Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    Mah Sing Healthcare Sdn. Bhd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (2023), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

    Tested Chemotherapy DrugsConcentrationMinimum Breakthrough Detection Time in minutes
    *Carmustine3.3mg/ml22.5
    Cisplatin1 mg/ml>240
    Cyclophosphamide20 mg/ml>240
    Dacarbazine10 mg/ml>240
    Doxorubicin, HCl2 mg/ml>240
    Etoposide20 mg/ml>240
    Fluorouracil50 mg/ml>240
    Methotrexate25 mg/ml>240
    Mitomycin0.5 mg/ml>240
    Oxaliplatin5 mg/ml>240
    Paclitaxel6 mg/ml>240
    *Thiotepa10 mg/ml34.5
    Vincristine1 mg/ml>240
    Tested Opioid DrugsConcentrationMinimum Breakthrough Detection Time in minutes
    Fentanyl Citrate Injection100mcg/2ml>240

    Warning: Do not use with Carmustine and Thiotepa.

    Note:
    Please note that the following drugs have low permeation times:
    (1) Carmustine - 22.5 minutes
    (2) Thiotepa - 34.5 minutes.

    Device Description

    The Nitrile Powder Free Examination Glove with Tremella Fuciformis Extract, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate - Blue are Class 1, Polymer Patient Examination Gloves and Chemotherapy Gloves.

    The glove is made of Nitrile Butadiene Rubber. They are ambidextrous with beaded cuffs, fingertip texture and come in different sizes - Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978-05(2023). The gloves are single use and provided non-sterile.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Nitrile Powder Free Examination Gloves, not a software-based medical device or AI system. Therefore, the questions regarding ground truth, expert adjudication, MRMC studies, standalone performance, training sets, and data provenance are not applicable to this document.

    However, based on the non-clinical performance testing detailed in the submission, we can address the acceptance criteria and demonstrated performance for the physical device.

    Acceptance Criteria and Reported Device Performance for Nitrile Powder Free Examination Gloves K243694

    The acceptance criteria for this medical device are based on various physical, chemical, and biological tests, primarily adhering to ASTM and ISO standards for examination gloves and their specific claims (chemotherapy and fentanyl resistance).

    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodAcceptance CriteriaReported Device Performance
    Physical Dimensions (ASTM D6319)Length: S: ≥220mm, M/L/XL: ≥230mm
    Width: S: 80±10 mm, M: 95±10 mm, L: 110±10 mm, XL: 120±10 mm
    Thickness: Palm: ≥0.05 mm, Finger: ≥0.05 mmLength: S: ≥220mm (Pass), M/L/XL: ≥230mm (Pass)
    Width: S: 86-88/Pass, M: 98-99/Pass, L: 105-110/Pass, XL: 117-119/Pass
    Thickness: Palm: S-XL: 0.06mm/Pass, Finger: S-L: 0.08mm/Pass, XL: 0.09mm/Pass
    Watertightness Test for Detection of Holes (ASTM D5151, ASTM D6319)Meet AQL 2.5 with G1Meet AQL 2.5 with G1 (Pass)
    Powder Content (ASTM D6124, ASTM D6319)Meet the requirement of ASTM D 6124-06 (≤2.0mg/glove)S: 0.63mg/glove, M: 1.02mg/glove, L: 0.28mg/glove, XL: 0.38mg/glove (Pass)
    Physical Properties (Tensile Strength & Elongation) (ASTM D412, ASTM D6319)Tensile Strength: before aging ≥14 MPa, after aging ≥14 MPa
    Ultimate Elongation: before aging ≥500 %, after aging ≥400%Tensile Strength: Before aging: Pass, After aging: Pass
    Ultimate Elongation: Before aging: Pass, After aging: Pass
    Chemotherapy drug and Fentanyl Citrate claim (ASTM D6978)The subject device was tested for use with chemotherapy drugs and Fentanyl Citrate. Implicit acceptance is that breakthrough times meet or exceed safe working periods, and any low permeation drugs are adequately warned.No breakthrough detected up to 240 minutes for: Cisplatin, Cyclophosphamide, Dacarbazine, Doxorubicin HCl, Etoposide, Fluorouracil, Methotrexate, Mitomycin, Oxaliplatin, Paclitaxel, Vincristine, Fentanyl Citrate Injection.
    Low permeation times: Carmustine (22.5 minutes), ThioTEPA (34.5 minutes) - Warning issued for these two.
    Cytotoxicity (ISO 10993-5)Non-cytotoxicUnder conditions of the study, device extract is cytotoxic (This seems to be a clerical error in the document as it proceeds to state 'Non-cytotoxic' in subsequent summary sections of the same document for the same test, and a device would not be cleared if cytotoxic. Given the context of clearance, it likely passed or the initial statement is a misphrasing where the high concentrations showed cytotoxicity but the diluted, in-use concentrations did not, aligning with the "Under the condition of this test, exhibited cytotoxicity reactivity from 100.0% extract concentration to 25.0% extract concentrations and exhibited no cytotoxicity reactivity from 12.5% extract concentration to 3.125% extract concentrations" within the Summary of Technological Characteristic section. The overall conclusion for biocompatibility in Table 1 Part 4 of 8 "Similar" for this test, and the clearance, suggests it was deemed acceptable).
    Irritation (ISO 10993-23)Non-irritatingUnder the conditions of the study, device is not an irritant. / Pass
    Sensitization (ISO 10993-10)Non-sensitizingUnder the conditions of the study, device is not a sensitizer. / Pass
    Acute Systemic Toxicity (ISO 10993-11)Non-acute systemic toxicityUnder the conditions of the study, device did not show acute systemic toxicity in vivo. / Pass
    Bioburden Determination (ISO 11737-1:2018)≤ 200 CFU/device80-119 CFU/glove (Pass)
    Material-Mediated Pyrogenicity (USP )Non-pyrogenicUnder the conditions of the study, non-pyrogenic (Pass)

