Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K230529
    Date Cleared
    2023-11-22

    (268 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Maxigen Biotech Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HealiAid® Dental Collagen Wound Dressing is indicated for the management of oral wounds and sores, including:

    1. Denture sores
    2. Oral ulcers (non-infected or viral)
    3. Periodontal surgical wounds
    4. Extraction sites
    5. Surgical wounds.
    6. Traumatic wounds
    Device Description

    The subject device is HealiAid® Dental Collagen Wound Dressing (HealiAid Dental) that is a white, porous, pliable and absorbable collagen wound dressing. HealiAid Dental is fabricated by fibrous collagen matrix purified from bovine Achilles tendons. HealiAid Dental is pliable and can be applied easily to oral wounds to protect the wound bed. The product is supplied in a sterile, nonpyrogenic package, and is indicated for single use only.

    AI/ML Overview

    The provided text describes the regulatory clearance (510(k)) for the HealiAid® Dental Collagen Wound Dressing. This is a medical device, not an AI/ML software device, and therefore the concepts of acceptance criteria for AI performance metrics (like sensitivity, specificity, AUC), sample size for test sets, ground truth establishment by experts, adjudication methods, MRMC studies, or training sets are not applicable here.

    Instead, the acceptance criteria for this medical device are related to its physical and biological properties, and its performance in pre-clinical studies to demonstrate substantial equivalence to a legally marketed predicate device.

    Here's a breakdown of the information provided, framed to address the prompt's categories where applicable for a medical device without AI/ML components:


    1. Table of Acceptance Criteria and Reported Device Performance

    For medical devices like wound dressings, "acceptance criteria" are typically defined by demonstrating compliance with established standards (like ISO 10993 for biocompatibility) and exhibiting comparable performance to predicate devices in pre-clinical tests.

    Acceptance Criteria CategorySpecific Test / PropertyReported Device Performance (HealiAid® Dental)
    BiocompatibilityCytotoxicity (ISO 10993-5)Non-cytotoxic
    Genotoxicity (Ames test, Mammalian Cell Gene Mutation, Micronucleus Test - ISO 10993-3)Non-genotoxic
    Systemic Toxicity (Acute & Subchronic - ISO 10993-11)No acute/subchronic systemic toxicity
    Irritation (Intracutaneous reactivity - ISO 10993-23)Non-irritant
    Sensitization (Guinea Pig Maximization - ISO 10993-10)Non-sensitizing
    Implantation (in porcine gingiva - ISO 10993-6)Very slight or non-tissue reaction
    Material-mediated pyrogenicity (USP 151)Non-pyrogenic
    Physical/Chemical PropertiesCollagen contentDemonstrated substantial equivalence
    Carbohydrate analysisDemonstrated substantial equivalence
    pH value analysisDemonstrated substantial equivalence
    Thermal stability analysisDemonstrated substantial equivalence
    Microstructure analysisDemonstrated substantial equivalence
    Moisture content analysisDemonstrated substantial equivalence
    Water absorption measurementDemonstrated substantial equivalence
    DegradationIn vitro/in vivo degradation testSimilar timecourse of degradation to predicate (within 30 days in oral cavity)
    Sterility & Shelf LifeSterilization (Gamma irradiation, SAL 10-6)Confirmed by testing
    Viral Inactivation (ISO 22442 series compliance)Compliant with ISO 22442
    Shelf life (3-year stability)Confirmed by real-time aging

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Pre-clinical Studies): The document does not specify exact sample sizes for each pre-clinical test (e.g., number of animals for toxicity or implantation studies). These studies are typically conducted according to standardized protocols outlined in documents like ISO 10993, which define appropriate sample sizes for each test type.
    • Data Provenance: The studies were pre-clinical (laboratory and animal studies) rather than human clinical trials. The bovine Achilles tendons are the source material.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • This is not applicable as the "ground truth" for this type of device is established through standardized laboratory tests and animal models, adhering to recognized international standards (e.g., ISO 10993, USP 151, ISO 22442). These tests have defined endpoints and criteria, and their interpretation is generally performed by qualified laboratory personnel following strict protocols, rather than a consensus of human "experts" evaluating an outcome like an image.

    4. Adjudication Method for the Test Set

    • Not applicable. The pre-clinical tests have objective measurements and pass/fail criteria based on established standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • Not applicable. MRMC studies are specific to diagnostic devices (often imaging-based AI) where multiple human readers interpret cases, and their performance is compared with and without AI assistance. The HealiAid® Dental Collagen Wound Dressing is a therapeutic/wound management device, not a diagnostic one.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical wound dressing and does not involve an algorithm or AI functionality.

