K Number
K153577Device Name
FREND Testosterone Test System
Manufacturer
NanoEnTek USA Inc.
Date Cleared
2016-10-14
(304 days)
Product Code
CDZ
Regulation Number
862.1680Why did this record match?
Applicant Name (Manufacturer) :
NanoEnTek USA Inc.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FREND™ Testosterone test is a fluorescent nanoparticle immunoassay designed for in vitro quantitative measurement of total testosterone in human serum and plasma (K3-EDTA and lithium heparin). Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, impotence in males, hirsutisim (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
The FREND™ Testosterone microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader. The FREND™ Testosterone Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.
Device Description
The FREND™ Testosterone is a rapid fluorescence immunoassay that measures testosterone levels in human serum and plasma using the FREND™ system.
The FREND™ Testosterone test is a single-use rapid "competitive" immunoassay utilizing fluorescent nanoparticle in microfluidic flow to capture and quantify total testosterone levels in human serum and plasma (KgEDTA and lithium-heparin) specimens using the FREND™ system. The FREND™ Testosterone Test is a two-step competitive immunoassay with gold micro-particles labeled with mouse monoclonal antitestosterone antibody, testosterone-biotin labeled with fluorescence nanoparticles and fluorescence detection by the FREND™ System.
The FREND™ Testosterone test utilizes microfluidic technology and detects immunecomplexes bound to testosterone. In Step 1, a 70 µL patient sample is first incubated for 5 minutes at 98.6 °F (37 °C) in the Testosterone Gold Antibody pretreatment tube, where the sample interacts with a proprietary mix of a pretreatment solution. In Step 2, the Test Cartridge is placed on the warming platform of the heating block and 35 µL of the mixture from Step 1 is manually loaded into the inlet of the cartridge. The cartridge remains on the warming platform for 30 seconds, while the sample hydrates the testosterone-biotin fluorescent bead conjugate and migrates along the test strip. During migration the bound testosterone in the sample and the testosterone-biotin fluorescent bead conjugates compete to form antigen-antibody complexes in the test zone. Unbound testosterone-biotin fluorescent conjugates flow through and bind to the antitestosterone antibody that is immobilized on the surface in the reference zone. The cartridge is inserted into the FREND instrument for analysis where fluorescent signals in the test and reference zones are measured, typically within 4 minutes.
Testosterone quantification is based upon the ratio of the intensity of the test and reference zones. The magnitude of the fluorescent ratio is inversely proportional to the amount of testosterone in the sample.
The measuring range of the FREND™ Testosterone Test System is 20 to 1500 ng/dL. Results are determined via a lot-specific calibration curve which is generated by the manufacturer using a six-point calibration determined from values averaged from 5 replicates at each level. The established curve is uploaded to the FREND™ System via the Testosterone Code-chip and is valid until the lot expiration date.
The FREND™ Testosterone test cartridge is a disposable plastic device that houses the reagents and contains a port or opening (inlet) where the sample is applied. Once the sample is applied, it will mix with the reagents and travel towards the detection area via capillary action.
The FREND™ System is a portable, automated FREND™ cartridge reader. The FREND™ System is based on quantitative immunoassay technology capable of quantifying single or multiple analytes by measuring laser-induced fluorescence in a single-use disposable reagent cartridge. The FREND™ cartridge utilizes micro-fluidics lateral flow technology where the analyte of interest in the sample forms immune complexes while moving through the fluidics pathway in the cartridge. The concentration of the analyte of interest in an unknown sample is calculated using the ratio of the fluorescent intensity of the test zone and the reference zone.
The FREND™ System is a bench top fluorescence reader containing a touch screen user interface. The System has a slot that accepts the FREND™ Testosterone Test Cartridge (which contains the reagents and sample), and is programmed to analyze the Test when the sample has fully reacted with the on-board in-cartridge reagents. Results of the test are displayed on the screen and can be printed on an optional printer.
The FREND™ System software controls the graphical user interface, communication with hardware, database manaqement and data analysis. The software also controls the functions of the mechanical components including the motor, laser, printer control and acquisition of data from the sensor. The user can set the time and date and enter patient ID through the graphic user interface. The user cannot make any changes to the software.
The FREND™ Testosterone Test System includes the following in the kit:
- . 20 FREND™ Testosterone cartridges
- 20 Testosterone Gold Antibody Pretreatment Tubes .
- Disposable pipette tips .
- . 1 FREND™ Testosterone Code Chip
- 1 FREND™ Testosterone Package Insert ●
The FREND™ System (previously cleared in K124056 (FREND™ PSA Plus) K131928 (FREND™ TSH), and K152422 (FREND™ Free T4)) is not provided with the kit but is required for the use of the FREND™ Testosterone test cartridge.