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific "sample size" for each individual test reported (e.g., how many gloves were tested for watertightness, how many for a specific drug permeation). However, these tests are typically performed on a statistically representative sample of manufactured gloves as per the respective ASTM/ISO standards.

    • Data Provenance: The testing was conducted as part of the 510(k) submission process for a device manufactured by Mah Sing Healthcare Sdn. Bhd. in Malaysia. The data would therefore be prospective for the purpose of demonstrating the device's adherence to the standards for this specific submission. The data is generated from laboratory testing of the physical product.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This relates to physical/chemical/biological testing of a medical glove, not the establishment of ground truth for an AI or imaging diagnostic system. The "ground truth" here is the objective measurement against established physical/chemical/biological standards.

    4. Adjudication method for the test set:

    Not applicable. Adjudication methods (like 2+1, 3+1) are relevant to human review of diagnostic medical data, often for establishing a consensus "ground truth" when a definitive objective truth is unavailable (e.g., interpreting medical images). For physical device testing, the results are typically objectively measured and do not require expert adjudication in this manner.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No. MRMC studies are used to evaluate the impact of a new diagnostic aid (like AI) on physician performance. This is for a physical medical device (examination gloves), not an AI system or diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This question pertains to AI algorithm performance.

    7. The type of ground truth used:

    Objective Experimental Measurement and Compliance with Standardized Protocols. For each test (e.g., tensile strength, chemical permeation, cytotoxicity), the "ground truth" is the quantitative measurement obtained through standardized laboratory procedures (e.g., ASTM D6319 for physical properties, ASTM D6978 for chemical permeation, ISO 10993 series for biocompatibility) and compared against the defined acceptance criteria within those standards.

    8. The sample size for the training set:

    Not applicable. This question pertains to AI/machine learning models. Examination gloves do not have a "training set."

    9. How the ground truth for the training set was established:

    Not applicable. Again, this relates to AI/machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230002
    Date Cleared
    2023-03-24

    (80 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Mah Sing Healthcare Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Nitrile Powder Free Black Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250. The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are black in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in five sizes (XS, S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Nitrile Powder Free Black Patient Examination Gloves, Non-sterile, Tested For Use With Chemotherapy Drugs, and does not pertain to an AI/ML-enabled medical device. Therefore, the information requested about acceptance criteria and studies proving the device meets them in the context of an AI/ML device (e.g., sample size for test sets, data provenance, a number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, training set details) is not applicable to this document.

    Instead, the document describes the acceptance criteria and performance of these medical gloves based on non-clinical bench testing to demonstrate substantial equivalence to a predicate device.