    7. The Type of Ground Truth Used

    • For the pre-clinical studies, the "ground truth" is established by the objective results of standardized biological and chemical tests (e.g., cell viability in cytotoxicity, presence/absence of mutagenic activity in genotoxicity, tissue reactions in implantation, absorption values, degradation rates). These are verifiable scientific outcomes, not interpretations of medical images or pathology.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not use AI/ML and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223126
    Date Cleared
    2023-10-11

    (373 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Maxigen Biotech, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SurgiAid® Collagen Wound Dressing is indicated for use in patients who have surgical wounds, donor sites/grafts, podiatric wound, wound dehiscence, traumatic wounds, abrasions, lacerations, partial thickness burns or skin tears. SurgiAid® Collagen Wound Dressing can be applied to wounds with depth less than 0.3 cm.

    Device Description

    SurgiAid Collagen Wound Dressing (SurgiAid) is white, porous pliable and absorbable collagen wound dressing. It is fabricated by fibrous collagen matrix which is purified from bovine Achilles tendon. SurgiAid is pliable and can be applied easily to clean wound. The product is supplied in sterile, non-pyrogenic package, and is indicated for single use only.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device called "SurgiAid® Collagen Wound Dressing" (K223126). This submission focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than a de novo approval requiring extensive clinical efficacy studies. Therefore, the "study" described and the "acceptance criteria" are primarily related to non-clinical performance data and comparison to the predicate device.

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparison table (Table 1) between the subject device (K223126), the predicate device (K100927), and a reference device (K122325). The acceptance criteria are implied by the "Comparison" column, which states "Identical to Predicate Device" or similar, indicating that the subject device's performance should match that of the predicate. For specifications (dimensions), the acceptance is that the new sizes "fall within the size range of the reference device."

    Item / Acceptance CriteriaReported Device Performance (Subject Device K223126)
    Manufacturer (Identical to Predicate Device)MAXIGEN BIOTECH INC. (Identical to Predicate Device)
    Product Code (Identical)KGN (Identical)
    Classification name (Identical)Collagen Wound Dressing (Identical)
    Indications for use (Identical to Predicate Device)SurgiAid® Collagen Wound Dressing is indicated for use in patients who have surgical wounds, donor sites/grafts, podiatric wound, wound dehiscence, traumatic wounds, abrasions, lacerations, partial thickness burns or skin tears. SurgiAid® Collagen Wound Dressing can be applied to wounds with depth less than 0.3 cm. (Identical to Predicate Device)
    Materials (Identical to Predicate Device)Bovine (Achilles) tendon type I collagen (Identical to Predicate Device)
    Form (Identical)Sheet (Identical)
    Specifications (Dimensions) (New dimensions fall within the size range of the reference device)15mm x 20mm x 0.5mm20mm x 30mm x 0.5mm25mm x 30mm x 0.5mm30mm x 40mm x 0.5mm100mm x 100mm x 0.5mm20mm x 40mm x 3.0mm30mm x 30mm x 3.0mm40mm x 50mm x 3.0mm50mm x 50mm x 3.0mm100mm x 100mm x 3.0mm (These new dimensions are stated to fall within the size range of the reference device - SkinTemp II Dressing, which has sizes 2"x2", 3"x4", 8"x12")
    Usage type (Identical)Single use only (Identical)
    Sterilization (Identical to Predicate Device)By gamma irradiation (Identical to Predicate Device)
    Design (Identical to Predicate Device)White, porous, pliable, absorbable, sterile, non-pyrogenic package collagen wound dressing and it is indicated for single use only. (Identical to Predicate Device)
    Biocompatibility (Identical)All biocompatibility evaluations of SurgiAid were conforms to the requirements specified in ISO10993 (Identical to Predicate Device, all results acceptable based on non-clinical data section).
    Shelf life (Identical to Predicate Device)Three years (Identical to Predicate Device)
    Sterile Barrier Packaging (Identical to Predicate Device)Double blister tray as Sterile Barrier System (Identical to Predicate Device)
    Storage condition (Identical to Predicate Device)Store in a dry place below 25°C (Identical to Predicate Device)
    Performance (Water absorption, Porosity, Density) (Identical to Predicate Device)Water absorption (>25 fold); Porosity (>80%); Density (About 37 mg/cm³) (Identical to Predicate Device, confirmed in non-clinical data section that "All tests were met acceptance criteria in according with related standards.")

    The Study that Proves the Device Meets the Acceptance Criteria:

    The "study" here is a series of non-clinical tests and a comparison to the predicate device to demonstrate substantial equivalence, rather than a clinical trial.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses non-clinical performance and material characteristics.

