Search Results
Found 2 results
510(k) Data Aggregation
(55 days)
Ningbo Caremed Medical Products Co., Ltd.
Lancet is intended for capillary blood sampling.
The Sterile Lancets for Single Use is composed two components: needle and needle holder. Needle holder is plastic part that enclosed the needle. The sterile barrier is the needle holder and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The lancet is together with a lancing device to puncture the skin to obtain a drop of capillary blood from fingertip or from alternative sites. The shelf-life of the product is 5 years.
The provided document is a 510(k) premarket notification for a medical device (Sterile Lancets for Single Use). It describes the device, its intended use, a comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" with numerical targets in a formal table corresponding to performance. Instead, it refers to "Performance testing is conducted according to Inspection Specification" and states that "the test sample has undergone accelerated aging for 5.5 years" and "passed all tests." The performance attributes tested are listed.
Here's an interpretation of the performance attributes and the implied acceptance criteria from the text:
Acceptance Criterion (Implied) | Reported Device Performance |
---|---|
Appearance: Acceptable visual inspection | The device passed tests after accelerated aging. (Implied: Appearance remained acceptable.) |
Dimension: Within specified tolerances | The device passed tests after accelerated aging. (Implied: Dimensions remained within specified tolerances.) |
Firmness of needle and main body: Adequate mechanical integrity | The device passed tests after accelerated aging. (Implied: Firmness remained adequate.) |
Puncture Performance: Satisfactory for capillary blood sampling | The device passed tests after accelerated aging. (Implied: Puncture performance remained satisfactory.) |
Matching: Compatibility with lancing devices | The device passed tests after accelerated aging. (Implied: Compatibility remained.) |
Launch Performance: Adequate for intended use | The device passed tests after accelerated aging. (Implied: Launch performance remained adequate.) |
Bacterial endotoxin: Below specified limits | The device passed tests after accelerated aging. (Implied: Bacterial endotoxin levels were below limits.) |
Sterilization (SAL=10-6): Achieved | Sterilized by Radiation SAL=10-6. The device passed tests after accelerated aging. (Implied: Sterility was maintained.) |
Biocompatibility: Conforming to ISO 10993 standards | Conforms to the requirements of ISO 10993 series standards, specifically: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity. |
Shelf-life: 5 years | After five years of accelerated aging, the product has passed all tests, so it can be guaranteed that the product can be used normally within five years. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify the numerical sample size used for performance testing. It generally refers to "the test sample."
- Data Provenance: The document indicates that the testing was conducted by Ningbo Caremed Medical Products Co., Ltd., based in China. The data would be retrospective in the sense that the tests were completed before the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable as the device is a medical lancet and the testing involves objective physical, chemical, and biological performance tests, not subjective interpretation by experts to establish a "ground truth" like in imaging or diagnostic AI. The "ground truth" for these tests would be the established scientific and engineering standards and methods.
4. Adjudication Method for the Test Set
This information is not applicable for the same reasons as point 3. Adjudication methods are typically used when subjective expert opinion is involved in ground truth establishment.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not applicable. An MRMC study is relevant for diagnostic devices, especially those involving human interpretation (e.g., radiologists interpreting images). This device is a sterile lancet for single-use capillary blood sampling, a physical instrument, and does not involve human readers interpreting output that would be improved by AI assistance.
6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)
This information is not applicable. This device is a physical medical instrument, not a software algorithm.
7. Type of Ground Truth Used
The "ground truth" for the performance testing of the lancet is based on:
- Established scientific and engineering standards: For dimensions, firmness, puncture performance, matching, launch performance.
- Specific ISO standards: For biocompatibility (ISO 10993 series) and sterilization (SAL of 10-6).
- Bacterial endotoxin limits: Established regulatory or compendial limits.
- Accelerated aging protocols: To simulate real-time shelf-life.
8. Sample Size for the Training Set
This information is not applicable. This device is not an AI/ML algorithm that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable as there is no training set for this device.
Ask a specific question about this device
(54 days)
Ningbo Caremed Medical Products Co. Ltd.
The safety lancet is intended for capillary blood sampling.
Sterile safety lancet consists of outer shell, middle sleeve, needle holder, spring, upper shell, back cover, small cover, needle. The sterile part of the safety lancet is the needle tip. The sterile barrier is the needle holder and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use and single person use only. The shelf-life of the product is 5 years. The device is for medical professionals and lay person use as an OTC use.
This document is a 510(k) Summary for a medical device called "MedtFine Safety Lancet". It details the device's characteristics, intended use, and comparison to a predicate device to demonstrate substantial equivalence.
Here's an analysis of the provided text for acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The document provides a "Characters comparison" table (Table 2) that includes performance parameters for both the proposed "MedtFine Safety Lancet" and the predicate device "Safety Lancet (K220370)". This table effectively serves as a comparison against acceptance criteria. The acceptance criterion for each performance characteristic is generally implied to be "same" or within the range of the predicate device, or "meet requirements."
Acceptance Criteria (Implied from Predicate/Requirement) | Reported Device Performance (MedtFine Safety Lancet) | Comments (from document) |
---|---|---|
Product Code: FMK | FMK | Same |
Indication for Use: Capillary blood sampling | The safety lancet is intended for capillary blood sampling. | Same |
OTC Use | OTC use | Same |
Specifications: Needle: 18G, 18G blade, 21G, 23G, 16G, 28G, 30G, 32G | Needle: 18G, 18G blade, 21G, 23G, 16G, 28G, 30G, 32G | Same, within range. |
Safety protection features: Yes | Yes | Same |
Reuse durability: Single use for single person | Single use for single person | Same |
User environments: Home, clinical | Home, clinical | Same |
Components and materials: (Specific to predicate; proposed device differs) | Outer shell: PP, Middle sleeve: ABS, Needle holder: PE, Spring: carbon steel, Upper Shell: PE, Back cover: PE, Small cover: PP, Needle: Stainless Steel | #1 (The proposed and predicate devices used different materials for manufacturing. The device structures kept the same. The performance testing and biocompatibility testing results demonstrated the proposed device met the requirements of various standards) |
Sterilization: Radiation SAL=10-6 | Radiation SAL=10-6 | Same |
Shelf life: 5 years | 5 years | Same |
Activation force: 4~15N | 4~7N | Same, within range. |
Puncture force: 18 |
8. The sample size for the training set
This information is not applicable. The MedtFine Safety Lancet is a mechanical medical device, not an AI/ML-based device that requires a "training set." Its design and manufacturing are based on engineering principles and material science, not data-driven learning.
9. How the ground truth for the training set was established
This information is not applicable for the reasons stated above (not an AI/ML device). The design and development process for such a device relies on established engineering specifications, material selection, and manufacturing processes, which are then validated through the performance and biocompatibility testing outlined.
Ask a specific question about this device
Page 1 of 1