Search Results
Found 10 results
510(k) Data Aggregation
(71 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa PR3 ANCA EIA kit is designed for the semi-quantitative and qualitative determination of proteinase 3 anti neutrophil cytoplasmic antibodies (PR3 ANCA) in human serum or plasma to aid in the diagnosis of Wegener's granulomatosis.
Varelisa PR3 ANCA is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of PR3 ANCA in human serum or plasma. The test kit contains microplate strips coated with human purified PR3 antigen, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, sample diluent and wash buffer.
Here's an analysis of the provided text to extract the requested information about acceptance criteria and the study proving device performance:
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes the device, its intended use, and its comparison to a predicate device. However, it does not explicitly state acceptance criteria in the form of specific performance metrics (e.g., sensitivity, specificity thresholds) that the new device (Varelisa PR3 ANCA) needed to meet.
Instead, the study's goal was to demonstrate substantial equivalence to the predicate device (INOVA QUANTA Lite™ PR-3) and that it "performs as expected from the medical literature" and "according to state-of-the-art expectations." The evaluation was based on a comparative study.
Therefore, the table will reflect this comparative nature rather than explicit acceptance criteria.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Substantial equivalence to the predicate device (INOVA QUANTA Lite™ PR-3) | Demonstrated comparability through: |
- Comparison study analyzing positive, equivocal, and negative sera | - Data show comparability in results for positive, equivocal, and negative sera with the predicate. |
- Results obtained for externally defined Calibrators and clinically defined sera | - Data show comparability in results for externally defined calibrators and clinically defined sera with the predicate. |
- Results obtained for samples from apparently healthy subjects (normal population) | - Data show expected results for samples from apparently healthy subjects. |
Performance according to state-of-the-art expectations and medical literature | "the assay performs as expected from the medical literature" and "performs according to state-of-the-art expectations." |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: Not explicitly stated in the provided text. The document mentions a "data set including results obtained within a comparison study analyzing positive, equivocal and negative sera," and "results obtained for externally defined Calibrators and clinically defined sera," and "results obtained for samples from apparently healthy subjects (normal population)." However, the exact number of samples for each category or overall is not provided.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "comparison study," which implies prospective collection for the purpose of the study. However, it's not definitively stated as retrospective or prospective, though the nature of comparing performance against an existing device suggests a concurrent, prospective analysis of samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Number of Experts: Not mentioned.
- Qualifications of Experts: Not mentioned.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not mentioned. The process of establishing "clinically defined sera" or the determination of "positive, equivocal and negative sera" is not detailed.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:
- MRMC Study: No. This study is an in-vitro diagnostic (IVD) assay comparison, not a multi-reader, multi-case study involving human readers interpreting images or data.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:
- Standalone Performance: Yes, in essence. The Varelisa PR3 ANCA is an automated in-vitro diagnostic (IVD) immunoassay kit. Its performance is evaluated directly based on its output (semi-quantitative and qualitative determination of PR3 ANCA) from human serum or plasma samples. There is no "human-in-the-loop" component in the performance of the device itself (though human laboratory personnel would run the test). The "algorithm" here is the biochemical reaction and detection system of the EIA.
7. The Type of Ground Truth Used:
The ground truth appears to be based on:
- Clinical Definition/Established Clinical States: "clinically defined sera." This likely refers to samples from patients with confirmed diagnoses of Wegener's granulomatosis (positive) and other relevant conditions or healthy controls (negative/equivocal).
- Predicate Device Results: The comparison study likely used the predicate device's results as a reference point for agreement for positive, equivocal, and negative sera.
- External Calibrators: Used to establish reference points within the assay itself.
- Medical Literature: Performance is deemed "as expected from the medical literature," implying existing knowledge about PR3 ANCA levels and their correlation with disease.
8. The Sample Size for the Training Set:
- Training Set Sample Size: Not applicable/not mentioned. This is an immunoassay kit, not a machine learning algorithm that requires a "training set" in the traditional sense. The device's components (antigen, controls, calibrators) are designed and manufactured based on biochemical principles rather than trained on a dataset.
9. How the Ground Truth for the Training Set Was Established:
- Ground Truth for Training: Not applicable. As noted above, this is not a machine learning model where a training set with established ground truth is used to train an algorithm. The "ground" for the development of such an IVD kit would be the established biochemical understanding of antigen-antibody interactions and the clinical relevance of PR3 ANCA.
Ask a specific question about this device
(46 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa Histone Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgG and IgM antibodies to histone in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE) or druginduced lupus erythematosus (DIL).
Varelisa Histone Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of histone antibodies in human serum or plasma. The wells of a microtiterplate are coated with human histone antigen. Antibodies specific for histones present in the patient sample bind to the antigen.
In a second step the enzyme labeled second antibody (conjugate) binds to the antigen-antibody complex which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution.
The rate of color formation from the chromogen is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
This submission is for an in-vitro diagnostic device, not an AI/ML powered device. Hence the provided sections of the 510k summary do not include specific "acceptance criteria" based on performance metrics like sensitivity, specificity, or AUC, nor do they detail a study designed to "prove" the device meets such criteria in the way one might expect for an AI/ML algorithm.
