K Number
K142054

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-09-29

(62 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. The Zura-EVO System is intended for the following applications: Cardiac and , Transesophageal. Measurement and calculation packages that provide information of anatomical structures that may be used by a physician for clinical diagnosis purposes.

  2. When used with the ClariTEE probe -- The episodic assessment of cardiac function using transesophageal echocardiography (TEE). It is indical for use in clinical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs.

  3. When used with the TTE probe is intended for imaging and assessment of Cardiac anatomy and function

Device Description

The Zura-EVO Imaging System is a multi-purpose mobile, software controlled diagnostic ultrasound system. Its function is to acquire ultrasound echo data and display it in B-Mode, or Color Doppler Mode on a Flat Panel Display. The user interface includes specialized controls, a minimized computer keyboard, and touch panel on an ergonomic console.

The system has an optional electrocardiography (ECG) display feature and support for a 3- lead ECG cable assembly. The systems provide measurement capabilities for anatomical structures that provide information used for clinical diagnostic purposes. The system has measurements and calculations, image storage and review, and recording capabilities. The systems include a Digital Imaging and Communications (DICOM) module which enables storage.

The Zura-EVO Imaging System consists of the following major components:

  1. Ultrasound Machine: A predicate device ultrasound machine (K102997) 2. Ultrasound Probes

a. ClariTEE Probe: A miniaturized TEE probe optimized for longer dwell time relative to standard TEE probes that enables long term use in clinical settings such as the ICU. This probe has predicate coverage under K080223 and K100969.

b. TTE probe SA4-2/24: Transthoracic probe has a predicate coverage under K102997

  1. Ultrasound Imaging Software: Ultrasound imaging functions are controlled via the ImaCor Zura software. The Graphical User Interface, which is the piece of software exposed to the user, communicates user inputs to the Ultrasound Imaging Platform which interfaces with the Ultrasound Transducer. The ImaCor Zura software also communicates with the probe directly to monitor the probe's lifespan. Orientation and position of the probe is controlled manually by the user.
AI/ML Overview

The provided document is a 510(k) summary for the Zura-EVO Imaging System. It describes the device, its intended use, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets acceptance criteria.

The document primarily focuses on regulatory approval based on demonstrating substantial equivalence to existing devices, rather than presenting a performance study with acceptance criteria. It mentions non-clinical testing for electrical safety, sterility, aged acceleration testing, and acoustic output testing against specified standards (IEC 60601, ISO 14971, ISO 10993). However, these are general compliance tests for device safety and performance according to harmonized standards, not specific clinical performance metrics with acceptance thresholds typically seen in studies for medical device approval.

Therefore, I cannot extract the requested information from the provided text. The document does not include:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes, data provenance, or details about test set anaysis.
  • Number or qualifications of experts for ground truth establishment.
  • Adjudication methods.
  • Information about MRMC comparative effectiveness studies or effect sizes.
  • Details on standalone algorithm performance.
  • The type of ground truth used for specific clinical assessments.
  • Training set sample sizes or how ground truth was established for training data.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2014

ImaCor, Inc. % Mr. James W. Monroe Director of QA/RA 839 Stewart Avenue, Suite 3 GARDEN CITY NY 11530

Re: K142054 Trade/Device Name: Zura-EVO Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: July 30, 2014 Received: July 31, 2014

Dear Mr. Monroe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the Zura-EVO Imaging System, as described in your premarket notification:

Transducer Model Number

ClariTEE Probe

TTE probe SA4-2/24

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

FDA signature

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142054

Device Name Zura-EVO Imaging System

Indications for Use (Describe)

  1. The Zura-EVO System is intended for the following applications: Cardiac and , Transesophageal. Measurement and calculation packages that provide information of anatomical structures that may be used by a physician for clinical diagnosis purposes.

  2. When used with the ClariTEE probe -- The episodic assessment of cardiac function using transesophageal echocardiography (TEE). It is indical for use in clinical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs.