    • Sample size: The specific sample sizes for each non-clinical test (e.g., how many dressings were tested for water absorption, porosity, sterilization validation, etc.) are not explicitly provided in the summary.
    • Data provenance: The document states the manufacturer is Maxigen Biotech Inc. in Taoyuan City, Taiwan. The non-clinical tests would have been conducted by or for this manufacturer. It does not specify whether these tests were performed in Taiwan or elsewhere. The data is "retrospective" in the sense that the new device's characteristics were compared to already established characteristics of the predicate and reference devices, and tested against established standards (ISO10993, etc.). These are typically laboratory tests, not patient data trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to this type of submission. The "ground truth" for non-clinical performance data (e.g., water absorption, porosity, biocompatibility) is established according to recognized international standards (e.g., ISO 10993 for biocompatibility) and validated laboratory methods. It does not involve expert consensus on medical images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods are typically used in clinical trials involving subjective interpretation of data (like imaging studies where multiple readers agree/disagree on a finding). For non-clinical tests, results are typically quantitative and objective measurements against predefined criteria/standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is relevant for diagnostic devices, especially those incorporating AI, where human readers interpret medical images. This document describes a collagen wound dressing, which is a therapeutic device, and does not mention any AI component or diagnostic function.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical wound dressing, not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Standards/Specifications: The "ground truth" for characteristics like water absorption, porosity, density, sterilization, packaging integrity, and biocompatibility is based on established international standards (e.g., ISO10993-1:2018) and predefined acceptance criteria derived from the predicate device's performance and general recognized medical device safety and efficacy principles.
    • Predicate Device Data: The performance of the predicate device (K100927) serves as a key "ground truth" for comparison to establish substantial equivalence.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of this 510(k) submission for a wound dressing. Training sets are used in machine learning for AI algorithms.

    9. How the ground truth for the training set was established

    This question is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100927
    Device Name
    SURGIAID
    Date Cleared
    2011-02-02

    (303 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    MAXIGEN BIOTECH, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SurgiAid Collagen Wound Dressing is intended for use in patients who have surgical wounds, donor sites/grafts podiatric wounds, wound dehiscence, traumatic wounds, abrasions, lacerations, partial thickness burns or skin tears. SurgiAid Collagen Wound Dressing can be applied to wounds with depth less than 0.3 cm.

    Device Description

    SugiAid Collagen Wound Dressing is white, porous, pliable and absorbable collagen wound dressings. It is fabricated by fibrous collagen matrix which is purified from bovine Achilles tendon. SugiAid is pliable and can be applied easily to clean wounds. The product is supplied in sterile, nonpyrogenic package, and is indicated for single use only.

    AI/ML Overview

    The provided text describes a 510(k) summary for the "SurgiAid Collagen Wound Dressing." This document focuses on demonstrating substantial equivalence to a predicate device rather than conducting a de novo clinical study to define and meet specific acceptance criteria for performance.

    Therefore, the study design elements typically associated with establishing acceptance criteria for a new AI/medical device (like sample size for test sets, ground truth methodology, MRMC studies, or standalone performance) are not applicable here. The submission's goal is to prove the new device is as safe and effective as a previously cleared device, not to prove its own independent performance against a set of statistical targets.

    Here's a breakdown based on the provided text, highlighting what is (and isn't) present:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Demonstration of Substantial Equivalence to Predicate)Reported Device Performance (as demonstrated by testing and comparison)
    Biocompatibility: Conformance to ISO 10993-1:2009 series requirements.Confirmed: Biocompatibility evaluations (cytotoxicity, hemolysis, sensitization, intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation, pyrogenicity) confirm conformance.
    Similar Physical Properties:Confirmed: Pre-clinical benchtop type evaluations including physical tests (e.g., Purity (Collagen content), Endotoxin level (Pyrogen test), Pepsin resistance, Collagenase resistance & Rehydration Test) show substantial equivalence.
    Similar Chemical Properties:Confirmed: Pre-clinical benchtop type evaluations including chemical tests show substantial equivalence.
    Same Intended Use:Confirmed: "SugiAid Collagen Wound Dressing has the same intended use."
    Similar Operating Principle:Confirmed: "using similar operating principle."
    Similar Technological Characteristics:Confirmed: "using similar technological characteristics." Bench testing demonstrates that differences do not raise new safety/effectiveness questions.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This submission relies on bench testing and biocompatibility assessments rather than a clinical trial with a "test set" of patient data in the typical sense for AI/diagnostic devices. The goal is to show the device itself meets material and safety standards and is similar to a predicate, not to test its performance on a dataset of clinical cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No human experts were used for "ground truth" establishment in this context, as it's a submission for a wound dressing, not an AI diagnostic device. The "ground truth" for material properties is established through standardized laboratory testing and specific ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As there is no "test set" in the context of clinical image or data interpretation, no adjudication method was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical medical device (collagen wound dressing), not an AI-assisted diagnostic tool. Therefore, MRMC studies are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" here is based on established scientific and regulatory standards for material properties, biocompatibility, and intended use. This includes:
      • ISO 10993-1:2009 series for biocompatibility (cytotoxicity, hemolysis, sensitization, etc.).
      • Validated laboratory methods for physical and chemical tests (Purity, Endotoxin level, Pepsin resistance, Collagenase resistance, Rehydration Test).
      • Comparison to the known characteristics of the predicate device (SkinTemp, K925545).

    8. The sample size for the training set

    • Not Applicable. This device is not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1