Instead, the submission focuses on demonstrating substantial equivalence to a previously legally marketed predicate device (INOVA QUANTA Lite™ Histone IgG). The "acceptance criteria" are implied by the demonstration of comparable performance and alignment with the predicate device regarding its intended use for aiding in the diagnosis of SLE or DIL.
Here's an interpretation based on the provided text, framed to address your requested points where applicable, and noting where information is not available due to the nature of the device:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is an in-vitro diagnostic (assay) device, the "acceptance criteria" are not explicitly stated as numerical performance targets (e.g., "sensitivity must be >90%"). Instead, they are interpreted as demonstrating comparable performance to the predicate device and expected results from medical literature. The "reported device performance" is qualitative and refers to the overall comparability.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Substantial Equivalence to Predicate Device: The device should perform comparably to the INOVA QUANTA Lite™ Histone IgG assay. | Laboratory equivalence supported by: |
- Results from a comparison study analyzing positive, equivocal, and negative sera.
- Results obtained for externally defined Calibrators.
- Results obtained for samples from apparently healthy subjects (normal population).
The data show that the assay performs as expected from the medical literature. |
| Accordance with Medical Literature: Performance should align with established medical understanding of histone antibodies in SLE/DIL diagnosis. | The data show that the assay performs as expected from the medical literature. |
| Intended Use Fulfilled: Designed for semiquantitative and qualitative determination of IgG and IgM antibodies to histone. | The device is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of histone antibodies in human serum or plasma. It detects Anti-Histone IgG and IgM antibodies. |
| Methodological Comparability: Indirect noncompetitive enzyme immunoassay, similar sample dilutions, comparable antigens, and detection systems. | Both assays are indirect noncompetitive enzyme immunoassays for the semiquantitative determination of antibodies against Histone in serum. Both recommend the same sample dilutions and use comparable antigens and detection systems. (Note: Varelisa detects IgG and IgM, predicate detects only IgG). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated. The text mentions "a comparison study analyzing positive, equivocal and negative sera" and "samples from apparently healthy subjects (normal population)," but no specific numbers are given for these cohorts.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- This information is not applicable and hence not provided for an in-vitro diagnostic assay comparison study. "Ground truth" for these types of tests typically refers to well-characterized patient samples with known clinical diagnoses or reference method results, rather than expert consensus on image interpretation.
4. Adjudication Method for the Test Set
- Not applicable and not provided for this type of device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not applicable. This is an in-vitro diagnostic assay, not an AI-assisted diagnostic tool that would involve human readers interpreting output.
6. Standalone Performance Study (Algorithm Only)
- Not applicable. This is an assay kit; its performance is its standalone performance when used in a laboratory setting. There isn't an "algorithm" in the typical sense of AI/ML. The "device performance" described in point 1 serves as its standalone performance.
7. Type of Ground Truth Used
- The "ground truth" implicitly used for this type of comparison would be clinical diagnosis (e.g., patients diagnosed with SLE/DIL vs. healthy controls) and/or reference laboratory methods/externally defined calibrators for the detection of histone antibodies, against which the assay results are correlated. The text mentions "positive, equivocal and negative sera" and "samples from apparently healthy subjects (normal population)," implying clinically characterized samples.
8. Sample Size for the Training Set
- Not applicable. This is an immunoassay kit whose reagents and protocol are developed based on scientific principles of antigen-antibody binding, not through a "training set" in the context of machine learning.
9. How Ground Truth for the Training Set Was Established
- Not applicable for the reasons mentioned in point 8.
Ask a specific question about this device
(45 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa ssDNA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of anti-single stranded DNA (ssDNA) autoantibodies in serum or plasma. In conjunction with the Varelisa dsDNA Antibodies kit it assists in the diagnosis of systemic lupus erythematosus (SLE) and certain other rheumatic diseases. The test is not definitive in isolation but has to be seen as one parameter in a multicriterion diagnostic process.
Varelisa ssDNA Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of ssDNA antibodies in human serum or plasma. Antibodies specific for ssDNA present in the patient sample bind to the antigen. The assay should be used in combination with the Varelisa dsDNA Antibodies.
The test kit contains microplate strips coated with synthetic ssDNA, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, Sample Diluent and wash buffer.
This document describes the Varelisa® ssDNA Antibodies, an indirect noncompetitive enzyme immunoassay for the semi-quantitative and qualitative determination of anti-single stranded DNA (ssDNA) autoantibodies in human serum or plasma. It is intended to assist in the diagnosis of systemic lupus erythematosus (SLE) and other rheumatic diseases, to be used in conjunction with a dsDNA antibodies kit.
1. Table of Acceptance Criteria and Reported Device Performance
The submission document states that "all available data support that the new device, PHARMACIA Varelisa ssDNA Antibodies Assay is substantially equivalent to the predicate device, INOVA QUANTA Lite™ ssDNA Assay, and that the new device performs according to state-of-the-art expectations." However, specific numerical acceptance criteria for performance metrics (such as sensitivity, specificity, accuracy) and their corresponding reported device performance values are not explicitly provided in the given text.
The data provided primarily focuses on demonstrating "comparability" and "substantial equivalence" to the predicate device rather than defining and meeting specific quantitative acceptance criteria. The text mentions "results obtained within a comparison study analyzing positive, equivocal and negative sera" and "results obtained for samples from apparently healthy subjects (normal population)" to support comparability.
