Search Results
Found 1 results
510(k) Data Aggregation
(364 days)
The STIWELL med4 is a neuromuscular electronic stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator the STIWELL med4 is indicated for the following conditions:
- Relaxation of muscle spasm
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
As a transcutaneous electrical nerve stimulator for pain relief the STIWELL med4 is indicated for the following conditions:
- Symptomatic relief and management of chronic (long-term), intractable pain
- Adjunctive treatment in the management of post-surgical pain and post traumatic acute pain
As a biofeedback device the STIWELL med4 is indicated for the following conditions:
- Biofeedback, relaxation and muscle re-education purposes
As an external functional neuromuscular stimulator the STIWELL med4 is indicated for the following conditions:
- Helps to relearn voluntary motor functions of the extremities
As a nonimplanted electrical continence device the STIWELL med4 is indicated for the following conditions:
- Acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Inhibition of the detruser muscles through reflexive mechanisms and strengthening of pelvic floor muscles
- Incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles such as the abdominal and the gluteus muscles
The STIWELL med4 is powered by rechargeable batteries. It has four stimulation channels and two EMG measurement channels.
The STIWELL med4 is intended for stationary use in a hospital as well as home use by the patient. The physician/therapist has the flexibility to adjust the programs and monitors the progress of the therapy. Statistics regarding the completed treatments can be retrieved from the STIWELL med4 and from the PC.
The provided document is a 510(k) premarket notification for the STIWELL med4, a neuromuscular electronic stimulator. It focuses on establishing substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information regarding clinical study design, performance metrics, and expert adjudication is not present in this type of submission.
Here's an analysis based on the available information:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are primarily based on demonstrating substantial equivalence to legally marketed predicate devices, meaning the new device has the same intended use and the same technological characteristics as the predicate device, or has different technological characteristics but does not raise different questions of safety and effectiveness and is as safe and effective as a legally marketed device.
The document provides extensive tables comparing the STIWELL med4's technical characteristics across various stimulation programs (Functional Electrical Stimulation, Incontinence, TENS, Biofeedback) with its predicate devices. The "performance" in this context is the successful demonstration that these characteristics are similar enough to existing devices to not introduce new safety or effectiveness concerns.
Summary Table of Equivalence (Acceptance Criteria are inferred from predicate device characteristics):
| Characteristic/Program Type | Acceptance Criteria (inferred by predicate) | STIWELL med4 Performance (Reported) | Notes |
|---|---|---|---|
| General Unit Characteristics | |||
| Power Source | Battery Pack (NiMH or Alkaline) | Battery Pack Li-Ion 11,1V | Different battery chemistry, but the submission implicitly argues it does not raise new safety/effectiveness concerns. |
| Method of Line Current Isolation | Medical Class II Power Adapter/N/A | Medical Class II Power Adapter | Consistent. |
| Patient Leakage Current | N/A (Battery) | N/A (Battery) | Consistent. |
| Number of Output Modes | 1 | 1 | Consistent. |
| Number of Output Channels | 2 or 4 | 4 | Consistent with some predicates, exceeds others (e.g., Elpha II3000 has 2). |
| Number of EMG (input) Channels | 1 or 2 (or N/A) | 2 | Consistent with predicates that have EMG (e.g., Myotrac Infinity, Mentamove). |
| EMG Sensitivity | <0.1μV or 1μV | 1μV | Within range of predicate devices. |
| EMG Sampling Rate | 2.048kHz or Unknown | 3kHz | Higher than Myotrac Infinity (2.048kHz), but within acceptable range for biofeedback functionality. |
| EMG detection | Bipolar | Bipolar | Consistent. |
| EMG Range | 0-2000μV or 2-2000μV | 1-2000μV | Within range of predicate devices. |
| EMG Bandwidth | 10Hz-1kHz or Unknown | 70-480Hz | Falls within the broader range of predicate devices. |
| EMG Signal Processing | RMS or Unknown | AVR (Average Rectified Value) | Different method from Myotrac Infinity (RMS), but implicitly argued to not raise new safety/effectiveness concerns. The submission suggests this is equivalent in function. |
| Software/Firmware/Microprocessor Contr | Yes | Yes | Consistent. |
| Automatic Overload Trip? | Yes | Yes | Consistent. |
| Automatic No-Load Trip? | Yes | Yes | Consistent. |
| Automatic Shut Off? | Yes (e.g., 10min) | Yes (10min) | Consistent. |
| Patient Override Control? | Yes (Stop Button) | Yes (Stop Button) | Consistent. |
| Display Indicators | Yes (On/Off, Low Batt, V/C Level) | Yes (On/Off, Low Batt, V/C Level) | Consistent. |
| Compliance with Voluntary Standards | IEC 60601-1, -1-2, -2-10 (etc.) | IEC 60601-1, -1-2, -2-10 (etc.) | The device passed the equivalent standards as the predicate devices and additional environmental/electrical tests. |
| Output Specifications (Stimulation) | |||
| Waveform | Biphasic symmetrical/asymmetrical | Biphasic symmetrical | Consistent with some predicates (e.g., Mentamove, Compex 2). Elpha models have asymmetrical. The selection of symmetrical is a technological characteristic; the submission implies it does not raise new safety or effectiveness concerns compared to asymmetrical. |
| Shape | Rectangular (with/without discharge) | Rectangular | Consistent. |
| Max Output Voltage (500Ω/2kΩ) | Similar to predicates | 50V/115V | Within ranges or similar to predicate devices. |
| Max Output Current (500Ω/2kΩ) | Similar to predicates | 100mA/58mA | Within ranges or similar to predicate devices. |
| Pulse Width | 50μs-400μs, etc. | 50-400μs (FES), 500μs/300μs/350-500μs (Inc.) | Within or overlapping ranges of predicate devices. Slight differences in incontinence programs are noted but are implicitly presented as equivalent. |
