Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142054
    Manufacturer
    Date Cleared
    2014-09-29

    (62 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Zura-EVO System is intended for the following applications: Cardiac and , Transesophageal. Measurement and calculation packages that provide information of anatomical structures that may be used by a physician for clinical diagnosis purposes.

    2. When used with the ClariTEE probe -- The episodic assessment of cardiac function using transesophageal echocardiography (TEE). It is indical for use in clinical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs.

    3. When used with the TTE probe is intended for imaging and assessment of Cardiac anatomy and function

    Device Description

    The Zura-EVO Imaging System is a multi-purpose mobile, software controlled diagnostic ultrasound system. Its function is to acquire ultrasound echo data and display it in B-Mode, or Color Doppler Mode on a Flat Panel Display. The user interface includes specialized controls, a minimized computer keyboard, and touch panel on an ergonomic console.

    The system has an optional electrocardiography (ECG) display feature and support for a 3- lead ECG cable assembly. The systems provide measurement capabilities for anatomical structures that provide information used for clinical diagnostic purposes. The system has measurements and calculations, image storage and review, and recording capabilities. The systems include a Digital Imaging and Communications (DICOM) module which enables storage.

    The Zura-EVO Imaging System consists of the following major components:

    1. Ultrasound Machine: A predicate device ultrasound machine (K102997) 2. Ultrasound Probes

    a. ClariTEE Probe: A miniaturized TEE probe optimized for longer dwell time relative to standard TEE probes that enables long term use in clinical settings such as the ICU. This probe has predicate coverage under K080223 and K100969.

    b. TTE probe SA4-2/24: Transthoracic probe has a predicate coverage under K102997

    1. Ultrasound Imaging Software: Ultrasound imaging functions are controlled via the ImaCor Zura software. The Graphical User Interface, which is the piece of software exposed to the user, communicates user inputs to the Ultrasound Imaging Platform which interfaces with the Ultrasound Transducer. The ImaCor Zura software also communicates with the probe directly to monitor the probe's lifespan. Orientation and position of the probe is controlled manually by the user.
    AI/ML Overview

    The provided document is a 510(k) summary for the Zura-EVO Imaging System. It describes the device, its intended use, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets acceptance criteria.

    The document primarily focuses on regulatory approval based on demonstrating substantial equivalence to existing devices, rather than presenting a performance study with acceptance criteria. It mentions non-clinical testing for electrical safety, sterility, aged acceleration testing, and acoustic output testing against specified standards (IEC 60601, ISO 14971, ISO 10993). However, these are general compliance tests for device safety and performance according to harmonized standards, not specific clinical performance metrics with acceptance thresholds typically seen in studies for medical device approval.

    Therefore, I cannot extract the requested information from the provided text. The document does not include:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes, data provenance, or details about test set anaysis.
    • Number or qualifications of experts for ground truth establishment.
    • Adjudication methods.
    • Information about MRMC comparative effectiveness studies or effect sizes.
    • Details on standalone algorithm performance.
    • The type of ground truth used for specific clinical assessments.
    • Training set sample sizes or how ground truth was established for training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100989
    Manufacturer
    Date Cleared
    2010-06-25

    (77 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ImaCor Zura system with ClariTEE probe is intended for use in the episodic assessment of cardiac function using transesophageal echocardiography. It is indicated for use in clinical settings, including long term settings such as the ICU. for an indwelling time period not to exceed 72 hours. The ImaCor Zura system with ClariTEE probe is not intended for pediatric use

    Device Description

    The ImaCor Zura system with ClariTEE probe consists of three main components:

    1. Ultrasound Machine: A TEE predicate device optimized for use with ImaCor ClariTEE probe.
    2. Ultrasound Probe (ClariTEE formerly known as the Blue Probe in K080223): A miniaturized TEE probe optimized for longer dwell time relative to standard TEE probes enables use in longer term clinical settings such as the ICU. The probe distal tip is flexed upward transiently to obtain standard TEE images.
    3. Ultrasound Imaging Software: The software controls standard ultrasound machine functions such as imaging, recording and measuring. Continuous imaging is limited by a 20 minute software interlock should the operator mistakenly leave the machine in continuous imaging mode, thus limiting the potential unintentional exposure of the patient's mucosal tissue to acoustic energy. Maximum probe face temperature is limited according to FDA consensus standard IEC 60601-2-37. Latest software version enables two modes of operation; type B and color flow Doppler. The subsequent version cleared for marketing under K080223 enabled type B only.
    AI/ML Overview

    The ImaCor Zura system with ClariTEE probe, a Transesophageal Echo Imaging System, received 510(k) clearance for software modification. The primary change was the addition of a color flow Doppler mode of operation.