Acceptance Criteria | Reported Device Performance |
---|---|
Not explicitly defined in numerical terms (e.g., sensitivity > X%, specificity > Y%) | Device performs "as expected from the medical literature" and is "suitable for serum and plasma samples." Confirmed to be "substantially equivalent" to the predicate device based on comparison studies across positive, equivocal, and negative sera, and healthy subjects. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample size used for the test set during the comparison study. It mentions a "data set including: results obtained within a comparison study analyzing positive, equivocal and negative sera" and "results obtained for samples from apparently healthy subjects (normal population)."
- Sample Size for Test Set: Not explicitly stated.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "comparison study" and involves "samples from apparently healthy subjects (normal population)," implying, but not confirming, a prospective collection for this specific study, or at least a retrospective analysis of collected samples for performance evaluation. It is not definitively stated if the data is retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not provide information on the number of experts used or their qualifications to establish the ground truth for the test set.
4. Adjudication Method for the Test Set
The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for the test set.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size with AI vs Without AI Assistance
This device is an in-vitro diagnostic (IVD) immunoassay kit, not an AI-powered diagnostic system involving human readers interpreting images or data with or without AI assistance. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance is not applicable and was not performed.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This refers to a standalone performance evaluation of the device itself (the immunoassay kit) in determining ssDNA antibodies. The document implies that the "comparison study analyzing positive, equivocal and negative sera" and "results obtained for samples from apparently healthy subjects" were indeed a standalone evaluation of the device against established classifications of these samples. The general principle of an immunoassay kit is to provide a result directly, so the performance described aligns with a "standalone" evaluation of the assay.
7. The Type of Ground Truth Used
The document states that the comparison study involved "positive, equivocal and negative sera" and "samples from apparently healthy subjects (normal population)." This indicates that the ground truth was based on:
- Clinical Diagnosis/Classification: Samples were likely categorized as positive, equivocal, or negative for ssDNA antibodies, or as coming from "apparently healthy subjects," based on established clinical criteria, potentially including other diagnostic tests, clinical symptoms, and expert consensus.
- External Calibrators: "Results obtained for externally defined Calibrators" were also used, suggesting the use of reference materials with known concentrations.
8. The Sample Size for the Training Set
The document does not explicitly mention a training set in the context of an AI/machine learning model. This device is an immunoassay kit for antibody detection, not a machine learning algorithm that requires a distinct training set. The "data set" mentioned seems to be for performance evaluation/comparison rather than training.
9. How the Ground Truth for the Training Set Was Established
As a training set for an AI/machine learning model is not applicable in this context, the method for establishing its ground truth is not mentioned.
Ask a specific question about this device
(81 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa ß2-Glycoprotein I IgA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of ß2-glycoprotein I IgA antibodies in serum or plasma.
The presence of B2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases.
The Varelisa B2-Glycoprotein I IgA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of ß2-glycoprotein I IgA antibodies in serum or plasma.
The test kit contains microplate strips coated with human purified ß2glycoprotein I, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, buffered diluent and wash buffer.
Here's an analysis of the provided text regarding the Varelisa® ß2 Glycoprotein I IgA Antibodies device, structured to address your specific questions about acceptance criteria and study details.
Please note: The provided text is a 510(k) summary, which often focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive de novo performance data with detailed acceptance criteria and standalone studies in the way a PMA might. As such, some of your requested information (e.g., specific quantitative acceptance criteria, standalone performance details, MRMC study results) is not explicitly present in the provided document. I will extract what is available and indicate when information is not provided.
Acceptance Criteria and Reported Device Performance
The provided 510(k) summary describes laboratory equivalence to a predicate device. It does not explicitly state quantitative acceptance criteria in a table format with specific thresholds. Instead, it relies on demonstrating comparable performance to the predicate device and expected behavior based on medical literature.
Acceptance Criteria (Implied/General) | Reported Device Performance (Summary) |
---|---|
Comparability with Predicate Device (INOVA QUANTA Lite™ ß2 GPI IgA) | - Results obtained within a comparison study analyzing positive, equivocal, and negative sera showed comparability. |
- Results for externally defined calibrators showed comparability.
- Results for samples from apparently healthy subjects (normal population) showed comparability. |
| Suitability for Serum and Plasma Samples | - The device is outlined for use with serum and plasma, with corresponding performance data underlining its effectiveness with plasma as a sample. |
| Performance in line with medical literature | - The assay performs as expected from the medical literature regarding the diagnosis of thrombotic disorders related to primary Antiphospholipid Syndrome or secondary to SLE/other autoimmune diseases. |
| Decision Point Evaluation Method | - The Varelisa assay uses an OD cutoff for evaluation, and corresponding performance data show the comparability of results to the predicate device which uses a decision point method. |
2. Sample Sizes and Data Provenance for Test Set
- Sample Size for Test Set: Not explicitly stated as a single number. The document mentions a "data set including results obtained within a comparison study analyzing positive, equivocal and negative sera," and "results obtained for samples from apparently healthy subjects (normal population)." The specific number of samples in these comparison studies is not detailed.
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be clinical laboratory evaluations. The document refers to "the medical literature" and "scientific knowledge" as a basis for the intended use and performance expectations, implying a broader, though unspecified, provenance for general knowledge. The manufacturer is based in Germany.