    Here's the relevant information from the document regarding the glove's performance:

    1. A table of acceptance criteria and the reported device performance (for the non-AI device: medical gloves):

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance (Result)
    ASTM D6124-06 (Residual Powder)To determine residual powder in glovesLess than 2mg / gloveXS: 0.23mg/glove, S: 0.20mg/glove, M: 0.25mg/glove, L: 0.25mg/glove, XL: 0.30mg/glove (All below 2mg/glove)
    ASTM D5151-19 (Detection of Holes)To determine holes in glovesG-I, AQL 2.5 (In accordance with ASTM D6319-19)Passed G-I, AQL 1.5
    ASTM D6319-19 (Length)To determine the length of glovesXS: ≥220mm, S: ≥220mm, M: ≥230mm, L: ≥230mm, XL: ≥230mmXS: 244-247mm, S: 243-248mm, M: 243-252mm, L: 243-249mm, XL: 243-249mm (All meet minimums)
    ASTM D6319-19 (Width)To determine the width of glovesXS: 70±10mm, S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: 120±10mmXS: 76-78mm, S: 81-88mm, M: 95-99mm, L: 102-111mm, XL: 111-119mm (All within specified ranges)
    ASTM D6319-19 (Palm Thickness)To determine the thickness of gloves (palm)Palm: ≥0.05mmXS, S, M, L, XL: 0.06-0.07mm (All meet minimum)
    ASTM D6319-19 (Finger Thickness)To determine the thickness of gloves (finger)Finger: ≥0.05mmXS, S, M, L, XL: 0.07-0.08mm (All meet minimum)
    ASTM D6319-19 (Tensile Strength - Before Aging)To determine physical properties - Tensile strength (before aging)≥14MPa for all sizesXS: 23.79 MPa, S: 24.89 MPa, M: 25.39 MPa, L: 26.44 MPa, XL: 26.05 MPa (All meet minimum)
    ASTM D6319-19 (Tensile Strength - After Aging)To determine physical properties - Tensile strength (after aging)≥14MPa for all sizesXS: 27.52 MPa, S: 26.28 MPa, M: 25.93 MPa, L: 26.21 MPa, XL: 25.98 MPa (All meet minimum)
    ASTM D6319-19 (Ultimate Elongation - Before Aging)To determine physical properties - Ultimate Elongation (before aging)≥500% for all sizesXS: 539%, S: 533%, M: 527%, L: 538%, XL: 526% (All meet minimum)
    ASTM D6319-19 (Ultimate Elongation - After Aging)To determine physical properties - Ultimate Elongation (after aging)≥400% for all sizesXS: 435%, S: 453%, M: 460%, L: 455%, XL: 470% (All meet minimum)
    ASTM D6897-05 (Chemotherapy Drug Permeation)To assess resistance to permeation by chemotherapy drugs>240 minutes for most drugs specified (Carmustine & Thiotepa are exceptions with lower times)Carmustine: 32.2 min, Cisplatin: >240 min, Cyclophosphamide: >240 min, Dacarbazine: >240 min, Doxorubicin, HCl: >240 min, Etoposide: >240 min, Fluorouracil: >240 min, Methotrexate: >240 min, Mitomycin: >240 min, Oxaliplatin: >240 min, Paclitaxel: >240 min, Thiotepa: 37.4 min, Vincristine: >240 min. (Meets expectations based on warnings for Carmustine and Thiotepa).
    ISO 10993-10 (Animal Irritation Test)To determine potential for dermal irritationNot an irritantPassed, Primary Irritation Index (PII) was "0", "negligible" irritation.
    ISO 10993-10 (Dermal Sensitization Assay)To determine skin sensitization potentialNot a sensitizerPassed, no sensitization induced.
    ISO 10993-11 (Acute Systemic Toxicity)To provide information on health hazards from short-term exposureNot induce acute systemic toxicityPassed, did not demonstrate any adverse toxic reaction.

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size for Testing: The document does not specify the exact sample sizes for each non-clinical test (e.g., how many gloves were tested for holes, tensile strength, or chemotherapy permeation for each batch/size). The results are presented as averages or ranges across sizes.
    • Data Provenance: The tests were non-clinical bench tests performed by Mah Sing Healthcare Sdn. Bhd. (Malaysia), as indicated by the sender and manufacturer information. There is no mention of country of origin of data in the sense of patient data, as this device does not use patient data for its function. All tests are by standard methods (ASTM, ISO). The studies are inherently "prospective" in the sense that they are conducted specifically for this submission against defined standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. Ground truth for these physical and chemical tests is established by the specified ASTM and ISO standard test methods, which are objective, reproducible procedures. No human experts are involved in establishing "ground truth" for the physical properties of a glove in the way they would for medical image interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. As these are objective bench tests, there is no need for human adjudication of results. The results are quantitative measurements or direct observations according to specified protocols.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device. The "standalone" performance here refers to the glove's physical and chemical properties and its resistance to chemotherapy drugs, which are measured directly.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this device, the "ground truth" is defined by the measurement outcomes according to internationally recognized standard test methods (ASTM and ISO) for physical properties (e.g., dimensions, tensile strength, freedom from holes, powder residue) and chemical resistance (chemotherapy drug permeation). For biocompatibility, it's defined by the in vivo animal study results according to ISO standards.

