performance standard.
8. The Sample Size for the Training Set
This concept is not applicable. The Allient System is not an AI/ML algorithm that requires a "training set" of data. Its design and operational parameters are determined through engineering principles, bench testing, and adherence to established medical device standards.
9. How the Ground Truth for the Training Set Was Established
This concept is not applicable as there is no "training set" for this type of medical device. The "ground truth" for its design and verification is established through a combination of engineering design principles, regulatory requirements, and established industry standards for hemodialysis systems.
Ask a specific question about this device
(163 days)
RENAL SOLUTIONS, INC.
The SORB+ and HISORB+ Cartridges are to be used only with Renal Solutions' REDY Sorbant Dialysis systems for the treatment of acute and chronic uremic patients where hemodialysis is prescribed by the physician.
Sorbent (or regenerative) dialysis systems use the chemical action of a sorbent cartridge to remove urea and other waste materials from the dialysate and return the fresh dialysate to the dialysate holding tank. This fresh dialysate then passes through a dialyzer, removes additional waste material, and re-circulates through the sorbent cartridge to continue the process. This contrasts to traditional "single pass" dialysis, which uses purified water passing across the dialysate side of a dializer membrane in a single flow-through system to remove urea and other waste materials from the dialysis patient's bloodstream. Wastewater is routed to a drain and not recirculated. By re-circulating and refreshing the dialysate, the Renal Solutions' REDY System uses 6L of water per dialysis treatment. A traditional single pass dialysis system uses at least 120L of purified water during the typical 4 hour dialysis session.
The SORB+ and HISORB+ Cartridges are intended to be used in a Renal Solutions REDY Sorbent Hemodialysis System as the chemical agents that absorb the urea and other waste material from the dialysate stream.
Here's an analysis of the provided text to extract the acceptance criteria and study information:
Acceptance Criteria and Device Performance Study for SORB+ and HISORB+ Cartridges
This document describes the 510(k) summary for the SORB+ and HISORB+ Cartridges, which are sorbent-regenerated dialysate delivery systems for hemodialysis. The focus is on demonstrating substantial equivalence to predicate devices (K811170 - SORB 3160 and K812869 - HISORB 3260).
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" with quantitative targets for each test, but rather states that the device "Pass"ed the tests by conforming to specified standards or predicate device performance. The de facto acceptance criterion for each test was a "Pass" result, indicating conformance.
Testing Topic | Reference Standards / Implied Acceptance Criteria | Reported Device Performance |
---|---|---|
Validation of Maximum Allowable Contamination Levels of Toxic Chemicals in Water Supply for the SORB+ and HISORB+ Cartridges | AAMI/ANSI, RD5-1992 and RD62-2000, RD62, 2001 (Implied acceptance: Conformance to these standards for safe levels of toxic chemicals in water for dialysis) | Pass |
Limit Test of Endotoxin and Bacteria Removal by SORB+ and HISORB+ Cartridges | AAMI/ANSI (Implied acceptance: Conformance to these standards for acceptable levels of endotoxin and bacterial removal) | Pass |
Biocompatibility Validation of SORB + and HISORB + Cartridges | ISO 10993 (cytotoxicity, irritation or intracutaneous reactivity, and systemic toxicity (acute)). (Implied acceptance: Conformance to ISO 10993 for these biocompatibility aspects) | Pass |
Validation Testing of SORB+ and HISORB+ Cartridges | Sorb Technologies published capacity specifications (Implied acceptance: The cartridges meet their published capacity specifications for urea-nitrogen removal. Specifically, SORB+ has a capacity of 9.5-23.5g urea-nitrogen, and HISORB+ has a capacity of 23.5-35.0g urea-nitrogen, for removing patient dialysis waste products within a usual dialysis procedure.) This also includes "effectiveness in purifying water or dialysate to safe levels for dialysis." | Pass |
Comparison of Performance to Current SORB and HISORB Cartridges (in-vitro testing) | Substantial equivalency to the predicate devices (SORB and HISORB Cartridges) in in-vitro testing. (Implied acceptance: Performance is comparable to the predicate devices.) | Pass |
Overall Conclusion: Substantial Equivalency to Predicate | "All testing verifies the substantial equivalency of these cartridges to the predicate, SORB and HISORB Cartridges, when used with the Renal Solutions Sorbent Dialysis system for acute and chronic hemodialysis." (Implied acceptance: No unexpected safety or effectiveness issues compared to the predicate.) | Pass |
2. Sample Size for the Test Set and Data Provenance
The document does not specify exact sample sizes for the "test set" (i.e., the number of cartridges tested for each validation). It mentions "various weights [of REDY Chem™ dialysate] used individually and in combination" which hints at a testing regime, but no specific numbers.
The studies were in-vitro testing and "simulated patient use conditions." This suggests the tests were laboratory-based, not involving human patients directly. The provenance is not explicitly stated as retrospective or prospective due to the nature of the in-vitro testing. It would be considered prospective for the specific in-vitro tests conducted. There's no indication of country of origin for the data provided in this summary, but the submitter's address is in Oklahoma City, USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is generally not applicable to the validation of a medical device like a sorbent cartridge, which relies on chemical and physical performance measurements rather than expert interpretation of images or clinical outcomes. The "ground truth" would be established by objective measurements against predefined chemical and physical standards or specifications, rather than expert consensus.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation or clinical events requiring consensus. For these device performance tests, the outcome is objective (e.g., meeting a chemical purity threshold, acceptable biocompatibility levels). Therefore, no adjudication method as described would be relevant or used in these validation tests. The results would be a direct measurement against a standard.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done as this device is not an AI diagnostic tool and does not involve human readers for interpretation. The device's performance is measured directly (e.g., chemical capacity, purification levels).
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)
This question is not applicable in the context of this device. The device itself is a cartridge, not an algorithm. Its performance is inherently standalone (i.e., the cartridge's chemical and physical function) in that it performs its intended function independently of direct human intervention during the purification process. The "human-in-the-loop" would be the physician prescribing its use and monitoring the patient, but not directly interacting with the cartridge's internal chemical processes. The validation tests (e.g., capacity, biocompatibility) directly assessed the cartridge's standalone performance.
7. The Type of Ground Truth Used
The ground truth used for these tests was primarily:
- Objective Chemical and Physical Standards: Conformance to AAMI/ANSI standards for water purity, endotoxin/bacteria removal.
- Biocompatibility Standards: Conformance to ISO 10993.
- Manufacturer's Published Specifications: Sorb Technologies' published capacity specifications for urea-nitrogen removal.
- Predicate Device Performance: Comparison to the performance of the previously cleared SORB and HISORB Cartridges.
8. The Sample Size for the Training Set
This concept is not applicable. The device is a physical product (sorbent cartridge), not a machine learning model that requires a "training set." The development of the cartridges would have involved R&D, but the validation described herein does not fit the training/testing paradigm of AI/ML.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" is not applicable for this type of device.
Ask a specific question about this device
Page 1 of 1