- Retrospective/Prospective: Not explicitly stated. Given it's a 510(k) comparison study, it's often retrospective use of banked samples, but this is not confirmed.
3. Number of Experts and Qualifications for Ground Truth (Test Set)
This device is an in-vitro diagnostic (IVD) immunoassay for autoantibodies. The "ground truth" for such devices typically relies on a combination of:
- Clinical diagnosis based on established medical criteria (e.g., for SLE, Antiphospholipid Syndrome).
- Results from a well-characterized predicate immunoassay or reference method.
- Clinical status of patients (e.g., "apparently healthy subjects").
The document does not mention the use of human experts (e.g., radiologists) establishing ground truth in the way it would for an imaging AI device. The ground truth seems to be implicitly derived from the clinical status of the patients and the results of the predicate device.
4. Adjudication Method for Test Set
No explicit adjudication method is described. The "ground truth" for an immunoassay comparison study is usually established by the clinical diagnosis of the patient and/or the result from the predicate device (or a gold standard if available), rather than a panel of expert reviewers adjudicating individual results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was done. This type of study (evaluating human reader performance with and without AI assistance) is relevant for imaging AI devices that assist human interpretation. This device is an immunoassay kit for laboratory use and does not involve human "readers" in the diagnostic interpretation in the same way.
6. Standalone (Algorithm Only) Performance Study
Yes, a standalone performance was done in the sense that the device's performance was evaluated independently and then compared to a predicate device. The comparison studies evaluating "positive, equivocal and negative sera," "externally defined Calibrators," and "samples from apparently healthy subjects" represent evaluations of the device's standalone performance characteristics against expected outcomes or the predicate device. However, a specific "standalone" performance metric like sensitivity/specificity against a definitive gold standard not involving the predicate is not explicitly presented, as the emphasis is on substantial equivalence.
7. Type of Ground Truth Used
The ground truth appears to be a combination of:
- Clinical Status/Diagnosis: The intended use links the presence of antibodies to aiding in the diagnosis of thrombotic disorders related to Antiphospholipid Syndrome, SLE, or other autoimmune diseases. Samples were collected from "apparently healthy subjects" and presumably patients with known clinical conditions.
- Predicate Device Results: The entire study is a "comparison study analyzing positive, equivocal and negative sera" in relation to the predicate device, thereby using the predicate device's established performance as a de facto reference for "truth" in the context of substantial equivalence.
8. Sample Size for the Training Set
This document does not specify a separate "training set" or its size. This is typical for traditional IVDs like immunoassays, where extensive "training" in the machine learning sense is not performed. Assay development involves optimizing reagents, protocols, and cutoffs, which are different from training an ML algorithm on a dataset. The studies described are more akin to verification and validation.
9. How Ground Truth for the Training Set Was Established
As there is no "training set" in the context of an AI/ML algorithm, this question is not applicable. The development and optimization of the immunoassay would rely on established laboratory practices, chemical principles, and clinical samples previously characterized by the predicate device or clinical diagnosis.
Ask a specific question about this device
(81 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa ß2-Glycoprotein I IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of ß2-glycoprotein I IgM antibodies in serum or plasma.
The presence of ß2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases.
The Varelisa B2-Glycoprotein I IgM Antibodies Assay is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of B2-glycoprotein I IgM antibodies in serum or plasma.
The test kit contains microplate strips coated with human purified ß2glycoprotein I, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, buffered diluent and wash buffer.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly state numerical acceptance criteria (e.g., sensitivity, specificity thresholds) for the Varelisa® B2-Glycoprotein I IgM Antibodies device. Instead, the study aims to demonstrate substantial equivalence to a predicate device, the INOVA QUANTA Lite™ ß2 GPI IgM. The reported performance is based on qualitative comparisons and the conclusion of substantial equivalence.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Comparability to predicate device (INOVA QUANTA Lite™ ß2 GPI IgM) | "The comparability of QUANTA LiteTM ß2 GPI IgM and Varelisa B2-Glycoprotein I IgM Antibodies is supported by a data set including: |
- results obtained within a comparison study analyzing positive, equivocal and negative sera.
- results obtained for externally defined Calibrators.
- results obtained for samples from apparently healthy subjects (normal population)."** |
| Effectiveness with serum samples | Implied to be effective as it's a common sample type for this assay type and compared to a predicate using serum. |
| Effectiveness with plasma samples | "Corresponding performance data underline the effectiveness of the assay with plasma as sample." |
| Performance as expected from medical literature | "The data show that the assay performs as expected from the medical literature." |
| Comparability of evaluation methods (OD-cutoff vs. decision point) | "Corresponding performance data show the comparability of the results." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The exact numerical sample size for the test set (comparison study analyzing positive, equivocal, and negative sera; externally defined calibrators; normal population) is not specified in the provided text.
- Data Provenance: The text does not specify the country of origin of the data. It mentions "externally defined Calibrators" and "samples from apparently healthy subjects (normal population)". It is implied to be a retrospective study as it's a comparison study to a predicate.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the text. The study focuses on demonstrating comparability between two assay kits, rather than establishing a gold standard diagnosis by human experts for the test set itself. The "ground truth" for the samples would have been their established status as positive, negative, or equivocal for the autoantibody.