  3. When used with the TTE probe is intended for imaging and assessment of Cardiac anatomy and function

Type of Use (Select one or both, as applicable)
---------------------------------------------------

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Zura-EVO Imaging System

Intended Use: When used with the ClariTEE probe -- The episodic assessment of cardian using transesophageal echocardiography (TEE). It is indical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs. When used with the SA4-2/24 Phased Array Transducer (TTE Probe) it is intended for Diagnostic Ultrasound imagining or fluid analysis of the human body as follows:

Mode of Operation
Clinical ApplicationProbeBMPWDopplerCWDopplerColorDopplerPowerDopplerCombinedModesOther(Notes)
Opthalmic
Fetal
Abdominal
Intraoperative¹
IntraoperativeNeurological
Pediatric
Small Organ²
Neonatal Cephalic
Adult Cephalic
CardiacSA4-2/24 PhasedArray Transducer(TTE Probe)NN
TransesophagealClariTEE ProbePP
Transrectal
Transvaginal
Transurethral
Transcranial
Peripheral Vascular
Laparoscopic
MSK Conventional
MSK Superficial
Vascular Access
Nerve Block
Other

N = New Indication; P = Previously cleared under K080223, K100989

(Division Sign-off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K K

{4}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATION FOR USE FORM

ClariTEE Probe

Intended Use: When used with the ClariTEE probe -- The episodic assessment of cardiac function using transesophageal echocardiography (TEE). It is indicated for use in clinical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs.

Mode of Operation
Clinical ApplicationBMPWDopplerCWDopplerColorDopplerPowerDopplerCombinedModesOther(Notes)
Opthalmic
Fetal
Abdominal
Intraoperative1
Intraoperative Neurological
Pediatric
Small Organ²
Neonatal Cephalic
Adult Cephalic
Cardiac
TransesophagealPP
Transrectal
Transvaginal
Transurethral
Transcranial
Peripheral Vascular
Laparoscopic
MSK Conventional
MSK Superficial
Vascular Access
Nerve Block
Other

N = New Indication; P = Previously cleared under K080223, K100989

{5}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATION FOR USE FORM

SA4-2/24 Phased Array Transducer (TTE Probe)

Mode of Operation
Clinical ApplicationBMPWDopplerCWDopplerColorDopplerPowerDopplerCombinedModesOther(Notes)
Opthalmic
Fetal
Abdominal
Intraoperative¹
Intraoperative Neurological
Pediatric
Small Organ²
Neonatal Cephalic
Adult Cephalic
CardiacNN
Transesophageal
Transrectal
Transvaginal
Transurethral
Transcranial
Peripheral Vascular
Laparoscopic
MSK Conventional
MSK Superficial
Vascular Access
Nerve Block
Other

Intended Use: Diagnostic Ultrasound imagining or fluid analysis of the human body as follows:

N = New Indication; P = Previously cleared under K080223, K100989

(Division Sign-off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K K

{6}------------------------------------------------

5. 510(K) Summary

510(K) SUMMARY

Zura-EVO Imaging System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:

ImaCor Inc 839 Stewart Ave Unit 3 Garden City, NY 11530 Phone: (732) 718-9199, Facsimile: (516) 393-0969 Contact Person: James W. Monroe; imonroe@imacorinc.com. 732-718-9199

Date Prepared: July 24, 2014

Name of Device and Name/Address of Sponsor

Zura-EVO Imaging System

ImaCor Inc 839 Stewart Ave Unit 3 Garden City, NY 11530

Common or Usual Name

Ultrasound Imaging System

Classification Name

Ultrasonic Pulsed Doppler Imaging System (892.1550) Ultrasonic Pulsed Echo Imaging System (892.1560) with a Diagnostic Ultrasonic Transducer (892.1570)

Product Codes IYO, IYN, ITX,

Device Class II

Predicate Devices ImaCor Zura K080223, K100989 Ultrasonix SonixTablet K102997

{7}------------------------------------------------

Intended Use / Indications for Use

  1. The Zura-EVO Imaging System is intended for the following applications: Cardiac and , Transesophageal. Measurement and calculation packages that provide information of anatomical structures that may be used by a physician for clinical diagnosis purposes.

  2. When used with the ClariTEE probe -- The episodic assessment of cardiac function using transesophageal echocardiography (TEE). It is indicated for use in clinical settings including long term settings such as the ICU for an indwelling time not to exceed 72 hrs.

  3. When used with the TTE probe is intended for imaging and assessment of Cardiac anatomy and function

Device Description

The Zura-EVO Imaging System is a multi-purpose mobile, software controlled diagnostic ultrasound system. Its function is to acquire ultrasound echo data and display it in B-Mode, or Color Doppler Mode on a Flat Panel Display. The user interface includes specialized controls, a minimized computer keyboard, and touch panel on an ergonomic console.

The system has an optional electrocardiography (ECG) display feature and support for a 3-

lead ECG cable assembly. The systems provide measurement capabilities for anatomical structures that provide information used for clinical diagnostic purposes. The system has measurements and calculations, image storage and review, and recording capabilities. The systems include a Digital Imaging and Communications (DICOM) module which enables storage.

