Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K170364
    Date Cleared
    2017-05-02

    (85 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Ivy Sports Medicine, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ivy Sports Medicine Collagen Meniscus Implant (CMI) is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the CMI must extend at least into the red/white zone of the meniscus to provide sufficient vascularization.

    The CMI reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The CMI is not a prosthetic device and is not intended to replace normal body structure.

    Device Description

    The Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL) is a resorbable collagen matrix comprised primarily of bovine type I collagen. The CMI is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue. The surgeon trims the device to the size necessary for repair of the damaged or weakened soft tissue. The CMI XL is provided sterile for single use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the Ivy Sports Medicine Collagen Meniscus Implant XL (CMI XL). It asserts substantial equivalence to a predicate device, the ReGen Collagen Scaffold (K082079).

    Based on the provided document, here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    The document states: "Testing performed on the Collagen Meniscus Implant XL demonstrated that all of the predetermined acceptance criteria were met with passing results." However, it does not explicitly list the specific acceptance criteria or the numerical performance results against those criteria. It only makes a general statement about meeting them.

    2. Sample size used for the test set and the data provenance

    The document does not provide details on the sample size used for the test set or the data provenance. It mentions "Design verification testing" and "Bench performance test data" but no specifics on how many devices were tested or where the components/data originated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the document indicates "Clinical evaluation is not required for this device." The testing performed appears to be bench testing rather than clinical studies requiring expert ground truth for interpretation.

    4. Adjudication method for the test set

    This information is not applicable as the document indicates "Clinical evaluation is not required for this device" and the testing described is bench performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a physical collagen implant, not an AI-assisted diagnostic or interpretative tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical collagen implant, and there is no algorithm or human-in-the-loop component described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Since clinical evaluation was not required and the testing was described as "Design verification testing" and "Bench performance test data," the ground truth would likely be established by engineering specifications, material science standards, and comparison to the predicate device's established performance. The document implicitly suggests that the ground truth for performance relied on the predicate device's characteristics.

    8. The sample size for the training set

    This information is not applicable. The device is a physical implant, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable. The device is a physical implant, not an AI model that requires a training set.

    In summary, the document focuses on demonstrating substantial equivalence through bench testing, asserting that the new device meets predefined, unlisted, acceptance criteria based on the predicate device's performance, and explicitly stating that clinical evaluation was not required.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153087
    Date Cleared
    2016-07-14

    (262 days)

    Product Code
    Regulation Number
    878.5000
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Ivy Sports Medicine, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IvyAIR Meniscus System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The lvyAIR Meniscus System is indicated for use in meniscal repairs and allograft transplant procedures. The lvyAlR Meniscus System device is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures.

    Device Description

    The IvyAIR Meniscus System is an all-inside meniscus repair device. Each device includes two non-absorbable polymer implants, pre-tied with 2-0 non-absorbable suture and pre-loaded into a needle delivery system. The IvyAIR System is available in three needle configurations: Curved. Straight and Reverse Curve to allow the surgeon a choice for access to the desired zone of the meniscus. The IvyAIR Meniscus System is provided sterile for single use only.

    AI/ML Overview

    This document describes the IvyAIR Meniscus System, a medical device for meniscal repair. The submission focuses on demonstrating substantial equivalence to a predicate device rather than providing a study proving performance against specific acceptance criteria for a novel AI device. Therefore, a direct response to some of the requested points ("AI assistance effect size," "sample size for the training set," "how ground truth for the training set was established") is not applicable or not explicitly detailed in the provided text, as the device is a physical surgical suture retention system, not an AI-powered diagnostic or assistive tool.

    However, I can extract information related to the performance data and "acceptance criteria" as they pertain to the device's engineering design and biocompatibility testing in comparison to its predicate.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All of the pre-determined acceptance criteria were met with passing results." However, it does not explicitly list the specific numerical acceptance criteria for "Implant strength test to failure" or "Dynamic and static tensile loading." It only lists the tests performed. The acceptance criteria for biocompatibility tests are generally established by regulatory standards (e.g., ISO 10993).

    Test TypeAcceptance Criteria (Not Explicitly Quantified)Reported Device Performance
    Bench Testing
    Implant strength test to failure(Implied: Meets or exceeds requirements for meniscal repair devices, likely compared to predicate performance)Met pre-determined criteria
    Dynamic and static tensile loading(Implied: Meets or exceeds requirements for meniscal repair devices, likely compared to predicate performance)Met pre-determined criteria
    Biocompatibility Testing
    MEM Elution Using L-929 Mouse Fibroblast Cells(Implied: Non-cytotoxic, as per ISO 10993 standards)Biocompatible
    Guinea Pig Maximization Sensitization Test(Implied: Non-sensitizing, as per ISO 10993 standards)Biocompatible
    Intracutaneous Reactivity Test(Implied: Non-irritant, as per ISO 10993 standards)Biocompatible
    Acute Systemic Injection Test(Implied: Non-toxic systemically, as per ISO 10993 standards)Biocompatible
    Material mediated pyrogen(Implied: Non-pyrogenic, as per ISO 10993 standards)Biocompatible
    Subacute (14 Day) Intraperitoneal Toxicity Study In Rats(Implied: No significant toxicity, as per ISO 10993 standards)Biocompatible
    Subchronic (14 Day) Intravenous Toxicity Study In Rats (including histopathology)(Implied: No significant toxicity, as per ISO 10993 standards)Biocompatible
    Bacterial Mutagenicity Test (Ames Assay)(Implied: Non-mutagenic, as per ISO 10993 standards)Biocompatible
    In Vitro Mouse Lymphoma Assay(Implied: Non-genotoxic, as per ISO 10993 standards)Biocompatible
    In Vivo Mouse Micronucleus Assay(Implied: Non-genotoxic, as per ISO 10993 standards)Biocompatible
    Intramuscular Implantation Tests (2 weeks and 13 weeks)(Implied: No adverse local tissue reaction, as per ISO 10993 standards)Biocompatible

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for the bench and biocompatibility tests. The data provenance is implied to be from internal testing conducted by Ivy Sports Medicine, LLC. The testing is prospective in the sense that it was conducted specifically to support this regulatory submission, but the tests themselves are standard laboratory/bench tests, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is not an AI/diagnostic device that requires expert ground truth for its performance evaluation (e.g., interpreting images). Its performance is evaluated through material science and biological compatibility testing against established standards.

    4. Adjudication Method for the Test Set

    Not applicable. This concept is typically relevant to clinical trials or AI performance evaluations involving human interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is a physical medical device (suture retention system), not an AI-powered system designed to assist human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device.

    7. The Type of Ground Truth Used

    For the bench testing (Implant strength, Dynamic and static tensile loading), the "ground truth" is typically defined by engineering specifications and comparative performance data against the predicate device, or industry standards for mechanical properties.

    For biocompatibility testing, the "ground truth" is the established biological response as defined by international standards (e.g., ISO 10993 series) and observed biological effects in the conducted studies.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1