4. Adjudication Method for the Test Set
This information is not provided in the text. Given it's a serological assay comparison, expert adjudication in the typical sense (e.g., for imaging interpretation) is unlikely to be the primary method for test set ground truth. The initial classification of samples (positive, negative, equivocal) would have been based on established laboratory methods or clinical diagnoses.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is more common for diagnostic imaging AI devices where human interpretation is a key component. The Varelisa® B2-Glycoprotein I IgM Antibodies device is an in-vitro diagnostic (IVD) assay kit for laboratory use, not an AI-powered diagnostic tool for human reader improvement.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone performance assessment was effectively done. The device itself is an in-vitro diagnostic assay that provides a result (semiquantitative or qualitative determination of antibodies) without direct human interpretation in the sense of an "algorithm only". The comparison study assesses the performance of this standalone assay against a predicate standalone assay. The results are based on the assay's output (e.g., optical density, antibody units), not on a human's interpretation of those results.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, Etc.)
The ground truth for the comparison study would likely have been established by:
- Clinical Diagnosis / Established Patient Status: Samples categorized as "positive, equivocal, and negative sera" would have their status determined by established clinical diagnoses of thrombotic disorders, SLE, or other autoimmune diseases, in conjunction with other laboratory findings.
- Reference Method Results: For the "externally defined Calibrators," their true value would be established by a reference method or traceable standard.
- Healthy Population Screening: "Samples from apparently healthy subjects (normal population)" would be considered negative by definition for the condition.
The text does not explicitly state "pathology" or "outcomes data" as the direct ground truth, but clinical diagnosis often relies on these.
8. The Sample Size for the Training Set
This information is not applicable / not provided for this type of medical device submission. The Varelisa® B2-Glycoprotein I IgM Antibodies is a chemical reagent-based assay kit, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The device's performance characteristics are established through analytical and clinical validation studies, not an algorithmic training process.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable / not provided as there is no "training set" for this type of device.
Ask a specific question about this device
(81 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa B2-Glycoprotein I IgG Antibodies EIA kit is designed for the semiguantitative and qualitative determination of ß2-glycoprotein I IgG antibodies in serum or plasma.
The presence of ß2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases.
Varelisa B .- Glycoprotein I IgG Antibodies Assay is an indirect The noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of ß2-glycoprotein I IgG antibodies in serum or plasma.
The test kit contains microplate strips coated with human purified ß2glycoprotein I, calibrators, positive and negative controls, enzyme-labeled conjugate, substrate and substrate stop solution, buffered diluent and wash buffer.
The provided text describes the Varelisa® ß2 Glycoprotein I IgG Antibodies EIA kit and its comparison to a predicate device, but it lacks specific acceptance criteria as quantitative thresholds and a detailed study report that explicitly states how these criteria were met.
However, based on the provided text, I can infer the implied acceptance criteria, the general nature of the study conducted, and the reported device performance.
Here's an analysis of the information requested, based on the text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical acceptance criteria. Instead, it focuses on the comparability to a predicate device and performance as expected from medical literature.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Substantial Equivalence to Predicate Device: | |
- Comparable performance to INOVA QUANTA Lite™ ß2 GPI IgG Assay for semi-quantitative and qualitative determination of IgG antibodies against ß2-glycoprotein I in serum. | - "The comparability of QUANTA LiteTM B2 GPI IgG and Varelisa B2-Glycoprotein I IgG Antibodies is supported by a data set including results obtained within a comparison study analyzing positive, equivocal and negative sera." |
- "Corresponding performance data show the comparability of the results" (referring to OD-cutoff vs. decision point method for evaluation).
- "All available data support that the new device...is substantially equivalent to the predicate device." |
| Suitability for Intended Use (Serums & Plasma): | |
| - Effective for use with serum samples. | - "The data show that the assay performs as expected from the medical literature." |
| - Effective for use with plasma samples. | - "Corresponding performance data underline the effectiveness of the assay with plasma as sample." - "The data show that the device is suitable for serum and plasma samples." |
| Accuracy/Reliability (General): | |
| - Performs as expected from medical literature. | - "The data show that the assay performs as expected from the medical literature." - "The new device performs according to state-of-the-art expectations." |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated. The text mentions "a comparison study analyzing positive, equivocal and negative sera" and "results obtained for samples from apparently healthy subjects (normal population)," but no specific numbers of samples are provided for these test sets.
- Data Provenance: Not explicitly stated. The manufacturer is based in Germany (Pharmacia Deutschland GmbH), suggesting the study could have been conducted in Germany, but this is not confirmed. It is a retrospective or prospective study is not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The method for establishing ground truth for the comparison study is not detailed.
4. Adjudication method for the test set
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not applicable here. This device is an in vitro diagnostic (IVD) immunoassay kit for laboratory analysis, not an AI-assisted diagnostic tool that helps human readers (e.g., radiologists) interpret images. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable as it's an IVD kit, not an algorithm that operates independently without human interaction (even lab technicians are "in the loop" for running the test and interpreting the results within the context of the patient's clinical picture). The device itself is the "standalone" diagnostic tool in a lab setting, providing a semi-quantitative or qualitative result.
