Search Results
Found 2 results
510(k) Data Aggregation
(58 days)
Medifun Corporation
Safety lancet is intended for capillary blood sampling.
The Medifun Safety Lancet (MSL1 series) is designed to obtain capillary blood samples from the fingertip. The intended users include healthcare personnel, patients, and lay persons. According to the needle and color differences, the safety lancets have 6 models : MSL1-30, MSL1-28, MSL1-26, MSL1-23, MSL1-21, and MSL1-18. The safety lancets consist of the needle seat, main body, release platform, spring, and protective cap. The sterile part of the safety lancet is the needle tip. The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf-life of the product is 5 years. The lancets contain a safety feature. After use, the puncture mechanism retracts, rendering it unusable for a second time.
The provided text is a 510(k) summary for the Medifun Safety Lancet (MSL1 series). It details the device's characteristics, its comparison to a predicate device, and the performance data that supports its substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Item | Acceptance Criteria | Reported Device Performance |
---|---|---|
Appearance | Consistent with the drawings. There should be no damage, dirt, or holes and the protective cap must not be pulled open. | Meet the requirement |
Dimension | Consistent with the drawings. | Meet the requirement |
Firmness | The needle must remain firmly attached to the plastic part of the needle seat. | Meet the requirement |
Needle protrusion length | Use calipers to measure and meet the requirements. | Meet the requirement |
Launch performance | It must be able to fire smoothly, and there should be no abnormal rebound noise after firing. | Meet the requirement |
Puncture force | The needle tip of the needle should have good puncture ability. | Meet the requirement |
Disposable | The safety lancet is only allowed to be puncture once, and it becomes invalid and cannot be fired again after it has been punctured. | Meet the requirement |
Safety Feature | The downward firing force |
Ask a specific question about this device
(155 days)
Medifun Corporation
Medifun Lancing Device, Model No. LD-E1 is a reusable lancing device for the single user. Itshould be used with the proper sterile lancet for the capillary blood sampling. It could automatically inject the lancet into and retract it from the fingertip to obtain a capillary blood sample for glucose monitoring or other test that require one or two drops of blood. A depth adjustable cap allows the best depth of skin penetration for each individual user. Alternative site testing (palm or forearm) can be performed by installing the clear cap on the lancing device.
Medifun Lancing Device, Model No. LD-E1 is a mechanical blood lancet holder. It is used with a single use sterile lancet to puncture the skin for collecting capillary whole blood sampled from the fingertip or alternate sites. Medifun Lancing Device LD-E1 can be adjusted for 6 levels of depth. This device is for single user only.
This 510(k) summary is for a lancing device, which is a mechanical device used to obtain blood samples. The document focuses on showing substantial equivalence to existing predicate devices, primarily through non-clinical bench testing. It does not involve AI or human-in-the-loop performance studies. Therefore, many of the requested categories related to such studies are not applicable.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion | Reported Device Performance | Study that Proves Device Meets Criteria |
---|---|---|
Biocompatibility | Meets ISO 10993-1, ISO 10993-5 (In vitro Cytotoxicity), ISO 10993-10 (Skin Sensitization), ISO 10993-23 (Skin Irritation) standards. | Non-clinical bench testing specifically for ISO 10993-5, ISO 10993-23, and ISO 10993-10. |
Cleaning Validation | Successful validation of cleaning procedures. | Non-clinical bench testing for cleaning validation. |
Risk Management | Adheres to ISO 14971:2019 standard. | Risk management activities performed according to ISO 14971:2019. |
Mechanical Design & Functionality | Meets specified requirements (e.g., 6 levels of depth adjustment from 0.5mm-1.7mm, 'lancet retract after use'). | Non-clinical bench testing including "Specification and functional tests" and "mechanical design verification and validation testing." |
Substantial Equivalence (Overall) | The device is deemed substantially equivalent to predicate devices in intended use, technology, operation principle, and similar technical characteristics, with all verification tests meeting required acceptance criteria. | Comparative analysis with predicate devices (RIGHTEST Lancing Device GD500, GE Lancing Device, iGlucose Lancing Device K221062) and comprehensive non-clinical bench testing. The document states that "all tests were verified to meet the required acceptance criteria." |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of clinical data or patient samples. The testing described is primarily non-clinical bench testing. Therefore, details about sample size (e.g., number of devices tested for mechanical properties) and data provenance (country of origin, retrospective/prospective) for these bench tests are not provided in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as the studies were non-clinical bench tests; there was no "test set" requiring expert ground truth in the clinical sense.
4. Adjudication method for the test set
This is not applicable for the same reasons as point 3.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a mechanical lancing device, not an AI-assisted diagnostic or therapeutic tool that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a mechanical lancing device, not an algorithm.
7. The type of ground truth used
For the non-clinical tests, the "ground truth" would be established by the criteria defined in the respective ISO standards (e.g., ISO 10993 for biocompatibility) and the device's own internal specifications for functional performance. This is not "expert consensus," "pathology," or "outcomes data" in the clinical sense.
8. The sample size for the training set
This is not applicable. There is no concept of a "training set" for a mechanical lancing device as there are no algorithms being trained.
9. How the ground truth for the training set was established
This is not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 1