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Implicit)Reported Device Performance
    Intended UseIdentical to predicate device (ImaCor Zura K080223)"The intended use of the ImaCor Zura with ClariTEE is identical to that of the ImaCor Zura (K080223)."
    Indications for UseFor episodic assessment of cardiac function using TEE; indicated for clinical settings including ICU for indwelling time not exceeding 72 hours; not for pediatric use."The ImaCor Zura system with ClariTEE probe is intended for use in the episodic assessment of cardiac function using transesophageal echocardiography. It is indicated for use in clinical settings, including long term settings such as the ICU, for an indwelling time period not to exceed 72 hours. The ImaCor Zura system with ClariTEE probe is not intended for pediatric use." (Verbatim from K100989)
    Technological CharacteristicsSimilar to predicate devices, with specified modifications related to the probe and software.System consists of ultrasound machine (predicate optimized for ClariTEE), ClariTEE probe (miniaturized TEE probe for longer dwell time, flexed for images), and ultrasound imaging software (controls functions, 20-minute continuous imaging interlock, max probe face temp limited by IEC 60601-2-37).
    Software ModesAddition of color flow Doppler mode to support cardiac function assessment."Latest software version enables two modes of operation; type B and color flow Doppler."
    Safety - Acoustic EnergyLimit potential unintentional exposure of patient's mucosal tissue to acoustic energy."Continuous imaging is limited by a 20 minute software interlock should the operator mistakenly leave the machine in continuous imaging mode, thus limiting the potential unintentional exposure of the patient's mucosal tissue to acoustic energy."
    Safety - Probe TemperatureMaximum probe face temperature limited by FDA consensus standard IEC 60601-2-37."Maximum probe face temperature is limited according to FDA consensus standard IEC 60601-2-37."
    Effectiveness (New Feature)Color flow mode effectively demonstrates design modification for assessing cardiac function."Performance data using a flow phantom demonstrates the effectiveness of the design modification; the addition of a color flow mode of operation to support the assessment of cardiac function."
    Overall Safety & EffectivenessAs safe and effective as predicate devices."The ImaCor Zura System with ClariTEE probe is as safe and effective as the predicate devices." (Claim of Substantial Equivalence)

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Performance data using a flow phantom demonstrates the effectiveness of the design modification; the addition of a color flow mode of operation to support the assessment of cardiac function."

    • Test Set Sample Size: Not specified for the flow phantom. Flow phantom studies typically involve simulating various flow conditions but don't involve human subject sample sizes in the traditional sense.
    • Data Provenance: The study was conducted on a "flow phantom," which is a controlled, laboratory-based setup designed to simulate fluid flow. This implies an in-vitro or ex-vivo experimental setting, not human subject data. The country of origin is not specified, but the submission is to the US FDA, so presumably the testing was done in a manner compliant with US regulatory expectations, likely in the US or a country with comparable standards. It is a prospective test specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The performance data was gathered from a "flow phantom" for the new color flow mode. This type of testing typically relies on engineering specifications and physical measurements (e.g., flow velocity, direction accuracy) rather than expert human interpretation to establish ground truth for the device's technical function.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the performance data cited is from a "flow phantom," which means the "ground truth" for evaluating the function of the color flow mode would be based on the known parameters of the phantom and direct measurements from the device, not on expert adjudication of medical images.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The document only mentions performance data from a "flow phantom" and a claim of substantial equivalence to predicate devices based on this and other characteristics.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    Yes, a standalone performance assessment was done for the device's new feature (color flow mode) using a "flow phantom." This is an algorithm-only evaluation in the sense that the device's ability to accurately depict flow was tested independently of human interpretation of clinical images. The data pertains to the technical function of the device itself.

    7. The Type of Ground Truth Used

    For the specific performance data cited (color flow mode), the ground truth was known physical properties and measurements of the flow phantom. This is typically based on pre-established parameters of the phantom and/or reference measurements from other calibrated equipment, not on expert consensus, pathology, or outcomes data, as those relate more to clinical accuracy studies.

    8. The Sample Size for the Training Set

    Not applicable to this 510(k) submission. This submission is for modifications to an existing device (software change to add color flow Doppler), not the initial development of an AI algorithm that would require a distinct training set. The "performance data" refers to validation testing, not model training.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no mention of a training set or an AI algorithm being independently trained in this 510(k) summary. The submission focuses on modifications to an existing, cleared ultrasound system and probe.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1