7. The type of ground truth used
The type of ground truth is implied to be clinical diagnosis or established antibody status. The mention of "positive, equivocal and negative sera" suggests that the samples had a pre-determined status based on clinical findings, other laboratory tests, or established reference standards for ß2-glycoprotein I IgG antibodies. The intended use statement also highlights its role "in conjunction with clinical findings and other laboratory tests to aid in the diagnosis," which points towards clinical correlation as the ultimate ground truth.
8. The sample size for the training set
This information is not provided in the document. The document describes a comparison study, but not a distinct "training set" in the context of machine learning or algorithm development. The development of the immunoassay itself would involve internal validation and optimization, but the term "training set" is not used or detailed.
9. How the ground truth for the training set was established
As no training set is explicitly mentioned, how its ground truth was established is not discussed in the document.
Ask a specific question about this device
(81 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa B .- Glycoprotein I Antibodies Screen EIA kit is designed for the qualitative determination of B2-Glycoprotein I antibodies in human serum or plasma.
The presence of B2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases.
The Varelisa ß2-Glycoprotein I Antibodies Screen is an indirect noncompetitive enzyme immunoassay for the qualitative determination of antibodies against ß glycoprotein I in serum or plasma.
The test kit contains microplate strips coated with human purified ß2glycoprotein I. calibrator, negative control, enzyme-labeled conjugate, substrate and substrate stop solution, buffered diluent and wash buffer.
The provided text describes a 510(k) summary for a medical device called Varelisa® ß2-Glycoprotein I Antibodies Screen. While it discusses the device's intended use and general performance, it does not explicitly state formal acceptance criteria in a quantifiable manner or detail a specific study designed to "prove" the device meets such criteria to the extent typically expected in a scientific study report.
Instead, the document focuses on demonstrating substantial equivalence to a predicate device (INOVA QUANTA Lite™ B2 GPI Screen) through comparative performance data. The "study" mentioned is a "comparison study analyzing positive, equivocal and negative sera," along with results for external calibrators and samples from healthy subjects.
Here's an attempt to extract the requested information based on the provided text, acknowledging limitations due to the nature of the 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
As specific quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or agreement percentages) are not explicitly stated in the provided text, this table will represent the general comparative findings presented to establish substantial equivalence.
Acceptance Criterion (Inferred from Substantial Equivalence Goal) | Reported Device Performance (Varelisa® ß2-Glycoprotein I Antibodies Screen vs. Predicate) |
---|---|
Qualitative determination of B2-Glycoprotein I antibodies in human serum. | The new device is "substantially equivalent" to predicate for qualitative determination in serum. |
Qualitative determination of B2-Glycoprotein I antibodies in human plasma. | Performance data "underline the effectiveness of the assay with plasma as sample," unlike the predicate which is "only recommended for use in serum." |
Comparability with predicate device across positive, equivocal, and negative sera. | "The comparability... is supported by a data set including results obtained within a comparison study analyzing positive, equivocal and negative sera." |
Performance with externally defined Calibrators. | "Results obtained for externally defined Calibrators" support comparability. |
Performance with samples from apparently healthy subjects (normal population). | "Results obtained for samples from apparently healthy subjects (normal population)" support comparability. |
Performance according to current scientific knowledge regarding B2-glycoprotein I antibodies and antiphospholipid syndrome. | "Intended use... was adapted to the current state of scientific knowledge." The device "performs as expected from the medical literature." |
OD cut-off method for evaluation. | "Corresponding performance data show the comparability of the results" with the predicate's decision point method. |
Note: The document states, "The data show that the assay performs as expected from the medical literature. Furthermore the performance data show that the device is suitable for serum and plasma samples." This implies that the 'acceptance criteria' were met by demonstrating similar performance to the legally marketed predicate and suitability for both serum and plasma.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated. The text mentions "a comparison study analyzing positive, equivocal and negative sera," "results obtained for externally defined Calibrators," and "results obtained for samples from apparently healthy subjects (normal population)." However, the specific number of samples in each category tested is not provided.
- Data Provenance: Not explicitly stated (e.g., specific country of origin or whether samples were prospective or retrospective). The manufacturer is based in Germany, so it's plausible the data collection predominantly occurred there or in Europe, but this is not confirmed. The study used "human serum or plasma."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This information is not provided in the document. For immunological assays like this, ground truth is typically established by clinical diagnosis, other laboratory tests, or established consensus among medical professionals. The document simply states the device aids in diagnosis "in conjunction with clinical findings and other laboratory tests."
4. Adjudication Method for the Test Set
This information is not provided. The text does not describe any expert adjudication process for the test results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The device is an in-vitro diagnostic (IVD) immunoassay designed for qualitative determination of antibodies, not an imaging device typically assessed with MRMC studies or involving human readers interpreting outputs in such a way. The focus is on the device's accuracy in detecting specific antibodies rather than human interpretation of complex data. Therefore, there is no mention of human readers or an effect size for human improvement with AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes. The data presented appears to refer to the standalone performance of the Varelisa® ß2-Glycoprotein I Antibodies Screen device itself, comparing its output to that of the predicate device and potentially to clinical expectations. As an automated immunoassay, its performance is inherently standalone in generating a result, which is then interpreted by a clinician in the context of other findings.
7. Type of Ground Truth Used
The type of ground truth used is not explicitly detailed for the specific comparison study. However, for an immunoassay, the implicit ground truth would typically be established based on:
- Clinical diagnosis: Patients with primary Antiphospholipid Syndrome or SLE.