    8. The sample size for the training set:

    • Not applicable. There is no training set for a non-AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no training set for a non-AI/ML device.

    In summary, this document demonstrates the safety and effectiveness of medical gloves through adherence to established material and performance standards, not through AI/ML model validation studies.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Mah Sing Healthcare Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs: Test Chemotherapy Drugs are as follows: Carmustine 3.3mg/ml, Cisplatin 1 mg/ml, Cyclophosphamide 20 mg/ml, Dacarbazine 10 mg/ml, Doxorubicin, HCl 2 mg/ml, Etoposide 20 mg/ml, Fluorouracil 50 mg/ml, Methotrexate 25 mg/ml, Mitomycin 0.5 mg/ml, Oxaliplatin 5 mg/ml, Paclitaxel 6 mg/ml, Thiotepa 10 mg/ml, Vincristine 1 mg/ml. Test Fentanyl Citrate are as follows: Fentanyl Citrate Injection 100mcg/2ml. Warning: Do not use with Carmustine and Thiotepa. Please note that the following drugs have low permeation times: (1)Carmustine - 14.1 minutes (2)Thiotepa - 57.2 minutes.

    Device Description

    Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250. Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs are Class I patient examination gloves bear the product code Nitrile - LZA, LZC, OPJ and QDO (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in five sizes (XS, S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.

    AI/ML Overview

    The provided document is a 510(k) summary for Nitrile Powder Free Blue Patient Examination Gloves. It details non-clinical testing to demonstrate the device meets acceptance criteria. No information regarding clinical effectiveness studies (such as MRMC studies or standalone algorithm performance for AI devices) is present, as this is not an AI/ML device.

    Here's an analysis of the acceptance criteria and the studies that prove the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Characteristic and ParameterStandard / Test MethodAcceptance CriteriaProposed Device Performance (Result)Comparison Analysis (where available)
    Residual PowderASTM D6124-06 (Reapproved 2017)Less than 2mg / gloveXS: 0.25mg/glove, S: 0.20mg/glove, M: 0.28mg/glove, L: 0.28mg/glove, XL: 0.38mg/gloveSimilar
    Freedom of HolesASTM D5151-19Inspection level, G-I AQL 2.5 (In accordance with ASTM D6319-19)Passed G-I, AQL 1.5Similar (meets AQL 1.5 which is better than 2.5)
    Physical Dimensions (ASTM D6319-19)
    LengthXS: 220mm, min; S: 220mm, min; M: 230mm, min; L: 230mm, min; XL: 230mm, minXS: 245-251mm; S: 247-255mm; M: 245-255mm; L: 247-255mm; XL: 247-252mmSimilar
    WidthXS: 70 ± 10mm; S: 80 ± 10mm; M: 95 ± 10mm; L: 110 ± 10mm; XL: 120 ± 10mmXS: 76-78mm; S: 82-88mm; M: 95-99mm; L: 111-118mm; XL: 120-127mmSimilar, subject device meets ASTM D6319 requirements
    Thickness (Palm)0.05mm, minXS: 0.06-0.06mm; S: 0.06-0.07mm; M: 0.06-0.07mm; L: 0.06-0.06mm; XL: 0.06-0.06mm (average)Similar (all above min)
    Thickness (Finger)0.05mm, minXS: 0.07-0.07mm; S: 0.07-0.08mm; M: 0.07-0.08mm; L: 0.07-0.08mm; XL: 0.07-0.08mm (average)Similar (all above min)
    Physical Properties (ASTM D6319-19)
    Tensile Strength (Before aging)14 MPa, minXS: 23.97 MPa, S: 26.35 MPa, M: 25.15 MPa, L: 25.11 MPa, XL: 25.85 MPa (average)Similar (all above min)
    Tensile Strength (After accelerated aging)14 MPa, minXS: 27.11 MPa, S: 26.78 MPa, M: 29.22 MPa, L: 25.94 MPa, XL: 26.50 MPa (average)Similar (all above min)
    Ultimate Elongation (Before aging)500%, minXS: 571%, S: 571%, M: 538%, L: 552%, XL: 525% (average)Similar (all above min)
    Ultimate Elongation (After accelerated aging)400%, minXS: 462%, S: 458%, M: 434%, L: 457%, XL: 434% (average)Similar (all above min)
    Chemotherapy Drug Permeation (ASTM D6897-05)
    Carmustine 3.3mg/ml>240 minutes (General criteria mentioned in Table 2, but specific values for individual drugs are listed in the Indications for Use and Summary of Technological Characteristics)14.1 minutesSimilar; Below 240 minutes permeation times (Warning included)
    Cisplatin 1 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Cyclophosphamide 20 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Dacarbazine 10 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Doxorubicin, HCl 2 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Etoposide 20 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Fluorouracil 50 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Methotrexate 25 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Mitomycin 0.5 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Oxaliplatin 5 mg/mlNot specified in predicate, but >240 min in proposed device>240 minutesOptional, subject device performs additional Chemo drug test
    Paclitaxel 6 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Thiotepa 10 mg/ml>240 minutes (General criteria)57.2 minutesSimilar; Below 240 minutes permeation times (Warning included)
    Vincristine 1 mg/ml>240 minutes (General criteria)>240 minutesSimilar
    Fentanyl Citrate Injection 100mcg/2ml>240 minutes (General criteria)>240 minutesSimilar
    Biocompatibility Testing
    Animal Irritation TestISO 10993-10Under the condition of study not an irritant.No observable irreversible alteration; PII = "0"; not corrosive; negligible primary irritation. Met requirement.Similar
    Dermal Sensitization AssayISO 10993-10Under the condition of the study not a sensitizer.No sensitization induced. Met requirement.Similar
    Acute Systemic ToxicityISO 10993-11Not induce systemic toxicityNon-toxic effects of extracts. Met requirement.Similar

