Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202402
    Date Cleared
    2021-03-01

    (192 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NITRILE EXAMINATION GLOVES (Blue, Violet Blue, \Vhite) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device, NITRILE EXAMINATION GLOVES (Blue, Violet Blue, White) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are NITRILE EXAMINATION GLOVES (Blue, Violet Blue, White) includes variations of different size and color. The colors of the proposed device are Blue, and White.

    AI/ML Overview

    The document describes the K202402 Nitrile Examination Gloves. Here's a breakdown of the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria (Standard)Test ParameterProposed Device Performance (Minimum)Predicate Device AQL (or value)
    ASTM D6319-15Pinhole AQL1.5N/A (implied 1.5 by ASTM)
    Tensile Strength (Before Aging)15 MPa15 MPa
    Ultimate Elongation (Before Aging)500 %500 %
    Tensile Strength (After Aging)14 MPa14 MPa
    Ultimate Elongation (After Aging)500 %400 %
    ASTM D5151-06Freedom from HolesFree from holesFree from holes
    ASTM D6124-17Powder ContentLess than 2 mg per gloveMeet ASTM D6319 requirements
    ISO 10993-10:2010IrritationNot an irritantNot an irritant
    SensitizationNot a sensitizerNot a sensitizer
    ISO 10993-5:2009In vitro cytotoxicityN/A (implied non-cytotoxic)N/A (implied non-cytotoxic)
    BiocompatibilityAcute Systemic ToxicityNot toxic (for proposed device)Not Available (for predicate)
    Physical (Dimensions)Length (XS, S, M)230 mm min230 mm min
    Length (L, XL)235 mm min230 mm min
    Width70, 80, 95, 110, 120 mm (±10 mm)70, 80, 95, 110, 120 mm (±10 mm)
    Finger Thickness0.05 mm min0.10-0.12 mm (±0.03 mm)
    Palm Thickness0.05 mm min0.08-0.10 mm (±0.03 mm)
    Cuff Thickness0.05 mm min0.06-0.09 mm (±0.03 mm)

    Note: The document explicitly states that "The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319." and that the bench tests "demonstrated that the proposed device complies with the following standards".

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test that was performed. It generally refers to "bench tests" for physical properties and "biocompatibility tests".

    • Provenance: No specific country of origin is mentioned for the test data itself, beyond the manufacturer being in China. The study is non-clinical bench testing, not human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical glove, and the "ground truth" for its performance is established by adherence to recognized international standards (ASTM, ISO) through objective physical and biocompatibility testing, not expert consensus on interpretations of complex medical images or outcomes.

    4. Adjudication method for the test set

    Not applicable, as the tests are objective measurements against standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical glove, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical glove, not an algorithm.

    7. The type of ground truth used

    The ground truth used for performance evaluation is defined by international consensus standards for medical gloves (ASTM D6319-15, ASTM D5151-06, ASTM D6124-17) and biocompatibility testing standards (ISO 10993-10:2010, ISO 10993-5:2009). The device is tested against the specified parameters and thresholds outlined in these standards.

    8. The sample size for the training set

    Not applicable. This is a physical medical device (gloves), not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device (gloves), not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1