- Other laboratory tests: Results from established assays for B2-glycoprotein I antibodies or other relevant markers.
- Disease status: Classification of samples as "positive, equivocal and negative sera" implies a prior determination of their status, likely based on a combination of clinical criteria and/or predicate device results.
8. Sample Size for the Training Set
The document is a 510(k) summary for an immunoassay kit. Immunoassays are typically "trained" during their development and optimization phases, but the process is different from machine learning algorithms. The text does not provide any information about a "training set" or its size in the context of device development or any potential algorithm used.
9. How the Ground Truth for the Training Set Was Established
As there is no mention of a training set in the provided text in the context of a machine learning algorithm, there is no information on how its ground truth would have been established. For traditional immunoassay development, the "ground truth" for calibrator and control materials is established through rigorous internal characterization and standardization, often against reference materials. However, details of this are not included here.
Ask a specific question about this device
(56 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa Cardiolipin IgA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
The Varelisa Cardiolipin IgA Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in human serum or plasma. The wells of a microplate are coated with bovine cardiolipin antigen. Antibodies specific for cardiolipin present in the patient sample bind to the antigen. In a second step an enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of Conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
The provided text describes a 510(k) submission for a modified medical device, the Varelisa® Cardiolipin IgA Antibodies assay. The submission primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a standalone study proving its performance against those criteria.
However, based on the information provided, we can infer some aspects related to the study conducted for comparison.
Here's an analysis of the provided information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical acceptance criteria (e.g., specific sensitivity, specificity, accuracy thresholds). The primary "acceptance criteria" for a 510(k) submission, as implied, is demonstrating substantial equivalence to a predicate device.
The reported device performance is described qualitatively in the "Laboratory equivalence" section:
Acceptance Criteria (Inferred from Substantial Equivalence) | Reported Device Performance |
---|---|
Comparability with predicate device (Varelisa® Cardiolipin (IgA) Antibodies) | Data from a comparison study analyzing positive, equivocal, and negative sera, externally defined Calibrators, and samples from apparently healthy subjects show comparability. |
Assay performs as expected from medical literature | "The data show that the assay performs as expected from the medical literature." |
Substantial equivalence to predicate device | "all available data support that the new/modified device... is substantially equivalent to the predicate/original device..." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not explicitly stated. The document mentions "positive, equivocal and negative sera," "externally defined Calibrators," and "samples from apparently healthy subjects (normal population)," but no specific numbers.
- Data Provenance: Not explicitly stated. It's likely the study was conducted in Germany, given the manufacturer's location (Pharmacia Deutschland GmbH) and the company contact person's address. The type of study (retrospective or prospective) is also not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. The ground truth for this type of in vitro diagnostic device (immunological test) is typically established by the results of the predicate device, known positive/negative samples, or established clinical definitions of the condition, rather than expert interpretation of images or clinical cases. There's no mention of experts establishing ground truth in the context of radiologists or similar roles.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This concept is relevant for studies where human interpretation or consensus is required to establish ground truth or classify cases. For an immunoassay, the results are quantitative or semi-quantitative and typically don't require adjudication between human readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an in vitro diagnostic (IVD) immunoassay, not an AI-assisted diagnostic tool that aids human readers. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, implicitly. The "comparison study" and "laboratory equivalence" assessment described refer to the performance of the device itself (the assay) as a standalone diagnostic tool. The purpose is to show that the modified assay yields comparable results to the predicate assay.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth appears to be established by:
- Reference to the predicate device (Varelisa® Cardiolipin (IgA) Antibodies) through comparison studies with positive, equivocal, and negative sera.
- Externally defined Calibrators, which are reference materials with known concentrations.
- Samples from apparently healthy subjects, which serve as negative controls.
- The overall understanding of the medical literature regarding anti-cardiolipin antibodies in the context of APS and SLE.
8. The sample size for the training set
Not applicable in the conventional sense of machine learning. This is an immunoassay, not a machine learning algorithm that requires a "training set." The assay itself is a chemical and biological measurement system.
9. How the ground truth for the training set was established
Not applicable, as it's not a machine learning model with a training set. The assay's "calibration" would involve using known standards and calibrators, as mentioned, but these are part of the assay's operational parameters rather than a "training set" for an algorithm.
Ask a specific question about this device
(56 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa Cardiolipin IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
The Varelisa Cardiolipin IgM Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in human serum or plasma. The wells of a microplate are coated with bovine cardiolipin antigen. Antibodies specific for cardiolipin present in the patient sample bind to the antigen. In a second step an enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex.
The provided text is a 510(k) summary for the Varelisa® Cardiolipin IgM Antibodies device. It describes the device's intended use, principle, and establishes substantial equivalence to a predicate device. However, it does not contain the detailed performance study results, acceptance criteria, or most of the specific information requested in your prompt.
Specifically, the document does not contain:
- A table of acceptance criteria and reported device performance.
- Sample size used for the test set or its provenance.
- Number of experts, their qualifications, or adjudication method for ground truth.
- Information on a multi-reader multi-case (MRMC) comparative effectiveness study or effect size.
- Details about a standalone (algorithm only) performance study.
- The type of ground truth used (beyond implying clinical correlation for the intended use).
- Sample size for the training set or how its ground truth was established.
