Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213044
    Date Cleared
    2022-01-21

    (121 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in six specifications: XS,S,M,L XL and XXL. The subject device is non-sterile.

    AI/ML Overview

    This document is a 510(k) premarket notification for Nitrile Examination Gloves (K213044). It describes the non-clinical testing performed to demonstrate that the device meets acceptance criteria and is substantially equivalent to a predicate device (K171422).

    Here's the requested information:

    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength (mm): XS/S: ≥ 220; M/L/XL/XXL: ≥ 230
    Width (mm): XS: 70 ± 10; S: 80 ± 10; M: 95 ± 10; L: 110 ± 10; XL: 120 ± 10; XXL: 130 ± 10
    Thickness (mm): Finger: ≥ 0.05; Palm: ≥ 0.05Length (mm): XS/S/M/L/XL/XXL: > 240/Pass
    Width (mm): XS: 74-78 /Pass; S: 82-88 /Pass; M: 95-99/ Pass; L: 106-109/ Pass; XL: 110-116/ Pass; XXL: 127-129/ Pass
    Thickness (mm): Finger: 0.16-0.18/Pass; Palm: 0.12-0.14/Pass
    ASTM D5151Watertightness Test for Detection of HolesBe free from holes when tested in accordance with ASTM D5151, AQL=2.5XS: 0/125/Pass; S: 0/125/Pass; M: 0/125/Pass; L: 1/124/Pass; XL: 0/125/Pass; XXL: 0/125/Pass (All sizes passed, as the AQL for holes is 2.5, implying a certain number of failures are permissible within the sample size.)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124
    Ask a Question

    Ask a specific question about this device

    K Number
    K211045
    Date Cleared
    2021-06-23

    (76 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for "Synthetic Vinyl Examination Gloves." This type of document is for medical devices that are deemed substantially equivalent to a predicate device, and thus, typically does not involve the rigorous clinical studies that would require extensive details on acceptance criteria and study methodologies as one might find for a novel device or software.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (K153028) by comparing performance against recognized ASTM and ISO standards for examination gloves. Therefore, the questions related to AI performance, reader studies, ground truth establishment, and training sets are not applicable to this device.

    Here's an analysis of the available information regarding acceptance criteria and testing, tailored to the context of a medical glove:


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the "Synthetic Vinyl Examination Gloves" are based on recognized industry standards for medical gloves, primarily ASTM D5250, ASTM D5151, and ASTM D6124, along with biocompatibility standards ISO 10993-5 and ISO 10993-10. The device's reported performance demonstrates compliance with these standards.

    Acceptance Criteria (Standard & Requirement)Reported Device Performance (Subject Device K211045)
    Physical Dimensions (ASTM D5250)
    Length, mm: min 230230 (all sizes XS-XL)
    Width, mm: ±5 from specified target78-115 (±5) (meets specification for each size)
    Thickness, mm (Finger): min 0.080.08
    Thickness, mm (Palm): min 0.080.08
    Physical Properties (ASTM D5250)
    Tensile Strength (Before Aging): 11 MPa, minComply with ASTM D5250 (Average 16.9MPa for predicate, subject device also complies)
    Ultimate Elongation (Before Aging): 300% minComply with ASTM D5250 (Average 550% for predicate, subject device also complies)
    Tensile Strength (After Aging): 11 MPa, minComply with ASTM D5250 (Average 14.4MPa for predicate, subject device also complies)
    Ultimate Elongation (After Aging): 300% minComply with ASTM D5250 (Average 550% for predicate, subject device also complies)
    Freedom from Holes (ASTM D5151)
    AQL=2.5 when tested in accordance with ASTM D5151Be free from holes when tested in accordance with ASTM D5151 AQL=2.5
    Powder Content (ASTM D6124)
    Meet the requirements of ASTM D6124Average 0.02 mg per glove
    Biocompatibility (ISO 10993-10)
    Not an irritantUnder the conditions of the study, not an irritant
    Not a sensitizerUnder conditions of the study, not a sensitizer
    Biocompatibility (ISO 10993-5)
    Did not show potential toxicity to L-929 cellsUnder conditions of the study, did not show potential toxicity to L-929 cells

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the exact sample sizes used for each of the non-clinical tests (e.g., number of gloves tested for tensile strength, holes, etc.). The tests refer to compliance with ASTM and ISO standards, which typically define their own sampling plans.
      • Data Provenance: The tests were conducted by the manufacturer, Jiangsu Huayuan Medical Technology Co., Ltd. The document does not explicitly state the country of origin of the data beyond the manufacturer's location in China. The testing appears to be prospective for the subject device to demonstrate compliance with the stated standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. For non-clinical tests of physical properties and biocompatibility of medical gloves, "ground truth" is established by laboratory measurement against predefined objective criteria in the standards, not by expert consensus or subjective assessment.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective diagnostic interpretations, especially in image analysis where human readers evaluate cases. For objective laboratory testing, results are determined by instrumentation and adherence to standard protocols.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This device is a medical glove, not an AI software or a device that involves human reader interpretation in a diagnostic context.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This device is a physical medical glove and does not involve any algorithms or AI for standalone performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the physical and performance tests, the "ground truth" is defined by the objective, quantitative specifications outlined in the referenced national and international standards (ASTM D5250, ASTM D5151, ASTM D6124).
      • For biocompatibility tests, the "ground truth" is established by the biological response observed in standardized in vitro (cytotoxicity) and in vivo (irritation, sensitization) tests as per ISO 10993-5 and ISO 10993-10, evaluated against predefined criteria for an acceptable biological reaction.
    7. The sample size for the training set:

      • Not Applicable. This device is a medical glove. It does not involve a "training set" in the context of machine learning or AI.
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1