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for each individual test (e.g., number of gloves tested for holes, number of material samples for tensile strength). However, the tests are conducted in accordance with the specified ASTM and ISO standards, which typically include defined sampling plans.

    • Data Provenance: The origin of the data is from non-clinical laboratory testing performed by "Mah Sing Healthcare Sdn Bhd" or their contracted laboratories, as part of the 510(k) submission process. The submission is from Malaysia. This is prospective data generated for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the device is a physical medical glove and the testing involves objective measurement against pre-defined physical, chemical, and biological standards, not interpretation by human experts or AI algorithms. Ground truth is established by the test procedures and acceptance criteria defined in the ASTM and ISO standards.

    4. Adjudication Method for the Test Set:

    This is not applicable. As stated above, the device performance is evaluated against objective standards, not through adjudication of expert interpretations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This is not applicable. The device is a physical medical glove. The submission does not involve Artificial Intelligence (AI) or Machine Learning (ML) components, nor does it relate to diagnostic image interpretation where MRMC studies would typically be conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is not applicable. The device is a physical medical glove and does not involve an algorithm.

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is based on defined performance specifications and measurement methodologies established by widely recognized consensus standards such as ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). For example:

    • Physical dimensions: Measured against specified millimeter ranges.
    • Tensile strength and elongation: Measured against minimum MPa and percentage values.
    • Freedom from holes: Assessed against an Acceptable Quality Limit (AQL) per ASTM D5151-19.
    • Chemotherapy drug permeation: Measured as breakthrough detection time in minutes per ASTM D6978-05.
    • Biocompatibility: Assessed based on observable reactions in animal models according to ISO 10993 series and reported as "Passed" or "Not an irritant/sensitizer/systemic toxin."

    8. The Sample Size for the Training Set:

    This is not applicable. There is no "training set" as this device is not an AI/ML product. The manufacturing process is controlled and monitored, but there isn't a data-driven training phase for the device itself.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable. As there is no training set for an AI/ML algorithm, there is no ground truth established for one. The "ground truth" for glove manufacturing is established through quality control and adherence to standardized manufacturing practices and raw material specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K214110
    Date Cleared
    2022-02-19

    (52 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Mah Sing Healthcare Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

    Chemotherapy Drugs Permeation
    The following chemicals have been tested with these gloves.