The Zura-EVO Imaging System consists of the following major components:

  1. Ultrasound Machine: A predicate device ultrasound machine (K102997) 2. Ultrasound Probes

a. ClariTEE Probe: A miniaturized TEE probe optimized for longer dwell time relative to standard TEE probes that enables long term use in clinical settings such as the ICU. This probe has predicate coverage under K080223 and K100969.

b. TTE probe SA4-2/24: Transthoracic probe has a predicate coverage under K102997

  1. Ultrasound Imaging Software: Ultrasound imaging functions are

controlled via the ImaCor Zura software. The Graphical User Interface, which is the piece of software exposed to the user, communicates user inputs to the Ultrasound Imaging Platform which interfaces with the Ultrasound Transducer. The ImaCor Zura

{8}------------------------------------------------

software also communicates with the probe directly to monitor the probe's lifespan. Orientation and position of the probe is controlled manually by the user.

Description of design modifications

The Zura-EVO is different from the original Zura design in that the EVO GUI is via a touch display screen. A miniaturized keyboard and mouse are provided but are optional. The EVO mains power supply includes a battery so that the machine does not need to be rebooted when moving from bedside to bedside in a critical care setting.

Finally, while the Zura is provided only with the 72 hr. miniaturized ClariTEE probe (ultrasound transducer). The EVO is provided with two transducers; the 72 hr. ClariTEE probe and an optional TTE probe.

The software controls standard ultrasound machine functions such as imaging, recording and measuring. Continuous imaging is limited by a 20 minute software interlock should the operator mistakenly leave the machine in continuous imaging mode, thus limiting the potential unintentional exposure of the patient's mucosal tissue to acoustic energy. Maximum probe face temperature is limited according to FDA consensus standard IEC 60601-2-37.

There are two modes of imaging; type B and color flow Doppler.

Non-Clinical Testing

The following non-clinical testing was conducted: electrical safety to meet the requirements of IEC 60601 3rd Edition, Sterility to demonstrate and SAL 10 106, aged acceleration testing, and acoustic output testing to meet the requirements of IEC 60601-2-37.

Comparison to Predicate Device

The Zura-EVO Ultrasound Scanner is substantially equivalent to the predicate devices listed below with respect to intended use/indications for use, principles of operation and technological characteristics.

The Zura-EVO Ultrasound Scanner includes a digital beam-former that is identical in function to the predicate devices beam-former. It allows transmitting and receiving signals through the

ultrasound transducers. The ultrasound transducers are identical to the ones used on predicate devices.

The backend processing is also similar to the predicate devices and vields an ultrasound image in realtime for diagnosis purposes.

{9}------------------------------------------------

Guidance ( guidancedocument section ref.)DesignCharacteristicFDA ConsensusStandardRecognition Number
Risk AssessmentISO 14971:2007Medical devices -Application of riskmanagement tomedical devices5-40
Acoustic Output(Track 1)EN 60601-2-37:2008Medical electricalequipment -- Part 2-37: Particularrequirements for thebasic safety andessential performanceof ultrasonic medicaldiagnostic andmonitoring equipment
ThermalMechanical andElectrical SafetyEN 60601-1:2006Medical electricalequipment -- Part 1:General requirementsfor safety5-77
EN 60601-2-37:2008Medical electricalequipment -- Part 2-37: Particularrequirements for thebasic safety andessential performanceof ultrasonic medicaldiagnostic andmonitoring equipment5-53
Patient ContactingMaterialsBiological evaluationof medical devicesEN ISO 10993-1 20092-179
EN ISO 10993-5 20092-153
EN ISO 10993-6 20092-120
EN ISO 10993-7 200714-335
EN ISO 10993-10 20102-173
EN ISO 10993-11 20092-176
EN ISO 10993-12 20122-191
CleaningDisinfection,Sterilization andPyrogenicityEN ISO 11135-1:2007Sterilization of healthcare products -Ethylene oxide - Part1: Requirements fordevelopment,validation and routinecontrol of asterilization processfor medical devices(ISO 11135-1:2007)14-228
Software/FirmwareEN 62304Medical DeviceSoftware - SoftwareLife Cycle Processes13-32
LabelingEMCEN 60601-1-2:2007Medical electricalequipment -- Part 1-2:General requirementsfor basic safety andessential performance- Collateral standard:Electromagneticcompatibility -Requirements andtests5-53