The document focuses on establishing substantial equivalence based on:
- Laboratory equivalence: "results obtained within a comparison study analyzing positive, equivocal and negative sera," "results obtained for externally defined Calibrators," and "results obtained for samples from apparently healthy subjects (normal population)."
The intended use is described as:
"The Varelisa Cardiolipin IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE)."
The "study" or evidence to support substantial equivalence is limited to a laboratory equivalence comparison between the modified device and the predicate device, analyzing various serum samples and calibrators. No specific performance metrics (like sensitivity, specificity, accuracy) or acceptance criteria are stated from this comparison. The document states that the "data show that the assay performs as expected from the medical literature," implying clinical relevance, but doesn't quantify this performance.
Ask a specific question about this device
(56 days)
PHARMACIA DEUTSCHLAND GMBH
The Varelisa Cardiolipin IgG Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgG antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
The Varelisa Cardiolipin IgG Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of IgG antibodies against cardiolipin in human serum or plasma. The wells of a microplate are coated with bovine cardiolipin antigen. Antibodies specific for cardiolipin present in the patient sample bind to the antigen. In a second step an enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of Conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
Here's a breakdown of the acceptance criteria and study information for the Varelisa® Cardiolipin IgG Antibodies device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly state specific numerical acceptance criteria for the device. Instead, it focuses on demonstrating "laboratory equivalence" and "substantial equivalence" to a predicate device. The performance is summarized qualitatively.
Acceptance Criteria Category | Reported Device Performance | Comments |
---|---|---|
Laboratory Equivalence | The new device performs as expected from medical literature. Comparisons showed that results obtained within a comparison study analyzing positive, equivocal, and negative sera, as well as results for externally defined Calibrators and Quality Assessment sera, and samples from apparently healthy subjects, supported comparability. Minor differing results noted were probably due to a changed blocking procedure, but the overall conclusion was substantial equivalence. | The criteria are implicitly met by demonstrating that the new device's performance aligns with the predicate and scientific understanding. Specific quantitative targets for agreement (e.g., % positive agreement, % negative agreement) are not provided. |
Intended Use | The device is designed for semiquantitative and qualitative determination of IgG antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE). | The intended use statement itself serves as a high-level criterion, and the study aims to show the device fulfills this use. |
Substantial Equivalence | All available data support that the new/modified device is substantially equivalent to the predicate/original device, and performs according to state-of-the-art expectations. | This is the overarching "acceptance criterion" for 510(k) clearance in this context. |
2. Sample Sizes and Data Provenance for the Test Set
- Sample Size: The document mentions "positive, equivocal and negative sera," "externally defined Calibrators and Quality Assessment sera," and "samples from apparently healthy subjects (normal population)." However, the specific number of samples in each category tested for the comparison study is not provided.
- Data Provenance: Not explicitly stated. Given the manufacturer (Pharmacia Deutschland GmbH) is in Germany, it is reasonable to infer that some data might be from Europe, but this is not confirmed. The document does not specify whether the data was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth
- The document does not mention the use of experts to establish ground truth for the test set. The nature of the comparison (comparing to a predicate device and "externally defined Calibrators and Quality Assessment sera") suggests that the "ground truth" for the test set was likely derived from the established values of these calibrators and known statuses of the clinical samples (positive, negative, equivocal) as determined by the predicate device or other reference methods.
4. Adjudication Method
- Not applicable/Not mentioned. There is no indication of an adjudication method used, as expert consensus for ground truth determination is not described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This document describes the validation of an in vitro diagnostic (IVD) test kit, which typically does not involve human readers interpreting results in the same way an imaging or pathology AI system would. The "reading" is done by the instrument quantifying antibody levels.
6. Standalone (Algorithm Only) Performance Study
- Yes, this is essentially a standalone performance assessment. The device (an EIA kit) operates without a human-in-the-loop interpreting the primary output. Its performance is evaluated based on its ability to detect and quantify antibodies in samples.
7. Type of Ground Truth Used
-
The ground truth appears to be based on:
- Comparison to a Predicate Device: The performance of the new device was compared against the Varelisa® Cardiolipin (IgG) Antibodies (the original/predicate device).
- Externally Defined Calibrators and Quality Assessment Sera: These materials would have established, reference values for the analytes.
- Known Clinical Status: Samples were categorized as "positive, equivocal and negative sera," implying that their actual status for cardiolipin antibodies was known through other means (likely the predicate or reference methods).
It does not explicitly state "pathology," "expert consensus," or "outcomes data" as the primary form of ground truth for analytical performance, though the clinical relevance (APS diagnosis, thrombotic risk) implies underlying clinical outcomes would have defined the initial understanding of cardiolipin antibodies.
8. Sample Size for the Training Set
- The document does not discuss a separate "training set" in the context of an AI/algorithm. This is an immunoassay kit, not a machine learning model that typically requires a distinct training phase. The " Comparability of the new and the old version" section describes the studies used to demonstrate equivalency.
9. How Ground Truth for the Training Set Was Established
- Not applicable. As this is an immunoassay kit, rather than an AI algorithm requiring a training set, the concept of establishing ground truth for a training set does not apply in the conventional sense. The "training" for such a device is in its biochemical design and optimization during development, validated against known standards and clinical samples.
Ask a specific question about this device
Page 1 of 1