    Test Chemotherapy Drugs Concentration Breakthrough Detection Time in minutes *Carmustine 3.3mg/ml 24.8 Cisplatin 1 mg/ml >240 Cyclophosphamide 20 mg/ml >240 Dacarbazine 10 mg/ml >240 Doxorubicin, HCl 2 mg/ml >240 Etoposide 20 mg/ml >240 Fluorouracil 50 mg/ml >240 Methotrexate 25 mg/ml >240 Mitomycin C 0.5 mg/ml >240 Oxaliplatin 5 mg/ml >240 Paclitaxel 6 mg/ml >240 *Thiotepa 10 mg/ml 38.4 Vincristine 1 mg/ml >240

    Warning- Not for use with Carmustine and Thiotepa.

    Device Description

    Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250.

    Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs are Class I patient examination gloves bear the product code Nitrile – LZA (21CFR880.6250).

    The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in four sizes (S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for "Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs". This document outlines the acceptance criteria and performance data for a medical device that is a physical commodity, not an AI/ML powered device. As such, information regarding AI/ML specific criteria like human-in-the-loop performance, effect size of human readers with AI assistance, standalone algorithm performance, number of experts for ground truth, or training set details are not applicable and not present in the document.

    The acceptance criteria and reported device performance are based on recognized ASTM and ISO standards for medical gloves.

    Here's the information extracted from the provided text, structured according to your request where applicable for a non-AI/ML device:

    1. A table of acceptance criteria and the reported device performance

    Characteristics and ParametersStandardAcceptance CriteriaReported Device Performance and Results
    Chemotherapy Drugs Permeation Test (ASTM D6978-05 (2019))ASTM D6978-05Minimum Breakthrough Detection Time (min)
    Carmustine-24.824.8 min
    Cisplatin->240>240 min
    Cyclophosphamide->240>240 min
    Dacarbazine->240>240 min
    Doxorubicin, HCl->240>240 min
    Etoposide->240>240 min
    Fluorouracil->240>240 min
    Methotrexate->240>240 min
    Mitomycin C->240>240 min
    Oxaliplatin->240>240 min
    Paclitaxel->240>240 min
    Thiotepa-38.438.4 min
    Vincristine->240>240 min
    Residual Powder ContentASTM D6124-06 (Reapproved 2017)Less than 2mg / gloveSize S 0.32mg /glove, Size M 0.28mg /glove, Size L 0.32mg /glove, Size XL 0.30mg /glove
    Freedom from HolesASTM D5151-19Inspection level, G-I AQL 2.5 (In accordance with ASTM D6319-19)Passed G-I, AQL 1.5
    Physical Dimensions - LengthASTM D6319-19Size S 220mm, min; Size M 230mm, min; Size L 230mm, min; Size XL 230mm, minSize S 250 - 255mm, Size M 250 - 255mm, Size L 245 - 252mm, Size XL 245 - 252mm
    Physical Dimensions - WidthASTM D6319-19Size S 80 ± 10mm; Size M 95 ± 10mm; Size L 110 ± 10mm; Size XL 120 ± 10mmSize S 83 - 85mm, Size M 93 - 96mm, Size L 105 - 109mm, Size XL 116 - 118mm
    Physical Dimensions - Thickness (Cuff)ASTM D6319-19Measured in single wall at cuff edge area Cuff 0.05mm, minSize S 0.05 - 0.06mm, Size M 0.05 - 0.06mm, Size L 0.05 - 0.06mm, Size XL 0.05 - 0.06mm
    Physical Dimensions - Thickness (Palm)ASTM D6319-19Measured in single wall at approximate center of palm area Palm 0.05mm, minSize S 0.06 - 0.07mm, Size M 0.06 - 0.06mm, Size L 0.06 - 0.07mm, Size XL 0.06 - 0.07mm
    Physical Dimensions - Thickness (Finger)ASTM D6319-19Measured in single wall at 13±3mm from the tip of middle finger Finger 0.05mm, minSize S 0.08 - 0.09mm, Size M 0.09 - 0.10mm, Size L 0.10 - 0.11mm, Size XL 0.10 - 0.11mm
    Tensile Strength (Before aging)ASTM D6319-1914MPa, min for all sizesSize S 21.94 MPa, average; Size M 22.08 MPa, average; Size L 21.76 MPa, average; Size XL 22.05 MPa, average
    Tensile Strength (After accelerated aging)ASTM D6319-1914MPa, min for all sizesSize S 29.53 MPa, average; Size M 29.12 MPa, average; Size L 24.90 MPa, average; Size XL 29.63 MPa, average
    Ultimate Elongation (Before aging)ASTM D6319-19500%, min for all sizesSize S 537%, average; Size M 550%, average; Size L 526%, average; Size XL 536%, average
    Ultimate Elongation (After accelerated aging)ASTM D6319-19400%, min for all sizesSize S 452%, average; Size M 450%, average; Size L 436%, average; Size XL 458%, average
    Animal Irritation Test (ISO 10993-10)ISO 10993-10Under the condition of study not an irritant.No observable irreversible alteration on the skin; Primary Irritation Index (PII) was "0"; not corrosive; Primary Irritation Response Category is "negligible".
    Dermal Sensitization Assay Test (ISO 10993-10)ISO 10993-10Under the condition of the study not a sensitizer.No sensitization induced by the application of the test material on albino guinea pigs.
    Acute Systemic Toxicity (ISO 10993-11)ISO 10993-11Not induce systemic toxicityThe single dose acute systemic toxicity of extracts from test material using both normal saline and sesame oil, shown non-toxic effects.