| Frequency | 1-140Hz, 1-10Hz, etc. | Ranges vary by program | Similar to, within, or overlapping ranges of predicate devices across different program types. |
| Net Charge (μC per pulse) | 0μC | 0μC | Consistent across all stimulation types, achieved by "some positive and negative impulse" or "output capacitor." |
| Max Phase Charge (μC) | Similar to predicates | Varies by program | Within or similar to predicate device values. For incontinence, New Device has 50μC and predicate has 40μC (program U1), and New Device has 30μC while predicate has 40μC (program U2). These differences are implicitly argued to be acceptable for substantial equivalence. |
| Max Current Density (mA/cm²) | Similar to predicates | Varies by program | Values are provided for the new device and compared to predicates. In some incontinence programs, the new device's current density (4.7mA/cm²) is lower than the predicate (Vaginal: 20.4mA/cm², Anal: 33.3mA/cm²), which could be considered within acceptable safety margins. |
| Max Power Density (mW/W/cm²) | Similar to predicates | Varies by program | Values are provided for the new device and compared to predicates. Similar to current density, in some incontinence programs, the new device's power density is significantly lower than the predicate (e.g., 23.5 μW/cm² vs. Vaginal: 2.3mW/cm², Anal: 3.8mW/cm²), indicating a potentially safer profile for these differences. |
| ON/OFF Times | Ranges similar to predicates | Ranges vary by program | Similar to or overlapping various predicate device ranges. |
Detailed Study Information from the Provided Text:
This submission is a 510(k) Premarket Notification, which primarily relies on demonstrating substantial equivalence to predicate devices already on the market. It typically does not involve prospective clinical trials to prove efficacy or performance against specific, pre-defined quantitative acceptance criteria in the same way a PMA (Premarket Approval) application would.
Therefore, many of the specific details requested below regarding clinical studies (sample size, data provenance, expert adjudication, MRMC studies, standalone performance, training sets) are generally not part of a 510(k) submission unless there are significant technological differences requiring new clinical data to establish safety and effectiveness. The document itself states, "Tests have been carried out according to the following standards," referring to various IEC and ISO standards for electrical safety, EMC, and environmental testing, confirming that the primary testing was focused on meeting recognized standards and comparing technical specifications.
-
A table of acceptance criteria and the reported device performance:
- See the table above for inferred acceptance criteria (based on predicate device characteristics) and the STIWELL med4's reported performance against those. The "acceptance criteria" here are implicitly defined by the technical specifications and safety features of the predicate devices.
-
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not Applicable (N/A) for clinical data. This document outlines a 510(k) submission based on substantial equivalence and engineering/safety testing, not new clinical trials.
- The "test set" primarily refers to the device itself being tested in a lab setting against various electrical and environmental standards. The document mentions tests according to IEC standards for general safety, EMC, nerve/muscle stimulators, environmental conditions (transport, cold, dry heat, damp heat, vibration), and 21 CFR 898 for electrode lead wires and patient cables.
- Data Provenance: The standards listed implicitly suggest the testing was done by the manufacturer (Otto Bock Healthcare Product GmbH, Austria) or a contracted lab to meet international and US regulatory requirements for device safety and performance according to engineering specifications.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- N/A. The "ground truth" in this context is defined by the technical specifications of the predicate devices and the requirements of the international safety and performance standards listed (e.g., IEC 60601 series). Compliance is assessed through engineering tests, not expert clinical evaluation.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A. Adjudication methods like 2+1 or 3+1 are used in clinical trials to establish ground truth or resolve discrepancies in expert interpretation of clinical data. This type of study was not performed as part of this 510(k) submission.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- N/A. The STIWELL med4 is an electrical stimulator and biofeedback device, not an AI-assisted diagnostic or therapeutic tool. Therefore, MRMC studies comparing human readers with and without AI assistance are irrelevant and were not performed.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- N/A. This device is hardware for neuromuscular electrical stimulation and biofeedback. It does not have a "standalone algorithm" in the sense of a software-only diagnostic or therapeutic AI. Its performance is defined by its electrical output characteristics and biomechanical sensing capabilities as verified by engineering and safety testing.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Engineering Specifications and Regulatory Standards. The "ground truth" for this 510(k) submission is that the device's technical specifications and safety features meet or are equivalent to those of legally marketed predicate devices, and that it complies with relevant electrical and medical device safety standards (e.g., IEC 60601 series, 21 CFR 898).
-
The sample size for the training set:
- N/A. There is no "training set" in the context of a machine learning model. The STIWELL med4 is a physical medical device whose design and manufacturing processes are validated, not "trained" on data.
-
How the ground truth for the training set was established:
- N/A. As there is no training set, this question is not applicable.
Ask a specific question about this device
Page 1 of 1