Table of FDA Consensus Standards

{10}------------------------------------------------

Substantial Equivalence

The new device (EVO) has significant predicate coverage

{11}------------------------------------------------

Zura -EVOSubject DeviceZura with ClariTEEProbe (K080223,K100989)Ultrasonix Sonix-TabletK102997
Intended UseIntended for the followingapplications: Cardiac and ,Transesophageal.Measurement andcalculation packages thatprovide information ofanatomical structures thatmay be used by a physicianfor clinical diagnosispurposes.2. When used with theClariTEE probe -- Theepisodic assessment ofcardiac function usingtransesophagealechocardiography (TEE). Itis indicated for use in clinicalsettings including long termsettings such as the ICU foran indwelling time not toexceed 72 hrs.3. When used with the TTEprobe is intended forimaging and assessment ofCardiac anatomy andfunctionIntended for use inthe episodicassessment of cardiacfunctionusingtransesophagealechocardiography. Itis indicated for use inclinical settings,includinglong-term settingssuch as the ICU, foran indwelling timeperiod not to exceed72 hours. The ImaCorDiagnostic ultrasoundimaging - TEEIntended for the followingapplications: Abdominal,Cardiac, I ntraolperativeNeurological, Fetal, Pediatric,Small Parts, Neonatal/ AdultCephalic, OB/GYN,Transesophageal,Transrectal,Transvaginal,Peripheral Vascular,Musculoskeletal conventional,Musculoskeletal superficial,Pelvic, Nerve Block, VascularAccess, Transcranial. Thesystem also provides theability to measure anatomicalstructures {fetal, abdominal,pediatric, small organ,cardiac, transrectal,transvaginal, peripheralvessel, musculoskeletal} andprovides calculation packagesthat provide information tothe clinician that may be usedadjunctively with othermedical data obtained by aphysician for clinicaldiagnosis purposes.Power SourceNominal 120 Vac, 50/60 hz, single phaseNominal 120 Vac, 50/60 hz, single phaseNominal 120 Vac, 50/60 hz, single phase
Diagnostic ultrasoundimaging multimodalityincludes TEE, TTM, LinearDiagnostic ultrasoundimaging multimodalityincludes TEE, TTM, LinearBiocompatibilityISO 10993ISO 10993ISO 10993
User PopulationAll except pediatric forTEEAll except pediatric forTEEAllSoftwareModerate ConcernModerate ConcernModerate Concern
TechnologicalCharacteristicsUltrasound diagnostic withminiaturized TEE, TTE, LinearSoftware controls type B modeimaging and color flow dopplerUltrasound diagnosticwithminiaturized TEE probeSoftware controls type Bmode imaging and colorflow dopplerUltrasound diagnostic withminiaturized probe/transducerSoftware controls type B modeimaging and color flow dopplerSterilizationETO; disposable TEE probe provided sterileOther probes - standard disinfectionETO; disposable probe provided sterileReusable, standard disinfectant procedure applied
Major SystemComponentsUltrasound machinecontaining beam-formingarchitectureDisposable single use probe(miniaturized)Effective diameter of 5.5 mm.Phased array transducer.Detachable handleSoftwareQuantitative software aidsTTE and Linear probeUltrasound machinecontaining beam-formingarchitectureDisposable single useprobe(miniaturized)Effective diameter of 5.5mm.Phased array transducer.Detachable handleSoftwareQuantitative softwareaidsUltrasound machinecontaining beam-formingarchitectureDisposable single use probe(miniaturized)Effective diameter of 5.5 mm.Phased array transducer.Detachable handleSoftwareQuantitative software aidsTransducer list includes identicalTTE, Linear probeAcoustic OutputAcoustic Output Track 1 deviceAcoustic Output Track 1 deviceAcoustic Output Track 3 device
AccessoryThree lead ECG (not for diagnostic purposes)Three lead ECG (not for diagnostic purposes)Three lead ECG (not for diagnostic purposes)

{12}------------------------------------------------

  1. The ImaCor EVO is substantially equivalent to the ImaCor Zura and ClariTEE probe cleared for marketing by the FDA under K080223 and K100989. These 510K coverages are for the ImaCor ClariTEE probe

  2. The EVO is substantially equivalent to the Sonix-Tablet, manufactured by Ultrasonix Medical Corporation and cleared for marketing by the FDA under K102997. Ultrasonix is the OEM for the EVO and the Sonix-Tablet is the EVO's twin. The Linear and TTE transducers provided with the EVO are covered with the Sonix-Tablet under K102997

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.