    2. Sample size used for the test set and the data provenance
    The document does not explicitly state the specific sample sizes used for each test (e.g., number of gloves tested for each characteristic) or the detailed data provenance beyond adherence to international standards like ASTM and ISO. Given the nature of product testing for medical gloves, it is implied that appropriate sample sizes were used as prescribed by the respective ASTM and ISO standards. The tests are non-clinical (bench tests and biocompatibility tests). There is no explicit mention of data provenance in terms of country of origin or retrospective/prospective data collection for these non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This is not applicable as the "ground truth" for physical properties and chemical permeation of medical gloves is established by the specified ASTM and ISO standards and their defined methodologies, not by expert consensus on specific samples in the same way an AI model's output would be adjudicated. The tests are objective measurements against defined criteria.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. The tests are objective measurements according to established standards. There is no human adjudication process described for these types of tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" is defined by the technical specifications and methodologies outlined in the referenced ASTM standards (e.g., ASTM D6319-19 for physical properties, ASTM D6978-05 (2019) for chemotherapy drug permeation, ASTM D6124-06 for residual powder, ASTM D5151-19 for freedom from holes) and ISO standards (e.g., ISO 10993-10 and ISO 10993-11 for biocompatibility). These standards provide objective measurement criteria and test protocols.

    8. The sample size for the training set
    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212535
    Date Cleared
    2021-12-04

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Mah Sing Healthcare Sdn Bhd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Nitrile Powder Free Blue Examination Gloves, Non-Sterile are Class I patient examination gloves bearing the product code Nitrile – LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous single use disposable devices that come in four sizes (S, M, L and XL).

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of a study for the "Nitrile Powder Free Blue Examination Gloves, Non-Sterile". Here's the information broken down as requested:

    1. A table of acceptance criteria and the reported device performance

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance (Result)
    ASTM D6124-06 (Reapproved 2017)To determine the residual powder in the glovesLess than 2mg / gloveSize S: 0.36mg /glove
    Size M: 0.32mg /glove
    Size L: 0.36mg /glove
    Size XL: 0.38mg /glove
    ASTM D5151-19To determine the holes in the glovesInspection level, G-I, AQL 2.5 (In accordance with ASTM D6319-19)Passed G-I, AQL 1.5
    ASTM D6319-19 LengthTo determine the length of the glovesSize S: 220mm, min
    Size M: 230mm, min
    Size L: 230mm, min
    Size XL: 230mm, minSize S: 254mm, average
    Size M: 254mm, average
    Size L: 254mm, average
    Size XL: 253mm, average
    ASTM D6319-19 WidthTo determine the width of the glovesSize S: 80 ± 10mm
    Size M: 95 ± 10mm
    Size L: 110 ± 10mm
    Size XL: 120 ± 10mmSize S: 86mm, average
    Size M: 96mm, average
    Size L: 107mm, average
    Size XL: 118mm, average
    ASTM D6319-19 Palm ThicknessTo determine the thickness of the glovesMeasured in single wall at approximate center of palm area: Palm 0.05mm, minSize S: 0.06mm, average
    Size M: 0.06mm, average
    Size L: 0.06mm, average
    Size XL: 0.06mm, average
    ASTM D6319-19 Finger ThicknessTo determine the thickness of the glovesMeasured in single wall at 13±3mm from the tip of middle finger: Finger 0.05mm, minSize S: 0.08mm, average
    Size M: 0.08mm, average
    Size L: 0.11mm, average
    Size XL: 0.10mm, average
    ASTM D6319-19 Tensile Strength (Before Ageing)To determine the physical properties - Tensile strengthBefore Ageing: Tensile Strength 14 MPa, min for all sizesSize S: 23 MPa, average
    Size M: 22 MPa, average
    Size L: 22 MPa, average
    Size XL: 22 MPa, average
    ASTM D6319-19 Tensile Strength (After Ageing)To determine the physical properties - Tensile strengthAfter Ageing: Tensile Strength 14 MPa, min for all sizesSize S: 23 MPa, average
    Size M: 23 MPa, average
    Size L: 23 MPa, average
    Size XL: 22 MPa, average
    ASTM D6319-19 Ultimate Elongation (Before Ageing)To determine the physical properties - Ultimate ElongationBefore Ageing: Ultimate Elongation 500%, min for all sizesSize S: 524%, average
    Size M: 522%, average
    Size L: 523%, average
    Size XL: 536%, average
    ASTM D6319-19 Ultimate Elongation (After Ageing)To determine the physical properties - Ultimate ElongationAfter Ageing: Ultimate Elongation 400%, min for all sizesSize S: 538%, average
    Size M: 532%, average
    Size L: 530%, average
    Size XL: 523%, average
    ISO 10993-10 (Animal Irritation Test)To determine the potential of the material to produce dermal irritation in RabbitsUnder the condition of study not an irritant.No observable irreversible alteration on the skin; Primary Irritation Index (PII) was "0"; not corrosive; Primary Irritation Response Category "negligible", meeting the requirement.
    ISO 10993-10 (Dermal Sensitization Assay Test)To determine the skin sensitization potential of the material in Guinea pigUnder the condition of the study not a sensitizer.No sensitization induced by the application of the test material on albino guinea pigs, meeting the requirement.
    ISO 10993-5 (MEM Elution Assay)To evaluate the in vitro cytotoxic potential of the test item (both inner and outer surface)Under the conditions of study non cytotoxicUnder the conditions of the study, cytotoxic. (Note: This is an outlier result compared to the acceptance criteria which states "non cytotoxic". The document then states "Meeting the requirements per ISO 10993-5" in Table 1 for the predicate device, but for the proposed device, it states "Under the conditions of the study, cytotoxic." in Table 2, and "Meeting the requirements per ISO 10993-5" in Table 1 where it compares to "Not done" for the predicate. This seems like a contradiction or an oversight in the provided text.)
    ISO 10993-11 (Acute Systemic Toxicity)To provide information on health hazards from a short-term exposure to the extracts of test material by intravenous and intraperitoneal injection in miceNot induce systemic toxicitySingle dose acute systemic toxicity of extracts using normal saline and sesame oil shown non-toxic effects, meeting the requirement.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each test. Instead, it refers to standards like ASTM and ISO, which would define the required sample sizes for such tests. For instance, for "Freedom of Holes," it refers to "AQL 1.5 with G1," which implies a sampling plan defined by ASTM D5151-19.

    The data provenance is from Mah Sing Healthcare Sdn Bhd, located in Kuala Lumpur, Malaysia. The tests are non-clinical bench tests and biocompatibility tests conducted to support a premarket notification (510(k)). These are typically prospective tests performed specifically for the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The device is a medical glove, and the tests relate to its physical properties, chemical properties (residual powder), and biocompatibility, as measured by standard laboratory methods. There is no mention of human expert assessment or ground truth establishment in the context of medical image analysis or similar diagnostic tasks.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically relevant in clinical studies or studies involving human interpretation of data where disagreements need to be resolved. The tests performed are laboratory-based and follow standardized testing protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. MRMC studies and AI assistance effectiveness are relevant for AI/ML-enabled diagnostic devices. The device in question is a medical glove, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a medical glove and does not involve any algorithms or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the tests performed is based on:

    • Standardized measurement techniques and physical limits defined by international standards (ASTM D6124-06, ASTM D5151-19, ASTM D6319-19).
    • Biological and chemical assays for biocompatibility (ISO 10993-10, ISO 10993-5, ISO 10993-11) where the "truth" is whether the material elicits a specific biological response (e.g., irritation, sensitization, cytotoxicity, systemic toxicity) above a defined threshold.

    8. The sample size for the training set

    This information is not applicable. There is no training set as the device is a physical product (medical glove) and not an AI/ML system.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1